FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• After cleaning, rinse blades thoroughly, and dry prior to disinfection/sterilization
• Inspect physical condition for: Foreign Substances, Damage or cracks, broken, loose or wear
• WARNING: ultrasonic cleaning is not recommended
• A minimum of High Level Disinfection is required
• If recommended by manufacture blades may be steam autoclaved
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• NOTE: Autoclaving with lamp removed will result in damage to the electrical circuit
• Standard battery handles are usually not compatible with steam autoclave sterilization
• Autoclave able handles can often be identified by the term ‘AUTOCLAVE’ written on the handle If they do not have the marking they ARE NOT autoclavable
• NOTE: ALWAYS FOLLOW MANUFACTURERS GUIDELINES
• Always wrap laryngoscope blades and handles unattached if autoclaving
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• NOTE: Do not exceed temperature of 134°C
• Flash and Hot air sterilization is not usually recommended
• If disinfecting refer to solution manufacturer’s instructions for recommended exposure times and solution concentrations,
• NOTE: Disinfecting with lamp removed will result in damage to the electrical circuit
• Prior to immersion, ensure that the lamp is secure
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• Rinse thoroughly in sterile water
• Dry with a non linting cloth
Test Procedure
• Once Disinfected/Autoclaved (unwrap pack) replace appropriate size batteries (as per manufacturers instructions) into Laryngoscope handle and replace bottom cap Stubby handle: insert battery pack with tab side down
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• Laryngoscope blades and handles should always be tested after cleaning/disinfection/sterilization and prior to use
• To check, connect the laryngoscope blade to the handle and pull open to the “on” position If the unit fails to light or flickers, check the lamp/ batteries
• Be sure adequate supplies of spare lamps, batteries, and replacement parts are readily available
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• Be sure the lamp’s glass envelope is clean and free of any fingerprints after assembly If necessary, the glass may be cleaned with a soft cloth or cotton ball moistened in alcohol
• Wrapping the reassembled laryngoscope should protect it from contamination until the item is to be used the
• A re-sealable plastic bag or other impermeable wrap may be used as the covering because the laryngoscope is clean not sterile Wrapping the blades in sterilization wrap or a sterilization peel pack is not recommended because this may lead the user to think that the blade is sterile
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
• Contamination may result from a clean blade coming into contact with a contaminated laryngoscope handle
• The item should be clearly labeled as being high-level disinfected and not sterile Labeling should also contain some method to indicate the date when the high-level disinfection occurred and the person responsible for completing the process
5. Reference Documents:
FAUJI FOUNDATION HOSPITAL LAHOREDocument Code
Decontamination And Management Of LaryngoscopesFFH/CSSD/SOP/032
Issue#: A Issue Date:13/05/2014 Rev#: 00 Rev. Date:
Procedure Manual
Standard Precautions
6. Distribution :
Copy No. 1 with AQMR & Copy No. 2 in the Dept.