Operating Instructions
Programat® EP 3000 G2
Valid as of
Software Version 6.0
2
3
Views of the Furnace, List of Parts 4 1. Introduction / Signs and Symbols 81.1 Preface1.2 Introduction1.3 NotesregardingtheOperatingInstructions1.4 Notesonthedifferentvoltageversions
2. Safety First 92.1 Indications2.2 Healthandsafetyinstructions3. Product Description 123.1 Components 3.2 Hazardousareasandsafetyequipment 3.3 Functionaldescription3.4 Accessories
4. Installation and Initial Start-Up 134.1 Unpackingandcheckingthecontents 4.2 Selectingthelocation4.3 Assembly4.4 Removingthefurnacehead4.5 Initialstart-up
5. Operation and Configuration 195.1 Introductiontotheoperation 5.2 Explanationofthekeyfunctions 5.3 Programstructure5.4 Adjustableparametersandpossiblevalueranges5.5 Settingsandinformation5.6 Explanationofthesymbolsonthedisplay5.7 Explanationofthespeakersignals
6. Practical Use 256.1 Switchingon/off6.2 Firingprograms 6.3 Pressprograms6.4 Otheroptionsandspecialfeaturesofthefurnace7. Maintenance, Cleaning and Diagnosis 297.1 Monitoringandmaintenance 7.2 Cleaning 7.3 Temperaturecalibration7.4 Servicenote7.5 Stand-by7.6 Replacingthepressplunger
8. What if... 328.1 Errormessages8.2 Technicalmalfunctions8.3 Repair8.4 Loadfactorysettings
9. Product Specifications 359.1 Deliveryform 9.2 Technicaldata 9.3 Acceptableoperatingconditions9.4 Acceptabletransportationandstorageconditions
10. Appendix 3610.1 Programtable10.2 Menustructure
Table of Contents
4
List of parts
1 Sealingsurface
2 Furnaceheadsealingring
3 Insulation
4 Thermocouple
5 Firingplate2
6 Display
7 Frameplate
8 QTKheatingmuffle
9 Housingbase
10 Keypad(membrane-sealed)
11 On/Offswitch
12 Heatingelementfuse
13 Vacuumpumpfuse
15 Fuseholder
16 Powercord
17 Powersocket
18 Vacuumpumpsocket
19 Ratingplate
20 Screwforfurnaceheadcover
21 Vacuumhoseconnection
22 Headinsulation
23 Rubberfeet
25 Housingfurnacehead
26 Thermocoupleplug
27 Plugfuse
28 Heaterplug
29 Heaterplugsocket
30 Thermocoupleplugsocket
32 Leafspring
33 Airvents(base)
34 Coolingtray
35 Screwforcoolingtray
36 Hood
37 Knurledscrewforhood
38 Airventsfurnacehead
39 Airventsrearpanel
40 Warnings
41 Furnaceheadmountingmark
42 Furnacebasemountingmark
43 Furnaceheadmounting
44 Quartz-glasstube
46 Vacuumhose
47 Siliconewasher
48 Firingplateholder
49 Thermocouplecable
50 Connectingrodaxis
53 USBinterface
54 Plug-inconsole
56 Coverforpressdrive
58 Furnacehead,compl.
59 Pressplunger120
60 Pressdriveplug
61 Presselectronics
62 Coverforpresselectronics
63 Fan
64 Splittapersocketforpressplunger
65 Terminalscrewforpressplunger
66 Pressdrivecable
67 Pressdriveplugsocket
PleasenotethatthelistofpartsappliestotheentireOperatingInstructions.Thesepartsandtheirnumbersareoftenreferredtoinlaterchapters.
5
List of parts
7
1
2
3
9
5
8
40
25
38
56
34
6
10 2335
36
58
20
63
33
6
32 43
59
22
64
62
65
22
6163
11
36
37
39
12 13 53 19
12
11
1315
19
53
21
1716
18
46
49 2666 30 60 67 29 27 28 11
7
Control unit:
70Programkey
71ESCkey
72ENTERkey
73STARTkey
74StartLED
75STOPkey
76Pluskey
77Minuskey
78Settings/Information
79Cursorkeyright
80Cursorkeyleft
81Stand-bytemperature
82Closingtime
83Temperatureincrease
84Holdingtemperature
85Holdingtime
86Vacuumon
87Vacuumoff
88Long-termcooling
89PowerSavingkey
90Openfurnacehead
91Closefurnacehead
92Numerickeys
93Firing/Pressing
94Homekey
8
59
4
44
3
50
54
41
42
100 Programatfiringtray
101 MetalpinA
102 MetalpinB
103MetalpinC
115 Coolinggrid(complete)110 USBdatacable
100
115
101103102
120 AutomaticTemperatureCheckingSet2–ATK2
120110
78
93
94
81
82
85
86
88
8487
89
83
79 80 70 77 76 72 71
73
74
75
91
92
90
8
1. Introduction / Signs and Symbols
1.1 Preface
Dear Customer
ThankyouforhavingpurchasedtheProgramatEP3000/G2.Itisastate-of-the-artfurnacefordentalapplications.
Thefurnacehasbeendesignedaccordingtothelatestindustrystandards.Inappropriateusemaydamagetheequipmentandbeharmfultopersonnel.PleaseobservetherelevantsafetyinstructionsandreadtheOperatingInstructionscarefully.
EnjoyworkingwiththeProgramatEP3000/G2.
1.2 Introduction
ThesignsandsymbolsintheseOperatingInstructionsandonthefurnacefacilitatethefindingofimportantpointsandhavethefollowingmeanings:
Risksanddangers
Importantinformation
Contraindication
Burnhazard
Riskofcrushing
TheOperatingInstructionsmustberead
1.3 Notes regarding the Operating Instructions
Furnaceconcerned:ProgramatEP3000/G2Targetgroup: Dentaltechnologists
TheseOperatingInstructionsfacilitatethecorrect,safe,andeconomicuseoftheProgramatEP3000/G2furnace.
ShouldyoulosetheOperatingInstructions,extracopiescanbeorderedatanominalfeefromyourlocalIvoclarVivadentServiceCenterordownladedfromwww.ivoclarvivadent.com/downloadcenter.
IntheOperatingInstructions,thefurnaceisdescribedinthe200–240Vvoltageversion.Pleasenotethatthevoltagerangeshownontheimages(e.g.ratingplate)maydifferdependingonthevoltageversionofyourfurnace.
1.4 Notes on the different voltage versions
Thefurnaceisavailablewithdifferentvoltageversions.
– 110–120V/50–60Hz– 200–240V/50–60Hz
IntheOperatingInstructions,thefurnaceisdescribedinthe200–240Vvoltageversion.
Pleasenotethatthevoltagerangeshownontheimages(e.g.ratingplate)maydifferdependingonthevoltageversionofyourfurnace.
9
2. Safety First
ThischapterisespeciallyimportantforpersonnelwhoworkwiththeProgramatEP3000/G2orwhohavetocarryoutmaintenanceorrepairwork.Thischaptermustbereadandthecorrespondinginstructionsfollowed.
2.1 Indications
TheProgramatEP3000/G2mustonlybeusedtofireand/orpressdentalceramicmaterialsanditshouldbeusedforthispurposeonly.Otherusesthantheonesstipulated,e.g.cookingoffood,firingofothermaterials,etc.,arecontraindicated.Themanufacturerdoesnotassumeanyliabilityfordamageresultingfrommisuse.TheuserissolelyresponsibleforanyriskresultingfromfailuretoobservetheseInstructions.
Furtherinstructionstoassureproperuseofthefurnace:– Theinstructions,regulations,andnotesintheseOperatingInstruc-
tionsmustbeobserved.– Theinstructions,regulations,andnotesinthematerial‘sInstruc-
tionsforUsemustbeobserved.– Thefurnacemustbeoperatedundertheindicatedenvironmental
andoperatingconditions(seeChapter9).– TheProgramatEP3000/G2mustbeproperlymaintained.
2.1.1
Risks and dangers
Thefurnaceheadshouldnotberemovedfromthefurnacebaseaslongasthefurnaceheadiscon-nectedbymeansoftheheatercable.
2.1.2
Risks and dangers
Makesurethatnoliquidsorotherforeignobjectsenterthefurnace.
2.1.3
Contraindication
Firingtraysmustnotbeplacedintheareasurroundingthefiringtable,sincethiswillobstructtheclosingofthefurnacehead.
2.1.4
Contraindication
Foreignobjectsmustnotbeplacedonthefurnaceheadorairvents.Makesurethatnoliquidsorotherforeignobjectsentertheairvents,sincethismayresultinanelectricalshock.
2.1.5
Risks and dangers, burn hazard
Neverplaceobjectsinthefiringchamberbyhand,sincethereisaburnhazard.Alwaysusethetongs(accessories)suppliedforthispurpose.Nevertouchthehotsurfaceofthefurnacehead,asthereisaburnhazard.Pleasealsorefertothewarningsonthefurnace.
10
2.1.6
Risks and dangers
Donotcarrythefurnacebythecoolingtray.
2.1.7
Risks and dangers
Donotcarrythefurnaceheadbythecables,sincethecablesand
connectionsmaybedamaged.
2.1.8
Risks and dangers
Thefurnaceheadisequippedwithanelectricdriveandmustbeoperatedbymeansoftheelectroniccontrols.Neveropenthefurnaceheadbyhand,sincethemechanismwillbedamaged.
2.1.9
Contraindication
Neverusethefurnacewithoutafiringtable.
2.1.10
Contraindication
Donottouchthethermocoupleorthequartztubeinthefiringchamber.Avoidcontactwiththeskin(greasecontamination),asthepartsmaybedamaged.
2.1.11
Risk of crushing and burn hazard
Neverreachunderthefurnaceheadwiththehandorotherpartsofthebodyduringoperation,sincethereisariskofcrushingandaburnhazard.
2.1.12
Contraindication
Donotinsertanyforeignobjectsintotheairvents.Thereisariskofelectricalshock.
2.1.13
Contraindication
Thisproductcontainsceramicfibresandmayreleasefibredust.Donotusecompressedair,orblowonthefurnacethusdistributingthedustintheenvironmentandobservetheadditionalnotesonpage11.
2.1.14
Contraindication
Donotcoolthehotinvestmentringonthecoolingtray.Useonlythecoolinggridforthatpurpose.
