This manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of
ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows:
WARNING A potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
This is a hazard alert mark. This mark is the IEC417 mark for "Caution. Hot surface". Do not touch parts affixed with this mark without adequate protection. This mark informs you about the operation of the product. This manual is subject to change without notice at any time to improve the
product. Product specifications are subject to change without any obligation on the part
of the manufacturer. Under the copyright laws, the instruction manual and the software (program)
described in it are copyrighted, with all rights reserved. The software may be installed into one computer and may not be installed into other computers without the prior written consent of A&D Company. Copying includes translation into another language, reproduction, conversion, photocopy and offer or loan to another person.
The purchaser may make one copy of the software for backup purposes.
The manual and the software may not be copied, in whole or part, except as described above.
Windows is a registered trademark of the Microsoft Corporation. 2004 All rights reserved.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 1
ContentsContentsContentsContents
1. Safety and Compliance..................................................................3
2. Precautions...................................................................................5 2.1. Installing the Analyzer ..................................................................... 5 2.2. During Use ...................................................................................... 6 2.3. After Use and Maintaining the Analyzer ........................................... 7
3. Outline and Features.....................................................................8
4. Packing List and Names of each part ...........................................10 4.1. Display and Keys............................................................................ 12
5. Preparations ...............................................................................14 5.1. Installing the Analyzer ................................................................... 14 5.2. Setting the Clock and Calendar ...................................................... 15
5.2.1. Operation.............................................................................................. 15 5.3. Proper Operation for Precision Measurement ................................. 16
5.3.1. Operation of the sample ....................................................................... 16 5.3.2. Operation of the analyzer ...................................................................... 16 5.3.3. The Glass Fiber Sheets ( of accessory)................................................... 17
6. Measurement Procedure..............................................................18 6.1. Standard Mode Operation .............................................................. 18
6.1.1. ACCURACY............................................................................................ 18 6.1.2. Operation.............................................................................................. 18
6.2. Quick Mode Operation ................................................................... 21 6.2.1. ACCURACY............................................................................................ 21 6.2.2. Operation.............................................................................................. 21
6.3. Program Number ........................................................................... 24 6.3.1. Storing a Measurement Program to a Program Number .......................... 24 6.3.2. Recalling a Measurement Program with a Program Number .................... 24
7. Measurement Programs ..............................................................25 7.1. List of Measurement Programs....................................................... 25
7.1.1. ACCURACY of the Standard Mode and Quick Mode ............................... 26 7.1.2. Analyzing mode of the Automatic Mode ............................................... 27 7.1.3. Analyzing mode of the Timer Mode ...................................................... 27 7.1.4. Drying Program (Heating Pattern and Drying Termperature) ................. 27 7.1.5. Measurement Unit .................................................................................. 28
7.2. Procedures to Store a Measurement Program ................................. 29 7.2.1. Standard Drying .................................................................................... 29 7.2.2. Ramp Drying ......................................................................................... 32 7.2.3. Step Drying ........................................................................................... 36
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
2 MS-70 / MX-50 / MF-50 / ML-50
8. Check Function ...........................................................................41 8.1. Self-Check Function....................................................................... 41
8.1.1. Operation.............................................................................................. 41 8.2. Test Sample (Sodium Tartrate Dihydrate) ....................................... 42
9. Connecting to a Printer ...............................................................43 9.1. Print samples................................................................................. 44
9.1.1. Example To Print The Whole Data At One Time..................................... 44 9.1.2. Example To Print Selected Items ........................................................... 45 9.1.3. Explanation for Print Item ..................................................................... 47
10. Connecting to a Computer .........................................................49 10.1. RS-232C Serial Interface................................................................ 50 10.2. Output Format............................................................................... 51 10.3. Command...................................................................................... 52
11. Data Memory Function ...............................................................53 11.1.1. Preparation ......................................................................................... 53 11.1.2. Output All Data at One Time............................................................... 54 11.1.3. Delete All Data at One Time................................................................ 54
12. Calibration.................................................................................55 12.1. Identification Number (ID No.) ....................................................... 55
12.1.1. Setting the ID Number ........................................................................ 55 12.2. Calibration of the Weighting Sensor............................................... 56
12.2.1. Operation............................................................................................ 56 12.3. Calibration of Drying Temperature (for MS-70 and MX-50) ........................... 58
12.3.1. Operation............................................................................................ 58
13. Function Table ...........................................................................60 13.1.1. Operation............................................................................................ 61
14. Maintenance ..............................................................................62 14.1. Cleaning the Heater Unit................................................................ 62 14.2. Replacement of the Halogen Lamp................................................. 63 14.3. Factory Settings............................................................................. 64
14.3.1. Operation............................................................................................ 64 14.4. Troubleshooting............................................................................ 65 14.5. Error Message................................................................................ 66
15. Specifications.............................................................................67 15.1. Dimensions ................................................................................... 68 15.2. Accessories and Peripheral Equipment........................................... 69
16. Index.........................................................................................70
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 3
1.1.1.1. Safety Safety Safety Safety and Complianceand Complianceand Complianceand Compliance
WARNING Do not use a sample that could make a dangerous chemical reaction and cause an
explosion or poisonous gas, when the sample is dried. Keep flammables away from the analyzer. Parts of the analyzer become very hot. Materials placed near it might catch fire. Do not use the analyzer in ambient ignitable gas. It may cause explosion and fire. Use a power source (voltage, frequency, outlet type) adapted to the specification of
the analyzer. If excessive voltage is used, the analyzer may overheat and be damage or cause a fire.
Turn off the power switch and remove the power cord from the power outlet, when replacing the halogen lamp. Touching an electrode of the halogen lamp connector carelessly, it may cause to receive an electric shock.
Do not disassemble the analyzer. It may cause an error, damage, receiving an electric shock or fire. If the analyzer needs service or repair, contact the local A&D dealer.
Avoid getting the analyzer wet. It is not a water-resistant structure. If there is leakage of liquid into the analyzer, it may cause damage to the analyzer or receiving electric shock.
Do not look at the active halogen lamp to protect your eyes from damage. Do not drop, hit or crack the glassware including the halogen lamp, to avoid an
injury. When the halogen lamp is used beyond 5000 hours, we recommend replacing the
lamp with a new one to avoid trouble. When discarding a halogen lamp, do not break it to avoid scattering glass and injury.
CAUTION Do not touch the heater cover, the halogen lamp, glass-housing, pan handle, sample
pan and sample without adequate protection, it could cause a burn or scald. Parts of the analyzer are very hot when a measurement finishes. For operation, use the specified grips of the heater cover and pan handle. Use the standard accessory tools.
Do not touch parts affixed with the mark, because they may get very hot and dangerous.
When the analyzer is used in a room where hot air does not diffuse, it may unexpectedly overheat. In this case, adjust the drying temperature or move the analyzer to a place with adequate ventilation.
Avoid leaving the analyzer in direct sunlight, as that could cause discoloration of the case or a malfunction.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
4 MS-70 / MX-50 / MF-50 / ML-50
Compliance with FCC RulesCompliance with FCC RulesCompliance with FCC RulesCompliance with FCC Rules Please note that this device generates, uses and can radiate radio frequency energy.
This device has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when this device is operated in a commercial environment. If this unit is operated in a residential area, it may cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
Compliance with CouCompliance with CouCompliance with CouCompliance with Council Directivesncil Directivesncil Directivesncil Directives This device features radio interference suppression and safety regulation in
compliance with the following Council Directives Council directive 89/336/EEC EN61326 EMC directive Council directive 73/23/EEC EN61010-1 Low voltage directive
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 5
2.2.2.2. PrecPrecPrecPrecautionsautionsautionsautions
2.1.2.1.2.1.2.1. Installing the AnalyzerInstalling the AnalyzerInstalling the AnalyzerInstalling the Analyzer
------------Caution for MCaution for MCaution for MCaution for Measureeasureeasureeasurement Safety ment Safety ment Safety ment Safety ------------ Do not install the analyzer in a dangerous place. Maintain the following ambient condition to operate the analyzer. 5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation) Keep flammables away from the analyzer. Do not put anything on the heater cover. Do not install the analyzer in a small airtight room. If the analyzer is used in an
airtight room, hot air does not diffuse, the sample may unexpectedly overheat. In this case, the safety circuit of the halogen lamp activates. Move the analyzer to a place with adequate ventilation or adjust the drying temperature.
