    Medicines and Healthcare products regulatory Agency logo

    Evaluation report

    Department of

    Health logo

    April 2005

    MHRA 05023High power

    electrosurgery review

    update 2005

    best choice best practice. www.mhra.gov.uk nww.medical-devices.nhs.ukE

    VA

    LUATION

    http://www.mhra.gov.uk/http://localhost/var/www/apps/conversion/tmp/scratch_7/nww.medical-devices.nhs.ukhttp://localhost/var/www/apps/conversion/tmp/scratch_7/nww.medical-devices.nhs.ukhttp://www.mhra.gov.uk/
  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    2/33

    About MHRA evaluation reports.

    What you can expect.

    The Device Evaluation Service (DES) aims to provide independent and objective

    evaluations of medical devices available on the UK market. Specialist centres,

    mainly in NHS Trusts, do the evaluations under long term contract to, and in

    accordance with protocols approved by, the Medicines and Healthcare products

    Regulatory Agency (MHRA). The evaluations are usually of a unit supplied by the

    manufacturer. We would expect this unit to be representative of the product on the

    market but cannot guarantee this. Prospective purchasers should satisfy themselves

    with respect to any modifications that might be made to the product type after

    MHRAs evaluation. The reports are intended to supplement, not replace, information

    already available to prospective purchasers.

    The MHRA DES does not have access to any information held by the Agency in its

    capacity as the Competent Authority for the UK, apart from any information alreadyin the public domain. The reports will contain data given by the manufacturer on the

    regulatory status of their devices but, apart from this, they are not an indicator of the

    regulatory status of a product. Occasionally, DES refers products to the regulatory

    arm of the MHRA for considerations of breaches of the legislation governing medical

    devices. DES plays no further part in any regulatory investigation that ensues and

    does not have advance notification of any regulatory action that may follow.

    How to obtain MHRA reports.

    To order evaluation reports, a copy of the publications catalogue or to sign up for our

    e-mail alert service contact:

    MHRA

    Orders Department,

    Room 1207,

    Hannibal House,

    Elephant & Castle,

    London,

    SE1 6TQ.

    Tel: 020 7972 8181.

    Fax: 020 7972 8105.

    E-mail: [email protected]

    Visit www.mhra.gov.uk for a comprehensive list of publications, details of forthcoming

    evaluations, services and contacts.

    Colour reports.

    Full colour versions of all reports published after 2002 are available from the NHSNet

    at:

    nww.medical-devices.nhs.uk

    http://localhost/var/www/apps/conversion/tmp/scratch_7/nww.medical-devices.nhs.ukhttp://localhost/var/www/apps/conversion/tmp/scratch_7/nww.medical-devices.nhs.uk
  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    3/33

    Steven Edwards, Norma Greene, Dr Diane Crawford.

    CEDAR

    Clinical Engineering Device Assessment and Reporting

    Cardiff Medicentre

    University Hospital of Wales

    Heath Park

    Cardiff

    CF14 4UJ

    Tel: 029 20682120

    Fax: 029 20750239

    E-mail: [email protected]

    For more information on CEDAR visit www.cedar.wales.nhs.uk

    High power electrosurgery

    review update 2005

    Crown Copyright 2005

    Apart from any fair dealing for the purposes of research or private study, or criticism, or review, as

    permitted under the Copyright, Designs & Patents Act, 1998, this publication may only be reproduced,

    stored, or transmitted in any form or by any means with the prior permission, in writing, of the

    Controller of Her Majestys Stationery Office (HMSO).

    Information on reproduction outside these terms can be found on the HMSO website

    (www.hmso.gov.uk) or e-mail: [email protected].

    The MHRA is an executive agency of the Department of Health.

    ISBN [1 84182 953 6]

    http://www.cedar.wales.nhs.uk/http://www.cedar.wales.nhs.uk/
  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    4/33

    Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

    Table of principal applications . . . . . . . . . . . . . . . . . . . . 3Table of principal features . . . . . . . . . . . . . . . . . . . . . . . 4

    Bovie IDS300 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Output voltage (Open circuit). . . . . . . . . . . . . . . . . 7

    Leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    Manufacturers comments . . . . . . . . . . . . . . . . . . . 8

    BOWA Arc350. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Output voltage (Open circuit) . . . . . . . . . . . . . . . . 11

    Leakage current . . . . . . . . . . . . . . . . . . . . . . . . . . 11

    Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Manufacturing standards . . . . . . . . . . . . . . . . . . . 12

    Manufacturers comments . . . . . . . . . . . . . . . . . . 12

    EMED ES350 SURGILOGIC . . . . . . . . . . . . . . . . . . . . 13

    Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

    Output voltage (Open circuit). . . . . . . . . . . . . . . . 15

    Leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . 15

    Safety features. . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    Manufacturers comments . . . . . . . . . . . . . . . . . . 16