2.1.15
Risks and dangers Thefurnacemustnotbeoperatedifthequartztubeortheinsulationinthefiringchamberisdamaged.Thereisariskofelectricshockuponcontactwiththeheatingwire.Avoidanydamageoftheinsulationbycontactwiththeinvestmenttongsorfiringtongs.Makesuretokeepthefiringplatefortheinvestmentringclean.
2.1.16
Contraindication OnlyuseoriginalringbasesfromIvoclarVivadent.Observethemaximumheight(57mm)anddiameter(35mm,50mmand65mm)oftheinvestmentring.
11
2.2 Health and safety instructions
ThisfurnacehasbeendesignedaccordingtoEN61010-1andhasbeenshippedfromthemanufacturerinexcellentconditionasfarassafetyregulationsareconcerned.Tomaintainthisconditionandtoensurerisk-freeoperation,theusermustobservethenotesandwarningscontainedintheseOperatingInstructions.
– Placefurnaceonafire-prooftable(observelocalregulations,e.g.distancetocombustiblesubstancesorobjects,etc.).
–Alwayskeeptheairventsattherearofthefurnacefreefromobstruction.
– Donottouchanypartsthatbecomehotduringoperationofthefurnace.Thereisaburnhazard.
–Cleanfurnaceonlywithadryorslightlymoistcloth.Donotuseanysolvents.Disconnectpowerbeforecleaning.
–Useoriginalpackagingfortransportationpurposes.
– Thefurnacemustbecoolbeforeitispackedfortransportation.
– Theusermustespeciallybecomefamiliarwiththewarningsandoperatingconditionstopreventinjurytopersonnelordamagetomaterials.ThemanufacturerisnotresponsiblefordamageresultingfrommisuseorfailuretoobservetheOperatingInstructions.Warrantyclaimscannotbeacceptedinsuchcases.
– Beforeswitchingonthefurnace,makesurethatthevoltageindicatedontheratingplatecomplieswithyourlocalpowersupply.
– Thepowersocketmustbeequippedwitharesidualcurrentcircuitbreaker.
– Thefurnacemustbepluggedintoasocketwithprotectedcontacts.
– Beforecalibration,maintenance,repair,orexchangeofparts,thepowermustbedisconnectedifthefurnaceistobeopened.
– Ifcalibration,maintenance,orrepairhastobecarriedoutwiththepowerconnectedandthefurnaceopen,onlyqualifiedpersonnel,whoarefamiliarwiththerisksanddangers,mayperformtheseprocedures.
– Aftermaintenance,therequiredsafetytests(highvoltageresistance,protectiveconductor,etc.)havetobecarriedout.
– Ensurethatonlyfusesoftheindicatedtypeandratedcurrentareused.
– Ifitisassumedthatsafeoperationisnolongerpossible,thepowermustbedisconnectedtoavoidaccidentaloperation.Safeoperationisnolongerpossibleif
– thefurnaceisvisiblydamaged – thefurnacedoesnotwork – thefurnacehasbeenstoredunderunfavourableconditions
overanextendedperiodoftime
– Useonlyoriginalspareparts.
– Thetemperaturerangeforfaultlessoperationis+5°Cto+40°C(41°Fto104°F).
– Ifthefurnacehasbeenstoredatverylowtemperaturesorhighatmospherichumidity,theheadhastobeopenedandtheunitdriedorlefttoadjusttoroomtemperatureforapprox.1hour(donotconnectthepoweryet).
– Thefurnacehasbeentestedforuseataltitudesofupto2000mabovesealevel.
– Thefurnacemayonlybeusedindoors.
– Donotrunthefurnaceviaanextensioncord.
– Whenplacingandremovingtheinvestmentring,makesurenottohittheinsulationofthefiringchamber.
– Thereisaburnhazardatthecoolingtrayifthefurnaceiscontinuouslyoperatedinthepressmode(stand-by=700°C).
Anydisruptionoftheprotectiveconductoreitherinsideoroutsidethefurnaceoranylooseningoftheprotectiveconductorconnectionmayleadtodangerfortheuserincaseofamalfunction.Deliberateinterruptionsarenottolerated.Materialsdevelopingharmfulgasesmustnotbefired.
Warnings regarding the removal of the heating muffle Thisproductcontainsceramicfibresandmayreleasefibre
dust.Fibredusthasprovedtobecarcinogenicinanimalexperiments.
TheheatingmufflemustonlybedisassembledbyacertifiedAfterSalesServiceCentre.InformationregardingtheSafetyDataSheetisalsoavailablefromyourAfterSalesServiceCentre.
Warning The insulation on this product contains refractory ceramic fibres (RCF) which pose a possible cancer hazard, if agitated and inhaled. May be irritating to the skin, eyes or respiratory tract if insulation is cracked or corrupted.
California Proposition 65 Warning: ”This product contains Refractory Ceramic Fibres, a substance known to the State of California to cause cancer.”
Disposal:Thefurnacesmustnotbedisposedinthenormaldomesticwaste.PleasecorrectlydisposeofoldfurnacesaccordingtothecorrespondingEUcouncildirective.InformationonthecorrectdisposalmayalsobefoundonyourlocalIvoclarVivadenthomepage.
12
3. Product Description
3.1 Components
TheProgramatEP3000/G2comprisesthefollowingcomponents:– Furnacebasewithelectronic
controls– Furnaceheadwithfiring
chamberandpressdrive– Firingplate– Coolingtray– Powercordandhosefor
vacuumpump– Vacuumpump(accessories)
3.2 Hazardous areas and safety equipment
Descriptionofthehazardousareasofthefurnace:
3.3 Functional description
Thefiring/pressingchambermaybeheateduptomax.1200°C(2192°F)bymeansofaheatingelement.Furthermore,thefiringchamberhasbeendesignedinsuchawaythatavacuummaybecreatedwithavacuumpump.Thepresssureforthepressprocedureisgeneratedbyapressdrive.Thefiring/pressingprogramsarecontrolledwiththecorrespondingelectroniccontrolsandsoftware.Moreover,thesetandactualtempera-turesarecontinuouslycompared.
3.4 Accessories (not part of the delivery form)
– AutomaticTemperatureCheckingSet2(ATK2)– ProgramatAccessoriesSet2(Programatfiring
trays,SiliconeNitrideFiringTray“K”,firingtongs,AutomaticTemperatureCheckingSet2)
– Vacuumpump
Descriptionofthesafetyequipmentofthefurnace:
Safety equipment Protective effect
Protectiveconductor Protectionfromelectricalshock
Electricalfuses Protectionfromelectricalshock
Housingandcovers Protectionfromelectricalshock,burnhazardandcrushing
Hazardous area Type of risk
Firingchamber Riskofburning
Opening/closingmechanism Riskofcrushing
Electricalcomponents Riskofelectricalshock
13
4. Installation and Initial Start-Up
4.1 Unpacking and checking the contents
Thepackagingprovidesthefollowingadvantages:– Reusablepackaging– Closingmechanismwithintegratedtransportationgrips– IdealprotectionbyStyrofoaminserts– Easyhandling/optimumunpacking– Thepackagingmaybeusedinseveralways(modules)Checkthedeliveryforcompleteness(seedeliveryforminChapter9)andtransportationdamage.Ifpartsaredamagedormissing,contactyourlocalIvoclarVivadentServiceCenter.
Therearenospecialtransportationgripsonthefurnace.Supportthebottomofthefurnacetocarryit.
Removefurnacecomponentsfromtheirpackagingandplacetheunitonasuitabletable.Pleaseobservetheinstructionsontheouterpackaging.
Packing and shipping of individual componentsThepackagingoftheEP3000/G2permitssimpleandsafeshippingofindividualcomponents.Simplyusethetwocorrespondinginserts.Foldthesideflapsandcombinethetwopartsbymeansofthetransportationflaps.Thepackagingmaybedisposedwiththeregularhouseholdrefuse.
3. Product Description
4.2 Selecting the location
Placethefurnaceonaflattableusingtherubberfeet.Makesurethatthefurnaceisnotplacedintheimmediatevicinityofheatersorothersourcesofheat.Makesurethatairmayproperlycirculatebetweenthewallandthefurnace.
Alsoensurethatthereisenoughspacebetweenthefurnaceandtheuser,asthefurnacereleasesheatduringtheopeningofthefurnacehead.
The furnace should neither be placed nor operated in areas where there is an explosion hazard.
4.3 Assembly
Makesurethevoltageindicatedontheratingplate(19)complieswiththelocalpowersupply.Ifthisisnotthecase,thefurnacemustnotbeconnected.
Werecommendkeepingtheoriginalpackagingforfutureserviceandtransportationpurposes.
Step 1: Assembling the cooling tray (34)Removebothscrews(35)includingthesiliconewasher(47)forthecoolingtray(34).
47
3547
35
19
14
Step 2: Mounting the furnace headThecompletefurnacehead(58)isbestmountedwiththerearpanelofthefurnacepointingtowardstheuser.Liftthefurnaceheadwithbothhands(seepicture)andcarefullypositionitonthefurnaceheadmounting(43).
Securethecoolingtray(34)withthetwoscrews(35)includingthesiliconewasher(47).
Placethecoolingtray(34)ontheframeplate(7).Makesurethatthecoolingtray(34)iscorrectlypositionedontheframeplate(7).
58
347
35
47
15
Step 3: Placing the firing plate for the investment ring (5)Thefiringplatefortheinvestmentring(5)cannowbeplacedonthefiringplateholder(48).
Ensurethatthefurnaceheadmountingmark(41)isalignedwiththefurnacebasemountingmark(42).
Makesurethatthefiringplateholder(48)isnotdamagedbymountingthefurnacehead.
Step 4: ConnectionsConnectthecablesofthefurnaceheadwiththefurnacebase.Proceedasfollows:
– Insertthethermocoupleplug(26)(makesurethatthepolarityoftheplugiscorrect)
– Inserttheheaterplug(28)– Insertthepressdriveplug(60)
548
28
2726
60
41
42
50
54
16
Securetheheaterplug(28)withtheplugfuse(27)byturningituntiltheheaterplug(28)hasbeensecured.
