Q-SUNOperation and Maintenance

1

Q-SUN Xenon Arc TestersOperation and Maintenance

Andy Francis – Marketing Director

Bill Tobin – Senior Technical Marketing Specialist

Dave Duecker – Technical Marketing Specialist

Sean Fowler – Senior Technical Director

Q-Lab CorporationClick here to view the morning presentation.

Click here to view the afternoon presentation.

Q-SUNOperation and Maintenance

2

HousekeepingYou’ll receive a follow-up email from info@email.q-lab.com with links to a survey, registration for future webinars, and to download the slides

• Our ongoing webinar series can be found at: q-lab.com/webinarseries

• Our archived webinars are hosted at:q-lab.com/webinars

• Use the Q&A feature in Zoom to ask us questions today!

Q-SUNOperation and Maintenance

3

Q-SUN Xenon Test Chambers

• Simulate light (outdoor direct sunlight, indoor filtered light); heat (elevated temperature), and water (humidity and water spray)

• Used to Evaluate durability and lightfastnessof a range of materials

Q-SUNOperation and Maintenance

4

Q-SUN Xenon Arc Models

Xe-3

Xe-1

Xe-2

Q-SUNOperation and Maintenance

5

• Safety

• Features and Functions

• Running a Test

• Calibration

• Maintenance

Q-SUN Topics

Q-SUNOperation and Maintenance

6

Safety InformationXenon Lamps

• Xenon Lamps give off UV light and should not be viewed directly

• Interlocks stop tester when door opened

Q-SUNOperation and Maintenance

7

Safety InformationInterlock Switches

Xe-1 Xe-2(overhead view)

Xe-3

Q-SUNOperation and Maintenance

8

Q-SUN Safety Hazards

Electrical Shock Hot Surface

Q-SUNOperation and Maintenance

9

• Safety

• Features and Functions

• Running a Test

• Calibration

• Maintenance

Topics

Q-SUNOperation and Maintenance

10

Q-SUN Overview1) Simple user interface

2) USB port for data transfer

3) Xenon lamps with irradiance control

4) Optical filters

5) Water spray

6) Onboard irradiance sensors

7) Black Panel Temp sensor

8) Specimen holders

9) Relative Humidity/CAT sensor

Q-SUNOperation and Maintenance

11

Q-SUN Optional Features

Q-SUNOperation and Maintenance

12

Q-SUN light delivery

Xenon arc lamps

Optical filters

Xe-2 lantern Xe-1 / Xe-3 flat filter

Q-SUNOperation and Maintenance

13

Optical Filters and Lamps

Xe-1 and Xe-3

Lamp Holder

Xenon Lamp

Filter Lantern

Xe-2

Q-SUNOperation and Maintenance

14

Xenon Arc Spectrum

0.0

0.5

1.0

1.5

2.0

2.5

250 300 350 400 450 500 550 600 650 700 750 800

Spectral Irradiance

(W/m2/nm)

Wavelength (nm)

Peak Sunlight (G177)

Q-SUN Daylight-Q

Q-SUNOperation and Maintenance

15

SOLAR EYE Irradiance Control System

• SOLAR EYE Irradiance Control maintains the same light output at all times

• Ballasts control lamp output

• Allows for excellent repeatability and reproducibility

Sensor

Lamp

Ballast

Q-SUNOperation and Maintenance

16

Irradiance Control SystemOnboard Sensors

Xe-1Xe-2

(overhead)

Xe-3

Q-SUNOperation and Maintenance

17

Temperature Control SystemBlack Panel (BP or IBP) Thermometer

Xe-1Xe-3

Q-SUNOperation and Maintenance

18

Temperature Control SystemBlack Panel Thermometer and CAT/RH Sensor

Black Panel Thermometer(IBP shown)

Chamber Air Temperature / Relative Humidity Sensor

Xe-2

Q-SUNOperation and Maintenance

19

Temperature SystemChamber Air Thermometer

• Optional in Xe-1

• Tester can control temperature by BP or CAT; the other is simply monitored

Q-SUNOperation and Maintenance

20

RH/Temperature Control SystemChamber Air Thermometer / RH Sensor

• One sensor monitors chamber temperature and RH in Xe-2 and Xe-3

• Disposable; replace annually

Q-SUNOperation and Maintenance

21

Chiller (Optional, Xe-1 / Xe-3)

