International Radiation Protection Association 11th ...F3F0B03D-43AA-44CF-A54E... · • Tube...

Post on 04-Aug-2020

1 views 0 download

transcript

International Radiation Protection Association11th International Congress

Madrid, Spain - May 23-28, 2004

Refresher Course

Radiation Protection in Cardiac and Interventional Procedures

C. Reek

IRPA 11 Refresher Course 2

Overview

• Introduction and Background• Regulations• Dose• Patient Dose Reduction-Factors• Special Cases• Limitation of Staff Dose• Good Practice• Recommendations

IRPA 11 Refresher Course 3

Sources of Radiation

IRPA 11 Refresher Course 4

November 8th 1895

Wilhelm Conrad Röntgen

IRPA 11 Refresher Course 5

• 23rd January 1896• Public lecture on X rays• Hand of von Kolliker

IRPA 11 Refresher Course 6

First British Radiographs

• 25th January 1896• BMJ Alan Campbell Swinton

IRPA 11 Refresher Course 7

British Medical Journal 18th April 1896

• First published report of the dangers of X rays

IRPA 11 Refresher Course 8

1921

• Deaths attributed to X rays

• No regulations prior to 1928

IRPA 11 Refresher Course 9

Stochastic (Random) Effects

1 in 83,333Majority of NHS staff (<0.3mSv/yr)

1 in 25,000Cancer from radiation exposure of 1 mSv

1 in 43,500Accidents at work

1 in 9500Accidents on road

1 in 850Natural causes (40 yr old)

1 in 200Smoking 10 cigarettes/day

IRPA 11 Refresher Course 10

Deterministic Effects

>25 to eye>250 to eye>5Cataract5050010Telangiectasia

9090018Dermal necrosis

7070014Dry desquamation

353507Permanent epilation

<<11002Transient erythema

Mins fluoro0.2Gy/min

Mins fluoro0.02Gy/min

Threshold Dose to Skin

Injury

IRPA 11 Refresher Course 11

IRPA 11 Refresher Course 12

Development of Interventional Cardiology

1929First documented human

cardiac catheterisationEberswald, GermanyDr Werner Forssman

1958Diagnostic coronary

angiogramDr Mason Sones

IRPA 11 Refresher Course 13

Development of Interventional Radiology

1964Transluminal angioplastyDr Charles Dotter

1967Judkins techniqueDr. Melvin Judkins

IRPA 11 Refresher Course 14

Development of Interventional Cardiology

1974

First peripheral balloon angioplasty

Dr Andreas Gruentzig

1977

First cath lab PTCA on awake patient

IRPA 11 Refresher Course 15

Development of Interventional Cardiology

1980Use of angioplasty in

evolving myocardial infarct

Dr Geoffrey Hartzler

1987First use of coronary stents

in human

IRPA 11 Refresher Course 16

• Transplant post-biopsy

9.9 x 24.6 mm det. balloon

Embolisation: detachable balloon occlusion

IRPA 11 Refresher Course 17

Regulations

• Ionising Radiation Regulations 1999-relate to public and staff safety

• Ionising Radiation (Medical Exposure) Regulations IR(ME)R 2000

-govern the fate of patients undergoing a medical exposure

IRPA 11 Refresher Course 18

Regulations

• Duties of all bodies- employer, practitioner, operator and referrer

• ALARA/ ALARP principle• Training• Diagnostic Reference Levels (DRLs)• Local Rules-policies, procedures, protocols• Other Bodies eg NRPB, ICRP, ACC, BCIS,

European Commission’s Radiation Protection Research Program

IRPA 11 Refresher Course 19

Dose Definitions

• Gray - energy absorbed per unit mass(in diagnostic = KERMA, energy transferred)

• Sievert - equivalent dose=absorbed dose x radiation weighting factor

• Effective dose – equivalent dose in each organ and tissue x tissue weighting factor and summed over whole body

IRPA 11 Refresher Course 20

Dose Definitions

• DAP - dose area product (Gycm²)incident dose x area of X ray field

• Entrance skin dose – absorbed dose in the skin at a given location on the patient (Gy)

