Design and Performance of a Localized Fiber Optic, Spectroscopic Prototype Device for the Detection of the Metabolic Status of “Vulnerable Plaque”: in-vitro Investigation of Human Carotid Plaque
Transcript
Page 1: 151 performance of a localized fiber optic

Design and Performance of a Localized Fiber Optic, Spectroscopic Prototype Device for the Detection of the

Metabolic Status of “Vulnerable Plaque”:

in-vitro Investigation of Human Carotid Plaque

Page 2: 151 performance of a localized fiber optic

2

OUTLINE• INTRODUCTION

– Problem identification, objectives, specific aims, hypotheses, and background review

• OPTICAL DESIGN– ~ 1 mm3 tissue volume interrogation achieved with optical probe

• METHODS– Laboratory setup– Data collection – Calibration model development

• RESULTS• DISCUSSION/CONCLUSIONS

– Limitations– Future work

Page 3: 151 performance of a localized fiber optic

3

PROBLEM IDENTIFICATION• Atherosclerotic cardiovascular disease 6.3 M deaths / yr

worldwide • Cardiovascular disease #1 killer in the U.S.• 1.5 M myocardial infarctions (MI) / yr in the U.S. • 250,000 / yr sudden cardiac deaths • $111.8 billion / yr health care costs (direct/indirect)• Major risk factors

– Smoking– High blood cholesterol (LDL/HDL ratio)– Physical inactivity– Overweight/Obesity – Diabetes mellitus

Source: American Heart Association. 2002 Heart and Stroke Statistical Update. 2001. http://www.americanheart.org

Page 4: 151 performance of a localized fiber optic

4

Everybody has atherosclerosis, the question is who has vulnerable plaque

Sudden Cardiac DeathAcute MI

VulnerablePlaque(s)

Page 5: 151 performance of a localized fiber optic

5

Unknown Diagnosis – Vulnerable Plaque

• The “precursor” that ultimately ends in acute thrombi (clots) of sudden death MI

• Inflammatory cells found preferentially in vulnerable plaque

• Activity sustained through anaerobic metabolism and lactate production

Page 6: 151 performance of a localized fiber optic

6

Morphology vs. Activity Imaging

Inactive and non-inflamed plaque

Active and

inflamed plaque Appear Similar in

IVUS OCT MRI w/o CM

Morphology

Show Different

Activity

Thermography, Spectroscopy, immunoscintigraphy, MRI with

targeted contrast media…

Page 7: 151 performance of a localized fiber optic

7

HISTOLOGY

LIPID CORETHROMBUS

FIBRO-CALCIFIC

J Am Coll Cardiol. 2001 Sep;38(3):718-23.Am J Pathol. 2000 Oct;157(4):1259-68.

Courtesy of Texas Heart Institute

Page 8: 151 performance of a localized fiber optic

8

LONG-TERM OBJECTIVES

• Develop an optical spectroscopy catheter system to determine the metabolic status of atherosclerotic vessels – No exogenous dyes– No ionizing radiation– Low cost addition to existing cardiac catheterization laboratory

• Locate and identify vulnerable plaque based on metabolic status with optical spectroscopy

Page 9: 151 performance of a localized fiber optic

9

SHORT-TERM GOALS• Demonstrate feasibility in-vitro of optical

spectroscopy to accurately determine metabolic status– Tissue lactate concentration– Tissue pH

Page 10: 151 performance of a localized fiber optic

10

Specification Tissue pH Tissue Lactate Concentration

[moles lactate / gram tissue]

in-vitro Temperature (C)

in-vitro Experimental Stability

< 0.03 change/hr

- <0.4C change/hr

Optical Calibration Range

6.80 – 7.60 2 – 20 32.0 – 38.0

Optical Calibration Accuracy/Prediction Error(R2 / SECV)

0.75 / 0.08pH units

0.75 / 1.0 umoles/ gram

tissue

-

Maximum # of Factors in Optical Calibration Model

5 to 6 independent samples per

factor

5 to 6 independent samples per

factor

-

Page 11: 151 performance of a localized fiber optic

11

SPECIFIC AIMS1) Design a reflectance-based fiber optic probe that uses

visible to near-infrared light optimally to interrogate a small volume of tissue.

