Date post: | 28-Dec-2015 |
Category: |
Documents |
Upload: | camron-booker |
View: | 214 times |
Download: | 1 times |
1
HERSCHEL: PMT status report
• The order to replicate PCB for divider of R1828-01 has been launched on 11-02-2014a) digital file for production (promised by 14-03-2014)b) production of PCB (not started, on critical path)
• Prototype base modifications a) old base like for BLS counters (ready, waiting for R1828 delivery
2pieces)b) new base for modified voltages sequence (waiting for ordered components -
FARNELL)
• Stand for study of PMTs with LEDsa) pulsing LED with signal recording (in use, scope LeCroy WaveRunner
104MXi-A)b) permanent illuminating LED (scheduled for coming week-end)
• Cosmic stand with prototype counters and PMTsa) trigger counters (ready) b) prototype counters (to be assembled during next
week - Raphael)c) oscilloscope “LeCroy WR 104MXi-A” (ready – 350CHF/month)
• Two LEDs to be embedded inside scintillator – transparency monitoring!(to be checked with prototype)
• 30 old PMTs “56AVP” available from UA1 thanks to Raphael and Heinrich!!!
• Extensively checked last few days (first results available)
RD 21/03/2014
2
HERSCHEL: PMT status report
RD 21/03/2014
STAND for PMT TESTS with LEDs
• base for test R1828
• old resistive base for 56AVP (UA1)
• scope LeCroy WR 104MXi-A (1GHz, 4 ch., WindowsXP )
• NIM crate with LED driver and trigger logic
•HV power supply and current monitoring
All 30 PMTs “56AVP” passed check :
• all are alive and working
• relative photocathode quantum efficiency measured
• Gain = A (HV) α measured to be ~12.2
4
HERSCHEL: PMT status report
RD 21/03/2014
Two LEDs allow plastic transmittance monitoring
LED – 1 LED – 2
Aim – to monitor ratio LED1/ LED2 during shut-down
Vacuum pipe
5
HERSCHEL: PMT status report
RD 21/03/2014
OLD PMTs from UA1
For the same LED brightness and HV=1850V the response varies from
min - 35mVmax – 729mV
All 30 “56AVP” are alive – responding to LED pulse.
6
HERSCHEL: PMT status report
RD 21/03/2014
OLD divider from UA1
• Magnetic steel shielding with 2 LEDs on top
• Mu-metal screening
• Electro-static screening @photocathode HV
• Pure resistive divider
Average pulse height .vs. HV (two LED intensity settings)
56AVP #22544
7
HERSCHEL: PMT status report
RD 21/03/2014
Comparison of 30 PMTs “56AVP”
Pulse height spread for
fixed LED intensity and same HV=1850V
•no stability guarantee, •few minutes averaging
8
HERSCHEL: PMT status report
RD 21/03/2014
Single Photo-electron spectra PMT 56AVP # 49667
(~1M sweeps, @2500V)
0
100
200
300
400
500
-10 0 10 20 30 40 50 60 70 80 90 100
Q, pC
LED Pedestal
9
HERSCHEL: PMT status report
RD 21/03/2014
Outcome:
• Divider production under large concern
• Stand for prototype testing ready by next week
• 56AVP could be considered as a back-up solution (more study required)
• Dividers from UA1 to be investigated (if available)
• COMPASS group from Mainz asked to search spare parts for divider housing
• LEDs worth to embed in to scintillator to be able to monitor its transparency
• The coming weeks are crucial to finalize the baseline solution for counters.