DOC053.52.00139.Jan06Data SheetPHOSPHAX sc
Technical data
We reserve the right to make changes without notice.
FusesConnect a maximum of 2 analysis instruments to one sc1000.
Always connect a residual current circuit breaker between mains and system in case of external supply of power.
If you operate controller and / or analysis instrument outdoors, provide overvoltage protection between mains and system in case of external power supply.
Ambient conditions, limits for operation and storageFor information on the ambient conditions, limits for operation and storage, temperatures, atmospheric humidity, dust, etc. please see the technical data, the stickers on the related modules or the rating plates on the related modules.
Table 1 PHOSPHAX sc specifications (in some cases optional extra)
Enclosure material ASA/PC UV-resistant
Measuring method 2-beam photometer (yellow method)
Measuring range0.05...15 mg/l PO4–P 1...50 mg/l PO4–P
With different verification limits and resolutions
Verification limit 0.05 mg/l, with standard solution 1.00 mg/l, with standard solution
Measuring accuracy(with standard solution) 2 % of the measured value + 0.05 mg/l 2 % of the measured value + 1.0 mg/l
Reproducibility(with standard solution)
2 % of the measured value + 0.05 mg/l, 2 % of the measured value + 1.0 mg/l,
Response time (90 %) < 5 minutes
Adjustable measuring interval 5...120 minutes
Power supply Power supply with power cable only using the sc1000 controller(analyser, filtration probe sc and drain tubing: 115 V versions or 230 V versions)
Data transmission Data transmission with data cable on the sc1000 controller
Electrical power, power consumption: 400 VA (mean), max. 1000 VA (with 10 m heated filtration probe tubing)
Electrical fuse protection Via sc1000 controller
Outputs Relay, current outputs, bus interface via sc1000 controller
Operating temperature –20 °C...40 °C (–4 °F...104 °F); 95 % relative humidity, non-condensing
Storage temperature –20 °C...50 °C (–4 °F...122 °F; 95 % relative humidity, non-condensing
Sample temperature +4 °C...+ 45 °C (39 °F...113 °F)
Permissible pH value of the sample 5...9
Dimensions (W × H × D) 540 × 720 × 390 mm (21.25 × 28.35 × 15.35 inches)
Cable lengths for data and power cable 2 m (80 inches) (from edge of enclosure)
Mass PHOSPHAX sc Approx. 31 kg, without filtration probe sc and without chemicals
2
Emissions
CAUTIONDepending on the frequency of cleaning and calibration, a varying quantity of lyes and acids will be produced. The lyes and acids are toxic, caustic and/or an irritant. Ensure only trained personnel handle these residues and that the applicable accident prevention regulations are observed.
Wear personal safety clothing when handling chemicals and sample flows of unknown composition!
Dispose of these substances left over from cleaning and calibration correctly and in accordance with the applicable regulations.
3
Drawing with dimensions
Figure 1 Drawing with dimensions
523 mm[20.47 in.]
400 mm[15.75 in.]
466 mm[18.11 in.]
71
0m
m[2
7.5
in.]
280 mm[11.02 in.]
40 mm[1.57 in.]
282 mm [11.1 in.]
1040 mm[40.94 in.]
85
0m
m[3
3.1
5in
.]
242 mm [9.53 in.]
4
Replacement parts, consumables, accessories
PHOSPHAX sc chemistryDescription 1 Description 2 Order No.
Reagent PHOSPHAX sc (2 L) for all measuring ranges LCW869
Cleaning solution PHOSPHAX sc (1 L) for all measuring ranges LCW870
Replacement partsItem Description 1 Description 2 Order No.
