NOx Monitors for SCR Applications - Cemtek · PDF fileNOx Monitors for SCR Applications ......

Post on 06-Mar-2018

226 views 1 download

transcript

NOx

Monitors for

SCR

Applications

a technical solution to

meet every need…

CEMTEK Instruments Inc.

2013 S. Wood Ave.

Linden, NJ 07036

800-400-0201

www.cemtekinstruments.com

Todays regulations are placing demands on all types

of Combustion sets for tighter control of their

emissions.

No one solution can solve all application demands but

there are several alternate measurements that can

help improve their chances without costly upgrades.

• While manufacturers are striving to find the ultimate

solution for the next generation of turbines, burners

engines, they are still reliant on a combination of

combustion technology & post combustion

technology to achieve the lowest possible

emissions.

• Tighter NOx & CO control, through combustion

improvements and better catalyst efficiency is a

target for many.

Regulatory Constraints.

NOx Monitoring

Emissions variables

NOx can be controlled by various methods and is by

far the largest variable.

CO being a product of combustion conditions is

dependent on the combustor and flame temperature.

By using continuous real time sensors to measure

these variables improvements in their control can be

achieved

g

NOx / CO Tradeoff

SCR NOx Monitoring

How does NOx monitoring help operations?

Better control for transient conditions?

Better control under low load conditions.

Indications of poor combustion.

Catalyst deterioration.

Improve efficiency SCR NH3 injection during above

conditions.

Accurate NOx numbers from sensor with fast

response time, < 10seconds?

SCR NOx Monitoring

How do utilities run their SCR sets?

What are the main criteria for best operating

performance?

Accurate NOx numbers?

Reliable response time?

How are you obtaining this information for control

and optimization?

Continuous monitoring with traditional NEMS NOx

monitoring?

Periodic testing with portable/test trailer?

Close Coupled NOx Probe

SCR NOx Monitoring

SCR NOx Monitoring

Extractive or Dilution NOx measurement:

Response time based on sample line run

Installation can be costly depending on system

location. May be located in CEMS shelter or

designed as stand alone cabinet

Maintenance can be likened to a CEMS since in

effect that is what it is.

CEMTEK NOx – O2 Analyzers

Rack mount Analyzers for standard NEMS applications

Single NOx Analyzer

Dual NOx Analyzer

Single O2 Analyzer

Combination NOx-O2 Analyzer

NOx is typically measured using a

Chemiluminescence sensor

Sensitivity: ppb easily measureable

Oxygen is measured using a Zirconia planar

sensor

CEMTEK NOx – O2 Analyzers

FULLY EXTRACTIVE (DRY BASIS)

DILUTION EXTRACTIVE (WET BASIS)

OTHERS:

•FULLY EXTRACTIVE (HOT/WET)

•IN-SITU (CROSS STACK or PROBE)

•EX-SITU (CLOSE COUPLED EXTRACTIVE)

NEMS TYPES

SCR NOx Monitoring

Sample Gas Dried To A Dew-point Of +2°C

Excellent Performance In Low

Concentrations

Good Performance In High Concentrations

Requires Preventative Maintenance

FULLY EXTRACTIVE - Dry Basis

Standard Extractive NEMS

SAMPLE

CONDITIONING BAY

Standard Extractive NEMS

Extractive Flow Schematic

NOx

O2

Cooler

Probe

Sample Line

CO

Pump

Inst. Air

Cal

Gas

Standard Extractive NEMS

Standard Dilution NEMS

Dilution measurement measures a gas

stream’s properties by extracting a

certain amount of gas sample from the

stack flow then diluting the sample with

instrument grade ambient air to some

exact proportion say 100:1. This makes

for a cleaner gas stream through the

analyzer’s pneumatics eliminating the

need for expensive sample conditioning

and producing less problems which

cause analyzer downtime.

Dilution System Flow Diagram

1

2

3

4 5

6

7

NOx

O2

Probe

Flow Panel

Dilution Air Cleanup

Panel

Filters, Scrubbers, Dryers

1-Dilution Air

2-Diluted Sample

3-Purge

4-Bypass Air

5-Pressure Comp

6-Vacuum

7-Cal Gas

CEMTEK NOx – O2 Analyzers

CEMTEK 8000 NOx- Monitor

Chemiluminescense:

Is recognized world wide as best NOx measurement technique

Dilution:

Well proven and accepted in EPA applications

Best technique for particulate laden flue gas applications.

The 8000 Combines proven technology for Coal Fired

Applications into an unconventional analyzer

CEMTEK 8000 NOx - O2 Monitor

This analyzer is specifically

designed for use in combustion

process applications:

SCR inlet/outlet monitoring

applications.

for low NOx burner tuning,

Consisting of 3 Components

1. The Analyzer Controller which

houses the electronics and the

NO Sensor.

