+ All Categories
Home > Documents > 229838882-exp-1-hplc

229838882-exp-1-hplc

Date post: 07-Aug-2018
Category:
Upload: muhammad-nursalam
View: 213 times
Download: 0 times
Share this document with a friend

of 10

Transcript
  • 8/20/2019 229838882-exp-1-hplc

    1/21

     TITLE:

    High Performance Liquid Chromatography (HPLC): Method

    Development

    OBJECTIVE:

    To study development for optimizing a separation of a mixture of 5 compounds

    which are caffeine, acetone, methyl benzoate, phenotate and phenanthrene using

    HPLC by varying the mobile phase composition.

    ABSTRACT

    A further refinement to HPLC method has been developed and validated for the

    determination of 5 compounds which are caffeine, acetone, methyl benzoate,

     phenotate and phenanthrene by varying the mobile phase composition during the

    analysis; this is known as gradient elution. Gradient elution is where the mobile

     phase compostion is change with time during the separation. The method was

    intended to decrease the retention of the later-eluting components so that they elute

    faster, giving a narrower (& taller) peaks for most components and improves the

     peak shape for tailed peaks. In Optimum resolution was achieved by gradient

    elution on an analytical column with the mobile phase consisting of a 

    acetonitrile:water (20:80 v:v at a flow rate of 2.0 mL/min. The retention times of

    caffeine, acetone, methyl benzoate, phenotate and phenanthrene were about 0.783,

    0.864, 2.049, 2.434 and 3.717 min, respectively. Data acquisition was carried out

    using a photo diode array detector in the wavelength 254 nm. Extraction of

    chromatograms was carried out by timed wavelength. Data obtained in these

    studies indicated that the method was suitable for the intended purpose. 

  • 8/20/2019 229838882-exp-1-hplc

    2/21

    PROCEDURE

    A.INSTRUMENT SET UP

    Detector wavelength: 254nm

    Flow rate: 2.00 mL min-1 

    Mobile phase: acetonitrile: water

  • 8/20/2019 229838882-exp-1-hplc

    3/21

     

    B. EFFECT OF MOBILE PHASE ONHPLC SPERATION

    The instrument was set up with mobile phase ratio of acetonitrile:water (50:50 v:v)and the sample is injected into the column.

    This ratio is repeated for three times to

    verify the results.

    After that, the mobile phase composition ischange to acetonitrile:water (70:30 v:v) and

    the sample is injected into the column.

    This ratio is also repeated for three times toverify the results.

    Then, the resolution for both composition iscalculated and compared .

    The suitable composition of mobile phaseof these copmounds is identified.

  • 8/20/2019 229838882-exp-1-hplc

    4/21

  • 8/20/2019 229838882-exp-1-hplc

    5/21

     

    D. SEPERATION USING GRADIENTELUTION

    Based from the seperation above, thegradient elution seperation is performed

    to improved the column efficiency.

    The suitable ratio of mobile phase is setup and the sample is injected into the

    column.

    This method is repeated until the sitableratio of mobile phase is identified and all peak is seperated nicely and short resist

    time.

  • 8/20/2019 229838882-exp-1-hplc

    6/21

    RESULT 

    All the chromatogram for this experiment has been analyzed and it is attached

     behind the lab report. The resolution for isocratic elution for mobile phase

    composition ACN:H2O (50:50 v:v) is tabulated in Table 1.1, 1.2, 1.3 and for

    mobile phase composition ACN:H2O (70:30 v:v) is tabulated in Table 2.1, 2.2 and

    2.3 . While, the average resolution of both mobile phase compositions is tabulated

    in Table 3. Lastly, resolution for gradient elution is tabulated in Table 4. 

