RIGOL TECHNOLOGIES,INC.

Introduction to S1210 EMI Pre-Compliance

Test Software

Edward Pan

edward.pan@rigol.com

Rev.1.0

2017-OCT

RIGOL TECHNOLOGIES, INC.

S1210 EMI Test Solution

S1210 EMI Pre-Compliance Test Software

RIGOL TECHNOLOGIES, INC.

• Implement EMI Pre-Compliance testing to find root causes of circuitry problem.

• Simplify the Pre-Compliance testing procedure based on complicated testing

environment.

• Predict Pass/Fail result in front of different kinds of EMI testing standards before

sending to formal compliance testing.

• Save your constrained budget under highly competitive market.

Product Highlights

RIGOL TECHNOLOGIES, INC.

• Amplitude Calibration Function

• Standard Limited Line Setting Function

• Sweep Function

• Peak Value Search Function

• Test Report Generation

• Graphic Type Operation

• Peak and Marker Table

• Multi-Display Mode

• Support Demo Mode

Product Features

RIGOL TECHNOLOGIES, INC.

• Pre-define and preview the calibration data and curve

• Self-edit limited line and support two limited line simultaneously

• Support inconsecutive segment sweep function

• Support both conducted and radiated inspection

• Support peak table and marker table

• User-friendly testing report generation

• Support two traces graphical comparison

• Demo Mode offering

Exclusive Advantages

RIGOL TECHNOLOGIES, INC.

• Easily observe the testing data via PC Laptop

• Intuitive testing and calibration concepts to implement different projects.

• Fit multiple EMI international standards via self-edited limit line

• Clear testing result and logical report

• Easily self-adjusted graphical environment to fit users observation

• Apparent and detailed analysis logistics and methodologies.

User Benefits

RIGOL TECHNOLOGIES, INC.

Demo Procedure

RIGOL TECHNOLOGIES, INC.

Required Equipment List

1. DSA875/832/832E/815/710/705 Spectrum Analyzer

2. EMI-DSA800 Option

3. Ultra Sigma Software

4. S1210 EMI Pre-Compliance Test Software

5. NFP-3 Near Field Probe ( For Radiated Test Only )

6. USB A Type to B Type Cable x 1

7. Personal Computer ( With WIN10/8/7/Vista System )

8. EMI DUT Source

RIGOL TECHNOLOGIES, INC.

Use “A to B” Type USB cable to connect Spectrum Analyzer and

PC

Spectrum Analyzer Pre-Setup

A Type Connector

B Type Connector

RIGOL TECHNOLOGIES, INC.

Use NFP-3 Near Field Probe and connect to DSA800 Spectrum

Analyzer

Spectrum Analyzer Pre-Setup

NFP-3 Near Field Probe Set

Connect to DSA800

RIGOL TECHNOLOGIES, INC.

Spectrum Analyzer Setup

1. Press “Preset” on DSA800 firstly to original settings

RIGOL TECHNOLOGIES, INC.

Spectrum Analyzer Setup

2. Press “ FREQ” and Setup “Start” (30MHz), “Stop”(1GHz) frequency on

DSA800

3. Press “BW/Det” and adjust the RBW value to 100KHz

RIGOL TECHNOLOGIES, INC.

Spectrum Analyzer Setup

4. Press “AMPT” on DSA800, and Set the Vertical Axis

Units to “dBuV”

RIGOL TECHNOLOGIES, INC.

5. Start the S1210 EMI Software

6. Press “File” and press ”New” to open a project “test_0”

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

7. Select “test_0” and go to “Device” to connect PC to Spectrum, as you can see

the green USB Address will be shown on the bottom side of the screen

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

8. Go to “Scan Config” and select the Limited Line 1 file

( EN55011_CAG1_Radiated_Qpeak_10m.lim )

9. Set the detector as “Peak”, RBW value to “100KHz”,

and select “Clear/Write” when complete

8

9

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

10. Go to “Segment Config” and Setup the frequency sweeping segment

Start from 30MHz to 1GHz, RBW to 100KHz, Sweep time to 0 sec.

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

11. Use NFP-3 Near Field Probe to approach the left side of DSA800 LCD

Display. ( To inspect the EMI source from CCFL Adaptor of LCD Back

light module)

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

12. Go to “Measure” and implement the “Pre-Scan”

13. The EMI Spectral Lines are shown on the screen

Spectrum Analyzer Setup

RIGOL TECHNOLOGIES, INC.

Back up

RIGOL TECHNOLOGIES, INC.

S1210 EMI Test Solution

Double click the mouse on the graphic and the markers will

shown on the screen and the marker table as well.

RIGOL TECHNOLOGIES, INC.

S1210 EMI Test Solution

Go to “Set Reference” and users can do two trace comparison

RIGOL TECHNOLOGIES, INC.

S1210 EMI Test Solution

Users can “Zoom In” the dedicated peak and observe inside

RIGOL TECHNOLOGIES, INC.

S1210 EMI Test Solution

The Zoom In Spectral Lines

RIGOL TECHNOLOGIES, INC.

How to edit the

Limit Line?

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

1. Press the “Data Manager” at the top of the display

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

2. Select the “limit.lim” file

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

3. Edit the limit line parameters at right side window.

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

4. Export and save the file

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

5. Select one folder to export and save file

RIGOL TECHNOLOGIES, INC.

How to edit the limit line?

6. User can see the limit line file inside and folder and can

import to S1210 Software by “Scan Config” setup item.

RIGOL TECHNOLOGIES, INC.