AE Port PE N L PE N L EUT Port CDN M016 M3 CAL U100B 100 Ω 100 Ω 100 Ω 100 Ω CAL U100B AE Port PE N L PE N L EUT Port CDN M016 M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 50 Ω load SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration with a CDN The calibration setup always refers to the type of CDN. The CDN user manuals and the NSG 4070 Control Program show the required setup. (Immunity Menu / Global setup / Show test setup pictures). Examples for the system calibration of CDN M016 and CDN S are shown in the following figures: Used as M2 CDN Used as M3 CDN CDN M016 RF out Amp in < +10 dBm Amp out ch.1 < +27 dBm ch.2 < +20 dBm ch.3 < +20 dBm Power meter NSG 4070 Power USB Tuning StSize StSize StSize Local Back Stop Run Hold 0 . 1 4 7 8 5 2 3 kHz dBm 6 9 MHz dBμV Hz V Enter Step 1 STO FRQ LVL RCL Step 2 Step 3 MOD 2nd Help RF ON/OFF SAR M300 adapter Termination Ground plane 6 dB Attenuator AE Port EUT Port CAL U100B 150 Ω / 50 adapter 50 100 CAL U100B 150 Ω / 50 adapter 100 CDN M016 RF out Amp in < +10 dBm Amp out ch.1 < +27 dBm ch.2 < +20 dBm ch.3 < +20 dBm Power meter NSG 4070 Power USB Tuning StSize StSize StSize Local Back Stop Run Hold 0 . 1 4 7 8 5 2 3 kHz dBm 6 9 MHz dBμV Hz V Enter Step 1 STO FRQ LVL RCL Step 2 Step 3 MOD 2nd Help RF ON/OFF SAR M300 adapter SAR MA31 adapter Termination Ground plane 6 dB Attenuator Switch position M3 Switch position M2 AE Port EUT Port CAL U100B 150 Ω / 50 adapter 50 100 CAL U100B 150 Ω / 50 adapter 100 Power meter ch. 1 Compact generator NSG 4070 with built-in power amplifier Power meter ch. 1 Compact generator NSG 4070 with built-in power amplifier Figure 15: Test setup calibration with CDN M016 (switchable M2 / M3) Figure 16: Setup details with CDN M016
Transcript
Page 1: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

AE Port

PENL

PENL

EUT Port

CDN M016M3

CAL U100B

100 Ω

100 Ω

100 Ω

100 Ω

CAL U100B AE Port

PENL

PENL

EUT Port

CDN M016M2

CAL U100B

CAL U100B

50 Ω load

SAR M300 SAR MA31

50 Ω load

SAR M300 SAR M300

120

NSG 4070B

8.2.15. Test setup calibration with a CDNThe calibration setup always refers to the type of CDN. The CDN user manuals and the NSG 4070 Control Program show the required setup. (Immunity Menu / Global setup / Show test setup pictures). Examples for the system calibration of CDN M016 and CDNSareshowninthefollowingfigures:

Used as M2 CDN

Used as M3 CDN

CDN M016

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

SAR M300 adapterTermination Ground plane

6 dBAttenuator

AE Port EUT Port

CAL U100B150 Ω / 50 Ω adapter

50 Ω

100

Ω

CAL U100B150 Ω / 50 Ω adapter

100

Ω

CDN M016

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

SAR M300 adapterSAR MA31 adapter

Termination Ground plane

6 dBAttenuator

Switch position M3

Switch position M2

AE Port EUT Port

CAL U100B150 Ω / 50 Ω adapter

50 Ω

100

Ω

CAL U100B150 Ω / 50 Ω adapter

100

Ω

Power meter ch. 1

Compact generator NSG 4070with built-in power amplifier

Power meter ch. 1

Compact generator NSG 4070with built-in power amplifier

Figure 15: Test setup calibration with CDN M016 (switchable M2 / M3)

Figure 16: Setup details with CDN M016

Page 2: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

121

85-253100 E07

8.2.16. EUT test setup with CDNAfter calibration the calibration adapter has to be removed from the setup. The EUT must be connected through the CDN. One general example for the test setup with EUT is shown below:

CDN S

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Ground plane

6 dBAttenuator

AE Port EUT Port

CAL U100B150 Ω / 50 Ω adapter

100

Ω

No load oradapter required

Common-mode adapter Power meter ch. 1

Compact generator NSG 4070with built-in power amplifier

Figure 17: Test setup calibration with CDN S

Compact generator NSG 4070with built-in power amplifier

Auxiliaryequipment

EUT

AE

Equipmentunder test

Ground plane Insulating

AE Port EUT Port

CDN

6 dB attenuator

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Figure 18: Test setup with EUT

Page 3: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

122

NSG 4070B

8.2.17. Test setup calibration with EM clampThe NSG 4070 is connected via a 6 dB attenuator to the RF port of the EM clamp. The EUT port of the EM clampisconnectedwitha150Ωto50Ωadaptertopowermeterchannel1.TheAEportisterminatedwith150Ω.Bothadaptersneedtobeconnectedwiththesuppliedcableofthecalibrationjig.Theclampneedstobeclickedonthiscable.Thesetupforthesystemcalibrationisshowninfigurebelow:

8.2.18. EUT test setup with EM clampAftercalibrationthe150Ωto50Ωadapter/150Ωloadhastoberemovedfromthesetup.TheEUTmustbeconnected through the EM clamp. One general example for the test setup with EUT is shown below:

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

EUTAE

KEMZ 801A10 kHz ... 1000 MHz

EM clamp / EM-KoppelzangeKEMZ 801A

EUT

AE-side EUT-side

Auxiliaryequipment

Equipmentunder test

Insulating Reference ground plane Insulating

6 dB attenuator

Compact generator NSG 4070with built-in power amplifier

Figure 19: Test setup calibration according IEC / EN 61000-4-6 with EM clamp

Figure 20: Test setup with EUT according IEC / EN 61000-4-6 with EM clamp

Compact generator NSG 4070with built-in power amplifier

AE-side EUT-side

CAL 801A adapter setReference ground plane

6 dB attenuator

Termination

Power meter

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

100 Ω100 Ω50 Ω KEMZ 801A10 kHz ... 1000 MHz

EM clamp / EM-KoppelzangeKEMZ 801A

EUT

Page 4: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

123

85-253100 E07

8.2.19. Test setup calibration with currentinjectionprobe(150Ωsystem)The test generator is connected via 6 dB attenuator to the RF port of the current injection probe. The probe isinsertedina50Ωjig.Thejigisconnectedwitha150Ωto50Ωadaptertoapowermeterwith50 Ωinputimpedance.Theothersideofthejigisterminatedwith150Ω.Thesetupforthesystemcalibrationisshowninthefigurebelow:

Figure 22: Test setup calibration according IEC / EN 61000-4-6 with current injection probe

Figure 21: Test setup with EUT according IEC / EN 61000-4-6 with EM clamp and monitoring probe

Calibrationjig

Current injection probe

Power meter ch. 1

Ground plane

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Compact generator NSG 4070with built-in power amplifier

150 Ω termination

150/50 Ω adapter

6 dB attenuator

The test setup with using the monitoring probe is shown below.

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

AE

KEMZ 801A10 kHz ... 1000 MHz

EM clamp / EM-KoppelzangeKEMZ 801A

EUT

EUTAE-side EUT-side

Auxiliaryequipment

Equipmentunder test

Insulating Reference ground plane InsulatingMonitoring probe MD 4070

6 dB attenuator

Compact generator NSG 4070with built-in power amplifier

Page 5: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

124

NSG 4070B

8.2.20. Test setup calibration with currentinjectionprobe(50Ωsystem)The test generator is connected via 6 dB attenuator to the RF port of the current injection probe. The probe is insertedina50Ωjig.Thejigisconnectedtoapowermeterchannel1.Theothersideofthejigisterminatedwith50Ω.AllIEC/EN61000-4-6setupsoftheNSG4070arebasedonthe150Ω.Aconversionfactorof-9.5 dBisneededbecauseofthe50Ωto150Ωrelation.Forcalibratingina50Ωsystemisneededtoadd-9.5 dBasadditionalattenuation.Testlevelsof10VEMFmayrequireanadditionalattenuatorof10or20dB.Theresultant factor should be used (Example: 20 dB -9.5 dB = 10.5 dB). See chapter 4.3.3 if using the front panel operationoftheNSG4070or5.9.3.7ifusingtheWindowssoftware.Thesetupisshowninthefigurebelow:

Thefollowingfigureshowstheuseofinternalandexternalamplifier(“MIXED“ mode) for having an extended frequency range.

