Post on 21-Aug-2018
transcript
April 2006/2
Uric Acid FS TBHBA Mindray BS300
Parameters
No # Test Uric Acid Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 546 Sec Wave 700
Sample Vol 6 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Uric Acid
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0.07 Factor 0
Upper 0 Upper 0 Upper 20.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 3021 .. .. …
Notes
1. Please refer to the package insert for Uric Acid FS TBHBA for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
-AMYLASE CC FS Mindray BS300
Parameters
No # Test AMY Method Kinetic Direction Ascend
Unit U/l Decimals 1 Prim. Wave 405 Sec Wave 700
Sample Vol 5 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Amylase
Reaction 8 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 2000 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0501 .. .. …
Notes
1. Please refer to the package insert for -Amylase CC FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Bicarbonate FS Mindray BS300
Parameters
No # Test BIC Method Endpoint Direction Descend
Unit mmol/l Decimals 1 Prim. Wave 405 Sec Wave 510
Sample Vol 15 R1 Vol. 200 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Bicarbonate
Reaction 3 - 15 Substrate 2.5 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 50 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0950 .. .. …
Notes
1. Please refer to the package insert for Bicarbonate FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Bilirubin Auto Direct FS Mindray BS300
Parameters
No # Test BiliD Method Endpoint Direction Ascend
Unit mg/dl Decimals 2 Prim. Wave 546 Sec Wave 670
Sample Vol 10 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Bili Direct
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 10.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0821 .. .. …
Notes
1. Please refer to the package insert for Bilirubin Auto Direct FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Bilirubin Auto Total FS Mindray BS300
Parameters
No # Test BiliT Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 546 Sec Wave 670
Sample Vol 6 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Bili Total
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 30.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0811 .. .. …
Notes
1. Please refer to the package insert for Bilirubin Auto Total FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Calcium AS FS Mindray BS300
Parameters
No # Test Ca AS Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 670 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Calcium
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0.04 Factor 0
Upper 0 Upper 0 Upper 25.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1130 .. .. …
Notes
1. Please refer to the package insert for Calcium AS FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
January 2008/1
CK-MB FS (IFCC mod.) Mindray BS300
Parameters
No # Test CKMB Method Kinetic Direction Asscend
Unit U/l Decimals 1 Prim. Wave 340 Sec Wave 405
Sample Vol 10 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name CK-MB
Reaction 15 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 1500 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1651 .. .. …
Notes
1. Please refer to the package insert for CK-MB FS (IFCC mod.) for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
January 2008/1
CK-NAC FS (IFCC mod.) Mindray BS300
Parameters
No # Test CK Method Kinetic Direction Asscend
Unit U/l Decimals 1 Prim. Wave 340 Sec Wave 405
Sample Vol 10 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name CK-NAC
Reaction 15 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 2000 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1601 .. .. …
Notes
1. Please refer to the package insert for CK-NAC FS (IFCC mod.) for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Chloride FS Mindray BS300
Parameters
No # Test Cl Method Endpoint Direction Ascend
Unit mmol/l Decimals 0 Prim. Wave 450 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Chloride
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 130 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1200 .. .. …
Notes
1. Please refer to the package insert for Chloride FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Cholesterol FS Mindray BS300
Parameters
No # Test CHOL Method Endpoint Direction Ascend
Unit mg/dl Decimals 0 Prim. Wave 510 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Cholesterol
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 3 Factor 0
Upper 0 Upper 0 Upper 750 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1300 .. .. …
Notes
1. Please refer to the package insert for Cholesterol FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
July 2006/3
Cholinesterase FS Mindray BS300
Parameters
No # Test CHE Method Kinetic Direction Descend
Unit U/l Decimals 0 Prim. Wave 405 Sec Wave 0
Sample Vol 4 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name CHE
Reaction 7 - 25 Substrate 2.5 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 20000 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1401 .. .. …
Notes
1. Please refer to the package insert for Cholinesterase FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Creatinine FS Mindray BS300
Parameters
No # Test Crea Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 510 Sec Wave 578
Sample Vol 15 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Creatinine
Reaction 5 - 13 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0.2 Factor 0
Upper 0 Upper 0 Upper 15.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1711 .. .. …
Notes
1. Please refer to the package insert for Creatinine FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Ferritin FS Mindray BS300
Parameters
No # Test Ferritin Method Fixed Time Direction Ascend