Reduces minimum BPT by:

~10 °C (Xe-3)

~20 °C (Xe-1)

Q-SUNOperation and Maintenance

22

Q-SUN Water Delivery

• Water Spray

– Front

– Back

– Dual (Auxiliary)

– Immersion

• Relative Humidity control (Xe-2 and Xe-3)

Q-SUNOperation and Maintenance

23

Q-SUN Water Quality

• Purified (RO/DI) water is required

• Spray water requires even higher purity and also low silica

• Q-Lab follows ASTM G151 recommendations

Water Delivery

Resistivity (Ω·cm)

Conductivity(μS/cm)

Silica(ppm)

Total Dissolved Solids (ppm)

pH

Spray > 5 M < 0.2 < 0.1 < 0.1 6-8

Humidity > 200 k < 5.0 No requirement < 2.5 6-8

Q-SUNOperation and Maintenance

24

Spray Systems (Optional)

Front

Back

Dual

Q-SUNOperation and Maintenance

25

Water Spray System

• Two nozzles, used for:

– Mist

– Thermal Shock

– Erosion

Xe-1 Xe-3

• Features

– Pulse Rate Control

– Automatic Fault Detector

Q-SUNOperation and Maintenance

26

Water Spray System

• One nozzle, used for:

– Mist

– Thermal Shock

• Features

– Pulse Rate Control

– Automatic Fault Detector

• Second Nozzle for optional Back Spray

Xe-2

Q-SUNOperation and Maintenance

27

Relative Humidity Control System

• Feedback Loop System

• RH/CAT Sensor

• Main Controller

• Humidity Generator

– Nebulizer in Xe-2

– Boiler in Xe-3

Heater

Chamber

Recirc. Air Blower

Fresh Air blower

Q-SUNOperation and Maintenance

28

Q-SUN Specimen Exposure Areas

Xe-1Xe-2

Xe-3

Q-SUNOperation and Maintenance

29

• Safety

• Functions of the Tester

• Running a Test

• Calibration

• Maintenance

Q-SUN Topics

Q-SUNOperation and Maintenance

30

Step 1: Select an Optical Filter

• Check your test method

– Almost all test methods describe the filter type

• If you don’t test to a specific method:

– Outdoor applications: Daylight - Q

– Indoor applications: Window - Q

Q-SUNOperation and Maintenance

31

Q-SUN Optical Filters

0.0

0.5

1.0

1.5

2.0

2.5

3.0

250 300 350 400 450 500 550 600 650 700 750 800

Spectral Irradiance

(W/m2/nm)

Wavelength (nm)

ASTM G177 Daylight

Daylight - Q

Extended UV - Q/B

Window - Q

ExtendedUV

Window filter

Natural sunlight

Daylight filter

Q-SUNOperation and Maintenance

32

Step 2: Select Irradiance Settings

• 340 nm Control Point

– For outdoor service environments

• 420 nm Control Point

– For indoor service environments

• TUV Control Point

– 300-400 nm (wideband, general service environments)

– Typically used for European standards

Q-SUNOperation and Maintenance

33

Select Irradiance Values

Achievable irradiance values vary by optical filter, tester type, and control point

Q-SUNOperation and Maintenance

34

Step 3: Select Black Panel Type

Panel ConstructionASTM Designation

ISO Designation

Temp Range (°C)

Black painted stainless steel

Uninsulated Black Panel

Black Panel 45-110

Black painted stainless steel mounted on0.6 cm white PVDF

InsulatedBlack Panel

Black Standard 50-120

Q-SUNOperation and Maintenance

35

Black Panel Temperature Control

• BP/IBP is typically mounted in holder (sometimes directly on tray)

• Irradiance, color, and thickness all affect specimen temperature

Q-SUNOperation and Maintenance

36

Step 4: Programming

• Light

• Dark

• Light + Spray

• Dark + Spray

• Dark + Spray Front and Back

• Dual Spray

• Light + Dual Spray

• Light + Immersion

• Dark + Immersion

Q-SUNOperation and Maintenance

37

Programming TestsLegacy Controller

Xe-3

Xe-1

Xe-2

Program key

Q-SUNOperation and Maintenance

38

Programming TestsNew Gen 4 Main Controller

Xe-3

Xe-1

Xe-2

USB/LED

Q-SUNOperation and Maintenance

39

Status and Menu Screens

Status screenSetpoint and actual controls

Test timers

Menu screenCycle management

CalibrationSettings

Q-SUNOperation and Maintenance

40

Status and Menu Screens

Status screenSetpoint and actual controls

Test timers

Menu screenCycle management

CalibrationSettings

Q-SUNOperation and Maintenance

41

Programming Tests

Q-SUNOperation and Maintenance

42

Step 5: Calibration

More on this later!

