Technology Transfer Workshop
Method Building Essentials• Source Parameter Optimization
– Extracting the solution ion from the mobile phase
• Compound Optimization for MRM– Getting parent ion to the collision cell and
smashing it to bits
• MS optimization for EPI– Producing reproducible full scan MS with lots of
detail.
• Linking MRM with EPI using IDA
Technology Transfer Workshop
Method Parameters• TurboSpray Parameters
– The goal is to maximize ion formation based on the liquid matrix.
– Set globally in the method.– In multi-compound methods should be generalized
• Compound Parameters– The goal is two fold
• 1st maximize parent ion reaching the collision cell• 2nd Optimize fragmentation
– MS experiment determines if generalized and set globally• MRM = drug specific• EPI = generalized and set globally.
• MS Parameters– The goal is optimized resolution and sensitivity.– Minimize scan time
Technology Transfer Workshop
Infusion and FIA• Infusion
– Syringe pump driven
– 10 mcL/min
– Compounds ~ 5mg/L in H20 mixed with a small amount of 50:50 A:B mobile phase.
• Flow Injection Analysis– Syringe pump still used
– No chromatography
– LC mobile phase added at mixing-T on source. • Syringe pump +A + B = 200 mcl/min• Mobile phase A & B: 0.95 mL/min each• Syringe Pump 10 mcl/min.
Technology Transfer Workshop
TurboSpray Settings
• Ion Spray Voltage [IS]• Nebulizer Gas [GS1]• The “hairdryers”
– Heater [TEM]
– Heater Gas [GS2]
• Curtain Gas [CUR]• Interface Heater [ihe]
– On or Off…we leave it on.
Technology Transfer Workshop
3200 Qtrap Overview: The Source
Curtain Plate
CUR
GS1 IS
TEM &GS2
ihe
Orifice Plate
Voltage, Temperature & Gas Parameters
Technology Transfer Workshop
Nebulizer Gas [GS1]
1020
30 5070 9
0
1020 30 50 70 90
Increasing the GS1 from 10 to 50 Increases cocaine ion flux ~20%
Infusion
FIA
Technology Transfer Workshop
Heater [TEM]
TEM= 0 (no heat)
TEM= 100
TEM= 200
TEM= 300
TEM= 400TEM= 500
TEM= 600
TEM= 0 (no heat)
TEM= 100
TEM= 200
TEM= 300 - 600
Infusion FIAIncreasing TEM 0 to 500Increases cocaine flux >350%
Technology Transfer Workshop
TurboSpray Parameters• The curtain gas [CUR] should be maximized to repel
neutrals and droplets without dramatically impacting sensitivity.– The geometry of the 3200 source makes it less
important than on older sources.• Mobile phase composition and flow affects source
parameters– Increasing water content will require higher temperature
and gas pressures– Increasing flow rate will require increased temperature
and gas pressures.• Using FIA reproduces mobile phase conditions at
compound elution and can help to maximize sensitivity.
Technology Transfer Workshop
Compound Optimization Overview• Select Mass [M+H]+
– Needs to be the exact mass of the most abundant isotope
• Optimize compound voltages – maximize parent ion abundance– DP,EP, & CEP
• Optimize compound fragmentation– Select and optimize transition masses– CE
• Quantitative Optimization– Instrument’s automatic routine
Technology Transfer Workshop
Compound Optimization Parameters
• Getting ions to the collision cell– Declustering Potential [DP]
– Entrance Potential [EP]
– Collision Cell Entrance Potential [CEP]
• Controlling Fragmentation– Collision Energy [CE]
– Collision Gas Setting [CAD]
– Collision Cell Exit Potential [CXP]• Usually 4V
– Collision Energy Spread [CES]
Technology Transfer Workshop
3200 Qtrap: Potentials
• Most potentials are relative to the entrance potential (EP).
• CEM is the setting for the Cascade Electron Multiplier
Q0 Q1 Q2 Q3
EP CEP CXPDP
CE CEM
Technology Transfer Workshop
Declustering Potential• “The potential applied to the orifice plate (OR)
has the greatest effect on the amount of declustering in the orifice region of the interface.”