    Contents

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    5/33

    Martin Maxium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

    Output voltage (Open circuit). . . . . . . . . . . . . . . . 19

    Leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . 19

    Safety features. . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

    Manufacturing standards . . . . . . . . . . . . . . . . . . . 20

    Manufacturers comments . . . . . . . . . . . . . . . . . . 20

    Olympus UES40 SurgMaster . . . . . . . . . . . . . . . . . . . . 21

    Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    Output voltage (Open circuit). . . . . . . . . . . . . . . . 23Leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . 23

    Safety features. . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Manufacturers comments . . . . . . . . . . . . . . . . . . 24

    Valleylab Force EZ - 8C . . . . . . . . . . . . . . . . . . . . . . . . 25

    Power curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    Output voltage (Open circuit). . . . . . . . . . . . . . . . 27

    Leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . 27

    Safety features. . . . . . . . . . . . . . . . . . . . . . . . . . . 27Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    Manufacturers comments . . . . . . . . . . . . . . . . . . 28

    Contents

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    6/33

    Electrosurgery enables surgeons to simultaneously cut and seal tissue and

    manufacturers are continually developing specialist features for minimal access

    surgery. This technology facilitates new ways of working in surgery and can help

    meet government objectives of reduced length of stay for patients in hospital.

    This is an independent source of information on six different electrosurgical units

    (ESU) not included in the review published in September 2004. It should be read in

    conjunction with the previous issues on surgical diathermy: Bipolar electrosurgery

    review 2001 (MHRA 01022); Low/medium electrosurgery review 2002 (MHRA

    02037) and, in particular, the High powered electrosurgery review 2004 (MHRA

    04080).

    The report is divided into two sections. The first section contains two comparative

    tables so the purchaser can quickly identify which ESU best suits their needs, bothin terms of clinical application and technical specification. The second section

    comprises a product summary for each generator. Information supplied by the

    manufacturer and/or UK supplier has been used to compile the comparative tables

    and 'Main features' part of each product summary. Power output, output voltage and

    leakage current data has been obtained from in-house testing, in accordance with

    BS EN 60601-2-2:2001 Medical electrical equipment Part 2-2: Particular

    requirements for the safety of high frequency surgical equipment. All other

    information has been obtained from the operator and service manuals.

    The six devices we evaluated all offered a range of features covering the commonly

    performed general surgical procedures. Our observations below summarise some ofthe key differences between these products. Prospective purchasers are

    encouraged to read the detailed information in this report before selecting devices

    for local assessment.

    The Bovie IDS300 offers a basic suite of output modes and is intuitive to use, as it is

    similar in design to many other commonly found generators. Small lettering and

    buttons on the front panel could lead to user errors, especially in poorly illuminated

    areas.

    The BOWA Arc350 has a good range of operating modes and a useful range of pre-programmed power settings, covering most of the commonly performed surgical

    procedures. However, it is not intuitive to use, requiring frequent reference to the

    operator's manual.

    The EMED ES350 Surgilogic is easy to use and has a good range of output modes.

    Currently high leakage currents exceed the limit set in the manufacturing standard

    and increase the risk, to staff and patients, of alternate site burns. The attachment of

    many laparoscopic accessories is prevented by the absence of an 8mm/4mm power

    output socket.

    Summary

    1

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    7/33

    The Martin Maxium is the second ESU in the UK market to use LCD technology and

    it was found to be easy to use. It offers 37 different pre-programmed output settings,

    which may be confusing for some users.

    The Olympus UES40 Surgmaster is specifically designed for resection under saline

    with the TCRiS resectoscope accessory. It has a good range of operating modes

    and is easy to operate.

    The Valleylab Force EZ-8C is simple to use for most surgical procedures. Specialist

    settings may be obtained by the simultaneous use of several buttons on the front

    panel.

    Summary

    2

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    8/33

    Table of principal applications

    3

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Surgical applications

    BovieIDS300

    BowaArc350

    EmedES350

    MartinMaxium

    OlympusUES4

    0

    ValleylabForce

    EZ-8C

    Cardiac

    Chiropody

    Colposcopy

    Dental & oral

    Dermatology

    ENT

    Endoscopy

    Gasteroenterology

    General surgery

    Genitourinary

    Gynaecology

    Laproscopic

    Maxillofacial

    Microsurgery

    Minor surgery

    Opthalmology

    Orthopaedics

    Neurosurgery

    Paediatric

    Plastic & reconstructive

    Thoracic

    Urology (general)

    Vascular

    Vasectomy

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    9/33

    Table of principal features

    4

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Modes, features & accessories

