Date post: | 01-Apr-2015 |
Category: |
Documents |
Upload: | lawrence-francy |
View: | 228 times |
Download: | 0 times |
CPAP/PSV
Objectives
Understand the difference between CPAP, PEEP and PSV
Know the breath types with PSV: supported breaths Know the phase variables of the PSV: trigger, limit
and cycle Understand the ETS cycling mechanisms: flow and
time Setting the Level of Pressure Support Adjustment of pressure support Using PSV with different modes: SIMV Using PSV for weaning Advantages and disadvantages of PSV
CPAP
Continuous Positive Airway Pressure: CPAP
Ventilators can provide CPAP for spontaneously breathing patients Helpful for improving oxygenation in
patients with refractory hypoxemia and a low FRC
CPAP setting is adjusted to provide the best oxygenation with the lowest positive pressure and the lowest FiO2
Spontaneous Breath
Time (sec)
Flow(L/m)
Pressure(cm H2O)
Volume(mL)
Spontaneous Breathing
Time (sec)
Flow(L/m)
Pressure(cm H2O)
Volume(mL)
CPAP Level
Continuous Positive Airway Pressure (CPAP)
CPAP
CPAP
CPAP
CPAP
PEEP
PEEP: Positive End Expiratory Pressure
The term PEEP is defined as positive pressure at the end of exhalation during either spontaneous breathing or mechanical ventilation. However, use of the term commonly implies that the patient is also receiving mandatory breaths from a ventilator.
PEEP becomes the baseline variable during mechanical ventilation
PEEP
Helps prevent early airway closure and alveolar collapse and the end of expiration by increasing (and normalizing) the functional residual capacity (FRC) of the lungs
Facilitates better oxygenation
NOTE: PEEP is intended to improve oxygenation, not to provide ventilation, which is the movement of air into the lungs followed by exhalation
Pressure Support VentilationPhase Variables, Breath Type and Sequence
Pressure Support Ventilation: PSV
Patient triggered, pressure targeted, flow cycled mode of ventilation
Requires a patient with a consistent spontaneous respiratory pattern
The ventilator provides a constant pressure during inspiration once it senses that the patient has made an inspiratory effort
Patient Trigger
Pressure Limit
Flow Cycled
Flow Cycling During PSV
Flow cycling occurs when the ventilator detects a decreasing flow, which represents the end of inspiration
This point is a percentage of peak flow measured during inspiration PB 7200 – 5 L/min Bear 1000 – 25% of peak flow Servo 300 – 5% of peak flow
No single flow-cycle percent is right for all patients
Flow Cycling During PSV Effect of changes in
termination flow
A: Low percentage (17%)
B: High percentage (57%)
Newer ventilators have an adjustable flow cycle criterion, which can range from 1% - 80%, depending on the ventilator
PSV: Time Cycling
NOTE: During pressure support ventilation (PSV), inspiration ends if the inspiratory time (TI) exceeds a certain value. This most often occurs with a leak in the circuit. For example, a deflated cuff causes a large leak. The flow through the circuit might never drop to the flow cycle criterion required by the ventilator. Therefore, inspiratory flow, if not stopped would continue indefinitely. For this reason, all ventilators that provide pressure support also have a maximum inspiratory time.
Supported Breaths
Pressure Support VentilationThe Mode
Triggered Modes of VentilationPressure Support Ventilation
Control Trigger Limit Target Cycle
Pressure Patient Pressure Flow
Patient Triggered, Pressure Limited, Flow
Cycled Ventilation
PSV
PSV
Time (sec)
Flow(L/m)
Pressure(cm H2O)
Volume(mL)
Set PS Level
Flow Cycling
Better Efforts
Longer Inspiration
Patient Triggered, Flow Cycled, Pressure limited Mode
Pressure Support Ventilation(PSV)
CPAP+PSVCPAP+PSV
Time (sec)
Flow(L/m)
Pressure(cm H2O)
Volume(mL)
Set PS Level
Flow Cycling
Patient Triggered, Flow Cycled, Pressure limited Mode
CPAP Level
Pressure Support VentilationSettings and Adjustment
Setting the Level of Pressure Support
Goal: To provide ventilatory support Spontaneous tidal volume is 10 – 12 mL/Kg of
ideal body weight Maintain spontaneous respiratory rate
<25/min
Goal: To overcome system resistance (ET Tube, circuit, etc.) in the spontaneous or IMV/SIMV mode Set pressure at (PIP – Pplateau) achieved in a
volume breath or at 5 – 10 cm H2O
Setting the Level of Pressure Support
Exercise: Using the PIP and the PPlateau from the pressure waveform below, recommend a pressure support setting for this patient (patient is in VC-SIMV mode) 35
25
Answer: 10 cm H2O
The Results of Your Work
35 cm H2O
10 cm H2O
PS=6, VT= 350
PS=10, VT= 600
Pressure Support VentilationIndications, Advantages and Disadvantages
Pressure Support Ventilation: Indications
Spontaneously breathing patients who require additional ventilatory support to help overcome WOB, CL, Raw Respiratory muscle weakness
Weaning (either by itself or in combination with SIMV)
Pressure Support Ventilation: Advantages
Full to partial ventilatory support Augments the patients spontaneous
VT Decreases the patient’s spontaneous
respiratory rate Decreases patient WOB by
overcoming the resistance of the artificial airway, vent circuit and demand valves
Allows patient control of TI, I, f and VT
Pressure Or Volume-Targeted Ventilation
ADVANTAGES
Full to partial ventilatory support Augments the patients spontaneous
VT Decreases the patient’s spontaneous
respiratory rate Decreases patient WOB by
overcoming the resistance of the artificial airway, vent circuit and demand valves
Allows patient control of TI, Vm , f and VT
Set peak pressure Prevents respiratory muscle atrophy Facilitates weaning Improves patient comfort and
reduces need for sedation May be applied in any mode that
allows spontaneous breathing, e.g., VC-SIMV, PC-SIMV
DISADVANTAGES
Requires consistent spontaneous ventilation
Patients in stand-alone mode should have back-up ventilation
VT variable and dependant on lung characteristics and synchrony
Low exhaled VE Fatigue and tachypnea if PS
level is set too low
PS with Other ModesSIMV
VC-SIMV: +Efforts
VC-SIMV/PSupp
PCV+ PSupp
Thank You