2) Estimate the depth penetration of fiber optic probe, based on theory and experiments.

3) Identify major interferents to the optical spectra and tissue reference measurements collection.

4) Collect and analyze fresh tissue from human carotid endarterectomies to create large optical calibration training set while maintaining tissue in a viable physiological state in-vitro.

Page 12: 151 performance of a localized fiber optic

12

HYPOTHESES1) A small fiber optic prototype can make optical

measurements in-vitro for the assessment of metabolically active plaque in a defined region of tissue (< 1 mm3 volume).

2) in-vitro experimental factors can be assessed to their importance in affecting the optical calibration accuracy. The tissue temperature, experiment time, and gross pathology are identified a priori.

3) Mathematical models can be developed which relate the corresponding optical spectra to the individually measured metabolic parameters (tissue pH and lactate concentration) in the presence of inherent pathological variability.

Page 13: 151 performance of a localized fiber optic

13

SPECTROSCOPY BASICSIn general, spectroscopy is the use of the electromagnetic spectrum to perform physical or chemical analysis

Page 14: 151 performance of a localized fiber optic

14

PREVIOUS WORK

• Optical spectroscopy proposed by Lodder (UKY), Feld (MIT) and Jaross (Germany) to characterize morphological properties of atherosclerotic plaques such as thin fibrous cap, large lipid core

Page 15: 151 performance of a localized fiber optic

15

LACTATE AND PLAQUE

• Metabolite produced by activated macrophages• Studies show lactate is present in plaque (Kirk, Zemplenyi)

Anaerobic glycolysis: LDHPyruvate + NADH Lactate + NAD+

Overall anaerobic process: Glucose + 2ADP + 2Pi 2 Lactate + 2ATP +2H20 + 2H+

Page 16: 151 performance of a localized fiber optic

16

NIR Spectrum

0.15

0.2

0.25

0.3

0.35

0.4

0.45

1700 1800 1900 2000 2100 2200 2300 2400 2500

wavelength (nm)

arbi

trary

uni

ts (a

bsor

banc

e)

Near infrared absorbance of lactic acid

Page 17: 151 performance of a localized fiber optic

17

PLAQUE pHLarge scale, ex-vivo study on carotid plaques demonstrated metabolic heterogeneity in grossly pathological areas (Grascu, 1999)

Inflamed regions of plaque are lower in pH in the atherosclerotic Watanabe rabbit and human carotid plaques; plaque pH heterogeneity demonstrated (Naghavi, 2002)

Page 18: 151 performance of a localized fiber optic

18

DR. SOLLER’S LAB• Tissue pH can be measured by NIR spectroscopy in heart muscle

(Soller, Zhang 1998)• Lessons learned: volume of optical measurement >> volume of

reference measurement• Heterogeneity in a “large” tissue volume may be solved with

smaller optical probe

Page 19: 151 performance of a localized fiber optic

OPTICAL DESIGN

Page 20: 151 performance of a localized fiber optic

20

DESIGN PROCESS

Define of optical probe requirements Theoretical considerations of tissue optical

properties Monte Carlo simulations interpretation Building and testing several optical probes Depth penetration assessment

Page 21: 151 performance of a localized fiber optic

21

Optical Catheter System Diagram

• Optical fibers carry light to tissue• Light is reflected and/or backscattered

toward fibers that return light to spectrometer and tissue absorbance calculated

• Catheter geometry and optical coupling important

• Small source-receiver separations: light penetrates tissue while restricting volume interrogated

~2 mm

Light inTo spectrometer

wavelength

Abs

orba

nce

tissueinterface

Page 22: 151 performance of a localized fiber optic

22

THEORY• Monte Carlo Simulations

– Estimate light paths in complex absorbing and scattering medium

– Random events: reflection, absorption, scattering, or transmission

– Define grid geometry, specify tissue optical properties

Probe – tissue interfacen1=n2

dr= 25 m

dz=5 m

SourceDiffuse Reflectance(r)