Step 5: Mounting the hood (36)Onceallcablesareproperlyconnectedtothefurnacebase,thehood(36)canbemounted.Subsequently,securethehoodwiththeknurledscrew(37).
Thefurnacemayonlybeoperatedwiththehoodmounted.
Step 6: Establishing additional connections
Power connection Pleasemakesurethatthevoltageindicatedontheratingplatecomplieswiththelocalpowersupply.Connectthepowercordwiththepowersocketofthefurnace(17).Vacuum pump connectionConnectthevacuumpumpplugwiththevacuumpumpsocket(18).
WerecommendusingonlytheVP4vacuumpumpfromIvoclarVivadent,sincethispumpisespeciallycoordinatedwiththefurnace.Ifotherpumpsareused,pleaseobserveanddonotexceedthemaximumpowerconsumption.
27
28
36
37
18
17
16
21
17
4.4 Removing the furnace head
Beforethehood(36)isremoved,thefurnacehastobeswitchedoffandthepowercorddisconnectedfromthepowersocket(17).
1. Loosenandremovetheknurledscrew(37)ofthehood(36)
2. Removethehood(36)3. Disconnectthepressdriveplug(60)4. Disconnectthethermocoupleplug(26)5. Disconnecttheheaterplug(28)6. Presstheleafspring(32)withafinger,liftoff
thefurnaceheadatthesametimeandremoveit
Makesurethefurnaceheadhascompletelycooleddownbeforeitisremoved(firehazard).
18
4.5 Initial start-up
1. Connectthepowercordwiththewallsocket.2. PuttheOn/Offswitch(11)attherearofthefurnaceonposition
„I“andconnectthevacuumpump.
4.5.1 Start screenImmediatelyafterswitchingon,thedisplaybrieflyshowsthestartscreen.
4.5.2 Self-testThefurnacewillautomaticallyconductaself-testafterstart.Theper-formanceofallfurnacecomponentsisautomaticallychecked.Thedisplayshowsthefollowingindicationsduringtheself-test:
1Softwareversion 2Statusbar 3Firinghours 4Currentpowersupply
Ifacomponentisdefective,thecorrespondingerrornumber(ERxxx)willbeindicatedinthedisplay.
4.5.3 Selecting the operation modeAftertheself-test,theselectionscreenfortheoperationmodewillbedisplayed.Itisusedtoselectthegeneraloperationmode(firingorpressing).TheoperationmodeselectioncanalsobeshownviatheFiring/Pressingkey(93)ifnoprogramisactive.
Thecursorposition(framearoundthesymbol)canbechangedbymeansoftheCursorkeys.ThemarkedoperationmodecanfinallybeselectedusingtheEnterkey.
Thisinformationisonlyshowniftheoperationmodehasbeenchanged.Makesurethatthefurnacerunsatthenewstand-bytemperaturelongenoughbeforeyoustartaprogram.
UsetheESCkeytoquitsuchmessages(Info,Hint,Error).TheacousticsignalcanbestoppedanytimebypressingtheSTOPkey.
4.5.4 Basic build-up of screensThedifferentareasofthescreenareshownwiththepressprogramstand-byandoperatingscreensasexamples.
Theuppermostlinemainlyshowsstatusinformation.Inthecentralandlargestarea,importantinformationoftherespectivescreenaredisplayed.Thelowestlineprovidesinformationaboutpossibleactivities(keystrokes).Thecommandkeys(Openfurnacehead,Closefurnacehead,STOP,STARTkeys)arenotshownforreasonsofclarity.
Recommendation area
Main area
Operation mode
Status area
Furnacetemperature
Possible keystrokes
Selected program, investment ring
1
2
4
3
19
5. Operation and Configuration
5.1 Introduction to the operation
TheProgramatEP3000/G2isequippedwithagraphicaldisplaywithbacklighting.Thefurnacecanbeoperatedbymeansofthemembrane-sealedkeypad.Inaddition,theparameterscanbeselecteddirectlybymeansoftheParameterfiringcurvewiththeParameterkeys.Thenumericandcommandkeyscanbeusedtoprogramandcontrolthefurnace.
5.2 Explanation of the key functions
– Firing / Pressing (93) Selectionofthefiringorpressingmode
– Program key (70) Thiskeyisusedtoswitchbetweentheparameterscreenandthe
stand-byscreen(oroperationscreen).
– START key (73) Startstheselectedprogram.Thefactthattheprogramisrunning
isindicatedbythegreenStartLED.Iftheprogramisinterrupted(1xSTOP),theStartLEDflashesuntilrenewedpressingofSTARTresultsintheprogrambeingresumed.
– STOP key (75) AprograminprogresscanbeinterruptedbypressingSTOPonce.
PressingSTOPtwicewillaborttheprogram.MovementofthefurnaceheadcanbestoppedatanytimebypressingSTOP.
ThebeepercanbeconfirmedbypressingtheSTOPkey.
– ESC key (71) Endsanentrywithoutacceptingthevalue.Returnfromthe
currenttothepreviousscreen.Confirmationoferrormessages.
– ENTER key (72) Confirmationofenterednumericvalue.Selectionofasettingora
testprogram.
– Numeric keys (92) Usedtoenternumericvalues.
– Cursor key left, right (80, 79) Thesekeyscanbeusedtomovethecursor(e.g.forselectingthe
operationmode,sizeoftheinvestmentring,parameterprogram-ming,etc.).Inaddition,thedisplayedprogram,informationorsettingcanbechanged.
– Minus and Plus keys (77, 76) Thesekeyscanbeusedtochangethenumericvaluedisplayedor
markedwiththecursor.
– Settings / Information (78) Thiskeyshowstheselection(Information/Settings).TheLeftkey
showstheinformationscreen.TheRightkeyshowsthesettingsscreen.
– Open furnace head (90) Openingofthefurnaceheadin5seconds.
– Close furnace head (91) Closingofthefurnaceheadin5seconds.
– T = Holding temperature (84) Showstheholdingtemperature(actualfiringtemperature).
– H = Holding time (85) Showstheholdingtime(actualfiringtime).
– S = Closing time (82) Indicatestheclosingtimeofthefurnacehead(preheatingtime).
– B = Stand-by temperature (81) Indicatesthestand-bytemperature.
– t = Temperature increase (83) Showsthetemperatureincreaseperminutefortheheating
process(°C/minor°F/min).
– V1 Vacuum on temp. (86) Showsthetemperatureatwhichthevacuumisswitchedon.
– V2 Vacuum off temp. (87) Showsthetemperatureatwhichthevacuumisswitchedoff.If
thistemperaturecorrespondstotheHoldingtemperatureT,thevacuumremainsonduringtheentireholdingtime.
Special case:IfthistemperatureV2isexactly1°C(or1°F)higherthantheHoldingtemperatureTwhilelong-termcoolingisactive,thevacuumremainsonduringtheentirelong-termcooling.
– L = Long-term cooling (88) Determinesthetemperaturepointatwhichthefurnacehead
shouldbeopenedaftercompletionoftheHoldingtimeandfreeorcontrolled(tL)cooling.
– Power-saving key (89) Power-savingfunctionactivated(onlypossiblewiththefurnace
headclosedandthefurnaceonidle).Thedisplayshowsthepower-savingicon.Pressinganykeyendsthepower-savingfunction.
– tL = Controlled cooling - temperature decrease rate (°C/min or °F/min)
– Home key (94) Returnto”ProgramselectionIndication“.
78
93
81 82 85
86
88
84
94
87
89
83
79 80 70 77 76 72 71
73
74
75
91
92
90
20
5.4 Adjustable parameters and possible value ranges
Symbol Parameter Value range °C Value range °F
P Programnumber 1-300
B Stand-bytemperature 100-700°C 212-1292°F
S Closingtime(min:sec) 00:18-30:00
t Temperatureincreaserate 10-140°C/min 18-252°F/min
T Holdingtemperature 100-1200°C 212-2192°F
H Holdingtime(min:sec) 00.01-60:00
V1 Vacuumon 0or1-1200°C 0or34-2192°F
V2 Vacuumoff 0or1-1200°C 0or34-2192°F
L Long-termcooling 0or50-1200°C 0or122-2192°F
tL Coolingtemperaturerate 0or1-50°C 0or2-90°F/min
t Temperatureincreaserate2ndstage 10-140°C/min 18-252°F/min
T2 Holdingtemperature2ndstage 100-1200°C/min 212-2192°F
H2 Holdingtime2ndstage(min:sec) 00.01-60:00
V1 2 Vacuumontemp.2ndstage 0or1-1200°C 0or34-2192°F
V2 2 Vacuumofftemp.2ndstage 0or1-1200°C 0or34-2192°F
Automatic plausibility checkThefurnaceisequippedwithanautomaticplausibilitycheckfunction.Theparameters(e.g.T960butL1000)arecheckeduponeachprogramstart.Incaseofcontradictoryparametercombinations,theprogramstopsautomaticallyandtherespectiveerrornumberisindicated.
5.3 Program structure
5.3.1 Firing programsAllthefiringprogramsareequivalentand,therefore,full-fledgedprograms.Ineachprogram,alltheparameterscanbeadjusted.
a) Ivoclar Vivadent firing programs for Ivoclar Vivadent materials
Whenthefurnaceisdeliveredexworks,theIvoclarVivadentfiringprogramsalreadycontaintherecommendedmaterialparametersettings.Moreover,theprogramsarewrite-protected.Consequently,itisnotpossibletoaccidentallyoverwritetheparameters.
Pleaserefertotherespectiveprogramtable(listofparameters)inChapter10.
However,theparametersaredesignedinsuchawaythattheycanbechangedandover-writtenatanytime,iftheprogramsaretobeusedforotherpurposes.Therefore,theseprogramsarealsoavailableasfree,individualprograms.
b) Individual firing programs Theindividualfiringprograms(atleast300)can
beprogrammedfreely.
5.3.2 Press programsa) Ivoclar Vivadent press programs for Ivoclar
Vivadent materials Whenthefurnaceisdeliveredexworks,the
standardpressprogramsalreadycontaintherecommendedmaterialparametersettings.Theycannotbeadjusted.
b) Individual press programs Theindividualpressprograms(20)canbe
programmedfreely.
21
5.5 Settings, special programs and information
ThisscreencanbeselectedviatheSettings/Informationkey(78)ifnoprogramisactive.