    BovieIDS300

    BowaArc350

    EmedES350

    MartinMaxium

    OlympusUES4

    0

    ValleylabForceEZ-8C

    Monopolar pure cut

    Monopolar blend cut

    Monopolar soft coag

    Monopolar forced coag

    Monopolar spray coag

    Bipolar cut

    Bipolar coag

    Bipolar auto stop

    Bipolar auto start

    Return electrode moni tor

    Contact quality moni tor

    LCD display

    LED display

    Argon unit

    Smoke evacuation unit

    Suction unit

    Dual foot pedals

    Single foot pedals

    Software upgradable

    Instrument auto-detection

    Pedastal

    Programmable

    Other features not listed

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    10/33

    Description

    The Bovie IDS300 ESU is a straight-forward, general purpose electrosurgical

    generator. The unit comprises, 2 pure cut modes, called cut 1 and 2, 10 blend

    settings, a pinpoint and spray coagulation mode as well as a single bipolar

    coagulation mode. Cut 1 generates a

    constant power over a wide range of

    tissue impedances, cut 2 generates a

    constant power over a smaller impedancerange allowing the surgeon to experience

    a suitable cutting experience in different

    types of tissue. CEDAR Comments:

    Generally the diathermy unit is easy to

    use. However, the buttons on the front

    panel are very small and so is the

    labelling, so it is possible to select the

    incorrect button under certain operating

    conditions.

    Bovie IDS300

    5

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 300W cut Max power: 80W spray coag

    Max power: 80W bipolar

    Fundamental freq: 490kHz

    Modulation freq: 25 - 35 kHz

    Class IIB

    CF type

    Output sockets: 1 monopolar, 1

    bipolar and 1, 8mm pin socket

    Power adjustment by push

    button Programmable with 10 separate memory locations

    Dual monopolar activation is not provided

    Dual bipolar and monopolar activation is not provided

    Bipolar auto start and stop is not provided

    Foot pedal selector switch on the footswitch to allow one pedal to control

    either bipolar or monopolar output, or even a combination of both

    A single bipolar coag mode

    Ten different blended cut settings

    Weight 8.75kg

    Dimensions 311(W) 153(H) 413(L) mm

    Price (ex VAT) 4850.00

    Supplier: Avail Medical

    Address: 8 Ninian Park

    Tame Valley

    Dosthill

    Tamworth

    Staffordshire

    B77 5ES

    Telephone: 01827 288883

    Fax: 01827 288334

    Website:

    Email: [email protected]

    Manufacturer Bovie

    Manufactured in: USA

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    11/33

    6

    Technical assessment

    Power curves

    Bovie IDS300

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    IDS300 Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Pure Cut Power (max) Pure Cut Power (Surgical Setting) Blend Cut Power (max) Blend Cut Power (Surgical Setting)

    IDS300 Coag Mode Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Contact Coag Power (max) Spray Coag Power (max) Contact Coag Power (Surgical Setting) Spray Coag Power (Surgical Setting)

    IDS300 Bipolar Power (RMS) Vs Load

    0

    10

    20

    30

    40

    50

    60

    70

    0 200 400 600 800 1000 1200 1400 1600 1800 2000Test Resistance Ohms

    PowerOutputWatts

    Coag Pow er ( max ) Coag Pow er ( Surgic al Setting)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    12/33

    7

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 115mA for pure cut mode and143mA for spray coagulation mode. Maximum permitted leakage current under the

    test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive leakage current

    increases the risk of alternate site burns.

    Safety Features

    Return electrode monitoring.

    Contact quality monitoring.

    Bovie IDS300

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    IDS300 Spray Coag Voltage Vs PowerSetting

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    0 10 20 30 40 50 60 70 80

    Power Setting Watts

    PeakVoltage

    IDS300 Pure Cut Mode Voltage Vs Power Setting

    0

    100

    200

    300

    400

    500

    600

    700

    800

    0 50 100 150 200 250 300

    Power Setting Watts

    PeakVo

    ltage

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    13/33

    8

    Main features

    Servicing

    Comprehensive cover plan 585.00 per annum. Service includes unlimited use of a

    loan device / RTB Avail Medical.

    Manufacturing standards

    Manufacturer claims compliance with EN60601-1:c1990, EN60601-1-2: 2001 and

    EN60601-2-2:2000.

    CE Mark: Yes Notified body: Semko AB Code: 0413

    Manufacturers comments

    The Bovie IDS 300 is a simple to use digitally controlled Electrosurgical Unit that

    performs well during all surgical procedures. The IDS 300 is a standard day to dayworkhorse for the busy Day Surgical Unit or main theatre complex.

    Bovie IDS300

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Cut 1 300W 300 2500V

    Cut 2 300W 300 1500V

    Blend (max) 120W 300 3300V

    Pinpoint coag 120W 500 3500V

    Spray coag 120W 500 7000V

    Bipolar coag 150W 150 1000V

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    14/33

    Description

    The Bowa Arc 350 is a general purpose generator with a 300W maximum output.

