+ All Categories
Home > Documents > 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

Date post: 25-Dec-2015
Category:
Upload: alexander-simpson
View: 215 times
Download: 1 times
Share this document with a friend
Popular Tags:
147
1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review
Transcript
Page 1: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

1

RHB FLUORO SYLLABUS

DATA EXCERTS

Created 2008

RT 244 wk 4 / RT255 Review

Page 2: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

2• Ch 1 – pg 1• 5% population have fluoro exams• 53% GI tract (1996 )• 120 sec fluoro ~= ese 5 – 15 rads• Comp to ABD ~= ese 100- 500 mrads• Fluoro used for dynamic studies• USING FLUORO TO POSTION

PATIENTS PRIOR TO TAKING FILM IS PROHIBITED

• Must have fluoro permit or Supervisor & Operators Certificate from state

Page 3: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

3• Ch 2 - review all outlined• Inc factors directly influencing dose:• Ma, Fluoro (0.5 – 5 ) (usually 1 – 3 ma)• Spot film ma (100 – 200)• Kvp ( 80 -130) FOR FLUORO • collimaiton ALWAYS SMALL AS POSSIBLE• Filtration 3.0 mm AL for 125-150kvp• time, (less is better) – 5 Min audible alarm• Doubling time = doubling exp to patient • Pulsed fluoro vs continuous• Image receptor speed (FASTER)• 8:1 grid ratio fluoro

Page 4: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

4

• Ch 2

• LIH or last image hold

• High kvp / low ma

• Lower frame rate

• Long tube to patient distance– 18 “ preferred in stationary (15” minimum)– Mobile – 12” minimum

• Short II to patient distance– moving II 12 to 18 “ away ↑ dose by (30-45%)

• Dead Man Switch – foot pedal

Page 5: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

5• ALLOWABLE EXPOSURE RATES• Before 1974 w/o AEC = 5R/min• After with AEC• Regular mode = 10 R/ hr• BOOST/MAG mode = 20 R/hr• The table top intensity should not exceed 2.2

R/min for each mA of current at 80 kVp

• ROOM LIGHTING• Ambient light low – see image better – less pt

dose• Bright light reduced eye to resolve detail• Otherwise increasing kvp/mas will increase pt

dose

Page 6: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

6Light Levels and Fluoroscopy

Page 7: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

7

• II – 2mm lead equiv = Primary Barrier

• Interlocked – prevents exposure when parked

• SHIELDING requirements (pg 5)

• BUCKLY SLOT COVER – .25 mm lead– bucky moved to end of table

• CURTAIN – in place .25 mm lead– w/o dose can be 500 mrads/hr to operator– Not requried on c-arm mobile units

Page 8: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

8

Page 9: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

9

BUCKY SLOT COVER

.25 MM LEAD

Page 10: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

10

Lead curtain & dose reduction

Page 11: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

11

• Image intensifiers – reduces pt dose

• Review parts & function

• BG = MG x FG

• Quantum Mottle (not enough photons)

• Blotchy, grainy appearance– Caused by too little exposure– Most commonly remedied by increasing mA

Page 12: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

12

Functioning of Image Intensifier

Page 13: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

13

Units of measurement

• INPUT PHOSPHOR – IS MEASURED IN _________________________________

• OUTPUT PHOSPHOR IS MEASURED IN

• ______________________________

Page 14: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

14

BG = MG X FG

• FLUX GAIN – increase of light brightness due to the conversion efficiency of the output screen

• 1 electron = 50 light photons is 50 FG

• Can decrease as II ages

• Output phosphor almost always 1 inch

• Zinc cadnium phosphot

• Flux gain is almost always 50

Page 15: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

15

Image Intensifier

• VACUUM TUBE• ENCASED IN A

LEAD HOUSING • = 2MM PB • (PRIMARY

BARRIER)

Page 16: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

16Contrast

• Contrast –– Subject – changed by contrast media (PE)– Equipment (camera tube, settings on monitor)

• Controlled by amplitude of video signal

• Affected by:– Scattered ionizing radiation– Penumbral light scatter

Page 17: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

17

• Resolution– 525 line TV (old) 1024 (newer digital)– Ability to resolve objects (detail)– Measured in lp/mm– MTF – expression of resolution for imaging

system (modulation transfer function)