    Resolution (Isocratic elution)

    Mobile phase composition: ACN:H2O (50:50 v:v)

    Peak Calculation Resolution, R s 

    1-2

     

    =2.912

    2-3

     

    = 22.618

    3-4

     = 15.776

    4-5

     

    = 37.441

    Table 1.1: Resolutions run 1

  • 8/20/2019 229838882-exp-1-hplc

    7/21

    Peak Calculation Resolution, R s 

    1-2

     

    = 2.161

    2-3

      = 23.273

    3-4

     

    =17.011

    4-5

     

    = 39.873

    Table 1.2: Resolutions run 2

    Peak Calculation Resolution, R s 

    1-2

     

    = 2.286

    2-3

     

    = 25.493

    3-4

     

    =18.525

    4-5

      = 42.599

    Table 1.3: Resolutions run 3

  • 8/20/2019 229838882-exp-1-hplc

    8/21

    Mobile phase composition: ACN:H2O (70:30 v:v)

    Peak Calculation Resolution, R s 

    1-2

     

    = 1.558

    2-3

     

    = 9.396

    3-4

     

    = 6.878

    4-5

     

    = 17.040

    Table 2.1: Resolutions run 1

    Peak Calculation Resolution, R s 

    1-2

     

    = 1.799

    2-3

     

    = 12.213

    3-4

      = 9.492

    4-5

     

    = 28.463

    Table 2.2: Resolutions run 2

  • 8/20/2019 229838882-exp-1-hplc

    9/21

    Peak Calculation Resolution, R s 

    1-2

     

    = 1.571

    2-3

      = 10.050

    3-4

     

    = 7.425

    4-5

     

    = 17.656

    Table 23: Resolutions run 3

    Mobile phase composition Average Resolution, R s

    ACN:H2O (50:50 v:v) 2.21

    ACN:H2O (70:30 v:v) 1.64

    Table 3: average resolution of both mobile phase compositions

    Resolution (Gradient elution)

    Peak Calculation Resolution, R s 

    1-2

     

    = 1.744

    2-3

     

    = 22.854

    3-4

     

    = 6.644

    4-5

     

    =18.744

    Table 4: Resolutions for gradient elution

  • 8/20/2019 229838882-exp-1-hplc

    10/21

    DISCUSSIONS

    HPLC is a technique for separation, identification and quantification of

    components in a mixture. It is especially suitable for compounds which are not

    easily volatilized, thermally unstable and have high molecular weights.

    High performance liquid chromatography is basically a highly improved

    form of column chromatography. Instead of a solvent being allowed to drip

    through a column under gravity, it is forced through under high pressures of up to

    400 atmospheres. That makes it much faster. It also allows us to use a very much

    smaller particle size for the column packing material which gives a much greater

    surface area for interactions between the stationary phase and the molecules

    flowing past it. This allows a much better separation of the components of the

    mixture. The other major improvement over column chromatography concerns the

    detection methods which can be used. These methods are highly automated and

    extremely sensitive.

    According to………….  for the column and the solvent, there are two

    variants in use in HPLC depending on the relative polarity of the solvent and the

    stationary phase which is normal phase HPLC and reversed phase HPLC.

     Normal phase HPLC is essentially just the same as in thin layer

    chromatography or column chromatography. Although it is described as normal, it

    isn't the most commonly used form of HPLC. The column is filled with tiny silica

     particles, and the solvent is non-polar - hexane, for example. A typical column has

    an internal diameter of 4.6 mm (and may be less than that), and a length of 150 to

    250 mm. Polar compounds in the mixture being passed through the column will

    stick longer to the polar silica than non-polar compounds will. The non-polar ones

    will therefore pass more quickly through the column.

  • 8/20/2019 229838882-exp-1-hplc

    11/21

     Next, for the reversed phase HPLC, in this case, the column size is the same,

     but the silica is modified to make it non-polar by attaching long hydrocarbon

    chains to its surface typically with either 8 or 18 carbon atoms in them. A polar

    solvent is used for example, a mixture of water and an alcohol such as methanol.

    Hence, there will be a strong attraction between the polar solvent and polar

    molecules in the mixture being passed through the column. There won't be as much

    attraction between the hydrocarbon chains attached to the silica (the stationary

     phase) and the polar molecules in the solution. Polar molecules in the mixture will

    therefore spend most of their time moving with the solvent.

     Non polar compounds in the mixture will tend to form attractions with the

    hydrocarbon groups because of van der Waals dispersion forces. They will also be

    less soluble in the solvent because of the need to break hydrogen bonds as they

    squeeze in between the water or methanol molecules, for example. They therefore

    spend less time in solution in the solvent and this will slow them down on their

    way through the column. That means, now it is the polar molecules that will travel

    through the column more quickly.