    The unit has a good range of operating modes including a range of bipolar cutting

    modes. The unit also has a selection of pre-programmed user settings that aid the

    use of electrosurgery for specific surgical

    procedures such as TURP. A dual

    footswitch may be used to control either

    the bipolar or monopolar output, or acombination of the two. The footswitch

    must be connected to socket 3 on the

    rear of the machine. Blend 0 is equivalent

    to a pure cut waveform. The power

    managment system is called arc control

    enabling a constant cutting speed to be

    achieved regardless of tissue impedance.

    CEDAR Comments: Generally the

    generator is not intuitive to use and

    consultation with the user manual isrecommended.

    BOWA Arc350

    9

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 300W cut Max power: 120W spray coag

    Max power: 150W bipolar

    Fundamental freq: 330kHz

    Modulation freq: 12.5kHz

    Class IIB

    CF type

    Output sockets: 2 monopolar, 2

    bipolar and 2, 8mm pin sockets

    Power adjustment by push

    button Programmable with 100 separate memory locations

    Dual monopolar activation is not provided

    Dual bipolar and monopolar activation is not provided

    Bipolar auto start and stop

    Foot pedal selector switch on the footswitch to allow one pedal to control

    either bipolar or monopolar output, or a combination of both

    Specialist pre-programmed user settings, specific for surgical procedures

    A range of bipolar cut modes

    Weight 14kg

    Dimensions 430(W) 180(H) 400(L) mm

    Price (ex VAT) 9500.00

    Supplier: Avail Medical

    Address: 8 Ninian Park

    Tame Valley

    Dosthill

    Tamworth

    Staffordshire

    B775ES

    Telephone: 01827 288883

    Fax: 01827 288334

    Website:

    Email: [email protected]

    Manufacturer BOWA

    Manufactured in: Germany

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    15/33

    10

    Technical assessment

    Power curves

    BOWA Arc 350

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Arc 350 Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Ble nd 0 Po wer (ma x) Ble nd 0 Po wer (Su rg ica l Set ting ) Ble nd Cu t 9 Po wer (ma x) Ble nd Cu t 9 Po wer (Su rg ica l Set ting )

    Arc 350 Coag Mode Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Contact Coag Power (max) Spray Coag Power (max) Contact Coag Power (Surgical Sett ing) Contact Coag Power (Surgical Sett ing)

    Arc 350 Bipol ar Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180

    0 200 400 600 800 1000 1200 1400 1600 1800 2000

    Test Resistance Ohms

    PowerOutputWatts

    Cut Power (max) Coag Power (max) Cut Power (Surgical Sett ing) Coag Power (Surgical Sett ing)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    16/33

    11

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 113mA for pure cut mode and119mA for spray coagulation mode. Maximum permitted leakage current under the

    test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive leakage current

    increases the risk of alternate site burns.

    Safety features

    Return electrode monitoring.

    Contact quality monitoring.

    BOWA Arc 350

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Arc 350 Pure Cut Mode Vol tage Vs Power Setti ng

    0

    50

    100

    150

    200

    250

    300

    350

    400

    450

    500

    550

    600

    650

    0 50 100 150 200 250 300

    Power Setting Watts

    PeakVoltage

    Arc 350 Spray Coag Volt age Vs PowerSett ing

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    5000

    5500

    6000

    6500

    0 20 40 60 80 100 120

    Power Setting Watts

    PeakVo

    ltage

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    17/33

    12

    Main features

    Servicing

    Comprehensive cover plan 1050.00 per annum. Service includes unlimited returns

    to the service department and use of a loan unit during repair.

    Manufacturing standards

    Manufacturer claims compliance with ISO 13485:2003.

    CE Mark: Yes Notified body: TUV Code: 0123

    Manufacturers comments

    The BOWA arc350 is a future proof electrosurgical unit, that delivers a superior

    performance in all surgical specialities. The generator has been designed with

    multiple output modes that provides the optimum tissue interaction from the various

    pre-set programs. The unit can be programmed with individual user names. The

    Arc350 will perform well in high impedance tissue from as little as 1 watt.

    BOWA Arc 350

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Pure cut/blend 0 300W 500 560V

    Blend 9 300W 500 850V

    Moderate contact

    coag120W 75 185V

    Forced coag 120W 500 3500V

    Spray coag 120W 500 6000V

    Bipolar cut 150W 500 560V

    Bipolar contact coag 120W 75 182V

    Bipolar ligation 200W 50 210V

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    18/33

    Description

    The ES350 Surgilogic is the first generator on the UK market from EMED. The unit

    is designed for general surgery with a 350W maximum output power. It offers a good

    range of operating modes including a range of bipolar cutting modes, (see page 15).

