Blood Pressure MachinePrototype Review Presentation

By Audrow Nash, Travis McNeill, Ilya Sandler

cfmedical.com

www.ascottechnologies.com

1. Design specifications2. Progress3. Continua Compliance4. Android Application5. Related future work6. Documentation7. Plan

Table of Contents

www.mountainside-medical.com

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Design Specification

Progress: Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

USB Hub: 5V Raspberry Pi: max 515 mA Bluetooth Dongle: 100mA

Touch Screen: 12v with 660mA Blood Pressure module: 12v at 645mA

Progress: Power Supply Analysis

Progress: Power Supply Analysis

Progress : Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Project Specification: System must be contained within an enclosure Dimensions of the Touch Screen display

dictate the overall size of the enclosure 3D printed

Progress: Enclosure Analysis

Began Analysis from cardboard model Constructed 3-D model in Solid Works Equipment Layout Analysis using 3-D model Printed Enclosure

Special thanks to Dr. Patterson for his many contributions to the project

Progress: Enclosure

Progress: Enclosure Demo

Progress: Enclosure Demo

Progress: System Assembly

Progress: Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Progress: BPP Interface

http://www.raspberrypi.org/ Serial Communication through Linux Terminal

Progress: Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Progress: GUI demo

Progress: Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Continua Compliance:Explanation

Continua Compliance:Semester Plan

Continua Compliance:Semester Plan

Planned for difficulty 2 weeks Fall 2013 8 weeks Spring 2014

Pushing into time allocated for debug 3 additional three weeks

Total ~13 weeks for Continua Compliance

Continua Compliance:Methods Attempted

RFCOMM : Redesign Bluetooth Antidote: Continua compliant service Oximeter HDP tutorial HDPy repository

AND Blood pressure pump to Linux computer Contact HDPy developers

Linux to Linux spoof oximeter data example Linux to Linux spoof blood pressure data

Continua Compliance:Demo

http://techon.nikkeibp.co.jp/article/NEWS/20090126/164588/CHA.jpg

Progress: Design Specification

1. External Power Supply2. Enclosure for device3. Interface with blood pressure pump (BPP)4. System uses a touch screen interface for user

control5. Continua compliant6. Android Application with user interface

Android Application

Android HDP Application

Android application that supports the Health Device Protocol standard with Bluetooth capability.

Likely has error in the stack

Wrap Continua Communication in C programming wrapper

Establish auto-pairing between manager and agent

Create Continua Compliant Android sink

Related Future Work

Source code Drawings and documentation User’s manual for prototype blood pressure

machine Documentation for FDA

Documentation

Submission of a PMN (510k) Must show substantial equivalence to a

chosen predicate device. A&D UA-767PBT A&D UA-767PBT-C

Equivalence is shown through: Indication Statements Intended Use Technological Characteristics Descriptive Characteristics

FDA Documentation

Plan: Semester Milestones

Task Name Duration Start Finish

Case Design 35 daysMon

1/13/14 Fri 2/28/14

Raspberry Pi Interface 35 daysMon

1/13/14 Fri 2/28/14

Android Interface 20 daysMon

1/20/14 Fri 2/14/14

Prepare for PSR Presentation 5 days

Mon 2/3/14 Fri 2/7/14

Prepare for PRP Presentation 4 days

Mon 3/17/14

Thu 3/20/14

HDP Communication 18 daysMon

3/10/14 Fri 4/16/14

Testing & Documentation 35 days

Mon 3/3/14 Fri 4/18/14

Design Expo Preparation 9 days

Mon 4/21/14

Thu 5/1/14

1. Accurate data transmission2. Function properly with extended runtime3. User setup and run without instructions4. Behavior with various data interruption5. Interface to Continua Devices

Plan: Testing

Test name: Accurate data transmission

Independent variables: Blood pressure pump

Dependent variables: Data readings (systolic pressure, diastolic pressure, heart

rate)

Controls:Environment of testing, person who’s blood pressure is beingmeasured

Plan: Testing

Test name: Function properly with extended runtime

Independent variables:Time elapsed

Dependent variables:Device functionality: Able to take and view blood

pressure reading

Controls:Testing environment

Plan: Testing

Test name: User setup and run without instructions

Independent variables:User selected

Dependent variables:Time for user to view their blood pressure

Controls:Withhold instructions, method of seeking feedback,testing environment

Plan: Testing

Test name: Behavior with various data interruption

Independent variables:Method of interruption

Dependent variables:After connect is reestablished, if data passed continues

tobe passed without an error

Controls:Test environment, trials per independent variable

Plan: Testing

Test name: Interface to Continua Devices

Independent variables:Device used as agent

Dependent variables:If correct data is passed from agent to manager

Controls:Test environment, managing device

Plan: Testing

Summary

1. Design specifications2. Progress3. Continua Compliance4. Android Application5. Related future work6. Documentation7. Plan

Our sponsors: Ascot Technologies Suntech Medical

Special thanks to our mentors: Dr. Hargrove Grayson Randall

Thank you!