There is the voltage label on the back panel of the analyzer. Confirm that voltage, frequency and outlet type is correct for your local voltage.
Confirm that the rated voltage of the halogen lamp is correct for your power supply voltage. (Refer to 14.4.Troubleshooting)
Voltage Label Power Supply Voltage The Rated Voltage of Halogen Lamp 100 - 120 V AC 100 V to AC 120 V AC 120 V 200 - 240 V AC 200V to AC 240 V AC 240 V
Ground the analyzer using the ground terminal of the power cord. Do not change the setting of the I/II switch on the rear of the analyzer. If the
incorrect setting is used, it may damage the analyzer or cause a fire.
------------Caution for Precision MeasurementCaution for Precision MeasurementCaution for Precision MeasurementCaution for Precision Measurement------------ Confirm the following condition, because the weighing sensor (S.H.S.) is very sensitive. The weighing surface should be solid and free from vibration, drafts and as level as
possible. Install the analyzer in a stable place avoiding vibration and shock. Install the analyzer where it not affected by heaters or air conditioners. Ensure a stable power source. Keep the analyzer away from equipment that generates magnetic fields. Discharge static electricity.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
6 MS-70 / MX-50 / MF-50 / ML-50
2.2.2.2.2.2.2.2. During UseDuring UseDuring UseDuring Use
------------Caution for MCaution for MCaution for MCaution for Measureeasureeasureeasurement Safety ment Safety ment Safety ment Safety ------------ Operate the analyzer using the following procedure. Put the sample pan in the correct position. Handle the grip of the heater cover to open and
close it. Use the pan handle to move the sample pan. Do not touch hot parts around the grass-housing,
when the cover is opened. The glass-housing is very hot. It may cause a burn,
if touched. The sample pan and pan handle is very hot, when
finishing measurement. Allow them cool down. Use the tweezers or spoon to move the sample. Grips and Hot Parts. Hot parts are as follows: Use the following grips to operate the analyzer. Do Not Measure a Dangerous Sample. Do not use an explosive, flammable or noxious substance as a sample. Do not use a sample that makes a dangerous substance by drying it. Do not use unknown substances. When a sample surface becomes dry first and the inner pressure increases, the
sample may explode. Do not use such a sample. Turn off the power switch if a sample catches fire. The case of the analyzer is made of a flame-retardant substance (UL94V0). Do Not Put any Flammable Matter Around the Analyzer. During and after measurement, parts of the analyzer become very hot. Do not put
flammable matter near the analyzer. Do not put any thing on the heater cover.
Grip of pan handle
Hot Hot
HotGrips of
heater cover
Grip of heater cover
Grip of pan handle
Breeze break ring
Sample pan
Glass-housing
Pan support
Pan handle
Put correctly
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 7
Caution for Heating (Drying). When the drying temperature is set to 200°C and measurement is started, the
thermostat of the halogen lamp may work after 30 minutes. When the halogen lamp has cooled down, the next measurement can be started. If necessary, change the drying time and temperature.
When a measurement is started and the time passes one hour, the maximum temperature is automatically regulated to 160°C for safety.
Operation to Stop Measurement During measurement, the STOP key is always effective. If there is an error or danger,
press the STOP key.
2.3.2.3.2.3.2.3. After Use and Maintaining the AnalyzerAfter Use and Maintaining the AnalyzerAfter Use and Maintaining the AnalyzerAfter Use and Maintaining the Analyzer Put dust cover on the analyzer after it is cool. Clean the glass-housing carefully. Clean fingerprints from the halogen lamp to keep
its life. Refer to "14.2.Replacement of the Halogen Lamp".
Avoid mechanical shock to the analyzer. Do not disassemble the analyzer. Protect the analyzer from excessive dust. Use the packing box (special container) to move
the analyzer. Clean the analyzer with a lint free cloth that is
moistened with warm water and a mild detergent. Do not use organic solvents to clean the analyzer. Do not disassemble or remodel the analyzer.
Glass-housing
Halogen lamp
Glass housing
Screw
Screw
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
8 MS-70 / MX-50 / MF-50 / ML-50
3.3.3.3. Outline and Outline and Outline and Outline and FeaturesFeaturesFeaturesFeatures
The moisture analyzer is built using a super hybrid sensor (S.H.S.) adopted in an
analytical balance. Therefore, the result is more precise and gets greater repeatability. An analyzer using the S.H.S. has high sensitivity, needs only a sample quantity of a
few grams, and the analysis time becomes shorter. A 400W halogen lamp is used as the heating source and the temperature on the
sample pan can reach 200°C within two minutes. There are five analysis modes. Standard mode..... The moisture content can be obtained with settings of the
drying temperature and accuracy. Quick mode........... Sample is heated up for approximately three minutes at 200°C so
that analysis time becomes shorter. The moisture content can be obtained with settings of the drying temperature and accuracy.
Automatic mode .... When the change of moisture content per one minute is less than the preset termination value, the measurement is automatically stopped and the result is obtained.
Timer mode ............ The sample is dried for a preset time and the result is obtained. Manual mode.......... This mode can stop the measurement by key operation and the
result is decided. The heating patterns can be used to analysis mode without quick mode.
(For ML-50, standard drying and quick drying can be used only) Standard drying ..... Maintains a constant drying temperature. Ramp drying........... Increases the drying temperature gently. Step drying ............. Uses multiple steps of the drying temperature. Quick drying........... Heats up 200°C for few minutes and uses a constant drying
temperature. The analyzer can store and recall proper individual settings for each sample using a
program number (PROG No.). MS-70 / MX-50 MF-50 ML-50 Maximum program number 20 sets 10 sets 5 sets
The data memory function can store results and output them at one time.
MS-70 / MX-50 MF-50 ML-50 Maximum number of storable results 100 50 30 The software "WinCT-Moisture", the standard accessory of the MS-70 and MX-50,
has a function that can make a graph of the change of moisture content in realtime and has an optimum temperature search program that judges heating at an appropriate temperature setting.
The software " WinCT", the standard accessory of the MF-50, is communication
software for transmitting data to a computer using Microsoft Windows.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 9
The analyzer is equipped with a serial interface as standard. It can be connected to a printer or computer.
The analyzer can calibrate the weighing sensor (Use special mass.) and drying
temperature (Use temperature calibrator for MS-70 and MX-50 except MF-50 and ML-50). The analyzer can output the data required at GLP, GMP and ISO at the end of the calibration.
The analyzer has a self check function that can detect function errors. The analyzer displays the current change of moisture content per one minute [%/ min]
in real-time. It can be used for the reference to find the analyzing mode. The sample pans included in the standard accessory can be used repeatedly. There is a test sample that is used to check the moisture accuracy. (The test sample
is a part of standard accessory except ML-50) The glass fiber sheets can be used to quick and precision measurement for liquid
sample. (The glass fiber sheets are a part of standard accessory except ML-50) A reference card is built into the bottom of the analyzer. Principle and UsePrinciple and UsePrinciple and UsePrinciple and Use The moisture analyzer, based on the principle of thermogravimetric analysis, dries a
sample using a halogen lamp and obtains the moisture content in % and other results by the difference between the wet weight and dry weight.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
10 MS-70 / MX-50 / MF-50 / ML-50
4.4.4.4. Packing List and Names of each partPacking List and Names of each partPacking List and Names of each partPacking List and Names of each part Keep the packing box to move the analyzer. Packing list as follows:
Heater cover
Display
Keys
Leveling foot
Grip of pan handle
Bubble spirit level Power switch
Grips of heater cover
Grips of heater cover
Reference card
Sample pan
Breeze break ring
Pan handle
Glass-housing
Halogen lamp
Pan support
Fuse, T6.3A 250V
RS-232C serial interface
Fuse, T100mA 250V
Power input
Power supply voltage label
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 11
All AccessoriesAll AccessoriesAll AccessoriesAll Accessories
Caution Please confirm that the analyzer is correct for your local voltage, receptacle type and power cord.
