Computer modeling of ocular injury in infants exposed to acceleration. N. Rangarajan, S. Kamalakkannnan, T. Shams. GESAC Inc, Boonsboro, MD. Alex Levin, MD, MHSc, FAAP, FAAO, FRCSC. Sickkids Hospital, Toronto. Carole Jenny, MD, MBA. Brown University, Providence, RI. Objective. - PowerPoint PPT Presentation

transcript

Computer modeling of ocular injury in infants exposed to acceleration

N. Rangarajan, S. Kamalakkannnan, T. Shams. GESAC Inc, Boonsboro, MD.

Alex Levin, MD, MHSc, FAAP, FAAO, FRCSC. Sickkids Hospital, Toronto.Carole Jenny, MD, MBA.

Brown University, Providence, RI.

04/22/23 GESAC, Inc 2

Objective

• To develop a finite element model of the infant eye.

• To evaluate the level of stresses and strains when the head is exposed to an acceleration pulse.

• Model to be exercised with acceleration pulses obtained from Aprica 2.5 kg infant dummy shaking tests.

04/22/23 GESAC, Inc 3

Presentation Road Map

• Description of the model• Parametric study• Discussion of results• Conclusions• Limitations• Work in progress

04/22/23 GESAC, Inc 4

The Model

• LS Dyna model consists of orbit, fatty tissue, extra-ocular muscles, sclera, retina, and vitreous.

04/22/23 GESAC, Inc 5

Model Description

• Key dimensions– Sclera, diameter = 20 mm – Rectus muscle length = 50 mm

• Contact definitions– Surfaces for orbit/fat/sclera/retina tied

together– Nodes on retina tied to vitreous surface

04/22/23 GESAC, Inc 6

Data for Prescribed Motion

Shaking experiment with Aprica 2.5 dummy. Click on picture to see baby being shaken. Large AVI file

04/22/23 GESAC, Inc 7

Prescribed motion

• Applied motion - rotational oscillations– Represents average data from shaking tests with

Aprica 2.5kg dummy

04/22/23 GESAC, Inc 8

Eye ball with muscles

04/22/23 GESAC, Inc 9

The Model – Vitreous and Retina Attachment

04/22/23 GESAC, Inc 10

Parametric Study

• Parametric study was conducted to evaluate the effect of variation in material models and properties of vitreous and fat on the maximum stress and stress distribution.

• Six cases were simulated.

04/22/23 GESAC, Inc 11

Simulation Matrix

Case 1 2 3 4 5 6

Vitreous

BK, G0 & Gi in MPa

V.E.BK=0.7PR=0.49G0=0.014

Gi=0.012

V.E.BK=0.7PR=0.49G0=0.014

Gi=0.012

V.E.BK=7.0PR=0.49G0=0.14

Gi=0.12

FluidVC=0.5PR=0.49

FluidVC=0.3PR=0.49

FluidVC=0.1PR=0.49

Fat

E, BK, G0 & Gi in MPa

ElasticE=0.047PR=0.49

V.E. (Viscoelastic)BK=0.7

PR=0.49G0=0.014

Gi=0.012

04/22/23 GESAC, Inc 12

Results

• Results of the simulation study are discussed – Stresses on retina – Stresses on the sclera

04/22/23 GESAC, Inc 13

Stress on Retina – Case 2

Stress on one of the elements in case 2

04/22/23 GESAC, Inc 14

Stress on Retina – Case 3

Stress on one of the elements in case 3

04/22/23 GESAC, Inc 15

Stress on Retina – Case 5

Stress on one of the elements in case 5

04/22/23 GESAC, Inc 16

Simulation – Case 2

Click on picture to see animation of simulation results. Large AVI file.

04/22/23 GESAC, Inc 17

Simulation – Case 5

• Click on picture to see animation of simulation results. Large AVI file.

04/22/23 GESAC, Inc 18

Stress Distribution on Sclera– Case 2

04/22/23 GESAC, Inc 19

Stress Distribution on Retina – Case 2

04/22/23 GESAC, Inc 20

Stress Distribution on Vitreous – Case 2

04/22/23 GESAC, Inc 21

Stress Distribution on Sclera – Case 5

04/22/23 GESAC, Inc 22

Stress Distribution on Retina – Case 5

04/22/23 GESAC, Inc 23

Stress Distribution on Vitreous – Case 5

04/22/23 GESAC, Inc 24

Comparison of Maximum Stress on Retina

* The calculation reaches an infinite loop at 0.94 sec, when case 5 has a maximum of 0.0964

Case 1 2 3 4 5 6

Maximum Stress (MPa)

0.01448 0.01476 0.00719 0.13899 0.13918 0.08775*

FatVitreous

ElasticV.E.

V.E.V.E.

V.E.V.E.

V.E.Fluid

V.E.Fluid

V.E.Fluid

04/22/23 GESAC, Inc 25

Conclusions - 1

• Stress on retina and sclera accumulates (increases) as the shaking continues for certain material models for vitreous, e.g. viscoelastic or fluid.

• Maximum stress occurs around the vitreous-retina contact area both in front and back.

• Property of vitreous has great effect on maximum stress and stress distribution.

04/22/23 GESAC, Inc 26

Conclusions - 2

• When using viscoelastic material for vitreous, smaller bulk modulus value shows a clearer stress accumulation effect. Rate of change of stress is more evident.

• When using fluid for vitreous, the viscosity coefficient does not show significant effect on maximum stress, stress accumulation effect, and stress distribution.

04/22/23 GESAC, Inc 27

Limitations - 1

• Current model does not include several important structures,such as lens, choroid, ciliary body, and cornea.

• Definition of distribution of fat and orbit geometry was approximated.

04/22/23 GESAC, Inc 28

Limitations - 2

• Input motion was purely rotational at centre of the orbit.

• All model input data were obtained from literature, material data not verified by experimentation.

• Material property data were scaled and appropriateness of scaling has to evaluated.

04/22/23 GESAC, Inc 29

Limitations - 3

• Effect of variation of mesh size, integration intervals and integration procedures have not been fully evaluated.

04/22/23 GESAC, Inc 30

Limitations - 4

• All the material models used in this study are currently available in LS-Dyna material library. It may be necessary to develop new material models to fully describe the material used in this model.

• This is a preliminary study to provide a qualitative picture of what happens within the infant eye under repeated motion and results should be interpreted with caution.

04/22/23 GESAC, Inc 31

Work in Progress - 1

• Study influence of additional features such as lens, choroid and ciliary body on model response.

• Develop a more accurate orbit geometry and fatty tissue distribution.

• Change center of motion to head CG. Both linear and angular motion will be used as input.

04/22/23 GESAC, Inc 32

Work in Progress - 2

• Evaluate effect of mesh size change.• Evaluate effect of meshing axis.• Examine response under purely linear

deceleration like a typical frontal crash pulse. This will be an indirect method of validation of the model.

04/22/23 GESAC, Inc 33

Component models under development - 1

Sclera and cornea

Choroid

RetinaPars plana and pars plicata

04/22/23 GESAC, Inc 34

Component models under development - 2

Fat

Lens

Orbit

Vitreous

04/22/23 GESAC, Inc 35

Component models under development - 3

Extra-ocular muscles Eye assembly

04/22/23 GESAC, Inc 36

Acknowledgements

• This work was supported by Aprica, inc. Japan. We thank Aprica project managers, Ms. P. Kawasaki and Dr. R.Bigge, MD, PhD.

• Dr. Levin, MD provided invaluable guidance and a push when needed!

04/22/23 GESAC, Inc 37