• “The working range of DP is typically 0 to 100 V, although it may be set higher.”
• Decluster what?– Example clusters include: [M+H3O]+, [M+Na]+,
[M+H+CH3OH]+
• If set too high DP can actually fragment a compound in the source… BAD for LC/MS/MS.
Technology Transfer Workshop
Cocaine DP Ramp 5.0 to 100v
Infusion (10 mcl/min)
FIA 200 mcl/min 50:50 H2O: MeOH
Technology Transfer Workshop
Entrance Potential [EP]
• The EP parameter controls the entrance potential, which guides and focuses the ions through the high-pressure Q0 region.
• It is typically set at 10 V (for positive ions) or
–10 V (for negative ions) and affects the value of all the other instrument voltages.
Technology Transfer Workshop
Collision Cell Entrance Potential
• CEP (Collision Cell Entrance Potential)– The CEP parameter controls the collision cell
entrance potential, which is the potential difference between Q0 and IQ2.
– It focuses ions into Q2 (collision cell). CEP is used in Q1, MS/MS-type, and LIT scans.
• Generally the most mass dependent.
Technology Transfer Workshop
Fragmentation Parameters
• Controlling Fragmentation– Collision Energy [CE]
– Collision Gas Setting [CAD]
– Collision Cell Exit Potential [CXP]• Usually 4V
– Collision Energy Spread [CES]
Technology Transfer Workshop
Collision Energy (CE)• Fragmentation of parent ion is caused by collision with
low pressure N2 in the collision cell.• The more energy generally results in greater
fragmentation.• The energy of that collision is controlled by the
Collision Energy (CE) setting.– CE is a voltage difference between the Q0 and Q2 (EP –
RO2). – CE can be optimized for each drug.
• (Quantitative Optimization)
– Higher CE results in greater fragmentation of the parent molecule.
• Consider dextromethorphan…
Technology Transfer Workshop
Fragment Ion Selection
• In MRM screening, generally a single transition is selected for each drug. – For example dextromethorphan 272 128– The sensitivity of the method is directly
proportional to the intensity of the fragment ion.– Attention should also be given to the
uniqueness or the transition.
• How do you select a transition ion to monitor?
Technology Transfer Workshop
Fragment Ion Selection• Quantitative Optimization
– Starts with compound optimization • DP, EP, & CEP
• Fragment parent mass at many different collision energies.
• Pick the four most intense ions – Avoid loss of H2O
• (fragments δm >19 amu of parent)
• Determine the collision energy that produces the maximum amount of each transition ion.
Technology Transfer Workshop
Optimizing 128, 147, 171, 215
128
147
171
215
Fragment CE Max
128 84
171 51
147 41
215 32
CE
Technology Transfer Workshop
Dwell Time
• Dwell time is the amount of time (msec) the instrument spends at each transition.
• Effects sensitivity up to a point.– Increasing the DT >50 msec produces little if
any additional sensitivity.
– For most drugs DT> 25 has little effect.
• Also longer dwell times lengthen total MRM scan times.
Technology Transfer Workshop
Method Building: MRM
• Optimization of source parameters
• Optimization of compound parameters– Compound
– Fragmentation
• Selection of drug transitions– Repeated 130 times…..
• Congratulations You’re Ready to Build an MRM Method!
Technology Transfer Workshop
MRM Method for DXM
Parent [M+H]+
Daughter Transition
Ion
DwellTime
MRM Parameters
Technology Transfer Workshop
MS Parameters
• Enhanced Product Ion – Global Parameters
– Collision Energy Spread [CES]
– Collision Cell Gas Pressure [CAD]
– LIT Settings• Fill Times• LIT Scan Speed• Q0 Trapping
Technology Transfer Workshop
Dextromethorphan EPI
Typical EPI output...CE =20
CES = 30Q0 trap = Off
Scan Speed = 1000Fill time fixed 20ms
Technology Transfer Workshop
EPI: Global Parameters• Source Settings (TEM, GS1 etc…) are set
globally and are the same for MRM and EPI.
