Post on 14-Feb-2019
transcript
1 | VARIAN ONCOLOGY SYSTEMS
Acuros® BVadvanced dose calculation for brachytherapy
2 | VARIAN ONCOLOGY SYSTEMS
Simple Cylinder: Acuros BV vs TG 43
3 | VARIAN ONCOLOGY SYSTEMS
Shielded Rectal: Acuros BV vs. TG43
4 | VARIAN ONCOLOGY SYSTEMS
Free-Hand Implant: Acuros BV vs TG43
Skin 1 ∆1 =~13%Target ∆2 =~10%Skin 2 ∆3 =~22%
5 | VARIAN ONCOLOGY SYSTEMS
Contura: Acuros BV vs TG43
Balloon ∆1 =~11%Skin ∆2= ~8%Surface ∆3 =~13%Surface ∆4 =~3%
6 | VARIAN ONCOLOGY SYSTEMS
SAVI: Acuros BV vs TG43
Bone ∆1 =~6%Cavity ∆2 =~2%Rx distance ∆3 =~0.3%Skin ∆4 =~10%
7 | VARIAN ONCOLOGY SYSTEMS
MammoSite: Acuros BV vs TG43
Skin ∆1 =~14%
Bone ∆2 =~6%
Tissue ∆3 =~13%
8 | VARIAN ONCOLOGY SYSTEMS
MammoSite TG43 vs. Acuros DVH
SkinLung
PTV