#1#1#1#1 Differences of AccessoryDifferences of AccessoryDifferences of AccessoryDifferences of Accessory
MS-70 MX-50 MF-50 ML-50
Dust cover Test sample Glass fiber sheets Spoon Tweezers
Included.
RS-232C serial interface cable CD-ROM WinCT-Moisture WinCT
Not included. They can be purchased by
your order.
Sample pan 20 10 Pan handle 2 1
Sample pan #1
Pan handle #1
Pan support
Breeze break ring
Display cover
Power cord
Test sample #1Sodium tartrate dihydrate, 30g
RS-232C serial interface cable #1
CD-ROM #1 Spoon #1
Tweezers #1
Dust cover #1
Instruction manual
Spare fuse, T100mA 250V
Spare fuse, T6.3A 250V
Glass fiber sheets #1
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
12 MS-70 / MX-50 / MF-50 / ML-50
4.1.4.1.4.1.4.1. Display and KeysDisplay and KeysDisplay and KeysDisplay and Keys
Name State and Meaning At gram display Preset time is displayed at timer mode 1 Time At measurement Analysis time At gram display Set temperature of sample pan 2 Temperature of sample pan At measurement Current temperature of sample pan
PROG: Program No. At gram display Program number of measurement program MEM: Data No. Storing data Data number of data memory function 3 Analysis mode At setting Symbols: 5td, quc, U-a, U-t, U-m
At gram display Sample quantity [g] 4 Value At measurement Current moisture content [%] 5 Accuracy Accuracy indicator of measurement
6
Operation indicator Indicator of heater cover, sample and drying process Sample needs at least 0.1 g or more to start measurement.
7 Level indicator Reference of sample quantity for standard mode and quick mode.
Standard mode Quick mode Target quantity of sample [g] Target quantity At gram display Automatic mode Preset termination value [%/min.]
8
Drying rate At measurement Current drying rate [%/min.] Measurement unit
% MOIST D D/W/D
Moisture content W - D is based on W W
% MOIST D D/W/D
Moisture content (Atro) W - D is based on D D
% MOIST D D/W/D
D W
% MOIST D D/W/D
W D
W: Wet sample mass D: Dried sample mass9
g
Heating pattern #1 Standard drying
Maintains a constant drying temperature.
Ramp drying Increases the drying temperature gently.
Step drying Uses multiple steps of the drying temperature.
10
Quick drying Quick mode
Lights when heater cover is closedBlinks during measurement. Disappears when not measuringSample mark: Lights when the sample is 0.1 g or more.
Dry content Ratio Gram value
x100 x100 x100 x100
7
8
654
3
2
1 10 #2
9
Proper sample quantity range
Drying temperature
Drying temperature
Drying temperature
Drying temperature 200°Capprox. 3min.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 13
#1 ML-50 can select "standard drying " and "quick drying" only. #2 ML-50 does not display "heating pattern". Display Samples for Analysis ModeDisplay Samples for Analysis ModeDisplay Samples for Analysis ModeDisplay Samples for Analysis Mode
Analysis mode Symbols ( during settings) Gram display
(after settings and before measurement)
Standard mode 5td
Quick mode quc
Automatic mode U-a
Timer mode U-t
Manual mode U-m
Key Operation and Key FunctionsKey Operation and Key FunctionsKey Operation and Key FunctionsKey Operation and Key Functions
Keys Function and action Stores or recalls measurement program with the program number during the gram display.
PROGRAM Selects a heating pattern while the drying temperature is selected.
SELECT Selects item in the measurement program.
, ↓ , ↑ Changes value of item in the measurement program.
ENTER Stores current condition in the measurement program number. Output data at measurement.
START Start prepared measurement Sample needs at least 0.1 g or more to start measurement.
STOP Stop current measurement
RESET Sets the display to zero in the unit of gram. Cancel key.
Target quantity
Preset termination value of analyzing mode
Preset time
Target quantity
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
14 MS-70 / MX-50 / MF-50 / ML-50
5.5.5.5. PreparationPreparationPreparationPreparationssss
5.1.5.1.5.1.5.1. Installing the AnalyzerInstalling the AnalyzerInstalling the AnalyzerInstalling the Analyzer
1. Select the place to install the analyzer.
Refer "2.1.Installing the Analyzer". 2. Level the analyzer by adjusting the leveling feet
and confirm it using the bubble spirit level. 3. Read the power supply voltage label on the back of
the heater cover. Confirm that voltage, frequency and outlet type
is correct for your local voltage. Confirm that the rated voltage of the halogen
lamp is correct to your power supply voltage.
Voltage Label Power Supply Voltage
The Rated Voltage of the Halogen Lamp
100 - 120 V AC 100 V to AC 120 V AC 120 V 200 - 240 V AC 200V to AC 240 V AC 240 V
4. Confirm that the power switch is
"OFF" position. 5. Connect the power cord. Ground
the analyzer with the earth terminal on the power cord.
6. Align the guide hole of the breeze
break ring to the guide on body. 7. Install the pan support.
Align together the ∆ signs on the pan support and body.
8. Put the sample pan on the pan
handle. And hook the pan handle on the
notch in the rim of the breeze break ring.
Leveling foot
Bubble spirit level
Voltage label
Earth terminal
Power cord
Outlet with a earth terminal Example
Sign ∆ on pan support
Sign ∆ on body
Guide hole
Sample pan
Pan handle
Pan support
Breeze break ring
Guide on body
Notch in the rim
Notch in the rim
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 15
5.2.5.2.5.2.5.2. Setting the Clock and CSetting the Clock and CSetting the Clock and CSetting the Clock and Calendaralendaralendaralendar
Adjust the built-in clock and calendar before use.
5.2.1.5.2.1.5.2.1.5.2.1. OperationOperationOperationOperation 1. Turn on the analyzer.
The gram unit (of weighing mode) is displayed. 2. Press and hold the SELECT key to display Cl adj . 3. Press the ENTER key to display the calendar. Example: 15th April, 2002 4. To skip the calendar settings. Press the ↓ or ↑ key to proceed step 5. To adjust the calendar settings. Press SELECT key. Adjust the calendar using the
following keys. SELECT key..... Selects a figure. ↓ , ↑ key .............. Selects a value for the figure. ENTER key ...... Stores the current date and
proceeds to step 5. RESET key....... Cancels the adjustment and
proceeds to step 5. Symbols and arrangement of the calendar
ymd ................... Year, month, day mdy ................... Month, day, year dmy ................... Day, month, year
The arrangement of the calendar is used for the report of GLP, GMP and ISO.
5. Time is displayed. 6. To finish the adjustment. Press the RESET key to proceed step 7. To adjust the clock. Press the SELECT key. Adjust the clock using the
following keys. SELECT key..... Selects a figure. ↓ , ↑ key .............. Selects a value for the figure. ENTER key ...... Stores time and proceeds to
step 7. RESET key....... Cancels the adjustment and
proceeds to step 7. 7. When finishing the adjustment, dp is displayed. Press the RESET key to return to the weighing mode.
0.000g
Cl adj
ymd 02.04.15
12:34:56
ymd 02.04.15
dp
0.000g
12:34:56
To time
Date adjustment
Time adjustment
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
16 MS-70 / MX-50 / MF-50 / ML-50
5.3.5.3.5.3.5.3. Proper Operation for Precision MeasurementProper Operation for Precision MeasurementProper Operation for Precision MeasurementProper Operation for Precision Measurement
5.3.1.5.3.1.5.3.1.5.3.1. Operation of the sampleOperation of the sampleOperation of the sampleOperation of the sample Use a proper sample quantity. If the quantity is small, precise results may not be possible. If the moisture content of the sample (example: plastic) can be estimated that is less
than 1%, the mass of moisture is not enough for the measurement, when measuring light sample, the result may not be accurate. Consider the following sample mass for the measurement. An estimate of moisture content 1% 0.5% 0.1% Necessary mass for measurement 2 g at least 5 g at least 20 g at least
If the measurement is repeated, maintain the same sample quantity. Crush grain samples to a small, uniform powder for a quick drying process. Spread the sample as evenly as possible. The analyzer is designed to measure the moisture content
of the sample by its weight change. If the sample includes volatile matter, it may vaporize during drying causing a measurement error.