Page 23: 151 performance of a localized fiber optic

23

0.00E+00

5.00E+01

1.00E+02

1.50E+02

2.00E+02

2.50E+02

3.00E+02

3.50E+02

0.000 0.005 0.010 0.015 0.020 0.025

radius (cm)

diff

use

refle

ctan

ce [1

/cm

2]

514 nm 633 nm 1064 nm

0.00E+00

1.00E+02

2.00E+02

3.00E+02

4.00E+02

5.00E+02

6.00E+02

7.00E+02

0.000 0.005 0.010 0.015 0.020 0.025

radius (cm)

diff

use

refle

ctan

ce (1

/cm

2)

470 nm 633 nm 514 nm 1050 nm 1064 nm

a) b)

normal atheroscleroticDiffuse Reflectance(radius)

Theoretical Depth PenetrationWavelength

(nm)Normal Aorta

(microns)Atherosclerotic Aorta (microns)

470 -- 417.5

514 1063 722

633 1338 1197

1050 1333 1187

1064 -- 927

Page 24: 151 performance of a localized fiber optic

24

OPTICAL EXPERIMENTS

• Compared signal-to-noise ratios (SNR) for several fiber types / configurations– Different core sizes / number of fibers– With or without optical windows– Source-receiver separations

Page 25: 151 performance of a localized fiber optic

25

PROBE GEOMETRY

Large OD

Probe 360 degree illumination w/ optical window

Forward illuminationNo optical window

Final probe W/ optical window

Page 26: 151 performance of a localized fiber optic

26

TISSUE PENETRATION STUDY1) Reference spectra collected for each optical configuration (50, 500 micron

separations).

2) Absorbance spectra collected with n-th slice of 50 micron tissue.

3) Second absorbance spectrum collected with n-th slice plus diffuse reflector. Both absorption and scattering attenuate tissue signal.

Optical Probe

1) 2) 3)

To spectrometer

Diffuse Reflector

~50 um slicesAortic tissue

Light source (source fibers)

Diffuse Reflector

Page 27: 151 performance of a localized fiber optic

27

0

200

400

600

800

1000

420 550 760 970 1450 1750 2250

wavelength (nm)

pene

tratio

n de

pth

(um

)

S-D separation = 0.05 mm S-D separation = 0.5 mm

Page 28: 151 performance of a localized fiber optic

28

FINAL PROBE DESIGN

• Using a source-receiver separation of 50 microns, adequate depth resolution could be achieved in plaque in both the visible and near-infrared

• Increasing the collection fiber core diameter size to 200 microns with improved transmission out to 2400 nm, higher signal-to-noise ratio is achieved by improving the fiber collection area by 4 times and collection efficiency

• Using a 0.5 mm thick quartz optical window fused on the common end, with forward-viewing optics, the signal-to-noise ratio would be further improved across all wavelengths

Page 29: 151 performance of a localized fiber optic

29

Fiber optic probe used for all optical determinations in this study.

1 cm

Page 30: 151 performance of a localized fiber optic

METHODS

Page 31: 151 performance of a localized fiber optic

31

Heating pad

Balance

Fiber optic probe

Gas lineMicro pH and reference junction electrodes

Thermistor

37C, >85% RH

Spectralon 50% reflectance standard

Tissue bath

Laboratory setup for all studies.

Page 32: 151 performance of a localized fiber optic

32

Humidified Incubator maintained at 37°C.

Page 33: 151 performance of a localized fiber optic

33

in-vitro Plaque Validation Study• Minimum Eagle’s Medium (MEM), pH 7.4, 5.6 mM glucose, 26.2 mM

NaHCO3, with non-essential amino acids was used (Invitrogen).

• Media equilibrated with 75% O2 / 5% CO2 gas mixture prior to plaque addition.

• Seven human carotid plaques (UMass Memorial IRB Approval #10041) were collected and placed immediately in 37°C media enclosed in a humidified incubator at 37°C.

• Two plaques that were not placed in the liquid media, only in the humidified air of the incubator, served as controls.

• Measurements were taken with a 0.5 mm OD multi-parameter sensor placed in the tissue (Diametrics, MN).

• Changes in tissue pH, temperature, PO2 and PCO2 over time were analyzed.

