+ All Categories
Home > Documents > 2012 Blood Waived Testing

2012 Blood Waived Testing

Date post: 06-Apr-2018
Category:
Upload: priscila-moreno
View: 225 times
Download: 0 times
Share this document with a friend

of 20

Transcript
  • 8/2/2019 2012 Blood Waived Testing

    1/20

    In addition to this module, will need to review:

    ** OUMC Policy 11-03 Waived Testing** OUMC Policy 11-20 Identification and Labeling of Specimens** Manufacturers Guide for each test

  • 8/2/2019 2012 Blood Waived Testing

    2/20

    Always make sure when performing any of

    these tests that you: Check two patient identifiers.

    Ensure QC has been completed prior to performing

    the test. Use standard precautions when performing test and

    obtaining the specimen.

    Check expiration dates on material that you will be

    using for the test or QC. Explain the procedure or test to the patient.

  • 8/2/2019 2012 Blood Waived Testing

    3/20

    Urine Dipstick/Ketones QC must be completed each time a new bottle of strips is

    opened.

    KOVA Liqua-Trol is to be used to test for normal and abnormal

    readings on first strip of each new bottle and then monthly. Remove test strip from bottle, check expiration date on bottle.

    Completely immerse reagent area in FRESH urine, ensure goodsaturation and remove.

    Use the specific color chart to visually interpret test results at the

    appropriate time as indicated by manufacturers instructions. Record test results in the patients medical record.

    Always follow the manufacturers recommendation whenperforming QC or test.

  • 8/2/2019 2012 Blood Waived Testing

    4/20

    Urine Pregnancy Test (HCG)Allows quick qualitativemeasurement of HCG atbedside from a urine

    specimen. To be performed if patient

    could possibly be pregnant.

    External controls must beperformed every time a newbox of test packs is opened.

  • 8/2/2019 2012 Blood Waived Testing

    5/20

    Urine Pregnancy (HCG)

    Every test pack has a control line thatmust be present for the test to be valid.This is not the same as a positive result.If no control line appears the test is

    invalid and must be discarded.Allow test to stand per manufacturers

    direction. A positive result will have twolines, the control line and another

    indicating the presence of HCG. Look on the package for expiration dates

    and discard expired packages.

  • 8/2/2019 2012 Blood Waived Testing

    6/20

    Nitazine paper to

    check pH Used to check pH in body fluids

    such as semen, amniotic fluid,and gastric aspirates. Use standard precautions.

    Collect fresh specimen. DipNitrazine paper into specimen,

    shake excess fluid off of paper.Refer to color chart to interpretpH.

    Dispose of specimen and paperproperly.

    Do not use expired paper. Donot use specimens inpreservative solution. Do not usea towel or paper napkin to blot

    excess fluid.

  • 8/2/2019 2012 Blood Waived Testing

    7/20

    Fecal Occult BloodUsed for detection of blood in feces.

    Hemoccult card, Hemoccult developer,and stool specimen required.

  • 8/2/2019 2012 Blood Waived Testing

    8/20

    Fecal Occult Blood Obtain stool specimen. Smear a small amount of stool

    onto the Hemoccult card in the boxes marked A and B.Close the flap and allow the Hemoccult card to sit for3-5 minutes to allow penetration through the paper.On the other side of the card, apply 2 drops ofHemoccult developer to each area over the smears.Apply 1 drop of Hemoccult developer between thepositive and negative controls near the bottom of thetest area. Read results in less than 60 seconds. Anytrace of blue on or at the edge of the smear is a positiveresult.

  • 8/2/2019 2012 Blood Waived Testing

    9/20

    Fecal Occult Blood Both Hemoccult card and Hemoccult developer

    must not be expired. Control positive must be

    positive and control negative must be negative. Ifnot, test is invalid and must be discarded. Use onlyHemoccult cards and Hemoccult developer.

    Gastroccult, Hemoccult ICT, and Hemoccult

    SENSA cards and developers are notinterchangeable for this test.

  • 8/2/2019 2012 Blood Waived Testing

    10/20

    Glucometer QC must be done every 24 hours.

    QC solution is good for 90 days

    after opening and should belabeled with expiration date whenthe vial is opened.