Q-SUNOperation and Maintenance

43

Step 6: Specimen Mounting

• Specimen Capacity– Xe-1: 17 (51 × 102 mm)

– Xe-2: 31 (45 × 132 mm)

– Xe-3: 55 (51 × 102 mm)

• Type of holder– Open-Backed (thick, rigid specimens)

– Solid-Backed (flexible specimens)

– Three-dimensional

• Specimen Tray– Solid

– Mesh (open)

• Masking– Commonly used in textile testing

Q-SUNOperation and Maintenance

44

Specimen Holders (Xe-1, Xe-3)

3D Specimen HoldersPanel Holders

Q-SUNOperation and Maintenance

45

Specimen Mounting Trays (Xe-1, Xe-3)

Solid Tray Open Mesh Tray

Q-SUNOperation and Maintenance

46

Q-SUN Xe-2 Specimen Holders (Xe-2)

Open-backed Close-backed Textile masks

Q-SUNOperation and Maintenance

47

Mounting Flexible Specimens

Textiles Thin Films

Q-SUNOperation and Maintenance

48

Step 7: Running the Test

• Specimen Repositioning

• Tester status and monitoring

• Data Logging via VIRTUAL STRIPCHART (optional)

• Calibration

Q-SUNOperation and Maintenance

49

Specimen Repositioning• Ensures best repeatability and reproducibility

• Perform at least 4 times per test

• Important for both rotating rack and flat array testers

Q-SUNOperation and Maintenance

50

LED Status Indicator

Q-SUNOperation and Maintenance

51

Tester Status

Q-SUNOperation and Maintenance

52

Monitoring the Test

Q-SUNOperation and Maintenance

53

• Safety

• Functions of the Tester

• Running a Test

• Calibration

• Maintenance

Q-SUN Topics

Q-SUNOperation and Maintenance

54

Q-SUN Calibration

• Irradiance sensor

– Every 500 light hours

• Black Panel temp sensor

– Every 6 months

• Chamber air temp sensor

– Every 12 months

• Calibration best practices

– Calibrate whenever test conditions are changed

– Calibrate at operating conditions

– Always calibrate the lamps before calibrating the Black Panel

Q-SUNOperation and Maintenance

55

Irradiance Calibration Devices

CR20 UC1 + UC20

Q-SUNOperation and Maintenance

56

Irradiance calibration ports

Xe-1Xe-2

Xe-3

Q-SUNOperation and Maintenance

57

Irradiance CalibrationNew Q-SUN Testers

Plug UC20 Smart Sensor directly into tester USB port

Select optical filter and calibrate

Q-SUNOperation and Maintenance

58

Black Panel Calibration Devices

CT202 UC1 + UC202

BP IBP

Q-SUNOperation and Maintenance

59

Black Panel temperature sensor calibrationTester black panels and calibrated black panel are placed

adjacent to each other in the chamber

Tester Black Panel

Calibration black panel

Calibrationdevice

Q-SUNOperation and Maintenance

60

Black Panel Temp Sensor Calibration

In holders On tray

Q-SUNOperation and Maintenance

61

Black Panel Thermometer Calibration

New Q-SUN:Let temperature stabilize

and press Calibrate

Legacy Q-SUN; enter value from UC202 or CT202

Q-SUNOperation and Maintenance

62

RH/CAT Sensor ReplacementYearly (Xe-2 and Xe-3)

Xe-3Xe-2

Q-SUNOperation and Maintenance

63

Chamber Air Temp Sensor Calibration

Q-SUNOperation and Maintenance

64

Yearly Reference Device Calibration

Replace/return reference calibration devices to Q-Lab or local representative for recalibration/replacement

– UC20 Irradiance Smart Sensors (replace/recalibrate)

– UC202 Temperature Smart Sensors (replace/recalibrate)

– Chamber Air Temp / RH sensor (replace)

– CR20 (recalibrate)

– CT202 (recalibrate)