    The liquid phase is pumped at a constant rate to the column packed with the

    stationary phase. Before entering the column the analysis sample is injected into

    the carrier stream. On reaching the column the sample components are selectively

    retained on the basis of physic chemical interactions between the analyte molecules

    and the stationary phase. The mobile phase moving at a steady rate elutes the

    components based on the operating conditions. Detection techniques are employedfor detection and quantification of the eluted components.

    The HPLC schematic diagrams are shown in Figure 1 and HPLC machine is

    shown in Figure 2. In HPLC system, there are about eight important components.

  • 8/20/2019 229838882-exp-1-hplc

    12/21

    Some of the significance and role of each component part of the HPLC system is

    discussed.

    Figure 1: HPLC schematic diagram

    Figure 2: HPLC machine

    First of all, the mobile phase. Mobile phase serves to transport the sample to the

    system. Essential criteria of mobile phase are inertness to the sample components.

    Pure solvents or buffer combinations are commonly used. The mobile phase should

     be free of particulate impurities and degassed before use.

  • 8/20/2019 229838882-exp-1-hplc

    13/21

     Next is mobile phase reservoir. These are inert containers for mobile phase

    storage and transport. Generally transparent glass bottles are used so that so as to

    facilitate visual inspection of mobile phase level inside the container. Stainless

    steel particulate filters are provided inside for removal of particulate impurities in

    the mobile phase if any.

    Other than that, variations in flow rates of the mobile phase effect elution

    time of sample components and result in errors. Pumps function in providing

    constant flow of mobile phase to the column under constant pressure and to

     pressurize the liquid mobile phase

     Next is the injector. Injectors are used to provide constant volume injection

    of sample into the mobile phase stream. Inertness and reproducibility of injection

    are necessary to maintain high level of accuracy.

    Column contains the bed of stationary phase. It is a vital component and

    should be maintained properly as per supplier instructions for getting

    reproducibility separation efficiency run after run.

    Other than column, the column oven is also important part. The variation of

    temperature during the analytical run can result in changes of retention time of the

    separated eluting components. A column oven maintains constant column

    temperature using air circulation. This ensures a constant flow rate of the mobile

     phase through the column

    A detector gives specific response for the components separated by the

    column and also provides the required sensitivity. It has to be independent of any

    changes in mobile phase composition. It is function to detect the presence of

    components as they exit the column and lastly the recorder. The recorders function

    to record the detector signal.

  • 8/20/2019 229838882-exp-1-hplc

    14/21

    Modern HPLC systems are computer based and software controls

    operational parameters such as mobile phase composition, temperature, flow rate,

    injection volume and sequence and also acquisition and treatment of output.

    Those are the main parts of a basic HPLC system more specialized

    equipment might also have solvent selection valves, vacuum degasser, auto

    samplers, column switchers, pre or post column derivatization and fraction

    collectors.

    As we mention earlier, there are two variants which is normal phase

    chromatography and reversed phase chromatography. For this analysis, we used

    reversed phase chromatography. In reversed phase chromatography, the stationary

     phase is non polar and the mobile phase is relative polar. The most polar

    component will elute first, and increasing the mobile phase polarity increase the

    elution time. Hence, because of the sample components interact with both the

    stationary phase and the mobile phase the method development tends to be more

    complex in liquid chromatography

    Interactive mobile phase requires proper equalization intermolecular forces

    among the three members in the separation process which is the solute, the mobile

     phase and the stationary phase in other to get successful chromatography. These

    intermolecular forces are describes in term of the relative polarity of the reactants.

    The polarities of various analytes functional groups are hydrocarbon

  • 8/20/2019 229838882-exp-1-hplc

    15/21

     phase usually cannot compete successfully for the sample components; the

    retention time becomes shorter for practical application. When the situation where

    the polarity of the analyte and the stationary phase are too alike and totally

    different from the mobile phase, thus the retention time becomes too long.

    Due to theories of mobile phase and stationary phase interaction at any given

    set of sample component are impacted, and at best, we can only narrower the

    choice of the stationary has to a general type. Hence, because of this choice, we

    then can perform a series of set trial and error experiment until a satisfactory

    separation is identified.

    During this experiment, a High Performance liquid chromatography (HPLC)

    Agillent G1311A equipped with DA detector, 5 mm reverse phase C18 column and

    10 microliter sample loop was used. At flow rate 2.0 ml/min and detector

    wavelength at 254nm, the mobile phase ratio was set up at 50% acetonitrile and

    50% water at the beginning at the experiment in other to analyze and observe the

    effect of mobile phase composition on the chromatography separation. After the

    standard mixture is injected, the process is run and the peak obtained is analyzed.

