ENGINEERING PANEL SOLUTIONS MEDICAL GAS EQUIPMENT Medical.pdf · 2019. 10. 1. · ISO 7396-1 52 rue...

Post on 07-Sep-2021

3 views 0 download

transcript

0499

ISO 7396-1

52 rue de Lagny93100 MONTREUIL - FRANCEE-mail : commercial@cahouet.com

Tel. : +33 (0)1 41 72 90 00Fax : +33 (0)1 41 72 90 01Site : www.cahouet.com

ISO 9001 / ISO 13485

05/2

015

ENGINEERING PANEL SOLUTIONS

ALARM

CE marked Custom made panels Ready to install

Example : 2 liquids + 1 gas 1 liquid + 2 gases 3 gases...

Up to 9 remotes 8 logic or analogic inputs CE Medical marked Alarm units are connected together and to the computer by RS 485 BUS Gas : O2 - Air - Air800 - CO2 N2O - Vacuum

Pressure transducer

MEDICALGAS EQUIPMENT

TypeInlet

pressure (bar)

Outlet pressure

(bar)

Flow (Nm3/h) Specifications

T2m 200 1-11 44 Automatic change-over unit

CENTREOLE 200 1-14 150 High flow automatic change-over unit

SS2m 200 1-9 44Automatic outlet device for the reserve supply

SSEOLE 200 1-9 150High flow outlet device for the reserve supply

HP FLEX 200 200 -Liner SS + SS braids + connection to request

PIGTAILS HP 200 200 - Copper or SS

HEATER 200 200 44 Electric heaters 500W or 1000W

HP STAND 200 200 -Connection from bundle to pipeline

EUM 50 / 200 12 12 44 / 150Emergency and maintenance inlet (VSP)

BALL VALVE 20 20 -Low pressure ball valve. From 3/8” to 2” diameter

D300MV 30 1-12 100 Dual line regulating assembly

L2D5 20 1-12 200 Dual line regulating assembly

BP RANGE 50 0.5-12100 - 300500 - 800

Low pressure line regulator

CSG 50/200 9 9 44 / 150Gas emergency and maintenance supply cabinet

D300ML 12 4-10 40 Dual second stage regulating assembly

ROTAREG 20 4-12 40 Second stage regulator assembly

MEDICAL GAS EQUIPMENT

Ball valve

CSG 50/200

D300ML

RotaregPressure

transducer

Alarm Theo Rem

HPflexible hose

Pigtails

HP stand

BP 500

EUM 50-200

Heater SS2m

Centreole

T2m

SSEOLE

L2D5

LPflexiblehose

Ceiling base

Ceiling medical outlet

AGSS

AGSS Venturi

Medical outlet

Flowmeter

Novaflow

Dynaflow

LPflexible hose

GAS SOURCES

CORRIDORS

OPERATING THEATER

HOSPITAL BEDROOM

Remote

D300MV

TypeInlet

pressure (bar)

Outlet pressure

(bar)Flow Specifications

SAM 200 0-10 13.5 Nm3/h Outlet for hose or terminal unit

NOVA 40 200 0-4 50 Nm3/h Preset. Outlet G1/4 or terminal unit

NOVABLOC 200 -0,1 to 15

lpmFlowmeter regulator

DYNABLOC 200 - 0,5 to 50 lpm Paddle action flowmeter regulator

NOVAFLOW 2,7 - 6 -0,1 to 15

lpmCalibrated orifices flowmeter

DYNAFLOW 4 -0,5 to 50

lpmPaddle action flowmeter

FLOWMETER 4,5 -0 to 15

lpmFlowmeter

LP FLEX 10 - - LP flexible hose

THEO MEDICAL OUTLET

10 10 -AFNOR outlet(wall, bedhead, ceiling pendant) (Maximum operating pressure 10 bar)

AGSS Vacuum - - Air powered or Vacuum

Compact integrated regulating valve V2d15

GAS & AIR (O2 / N2O 50%/50%)

Medical O2 - O2/CO2

200 bar

Up to 50 lpm

Weight : 1,2 kg

MEDICAL GAS EQUIPMENT

INTERNATIONAL STANDARD CONNECTIONSCAHOUET can design all equipments

with your own standard (CGA - DIN - AFNOR - UNI - MIE

BS - ABNT - IRAM - JIS...)

Don’t hesitate to contact us : commercial@cahouet.com

15 bar

10 bar

200 bar

200 bar

10 bar

10 bar

10 bar 4 +10 bar

4 +10 bar

4 +10 bar