Page 34: 151 performance of a localized fiber optic

34

Control plaque with multi-parameter sensor in place.

Page 35: 151 performance of a localized fiber optic

35

Box-whisker plots for ∆pH / hour and ∆temperature / hour (top and bottom, respectively) for the control and test plaques. The change in pH and temperature over time is greater in the controls than the test plaques.

Page 36: 151 performance of a localized fiber optic

36

STABILITY REVIEW• Experiment time fixed max. 4 hrs • in-vitro experimental stability criteria met

– <0.03 pH units/hr and < 0.4°C /hr

• Tissue temperature in media > 32°C to ensure tissue viability

• Tissue values stable and different from media, controls• Plaques in oxygenated media had higher pO2 readings

versus control plaques• Thickness of plaque affected magnitude of pO2 readings• Unable to measure calcified areas over time

Page 37: 151 performance of a localized fiber optic

37

OPTICAL CALIBRATION • ~24 additional human carotid plaques were collected and placed in-vitro.

• Absorbance spectra (667 – 2500 nm) of each area were taken using Nicolet FTIR 670 spectrometer with fiber optic probe (Remspec, MA) for optical lactate determination.

• Tissue biopsies of the same area were taken using a 4-mm punch biopsy and immediately frozen in liquid nitrogen.

• Reference tissue lactate (LA) assayed using micro-enzymatic methods. Values reported as micromole LA per gram wet tissue.

• Matching spectra and reference values modeled by multivariate calibration techniques. R2 and the standard error of cross-validation (SECV) used to assess model accuracy.

Page 38: 151 performance of a localized fiber optic

38

OPTICAL CALIBRATION • Absorbance spectra (400-1100 nm) were collected for a smaller subset

of 14 plaques using a Control Development spectrometer and same optical probe for optical tissue pH determination.

• Reference tissue pH was measured using 750 um diameter micro-pH electrodes.

• Matching spectra and reference values modeled by multivariate calibration techniques. R2 and the standard error of cross-validation (SECV) used to assess model accuracy.

Page 39: 151 performance of a localized fiber optic

39

MODEL DEVELOPMENT

• Partial least-squares, factor analysis– Create calibration with as many points as possible

• Cluster analysis– Investigate (in)homogeneity of spectra

Page 40: 151 performance of a localized fiber optic

RESULTS

Page 41: 151 performance of a localized fiber optic

41

SPECTRA COLLECTION: Lactate• 82 raw absorbance spectra shown below (667-2400 nm)• Key features are water (~970, 1450 and 2000 nm), cholesterol and its

esters (~1750 nm).

“Thrombus/Red”

n=22

“Fatty/Yellow”

n=41

“Calcified/White”

n=19

Page 42: 151 performance of a localized fiber optic

42

SPECTRA COLLECTION: Tissue pH• 48 raw absorbance spectra shown below (400-1100 nm)• Key features are hemoglobin (550 – 575 nm) and water (~970 nm)

absorption

400 500 600 700 800 900 1000

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

wavelength (nm)

Abs

orba

nce

(arb

itrar

y un

its)

400 500 600 700 800 900 10000

0.5

1

1.5

2

2.5

3

wavelength (nm)

Abs

orba

nce

(arb

itrar

y un

its)

400 500 600 700 800 900 10000

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

wavelength (nm)

Abs

orba

nce

(arb

itrar

y un

its)

“Thrombus/Red”

n=11

“Fatty/Yellow”

n=23

“Calcified/White”

n=14

Page 43: 151 performance of a localized fiber optic

43

Histogram of lactate reference measurements3.2 2.7 (mean SD)

n=82

0

5

10

15

20

25

30

35

0 0.2 0.5 1 2 5 8 10 More

lactate concentration (micromoles/gram tissue)

frequ

ency

Page 44: 151 performance of a localized fiber optic

44

Histogram of tissue pH reference measurements7.33 0.21 (mean SD)

n=48

0

2

4

6

8

10

12

14

6.70 6.80 6.90 7.00 7.10 7.20 7.30 7.40 7.50 7.60 7.70 More

pH units

frequ

ency

Page 45: 151 performance of a localized fiber optic

45

REFERENCE MEASUREMENTS

• No spurious correlations between measured variables• Tissue temperature, experiment times within validated

experiment parameters• Pathology subjective

Variables Correlation Coefficient between data

Tissue Lactate – Tissue Temperature

0.03 (n=82)