Figure 23: Test setup calibration with currentinjectionprobeina50Ωsystem

Figure 24: Test setup calibration with currentinjectionprobeina50Ωsystem,2amps

Calibrationjig

Current injection probe

Power meter ch. 1

Ground plane

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Compact generator NSG 4070with built-in power amplifier

50 Ω termination

6 dB attenuator

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinput

Power amplifieroutputs

Calibration jig Current injection probe

Ground plane

Interlock

6 dB attenuator

Switch (part of SW 4070)

Switch (part of SW 4070)

2

B

A

1

A

B

DIRECTIONAL COUPLER

Page 6: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

125

85-253100 E07

8.2.21. EUT test setup with current injection probeAfter calibration the jig and the adapters must be removed from the setup. The EUT must be connected through the current injection probe. A general example for the test setup with EUT is shown below:

AttenuatorAuxiliaryequipment

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Compact generator NSG 4070with built-in power amplifier

EUTAE

Equipmentunder test

Insulating Current injection probeGround plane Insulating

Attenuator

Power meter ch.1

Auxiliaryequipment

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Compact generator NSG 4070with built-in power amplifier

EUTAE

Equipmentunder test

InsulatingCurrent injection probe

Monitoring probeGround plane Insulating

Figure 25: Test setup with EUT according IEC / EN 61000-4-6 with current injection probe

Figure 26: Test setup with EUT according IEC / EN 61000-4-6 with current injection probe and monitoring probe

The test setup with using the monitoring probe is shown below.

Page 7: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

126

NSG 4070B

The calibration setup for the monitoring probe with an externalpoweramplifierisshownbelow.

Calibration jig Monitoring probe

Power meter ch. 1

Ground plane

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Compact generator NSG 4070with built-in power amplifier

50 ΩTermination

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF Power meter

ch. 1 response powerch. 2 forward powerch. 3 n. c.

Amp in Amp out

OutputPower amplifier input50 Ω Termination

Directional coupler

NSG 4070RF output

Calibration jig Monitoring probe Ground plane

DIRECTIONAL COUPLER

Figure 27: Calibration setup of the monitoring probe

Figure 28: Calibration setup of the monitoring probe with external power amplifier and direc-tional coupler

8.2.22. Calibration of the monitoring probeTeseq recommends using the MD 4070 monitoring probe due to the operating range of the power meter. The active mode of MD 4070 allows also the operation with low stress levels. Details given in Table 11. The active mode requires to connect the power supply with the MD 4070. Using the whole dynamic range requires to calibrate the MD 4070 in the active and passive mode. The remote controlled switching requires connection with the NSG 4070 user port. The probecalibrationmeasurestheinsertionlossina50Ωsystem.

CAUTION: Usingpoweramplifierswithmorethan100Wattsrequirestoprotectpower meter channel 1 with an attenuator of at least 10 dB.

Page 8: 120 8.2.15. Test setup calibration with a CDN The ... · M2 CAL U100B CAL U100B 50 Ω load SAR M300 SAR MA31 0 Ωload SAR M300 SAR M300 120 NSG 4070B 8.2.15. Test setup calibration

127

85-253100 E07

8.2.23. Setup with external power amplifier and directional couplerAn externalpoweramplifiercanbeconnectedtotheNSG4070asshowninthegeneralexamplebelow.AnexternalpoweramplifierrequiresalwaysanexternaldirectionalcouplerregardlessofthetypeofNSG4070used.ForthecombinationinternalandexternalamplifierisneededtheoptionallyavailableRFswitchSW 4070.Asetupisshowninfigure24.

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

Power meter 50 Ωch. 1 calibration level/ monitoring probech. 2 forward powerch. 3 reverse power

Amp in Amp out

OutputPower amplifier input

From cal. adapter/ monitoring probe

Directional couplerTo coupling device

0.01 ...

DCP 0100

1000 MHz

DIRECTIONAL COUPLER

NSG 4070RF output

Figure 29: Setup with external power amplifier and directional coupler