• Distortion – size, shape (pg 9)– PINCUSION may comprise 8 – 10 % of

image area

Page 18: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

18

Image distortion

PINCUSHION EFFECT

Page 19: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

19

Size Distortion

• Affected by same parameters as static radiography– Primarily OID– Can be combated by bringing image

intensifier as close to patient as possible

Page 20: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

20

Shape Distortion

• Geometric problems in shape of input screen– Concave shape helps reduce shape

distortion, but does not remove it all– Vignetting or pin cushion effect

Page 21: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

21 • Pg 9

• LAG – blurring occurs from moving II too quickly – can’t capture photons– (amt of time for camera tube to capture)

• VIGNETTING – fall off of brightness at periphery of image = loss of brightness

• II TUBES = MULIT MODE

• BI FOCUS (2 sizes) TriI (3 sizes)

• Smaller surface area= MAG MODE

• ↓ suface area ↑ mag = ↑ patient dose

• II ~= 4 lp/mm Mag ~= 6 lp/mm

Page 22: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

22Intensifier Format and Modes

Note focal point moves farther from output in mag mode

Page 23: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

23• Pg 9

• TV – closed circuit system via video cables

• TV CAMERA/ PICK UP TUBES:

• VIDICON

• PLUMBICON

• IMAGE ORTHOCON

• TRANSFERS IMAGE FROM OUTPUT PHOSPHOR TO TV MONITOR

• CONNECTED BY FIBER OPTICS

Page 24: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

24TV camera and video signal

(II)• VIDICON tube uses antimony trisulphide

• PLUMBICON tube made out of lead oxide,

Page 25: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

25

VIDEO/CAMERA TUBE

• VIDICON MOST COMMOM– LOWEST PATIENT DOSE

• PLUMICON – BETTER RESOLUTION

• ORTHOCON – VERY $$$$– BEST RESOLUTION … but….

• (II & camera tube combined (not used)

Page 26: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

26

Type of TV camera VIDICON TV camera

improvement of contrast improvement of signal to noise ratio high image lag LOWER PATIENT DOSE

PLUMBICON TV camera (suitable for cardiology)

lower image lag (follow up of organ motions)

higher quantum noise level

CCD TV camera (digital fluoroscopy)

digital fluoroscopy spot films are limited in resolution, since they depend on the TV camera (no better than about 2 lp/mm) for a 1000 line TV system

Page 27: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

27

Vidicon (tube) TV Camera

Page 28: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

28

• TV CAMERA TUBE –

• 525 = TOTAL # OF SCAN LINES

Page 29: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

29Basic Componets of “old”

Fluoroscopy “Imaging Chain”Fluoro TUBE

Primary

Radiation PATIENT

EXIT Radiation

Image Intensifier

ABC Image Recording Devices

Fiber Optics OR

105 Photospot

CINE

Cas

sett

e

VIDICON

Camera Tube

CONTROL

UNITTV

Page 30: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

30• PG 10

• CAMERA CONTROL UNIT• Synchronizes the video signal between camera

& monitor

• TV –MONITOR– CRT – Cathode Ray Tube– Much larger than camera tube – but similar function– The electrons are synchronized by the control unit –

so they are of the same intensity and location as the electrons generated by the pick up (camera) tube

Page 31: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

31• CINE/ CINEFLUORGRAPHY PG 10

• Synchronized with camera shutters – no exp while closed

– Movie camera intercepts image• 16 mm and 35 mm formats• Record series of static exposures at high speed• 30 – 60 frames per second• Offer increased resolution

– At the cost of increased patient dose

• FRAMING FREQUENCY– Division of 60 (7.5,15,30,60,90,120)– ↑# frames ↑ patient dose– 2mr/frame ~30f/sec

Page 32: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

32• VIDEO DISC RECORDING• (like a laser disc – size of an LP record ???)