    The unit has specialist settings that aid

    the use of electrosurgery in difficult

    surgical procedures, such as cutting

    within a saline solution. CEDAR

    Comments: The unit is easy to operate

    with responsive operating buttons in a

    logical design. It should be noted that

    there is no 8mm pin socket so connection

    for laparoscopic instruments may prove

    difficult. The manufacturer claims

    compliance with EN60601-1-2:2000, but

    when tested against this standard it was

    found that the high frequency leakage

    current was in excess of the 150mA limit.

    EMED ES350 SURGILOGIC

    13

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 350W cut Max power: 150W spray coag

    Max power: 120W bipolar

    Fundamental freq: 350kHz

    Modulation freq: 100kHz

    Class IIB

    CF type

    Output sockets: 1 monopolar

    and 1 bipolar. 8mm pin socket is

    not provided

    Power adjustment by pushbutton

    Programmable with 5 separate memory locations

    Dual monopolar activation is not provided

    Dual bipolar and monopolar activation is not provided

    Bipolar auto start and stop

    Foot pedal selector switch on front panel to allow one pedal to control either

    bipolar or monopolar output

    Specialist monopolar mode (K4) for urology and for procedures involving

    cutting through saline

    A range of bipolar cut modes Weight 9.0kg

    Dimensions 420(W) 160(H) 380(L) mm

    Price (ex VAT) 2660.00

    Supplier: Medline

    Address: Unit 33

    Monument Business

    Park

    Warpsgrove Lane

    Chalgrove

    Oxfordshire

    OX44 7RW

    Telephone: 01865 400321

    Fax: 01865 400736

    Website: www.medlineuk.com

    Email: [email protected]

    Manufacturer EMEDManufactured in: Poland

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    19/33

    14

    Technical assessment

    Power curves

    EMED ES350 Surgilogic

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    ES350 Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    P

    owerOutputWatts

    Pure Cut Power (max) Pure Cut Power (Surgical Setting) Blend Cut Power (max) Blend Cut Power (Surgical Setting)

    ES350 Coag Mode Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Contact Coag Power (max) Spray Coag Power (max) Contact Coag Power (Surgical Setting) Contact Coag Power (Surgical Setting)

    ES350 Bipolar Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    0 250 500 750 1000 1250 1500 1750 2000

    Test Resistance Ohms

    PowerOutputWatts

    Cut Power (max) Coag Power (max) Cut Power (Surgical Sett ing) Coag Power (Surgical Sett ing)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    20/33

    15

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 78mA for pure cut mode and164mA for spray coagulation mode. Maximum permitted leakage current under the

    test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive leakage current

    increases the risk of alternate site burns.

    Safety features

    Return electrode monitoring.

    Contact quality monitoring.

    EMED ES350 Surgilogic

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    ES350 Pure Cut Mode Voltage Vs Power Setting

    0

    100

    200

    300

    400

    500

    600

    700

    800

    0 50 100 150 200 250 300 350

    Power Setting Watts

    PeakVoltage

    ES350 Spray Coag Voltage Vs PowerSetting

    0

    1000

    2000

    3000

    4000

    5000

    6000

    7000

    0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150

    Power Setting Watts

    PeakVoltage

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    21/33

    16

    Main features

    Servicing

    Comprehensive cover plan 480.00 per annum. Emergency call out rate 60 per

    hour. Non contract call out rate 85.00 per hour.

    Manufacturing standards

    ISO9001:2000, EN60601-1:1990. Manufacturer claims compliance with EN60601-2-

    2:1998.

    CE Mark: Yes Notified body: ORKI Code: A1947

    Manufacturers comments

    New to the UK market, the EMED range of diathermy machines has been designed

    to provide fully featured machines at a competitive price level. The ES350 is a multi-

    purpose, fully equipped system for general theatre use. 5 memory settings plus a

    special programme for Polypectomy and Papillotomy is an additional feature.

    ES350 is also available with argon beam.

    Comments made within this report with regard to high frequency leakage current

    have been noted and modifications will be made to ensure compliance with

    EN60601-1-2:2000.

    EMED ES350 Surgilogic

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Pure cut 350W 300 350V

    Blend 1 180W 300 1000V

    Blend 2 1500W 100 500V

    Soft coag 180W 150 400V

    Forced coag 180W 300 1000V

    Spray coag 150W 1500 3800V

    Bipolar cut 100W 300 320V

    Bipolar coag 120W 40 230V

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    22/33

    Description

    The Maxium is the latest generator from Martin. The unit is a general purpose

    generator with a 400W maximum output power. It offers an extensive range of

    specialist bipolar and monopolar

    operating modes, 37 modes in total.