Tissue pH – Tissue Temperature

0.00004 (n=48)

Tissue Lactate – Tissue pH 0.0008 (n=48)

Page 46: 151 performance of a localized fiber optic

46

RESULTS: Tissue Lactate• 6-Factor model from 17 points• Wavelength regions

contributing to model: – 2030 – 2330 nm

• The R2 of the determination for optical lactate (LA) calibration = 0.83.

• Estimated accuracy ~ 1.4 micromoles LA/gram tissue.

Page 47: 151 performance of a localized fiber optic

47

2000 2050 2100 2150 2200 2250 2300 23501

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

2000 2050 2100 2150 2200 2250 2300 23501.35

1.4

1.45

1.5

1.55

1.6

1.65

1.7

1.75

1.8

1.85

2000 2050 2100 2150 2200 2250 2300 23500.8

0.9

1

1.1

1.2

1.3

1.4

1.5

A

B

C

Clustering solution for 82 spectra for the optical determination of lactate. Cluster A – 45 spectra, B – 31 spectra, and C – 6 spectra. Cluster A contained the first 21 calibration spectra collected.

Page 48: 151 performance of a localized fiber optic

48

RESULTS: Tissue pH• 3-Factor model from 17 points• Wavelength regions contributing

to model: – 1: 400 – 615 nm– 2: 925 – 1890 nm– 3: 2044 – 2342 nm

• The R2 of the determination for optical tissue pH calibration = 0.75.

• Estimated accuracy ~ 0.09 pH units.

Region 1

Region 2 Region 3

6.80

7.00

7.20

7.40

7.60

7.80

6.80 7.00 7.20 7.40 7.60 7.80

electrode pH

NIR

pH

Page 49: 151 performance of a localized fiber optic

49

400 500 600 700 800 900 1000 11000

0.5

1

1.5

2

2.5

400 500 600 700 800 900 1000 11000

0.2

0.4

0.6

0.8

1

A

B

Clustering solution for optical determination of tissue pH. Two clusters: A contains 39 spectra, B contains 9 spectra. The underlying pathology in cluster B was identified as all thrombotic points.

Page 50: 151 performance of a localized fiber optic

50

DISCUSSION

• Lactate model on portion of entire data set– Further factor analysis showed spectra weakly

associated with theoretical lactate peaks – Number of factors in model too high for # of samples

used; need more samples• Tissue pH model on portion of entire data set

– Further factor analysis showed spectra associated with Hb and water peaks, evidence of pH-induced shift

– Number of factors acceptable for # of samples used

Page 51: 151 performance of a localized fiber optic

51

CONCLUSIONS1) A small fiber optic prototype can make optical measurements

in-vitro for the assessment of metabolically active plaque in a defined region of tissue (< 1 mm3 volume).

• Hypothesis accepted

2) in-vitro experimental factors can be assessed to their importance in affecting the optical calibration accuracy. The tissue temperature, experiment time, and gross pathology are identified a priori.

• Hypothesis accepted

Page 52: 151 performance of a localized fiber optic

52

CONCLUSIONS3) Mathematical models can be developed which relate the

corresponding optical spectra to the individually measured metabolic parameters (tissue pH and lactate concentration) in the presence of inherent pathological variability.

• Hypothesis rejected for large n; pending work

– Limited feasibility of models generated

– Pathological variability large

– Unmodeled tissue variability

– Lactate reference method precision

– Optical tissue volume >> real tissue pH heterogeneity

– Long-term spectrometer drift could not be ruled out

Page 53: 151 performance of a localized fiber optic

53

FUTURE WORK

• Considerable in-vitro work needs to continue• Other clustering algorithms, pre-processing methods• Reference lactate measurement precision• Reduce unmodeled variability, better tissue model• Larger data sets• Spectrometer stability • Rigorous acceptance criteria must be met before use

in-vivo animals or humans