Thecursorposition(framearoundthesymbol)canbechangedbymeansoftheCursorkeys.Themarkedsymbol(InformationorSet-tings)canfinallybeselectedusingtheEnterkey.
5.5.1 Settings OncetheSettingshavebeenselected,thefirstpageoftheSettingsisdisplayed.Thecurrentpagenumberisshownintheupperline.
Thedisplayedpage(Setting)canbechangedusingtheCursorkeys.
IftheMinus/Pluskeysareshown,thedisplayedSetting(e.g.Contrast)canbeeditedwiththeMinus/Pluskeys.IftheENTERkeyisshown,thedisplayedSettingorTestprogram(e.g.Calibration)canbeconfirmedwiththeENTERkey.
TheESCkeycanbeusedtoreturntothepreviousscreen.
Page Description
1/24 Contrast
2/24 Temperature mode
3/24 Calibration program ATK2
4/24 Volume
5/24 Buzzertunes
6/24 Time
Indication on display Short description
Thecontrastcanbesetwithinthedis-playedlimitingvaluesusingtheMinus/Pluskeys.
TheMinus/Pluskeyscanbeusedtotogglebetween°Cand°F.
Thisprogramisusedtoconducttheauto-matictemperaturecalibrationwiththeATK2.PleaseobservethenotesinChapter7.4.
TheMinus/Pluskeyscanbeusedtosetthedesiredvolume.
TheMinus/Pluskeyscanbeusedtosetthedesiredbuzzertunes.
ThetimecanbeenteredusingtheNumerickeys.
Page Description
7/24 Date
8/24 General write protection
9/24 „Ivoclar Vivadent optimized temperature control“
10/24 Power-Saving mode
11/24 Serviceinterval
12/24 Protocolling
13/24 Vacuum testprogram
14/24 Heater testprogram
15/24 Keypad testprogram
16/24 Cleaning program
17/24 Dehumidifica- tionprogram
Indication on display Short description
ThedatecanbeenteredusingtheNumerickeys.
Thegeneralwriteprotectioncanbeactivatedorde-activatedusingtheMinus/Pluskeys.Thegeneralwriteprotec-tionlocksallfiringprograms.Theusercodeisrequired.
TheSTDcodeisrequired.
Theautomaticpower-savingmodecanbeactivatedanddeactivatedwiththe+or–key(foraddi-tionalinformation,seesection6.3.10)
Hereyoucansettheintervalfortheservicenotestobedisplayed(Hint1700).
Theautomaticproto-colfunctioncanbeactivatedordeactivatedwiththe+or–key.
SeeChapter5.5.2Specialprograms.
SeeChapter5.5.2Specialprograms
SeeChapter5.5.2Specialprograms
SeeChapter5.5.2Specialprograms
SeeChapter5.5.2Specialprograms
22
Theusercode(6725)isrequiredforsomeSettings.
5.5.2 Special programs
5.5.2.1 Vacuum test programWiththisvacuumtestprogram,theperformanceofthevacuumsys-temcanbeautomaticallytested.Forthatpurpose,theachieved(minimum)pressureinmbarismeasuredandindicated.Ifthepressurevalueisbelow80mbar(hPa),thevacuumperformanceofthesystemisadequate.5.5.2.2 Heater test program Thequalityoftheheatingmufflemaybeautomaticallycheckedbymeansoftheheatertest(duration:approx.7min.).Theheatertestshouldonlybeconductedwiththeemptyfiringchamber,sinceanobjectinthechamber(e.g.firingtray)mayinfluencethetestresult.Conducttheheatertestimmediatelyafterswitchingonthefurnaceandbeforeanyactualfiringprocedures.Ifthefurnaceistoohot,anincorrectheatingmufflequalitywillbeindicated.Iftheheatingelementqualityfallsbelow50%,replacingtheheatingelementisrecommended.
5.5.2.3 Keypad test Eachtimethekeypadispressed,ashortbeepsounds.ThekeypadtestcanbeendedbypressingESC.
5.5.2.4 Cleaning program Thecleaningprogramisusedto„clean“theheatingmuffle(duration:approx.17min.)Afteracleaningprogram,itisrecommendedtocalibratethefurnace.Incaseofproblemswithdiscolourationoftheceramic,werecommendreplacingthefiringtraymaterial.
5.5.2.5 Dehumidification program Thecondensationofwaterintheinsulationofthefiringchamberandthevacuumpumpwillresultinalowervacuumandthustoimpairedfiringresults.Forthatreason,thefurnaceheadshouldbekeptclosedwhenthefurnaceisswitchedofforisbelow100°C,inordertopreventtheabsorptionofhumidity.Startthedehumidifica-tionprogramifrequired(humidityintheinsulation).
5.5.3 InformationOncetheInformationhasbeenselectedviatheSettings/Informa-tionkey(78),thefirstpageoftheInformationisdisplayed.Thecurrentpagenumberisshownintheupperline.
Thedisplayedpage(Information)canbechangedusingtheCursorkeys.TheESCkeycanbeusedtoreturntothepreviousscreen.
Page Description
18/24 Load factory settings
19/24 Set thefiring hoursofthe furnacehead tozero20/24 Setthe
vacuumpumphourstozero
21/24 Calibration interval
22/24 Press calibration interval
23/24 Dehumidifica- tioninterval
24/24 Ivoclar VivadentAG
Indication on display
Short description
Resetsthevaluesandparameterstothefactorysettings.NOTE:Allindividualprogramsyouhavecreatedandsavedwillbedeletedwiththisfunction.Theusercodeisrequired.
Resetsthedeterminedfiringhours.Theusercodeisrequired.
Resetsthedeterminedvacuumpumphours.Theusercodeisrequired.
Exworks:12months(another1,3and6monthsarepossible).After-wards,areminderfortemperaturecalibrationappears.
Exworks:100presscycles.Afterwards,areminderforthetemperaturecalibrationappears.
Exworks:12months.After-wards,areminderforthedehumidifica-tionprogramappears.
OnlyfortheAfterSalesService.
23
5.6 Explanation of the symbols on the displayPage Description
1/11 Serialnumber
2/11 Software version
3/11 Firinghours ofthefurnace head
4/11 Operating hoursofthe furnace
5/11 Operating hoursofthe vacuumpump
6/11 Laststart ofthe calibration program
7/11 Calibration values
8/11 Press procedures sincethelast calibration
9/11 Laststart ofthe dehumidifica- tionprogram
10/11 Powersupply
11/11 Errormessages
Indication on display
Short description
Serialnumberofthefurnace,seealsorat-ingplate.
Calibrationvaluesat660°Cand962°C.
Numberofpresscyclessincethelaststartofthecalibrationprogram.
Showsthecurrentsupplyvoltage.
Showsthelasterrormessages
Symbol name
„One-stageprogram“
„Two-stageprogram“
„Two-stageprogram“
„Standardopeningofthefurnacehead“(togglewithMinus/Pluskeys)
„Quickopeningofthefurnacehead“(togglewithMinus/Pluskeys)
Predrying
„ThermoShockProtection“
Standardclosingofthefurnacehead
Openlock
Closedlock
„Generalwritepro-tectionactive“
„Paging“
OperationmodeFiring
Pressingornoteonactivepressprocess
Meaning
Showsthatastandardone-stagefiringprogramisused
Showsthatatwo-stagespecialprogramisused.Theboldlineshowsthatthevaluesofthefirststagearedisplayed.
Showsthatatwo-stagespecialprogramisused.Theboldlineshowsthatthevaluesofthesecondstagearedisplayed.
Showsthatthefurnaceheadopenswithstandardspeedafterfiring.
Showsthatthefurnaceheadopensquicklyafterfiring.
Showsthattheoption„Predrying“wasactivated.
ThermoShockProtectionactive
PredryingandThermoShockProtectionareinactive
„Individualwriteprotectioninactive“
„Individualwriteprotectionactive“
Allprogramsarewrite-protected.
TochangepagesintheParameterscreen;onlyfortwo-stagefiringprograms.
ForselectingtheFiringmode.
ForselectingthePressmode.Ifthepressprocesshasbeenstarted,thissymbolisshownnexttotheremainingpressingtime.
Symbol
24
Symbol name
Information
Settings
Meaning
ForselectingtheInfor-mationscreen.
ForselectingtheSettingsscreen.
Symbol
ForadditionalexplanationsonthesymbolsandthecorrespondingfunctionspleaserefertoChapter6.4„Otheroptionsandspecialfeaturesofthefurnace“.
5.7 Explanation of the beeper signals
Basically,thebuzzertunesandvolumesetbytheuserareusedforallacousticsignals.
ThebeepercanonlybeendedbypressingtheSTOPkey.
1 After the self-test is completed Toinformtheuserthattheautomaticself-testhasbeen
successfullycompleted,theselectedmelodyisplayed.
2 Furnace head open and temperature below 550 °C / 1022 °F Toinformtheuserthatthetemperatureintheopenfurnacehead
hasdroppedbelow550°C/1022°F,theselectedmelodyisplayed(5seconds).Inotherwords,thefurnaceheadisbasicallycoolenoughforthenextprogramstart.
3 Furnace head open and temperature below 320 °C / 608 °F Toinformtheuserthatthetemperatureintheopenfurnacehead
hasdroppedbelow320°C/608°F,theselectedmelodyisplayed.
Ifthefirstplayback(10seconds)isnotacknowledgedwiththeSTOPkey,asecondplaybacksoundsafter5minutes(for5minutes).Afterthat,nofurthersignalisplayed.
IfoneofthetwoplaybacksisacknowledgedwiththeSTOPkey,thesignaltransmitterisswitchedoffimmediatelyandnofurthersignalsindicatingthecooledfurnaceheadwillbesounded.
4 For error messages Errormessagesareacousticallysupportedwiththe‚errormelody‘
(endlessbeep).ThesignaltransmittermaybeswitchedoffwiththeSTOPkey,whiletheerrormessagestillremainsvisible.IftheerrormessageisacknowledgewiththeESCkey,thesignaltransmitterisalsoswitchedoff.
25
6. Practical Use
6.1 Switching on/off
PutON/OFFswitchonposition“I”.Thefurnaceconductsanauto-maticself-test,whichwillbeindicatedinthebeginning.Subsequently,astatusbarshowshowmany%oftheself-testhavebeencompleted.Makesurethatthefurnaceisnotmanipulateddur-ingthistime.Aftersuccessfulcompletionoftheself-test,themainmenuisshowninthedisplay.