2. The Probe which contains the

dilution block and optional

Oxygen Sensor along with the

NO2 converter if required.

3. The self-limiting heated

interconnect umbilical which

contains the sample line,

calibration gas line,

thermocouple cables and

power cables.

Difference between PMT & Photo Diode

measurement technique

Most NOx measurements are done using a Photo Multiplier Tube

(PMT).

PMT configuration is complex, more expensive and susceptible to

temperature and moisture changes.

Our analyzer uses a Photo Diode for NOx detection.

Its means of detection consists of a reaction cell and detector, the

photo diode and makes for a much simpler configuration.

Is less susceptible to ambient environmental changes.

Advantages of Photo Diode measurement

technique

Dramatically smaller

Eliminates high voltage power supply

All solid state, much more rugged

Eliminates need for large TE coolers and drivers

Can be close-coupled to reaction, no optics other

than a window needed

Not affected by temperature swings

Linear, accurate, repeatable

Minimal sample conditioning

No air conditioned instrument shelters

The NOx Probe

Highly integrated pneumatics combines analyzer and low flow /

low dilution probe into a single instrument.

Heated to ~350 C to avoid condensation of acid gases before

dilution, also provides orifice temperature control

Heated dual-stage filtration and low flow avoids pluggage.

Through Probe Calibration for full system integrity.

In Stack Dilution block/probe design

The NOx Probe

Fully immersed, temperature

controlled sample orifice

Located at tip of probe for very fast

response and virtually no time lag

Less than 10 seconds combined

Solid state detector for fast accurate

measurement in harsh conditions.

8000 Flow Schematic

NOx Chemiluminescense Sensor

Chemiluminescense NOx

Sensor mounted on main

PCB, input from gas sample

and O3 generator into the

reaction cell, controlled by

exhaust eductor and air

pump

Filter Assembly

This component consists of

the ammonia scrubber and

in-line filter components. The

ammonia scrubber removes

NH3 from the sample gas. As

the in-line filter conditions the

sample gas prior to the NOx

cell. The assembly is located

before the NOx cell in the

controller.

NOx Controller Interface

NOx Controller Interface

Multi-unit Installation

NOx- Monitor Installation

NOx- Monitor Installation

Monitor installation requirements:

All connections between the Probe and the Controller are via the 1.5” Self

Limiting Heated Sample Line, typical length 6-20ft.

Customer Interface for Power, Signal & Pneumatics:

One ¼” Swagelok port for calibration gas, concentration is dependent

on range, xxppm NO in N2

one ¼ Swagelok port for instrument air (2.5 SCFM @ 70 psi). Clean dry

air @ -40F dew point, for dilution air & zero gas.

Air Clean up Panel is available as an option when compressed air is

unavailable

Signals to customer control are isolated 4-20 mA analog,

General Fault alarm – analyzer, System in Calibration alarm- powered

24v dc.

Power consumption of 5A (max on startup) at 120VAC.

SCR PERFORMANCE OPTIMIZATION

Probe install:

13th October 2010, approximately 1/2day work.

1400 MW Coal Fired SCR Application:

Eastern Bituminous High Sulfur

SO2 levels 2000-3000ppm

NH3 Slip 0-5ppm

SCR PERFORMANCE OPTIMIZATION

8000 NOx Probe tracking Dilution Extractive System on Split Outlet of SCR

SCR PERFORMANCE OPTIMIZATION

Expanded graph shows 8000 leading Dilution analyzer by approximately 90sec

SCR Control using CEMTEK 8000

Data taken from SCR NOx-O2 Monitors on a 600 MW power generator, using PRB coal and an SCR with

direct ammonia vaporization. Inlet is located just before ammonia injection grid. Outlet is located just after

SCR and before electrostatic precipitator.

SCR Monitor

Inlet NOx (red)

SCR Monitor

Outlet NOx (blue)

Ammonia Injection (black)

CEMS/Stack

NOx (green)

Control using Inlet

and Outlet Monitors

Control using

CEMS NOx

Target NOx Output Level: about 30 PPM

0 5 10 15 20 25 30 35 40 45 50 55 60

1.2 minutes

2.7 minutes 15.6 minutes

While using the SCR NOx-O2 inlet and outlet monitors, the control system

was able to reduce NOx output to the control level in just over 1 minute

even while NOx input levels remained high. After the combustion excursion

had subsided, the ammonia was quickly returned to “normal” levels and the

NOx output was brought back to the target NOx level in under 3 minutes.