    These specialist settings aim to aid the

    use of electrosurgery in difficult surgical

    procedures such as coagulating blood

    vessels and performing low hemorrhage

    resection of pedunculated polyps. The

    unit has an automatically controlled

    vessel sealing system called Seal Safe

    when used with the marClampTM series

    bipolar clamp. CEDAR Comments: The

    generator is easy to operate with

    responsive operating buttons in a logical

    LCD screen design. The large number of

    operating modes may lead to someconfusion.

    Martin Maxium

    17

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 400W cut Max power: 150W spray coag

    Max power: 120W bipolar

    Fundamental freq: 300 - 600kHz

    Modulation freq: 18.5kHz

    Class IIB

    CF type

    Output sockets: 2 monopolar

    and 2 bipolar and 2, 8mm pin

    socket

    Power adjustment by rotatingdial

    Programmable with 5 separate memory locations

    Dual monopole activation is provided

    Dual bipolar and monopolar activation is provided

    Bipolar auto start and stop, with Seal Safe auto control vessel sealing

    technology

    Foot pedal selector switch on front screen to allow one pedal to control either

    bipolar or monopolar output

    Over 35 specialist monopolar and bipolar modes for various different

    procedures Weight 8.3kg

    Dimensions 390(W) 182(H) 435(L) mm

    Price (ex VAT) 10500.00

    Supplier: Melyd Medical Ltd

    Address: Unit 1

    21 Kings Avenue

    Prestatyn

    Denbigshire

    LL19 9AA

    Telephone: 01745 853178

    Fax: 01745 888826

    Website: www.melydmedical

    .com

    Email: [email protected]

    Manufacturer Martin

    Manufactured in: Germany

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    23/33

    18

    Technical assessment

    Power curves

    Martin Maxium

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Maxium Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    400

    450

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    Pow

    erOutputWatts

    Pure Cut Power (max) Pure Cut Power (Surgical Setting) Blend Cut Power (max) Blend Cut Power (Surgical Setting)

    Maxium Coag Mode Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Contact Coag Pow er (max) Macro Spray Coag Pow er (max)

    Contac t Coag Pow er (Sur gic al Setting) Mac ro Spr ay Coag Pow er ( Surgic al Setting)

    Maxium Bipolar Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    160

    0 200 400 600 800 1000 1200 1400 1600 1800 2000

    Test Resistance Ohms

    PowerOutputWatts

    Pure Cut Power (max) Macro Coag Power (max) Pure Cut Power (Surgical Setting) Macro Coag Power (Surgical Setting)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    24/33

    19

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 98mA for pure cut mode and

    135mA for the macro spray coagulation mode. Maximum permitted leakage currentunder the test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive

    leakage current increases the risk of alternate site burns.

    Safety features

    Return electrode monitoring.

    Contact quality monitoring.

    Martin Maxium

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Maxium Spray Coag Voltage Vs PowerSetting

    0

    1000

    2000

    3000

    4000

    5000

    6000

    7000

    8000

    9000

    10000

    0 20 40 60 80 100 120 140

    Power Setting Watts

    PeakVoltage

    Maxium Pure Cut Mode Voltage Vs Power Setting

    0

    200

    400

    600

    800

    1000

    1200

    1400

    1600

    0 50 100 150 200 250 300

    Power Setting Watts

    PeakVoltag

    e

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    25/33

    20

    Main features

    In total there are 26 monopolar modes and 11 bipolar modes provided in this unit.

    Servicing

    Option (a) Fully comprehensive cover plan 600.00 per annum. Service includes all

    parts and labour, all emergency callouts and 1 service visit per annum. Option (b)Service contract as above excluding emergency callout charges 350.00 Option (c)

    Service contract inclusive of labour costs and 1 service visit per annum, exclusive of

    parts and emergency call outs 200.00.

    Manufacturing standards

    Manufacturer claims compliance with EN60601-1 and EN60601-2-2.

    CE Mark: Yes Notified body: DQS Code: CE 0297

    Manufacturers comments

    'Manufacturer did not comment'

    Martin Maxium

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Pure cut 400W 400 2500V

    Blend cut 320W 400 3100V

    Super Blend 250W 400 3100V

    Contact coag 250W 100 2800V

    Forced coag 150W 300 5500V

    Macro spray coag 150W 1200

    11000V

    Bipolar pure cut 120W 300 1000V

    Bipolar macro coag 120W 50 650V

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    26/33

    Description

    The UES40 SurgMaster is Olympuss latest electrosurgical unit and was launched in

    2004. The unit comprises a novel TCRis resectoscope, a bipolar type device for

    gynaecological applications which does not require a patient plate. The

    resectoscope is used with physiological saline, which the manufacturer claims

    greatly reduces the risk of TUR syndrome

    and minimises the depth of tissuedenaturation. The URO cutting mode is for

    producing incisions under a non-electrolyte

    solution. The saline setting is for use when

    operating through a physiological saline

    solution usually in conjunction with the

    TCRis resectoscope. CEDAR Comments:

    The unit is intuitive to use and easy to set

    up. TCRis stands for trans cervical

    resection in saline and TURis stands for

    trans urethral resection in saline.