When measuring a liquid or liquid state sample that may make a film on the surface, we recommend you use a glass fiber sheet (AX-MX-32-2).
5.3.2.5.3.2.5.3.2.5.3.2. Operation of the analyzerOperation of the analyzerOperation of the analyzerOperation of the analyzer Press the RESET key to display the zero value before each
measurement. Check that the displayed sample weight is stable before
measurement. Press the START key to start a measurement.
Select the proper Analysis Mode to finish a measurement. Use the change of moisture content per one minute [%/min] that is displayed during measurement as a reference value.
The analyzer needs a pre-heating process before measurement. When measuring samples repeatedly or continuously, the first result is always different from the others.
The pre-heating process is as follows: Put a sample pan, instead of a sample, on the pan. Press the START key to heat it. The analyzer temperature becomes equilibrium.
Use a sample on the sample pan that has cooled to room temperature. When a sample is put on a hot sample pan, the moisture content is diffused before measurement, and precise results are not possible. We recommend you use multiple sample pans.
Do not pile up sample pans during a measurement. Avoid drift and vibration of air conditioners. It may cause "measurement error" and
"unstable value". In particular, because MS-70 is sensitive instrument, it is necessary to consider these influences.
When the difference between ambient temperature and sample temperature is small, it may cause temperature control error. Example: If the heating temperature is set to 50°C, it is affected by an room temperature.
Check the activation of the halogen lamp with the operation indicator. Example: If low heating temperature is set, the brightness of halogen lamp becomes dark.
0.000g
5.43% 2.02
%/min
Result display
No good Good
Check the activation with blinking
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 17
5.3.3.5.3.3.5.3.3.5.3.3. The Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber SheetThe Glass Fiber Sheetssss Use the glass fiber sheet to measure the moisture content
that is included in the following samples. When this sheet is used, vaporization of moisture is speeded up and moisture measurement result becomes quick and precise.
This accessory (AX-MX-32-1) is a package of 100 glass fiber sheets. This accessory is inclued in the packing for MS-70, MX-50 and MF-50. Purchase this accessory by your order for ML-50. Use the glass fiber sheet (AX-MX32-1) for high surface tension liquid sample.
Example 1 : "Liquid Sample" Or "Meltable Sample" When the glass fiber sheet is used for these measurements, moisture is more apt to
vaporize because of expanding the surface area and space. And the glass fiber sheet has the effect of preventing a surface film from forming at drying process.
A sample including a lot of moisture : Milk, yogurt, soybean milk, condensed milk, ketchup, resin, liquid paste, hand soap, etc.
A sample that melts and adheres to the sample pan. Example: Chewing gums, caramel, hony, etc.
ProcedureProcedureProcedureProcedure (Preparation before heating) (Preparation before heating) (Preparation before heating) (Preparation before heating)
1 Put the glass fiber sheet on the sample pan. 2 Press the RESET key to display zero. 3 Soak the sample into the glass fiber sheet or put the
sample on the glass fiber sheet. 4 Press the START key to start the measurement.
Example 2 : "If The Sample's Surface Is Apt To Carbonize" When the sample is covered with a glass fiber sheet, carbonization of the sample surface
decreases. Therefore the measurement result becomes repeatable and precise. A sample including a lot of sugar, protein and oil Example: Hony, soybean powder, green tea leaves, cookies, etc. A sample that has dark surface and is apt to carbonize Example: Coffee, peanuts, etc.
Procedure Procedure Procedure Procedure (Preparation before heating)(Preparation before heating)(Preparation before heating)(Preparation before heating) 1 Put a glass fiber sheet on the sample pan. 2 Press the RESET key to display zero. 3 Remove the glass fiber sheet. 4 Put the sample on the pan.
Cover the sample with the glass fiber sheet. 5 Press the START key to start the measurement.
0.000g
Glass fiber sheet
Sample pan
Liquid sample
Zero display
0.000g
Sample
Glass fiber
Sample pan
Zero display
No good Good Spread the sample as
evenly as possible.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
18 MS-70 / MX-50 / MF-50 / ML-50
6.6.6.6. Measurement ProcedureMeasurement ProcedureMeasurement ProcedureMeasurement Procedure
6.1.6.1.6.1.6.1. Standard Mode OperationStandard Mode OperationStandard Mode OperationStandard Mode Operation The standard mode can obtain the moisture content with settings of ACCURACY
and drying program (heating pattern, drying temperature).
6.1.1.6.1.1.6.1.1.6.1.1. ACCURACYACCURACYACCURACYACCURACY ACCURACY of measurement can be set either HI, MID. or LO. The sample quantity is automatically selected by ACCURACY. The termination value of the analyzing mode is automatically selected by
ACCURACY and minimum scale value of % display. The analysis mode is the program to finish the measurement when a change of
moisture content per one minute becomes smaller than a preset termination value. The settings are as follows: Specify an ACCURACY.
ACCURACY Model Minimum scale HI MID. LO
0.001 % 0.01 %/min 0.02 %/min 0.05 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/minMS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/minMX-50 0.1 % 0.10 %/min 0.20 %/min 0.05 % 0.05 %/min 0.10 %/min 0.1 % MF-50 1 % 0.1 %
Preset Termination value
ML-50 1 %
0.10 %/min 0.20 %/min 0.50 %/min
Sample quantity 10 g 5 g 1 g Use Precise result ↔ Quick measurement
6.1.2.6.1.2.6.1.2.6.1.2. Operation Operation Operation Operation This operation explanation uses the following example of the MX-50: Refer to "7.Selection of Measurement Method" for detail. Input Parameters Analysis mode ...................................Standard mode Drying temperature ............................ 130 °C ACCURACY....................................... LO Sample quantity .................................Approximately 1 g (Automatic selection) Analysis mode ................................... 0.10 %/min (Automatic selection) Stored Parameters (Factory Settings for the MX-50) Heating pattern ..................................Standard drying ( ) Measurement unit ...............................Moisture content is based
on a wet sample Minimum scale value of % display........ 0.01 % Minimum scale value of gram display ... 0.001 g Data memory function........................Not used
% MOIST D/W/D
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 19
1. Turn on the analyzer. The gram unit (of weighing mode) is displayed.
Enter the Standard ModeEnter the Standard ModeEnter the Standard ModeEnter the Standard Mode 2. Press the SELECT key to display an analysis
mode and press the ↓ or ↑ key to select 5td .
Select ACCURACYSelect ACCURACYSelect ACCURACYSelect ACCURACY 3. Press the SELECT key to select ACCURACY. When ACCURACY is selected, HI, MID. or LO blinks. 4. Press the ↓ or ↑ key to select LO of ACCURACY.
Set Drying Temperature at the Sample PanSet Drying Temperature at the Sample PanSet Drying Temperature at the Sample PanSet Drying Temperature at the Sample Pan 5. Press the SELECT key to select the drying
temperature. 6. Press the ↓ or ↑ key to set 130 °C.
Store the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the Operation 7. Press the ENTER key to store the parameters. The weighing mode is automatically displayed.
Put a Sample on the PanPut a Sample on the PanPut a Sample on the PanPut a Sample on the Pan 8. Put the breeze break ring, pan support, pan handle
and sample pan in order. (With no sample.) 9. Close the heater cover. 10. When displaying a stable value, Press the RESET
key. Avoid mechanical vibration, breeze and environmental noise during measurement. If it deviates from zero display, press the RESET key.
Close
Preparation
Standard mode
ACCURACY
Temperature
Standard drying ML-50 does not display
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
20 MS-70 / MX-50 / MF-50 / ML-50
11. Open the heater cover. Put in a sample using the level indicator.
Caution The sample needs to be more than 0.1 g. Spread the sample evenly. 12. If the ↑ key is pressed during gram display, analysis
mode, measurement unit and minimum scale value are displayed. If the ENTER key is pressed, the sample mass is output.