While using the CEMS/Stack NOx, the control system was NOT

able to reduce NOx output before the combustion excursion had

subsided. In fact, the ammonia valve was opened after the excur-

sion subsided and remained open too long, over-injecting ammonia

and pushing NOx well below the target level for over 15 minutes.

Combustion

Excursions

2 minute delay

Time (minutes)

3 to 4 PPM

450 PPM

300 PPM

100 PPM

CT Performance Optimization

SCR INLET

CT OUTLET

CT Trial for GE. 7FNB DLN 150MW

Testing a new Fast Response Close Coupled

NOx / O2 probe:

Comparative measurements to Utility Test

Trailer with conventional CEMS:

Prove accuracy, availability, speed of

response.

Usefulness of real-time continuous monitoring

for feed back.

CT Trial for GE. 7FB DLN 150MW

Testing of CEMTEK 8000 NOx/O2 Probe:

Outlet of 150 mw GE Turbine prior to HRSG.

Gas Temp- 1200F, Duct Pressure - 20”wc.

PROBE length 20”, with 10” extension.

Sample Line 12ft.

Response time < 10 sec

CT Trial for GE. Frame 7 150MW

Testing for CEMTEK 8000 NOx/O2 Probe:

During initial burn in ( 2 wks) daily calibration

checks to prove RA.

NOx values corrected to 15% O2 using

analyzer measurement.

Corrected to dry number using trailer test

data for moisture.

Calibration Drift logged every 24 hours.

NOx Raw Signal WET vs. Dry

CT Outlet Corrected to 15% O2

CT Trial for GE. Frame 7 150MW

Resultant average values between Test trailer

and Fast Response Close Coupled NOx / O2

probe:

Relative Accuracy shown:

Trailer @ 15% O2 = 15.86ppm

8000 Probe @ 15% = 15.4ppm

RA differential 0.46ppm

Response differential 60 seconds

Daily Cal Check Response

NOx Calibration Gas = 25.17ppm Nox Range = 0-50ppm

O2 Gas = 20.9% O2 Range = 0-25%

Sytem Responses Date Day NOx Zero NOx Span O2 Zero O2 Span

9/9/2011 1 0.67 26.09 0.03 20.82

9/10/2011 2 0.60 25.30 0.00 20.80

9/11/2011 3 0.90 26.70 0.00 20.80

9/12/2011 4 0.60 25.40 0.00 20.80

9/13/2011 5 0.40 25.30 0.00 20.80

9/14/2011 6 0.70 24.40 0.00 20.80

9/15/2011 7 0.20 24.90 0.00 20.80

9/16/2011 8 -0.10 24.40 0.00 20.90

9/17/2011 9 0.50 25.50 0.00 21.00

9/18/2011 10 0.20 25.80 0.00 20.70

9/19/2011 11 0.60 25.40 0.00 20.70

9/20/2011 12 0.50 25.00 0.00 20.60

9/21/2011 13 0.40 24.50 0.00 20.80

9/22/2011 14 0.30 24.30 0.00 20.80

9/23/2011 15 1.10 24.00 0.00 20.90

9/24/2011 16 0.30 23.50 0.00 20.80

9/25/2011 17 0.40 23.20 0.00 20.70

9/26/2011 18 0.20 23.50 0.00 20.70

Analyzer Zero Drift

Drift Calculations (Difference From Bottle)

Date Day NOx Zero NOx Span O2 Zero O2 Span

9/9/2011 1 0.67 26.09 0.03 -20.82

9/10/2011 2 0.60 25.30 0.00 -20.80

9/11/2011 3 0.90 26.70 0.00 -20.80

9/12/2011 4 0.60 25.40 0.00 -20.80

9/13/2011 5 0.40 25.30 0.00 -20.80

9/14/2011 6 0.70 24.40 0.00 -20.80

9/15/2011 7 0.20 24.90 0.00 -20.80

9/16/2011 8 -0.10 24.40 0.00 -20.90

9/17/2011 9 0.50 25.50 0.00 -21.00

9/18/2011 10 0.20 25.80 0.00 -20.70

9/19/2011 11 0.60 25.40 0.00 -20.70

9/20/2011 12 0.50 25.00 0.00 -20.60

9/21/2011 13 0.40 24.50 0.00 -20.80

9/22/2011 14 0.30 24.30 0.00 -20.80

9/23/2011 15 1.10 24.00 0.00 -20.90

9/24/2011 16 0.30 23.50 0.00 -20.80

9/25/2011 17 0.40 23.20 0.00 -20.70

9/26/2011 18 0.20 23.50 0.00 -20.70

Target Drift Rates Not to Exceed

NOx = 2.5 ppm

O2 = 0.5 %