• Combines radiography with fluoro• Image is stored as a single frame on a video

disc recorder - newer uses CD• 95% dose reduction when utilizing video disc

recording during fluoro• VIDEO TAPE RECORDING• Provides a permanent record w/o ↑ pt dose• Instant play back - Real time images• Poor image quality

Page 33: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

33• SPOT FILMS – CASSETTES PG 11

• BETWEEN PT & II• ↑ Ma (100 -200) ↓ time• High resolution

• PHOTOSPOT –• ½ - 1/3 DOSE OF FILM CASSETTES• TAKEN FROM OP OF II• ↑ Ma , ↓ time (LESS THAN CASSETTE)

• ↓ image quality than cassettes

Page 34: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

34CASSETTE SPOT FILMINGvs PHOTOSPOT FILMING

• First type of recording used • 9x9 cassettes then later up to 14x 14• 9 on 1, 4 on 1, 2 on 1• Delay while filming (anatomy still moving)• Radiographic mA - must boost up to • 100 – 200 mA for filming• And moving cassettes around inside tower• Higher patient dose• Replaced by Photospot (f/sec) filming

Page 35: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

35CASSETTE SPOT FILMINGvs PHOTOSPOT FILMING

• Photospot (f/sec) filming –

• Set at control panel from 1 f/sec – 12 f/sec

• Used for rapid sequence:– Upper Esophogram– Voiding Cystourethrograms (Peds)

• Lower patient dose

Page 36: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

36

• ACCESSORIES –• GONAD SHILEDS – 0.5 MM PB

• GRIDS - 8:1 RATIO

• CASSETTES

– front – low z#

–Higher speed = lower pt dose = lower resolution

Page 37: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

37

PATIENT (BODY) = Scatter to RTDOSE influenced by:PhysiquePathologyHabitusType of tissue exposedTissue densityREMNANT RADITIATION hits II(review ESE – tube, II)

Page 38: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

38

• EQUIPMENT REVIEW

• PG 16 – 26

• REVIEW OF FORMULAS TO KNOW:

Page 39: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

39

Intensifier Performance Conversion factor is the ratio of output phosphor

image luminance (candelas/m2) to x-ray exposure rate entering the image intensifier (mR/second).

• Very difficult to measure: no access to output phosphor

• No absolute performance criteria

Page 40: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

40

Intensifier Brightness Gain (BG)

BG =

Minification Gain x Flux Gain

• mG = (Input Diameter )2

(Output Diameter)2

Page 41: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

41

Intensifier Brightness Gain• Flux Gain (FG): Produced by accelerating the

photoelectrons across a high voltage (>20 keV), thus allowing each electron to produce many more light photons in the output phosphor than was required to eject them from the photcathode.

• Summary: Combining minification and flux gains:

Page 42: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

42

MAG MODE FORMULA

IP OLD SIZE

IP NEW SIZE = %mag

Page 43: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

43

PT dose in MAG MODE

IP OLD SIZE 2

IP NEW SIZE 2 = ↑pt dose

Page 44: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

44

The image intensifier (I.I.)

+

I.I. Input Screen

I.I.Output Screen

Photocathode

Electrode E1

Electrode E3

Electrode E2

Electrons Path

Page 45: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

45

VIDICONVIDICON

FILMFILM PMPM REFERENCEREFERENCEkVkV

CONTROLLERCONTROLLER

X Ray TUBEX Ray TUBE kVkV

GENERAL SCHEME OF FLUOROSCOPYGENERAL SCHEME OF FLUOROSCOPY

Page 46: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

46Modern fluoroscopic system

components

Page 47: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

47 vignetting

• Falloff of brightness at edges of TV screen

• Unequal mag can cause unequal illumination

• Center of screen is brighter than periphery

• Resolution is also better in center

End of Wk 4 RHB review

Page 48: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

48

• PG 23

• VEILING GLARE•   Scatter in the form of

x-rays, light & electrons can

• reduce contrast • of an image intensifier

tube.

Page 49: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

49

• PG 23

• ABC

• Automatic brightness control allows Radiologist to select brightness level on screen by ↑ kVp or ↑ mAs

• Automatic dose control

• Located just beyond the Output Phosphor

• Will adjust according to pt thickness

Page 50: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

50Automatic Brightness Control

• Monitoring Image Brightness– Photocell viewing (portion of) output phosphor– TV signal (voltage proportional to brightness)

• Brightness Control: Generator feedback loop– kVp variable– mA variable/kV override– kV+mA variable – Pulse width variable (cine and pulsed fluoro)– less dose with pulsed vs continous fluoro

Page 51: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

51

• ABS - look at bullets pg 23– Hold image brightness– Ignore info at edge (vignetting)– Keep pt ese to less than 10R/min reg– Can shut off – use manual mode