Start the MeasurementStart the MeasurementStart the MeasurementStart the Measurement 13. Close the heater cover. Press the START key
after a stable value is displayed. 14. If the SELECT key is pressed during measurement,
other units can be displayed temporarily and its measurement values can be output.
Caution Do not press the SELECT key while
sampling data.
Finish the MeasurementFinish the MeasurementFinish the MeasurementFinish the Measurement 15. When the change of moisture content per one
minute (drying rate) reaches the preset termination value, the measurement is completed and the buzzer beeps.
16. Open the heater cover and remove the sample
using the pan handle. 17. Press the following key to return to gram display. ENTER key..... Outputs (Prints) result. SELECT key ... Returns to the weighing mode. RESET key ..... Returns to the weighing mode and
displays zero. 18. If the same condition is used, proceed to step 8. If changing the condition, proceed to step 2. The sample pans can be washed and reused. There is the Reference card on the bottom of the analyzer.
Sample needs more than 0.1 g. Significant sample is indicated segment.
Range of proper sampleNo good Good
Close
Remove sample
Drying temperature
Analysis time
Drying rate
Result
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 21
6.2.6.2.6.2.6.2. Quick Mode OperationQuick Mode OperationQuick Mode OperationQuick Mode Operation
The quick mode can obtain the moisture content with
settings of ACCURACY and drying temperature. Sample is heated up at 200°C for approximately three
minutes so that moisture content is measured quickly.
6.2.1.6.2.1.6.2.1.6.2.1. ACCURACYACCURACYACCURACYACCURACY ACCURACY of measurement can be set either HI, MID. or LO. The sample quantity is automatically selected by ACCURACY. The termination value of the analysis mode is automatically selected by
ACCURACY and minimum scale value of % display. Analysis mode is the program to finish the measurement when a change of moisture
content per one minute becomes smaller than a preset termination value. The settings are as follows: Specify an ACCURACY.
ACCURACY Model Minimum scale HI MID. LO
0.001 % 0.02 %/min 0.05 %/min 0.05 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/minMS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/minMX-50 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.05 % 0.1 % MF-50 1 %
0.10 %/min 0.20 %/min 0.50 %/min
0.1 %
Preset Termination value
ML-50 1 % 0.20 %/min 0.50 %/min 1.00 %/min
Sample quantity 5 g 2 g 1 g Use Precise result ↔ Quick measurement
6.2.2.6.2.2.6.2.2.6.2.2. Operation Operation Operation Operation This operation explanation uses the following example of the MX-50 : Refer to "7.Selection of Measurement Method" for detail. Input Parameters Analysis mode ...................................Quick mode Heating pattern ..................................Quick drying ( , Automatic selection) Drying temperature ............................ 130 °C ACCURACY....................................... LO Sample quantity .................................Approximately 1 g (Automatic selection) Analyzing mode ................................. 0.20 %/min (Automatic selection) Stored Parameters (Factory Settings for the MX-50) Measurement unit ...............................Moisture content is based
on a wet sample Minimum scale value of % display........ 0.01 % Minimum scale value of gram display ... 0.001 g Data memory function........................Not used
% MOIST D/W/D
200°CApprox. 3 min. Drying temperature
Heating pattern
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
22 MS-70 / MX-50 / MF-50 / ML-50
1. Turn on the analyzer. The gram unit (of weighing mode) is displayed.
Enter the Quick ModeEnter the Quick ModeEnter the Quick ModeEnter the Quick Mode 2. Press the SELECT key to display an analysis
mode and press the ↓ or ↑ key to select quc .
Select ACCURACYSelect ACCURACYSelect ACCURACYSelect ACCURACY 3. Press the SELECT key to select ACCURACY. When ACCURACY is selected, HI, MID. or LO blinks. 4. Press the ↓ or ↑ key to select LO of ACCURACY.
Set Drying Temperature at the Sample PanSet Drying Temperature at the Sample PanSet Drying Temperature at the Sample PanSet Drying Temperature at the Sample Pan 5. Press the SELECT key to select the drying
temperature. 6. Press the ↓ or ↑ key to set 130 °C.
Store Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the Operation 7. Press the ENTER key to store the parameters. The weighing mode is automatically displayed.
Put Put Put Put a a a a Sample on the PanSample on the PanSample on the PanSample on the Pan 8. Put the breeze break ring, pan support, pan handle
and sample pan in order. (With no sample.) 9. Close the heater cover. 10. When displaying a stable value, Press the RESET key.
Avoid mechanical vibration, breeze and environmental noise during measurement. If it deviates from zero display, press the RESET key.
Close
Preparation
Quick mode
ACCURACY
Temperature
Quick drying ML-50 does not display
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 23
11. Open the heater cover. Put in a sample using the level indicator.
Caution The sample needs to be more than 0.1 g. Spread the sample evenly. 12. If the ↑ key is pressed during gram display, analysis
mode, measurement unit and minimum scale value are displayed. If the ENTER key is pressed, sample mass is output.
StarStarStarStart the Measurementt the Measurementt the Measurementt the Measurement 13. Close the heater cover. Press the START key
after a stable value is displayed. 14. If the SELECT key is pressed during measurement,
other units can be displayed temporarily and its measurement values can be output.
Caution Do not press the SELECT key while
sampling data.
Finish the MeasurementFinish the MeasurementFinish the MeasurementFinish the Measurement 15. When the change of moisture content per one
minute (drying rate) reaches the preset termination value, the measurement is completed and the buzzer beeps.
16. Open the heater cover and remove the sample
using the pan handle. 17. Press the following key to return to gram display. ENTER key..... Outputs (Prints) result. SELECT key ... Returns to the weighing mode. RESET key ..... Returns to the weighing mode and
displays zero. 18. If the same condition is used, proceed to step 8. If changing the condition, proceed to step 2. The sample pans can be washed and reused. There is the Reference card on the bottom of the analyzer.
No good Good
Close
Remove sample
Drying temperature
Analysis time
Drying rate
Result
Sample needs more than 0.1 g. Significant sample is indicated segment.
Range of proper sample
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
24 MS-70 / MX-50 / MF-50 / ML-50
6.3.6.3.6.3.6.3. Program NumberProgram NumberProgram NumberProgram Number The measurement conditions of all program numbers are set to the standard mode at the
factory. The analyzer can store and recall proper individual settings for each sample with the program number (PROG No.).
MS-70 / MX-50 20 sets PROG 1 to 20 MF-50 10 sets PROG 1 to 10 ML-50 05 sets PROG 1 to 5 The same measurement program is stored in all program
numbers with factory settings. Analysis mode.....................Standard mode Heating pattern....................Standard drying Caution If the data memory function is active, the data memory number (MEM) is
displayed, in place of the program number (PROG).
6.3.1.6.3.1.6.3.1.6.3.1. Storing a Storing a Storing a Storing a MeasurementMeasurementMeasurementMeasurement Program Program Program Program to a Program Number to a Program Number to a Program Number to a Program Number Displaying or recalling a PROG number, a measurement program can be renewed. 1. Press and hold the PROGRAM key in gram display. 2. Press the ↓ or ↑ key to select a program number 3. Press the ENTER key to use the selected number. 4. Press the SELECT key to edit the parameters. 5. Edit parameters of a measurement program.
Refer to "7.Selection of Measurement Method" 6. Press the following key to return to the gram display. ENTER key..... Stores parameters to the selected number. RESET key ..... Cancels the operation and returns to the
weighing mode.
6.3.2.6.3.2.6.3.2.6.3.2. Recalling a Recalling a Recalling a Recalling a MeasurementMeasurementMeasurementMeasurement Program Program Program Program with a Pro with a Pro with a Pro with a Program Numbergram Numbergram Numbergram Number Stored measurement programs can be recalled with a program number. 1. Press and hold the PROGRAM key in the gram display. 2. Press the ↓ or ↑ key to select a program number 3. Press the ENTER key to use the selected number.
Program number
Memory number
PROG "1 0.000g
PROG "2
end PROG 2 0.000g
Program number
Edit measurement program.
end PROG 2 0.000g
PROG "1 0.000g
PROG "2
end PROG "2 0.000g
Program number
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 25
7.7.7.7. MeasurementMeasurementMeasurementMeasurement Programs Programs Programs Programs
7.1.7.1.7.1.7.1. List of List of List of List of MeasurementMeasurementMeasurementMeasurement Programs Programs Programs Programs There are five analysis modes.