    The process is repeated 3 times to get ideal result. Next, the resolution of the

     process is identified by calculating using formula below:

    Where t is the retention time and W is the peak width.

    After performing calculation for resolution as shown in Table 1.1, 1.2, and 1.3 we

    get average resolution of 2.213.

  • 8/20/2019 229838882-exp-1-hplc

    16/21

     Next, with the same requirement of flow rate and detector wavelength, the

    mobile phase composition was changed to 70% acetonitrile and 30% water. The

    same steps have been done and the process is also repeated for 3 times to get ideal

    result. After performing calculation for resolution as shown in Table 2.1, 2.2, and

    2.3 we get average resolution of 1.643.

    Resolution describes the ability of a column to separate the peaks of interest,

    and so the higher the resolution, the easier it is to achieve baseline separation

     between two peaks. Resolution takes into consideration efficiency, selectivity and

    retention. It is useful to relate the resolution to the number of plates in the column,

    the selectivity factor and the retention factors of the two solutes:

    One can improve resolution by improving any one of these parameters. To

    obtain high resolution, the three terms must be maximized. An increase in N, the

    number of theoretical plates, by lengthening the column leads to an increase in

    retention time and increased band broadening which may not be desirable. Instead,

    to increase the number of plates, the height equivalent to a theoretical plate can be

    reduced by reducing the size of the stationary phase particles.

    It is often found that by controlling the capacity factor, k', separations can be

    greatly improved. This can be achieved by changing the composition of the mobile

     phase (in Liquid Chromatography).

  • 8/20/2019 229838882-exp-1-hplc

    17/21

    The selectivity factor, α, can also be manipulated to improve separations.

    When α is close to unity, optimising k' and increasing N is not sufficient to give

    good separation in a reasonable time.

    A value of 1 is the minimum for a measurable separation to occur and to

    allow adequate quantitation. A value of 0.6 is required to discern a valley between

    two equal-height peaks. Values of 1.7 or greater generally are desirable for rugged

    methods. A value of 1.5 is considered to be a baseline separation and ensures the

    most accurate quantitative result.

    By comparing the average resolution for both mobile phase compositions,

    we can see that with mobile phase ratio of 70% acetonitrile and 30% water gives

    the best resolution which is 1.64. It is indicate that the resolution is good and

    efficiency of separation is increase. Meanwhile, for the mobile phase ratio of 50%

    acetonitrile and 50% water, the resolution is 2.21 shows that it has good resolution,

     but the efficiency of the separation is very weak.

    After that, the components in standard mixture is identified when each of the

    component is injected individually by using mobile phase compositions of 70%

    acetonitrile and 30% water as the best composition baseline separation. The

    compound in the standard mixture is identified by comparing the retention time of

    standard mixture with the retention time of individual compound. From the result

    obtained, peak 1 indicate compound of caffeine, peak 2 is acetone, peak 3 is

    methyl benzoate, peak 4 is phenatole and the last peak is phenantrene.

    Other than that, the rate theory of chromatography / Van Deemter equation

    is related in HPLC broadening peak. Below are Van Deemeter equations:

    H = A + B/v + Cv

  • 8/20/2019 229838882-exp-1-hplc

    18/21

    In HPLC the B term can be neglected due to diffusion is 100 times less in

    liquids than in the gas. While the C term is the largest contribution to H. Consider

    the following mobile phase mass transfer coefficient:

    CMv = (f M(k’)d p2/DM)v

    Where d p is particle diameter of packing and DM is mobile phase diffusion

    coefficient. CMv is less if d p is smaller hence greater surface area or the solute

    diffusion coefficient in the mobile phase, DM is larger.

     Next, a gradient elution separation method is performed in other to improve

    te efficiency of the column. Meaning that, isocratic elution is performed with a

    single solvent or constant solvent mixture. If one solvent does not provide

    sufficiently rapid elution of all components, then gradient elution can be used.

    In this case, by increasing amount of water was added to acetonotrile to

    create a continuous gradient. In this experiment, using acetonitrile and water

    gradient, the more hydrophobic components will elute when the mobile phase

    consist mostly of acetonitrile which giving a relatively hydrophobic mobile phase.