Page 52: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

52• Pg 24 ABS• Variations with X-ray factors:• Brightness of image varies directly with

mA & approx the square of kVp– kVp ↑ 80 to 88 – (10% change) 10 2 = 100% ↑ in brightness

• kVp ↑ = ↓ image resolution but also• ↓ patient dose • ∆ USE optimal kVp:• to ↑ image resolution & ↓ patient dose

Page 53: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

53

TV

MONITOR

Pg 27

TV Image Quality

Page 54: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

54

• TV monitor made of 1000’s of tiny dots of differing brightness – arranged in horizontal scan lines

• 525 system= total # of line – regardless of screen size

• (one dot at a time transmitted from tube)

• 1000+ systems = double resolution• Resolution/quality determined by:• affected by # of scan lines = BANDPASS of system• Horizontal resolution / Vertical resolution• Contrast / Brightness / Lag

Page 55: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

55

Page 56: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

56TV SYSTEMS

• Images are displayed on the monitor as individual frames – which tricks the eye into thinking the image is in motion (motion integration)

• 15 f/sec – eye can still see previous image• 32/f harder to detect with eye

• Weakest Link - 2 lp /mm resolution• No Permanent Record (cannot make dx)• Real Time

Page 57: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

57

• Bandwidth or Bandpass- total # of cycles/sec

– Product of scan lines, scan rate & frequency

• Horizontal resolution – resolve dots

• ↑ bandwidth = camera tube to turn off & on more times/sec

• ∆ ↑ bandwidth ↑ resolution

Page 58: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

58 TV RESOLUTION-Horizontal

• Along a TV line, resolution is limited by how fast the camera electronic signal and monitor’s electron beam intensity can change from minimum to maximum.

• This is bandwidth. For similar horiz and vertical resolution, need 525 changes (262 ½

• full cycles) per line. • Example (at 30 frames/second):

262 cycles/line x 525 lines/frame x 30 frames/second

Page 59: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

59

KELL FACTOR pg 28

VERTICAL RESOLUTIONABILITY TO RESOLVE OBJECTS SPACED APART IN A VERTICAL DIRECTION

MORE DOTS(GLOBULES) = MORE SCAN LINES = MORE/BETTER RESOLUTION

• RATIO OF VERTICAL RESOLUITON• # OF SCAN LINES

• KELL FACTOR FOR 525 LINE SYSTEM

• IS 0.7

Page 60: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

60

• More on camera tubes – pg 28

• Plumbicon - ♥ cath

• Image Orthocon

• CCD – replaced conventional tubes– Fast discarge time – eliminates image lag

– Repeat info pg 29 -30

Page 61: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

61

• CINE – FRAMING & PT DOSE PG 31• Number of frames per second• Cine – division of 60 (7.5, 15,30,90,120)• Organ if interest determines f/s rate• Patient exposure• Under framing• Exact framing• Overframing• Total overframing

CINE RESULTS IN APPROX 10 x HIGHER DOSE THAN CONV FLUORO

Page 62: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

62

• STATIC IMAGE RECORDING PG 31-32

• Video Disk – LIH (95% dose ↓)

• ABD fluoro ~ 10 R ↓ to 500 mR

• Speed 1image/sec – 30 image/sec

• Best to use in “real time” not needed

• PHOTOSPOT

• 70 MM – 105 MM FILM

• LARGER FILM = ↑ IMAGE QUALITY

• ABOUT ½ DOSE THAN CASSETTES-

• CASSETTES HAVE ↑ IMAGE QUALITY• ?? BOTTOM OF PG 32 (20-50X ↑) ???

Page 63: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

63

• DIGITAL FLUORO

• IMAGE CONVERTED ADC – DAC

• CAN PRINT IMAGE ON FILM

• ADVANTAGES: (pg 34) (more later)

• Mask –mode subtraction

• Post processing

• ?? Resolution of digital images lower than film??

