EEG - Montages, Equipment and Basic Physics

Post on 01-Dec-2014

1,561 views 15 download

description

This presentation discusses the 10-20 system of electrode placement, with its modifications. Also discussed are the Equipment Specifications, basic Physics and sources of interference

transcript

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

Recording technique !!

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

EEG measurement setup

• 10-20 Lead system is most widely clinically accepted

• Certain physiological featuresare used as reference points

• Brain research utilizes even 256 or 512 channel EEG hats

CORTICAL CONFIGURATION

A TYPICAL RESEARCH CAP CONFIGURATION

HAS 64-256 ELECTRODES

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

Electrodes – Basics• High-quality biopotential measurements require– Good amplifier design– Use of good electrodes and their proper placement on the

patient– Good laboratory and clinical practices

• Electrodes should be chosen according to the application• Basic electrode structure includes:– The body and casing– Electrode made of high-conductivity material– Wire connector– Cavity or similar for electrolytic gel– Adhesive rim

Ag-AgCl, Silver-Silver Chloride Electrodes

• The most commonly used electrode type

• Silver is interfaced with its salt silver-chloride

• Choice of materials helps to reduce junction potentials

• Electrolytic gel enhances conductivity and also reduces junction potentials

• The gel is typically soaked into a foam pad or applied directly in a pocket produced by electrode housing

• Relatively low-cost and general purpose electrode

• Particularly suited for ambulatory or long term use

A silver/silver chloride electrode, shown in cross section.

Gold Electrodes• Very high conductivity suitable for low-noise meas.• Inertness suitable for reusable electrodes• Body forms cavity which is filled with electrolytic gel• Compared to Ag-AgCL: greater expense, higher

junction potentials and motion artifacts• Often used in EEG, sometimes in EMG

Conductive polymer electrodes

• Made out of material that is simultaneously conductive and adhesive• Polymer is made conductive by adding monovalent metallic ions• Aluminum foil allows contact to external instrumentation• No need for gel or other adhesive substance• High resistivity makes unsuitable for low-noise meas.• Not as good connection as with traditional electrodes

Metal or carbon electrodes• Other metals are seldom used as high-quality noble

metal electrodes or low-cost carbon or polymericelectrodes are so readily available

• Historical value. Bulky and awkward to use• Carbon electrodes have high resistivity and are noisier

but they are also flexibleand reusable• Applications in electrical stimulation and impedance

plethysmography

Needle electrodes

• Obviously invasive electrodes• Used when measurements have to be taken from the organ itself• Small signals such as motor unit potentials can be measured• Needle is often a steel wire with hooked tip

Electrode-electrolyte interface The current crosses it from left to right. The electrode consists of metallic atoms C. The electrolyte is an aqueous solution containing cations of the

electrode metal C+ and anions A-.

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

In clinical EEG recording, the waveforms are recorded in a series of Montages.

What is a montage?

Each EEG trace is generated from an active and a reference electrode. Different patterns of electrodes are selected and the traces grouped to provide data from different areas of the scalp.

Bipolar or unipolar electrodes can be used in the EEG measurement.

In the unipolar method the potential difference between a pair of electrodes is measured.

In the bipolar method the potential of each electrode is compared either to a neutral electrode or to the average of all electrodes

Unipolar and Bipolar EEG measurement

Road Map for the Session

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing

Why do we need Filtering

Types of Filter

1. Low-pass – attenuate high frequencies

2. High-pass – attenuate low frequencies

3. Band-pass – attenuate both

4. Notch – attenuate a narrow band

Properties of Filters

• “Transfer function”

1. Effect on amplitude at each frequency

2. Effect on phase at each frequency

• “Half Amp. Cutoff”

1. Frequency at which amp is reduced by 50%

High-pass

Low-pass

Band-pass and Notch

Problems with Filters

• Original waveform, band pass of .01 – 80Hz

• Low-pass filtered, half-amp cutofff = ~40Hz

• Low-pass filtered, half-amp cutofff = ~20Hz

• Low-pass filtered, half-amp cutofff = ~10Hz

Filtering Artefacts

• “Precision in the time domain is inversely related to precision in the frequency domain.”

Time constant of Low pass Filter

o Want the majority of voltage of signal to be measured across the Co It takes time to charge the C ()o It takes longer to charge a capacitor with a big capacitance

(can hold a lot of chargeo It takes longer to charge if R is big b/c it slows the currento We only have until the peak of the half cycleo Want the time to peak of half cycle (or longer) to charge the capacitor

Summary

• Introduction

• 10 – 20 System

• Electrodes

• Montages

• Post Processing