Unit µg/dl Decimals 1 Prim. Wave 578 Sec Wave 700
Sample Vol 9 R1 Vol. 180 R2 Vol. 90 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Ferritin
Reaction 3 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 5 Factor 0
Upper 0 Upper 0 Upper 1000 Speed 0
Calibration
Request
Rule 5P Logit Log
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 7059 .. .. …
Notes
1. Please refer to the package insert for Ferritin FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Iron FS Ferene Mindray BS300
Parameters
No # Test Fe Method Endpoint Direction Ascend
Unit µg/dl Decimals 1 Prim. Wave 630 Sec Wave 700
Sample Vol 15 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Iron
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 5 Factor 0
Upper 0 Upper 0 Upper 1000 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 1911 .. .. …
Notes
1. Please refer to the package insert for Iron FS Ferene for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
Alkaline Phosphatase FS IFCC 37°C Mindray BS300
Parameters
No # Test ALP IFCC Method Kinetic Direction Ascend
Unit U/l Decimals 1 Prim. Wave 405 Sec Wave 700
Sample Vol 4 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name ALP IFCC
Reaction 6 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 700 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0441 .. .. …
Notes
1. Please refer to the package insert for Alkaline Phosphatase FS IFCC 37°C for detailed information about the test on
the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
April 2006/2
Gamma GT FS (Szasz mod.) Mindray BS300
Parameters
No # Test GGT Method Kinetic Direction Ascend
Unit U/l Decimals 1 Prim. Wave 405 Sec Wave 700
Sample Vol 8 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Gamma GT
Reaction 6 - 21 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 570 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 2801 .. .. …
Notes
1. Please refer to the package insert for Gamma GT FS (Szasz mod.) for detailed information about the test on the
following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Glucose GOD FS Mindray BS300
Parameters
No # Test GLUC Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 510 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Glucose
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 400 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 2500 .. .. …
Notes
1. Please refer to the package insert for Glucose GOD FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
January 2009/3
oneHbA1c FS Mindray BS300
Parameters
No # Test HbA1c Method Endpoint Direction Ascend
Unit # Decimals 2 Prim. Wave 670 Sec Wave 0
Sample Vol 4 R1 Vol. 180 R2 Vol. 90 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name HbA1c
Reaction -20 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower # Factor 0
Upper 0 Upper 0 Upper # Speed 0
Calibration
Request
Rule Spline
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 3329 .. .. …
Notes
1. Please refer to the package insert for oneHbA1c FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
HDL-C Immuno FS Mindray BS300
Parameters
No # Test HDL-C Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 630 Sec Wave 700
Sample Vol 3 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name HDL-C
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 180 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 3521 .. .. …
Notes
1. Please refer to the package insert for HDL-C Immuno FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Lactate FS Mindray BS300
Parameters
No # Test LACT Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 340 Sec Wave 700
Sample Vol 4 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Lactate
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 120 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 4001 .. .. …
Notes
1. Please refer to the package insert for Lactate FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
LDH FS DGKC Mindray BS300
Parameters
No # Test LDH DGKC Method Kinetic Direction Descend
Unit U/l Decimals 0 Prim. Wave 340 Sec Wave 405
Sample Vol 7 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name LDH DGKC
Reaction 7 - 25 Substrate 2.5 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 2400 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 4201 .. .. …
Notes
1. Please refer to the package insert for LDH FS DGKC for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
LDL-C Select FS Mindray BS300
Parameters
No # Test LDL-C Method Endpoint Direction Ascend
Unit mg/dl Decimals 1 Prim. Wave 630 Sec Wave 700
Sample Vol 3 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name LDL-C
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 400 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 4121 .. .. …
Notes
1. Please refer to the package insert for LDL-C Select FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Lipase DC FS Mindray BS300
Parameters
No # Test LIP Method Kinetic Direction Ascend
Unit U/l Decimals 1 Prim. Wave 578 Sec Wave 700
Sample Vol 5 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 10 Antigen Check Antigen Lower 0 Full Name Lipase
Reaction 8 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 300 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 4321 .. .. …
Notes
1. Please refer to the package insert for Lipase DC FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
November 2008/1
Lp(a) 21 FS Mindray BS300
Parameters
No # Test LPA Method Endpoint Direction Ascend
Unit mg/dL Decimals 1 Prim. Wave 700 Sec Wave 0
Sample Vol 5 R1 Vol. 200 R2 Vol. 100 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Lp(a)
Reaction 6 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 3 Factor 0
Upper 0 Upper 0 Upper 130 Speed 0
Calibration
Request
Rule Spline