Page 65: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

65• Pg 34 – 35• PULSED FLUORO - ↓ PT/RT DOSE• LOWER FRAME RATE• NOT USED FOR MOVING OBJECTS

• HIGH LEVEL BOOST (Interventional)• Uses 10-20 mA (40 mA)• 10 – 20 X higher pt dose• 10 – 50 R/min dose rate at tabletop• (20R / MIN – 1994 regs limits w/o recording

devices such as video )

Page 66: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

66RECORDING DEVICESRESOLUTION P 542 (3rd ed)

OPTICAL MIRROR – BEST BUT NOT PERMANENT RECORDING MEDIUM

• SPOT FILM CASSETTES 6LP/MM• PHOTO SPOT 105 / 70 • CINE 35 MM / 16 MM• DIGITAL (?) (VS FILM)• VIDEO – VIEWING REALTIME• VIDEO TAPE - PLAYBACK

Page 67: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

67

Pg 35

Digital static real-time 7 th highest short

8th

Page 68: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

68

• Contast Media (pg 36)

• Works because of photoelectric effect

• (absorbtion of higher z# - shows differences in contrast of structures

Page 69: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

69

Page 70: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

70

• Ch 4 – pg 37 Fluoro Exam (repeat info)

↑ ↑ directly proportional to dose (pt & Rt)

• Operator dose & pt dose

• Image brightness & rad dose

• mA, kvp, collimation, filtration, time, TPD

↑ ↓indirectly proportional to dose (pt & Rt)

• Poor image quality

• Room lighting

• Tabletop absorbtion

Page 71: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

71

• Ch 4 – pg 37 Fluoro Exam (repeat info)

Patient & Operator (MD/RT)shielding:

• Gonad Shield

• Bucky slot cover

• 3 φ generators

mA ↓from 5 to 3 = dose ↓40 %– image will not be brighter if field size is larger

Page 72: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

72

• Pg 38

• ↑ kVp – ↑ internal dose of organs (slight)

• But ↓ ESE dose

• Collimation-

• Must be seen when carriage is 14” above table top & shutters fully open

• Open collimation - ↑ pt dose &

• ↑ patient integral dose

Page 73: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

73

• Pg 39 Integral Dose

• Total energy absorbed from the beam

• Unit is GRAM RAD (1 gm rad = 100 ergs)

Page 74: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

74

• Pg 39 Filtration (repeat)

• Minimum 2.5 mm al eq

• Fluoro – 3 mm al eq

• Reduces low energy scatter

• Improves quality of beam

• Inherent + added

• Fluoro kvp ranges from 80 – 120

• TABLETOP INTENSITY OF FLUORO –

• NOT EXCEED 2.2 R/MIN /MA AT 80KVP

Page 75: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

75

• PG 39

• HVL –

• Thickness of material to reduce exposure by ½

• Also used for sheilding

• Ex:

• To reduce 100mr/min to 25 mr/min – need 2 HVL

• ?? How many for 10 mr/min ?????

Page 76: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

76• Pg 40 Exposure Time• ↓ beam on time = ↓ pt exposure

• 5 “looks” @ 12 sec / look ~ =1 min exposure

• ~= 400 mrad / look (@ 5R/min) ?• 5 min timer – audible alarm• Allowable Exposure Rates – repeat• 1974 • with AEC 10 R/min; 20R/min Boost • without AEC 5 R/min

Page 77: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

77 Recording the

Patient DOSE

FLUOROBEAM ON

TIME

Page 78: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

78

• Pg 41

• AEC –

• Must MONITOR tube current & potiental at least ONCE EACH WEEKLY

• (QC – recorded DAILY)

• Using a PHANTOM (H2O or lucite)

– Physicist 1/year w/o aec once/3 years

Page 79: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

79 SSD – TUBE TO SKIN DISTANCETPD –

• FIXED UNITS (pg 41)

• 18” PREFERRED• 15 “ MINIMUM

MOBILE UNITS ( C-ARMS)• 12’ MINIMUM TPD

• Increasing TPD 12 – 18 “

• ESE is ↓ 30 % (45%)???