    Assure that strips and code key lot

    number match. Strips are good until the date on

    the side of the strip vial.

  • 8/2/2019 2012 Blood Waived Testing

    11/20

    If the meter displays HI, glucose isgreater than 600. If it displays LO,glucose is less than 10.

    OUMC Normal range is 70-110.

    OUMC critical range is 450.When the circumstances indicate that

    you should contact a physician, ensurethat you document the results and full

    name of the physician that younotified.

  • 8/2/2019 2012 Blood Waived Testing

    12/20

    Strep Group A Test

    Used for quick diagnosis at bedside ofStreptococcus Group A infection of thethroat.

    Reagents, tube, tray, and swab provided bymanufacturer.

  • 8/2/2019 2012 Blood Waived Testing

    13/20

    Strep Group A Test Obtain pharyngeal specimen from patient using sterile

    rayon swab provided by manufacturer. Test specimenimmediately or prepare for transport in modified Stuarts

    liquid transport media. Place test tube into tray, keepingtube upright. Add four drops each of Reagent 1 and Reagent2. Insert specimen swab into mixed reagent and twirl swabto mix. Let stand for 1 minute to ensure saturation. Remove

    swab while squeezing tube to drain specimen from swab.Insert test strip into tube. Let stand for 5 minutes. Twored/pink lines indicate a positive result. One red line nearthe C indicates a negative result. One red/pink line nearthe T indicates an invalid result and must be discarded.

  • 8/2/2019 2012 Blood Waived Testing

    14/20

    Strep Group A Test False positives/negatives can be caused by:

    Using swab other than the provided rayon swab

    Obtaining specimen from source other than the back ofthe patients throat

    Using non-approved transport media,

    Using expired equipment,

    Using incorrect reagents,

    Incorrectly mixing reagents, or Removing swab or test strip before the proper amount of

    time has elapsed.

    Positive and negative controls must be tested with

    every kit. Equipment must be stored at 15C-30C.

  • 8/2/2019 2012 Blood Waived Testing

    15/20

    I-Stat Quality Control with the electronic simulator must be

    performed every 8 hours.

    Aqueous cartridge QC must be done monthly. Individual cartridges must be at warm to room

    temperature for 5 minutes. An entire box must be atwarm to room temperature for 1 hour.

    Cartridges are good once brought to room temperaturefor 2 weeks. The expiration date should be writteneither on the box or the individual cartridge.

  • 8/2/2019 2012 Blood Waived Testing

    16/20

    I-Stat Care should be taken not to contaminant the specimen

    during collection. Make sure that sample is free fromIV fluids and that the sample is mixed well to prevent

    separation. If sample is taken outside of the patients room it must

    be first labeled with a patients label.

    Each staff member will receive their unique operator

    ID after training. This will be used when running a testor QC.

    The I-Stat should be docked and download aminimum of once a shift.

  • 8/2/2019 2012 Blood Waived Testing

    17/20

    Biosite The BioSite allows blood to be quickly tested for

    cardiac markers at the bedside.

    Anyone with suspected cardiac issues will be tested bythe BioSite.

    Most commonly, Myoglobin, CK-MB, and Troponinare tested all at the same time. BNP can also be

    checked simultaneously to evaluate CHF.

  • 8/2/2019 2012 Blood Waived Testing

    18/20

    BiositeThe most common problems are using

    the wrong tube for blood specimens

    (purple/lavender must be used) andallowing the calibration to lapse.

    The machine must be calibrated with aspecial cartridge every 24 hours.

  • 8/2/2019 2012 Blood Waived Testing

    19/20

    Biosite To use the BioSite, collect a specimen in a purple

    topped tube. Enter the patients account number

    into the BioSite machine. Use the provideddropper to put blood into the test area of thecartridge. Insert the cartridge and press the blueenter button.

    The cartridge will automatically eject after the testis complete. Press the red cancel button to eject acartridge currently being tested.

  • 8/2/2019 2012 Blood Waived Testing

    20/20

    References: OUMC Policy 11-03 Waived Testing

    OUMC Policy 11-20 Identification and Labeling ofSpecimens

    Manufacturers Guide for each test.

    Core Lab Policy: Point of Care


Recommended