IRPA 11 Refresher Course 21

Annual Dose Limits

1 mSv1 mSvFetus50 mSv150 mSv500 mSvOrgans

15 mSv50 mSv150 mSvEyes

1 mSV6 mSv20 mSvWhole body

PublicUnclassified / trainees

Classified Staff

IRPA 11 Refresher Course 22

Monitoring

Ensure

• Film badges and TLDs are easily available

• Results of monitoring are available to all

• Reminders to wear them are in appropriate places

IRPA 11 Refresher Course 23

Dose to patient

13.07070CA + ad hoc PTCA + stent

9.04920PTCA + stent

9.35060CA + ad hoc PTCA

6.93760PTCA

5.63040Coronary angiography

ED mSVDAP cGycm²Procedure

IRPA 11 Refresher Course 24

Diagnostic Reference Levels

• Set for categories of procedures – 20/annum• At least 100 cases• DAP (dose area product) or screening time and

mAs• Level set - 90th percentile• No national DRLs for coronary angiography• NRPB – proposed 36 Gcm²• European DIMOND – proposed 45 Gcm²

IRPA 11 Refresher Course 25

Dose Reduction - Equipment

• Frame rate selection• Pulsed fluoroscopy• Fluoroscopy and image

acquisition dose rate selection

• Last image hold• ‘Replay fluoro’• Electronic magnification• Image processing• Flat panel detectors

IRPA 11 Refresher Course 26

Dose reduction -Equipment

• Road map• Reload facility• Virtual collimation• Intelligent filtration• Dose display

BUT• Most dose reduction

features optional• Improved imaging allows

more complex cases

IRPA 11 Refresher Course 27

Dose Reduction - Equipment

• Modification of existing equipment• Equipment maintenance • Quality assurance

regular IQ and dose checksmanufacturermedical physicists

IRPA 11 Refresher Course 28

Factors affecting exposure

IRPA 11 Refresher Course 29

Inspiration

IRPA 11 Refresher Course 30

Operational Factors

• Arterial access• Oblique views

IRPA 11 Refresher Course 31

Operational Factors

• Suitable kV and mA

• Centring

• Diaphragms-ROI

• Field size -panning

• Magnification

IRPA 11 Refresher Course 32

Appropriate Views

Views should not be prescriptive-depend on patient

IRPA 11 Refresher Course 33

Lateral View

IRPA 11 Refresher Course 34

Operational

• Stenting strategy

IRPA 11 Refresher Course 35

Behavioural

• Table and detector in correct position before screening

• Screening time• Image acquisition –operator / radiographer?• Use of equipment features-dose reduction

programmes etc• Prolonged procedures – reduce skin dose• Operator fatigue

IRPA 11 Refresher Course 36

Radiation Awareness

• Regular audit• Feedback on dose information

patientsstaff

• High standards of equipment maintenance and quality assurance

• Appropriate theoretical training with annual updates

• Practical training and regular re-assessment

IRPA 11 Refresher Course 37

Audit - Operator key

0

1

2

3

4

5

6

7

8

9

Ave . scr. Tim e

Operator Key

1 2

8 9 10 11 12 13 14

15 16 17 18 19 20 21

22 23 24 25 26 27 28

29 30 31 32 33 34 35

36 37 38 39 40 41 42

TOTAL - REG TOTAL - ALL

3 4 5 6 7

IRPA 11 Refresher Course 38

Audit

1513

0 0

23

323740

15

24

46

37

27

0

32

9

1

434342

030 0

47

0

25

40

79

3 2

21

2927

6 5

55

4

0

10

20

30

40

50

60

70

80

No.