    Olympus UES40 SurgMaster

    21

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 300W cut Max power: 120W spray coag

    Max power: 160W bipolar

    Fundamental freq: 350kHz

    Modulation freq: 100kHz

    Class IIB

    CF type

    Output sockets: 2 monopolar

    with 1, 8mm pin socket and 2

    bipolar sockets

    Power adjustment by pushbutton

    Programmable with 5 separate memory locations

    Dual monopole activation is not provided

    Dual bipolar and monopolar activation is not provided

    Foot pedal selector switch on front panel allows one pedal to control either

    bipolar or monopolar output

    Specialist monopolar modes for urology and for procedures involving cutting

    through saline

    Specialist bipolar modes including cutting and heavy duty coagulation

    Weight 12.0kg Dimensions 350(W) 150(H) 400(L) mm

    Price (ex VAT) 10495.00

    Supplier: KeyMed Ltd

    Address: KeyMed House

    Stock Road

    Southend on Sea

    Essex

    SS2 5QH

    Telephone: 01702 616333

    Fax: 01702 465677

    Website: www.keymed.co.uk

    Email:

    Manufacturer OlympusManufactured in: Japan

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    27/33

    22

    Technical assessment

    Power curves

    Olympus UES40 SurgMaster

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    22

    UES40 Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    Po

    werOutputWatts

    Pure Cut Power (max) Blend Cut Power (max) Pure Cut Power (Surgical Setting) Blend Cut Power (Surgical Setting)

    UES40 Bipolar Power (RMS) Vs Load

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    0 200 400 600 800 1000 1200 1400 1600 1800 2000

    Test Resistance Ohms

    PowerOutputWatts

    Cut Power (max) Soft Coag 2 Power (max) Cut Power (Surgical Setting) Soft Coag 2 Power (Surgical Setting)

    UES40 Coag Mode Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    160

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Coag 1 Power (max) Spray Coag Power (max) Coag 1 Power (Surgical Setting) Spray Coag Power (Surgical Setting)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    28/33

    23

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 74mA for pure cut mode and82mA for spray coagulation mode. Maximum permitted leakage current under the

    test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive leakage current

    increases the risk of alternate site burns.

    Safety features

    Return electrode monitoring.

    Contact quality monitoring.

    Olympus UES40 SurgMaster

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    UES40 Spray Coag Voltage Vs PowerSetting

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    0 20 40 60 80 100 120

    Power Setting Watts

    PeakVoltage

    UES40 Pure Cut Mode Voltage Vs Power Setting

    0

    50

    100

    150

    200

    250

    300

    350

    400

    450

    500

    550

    0 50 100 150 200 250 300

    Power Setting Watts

    PeakVoltage

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    29/33

    24

    Main features

    Servicing

    Comprehensive cover plan 784.00 per annum.

    Manufacturing standards

    Manufacturer claims compliance with ISO9001:2000 and EN60601-2-2.

    CE Mark: Yes Notified body: TUV Code: 0197

    Manufacturers comments

    'Manufacturer did not comment'

    Olympus UES40 SurgMaster

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Pure cut 300W 200 1200V

    Blend 250W 200 1200V

    URO 300W 200 1200V

    Coag 1 120W 200 2200V

    Coag 2 120W 200 3200V

    Spray 120W 200 3300V

    Bipolar cut 90W 100 450V

    Bipolar soft ooag 1 90W 100 220V

    Bipolar soft coag 2 90W 100 220V

    Bipolar coag hard 160W 100 240V

    Saline pure cut 320W 100 500V

    Saline blend cut 320W 100 900V

    Saline coag 1 200W 100 250V

    Saline coag 2 80W 100 880V

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    30/33

    Description

    The Force EZ incorporates the Valleylab Instant Response Technology power

    managment system. When using the bipolar mode, the cut modes, desiccate 2 and

    3 coagulation modes, the output voltage is automatically controlled by a

    microprocessor within the generator to compensate for the changing impedance at

    the site of operation. This leads to the flatter shaped power curves visible in the

    technical assessment (see page 25). The Instant Response Technology is not

    applied to the desiccate 1 and fulgurate 1 and 2 coagulation modes. Dessicate

    refers to a contact coagulation mode and

    fulgurate to a spray type coagulation. Itshould also be noted that dessicate

    modes 1,2,3 and fulgurate modes 1 and 2

    are special modes and are not

    automatically set from the front panel. To

    adjust between dessicate 1,2 and 3 and

    fulgurate 1 and 2 the footswitch button

    and either the fulgurate or dessicate blue

    button must be pushed simultaneously.