Page 80: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

80

• Lighting-

• Visual Acuity – percieve fine detail

• ↓ room light ↑ acuity

• Photopic 10 x greater than Scotopic

• Normal viewing 12 – 15 inches

• Integration time – recognition time is 0.2 sec

Page 81: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

81

• Sheilding (again!) pg 42 -43

• Gonadal

• Bucky slot cover

• 3 φ generators note bottom of paragraph pg 43 = no difference when tech adjusted

Page 82: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

82

• Ch 5 – bullets (repeat info)

• Ch 6 – PEDS

• More sensitive to radiation

• Motion – problem with peds

• Artifacts - diapers

• Immobilization

• AEC – contrast filled - may over expose

• (ex cystogram- turn off AEC)

Page 83: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

83

• Ch 6 – PEDS (more)

• Collimation reduced

• (can dramatically ↑ dose when open)

• Shield gonads when possible

• Photospot best recording for children

• Persons holding – scatter use sheilds

• Monitor if frequently exposed”

• Close II to pt

Page 84: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

84KEEP I.I. CLOSE TO

PATIENT

Page 85: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

85RADIATION PROTECTIONThe Patient is the largest

scattering object

• Lower at a 90 DEGREE ANGLE from the patient + PRIMARY BEAM

AT 1 METER DISTANCE -• 1/1000 OF INTENSITY

PRIMARY XRAY or 0.1%

Page 86: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

86

• Ch 7 – Mobile Equipment

• (more later)

• Bullets

• Space to prevent closer than 12” (30cm)

• Provide sheilding for ALL

• Ch 8 Responsibilities of Supervisor

• Fluoro – Permit Holders only operate

Page 87: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

87

PERSONNEL PROTECTIONpg 50

• PROTECTIVE APRONS –

• 0.25 PB = 97% ↓ TO SCATTER

• 0.5 PB = 99.9% ↓ TO SCATTER

• THYROID SHEILDS (0.25 & 0.5)

• GLOVES (0.25 & 0.5)

• Hanging screens, Sliding Sheilds

• Lead glasses (interventional)

Page 88: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

88

• Display of Documents

• CRT/ FLUORO – MUST BE DISPLAYED

• NOTICE TO EMPLOYEES

• Dose – Recorded monthly/quarterly– DOSE LIMITS– WHOLE BODY– EYES – EXTREMITIES (BELOW ELBOW/KNEES)

Page 89: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

89Report at least every quarter

Preserved for a minimum of 3 years (employer)

indefinitely - individual

Page 90: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

90

RHB “RULES” RHB RP PG61

• LICENTIATES OF THE HEALING ARTS(MD, DO, DC, DPM)

• MUST HAVE A • RADIOLOGY SUPERVISOR &

OPERATORS PERMIT & CERTIFICATE• TO OPERATE OR SUPERVISE THE USE

OF X-RAYS ON HUMANS• SUPEVISORS MUST POST THEIR

LICENSES

Page 91: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

91RHB NOTIFICATION (EXP IN 24 HOURS)

(Fluoro – pg 53 & RP Syllabus – pg 68)

• Equipment Purchase – 30 days

• Change of Name /Address 30 days

• Illegal to Deliberate Expose individual

• ALL XRAYS MUST BE ORDERED BY A PHYSICIAN

• VERBAL OR WRITTEN PRESCRIPTION

Page 92: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

92

• Ch 9 - pg 56 (more repeated info)

• Fluoro Permit Required

• ALARA:

• Know exam

• Remove unecessary people

• COMMUNICATE /COLLIMATE/SHIELD– ↓motion/ artifacts

• TECHNIQUE ↑ kvp ↓mas (practical)

• POSITIONING – AP/PA & GOOD!

Page 93: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

93

• Pg 57

• Direct Supervision

• Permit Holders:

• Review list

• Restrictions:

• Perform w/o order

• Make a dx, telling to patient

Page 94: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

94

• Ch. 10 – Health Effects

• Somatic Dose– Measured bone marrow, skin,thyroid, gonad– (not when organ is sheilded)

• Genetic Dose– Future generations

Ch 11 – Biologic Effects (other class)

Page 95: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

95 RHB – Rad Prot – CH. IX p 51• ALARA (no minimum threshold)

• STOCHASTIC EFFECTS –NON TRESHOLD (CA + GENETIC)

• NON STHOCAHSTIC (DETERMINISTIC)SEVERITY OF EFFECTS VARIES WITH RADIATION DOSE (THRESHOLD)

(CATARACTS, SKIN, BONE MARROW, STERILITY

Page 96: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

96

• Ch 11 – Biologic Effects

• Dose rate to tissue

• Total dose

• Type of cell irradiated

• Latent period (no visable effect)