Measurement Programs Drying Program
Heating pattern Analyzing mode to Finish Measurement
MS-70, MX-50, MF-50 ML-50
Drying Temperature
Standard Mode 5td
Standard dryingRamp drying Step drying
Standard drying
Quick Mode quc
Measurement condition is automatically selected by ACCURACY and minimum value of % display. When drying rate is less than preset termination value, measurement is completed automatically.
Quick drying
Automatic Mode U-a
When drying rate is less than preset termination value, measurement is completed automatically.
Timer Mode U-t
Sample is dried for a preset time. 1min. to 480min.
Manual Mode U-m
Measurement is completed by the STOP key. Max. 480 min.
Standard dryingRamp drying Step drying
Standard drying
50°C to
200°C
Drying rate: Change of moisture content per one minute [%/min]
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
26 MS-70 / MX-50 / MF-50 / ML-50
7.1.1.7.1.1.7.1.1.7.1.1. ACCURACY of the Standard Mode and Quick ModeACCURACY of the Standard Mode and Quick ModeACCURACY of the Standard Mode and Quick ModeACCURACY of the Standard Mode and Quick Mode ACCURACY of measurement can be set either HI , MID. or LO . The sample quantity is automatically selected by ACCURACY. The termination value of the analyzing mode is automatically selected by ACCURACY
and minimum scale value of % display. The analyzing mode is the program to finish the measurement when a change of
moisture content per one minute becomes smaller than a preset termination value. The settings are as follows: Specify an ACCURACY. Drying rate: Change of moisture content per one minute [%/min] Standard ModeStandard ModeStandard ModeStandard Mode
ACCURACY Model Minimum scale HI MID. LO
0.001 % 0.01 %/min 0.02 %/min 0.05 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/minMS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/minMX-50 0.1 % 0.10 %/min 0.20 %/min 0.05 % 0.05 %/min 0.10 %/min 0.1 % MF-50 1 % 0.1 %
Preset Termination value
ML-50 1 %
0.10 %/min 0.20 %/min 0.50 %/min
Sample quantity 10 g 5 g 1 g Use Precise result ↔ Quick measurement
Quick ModeQuick ModeQuick ModeQuick Mode
ACCURACY Model Minimum scale HI MID. LO
0.001 % 0.02 %/min 0.05 %/min 0.05 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/minMS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/minMX-50 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.05 % 0.1 % MF-50 1 %
0.10 %/min 0.20 %/min 0.50 %/min
0.1 %
Preset Termination value
ML-50 1 % 0.20 %/min 0.50 %/min 1.00 %/min
Sample quantity 5 g 2 g 1 g Use Precise result ↔ Quick measurement
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 27
7.1.2.7.1.2.7.1.2.7.1.2. Analyzing mode of the Automatic ModeAnalyzing mode of the Automatic ModeAnalyzing mode of the Automatic ModeAnalyzing mode of the Automatic Mode When the change of moisture content per one minute is less than the preset value,
the measurement is automatically completed. Range Preset Termination
value to complete measurement MS-70 MX-50 MF-50 ML-50
2.00 %/min 1.00 %/min 0.50 %/min Available Available 0.20 %/min Available range range 0.10 %/min Available range (Factory setting) (Factory setting) 0.05 %/min range (Factory setting) 0.02 %/min (Factory setting) 0.01 %/min 0.005 %/min 0.002 %/min 0.001 %/min Unavailable item
7.1.3.7.1.3.7.1.3.7.1.3. Analyzing mode of the Timer ModeAnalyzing mode of the Timer ModeAnalyzing mode of the Timer ModeAnalyzing mode of the Timer Mode Sample is dried for a preset analysis time. Drying Time 1 minute to 480 minutes. Setting interval : 1 minute during 1 minute to 60 minutes. 5 minutes during 60 minutes to 480 minutes. Factory setting: 10 minutes.
7.1.4.7.1.4.7.1.4.7.1.4. Drying Program (Drying Program (Drying Program (Drying Program (Heating Pattern and Drying Termperature)Heating Pattern and Drying Termperature)Heating Pattern and Drying Termperature)Heating Pattern and Drying Termperature)
Heating patternHeating patternHeating patternHeating pattern
Standard Mode, Automatic Mode,
Manual Mode Timer Mode Quick Mode
Standard drying
Ramp drying
Step drying
Quick drying
"Temperature 1" can be set higher than "temperature 2" in step drying. ML-50, can select standard drying and quick drying only, does not display heating pattern.
Temperature TemperatureTime
Temperature Temperature
Time1Time250°C50°C Time
Stage 1
Temperature1
Time
Temperature2
Stage 2
Temperature1Time1
Temperature2
Stage 1 Stage 2Time2
Temperature
200°C 3 min.
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
28 MS-70 / MX-50 / MF-50 / ML-50
How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Select quick mode (Extracts)How to Select quick mode (Extracts) Press the SELECT key during the gram display. Display quc using the ↓ or ↑ key. ML-50 is does not display heating pattern. How to Select a How to Select a How to Select a How to Select a Heating patternHeating patternHeating patternHeating pattern without quick mode (Extracts) without quick mode (Extracts) without quick mode (Extracts) without quick mode (Extracts) Press the SELECT key to select temperature
during the gram display. Press the PROGRAM key to select a symbol of
heating pattern are displayed in order. ML-50 can select standard drying and quick drying only. Display examples during settingsDisplay examples during settingsDisplay examples during settingsDisplay examples during settings
DDDDrying Temperature at the Sample Panrying Temperature at the Sample Panrying Temperature at the Sample Panrying Temperature at the Sample Pan Drying temperature range.................50°C to 200°C, 1°C interval. When a measurement is started and one hour passes, the maximum temperature is
automatically regulated to 160°C for safety.
7.1.5.7.1.5.7.1.5.7.1.5. Measurement Measurement Measurement Measurement UnitUnitUnitUnit Unit Formula Display
Moisture content is based on wet sample mass *1
W - D W
% MOIST D /W/D
Moisture content (Atro) is based on dried sample mass
W - D D
% MOIST D /D
Dry content D W % RATIO D/W/D
Ratio *2 W D % RATIO W/D
Gram value g W: Wet sample mass D: Dried sample mass *1: Factory settings *2: When result reaches to 999%, measurement is stopped.
x 100 x 100 x 100 x 100
Heating pattern Time Time2
Time1
Temperature Temperature2
Temperature
Heating pattern
Heating patternTemperature
"q"u"c
Quick mode
Heating pattern
"1"0"5°C
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 29
7.2.7.2.7.2.7.2. Procedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to Store a Measurement ProgramProcedures to Store a Measurement Program
7.2.1.7.2.1.7.2.1.7.2.1. Standard DryingStandard DryingStandard DryingStandard Drying This explanation uses the following parameters and sample displays of MX-50. Common Items Program number.......................................... 2 ( PROG 2 ) Drying program Heating pattern ....................Standard drying ( ) Drying temperature..............160°C Measurement unit ...............................................Moisture content Minimum scale value during measurement.. 0.01 % Minimum scale value of gram display .......... 0.001 g Sample quantity ........................................... Approximately 5 g Data memory function.................................. Not used Items for Standard Mode Analysis mode ............................................. Standard mode (Symbol: 5td ) ACCURACY................................................. MID. Analyzing mode to finish a measurement .........Automatic setting by ACCURACY Items for Automatic Mode Analysis mode ............................................. Automatic mode (Symbol: U-a ) Analyzing mode to finish measurement ............0.05 %/min Items for Timer Mode Analysis mode ............................................. Timer mode (Symbol: U-t ) Analyzing mode to finish measurement ............10 minutes Items for Manual Mode Analysis mode ............................................. Manual mode (Symbol: U-m )
1. Display the gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement Program 2. Press the PROGRAM key and press the ↓ or
↑ key to select a program number. 3. Press the ENTER key to use the number. 4. The analyzer displays end and returns to the
weighing mode. Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the program number (PROG).
