MW82119B-703 QRC

Loop Cable Connected

Copyright 2020 Anritsu Company. All rights reserved.

1-Port PIM Master Calibration Quick Reference CardTM

1. Connect Port 2 Out to Port 2 Returnwith supplied cable.

4. Press Enter to calibrate,or Escape to exit calibration.

3. Connect a low PIM terminationdevice to the PIM Standard device.

2. Connect a PIM Standard to thePIM Test Port.

5. Remove the PIM Standard fromthe PIM Test Port.

6. Connect a low PIM terminationdevice to the PIM Test Port.

7. Press Enter to calibrate,or Escape to exit calibration.

8. Remove termination devicefrom instrument.

10. Press Enter to calibrate, or Escape to exit calibration.

9. Leave PIM Test Port open.

7/16”7/16”

PIM Standard

Low PIM Load

Loop Cable Connected

Low PIM Load

Loop Cable Connected

Loop Cable Connected

TM

1. Connect Port 2 Out to Port 2 Returnwith supplied cable.

4. Press Enter to calibrate,or Escape to exit calibration.

3. Connect a low PIM terminationdevice to the PIM Standard device.

2. Connect a PIM Standard to thePIM Test Port.

5. Remove the PIM Standard fromthe PIM Test Port.

6. Connect a low PIM terminationdevice to the PIM Test Port.

7. Press Enter to calibrate,or Escape to exit calibration.

8. Remove termination devicefrom instrument.

10. Press Enter to calibrate,or Escape to exit calibration.

9. Leave PIM Test Port open.

7/16”7/16”

PIM Standard

Low PIM Load

Loop Cable Connected

Low PIM Load

Loop Cable Connected

10580-00472 Rev. A

MW82119B-703 QRC Copyright 2020 Anritsu Company. All rights reserved.10580-00472 Rev. A

1-Port PIM Master Calibration Quick Reference Card

9. Remove termination devices from instrument.

PIM Standard

Low PIM Load

Low PIM Loads

Loop Cable Connected

Loop Cable Removed

4. Press Enter to calibrate, or Escape to exit calibration.

3. Connect a low PIM terminationdevice to the PIM Standard device.

2. Connect a PIM Standard to thePIM Test Port.

5. Remove the PIM Standard from the termination device and the PIM Test Port.

7. Connect a low PIM terminationdevice to the PIM Test Port and tothe Port 2 Out port.

8. Press Enter to calibrate, or Escape to exit calibration.

11. Press Enter to calibrate, or Escape to exit calibration.

1. Connect Port 2 Out to Port 2 returnwith supplied cable

10. Leave PIM Test Port open.6. Remove the loop cable.

4.3-107/16”7/16”

Loop Cable Removed

9. Remove termination devices from instrument.

PIM Standard

Low PIM Load

Low PIM Loads

Loop Cable Connected

Loop Cable Removed

4. Press Enter to calibrate, or Escape to exit calibration.

3. Connect a low PIM terminationdevice to the PIM Standard device.

2. Connect a PIM Standard to thePIM Test Port.

5. Remove the PIM Standard from the termination device and the PIM Test Port.

7. Connect a low PIM terminationdevice to the PIM Test Port and tothe Port 2 Out port.

8. Press Enter to calibrate, or Escape to exit calibration.

11. Press Enter to calibrate, or Escape to exit calibration.

1. Connect Port 2 Out to Port 2 returnwith supplied cable

10. Leave PIM Test Port open.6. Remove the loop cable.

4.3-107/16”7/16”

Loop Cable Removed

TM

TM2-Port PIM Master Calibration Quick Reference Card

2-Port PIM Master Calibration Quick Reference Card