Page 97: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

97

Page 98: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

98

SOMATIC & GENETICSTOCHASTIC VS NON STOCHASTIC

• A = STOCHASTIC• “CHANCE” EFFECTS GENETIC, LEUKEMIA,

CANCERDIAGNOSTIC RADIOLOGY

B= NON-STOCHASTICTHRESHOLD EFFECTSDETERMINISTICSOMATIC EFFECTSSKIN ERYTHEMA,

CATARACTS, STERILITYRAD -MALIGNANCIES

Page 99: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

99

• 1 – LINEAR –

NON THRESHOLD

DIAGNOSTIC RAD

• 2 – LINEAR

THRESHOLD

• 3 – SIGMOIDAL

RAD THERAPY

Page 100: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

100

Page 101: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

101

Page 102: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

102 • Cell Sensitivity to Radiation (pg 64)

• Lymphocytes most sensitive

• Pg 65

• Radiation induced cancers –

• Bone #1, Breast (last) ??

Page 103: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

103 PERSONNEL PROTECTION

CH 12 PG 68 Permissible Occupational Dose

• Annual dose:• 5 Rem / year 50 mSv / year

• Cumulative Dose• 1rem x age 10mSv X age

Page 104: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

104

Page 105: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

105

Personnel Monitoring Devices

•Film Badges•TLD •Pocket Dosimeter•Ring Badge

Page 106: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

106

Personnel Monitoring Devices PG

68-75 ACCURACY

• Film Badges

• TLD

• POSL

• Pocket Dosimeter

• Ring Badge

• 10 mrem

• 5 mrem

• 1 mrem

• ?

Page 107: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

107

fb

• FILM BADGE• TLD/POSL

• POCKET DOSIMETER

• MONTHLY• QUARTERLY

• DAILY

Page 108: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

108

MEASUREMENT

• POCKET DOSIMETER

• NOT ACCURATE

• Or permanent

Page 109: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

109 Gonad shielding & dose

• ♀ receive 3x more dose than

• ♂ for pelvic x-rays

• 1 mm lead will reduce exposure

(primary) by about 50% ♀

• by about 90 – 95 % ♂

Page 110: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

110

• It may be advisable to wear a second personal monitoring device if a worker is:

• 1. performing routine radiological procedures

• 2. pregnant

• 3. a student

• 4. performing special procedure

• examinations

Page 111: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

111

Q = t x ּת

CUTIE PIE

Page 112: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

112 Ppg 75 & 76

Monitoring requirments

Page 113: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

113

• HIGH RADIAITON AREA –

• 100 mRem ( 0.1 rem / (1 msV)– @ 30 cm from the source of radiaton

• RADIAITON AREA –

• RHB: 5 mRem ( 0.005 rem / (.05 msV)– @ 30 cm from the source of radiation

• PUBLIC 2 mrem per week* (STAT)

Page 114: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

114 MONITORING PG 75

• CONTROLLED AREA – Used by occupationaly exposed personnel (monitored)

• 100mrem / WEEK

• UNCONTROLLED AREA – PUBLIC

• 2 mrem per week*

Page 115: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

115 • APPENDIX

• #1 – THE EYE PHYSIOLOGY

• #2 COMPETENCY SHEET

• # 3 Units of Dose

• #4 Time Distance Sheilding

• #5 Rad Innjury

• #6 Pregnancy

• #7 Occupationally Exposed Female

• # 8 Gonad Sheilding

Page 116: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

116 • MORE –

• App # 9 Overview of Regs 89 -96

• QC pg 97 – 113 (more later)• Daily / Semi Annual / Annual reqs

• App 12 ISOEXPOSURE CURVES– pg 115-117

– ORGAN DOSES – PG 118– MORE REGS - to pg 125– Glossary 125 - 160

Page 117: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

117

Over vs under the table fluoro tubes

pg 116 dose to operator

Page 118: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

118

Page 119: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

119 Units of Dose – App #3 pg 79

100 R = 1 C/kg1Rad = 1/100 Gray1 Rem = .01 Sievert(Rad + QF = REM)1 REM = 10 mSv

1 Rem = 1000mRem

Page 120: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

120

Page 121: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

121 Pregnancy & Embryo

Mother –

occupational worker (5 rem)

• Baby – (500 mRem)