Drying temperature Time
Drying temperature
Standard Mode, Automatic Mode, Manual Mode Timer Mode
% MOISTD /W/D
PROG 1 0.000g
PROG "2
end PROG 2 0.000g
Program number
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
30 MS-70 / MX-50 / MF-50 / ML-50
Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELECT key to select a symbol, it will
blink. Select an analysis mode using the ↓ or ↑ key. If standard mode is used, select 5td . If automatic mode is used, select U-a . If timer mode is used, select U-t . If manual mode is used, select U-m .
Set ACCURACY for the Standard ModeSet ACCURACY for the Standard ModeSet ACCURACY for the Standard ModeSet ACCURACY for the Standard Mode 6. Select ACCURACY with the SELECT key.
Select MID. with the ↓ or ↑ key. When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating PatterSet the Heating PatterSet the Heating PatterSet the Heating Patternnnn 7. Select drying temperature with the SELECT key.
Select standard drying of the heating pattern with the PROGRAM key.
Set the Drying TemperatureSet the Drying TemperatureSet the Drying TemperatureSet the Drying Temperature 8. Set 160°C with the ↓ or ↑ key.
Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for AutomaticAutomaticAutomaticAutomatic MMMModeodeodeode 9. Select the preset termination value with the
SELECT key. Select 0.05 [%/min] with the ↓ or ↑ key.
Set the Analysis Time for the Set the Analysis Time for the Set the Analysis Time for the Set the Analysis Time for the TimerTimerTimerTimer MMMModeodeodeode 10. Select time with the SELECT key.
Select 10.0 [min] with the ↓ or ↑ key.
Set the Set the Set the Set the MeasurementMeasurementMeasurementMeasurement Unit Unit Unit Unit 11. Select a measurement unit with the SELECT
key. Select a moisture content (based on a wet sample ) with the ↓ or ↑ key.
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display 12. Select % display with the SELECT key.
Select 0.01 [%] with the ↓ or ↑ key.
0.000g "1"0."0 min
Analysis Time
ACCURACY MID.
ACCURACY
"0."0"5 %/min Preset termination value
Standard mode
Automatic mode
Timer mode
Standard mode Manual mode
"U"-"a 0.000g
Analysis mode
0.000g "1"6"0°C
0"% "M"O"I"S"T "/"W
Measurement unit
"0".0"1%
Minimum scale value of %
display
0.000g Standard drying
Drying temperature
Automatic mode Timer mode Manual mode
"1"0"5°C
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 31
Set Set Set Set the Minimum Scale Value of the Gram Displaythe Minimum Scale Value of the Gram Displaythe Minimum Scale Value of the Gram Displaythe Minimum Scale Value of the Gram Display 13. Select gram display with the SELECT key.
Select 0.001 [g] with the ↓ or ↑ key.
Store the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the Operation 14. Press the ENTER key to store the new parameters
for the measurement program to program number 2. Pressing the key, the weighing mode is automatically displayed. When PROG 2 is recalled, the settings can be used.
To cancel the new parameters and return to the
weighing mode, press the RESET key.
0.000g
Gram display ofweighing mode
"0."0"0"1g Minimum scalevalue of gramdisplay
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
32 MS-70 / MX-50 / MF-50 / ML-50
7.2.2.7.2.2.7.2.2.7.2.2. Ramp DryingRamp DryingRamp DryingRamp Drying
Standard Standard Standard Standard Mode, Automatic Mode or Mode, Automatic Mode or Mode, Automatic Mode or Mode, Automatic Mode or Manual Manual Manual Manual MMMModeodeodeode Refer to page 34 for Timer Mode
This explanation uses the following parameters and sample displays of MX-50. Common Items Program number.......................................... 3 ( PROG 3 ) Drying program Heating pattern ....................Ramp drying ( ) Final drying temperature.....160°C Ramp time............................5.0 minutes Measurement Unit...............................................Moisture content Minimum scale value during measurement.. 0.01 % Minimum scale value of gram display .......... 0.001 g Sample quantity ........................................... Approximately 5 g Data memory function.................................. Not used Items for Standard Mode Analysis mode ............................................. Standard mode (Symbol: 5td ) ACCURACY................................................. MID. Analyzing mode to finish measurement ............Automatic setting by ACCURACY Items for Automatic Mode Analysis mode ............................................. Automatic mode (Symbol: U-a ) Analyzing mode to finish measurement ............0.05 %/min Items for Manual Mode Analysis mode ............................................. Manual mode (Symbol: U-m )
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement Program 2. Press the PROGRAM key and press the ↓ or
↑ key to select a program number. 3. Press the ENTER key to use the number. 4. The analyzer displays end and returns to the
weighing mode. Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the program number (PROG).
PROG 1 0.000g
PROG "3
end PROG 3 0.000g
Program number
% MOISTD /W/D
50°CRamp time
Final drying temperature
(ML-50 can not select ramp drying)
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 33
Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELECT key to select a symbol, it will
blink. Select an analysis mode using the ↓ or ↑ key.
If standard mode is used, select 5td . If automatic mode is used, select U-a . If manual mode is used, select U-m .
Set ACCURACY for Standard ModeSet ACCURACY for Standard ModeSet ACCURACY for Standard ModeSet ACCURACY for Standard Mode 6. Select ACCURACY with the SELECT key.
Select MID. with the ↓ or ↑ key. When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating PatternSet the Heating PatternSet the Heating PatternSet the Heating Pattern
7. Select drying temperature with the SELECT key. Select ramp drying of heating pattern with the PROGRAM key.
Set the Final Drying TemperatureSet the Final Drying TemperatureSet the Final Drying TemperatureSet the Final Drying Temperature 8. Set 160°C with the ↓ or ↑ key.
Set the Ramp TimeSet the Ramp TimeSet the Ramp TimeSet the Ramp Time 9. Set 5.0 minutes with the ↓ or ↑ key.
Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for AutomaticAutomaticAutomaticAutomatic MMMModeodeodeode 10. Select preset termination value with the SELECT
key. Select 0.05 [%/min] with the ↓ or ↑ key.
Set the Set the Set the Set the Measurement Measurement Measurement Measurement UnitUnitUnitUnit 11. Select measurement unit with the SELECT key.
Select a moisture content (based on a wet sample ) with the ↓ or ↑ key.
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display 12. Select % display with the SELECT key.
Select 0.01 [%] with the ↓ or ↑ key.
Set the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram Display 13. Select gram display with the SELECT key.
Select 0.001 [g] with the ↓ or ↑ key.
Automatic mode
0"% "M"O"I"S"T "/"W
"0."0"1%
"0."0"0"1g
0.000g "1"6"0°C
ACCURACY MID.
"0."0"5 %/min
0.000g 5.0 min
0.000g ""1"0"5°C
Measurement unit
Minimum scale value of % display
Minimum scale value of gram display
Final drying temperature
Ramp time
Ramp drying
ACCURACY
Preset termination value
Standard mode Manual mode
"U"-"a 0.000g
Standard mode
Automatic mode Manual mode
Analysis mode
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
34 MS-70 / MX-50 / MF-50 / ML-50
Store Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the OperationStore Parameters and Finish the Operation 14. Press the ENTER key to store the new parameters of
the measurement program to program number 3. Pressing the key, the weighing mode is automatically displayed. When PROG 3 is recalled, the settings can be used.
To cancel the new parameters and return to the
weighing mode, press the RESET key.
Timer ModeTimer ModeTimer ModeTimer Mode Refer to page 32 for Standard Mode, Automatic Mode or Manual Mode
This explanation uses the following parameters and sample displays of MX-50. Common Items Program number.......................................... 4 ( PROG 4 ) Analysis mode ............................................. Timer mode (Symbol: U-t ) Drying program Heating pattern ....................Ramp drying ( ) Final drying temperature.....160°C Ramp time............................5.0 minutes Drying time...........................10.0 minutes Measurement unit ...............................................Moisture content Minimum scale value during measurement.. 0.01 % Minimum scale value of gram display .......... 0.001 g Sample quantity ........................................... Approximately 5 g Data memory function.................................. Not used
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement Program 2. Press the PROGRAM key and press the ↓ or
↑ key to select a program number. 3. Press the ENTER key to use the number. 4. The analyzer displays end and returns to the
weighing mode. Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the program number (PROG).