    The more hydrophilic compounds will elute under conditions of relatively low

    acetontrile and high water. Often a series of test are performed on the analyte and

    the number of the trial runs has been done in other to find the HPLC method which

    gives the best separations peak. From the gradient elution method, the average

    resolution obtained is 1.74 and it is indicate as good resolution and efficiency of

    separation is high.

    From the chromatogram obtain from both method which are isocratic elution

    method and gradient elution method, there are some differences can be obtain from

    the separation condition. In isocratic elution method, peak width increases with

    retention time linearly according to the equation for N, the number of theoretical

  • 8/20/2019 229838882-exp-1-hplc

    19/21

     plates. This leads to late eluting and peaks get a little bit flat and broad. In isocratic

    elution method, the retention times of caffeine, acetone, methyl benzoate,

     phenotate and phenanthrene were about 0.739, 0.812, 1.318, 1.763 and 3.864 min.

    Meanwhile, the gradient elution method decreases the retention of the later

    eluting components so that they elute faster, giving narrower, taller and sharp

     peaks for most components. The retention times of caffeine, acetone, methyl

     benzoate, phenotate and phenanthrene were about 0.783, 0.864, 2.049, 2.434 and

    3.717 min. This also improves the peak shape for tailed peaks, as the increasing

    concentration of the organic eluent pushes the tailing part of a peak forward. This

    also increases the peak height which where the peak look sharp which is important

    in trace analysis.

    Thus, the gradient elution method gave a shorter overall analysis with

    similar resolution of the critical pair compared to isocratic elution without

    sacrificing repeatability in retention time, peak area and peak height.

    From the result, it shows that almost the entire peak is well separated from

    the neighbor compound. But what was happened is only the peak 1 and the peak 2

    is not well separated. Peak 1 indicates the caffeine compound and the peak 2

    indicate acetone compound. The mobile phase composition used is 20%

    acetonitrile and 80% water. It shows that, just a bit separation has occurred. As we

    know, the quantitative analysis in separation method depends upon direct

    relationship between the area under a peak or height in the chromatogram and the

    amount of compound corresponding to that peak in the analyzed sample.

    Therefore, each peak should be totally resolved from any neighboring peaks.

    There are some factors leads to problem stated above. The mobile phase

    must be degassed properly. It is because mobile phases entrap air from the

  • 8/20/2019 229838882-exp-1-hplc

    20/21

    atmosphere and this trapped air gets released as small bubbles under high pressures

    encountered during the HPLC analysis. Such bubbles can lead to noise in detector

    response or hinder flow of mobile phase through columns. In order to overcome

    such problems degassing of mobile phase becomes essential. It is require a very

    clean & pure HPLC grade solvent to prevent column degradation with impurities

    & to minimize detector background signals from contaminants (usually UV

    transparent). The gas that interference the HPLC analysis, such as N2, O2, and CO2.

    The gas bubbles create difficulties with pumps destabilize pressure, columns bad

    separation & detectors. Therefore we need to degasse the mobile phase in other to

    get good analysis.

    There are few methods in degassing the mobile phase. Vacuum devices

    (vacuum applied to headspace above solvent). Ultra sonication (high frequency

    vibration drives gasses out of solvent) Heating (decreases solubility of gases). He

    sparging. Sparging is a process in which dissolved gases are swept out of a solvent.

    There are some step is carefully taken when doing the experiment. When

    injecting the sample into the loop, the sample volume is no more than volume

    indicated on a loop. Besides, sample injection only with flat end needle to prevent

    damage to the injection port.

    Other than that, the syringe is washed at least 5 times with washing solution

    and wash 3 times with sample before the sample is inejcted to the column to get

    desire peak and no contiminants. When filling the syringe with the sample, the

    maximum value should be 20μL. 

  • 8/20/2019 229838882-exp-1-hplc

    21/21

     

    CONCLUSION

    From the experiment, the concept and method development of optimizing aseparation of a standard compounds using hplc by varying the mobile phase

    composition is studied and determined.

    REFERENCES

    1.  Skoog, Hooler and Nierman, 5th Editon. Principles of Instrumental Analysis.

    Thomson Learning 1998.

    2. 

     Nor’ ashikin, Ruziyati and Mardiana, 2nd Edition. Analytical Separation

    Methods Laboratory Guide


Recommended