Q-SUNOperation and Maintenance

65

• Safety

• Functions of the Tester

• Running a Test

• Calibration

• Maintenance

Q-SUN Topics

Q-SUNOperation and Maintenance

66

Xenon Lamp Replacement Schedule

Q-SUN tester IrradianceWarrantedLamp Life (hrs)

Legacy Typical 1500

NEW “E” Model Typical 3000

NEW “E” Model Maximum 1000

• Xenon lamps need to be replaced because they “age” (spectral shift to less UV)

• Q-SUN Optical Filters do not age

Q-SUNOperation and Maintenance

67

Q-SUN Xe-1, Xe-3 Lamp Replacement

(2) Remove lamp and clean housing

(1) Remove lamp housing from tester

(3) Clean UV filters

Q-SUNOperation and Maintenance

68

Q-SUN Xe-1, Xe-3 Lamp Replacement

(4) Replace lamp

(7) Install replacement lamp

(6) Final lamp check

(5) Verify trigger finger contact

Q-SUNOperation and Maintenance

69

Q-SUN Xe-2 Lamp Replacement

(1) Remove old lamp

(2) Remove and clean filter lantern with ammonia

Q-SUNOperation and Maintenance

70

Q-SUN Xe-2 Lamp Replacement

(4) Verify trigger finger contact

(3) Install replacement lamp

Q-SUNOperation and Maintenance

71

Routine Maintenance

• Clean / Replace Air Filters (Monthly)

• Clean humidifier

• Inspect Water Filter

• Clean Spray Nozzles

• Inspect Chamber Wall Reflectors

• Clean slip ring track (Xe-2)

Every 6 months

Q-SUNOperation and Maintenance

72

Clean or Replace Air Filters

Disposable Washable

Q-SUNOperation and Maintenance

73

Clean or Replace Air Filters

Xe-3 chamber air filter

Xe-3 Ballast Blower air filter

Xe-1 air filter

Chiller air filter

Q-SUNOperation and Maintenance

74

Clean or Replace Air Filters

Xe-2 Chamber Air Filter

Xe-2 LampCoolingAir Filter

Q-SUNOperation and Maintenance

75

Clean humidifier

Water pan Reservoir

• Disassemble humidifier water pan and reservoir

• Clean with alcohol or mild detergent

Q-SUNOperation and Maintenance

76

Inspect Water Filter

Xe-3Xe-2

Q-SUNOperation and Maintenance

77

Clean Spray Nozzles

• Remove and disassemble for cleaning

• Clean in ultra-sonic cleaner or …

• Clean with anti-scaling cleaner (to remove calcium and magnesium deposits) or…

• Thorough washing and rinsing in detergent

Q-SUNOperation and Maintenance

78

Inspect Chamber Reflectors

Clean - OK Cloudy - Replace

Q-SUNOperation and Maintenance

79

Clean slip ring track (Xe-2)

• Wipe down with scouring pads

• Clean with isopropyl alcoholNew contactless system

for BP Temp Sensor!

Q-SUNOperation and Maintenance

80

Q-Lab Tester Education Program• This presentation is a condensed version of our full

Q-SUN operator training program

• Q-Lab offers on-site and virtual training, including hands-on sessions and more in-depth review of tester construction and operation!

• Q-Lab offers customized training plans including:– Hands-on operator training for your existing tester.

– Lecture-style presentations

– Customizable subject matter

– Content for all experience levels

– Affordable, full day sessions with all-in-one pricing

– Private or group sessions at your site or a shared site

• Contact info@q-lab.com for more info

Smrithi KumarCustomer Education Specialist

Q-SUNOperation and Maintenance

81

Q-Lab Repair and Tester Support• Free email/phone support for North America and remote offices worldwide

– Call +1 440 835 8700 (Mon-Fri 8-5 EST)

– Email repair@q-Lab.com

• On-site emergency repair

• ISO 17025 Accredited on-site calibration audits for all models (current and legacy)– Includes a comprehensive tester calibration

– Thorough inspection / data collection for functionality and on-site training

– Endorsed by A2LA/ILAC for temperature, humidity, irradiance, pressure and orifice (MTG)

• New tester commissioning and re-commissioning

• Accessory installation– Add Water Repur, Chiller, or Space Saver; verify commissioning of relocated tester

• Call Q-Lab Repair or email repair@q-lab.com for a quote today!

Q-SUNOperation and Maintenance

82

Questions?