6.2 Firing programs
6.2.1 Selecting the type of firing programOncetheFiringmodehasbeenselected,thefiringprogramtypescreenisdisplayed.ThereisachoicebetweenIvoclarVivadentfiringprogramsforIvoclarVivadentmaterialsandfree,individualfiringprograms.Thecursorposition(frame)canbechangedbymeansoftheCursorkeys.ThemarkedtypeoffiringprogramcanfinallybeselectedusingtheEnterkey.
6.2.2 Selecting the type of firing program and stand-by screen (firing program)Oncethedesiredtypeoffiringprogramhasbeenselected,thefiringprogramscreenisdisplayed.Thisscreencorrespondstothestand-byscreenofthefiringprograms.
– Thedisplayed(selectable)firingprogramcanbechangedusingtheCursorkeys.Thedisplayedfiringprogramcanfinallybeselect-edusingtheEnterkey.
– TheProgramkeycanbeusedtochangetotheparameterscreen(firingprogram).
– TheESCkeycanbeusedtochangetotheselectionoffiringprogramtypes.
6.2.3 Parameter screen (firing program)TheProgramkeycanbeusedtochangetotheparameterscreen(fir-ingprogram)anytime.EverytimetheProgramkeyispressed,itispossibletotogglebetweenthestand-byoroperationscreenandtheparameterscreen.
ThecursorpositioncanbemovedbetweentheparametersbymeansoftheCursorkeys.Themarkedparameter(orfunctionsymbol)canbeeditedwiththeMinus/PluskeysorviathenumerickeysandconfirmedbypressingENTER.Examples:
TheESCkeycanbeusedtoreturntothepreviousscreen.
– Program write protection and general write protection ThestandardfiringprogramsforIvoclarVivadentmaterialshave
beenlockedexworkswithanactiveprogramwriteprotection(closedlock),whichcanbedeactivatedbytheuserforeachprogramusingtheMinus/Pluskeys.
Theindividualfiringprogramsareopenexworks,i.e.theprogramwriteprotectionisnotactive(openlock).
Ifthesuperordinate„Generalwriteprotection“wasactivatedviaSettingsandusercode,the„Generalwriteprotectionsymbol“isdisplayedinstead.
– Indication of invalid entry Theblinkingexclamationmark(!)indicatesaninvalidentry.For
furthernotespleaserefertoChapter6.3.2.
– Predrying Ifthepredryingfunctionwasactivated,therespectivesymbolis
shownintheparameterscreen.ForfurthernotespleaserefertoChapter6.3.6.
– Standard / quick opening of the furnace head Theopeningtimeofthefurnaceheadattheendofthefiring
program(standard:60seconds,quick:18seconds)canbechangedbymeansoftheMinus/Pluskeys.ForfurthernotespleaserefertoChapter6.3.7.
– One-/two-stage program Ifthecursorispositionedonthe„Programstagesymbol“,the
programcanbeselectedasone-ortwo-stageprogramusingtheMinus/Pluskeys.ForfurthernotespleaserefertoChapter6.3.8.
– Displaying the parameters of the first/second stage (only for two-stage programs)
Ifthecursorispositionedonthe„Pagingsymbol“,itispossibletotogglebetweentheparametersofthefirstandsecondstageusingtheENTERkey.
6.2.4 Parameter detail screen (firing program)Theseparameterdetailscanbedisplayedinthefiringmodebypressingthecorrespondingkeys().
ThedisplayedparametercanbeeditedwiththeMinus/PluskeysorviathenumerickeysandconfirmedbypressingENTER.Theacceptablevaluerangeisshownforinformationpurposestoavoidincorrectentries.
6.2.5 Operation screen (firing program)Operationscreenwhileafiringprogramisinprogress.
Thefollowinginformationisshowninthisscreen:a)Programgroupb)Programnumberc)Remainingtimed)Currenttemperaturee)Statusofvacuum*)f)Statusbarinthefiringcurve
f)
a)
e)
b)
d)
c)
6.3 Press programs
6.3.1 Pressing using an Ivoclar Vivadent program
6.3.2 Selecting the press program typeOncethePressingmodehasbeenselected,thepressprogramtypescreenisdisplayed.ThereisachoicebetweenIvoclarVivadentpressprogramsforIvoclarVivadentmaterialsandfree,individualpressprograms.
Thecursorposition(frame)canbechangedbymeansoftheCursorkeys.ThemarkedtypeofpressprogramcanfinallybeselectedusingtheEnterkey.
6.3.3 Selecting the press programOncethedesiredpressprogramhasbeenselected,thepressprogramscreenisdisplayed.
TheCursorkeyscanbeusedtotogglebetweenthepossiblepressprograms.ThedisplayedpressprogramcanfinallybeselectedusingtheEnterkey.TheESCkeycanbeusedtochangetotheselectionofpressprogramtypes.
26
Ifatwo-stageprogramisselected,thefiringcurveisshownintwostages.Duringafiringprograminprogress,theparameterscreenoroperationscreenmaybedisplayedatanytimeforinformationpur-posesbypressingthePkey.However,theparametersmayonlybechangedwiththeprogramstoppedorthefurnaceinstand-bymode.
*)Thevacuumindicationisfadedoutifnovacuumisneeded.
6.2.6 Firing using an Ivoclar Vivadent program Step 1:Selectthedesiredfiringprogramaccordingtotheprogramtable(Chapter10).
Step 2:Openthefurnaceheadwiththe„Openfurnacehead“key(90)andplacethefiringtraywiththeobjecttobefiredinthefurnace.
Step 3:PressSTART(73)tostarttheselectedprogram.Thestatusisindicatedintheoperationscreen.
6.2.7 Firing using an individual program Step 1:Selectafree,individualfiringprogram.SeeChapter6.1.1,...
Step 2:Setthedesiredparameterseitherintheparameterscreen(seeChapter6.1.3)orbymeansoftheparameterdetailscreens(seeChapter6.1.4).
Step 3:Openthefurnaceheadwiththe„Openfurnacehead“key(90)andplacethefiringtraywiththeobjecttobefiredinthefurnace.
Step 4:PressSTART(73)tostarttheselectedprogram.Thestatusisindicatedintheoperationscreen.
Selecting the operation mode
Selecting the firing program type
Selecting the firing program and stand-by screen
Operation screen
Selecting the operation mode
Selecting the press program type
Selecting the press program
Investment ringselection
Stand-by screen
Operation screen
Parameter screen
Parameter screen
27
6.3.9 Notes on the individual press programs
Fortheabortspeed,werecommendusingavalueof300µm/mininthelayeringtechniqueand150µm/mininthestainingtechnique.Ahighervalue(abortspeede.g.300µm/min)abortsthepresspro-cedureearlier,whilealowervalue(abortspeede.g.100µm/min)abortsthepressprocedurelaterandprolongsthepressprocedure.
Fortheall-ceramicsystemsfromIvoclarVivadent(IPSe.max,IPSEmpressEsthetic),onlytheoriginalstandardpressprogramswhichareespeciallycoordinatedwiththematerialsmustbeused.
6.4 Other options and special features of the furnace
6.4.1 Quick selection of the firing programEachfiringprogramcanbedirectlyselectedbyitsprogramnumber.Toquicklyselectthefiringprogram,presstheProgramkeyandentertheprogramnumber.ConfirmwiththeENTERkey.Inaddition,theCursorkeyLeft/RightorMinus/Pluskeyscanbeusedinthestand-byscreentonavigatethroughthefiringprograms.
6.4.2 Indication of invalid entryIfaninvalidvalueisenteredbymeansofthenumerickeys(outsidetheacceptablevaluerange),theinvalidentrystillblinksafterconfirmation.Aserrormessage,anexclamationmark(!)blinksinthebottomlineoftheparameterordetailscreenuntilthenextvalueisenteredandsuccessfullyconfirmedortheprocessisabortedwithESC.Theold,validvaluereappears.
6.4.3 Error message symbolTheerrormessagesymbolshouldsupplyafirstindicationofthetypeoferror:
Note, information Entry errorInformation symbol Exclamation mark symbol
Technical error Fork wrench symbol
6.3.4 Selecting the investment ringOncethedesiredtypeofpressprogramhasbeenselected,theinvestmentringisselected.
Thecursorposition(framearoundthesymbol)canbechangedbymeansoftheCursorkeys.ThemarkedinvestmentringsizecanfinallybeselectedusingtheEnterkey.TheESCkeycanbeusedtochangetotheselectionofpressprograms.
6.3.5 Stand-by screen (press program)Oncetheinvestmentringhasbeenselected,thestand-byscreenappears(pressprogram).
TheProgramkeycanbeusedtochangetotheparameterscreen(pressprogram).
TheESCkeycanbeusedtochangetotheselectionoftheinvest-mentring.
6.3.6 Parameter screen (press program)TheProgramkeycanbeusedtochangetotheparameterscreen(pressprogram).
TheESCkeycanbeusedtoreturnfromtheparameterscreentothestand-byscreenortothepreviousscreen.
ThecursorpositioncanbemovedbetweentheparametersbymeansoftheCursorkeys.ThemarkedparametercanbeeditedwiththeMinus/PluskeysorviathenumerickeysandconfirmedbypressingENTER.
6.3.7 Operation screen (press program)Thefollowingoperationscreenappearswhileapressprogramisintheheatingandholdingtimephase.
Duringtheactualpressprocessthefollowingoperationscreenappears.
Theanimatedarrowindicatesthatthepressplungermovesdown-wards.Theentiretimeofthepresscycleisdisplayedonceandthepathwhichthepressplungerhasalreadycoveredsincethestartofthepressprocedure.
6.3.8 Notes on the Ivoclar Vivadent press programsTheProgramatEP3000/G2hasbeenespeciallycoordinatedwiththematerialssystemsfromIvoclarVivadent.Therefore,therespectiveparametersofthedifferentprogramshavealreadybeensetexworks.Youonlyhavetoselectthedesiredprogramforthecorrespondingmaterial.