No. of cases

Left Heart Catheters - January - March 2001 (No. of Cases)

IRPA 11 Refresher Course 39

Audit

0

500

1000

1500

2000

2500

3000

3500

CGycm2

Ave dose

Left Heart Catheters - January - M arch 2001 (Ave Dose)

IRPA 11 Refresher Course 40

Audit

0 0

13

6

1110

7

4

2 3

19

25

11

0

7

16

19

0

2

4

2524

0

24

17

22

0 0

13

0

15

12

0

5

10

15

20

25

%

% above DRL

Left Heart Catheters - January - March 2001 (% Above DRL)

IRPA 11 Refresher Course 41

Special Cases-Cardiology

• EPS/RFA• Brachytherapy• CT angiography• Radio-isotopes• Echocardiography• Magnetic Resonance Imaging

IRPA 11 Refresher Course 42

Electrophysiology

IRPA 11 Refresher Course 43

Brachytherapy

60mm seed delivery catheter Catheter in diagonal 6 month follow up result

in LAD

Bifurcation in-stent restenosis Two wires Cutting balloon final result

IRPA 11 Refresher Course 44

CT Angiography

• Anomalous right coronary artery arising from LC sinus

• Occlusion of RCA

IRPA 11 Refresher Course 45

Magnetic ResonanceImaging

IRPA 11 Refresher Course 46

Special Cases-Radiology

Per-operative arterial rupture

Covered stent deployment

IRPA 11 Refresher Course 47

Radiology

• Dose ranges depend on type and site of intervention

• Dose reduction to patient -principles as cardiology

• Tube movements less• Position of operator more variable• Use ultrasound guidance where possible• CT fluoroscopy-decreased operation times

IRPA 11 Refresher Course 48

Risk

0.00

0.04

0.08

0.12

0.16

0.20

0 15 30 45 60 75 90

Age at Exposure

Death per Sievert

(Sv)

FemaleMale

IRPA 11 Refresher Course 49

Doses

3.8450Closure of persistent ductusarteriosus

8.4670Tetralogy of Fallot

ED mSvDAP cGcm²

Dose from common paediatric procedures

IRPA 11 Refresher Course 50

Paediatric Intervention

IRPA 11 Refresher Course 51

Limitation of Staff Exposure

• Reduce patient exposure - reduce staff exposure

• Radiation awareness and education• Dedicated interventional equipment• Reduce time of exposure• Use inverse square law• Use shielding by barrier

IRPA 11 Refresher Course 52

Limitation of Staff Dose

IRPA 11 Refresher Course 53

Use lead shields

IRPA 11 Refresher Course 54

Wear appropriate lead protection

• Lead glasses• Thyroid shield• Lead apron – 3.5 - 5mm

lead equivalent• Monitoring

IRPA 11 Refresher Course 55

Use mobile barriers

• Lead shield for operator

• Mobile barrier for radiographer

• Other staff appropriately positioned

IRPA 11 Refresher Course 56

• Position monitors so that operator looks away from beam

IRPA 11 Refresher Course 57

Paediatrics-always a special case

IRPA 11 Refresher Course 58

Good Practice 1

• Follow Local Rules, procedures and protocols• Have all available information about patient

eg. previous grafts, echo data etc• Check patient identity, exposure justification,

consent• Position patient• Ensure all appropriate staff in room are

protected and wearing monitors• Use all lead shielding

IRPA 11 Refresher Course 59

Good Practice 2

• Position table before screening• Keep mA low-kV high (60-90kV for coronaries)• X ray tube at max and II at min distance from

patient• Check staff position• Use dose reduction programmes if possible• Acquire images on full inspiration where

possible• Collimate to area of interest and choose views

carefully

IRPA 11 Refresher Course 60

Good Practice 3

• Prolonged procedure-change beam angulation• Minimise fluoroscopy, high dose rate time,

number of acquisitions• Remember software features to reduce dose eg

replay fluoro• Don’t over use magnification• Remove grid for small patients• Check and record screening time and DAP cf

DRL

IRPA 11 Refresher Course 61

Recommendations

• Dedicated interventional equipment• Radiation dose reduction packages should be mandatory• Radiation dose should be displayed on monitors• Radiation awareness should be promoted by audit and

regular feedback• Local standards-regular review and improvement• Continuing education-including practical training with

annual updates and testing• Research continued-to develop international standards