1 Door for sc-analyser enclosure incl. 4 instrument labels LZY143
1 Set of instrument labels (4 pieces) LZY144
2 Door catch LZY148
3 Enclosure for sc-analyser without door LZY145
4 Set of filter pads (2 pieces) LZY154
5 Clamping lock for sc-analyser LZY147
6 Fence incl. screws M3 x 6 LZY157
7 Compressor switchable 115 V/230 V LZY149
8 Air tubing for compressor incl. non-return valve, fitting, LZY151
9 Fan for incoming air LZY152
10 Collecting tray for sc-analyser LZY146
11 Hinge incl. screws LZY155
12 Gasket for door of sc-analyser LZY187
13 Door lock sc-analyser LZY188
14 Mounting plate for sc-analyser LZY161
15 Analytical section of PHOSPHAXsc MR: 0.05-15 mg/l LZY163
16 Analytical section of PHOSPHAXsc MR: 1-50 mg/l LZY164
17 Measuring cell PHOSPHAX sc MR: 0.05 -15 mg/l LZY185
18 Measuring cell PHOSPHAX sc MR: 1 -50 mg/l LZY186
19 Valve block PHOSPHAX sc incl. valves; MR: 0.05 -15 mg/l LZY183
20 Valve block PHOSPHAX sc incl. valves; MR: 1 -50 mg/l LZY160
21 Valve block PHOSPHAX sc MR: 0.05 -15 mg/l LZY271
22 Valve block PHOSPHAX sc MR: 1 -50 mg/l LZY272
23 Upper part of valve block LZY174
24 Upper part of valve block with valve LZY175
25 Valve 2/2 ways LZY168
26 Tubing 3.2 mm (2 m) sc-analyser LZY195
27 Set of fittings 3.2 mm (4 pieces) LZY111
28 Fitting for tubing DN4/6 LZY134
29 Blind plug LZY193
30 Valve block 2-channel switch incl. valve for sc-analyser LZY267
31 Valve block 2-channel switch for sc-analyser LZY172
32 Valve 3/2 ways LZY171
33 Conversion kit from 1-channel > filtration probe sc-analyser AMTAX sc/PHOSPHAX sc LZY242
34 Bottom part of overflow vessel LZY165
5
35 Upper part of overflow vessel LZY166
36 Upper part of overflow vessel incl. valve LZY167
37 Locking screw LZY150
38 Conversion kit from filtration probe > 1-channel sc-analyser AMTAX sc/PHOSPHAX sc LZY241
39 Upper part of overflow vessel for 1 or 2-channel instrument LZY268
40 Overflow vessel for 2-channel instrument LZY269
41 Safety plate for sc-analyser LZY179
42 Insulation cover for photometer PHOSPHAX sc MR 0.05-15 mg/l LZY225
43 Insulation cover for photometer PHOSPHAX sc MR 1-50 mg/l LZY226
44 Holder for piston pump LZY180
45 Pump head for air pump 10 mL LZY181
46 Piston pump for sc-analyser LZY177
47 Cover for reagent pump LZY178
48 Set of fittings 1.6 mm (4 pieces) LZY192
49 Tubing 1.6 mm (2 m) sc-analyser LZY194
50 Reagent pump for sc-analyser (valve pump) LZY176
51 Fan for air recirculation for sc-analyser LZY153
52 Heating for analyser enclosure incl. connectors LZY156
53 Cover LZY270
54 Cover for processor card LZY159
55 Cover for power supply LZY158
56 Power supply 100-240 V AC XMU434 YAB039
57 Amplifier card for PHOSPHAX sc YAB045
58 Processor card for sc-analyser OAB028, ZBA803 YAB043
59 Set of seals for PHOSPHAX sc measuring cell (4 O-rings) LZY197
60 Sealing for valve block LZY199
61 Sealing for overflow vessel LZY198
62 Card with temperature sensor AMTAX sc/PHOSPHAX sc YAB089
17+19+42 Conversion kit PHOSPHAX sc from HR to LR LZY313
18+20+43 Conversion kit PHOSPHAX sc from LR to HR LZY312
30+40 Conversion kit from 1-channel into 2-channel sc-analyser LZY170
AccessoriesDescription 1 Description 2 Order No.
Set of plugs (rubber) for sc-analyser (3*type 1, 1*type 2, 3*type 3) LZY007
Accessories for AMTAX/PHOSPHAXsc for continuous sampling (1/2-channel) LZY189
Cutter for tubing LZY201
Special tool for electrode AMTAX sc LZY200
Heated drain/connecting hose 230 V LZY302
Heated drain/connecting hose 115 V LZY303
Set of connectors for sc-analysers LZY190
Set of screws for sc-analyser 4 x M3x6, 2 x M3x25; 2 x M3x50 LZY191
Replacement parts (continued)Item Description 1 Description 2 Order No.
6
FasteningDescription 1 Description 2 Order No.