    This is described within the operating

    manual. CEDAR comments: The unit is

    intuitive to use and easy to set up, butdoes have hidden surgical mode settings.

    Valleylab Force EZ - 8C

    25

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Summary

    Max power: 300W cut Max power: 120W spray coag

    Max power: 70W bipolar

    Fundamental freq: 393 - 470kHz

    Modulation freq: 30 - 57kHz

    Class IIB

    CF type

    Output sockets: 1 monopolar with

    1, 8mm pin socket and 1 bipolar

    Power adjustment by rotating

    knob Non programmable

    Dual monopolar activation not provided

    Dual bipolar and monopolar activation not provided

    Foot pedal selector switch on front panel allows the same pedal to control

    either bipolar or monopolar output

    Specialist monopolar modes for fulguration and desiccation

    Price (ex VAT) 7564.00

    Supplier: Valleylab

    Address: 154 Fareham Road

    Gosport

    Hampshire

    Chalgrove

    PO13 0AS

    Telephone: 01329 224115

    Fax: 01329 224260

    Website: www.valleylab.com

    Email:

    Manufacturer Tyco HealthcareManufactured in: USA

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    31/33

    26

    Technical assessment

    Power curves

    Valleylab Force EZ-8C

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Force EZ Cut Modes Power (RMS) Vs Load

    0

    50

    100

    150

    200

    250

    300

    350

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohms

    PowerOutputWatts

    Pure Cut Power (max) Blend Cut Power (max) Pure Cut Power (Surgical Setting) Blend Cut Power (Surgical Setting)

    Force EZ Bipolar Power (RMS) Vs Load

    0

    10

    20

    30

    40

    50

    60

    70

    80

    0 200 400 600 800 1000 1200 1400 1600 1800 2000

    Test Resistance Ohms

    PowerOutputWatts

    Coag Power (max) Coag Power (Surgical Setting)

    Force EZ Coag Mode Power (RMS) Vs Load

    0

    20

    40

    60

    80

    100

    120

    140

    0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000

    Test Resistance Ohm)

    PowerOutputWatts

    Contact Coag Power (max) Spray Coag Power (max) Contact Coag Power (Surgical Set ting) Contact Coag Power (Surgical Sett ing)

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    32/33

    27

    Output voltage (Open circuit)

    Leakage current

    Maximum level of high frequency leakage current was 108mA for pure cut mode and100mA for spray coagulation mode. Maximum permitted leakage current under the

    test conditions specified in EN60601-2-2: 2001 is 150mA. Excessive leakage current

    increases the risk of alternate site burns.

    Safety features

    Return electrode monitoring.

    Contact quality monitoring.

    Valleylab Force EZ-8C

    MHRA

    05023Highpowerelectrosurgeryreview

    update2005

    Force EZ Pure Cut Mode Voltage Vs Power Setting

    0

    100

    200

    300

    400

    500

    600

    700

    800

    900

    1000

    1100

    0 50 100 150 200 250 300

    Power Setting Watts

    PeakVoltage

    Force EZ Spray Coag Voltage Vs PowerSetting

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    0 20 40 60 80 100 120

    Power Setting Watts

    PeakVoltage

  • 8/12/2019 05023 High Power Electrosurgery Review Update 2005

    33/33

    Main features

    Servicing

    PPM contract 350.00 per annum for 1-3 units. Extra units charged at 100 per unit

    to a maximum of 3 units.

    Manufacturing standards

    ISO9001:1994, EN60601-1:1995. Manufacturer claims compliance with EN60601-2-

    2:2000.

    CE Mark: Yes Notified body: BSi Code: 0086

    Manufacturers comments

    The 'hidden' modalities should be viewed as an opportunity for a customer, whose

    surgical needs are different to the factory default settings, to be able to customise their

    output if necessary. The factory default settings were chosen as these will meet the

    surgical needs of the surgeon in almost every surgical procedure performed. Very

    occasionally a surgeon will have an unique requirement. The ability to change the

    factory default settings is therefore an option that is available, but rarely used,

    particularly in the UK. In effect the Force EZ has more versatility than the standard

    outputs offered on the front panel if required. The reason that these are 'hidden' is to

    make the unit as simple and easy to use as possible and not to complicate the front

    panel.

    Valleylab Force EZ-8C

    HRA

    05023Highpowerelectrosurgeryreview

    update2005

    Facility Max power Rated load Peak voltage

    Pure cut 300W 300 1000V

    Blend 200W 300 1700V

    Bipolar 70W 100 150V

    Dessicate 1 120W 500 1750V

    Dessicate 2 120W 300 330V

    Dessicate 3 120W 300 550V

    Fulgurate 1 120W 500 3000V

    Fulgurate 2 120W 500 4250V


Recommended