% MOISTD /W/D
PROG 1 0.000g
PROG "4
end PROG 4 0.000g
Program number
Final drying temperature50°C Ramp time
Drying time
0.000g 160°C
Gram display of weighing mode
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 35
Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELECT key to select a symbol, it
will blink. Select U-t of the timer mode with the ↓ or ↑ key. (Press these keys to select it)
Set the Heating PatternSet the Heating PatternSet the Heating PatternSet the Heating Pattern 6. Select drying temperature with the SELECT key.
Select ramp drying of the heating pattern with the PROGRAM key.
Set the Final Drying TemperatureSet the Final Drying TemperatureSet the Final Drying TemperatureSet the Final Drying Temperature 7. Set 160°C with the ↓ or ↑ key.
Set the Ramp TimeSet the Ramp TimeSet the Ramp TimeSet the Ramp Time 8. Set 5.0 minutes with the ↓ or ↑ key.
Set the Analysis TimeSet the Analysis TimeSet the Analysis TimeSet the Analysis Time 9. Set 10.0 minutes with the ↓ or ↑ key.
Set a Measurement UnitSet a Measurement UnitSet a Measurement UnitSet a Measurement Unit 10. Select a measurement unit with the SELECT
key. Select moisture content (based on a wet sample ) with the ↓ or ↑ key.
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display 11. Select the % display with the SELECT key.
Select 0.01 [%] with the ↓ or ↑ key.
Set the Minimum Scale Value of the GSet the Minimum Scale Value of the GSet the Minimum Scale Value of the GSet the Minimum Scale Value of the Gram Displayram Displayram Displayram Display 12. Select gram display with the SELECT key.
Select 0.001 [g] with the ↓ or ↑ key.
Store the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the Operation 13. Press the ENTER key to store the new parameters
of the measurement program to program number 4. Pressing the key, the weighing mode is automatically displayed. When PROG 4 is recalled, the settings can be used.
To cancel the new parameters and return to the weighing mode, press the RESET key.
"U"-"t 0.000g
Analysis mode
1
0.000g "1"0"5°C
Ramp Time
"5."0 min
2
0"% "M"O"I"S"T "/"W
Measurement unit
"0."0"1% Minimum scale value of % display
"0."0"0"1g Minimum scale value of gram display
0.000g 160°C 10.0 min
"1"0."0 min
Analysis time
0.000g "1"6"0°C
Final drying temperature
Ramp drying
Gram display of weighing mode
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
36 MS-70 / MX-50 / MF-50 / ML-50
7.2.3.7.2.3.7.2.3.7.2.3. Step DryingStep DryingStep DryingStep Drying
Standard Mode, Automatic Mode or Standard Mode, Automatic Mode or Standard Mode, Automatic Mode or Standard Mode, Automatic Mode or Manual Manual Manual Manual MMMModeodeodeode Refer to page 38 for Timer Mode
This explanation uses the following parameters and sample displays of MX-50. "Temperature 1" can be set higher than "temperature 2" in step drying. Common Items Program number.......................................... 5 ( PROG 5 ) Drying program Heating pattern ....................Step drying ( ) Drying temperature 1 ..........120°C Drying temperature 2 ..........160°C Time......................................5.0 minutes Measurement Unit...............................................Moisture content Minimum scale value during measurement.. 0.01 % Minimum scale value of gram display .......... 0.001 g Sample quantity ........................................... Approximately 5 g Data memory function.................................. Not used Items for Standard Mode Analysis mode ............................................. Standard mode (Symbol: 5td ) ACCURACY................................................. MID. Analyzing mode to finish measurement ............Automatic setting by ACCURACY Items for Automatic Mode Analysis mode ............................................. Automatic mode (Symbol: U-a ) Analyzing mode to finish measurement ............0.05 %/min Items for Manual Mode Analysis mode ............................................. Manual mode (Symbol: U-m )
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement ProgramSelect a Program Number to Edit the Measurement Program 2. Press the PROGRAM key and press the ↓ or
↑ key to select a program number. 3. Press the ENTER key to use the number. 4. The analyzer displays end and returns to the
weighing mode. Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the program number (PROG).
% MOISTD /W/D
PROG 1 0.000g
PROG "5
end PROG 5 0.000g
Program number
Time
Drying temperature 2 Drying temperature 1
Stage 1 Stage 2
(ML-50 can not select step drying)
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
MS-70 / MX-50 / MF-50 / ML-50 37
Select an Analysis ModeSelect an Analysis ModeSelect an Analysis ModeSelect an Analysis Mode 5. Press the SELECT key to blink a symbol, it will
blink. Select an analysis mode using the ↓ or ↑ key. If standard mode is used, select 5td . If automatic mode is used, select U-a . If manual mode is used, select U-m .
Set ACCURACY for the Standard ModeSet ACCURACY for the Standard ModeSet ACCURACY for the Standard ModeSet ACCURACY for the Standard Mode 6. Select ACCURACY with the SELECT key.
Select MID. with the ↓ or ↑ key. When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating PatternSet the Heating PatternSet the Heating PatternSet the Heating Pattern
7. Select drying temperature 1 with the SELECT key. Select step drying of the heating pattern with the PROGRAM key.
Set Drying Temperature 1Set Drying Temperature 1Set Drying Temperature 1Set Drying Temperature 1 8. Set 120°C to drying temperature 1 with
the ↓ or ↑ key.
Set the TimeSet the TimeSet the TimeSet the Time 9. Select time with the SELECT key.
Set 5.0 minutes with the ↓ or ↑ key.
Set Drying Temperature 2Set Drying Temperature 2Set Drying Temperature 2Set Drying Temperature 2 10. Select drying Temperature 2 with
the SELECT key. Set 160°C with the ↓ or ↑ key.
Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for Set the Preset Termination Value for AutomaticAutomaticAutomaticAutomatic MMMModeodeodeode 11. Select a preset termination value with the
SELECT key. Select 0.05 [%/min] with the ↓ or ↑ key.
Set Set Set Set Measurement Measurement Measurement Measurement UnitUnitUnitUnit 12. Select measurement unit with the SELECT key.
Select moisture content (based on a wet sample ) with the ↓ or ↑ key.
ACCURACY MID.
Standard mode
"0."0"5 %/min
Automatic mode
"U"-"a 0.000g
0"% "M"O"I"S"T "/"W
"5."0 min
2 "1"6"0°C
"1"0"5°C
1 "1"2"0°C
Standard mode Manual mode
Automatic modeManual mode
ACCURACY
Preset termination value
Analysis
Measurement unit
Drying temperature 2
Drying temperature 1
Step drying
Time
Purchase A&D Weighing Balaces @ www.balances.com
Learn more at http://scaleman.com/balances/moisture-balances.html
38 MS-70 / MX-50 / MF-50 / ML-50
Set the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % DisplaySet the Minimum Scale Value of the % Display 13. Select the % display with the SELECT key.
Select 0.01 [%] with the ↓ or ↑ key.
Set the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram DisplaySet the Minimum Scale Value of the Gram Display 14. Select the gram display with the SELECT key.
Select 0.001 [g] with the ↓ or ↑ key.
Store the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the OperationStore the Parameters and Finish the Operation 15. Press the ENTER key to store the new parameters of
the measurement program to program number 3. Pressing the key, the weighing mode is automatically displayed. When PROG 3 is recalled, the settings can be used.
To cancel new parameters and return to weighing
mode, press the RESET key.
Timer ModeTimer ModeTimer ModeTimer Mode Refer to page 36 for Standard Mode, Automatic Mode or Manual Mode
This explanation uses the following parameters and sample displays of MX-50. "Temperature 1" can be set higher than "temperature 2" in step drying. Common Items Program number.......................................... 6 ( PROG 6 ) Analysis mode ............................................. Timer mode (Symbol: U-t ) Drying program Heating pattern ....................Step drying ( ) Drying temperature 1 ..........160°C Drying temperature 2 ..........120°C Time 1 .....