Wall mounting kit 4 screws 5*60 incl. 4 anchors LZX355
Mounting kit for sc-analyser incl. fastening and angle bracket, screws LZY044
Set of screws for fastening and angle bracket LZY216
Set of screws for sc-analyser LZY223
Set of screws for rail mounting LZY285 and LZY316 LZY220
Rail mounting for analyser with controller LZY285
Rail mounting for analyser without controller LZY316
Stand mounting for sc-analyser with controller LZY286
Stand mounting for sc-analyser without controller LZY287
11
Contact
HACH LANGE GmbH
Willstätterstraße 11D-40549 DüsseldorfTel. +49 (0) 211- 52 88 - 0Fax +49 (0) 211- 52 88 - 143info@hach-lange.dewww.hach-lange.de
HACH LANGE LTDPacific WaySalfordManchester, M50 1DLTel. +44 (0)161 8 72 14 87Fax +44 (0)161 8 48 73 24info@hach-lange.co.ukwww.hach-lange.co.uk
HACH LANGEHACH SAS33, Rue du BallonF-93165 Noisy Le GrandTél. +33 (0)1 48 15 68 70Fax +33 (0)1 48 15 80 00info@hach-lange.frwww.hach-lange.fr
DR. BRUNO LANGEGES. MBHIndustriestraße 12A-3200 ObergrafendorfTel. +43 (0) 2747 - 74 12Fax +43 (0) 2747 - 42 18info@hach-lange.atwww.hach-lange.de
DR. BRUNO LANGE AG
Juchstrasse 1CH-8604 HegnauTel. +41 (0)44- 9 45 66 10Fax +41 (0)44- 9 45 66 76info@hach-lange.chwww.hach-lange.ch
HACH LANGE SA
Motstraat 54B-2800 MechelenTél. +32 (0)15 42 35 00Fax +32 (0)15 41 61 20info@hach-lange.bewww.hach-lange.be
DR. LANGE NEDERLAND B.V.Laan van Westroijen 2aNL-4003 AZ TielTel. +31(0)3 44 63 11 30Fax +31(0)3 44 63 11 50info@hach-lange.nlwww.hach-lange.nl
HACH LANGE ABVinthundsvägen159ASE-128 62 SKÖNDALTel. +46 (0)8 7 98 05 00Fax +46 (0)8 7 98 05 30info@hach-lange.sewww.hach-lange.se
HACH LANGE A/SÅkandevej 21DK-2700 BrønshøjTel. +45 36 77 29 11Fax +45 36 77 49 11info@hach-lange.dkwww.hach-lange.dk
HACH LANGE S.L.U.C/Araba 45, Apdo. 220E-20800 Zarautz/GuipúzcoaTel. +34 9 43 89 43 79Fax +34 9 43 13 02 41info@hach-lange.eswww.hach-lange.es
HACH LANGE SP.ZO.O.ul. Opolska 143 aPL-52-013 WroclawTel. +48 71 3 42 10-81Fax +48 71 3 42 10-79info@hach-lange.plwww.hach-lange.pl
HACH LANGE S.R.L.Via Riccione, 14I-20156 MilanoTel. +39 02 39 23 14-1Fax +39 02 39 23 14-39info@hach-lange.itwww.hach-lange.it
HACH LANGE S.R.O.Lešanská 2a/1176CZ-141 00 Praha 4Tel. +420 272 12 45 45Fax +420 272 12 45 46info@hach-lange.czwww.hach-lange.cz
HACH LANGE S.R.O.Sabinovská 10SK-821 02 BratislavaTel. +421 2 4820 9091Fax +421 2 4820 9093info@hach-lange.skwww.hach-lange.sk
HACH LANGE LDARua dos Malhões,Edif. D. Pedro IP-2770-071 Paço D'ArcosTel. +351 210 00 1750Fax +351 210 00 8140info@hach-lange.ptwww.hach-lange.pt
HACH LANGE S.R.L.str. Aviator Teodor Iliescu nr.37Bucuresti, Sector 1RO-011672Tel. +40 (0)21 2 08 95 78Fax +40 (0)21 2 08 95 78info@hach-lange.comwww.hach-lange.com
HACH LANGE KFT.
Íves u. 2.H-9027 GyőrTel. +36 (06)96 511 400Fax +36 (06)96 329 981info@hach-lange.comwww.hach-lange.hu
HACH LANGE
8, Kr. Sarafov str.BG-1164 SofiaTel. +359 (0)2 963 19 21Fax +359 (0)2 866 04 47info@hach-lange.comwww.hach-lange.com