• .5 rem/ year .05 rem/month

• 5 mSv .5 mSv / month

Page 122: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

122

Page 123: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

123 • PREGNANCY

• Pregnant Patient - PG 84

Under 5 rad – negligible risk

Risk increases above 15 rad

Recommend abortion (spontaneous) 25 rad

• (“Baby exposure” approx 1/50 ? of ESE)

• www.ntc.gov/NRC/RG/08/08-013.html

Page 124: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

124 GERM CELLS in Females

(present at birth)• HIGHLY RADIOSENSITIVE = DEPENDS

OF STAGE OF DEVELOPMENT• Mature ovum do not divide frequently• (20-30 yrs old - least sensitive)• Immature very sensitive• If exposed ova meets sperm – may

contain damaged chromosomes – passing genetic damage to offspring =

• CONGENTIAL ABNORMALITIES

Page 125: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

125 FETAL EFFECTS

• Most sensitive in first trimester - large number or stem cells

• First two weeks – death by spontaneous abortion

• A dose of 10 rad (.10 gy) - expected death rate occurs at 10% higher than that would normally exist

Page 126: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

126

Page 127: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

127 Declared Pregnant Worker

• Must declare pregnancy – 2 badges provided

• 1 worn at collar (Mother’s exposure)

• 1 worn inside apron at waist level

Page 128: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

128

GONAD SHIELDING pg 87

• MUST BE . 5 MM OF LEAD

• MUST BE USED WHEN GONADS WILL LIE WITHING 5 CM OF THE COLLIMATED AREA (RHB)

• FLAT / CONTACT / SHADOW

Page 129: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

129

Page 130: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

130

RHB- RAD PROT SYLLABUS

Page 131: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

131 • HIGHLIGHTS RE FLUORO & RAD PROT

• PG 43- lymphocytes most depressed

• 300 rads ♀ OVARIES = TEMP STERILITY

• 30 rads ♂ TESTIES = TEMP STERILITY

• PG 45 – REPEAT INFO

Page 132: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

132 Gonad shielding & dose

• ♀ receive 3x more dose than

• ♂ for pelvic x-rays

• 1 mm lead will reduce exposure

(primary) by about 50% ♀

• by about 90 – 95 % ♂

Page 133: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

133

ROOM SHIELDING

• PRIMARY SHIELD – • PRIMARY BEAM DIRECTED AT

WALL• 1/16 LEAD - 7 FEET HIGH

Page 134: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

134

ROOM SHIELDING

• SECONDARY – NO PRIMARY BEAM

• 1/32 LEAD

• CONTROL BOOTH (SECONDARY)

• BEAM SCATTERS 2X BEFORE HITTING

• LEAD WINDOW – 1.5MM LEAD EQ

Page 135: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

135

• OFF FOCUS RADIATION

• & Stray Radiation

• Pg 51 RHB syllabus

Page 136: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

136 LEAKAGE RADIATION

may not EXCEED

•TUBE HOUSING 100mR / HR

@ 1 meter

Page 137: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

137 OFF FOCUS

RADIATION

Page 138: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

138

Page 139: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

139

Page 140: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

140

Page 141: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

141

SHADOW OF

SOMEONE’S HEAD =

OFF FOCUS FROM TUBE

Page 142: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

142 PERSONNEL PROTECTION

pg 52 RP RHB• STANDING BEHIND A PROTECTIVE

PRIMARY (1/16TH pb) BARRIER:

• PRIMARY RADIATION EXPOSURE – 99.87% REDUCED

• PORTABLE BARRIER = 99 % REDUCTION

Page 143: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

143 • PERSONNEL PROTECTION

pg 52 RP RHB

• SCATTER RADIATION

• .25 PB = 97 % ↓

• 0.5 MM PB = 99% ↓

Page 144: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

144 • (VENIPUNCTURE = PG 64)

• AppendiX # 9 ESE DOSE RATES PG84

• PG 87 NOMIGRAM

Page 145: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

145

Why did the bunny die??

BUNNY A

• Received 200 rads

BUNNY B

• Received 200 rads

Page 146: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

146

Why did the bunny die??

BUNNY A

200 rads of X-RAY = 200 RADS

BUNNY B

200 rads of alpha =

4000 rads

Page 147: 1 RHB FLUORO SYLLABUS DATA EXCERTS Created 2008 RT 244 wk 4 / RT255 Review.

147

The END – Be Careful !!!


Recommended