2 0 0 8 - 0 4 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology
Transcript
Page 1: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

INJEX Pharma GmbH

INJEX Operation

Needle-Free Injection Technology

Page 2: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of Transporter (I)

- turn the green knurled nut of the Transporter, as shown, anticlockwise, until the cartridge can be placed in the Transporter´s compartment.

Transporter

Page 3: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of Transporter (II)

- take the cartridge and put it into the Transporter´s compartment.

cartridge

Page 4: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of Transporter (III)

- for a safe and firm seat of the cartridge, turn the green knurled nut of the Transporter clockwise, as shown, until the plunger softly touches the rubber end of the cartridge.

Page 5: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of Transporter (IV)

- take the Transporter Adapter out of it´s sterile packaging.

- screw the Transporter Adapter on to the Transporter. Ensure that the adapter needle pierces the rubber seal of the cartridge.

Page 6: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Filling of Ampoule (I)

- take an INJEX-Ampoule out of it´s sterile packaging and press the plunger completely into the INJEX-Ampoule to force the air out of the Ampoule.

Ampoule

plunger

Page 7: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

- open the protective cap of the Transporter Adapter.

- screw the INJEX-Ampoule, clockwise, as shown, into the open end of the Adapter until it fits tightly.

Filling of Ampoule (II)

Page 8: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Filling of Ampoule (III)

- turn the green knurled nut at the upper end of the Transporter, clockwise, as shown, until the desired dose of the pharmaceutical agent is in the INJEX-Ampoule.

Page 9: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

- unscrew the filled INJEX-Ampoule from the Adapter in an anticlockwise direction and close the protective cap.

Filling of Ampoule (IV)

Page 10: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of the Injector (I)

- take the loaded and safeguarded Injector out of the Reset Box.

Injector

Reset-Box

Page 11: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of the Injector (II)

- screw the filled INJEX-Ampoule clockwise into the open end of the Injector (the last round goes a little bit harder, this is normal).

Page 12: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

Preparation of the Injector (III)

- push the green safety ring towards the Ampoule into „Safe off“ position. Now the Injector is ready for use.

Page 13: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

The Injection- hold the Injector under firm pressure perpendicularly to the

injection site (i.e. abdomen), until a clear indentation forms on the skin. Press the trigger and release the injection. Hold the Injector on the injection site for another 1 – 2 sec.

Page 14: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

- unscrew the INJEX-Ampoule and discard it – domestic waste -.

After the Injection (I)

Page 15: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

- push the green safety ring back into the „Safe“ position.

- store the secured and non-reset Injector and its accessories in the INJEX Case.

After the Injection (II)

Page 16: 2008-04 INJEX Pharma GmbH INJEX Operation Needle-Free Injection Technology.

2008-04

INJEX Pharma GmbHBuckower Damm 11412349 Berlin, Germany+49-30-667 915 63www.injex.de