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user NOTE – TruCal TRUCAL LP(A) 21 Level 1 is not used for Calibration (Only 5 levels including blank allowed )
Order information Cat. No. 1 7139 .. .. …
Notes
1. Please refer to the package insert for Lp(a) FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Magnesium XL FS Mindray BS300
Parameters
No # Test MG Method Endpoint Direction Ascend
Unit mg/dl Decimals 2 Prim. Wave 546 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Magnesium
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0.05 Factor 0
Upper 0 Upper 0 Upper 5.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 4610 .. .. …
Notes
1. Please refer to the package insert for Magnesium XL FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
May 2006/2
CRP U-hs Universal (U) application Mindray BS300
Parameters
No # Test CRP U Method Endpoint Direction Ascent
Unit mg/l Decimals 2 Prim. Wave 510 Sec Wave 0
Sample Vol 2.5 R1 Vol. 180 R2 Vol. 180 Line Limit 20
Response
Incubation 1 Antigen Check Antigen Lower 0 Full Name CRP U
Reaction 1 - 14 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 0 Speed 0
Calibration
Request
Rule 5P logitlog
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user NOTE – TruCal CRP U Level 4 is not used for Calibration (Only 5 levels including blank allowed )
Order information Cat. No. 1 7045 .. .. …
Notes
1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
May 2006/2
CRP U-hs High sensitive (hs) application Mindray BS300
Parameters
No # Test CRP hs Method Endpoint Direction Ascent
Unit mg/dl Decimals 2 Prim. Wave 510 Sec Wave 0
Sample Vol 18 R1 Vol. 180 R2 Vol. 180 Line Limit 20
Response
Incubation 1 Antigen Check Antigen Lower 0 Full Name CRP hs
Reaction 1 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 0 Speed 0
Calibration
Request
Rule 5P logitlog
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user NOTE – TruCal CRP hs Level 4 is not used for Calibration (Only 5 levels including blank allowed )
Order information Cat. No. 1 7045 .. .. …
Notes
1. Please refer to the package insert for CRP U-hs for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
Total Protein FS Mindray BS300
Parameters
No # Test TP Method Endpoint Direction Ascend
Unit g/dl Decimals 1 Prim. Wave 546 Sec Wave 0
Sample Vol 6 R1 Vol. 240 R2 Vol. 60 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Total Protein
Reaction -2 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 15.0 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 2311 .. .. …
Notes
1. Please refer to the package insert for Toatl Protein FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
Total Protein UC FS Mindray BS300
Parameters
No # Test TPU Method Endpoint Direction Ascend
Unit mg/l Decimals 0 Prim. Wave 578 Sec Wave 700
Sample Vol 5 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name TPU
Reaction 0 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 20 Factor 0
Upper 0 Upper 0 Upper 300 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 0210 .. .. …
Notes
1. Please refer to the package insert for Total Protein UC FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
ASAT (GOT) FS (IFCC mod.) Mindray BS300
Parameters
No # Test ASAT Method Kinetic Direction Descend
Unit U/l Decimals 1 Prim. Wave 340 Sec Wave 405
Sample Vol 15 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name ASAT
Reaction 6 - 25 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 450 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 2601 .. .. …
Notes
1. Please refer to the package insert for ASAT (GOT) FS (IFCC mod.) for detailed information about the test on the
following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
ALAT (GPT) FS (IFCC mod.) Mindray BS300
Parameters
No # Test ALAT Method Kinetic Direction Descend
Unit U/l Decimals 1 Prim. Wave 340 Sec Wave 405
Sample Vol 15 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name ALAT
Reaction 8 - 23 Substrate 2.5 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 450 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 2701 .. .. …
Notes
1. Please refer to the package insert for ALAT (GPT) FS (IFCC mod.) for detailed information about the test on the
following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
December 2006/2
Triglcerides FS Mindray BS300
Parameters
No # Test TRIG Method Endpoint Direction Ascend
Unit mg/dl Decimals 0 Prim. Wave 510 Sec Wave 700
Sample Vol 3 R1 Vol. 250 R2 Vol. 0 Line Limit 20
Response
Incubation 0 Antigen Check Antigen Lower 0 Full Name Triglycerides
Reaction 0 - 5 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 1 Factor 0
Upper 0 Upper 0 Upper 1000 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 5760 .. .. …
Notes
1. Please refer to the package insert for Triglycerides FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.
April 2006/2
Urea FS Mindray BS300
Parameters
No # Test Urea Method Kinetic Direction Descend
Unit mg/dl Decimals 1 Prim. Wave 340 Sec Wave 405
Sample Vol 2 R1 Vol. 200 R2 Vol. 50 Line Limit 20
Response
Incubation 25 Antigen Check Antigen Lower 0 Full Name Urea
Reaction 3 - 10 Substrate 0 Upper 0 Print No. #
R1 Blank Mix R Blank Linearity
Lower 0 Lower 0 Lower 2 Factor 0
Upper 0 Upper 0 Upper 300 Speed 0
Calibration
Request
Rule Two point Linear
Replicates 2 Difference 0
Interval 0 Blank Response 0 0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0
# entered by the user
Order information Cat. No. 1 3101 .. .. …
Notes
1. Please refer to the package insert for Urea FS for detailed information about the test on the following:
Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature
2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.
**This application proposal is for guidelines only. To avoid misinterpretation measured results have to be validated and assessed with caution.