Symbol Parameter Value range Value range
B Stand-bytemperature 100-700°C 212-1292°F
t Temperatureincreaserate 10-140°C/min 18-252°F/min
T Holdingtemperature 100-1200°C 212-2192°F
H Holdingtime(min:sec) 00:00-60:00
E Abortspeed 0-100000µm/min
28
6.4.4 Stopping the running programPresstheSTOPkeyoncetopausearunningprogram.ThegreenLEDintheSTARTkeyblinks.PresstheSTOPkeytwicetocompletelystoptheprogramorpressSTARTtocontinue.
6.4.5 Changing the parameters while the program is interruptedAllparametersoftheprogram,whichhavenotyetbeenexecuted,canbechangedwhiletheprogramisinterrupted.
6.4.6 Closing of the furnace head Variousfunctionsareavailablefortheclosingofthefurnacehead.Ifthecursorisonthespotintheparameterlistshowninthefigure,youcantogglebetweenthefollowingfunctionsbyusingthe+or–key:
– ThermoShockProtection
– Vortrocknen
– keineFunktionaktiviert
– TSP – Thermo Shock Protection TheTSPfunctionpreventstheobjectfromtoohighof
temperaturesduringtheclosingprocess.Forthispurpose,theTSPfunctiongaugesthetemperatureofthefiringchamberinthefurnaceheaduponstartofthefiringprogram.Ifrequired,theclosingpathwithinthesetclosingtimeSisadjusted.TSPisonlyactiveifnoactivepredryinghasbeenselectedorifthetempera-tureinthefiringchamberistoohighwhentheprogramisstarted.Inaddition,TSPworksonlyifthestand-bytemperatureofB=403°C/757°FrequiredforIvoclarVivadentmaterialsisused.TheactiveTSPfunctionisshownonthedisplaybymeansofthesymbol„TSP“.Ifprogramsarestartedwithatemperatureofmorethan680°C/1256°F,anerrormessageisproduced.
– Pre-drying Ifthepre-dryingfunctionisactive,the„pre-dryingtemperature“
issetafterthefiringprogramstartwiththefurnaceheadopen(heatingorcooling).
This„pre-dryingtemperature“correspondstothestand-bytemperatureoftheactivefiringprogram.Oncethistemperatureisreached,thefurnaceheadisclosedwithinthedesiredclosingtime.
6.4.7 Quick opening of the furnace headTheopeningmodeofthefurnaceheadcanbeselectedintheparameterscreenofthefiringprograms.Ifthecursorisseton„Standardfurnaceheadopening“youcantoggleto„Quickopeningofthefurnacehead“andviceversabymeansoftheMinus/Pluskeys(standardopeningofthefurnacehead:in60seconds,quickopeningofthefurnacehead:in18seconds).
6.4.8 One-stage /two-stage programsIntheparameterscreen,thefiringprogramcanbesetasaone-stageortwo-stageprogram.Ifthecursorissetonthe„one-stagesymbol“,pressingtheMinus/Pluskeysresultsinthesymboltochangetothe„two-stagesymbol“.Atthesametime,theprogramisalsochangedtobecomea„two-stageprogram“.Ifthecursorissetonthe„two-stagesymbol“,pressingtheMinus/Pluskeysresultsinthesymboltochangetothe„one-stagesymbol“.Atthesametime,theprogramisalsochangedtobecomea„one-stageprogram“.
6.4.9 Software updateTheuserwillbeabletoconductasoftwareupdatebyPCanddown-loadcable.Forthatpurpose,thesoftwaredownloadmodeofthefurnaceisactivatedbypressingtwospecialkeyssimultaneouslywhilethepowersupplyisswitchedon.Forfurtherdetails,pleaserefertotheSoftwareUpdateInstructions(www.ivoclarvivadent.com/downloadcenter).
6.4.10 Power-save modeIfthepower-savemodeisactivatedandthefurnaceheadclosed,thisfunctionisautomaticallyactivatedafter30minutesifthefurnaceisidleandnokeyispressedduringthistime.ThePowerSavingiconappearsonthedisplay.Thepower-savemodeisterminatedbypressinganykey.
Theenergysavingmode(PowerSavingTechnology)isavailableintheEP3000/G2onlyinthefiringmode.Inthepressmode,theenergysavingfunctionisdisabled,asthefurnacemustfulfiladditionalrequirements.
29
7. Maintenance, Cleaning, and Diagnosis
ThischapterdescribestheusermaintenanceandcleaningproceduresfortheProgramatEP3000/G2.Onlythosetasksarelistedthatmaybeperformedbydentalprofessionals.Allothertasksmustbeper-formedbyqualifiedservicepersonnelatacertifiedIvoclarVivadentServiceCenter.
7.1 Monitoring and maintenance
Thetimeforthesemaintenanceproceduresdependsonthefrequencyofuseandtheworkinghabitsoftheusers.Forthatreason,therecommendedtimesareonlyapproximations.
Thisfurnacehasbeendevelopedfortypicaluseindentallaboratories.Iftheproductisusedinaproductionenterprise,forindustrialapplications,andforcontinuoususe,prematureageingoftheexpendablepartshastobeexpected.
Theexpendablepartsareasfollows:– Heatingmuffle– Insulationmaterial
Expandablepartsarenotcoveredbythewarranty.Pleasealsoobservetheshorterserviceandmaintenanceintervals.
Ingeneral,thefurnaceheadshouldnotbereplacedsincethecomponents(furnaceheadandfurnacebase)havebeencoordinatedwitheachother.However,ifthefurnaceheadmustbereplacedformaintenancereasons,subsequenttemperaturecalibrationisrequired.
What Part When
Checkallplug-inconnectionsforcorrectfit Var.externalconnections weekly
Checkifthefurnaceheadopenssmoothlyandwithoutexcessivenoise. Openingmechanism monthly
Checkifthethermocoupleisstraightandintherightplace. Thermocouple(4) weekly
Checktheinsulationforcracksanddamages.IftheinsulationisworndownithastobereplacedbyacertifiedIvoclarVivadentServiceCenter.Finehair-linecracksonthesurfaceoftheinsulationareharmlessanddonotinflu-encethefunctionofthefurnaceinanegativefashion.
Insulation(3) monthly
Checkifthesealingrimsofthefurnaceheadandthefurnacebasearecleanandundamaged.
Sealingrimsofthefurnacehead(2)andthefurnacebase(1)
weekly
Checkthekeypadforvisibledamage.Ifthekeypadisdamaged,ithastobereplacedbyacertifiedIvoclarVivadentServiceCenter.
Keypad(10) weekly
Checktemperature.Usethetemperaturecheckingsettocheckandadjustthetemperatureinthefurnace.
Firingchamber twiceayear
Checkthequartzglasscylindertomakesurethequartzglassisnotdefective.
Firingchamber daily
30
7.2 Cleaning
Thefurnacemayonlybecleanedwhenitiscool,sincethereisaburnhazard.Donotuseanycleaningsolutions.
Thefollowingpartshavetobecleanedfromtimetotime:
7.3 Temperature calibration
1. Selectthecalibrationprogram.
2. Removethefiringplatefromthefurnaceusingthefurnacetongsandplaceitonthecoolingtray.
3. CarefullygriptheupperpartoftheATK2usingthefurnacetongs(Caution:Fractureriskoftheceramic)andinsertitintotheholesdesignatedforthispurposeuntilitsnapsintoplace.Theorientationofthecalibrationsample(leftorright)isnotimportant.
4. Ifnecessary,usethefurnacetongstoapplyslightpressuretothecenterofthecalibrationbaseuntilthecalibrationsampleclicksintoplace.
Observethecorrespondingmarkings.
5. Startthecalibrationprogram.
6. Attheendoftheprogram,openthefurnaceheadandcarefullyremovetheATK2usingthefurnacetongsandplaceitonthecoolingtraytoallowittocool.
7. Replacethefiringplateusingthefurnacetongs.
8. Closethefurnaceheadandselectafiringprogram.
9. TheATK2canonlybeusedonce.Useanewcalibrationsetforthenextcalibrationprocedure.
7.4 Service note
Whentheservicenote(Hint1700)appearsforthefirsttime,twoyearshavepassedortheheatingmufflehasmorethan1200firinghours.Forthatreason,IvoclarVivadentrecommendstohavethefur-naceserviced.PleaseseeyourEquipmentServicePassforfurtherinformation.TheintervaluntilthenextservicenotecanbeselectedonceintheSettings(seeChapter5.5.1).
7.5 Stand-by
Werecommendkeepingthefurnaceheadclosed,particularlyifthetemperaturedropsbelow100°C/212°F.Thiswillpreventunintentionalmoistureabsorptionandformationofcondensateinthefiringchamber.Consequently,vacuumproblemsareavoidedandtheservicelifeoftheheatingelementisprolonged.
What: When: Cleaning material:
Housing(9)andfurnacehead(25)
ifrequired soft,drycloth
Keypad(10) weekly soft,drycloth
Coolingtray(34) daily cleaningbrush*
Insulation(3) daily cleaningbrush*
Sealingrimofthefurnacehead(2)andsealingsurface(1)
daily cleaningbrushandasoftcloth
Firingplate ifrequired cleaningbrushorvacuumcleaner
*Nevercleanwithcompressedair!
31
7.6 Replacing the press plunger
Inordertofacilitatereplacingthepressplunger,thefollowingprocedureisrecommended.
1.Removethescrew(20)andthepressdrivecover(56)whilethefurnaceheadisclosed.
2.Loosentheterminalscrew(65)fromthepressplungerbyabouthalfarotation.
3.Openthefurnaceheadbymeansoftherespectivekey(90).Oncethefurnaceheadiswideopen,switchoffthefurnace,disconnectthepower,andallowthefurnacetocooltoroomtemperature.
4.Pushthepressplunger(59)withslightlyrotatingmovementsintothefiringchamberwithonehandandpullfrombelowwiththeotherhand.
Contraindication: Donottouchthethermocouplewhenreplacingthepress
plunger.
5.Pushthewhitepressplunger(59)withthetaperaheadintotheguidebush.Pushthepressplungerwithslightlyrotatingmovementsintoitssplittapersocket(64)andfastenthescrew(65).
Contraindication: Neverreachintothepressdriveduring
operation.Thereisariskofburnhazardandcrushing.
Mountthepressdrivecover(56)andfastenwithscrew.
6.ConnectthepowerplugandswitchonthefurnacewiththeI/Oswitch.
