2D-CSI in a fraction of the time using multiple receiver coils. Simon J. Doran 1 , Adam J. Schwarz 2 and Martin O. Leach 2 1 Department of Physics, University of Surrey, Guildford, Surrey, UK 2 Institute of Cancer Research & Royal Marsden NHS Trust, Sutton, Surrey, U.K. Acknowledgements. - PowerPoint PPT Presentation
Popular Tags:
18
2D-CSI in a fraction of the time using multiple receiver coils Simon J. Doran 1 , Adam J. Schwarz 2 and Martin O. Leach 2 1 Department of Physics, University of Surrey, Guildford, Surrey, UK 2 Institute of Cancer Research & Royal Marsden NHS Trust, Sutton, Surrey, U.K.
Transcript
Page 1: 2D-CSI in a fraction of the time  using multiple receiver coils

2D-CSI in a fraction of the time using multiple receiver coils

Simon J. Doran1, Adam J. Schwarz2

and Martin O. Leach2

1Department of Physics, University of Surrey,

Guildford, Surrey, UK2Institute of Cancer Research & Royal Marsden NHS

Trust, Sutton, Surrey, U.K.

Page 2: 2D-CSI in a fraction of the time  using multiple receiver coils

Funding from Cancer Research Campaign [CRC]National Institutes of Health (NIH)NHS (South Thames)

Acknowledgements

Page 3: 2D-CSI in a fraction of the time  using multiple receiver coils

Background

• Use of phased array receiver coils in surface coil studies - improves SNR- improves coverage (e.g. ~volumetric coverage)

• Spatial receive profiles of the coil array elements already employed in imaging schemes to reduce the total acquisition time

• Aim of this study is to develop such techniques for application to chemical shift imaging (CSI) spectroscopy acquisitions, primarily for application to extra-cranial tumours.

Page 4: 2D-CSI in a fraction of the time  using multiple receiver coils

Reduced imaging time with phased arrays: Ra-Rim method

Advantages:

• Works for arbitrary array sensitivity profiles. (Orthogonal coil basis functions [as in SMASH] not required).

• Don’t require pure coil sensitivity maps. Any sequence can be used for the reference.

Page 5: 2D-CSI in a fraction of the time  using multiple receiver coils

Reduced imaging time with phased arrays: Ra-Rim method1

1 Ra and Rim (1993) MRM 30 142-145

For C coils, acquire:

MxN reference image: R ( M x N x C )

Mx(N/L) folded image: I ( M x N/L x C )

These are related to each other via the array sensitivity function, which is the same for both.

Page 6: 2D-CSI in a fraction of the time  using multiple receiver coils

Folded images acquired in fraction of time

N N / L

E.g. L = 2

Page 7: 2D-CSI in a fraction of the time  using multiple receiver coils

Reconstruction scheme

To find relationship x jl between desired (unfolded) and

reference image, solve the matrix equation

I jc = R j

c l x jl for x j

l, for each j

(i.e. N/L times).

Desired image obtained as Icfinal = Rc x

Foldedimage

Referenceimage

Unknown to be found

Pixel numberin fold direction

Coil number Foldingparameter

Page 8: 2D-CSI in a fraction of the time  using multiple receiver coils

Application to CSI

• Additional complex, chemical shift dimension in undersampled data set.

• Use imaging (rapidly acquired gradient echo) reference data for coil sensitivity maps.

16 x 16 x 1024 x 4 2D-CSI [for comparison] 8 x 16 x 1024 x 4 undersampled 2D-CSI 16 x 16 x 1 x 4 reference images

(in practice rebinned from clinical 256 x 256 images)

• Three compartment phantom

• Siemens Vision, body-phased-array (C=4)

• (No k-space apodisation)

Page 9: 2D-CSI in a fraction of the time  using multiple receiver coils

Extension to CSI

Spatial dimensions 16 x 16

Spectral dimension1024

Repeat the basic Ra-Rim1024 times, once on eachplane.

Page 10: 2D-CSI in a fraction of the time  using multiple receiver coils

Phantom & coil arrangement

Transverse section Side view

20mMCho

50mM Cho

oil

Coil elements

Page 11: 2D-CSI in a fraction of the time  using multiple receiver coils

Reconstruction of undersampled images

Folded T1w

Reference

Reconstructed T1w

Page 12: 2D-CSI in a fraction of the time  using multiple receiver coils

Reference images from individual elements

Acquired at 256x256

k-space truncated to 16x16 for registration with CSI

Page 13: 2D-CSI in a fraction of the time  using multiple receiver coils

CSI metabolite images : oil

Folded CSI (oil CH2) Reconstructed

Full CSI

Rebinned ref. image

Page 14: 2D-CSI in a fraction of the time  using multiple receiver coils

CSI metabolite images : water

Folded CSI (water)

Rebinned ref. image

Reconstructed

Full CSI

Page 15: 2D-CSI in a fraction of the time  using multiple receiver coils

Folded spectra from within oil and choline balls

Page 16: 2D-CSI in a fraction of the time  using multiple receiver coils

Unfolded Reference from full 16x16 CSI

Unfolded spectrum from within oil and choline balls

CH2

CH3

(water)

choline residuallipid

Page 17: 2D-CSI in a fraction of the time  using multiple receiver coils

Conclusions

• Ra-Rim method has been extended to reduce acquisition time in CSI spectroscopic studies in (pseudo-) abdominal sites, using product coil and rapidly acquired image data as coil sensitivity reference.

• ‘Unfolding’ process moves aliased signal to its correct location.

• Further work is need both to quantify the minor differences seen between restored (unfolded) and reference CSI signals and to improve the method.

Page 18: 2D-CSI in a fraction of the time  using multiple receiver coils

Funding from Cancer Research Campaign [CRC]National Institutes of Health (NIH)NHS (South Thames)

Acknowledgements