0.1bd_ts_06664 AB_01
2003 Philips Medical SystemsALL RIGHTS RESERVED
4512 984 06664 REV AB
RADIOGRAPHY
LIST OF DOCUMENTATION IN THIS BINDER:
⊗ Subsystem Manual bucky DIAGNOST TS
INALFA non sensing bucky unit
Manual Collimator Ralco
Note: indicates document present
LIST OF ALL BINDERS FOR BUCKY DIAGNOST TS
-- bucky DIAGNOST TS
-- OPTIMUS generator
-- X-ray tube
-- bucky DIAGNOST VE/VT
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
1
2
3
4
5
6
7
8
P
Z1-
Z2-
INTRODUCTION AND TECHNICAL DATA
INSTALLATION
FAULT FINDING
REPLACEMENT
PROGRAMMINGS
ADJUSTMENTS
ACCEPTANCE
SERVICE INFORMATION
PARTS LIST
SCHEMATIC DIAGRAMS
WIRING DIAGRAMS
E WPhilipsMedical SystemsDevelopment and Manufacturing Centre
4512 984 06664 REV AB 1bd_ts_06664_AB_titel
SERVICE MANUAL704
SUBSYSTEM
bucky DIAGNOST TS
9848 600 00061 Fix
9848 600 00071 Elevation
9848 600 00081 Tomo
bucky table for routine radiography diagnostics
DMC Hamburg
Printed in HamburgFederal Republic of Germany
2003 Philips Medical SystemsALL RIGHTS RESERVED
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Bucky DIAGNOST TS
4512 984 06664 REV AB 3.1 2003 Philips Medical SystemsALL RIGHTS RESERVEDbd_ts_06664 AB_lopad
SERVICE MANUAL -- SUBSYSTEM
Bucky DIAGNOST TS Author: V. Neumann
Type No: 9848 600 000619848 600 000719848 600 00081
In case there are any questions concerning this manual,please send this LOPAD via fax to 49/(0)40/5078 2481
File:Bucky Diagnost_TS_06664 AB
List of pages and drawings (LOPAD) Manual Order No: 4512 984 06664
released: 12/2003
0.1 A4 (Rosa Karton)
13.13.2
_________________________________________
1-0.1/0.2 (99.0)
1-1 ... 1-15 (99.0)
1Z--1.1 (99.0) A31Z--1.2 (99.0) A31Z--1.3 (97.1) A31Z--1.4 (99.0) A31Z--1.5 (99.0) A3
1Z--2.1 (99.0) A3
1Z--2.2.1 (99.0) A31Z--2.2.2 (99.0) A3
1Z--3.1 (97.1) A31Z--3.2 (97.1) A3
1Z--4.1 (97.1) A31Z--4.2 (97.1) A3
1Z--5.1 (99.0) A31Z--5.2 (99.0) A3
1Z--6.1 (99.0) A31Z--6.2 (99.0) A31Z--6.3 (00.0) A3
1Z--7 (00.0) A3
1Z--8 (99.0) A3
1Z--9 (99.1) A3
1Z--10 (00.0) A3
Z--50 (00.0) A3 4512 982 00081
_________________________________________
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS
4512 984 06664 REV AB3.2 2003 Philips Medical SystemsALL RIGHTS RESERVED
bd_ts_06664 AB_lopad
2-0.1 (99.0)
2-1 ...2- 36 (99.0)
2Z--2.0 (01.0) A4
2Z--10 (b/03.0) A3 4512 983 06151_________________________________________
3-0.1 (99.0)
3-1 ... 3-9 (99.0)_________________________________________
Z0--1 (99.0)
Z1--1 (99.0) A3Z1--1.1 (99.0) A3
Z1--2 (99.0) A3Z1--2.1 (99.0) A3
Z1--3 (99.0) A3Z1--3.1 (01.1) A3
Z1--4 (99.0) A3Z1--4.1 (99.0) A3Z1--4.2 (00.0) A3Z1--4.2 (a/00.0) A3
Z1--5 (a/00.0) A3Z1--5.1 (a/00.0) A3Z1--5.1.1 (00.0) A3
Z1--6.1 (99.0) A3
Z1--7.1 (a/01.0) A3_________________________________________
Z0--2 (99.0)
Z2--1 (00.0) A4
Z2--2.1 (99.0) A3Z2--2.2 (99.0) A3
Z2--3.1 (99.0) A3Z2--3.2 (99.0) A3Z2--3.3 (99.0) A3
Z2--4.1 (99.0) A3Z2--4.2 (99.0) A3
Z2--5 (a/00.0) A3
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1999 Philips Medical Systems
ALL RIGHTS RESERVEDINH
1--0.1
INTRODUCTION AND TECHNICAL DATA
TEXT
Contents 1--0.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1. Technical data 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1. System bucky DIAGNOST TS 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1.1. Standard Table 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1.2. Tomography table 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1.3. X-ray generator 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1.4. Applications 1--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.2. Tubes 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.3. Mechanical data 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.3.1. Dimensions 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.3.2. Movements 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4. Electrical data 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4.1. Table 1--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4.2. X-ray generator OPTIMUS 50 / 65 / 80 1--3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5. Power supply 1--4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5.1. Type of power supply 1--4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5.2. Calculating the mains resistances 1--5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6. Earth-leakage circuit breaker 1--5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.7. Emergency-OFF device 1--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.8. Heat emission 1--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Environmental conditions 1--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.1. Table 1--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.2. X-ray generator 1--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3. Measurements and weights 1--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.1. Pre-installed system 1--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2. Transport data 1--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2.1. Box dimensions 1--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2.2. Roadway 1--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2.3. Unpacking 1--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. Scope of delivery 1--10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.1. ”Fixed table”