20
56
65
6459
32
8. What if ...
Thischapterwillhelpyoutorecognizemalfunctionsandtakeappropriatemeasuresor,ifpossibleandacceptable,toperformsomesimplerepairs.
8.1 Error messages
Thefurnacecontinuouslychecksallfunctionsduringoperation.Ifanerrorisdetected,therespectiveerrormessageisdisplayed.
Thefollowingerrormessagesmaybedisplayed.Ifthereareanyquestions,pleasecontacttheIvoclarVivadentAfterSalesService.
Error /HintNo.
Continua-tion pos-sible
Error Error Message Text
2 T<B EnteralogicalvalueforT.
8 L>T Enteralogicalvalueforlong-termcoolingL.
9 V2x<=V1x Enteralogicalvalueforthevacuum-ontemperatureVx1orthevacuum-offtemperatureVx2.
10 V2x>Tx+1°C EnteralogicalvalueforV1x,V2x
11 IncorrectvaluesforV1x,V2x GebensieplausibleWertefürV1x,V2xein.
13*,**
CurrenttemperatureafterStart>Tx+80°C
Excesstemperature!Programaborted,furnaceheadopenstoallowthefurnacetocooldown!
16 T2<T1 EnteralowervalueforT1orahighervalueforT2.
17 Powerfailure>10sduringafiringprograminprogress
Afiringprograminprogresswasinterruptedformorethan10s.Theprogramcannotbecontinued!
18 T1>V12 EnteralowervalueforT1orahighervalueforV12.
19 vVset,butV2ismissingorinvalid
Pre-vacuumactivated!V2mustbehigherthanB.
20**
no Errorintheheatingsystem Checktheheaterfuse.Ifthefuseisi.o.contactyourServiceCenter.
23 Heatingmuffleveryold Theheatingmuffleisveryold.Itisrecommendedtoreplaceit.Aftertheerrormessagehasbeenacknowledged,afiringprogrammaystillbestarted.
24 Heatingmuffledefective Theconditionofthemuffleissopoorthatishastobereplacedimmediately.
26 Tis>B+160°Catthestartofafiringprogram
Firingchambertoohottostartafiringprogram.
27**,***
no Furnaceheadcannotbeinitialized
Thefurnaceheadcannotbemovedtothefinalposition.Itmightbeblockedbyanexternalmechanicalsource!IfnotcontactyourServiceCenter!
28**
Thefurnaceheaddoesnotreachthetargetposition
Thefurnaceheaddoesnotopen/closecorrectly.Thefurnaceheadwasmanuallymovedorisobstructed.Thefurnaceheadmustonlybemovedusingthekeysintendedforthispurpose!
32**
no Thevacuumisnotreleased Thevacuumcannotbereleased.Thevacuumvalvemightbedirtyorstuck.ContactyourServiceCenter.
33 Necessaryvacuum(xxxmbar)isnotreachedwithin1min.
Thevacuumcannotbeestablished.Checkthesealofthefiringchamber,vacuumhose,vacuumpump,pumpfuse.
103 Programstartblocked Startingaprogramisnotpossibleduetoatechnicalmalfunction.
107 Incorrecttimesettings(date/time)
Theclocksettingsareincorrect.Pleasesetacorrectdateandacorrecttime!
110 HV>H(H2) EnteralowervalueforHVorahighervalueforH(H2).
120 „Shareoftheholdingtimewithvacuum“isactivated,butVx2doesnotcorrespondtoTxorTx+1
ActivatethevacuumduringtheholdingtimeTxordeactivateHV.
500 Errorpressingtime Max.pressingtimeexceeded.
504 Errorpressposition Max.positionexceeded.
505 Errorpressforce Max.pressforceexceeded.
513 Errorpressdriveinitalized Pressdriveisnotinitialized.Pleaseswitchthefurnaceoffandonagain.
514 Errorpressdrive Technicalerrorinpressdrive.
33
PleasecontacttheIvoclarVivadentAfterSalesService,ifoneofthefollowingerrormessagesisbeingdisplayed:
25,29
43,44,45,46,47,48
54,56
103,107
143,144,145,146,147,148
700,701,703,704,705,706,707
1010,1011,1012,1013,1014,1015,1016
1024,1025,1026,1028
1143,1144,1145,1146,1147,1148
1202,1203,1204,1205,1206,1207
1400,1401,1402
1500
* Furnaceheadopenswhenthiserroroccurs.** Aprograminprogressisstopped.*** Theerrorcannotbeacknowledged;theprogramscannotbestarted.
Error /HintNo.
Continua-tion pos-sible
Error Error Message Text
520 Errormufflecrack CDSCrackDetectionSystemhasbeenactivated.Theprogramhasbeenabortedandthepressplungerhasbeenmovedbackwards.CDScouldprobablysaveyourrestorationsfrommufflecracks.Pleasecheckyourrestorationsbeforeyoucontinueyourworkingprogress.
521 Errormufflecrack CDSCrackDetectionSystemhasbeenactivated.Theprogramhasbeenabortedandthepressplungerhasbeenmovedbackwards.CDScouldprobablysaveyourrestorationsfrommufflecracks.Pleasecheckyourrestorationsbeforeyoucontinueyourworkingprogress.
522 Errormufflecrack CDSCrackDetectionSystemhasbeenactivated.Theprogramhasbeenabortedandthepressplungerhasbeenmovedbackwards.CDScouldprobablysaveyourrestorationsfrommufflecracks.Pleasecheckyourrestorationsbeforeyoucontinueyourworkingprogress.
530 Errorduringloggingofpressprogramdata
Anerrorhasoccurredduringloggingofpressprogramdata.Thestoragemediummightbefull.
700 Supplyvoltageoutsidetheacceptablerange
Thesupplyvoltageisoutsidetheacceptablerange.Checkthesupplyvoltage.
701***
no Start-upabortedduetoanerror Theself-testofthefurnacewasinterruptedbyanerror.Itisnotpossibletoworkwiththefurnace!Switchthefurnaceoffandonagain,oncetheerrorhasbeenrectified.
702 Briefpowerfailureduringaprograminprogress
Afiringprograminprogresswasinterruptedbyabriefpowerfailure.Theprogramiscontinued!
704 Powerfailureduringanover-nightprograminprogress
Anovernightprograminprocesswasinterruptedbyapowerfailure.Theprogramiscontinued!
707 Incorrectsupplyvoltage Thefurnaceisoperatedwiththeincorrectsupplyvoltage.Makesurethatthefurnaceisoperatedwiththesupplyvoltageindicatedontheratingplate.
800 Finalvacuumvaluenotreached Therequiredfinalvacuumvaluecannotbereached.Checkthevacuumpump.
801 Vacuumdrop Anunacceptablevacuumdrophasoccurred
802 Thevacuumdoesnotincrease(self-test)
Novacuumincreasecouldbemeasured.Checkthefollowingpoints:Isthefiringchambertight(nocontaminationonthesealingsurfaces)?Isthevacuumhoseconnected?Isthevacuumpumpcon-nected?IsthefuseF1o.k.?
1310 Calibrationreminder Sometimehaspassedsincethelastcalibrationprocedure.Calibratethefurancesoon.
1311 Calibrationreminder–presscycles
Somepresscycleshasbeendonesincethelastcalibrationprocedure.Calibratethefurancesoon.
1312 Dehumidificationreminder Sometimehaspassedsincethelastdehumidification.Conductadehumidificationinthenearfuture.
1550 Changeoperationmode Theoperationmodehasbeenchanged!Makesurethatthefurnacerunsatthenewstand-bytem-peraturelongenoughbeforeyoustartaprogram.
1700 Servicenote Twoyearshavepassedortheheatingmufflehasbeeninuseformorethan1200firinghourssincethelastinspectionofthefurnace.IvoclarVivadentthereforerecommendsthatyourfurnaceshouldbeinspected.PleaserefertotheEquipmentServicePassportortheOperatingInstructionsforfurtherinformation.TheintervaluntilthenextappearanceofthenotecanbesetintheSettings.
34
8.2 Technical malfunctions
Thesemalfunctionsmayoccurwithoutanerrormessagebeingdisplayed.*Ifthereareanyquestions,pleasecontacttheIvoclarVivadentAfterSalesService.
8.3 Repair
RepairsmayonlybecarriedoutbyacertifiedIvoclarVivadentServiceCenter.PleaserefertotheaddressesonthelastpageoftheseOperatingInstructions.
IfrepairsduringthewarrantyperiodarenotcarriedoutbyacertifiedIvoclarVivadentServiceCenter,thewarrantywillbevoidedimmediately.Pleasealsorefertothecorrespondingwarrantyregulations.
8.4 Load factory settings
Ifyouwanttoresetthefurnacetoitsoriginalsettings,chooseSettings–FactorySettings(seeChapter5.5.1).Allprograms(bothIvoclarVivadentandindividualprograms)andfurnacesettingswillberesettothefactorysettings.
Fehler Kontrollfrage Massnahme
Vacuum is not released or only very slowlyIsthevacuumreleasedwithinapproximately30seconds?
Waituntilthevacuumisreleased,removeobject.Switchthefurnaceonandoffagain.*
Indication on display incomplete Activatethedisplaytestprogram *
Writing in the display is very hard to read Isthecontrastproperlyset? Adjustcontrast
Display not illuminatedIsthefurnaceproperlyconnectedaccordingtotheOperatingInstructionsandswitchedon?
Correctlyconnectthefurnaceandswitchiton
Buzzer does not sound Isthebuzzerswitchedoff(Tune=0)? Selecttune1–5
Furnace head does not open
Wasthefurnaceheadmovedmanually? Openthefurnaceheadonlybyusingthecorres-pondingkeys.Switchthefurnaceonandoffagain.
Hasthevacuumalreadybeenreleased? Istheprogramstillrunning?Waituntiltheprogramiscomplete.Switchfurnaceoffandonagain.*
Vacuum pump does not start working
Isthevacuumpumpfusedefective? Checkfuseandreplaceifnecessary.
Wasthemaximumpowerconsumptionexceeded? UseonlythevacuumpumprecommendedbyIvoclarVivadent
Isthevacuumpumpplugcorrectlyconnected? Correctlyconnectthevacuumpumptothefurnacebase.