• Compound parameters individualized to optimize transition ion formation (MRM) must be set globally for EPI. (DP,EP,CEP,CE,CXP)
• The goal of EPI optimization is to find the best settings to produce good MS data that is reproducible and library searchable.
Technology Transfer Workshop
Collision Energy Spread [CES]• “The CES parameter controls the spread of collision
energies used when filling the LIT. It is used in conjunction with the Collision Energy (CE) parameter.” – “The advantage of using a collision energy spread is
that you do not have to optimize the collision energy.”
• “By specifying the CE and CES parameters, low, medium, and high collision energies are used in a single scan to provide maximum information in the product ion spectra (low and high mass fragments).”
• We have chosen CE 20:CES 30 (20/30) or (25/30)– Provided the best fragmentation across the widest
number of drugs.
Technology Transfer Workshop
Collision Gas Pressure [CAD]
• Controls the pressure of the N2 gas in the collision cell
• In simplified mode you get three settings: – Variable by instrument
• Low: 1.9x10-5 Torr• Medium: 2.6x10-5 Torr• High: 3.3x10-5 Torr
• Collision with the gas causes fragmentation, but also helps to “cool” fragment ions and focus them into the LIT.
Technology Transfer Workshop
Collision Gas Pressure [CAD]
CAD = High (3.3x10-5 Torr)
CAD = Medium (2.6x10-5 Torr)
CAD = Low (1.9x10-5 Torr)
Technology Transfer Workshop
LIT Settings• The LIT has four important settings
1. Mass Range • Set to cover the range dictated by compounds
covered (e.g.50 to 500 amu)
2. Scan Speed• Slower Scan Speed give better mass resolution
but cost time and sensitivity.• Qtrap has three 250, 1000 and 4000 amu/s
Technology Transfer Workshop
LIT Settings• The LIT has four important settings
3. Fill Time– The amount of time the trap remains open to
accept ions. – Can be fixed time (e.g. 50 msec) or Dynamic– Dynamic fill time (DFT)
– Fill time based on presampling of incoming ion flux– Prevents over or under filling of the LIT.
4. Q0 trapping [on or off]– Like the LIT Q0 can store ions coming in from
the source while the LIT is closed.– Can increase sensitivity
Technology Transfer Workshop
Fixed vs. Dynamic Fill Times
Dynamic Fill Time
Fixed 20ms Fill Time Q0 OFF
Fixed 20ms Fill Time Q0 On
Technology Transfer Workshop
Building a Qualitative Method• Use MRM as the survey method
– Each compound setting optimized for sensitivity and specificity
• Use EPI to generate a full scan mass spectrum with a lot of detail to facilitate positive identification.– Use standard parameters for reproducibility and
library searching.
• Use Information Dependent Acquisition (IDA) to switch between MRM and EPI.
Technology Transfer Workshop
Information Dependent Acquisition• Information Dependent Acquisition
– Allows on-the-fly software switching between MS modes.
• IDA examines– The results of the MRM cycle to see if any
drugs are above the a threshold.• IDA determines
– IF any drugs will be selected for EPI– Which drugs will be selected
• IDA switches – The instrument from MRM mode to EPI mode
Technology Transfer Workshop
IDA Example• For Example
– Experiment 1: MRM• Survey Scan of 100+ drugs
– IDA (Decision Maker)• When a transition is above a threshold…• And is the most intense transition…• And is not on an exclusion list… • Then trigger a second experiment
– Experiment 2: Product Ion Scan• Full scan mass spectrum for compound
identification.
Technology Transfer Workshop
IDA: Dynamic Exclusion• A software technique to handle coeluting peaks
especially when one is much less intense.
Threshold
Dynamic Exclusion
Time
Product Ion Scans
Technology Transfer Workshop
IDA Method Flowchart
130 MRM Transitions
Any over 1100
cps?
Add to exclusion list for 15 sec.
On exclusio
n list ?
3rd EPI?
Do EPI
YES
YES
YES
NO
NO
NO
IDAIDA
Dynamic ExclusionDynamic Exclusion
Technology Transfer Workshop
Putting it all together…
XIC of oxycodone MRM
MRM Triggered EPI
Overlay of 130 MRM transitions