Final vacuum is not reached
IsthevacuumhoseOK? Checkvacuumhoseandhoseconnection
IsthepumpoutputOK? Startthevacuumtestprogram
Humidity/condensationinthevacuumhose? Startdehumidificationprogram
Incorrect or illogical temperature indication
Isthethermocouplebentorfractured? *
Isthethermocouplecorrectlyconnected? Correctlyconnectthermocouple
Isthethermocoupleplugdefective? *
Hairline cracks in the heating muffle
Arethecracksverysmallandinsignificant(hairlinecracks)?
Smallcracksinthemufflearenormalanddonotnegativelyinfluencethefunctionofthefurnace
Arethecrackslargeorhavepartsoftheheatingmufflebrokenoff?
*
Cracks in the insulation
Arethecracksverysmallandinsignificant(hairlinecracks)?
Smallcracksintheinsulationdonotnegativelyinfluencethefurnace
Arethecrackslargeorhavepartsoftheinsulationbrokenoff?
*
Cracks in the quartz glass / heating elementAretherecracksinthequartzglassoristhequartzglasssheathingtheheatingwirebroken?
Switchoffthefurnace*
35
9. Product Specifications
9.1 Delivery form
– ProgramatEP3000/G2– ProgramatfringtrayKit– Firingplate– AutomaticTemperaturCheckingSet2ATK2(TestSet)– USBdownloadcable– USBStickProgramat– PrograBase2Software– Servicepassportequipment– OperatingInstructions– Coolinggrid
9.1.2 Recommended accessories – ProgramatAccessoriesSet– AutomaticTemperatureCheckingSet2(ATK2)– VacuumpumpVP4
9.2 Technical data
Powersupply 110–120V/50–60Hz 200–240V/50–60Hz
OvervoltagecategoryIIContaminationlevel2
Toleratedvoltagefluctuations +/-10%
Max.powerconsumption 12Aat110-120V 8.5Aat200-240VAcceptabledataforvacuumpumpsfromothermanufacturersMax.output: 250W/max.leakagecurrent0.75mAFinalvacuum: <50mbar Useonlytestedpumps
Electricalfuses: 110–120V: 250V/T15A(heatingcircuit) 250V/T5A(vacuumpump) 200–240V: 250V/T8A(heatingcircuit) 250V/T3.15A(vacuumpump)
Dimensionsofelectricalfuses: 110–120V: Diameter6.3x32mm 200–240V: Diameter5x20mm
Dimensionsoftheclosedfurnace: Depth:430mm/Width:305mm/410mm(withcoolingtray) Height:565mm
Usablesizeofthefiringchamber: Diameter80mm Height48mm
Max.firingtemperature: 1200°C
Weight: Furnacebase:12.0kg Furnacehead:7.0kg
Safety information TheProgramatEP3000/G2complieswiththefollowingguidelines:– EN61010-1:20012ndEd.– IEC61010-1:20012ndEd.– UL/CSA61010-1:20042ndEd.– EN61010-2-010:20032ndEd.– IEC61010-2-010:20032ndEd.– CSA61010-2-010:20042ndEd
Radioprotection/electromagneticcompatibility EMCtested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:+5°Cto+40°C/+41°Fto+104°FAcceptable humidity range:80%maximumrelativehumidityfortemperaturesupto31°C(87.8°F)graduallydecreasingto50%relativehumidityat40°C(104°F);condensationexcluded.Acceptable ambient pressure: Thefurnacehasbeentestedforuseataltitudesofupto2,000m(6562ft.)abovesealevel.
9.4 Acceptable transportation and storage conditions
Acceptabletemperaturerange: -20°Cto+65°C/-4°Fto+149°FAcceptablehumidityrange: Max.80%relativehumidityAcceptableambientpressure: 500mbarto1060mbar
UseonlytheoriginalpackagingoftheProgramatEP3000/G2togetherwiththecorrespondingfoammaterialforshippingpurposes.
10. Appendix
10.1 Program table
Twoprogramtables(°C/°F)areenclosedintheseOperatingInstructions.Ifnot,pleasecontactyourlocalIvoclarVivadentServiceCenter.
Important information Thecurrentprogramtablesarealsoavailableat: www.ivoclarvivadent.com/downloadcenter
TheprogramtablescanbedownloadedfromtheInternetasPDFfiles.Pleasemakesurethatyourprogramtablecomplieswiththesoftwareversionyouuse,asthetableiscoordinatedwiththerespectivesoftwareversion.
36
37
10.2 Menu structure
10.2.1 Selecting the operation mode and program type
10.2.2 Firing programs
10.2.3 Press programs
38
10.2.4 Settings / Information
39
Ivoclar Vivadent AGBendererstrasse 2FL-9494 SchaanLiechtensteinTel. +423 235 35 35Fax +423 235 33 60www.ivoclarvivadent.com
Ivoclar Vivadent Pty. Ltd. 1 – 5 Overseas DriveP.O. Box 367Noble Park, Vic. 3174AustraliaTel. +61 3 979 595 99Fax +61 3 979 596 45www.ivoclarvivadent.com.au
Ivoclar Vivadent GmbHBremschlstr. 16Postfach 223A-6706 BürsAustriaTel. +43 5552 624 49Fax +43 5552 675 15www.ivoclarvivadent.com
Ivoclar Vivadent Ltda.Rua Geraldo Flausino Gomes, 78 – 6.º andar Cjs. 61/62Bairro: Brooklin NovoCEP: 04575-060 São Paulo – SPBrazilTel. +55 11 3466 0800Fax +55 11 3466 0840www.ivoclarvivadent.com.br
Ivoclar Vivadent Inc.2785 Skymark Avenue, Unit 1MississaugaOntario L4W 4Y3CanadaTel. +1 905 238 5700Fax +1 905 238 5711www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd.Rm 603 Kuen Yang International Business PlazaNo. 798 Zhao Jia Bang RoadShanghai 200030ChinaTel. +86 21 5456 0776Fax +86 21 6445 1561www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd.Calle 134 No. 7-B-83, Of. 520BogotáColombiaTel. +57 1 627 33 99Fax +57 1 633 16 63www.ivoclarvivadent.com
Ivoclar Vivadent SASB.P. 118F-74410 Saint-JoriozFranceTel. +33 450 88 64 00Fax +33 450 68 91 52www.ivoclarvivadent.fr
Ivoclar Vivadent GmbH Dr. Adolf-Schneider-Str. 2D-73479 Ellwangen, JagstGermanyTel. +49 (0) 79 61 / 8 89-0Fax +49 (0) 79 61 / 63 26www.ivoclarvivadent.de
Ivoclar Vivadent Marketing Ltd. (Liaison Office) 503/504 Raheja Plaza 15 B Shah Industrial Estate Veera Desai Road, Andheri (West) Mumbai, 400 053 IndiaTel. +91 (22) 2673 0302 Fax +91 (22) 2673 0301 www.ivoclarvivadent.com
Ivoclar Vivadent s.r.l. Via Gustav Flora, 3239025 Naturno (BZ)ItalyTel. +39 0473 67 01 11Fax +39 0473 66 77 80www.ivoclarvivadent.it
Ivoclar Vivadent K.K.1-28-24-4F HongoBunkyo-ku Tokyo 113-0033JapanTel. +81 3 6903 3535Fax +81 3 5844 3657www.ivoclarvivadent.jp
Ivoclar Vivadent S.A. de C.V.Av. Insurgentes Sur No. 863Piso 14, Col. Napoles03810 México, D.F.MéxicoTel. +52 (55) 50 62 10 00Fax +52 (55) 50 62 10 29www.ivoclarvivadent.com.mx
Ivoclar Vivadent Ltd.12 Omega St, AlbanyPO Box 5243 Wellesley StAuckland, New ZealandTel. +64 9 914 9999Fax +64 9 814 9990www.ivoclarvivadent.co.nz
Ivoclar Vivadent Polska Sp. z o.o.Al. Jana Pawla II 7800-175 WarszawaPolandTel. +48 22 635 54 96Fax +48 22 635 54 69www.ivoclarvivadent.pl
Ivoclar Vivadent Marketing Ltd. Derbenevskaja Nabereshnaya 11, Geb. W 115114 Moscow Russia Tel. +7 495 913 66 19 Fax +7 495 913 66 15 www.ivoclarvivadent.ru
Ivoclar Vivadent Marketing Ltd.171 Chin Swee Road#02-01 San CentreSingapore 169877 Tel. +65 6535 6775Fax +65 6535 4991www.ivoclarvivadent.com
Ivoclar Vivadent S.L.U.c/ Emilio Muñoz Nº 15 Entrada c/ AlbarracinE-28037 MadridSpainTel. + 34 91 375 78 20Fax + 34 91 375 78 38www.ivoclarvivadent.es
Ivoclar Vivadent ABDalvägen 14S-169 56 SolnaSwedenTel. +46 (0) 8 514 93 930Fax +46 (0) 8 514 93 940www.ivoclarvivadent.se
Ivoclar Vivadent Liaison OfficeAhi Evran Caddesi No 1Polaris Is Merkezi Kat: 780670 MaslakIstanbul Turkey Tel. +90 212 346 04 04 Fax +90 212 346 04 24www.ivoclarvivadent.com
Ivoclar Vivadent LimitedGround Floor Compass BuildingFeldspar CloseWarrens Business ParkEnderbyLeicester LE19 4SEUnited KingdomTel. +44 116 284 78 80Fax +44 116 284 78 81www.ivoclarvivadent.com
Ivoclar Vivadent, Inc. 175 Pineview DriveAmherst, N.Y. 14228USATel. +1 800 533 6825Fax +1 716 691 2285www.ivoclarvivadent.com
Version: 1Date information prepared: 03/2011Valid as of Software V6.0
Thisapparatushasbeendevelopedsolelyforuseindentistry.Start-upandoperationshouldbecarriedoutstrictlyaccordingtotheOperatingInstructions.LiabilitycannotbeacceptedfordamagesresultingfrommisuseorfailuretoobservetheInstructions.TheuserissolelyresponsiblefortestingtheapparatusforitssuitabilityforanypurposenotexplicitlystatedintheInstructions.Descriptionsanddataconstitutenowarrantyofattributes.
PrintedinAustria©IvoclarVivadentAG,Schaan/Liechtenstein638623/0311/e