9848 600 00061 – bucky DIAGNOST TS with floating table top 1--10. . . . . . . . . . . . . .4.2. Height adjustable
9848 600 00071 bucky DIAGNOST TS with floating table and elevator 1--11. . . . . . . .4.3. Tomography
9848 600 00081 bucky DIAGNOST TSwith floating table and elevator and tomo extension 1--11. . . . . . . . . . . . . . . . . . . . . . . .
4.4. Options 1--11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.4.1. bucky DIAGNOST TS 1--11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.4.2. bucky DIAGNOST VE / VT 1--12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. Site planning 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.1. Planning data 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.2. Determining the site of installation 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.3. Floor fixing 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.4. Loading capacity of floor 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.5. Height of the room 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5.6. Electrical data 1--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--0.2 1999 Philips Medical Systems
ALL RIGHTS RESERVEDINH
6. Tools 1--14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7. Installation materials required 1--14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7.1. Cables 1--15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DRAWINGS
bucky DIAGNOST TSelevated or tomo version for wall stand left, mechanical dimensions 1Z--1.1. . . . . . . . .bucky DIAGNOST TSelevated or tomo version for wall stand right, mechanical dimensions 1Z--1.2. . . . . . . .OPTIMUS L 50, mechanical dimensions 1Z--1.3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TSfixed table version for wall stand left, mechanical dimensions 1Z--1.4. . . . . . . . . . . . . . .bucky DIAGNOST TSfixed table version for wall stand right, mechanical dimensions 1Z--1.5. . . . . . . . . . . . . .bucky DIAGNOST TSelevated or tomo version, wall clearance 1Z--2.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS fixed table version 1Z--2.2.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS fixed table version 1Z--2.2.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VEtypical room layout 1Z--3.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VEtypical room layout 1Z--3.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VErelation data 1Z--4.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VErelation data 1Z--4.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TSelevated or tomo version, connecting surface / floor fixing 1Z--5.1. . . . . . . . . . . . . . . . . .bucky DIAGNOST TSfixed table version, connecting surface / floor fixing 1Z--5.2. . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS, VE, OPTIMUS L 50standard version, connection diagram 1Z--6.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS, VE, OPTIMUS L 50tomo version, connection diagram 1Z--6.2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS, VE, OPTIMUS L 50earthing diagram 1Z--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .bucky DIAGNOST TS, VEsystem installation and cable length 1Z--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Wall connection box bucky DIAGNOST TS 1Z--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--1 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1. Technical data
1.1. System bucky DIAGNOST TS
1.1.1. Standard Table
The bucky DIAGNOST TS is installed with the X-raygenerator OPTIMUS L 50. The bucky DIAGNOST TS ischaracterized by:
-- X-ray exposures at the horizontal patient with buckyunit.
-- X-ray exposures at the horizontal patient withoutbucky unit.
-- Patient area of table top2400mm x 600mm (excl. the Alu profile)
-- SID 1000mm to 1250mm.
1.1.2. Tomography table
Linear tomography at the horizontal patient.
-- Angle: 8° ... 40°
-- Time: 0.5 ... 4sec.
-- Layer: 0° ... 250mm
-- SID: 1000 ... 1250mm, 1100mm default
1.1.3. X-ray generator
The X-ray generator OPTIMUS L 50 for radiography is based on computer-controlled converter technology.The converter operates in the non-audible frequency range.Applicational options are essentially achieved by releasing software modules.
1.1.4. Applications
-- Radiography
-- Tomography
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--2 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1.2. Tubes
Versions of the X-ray generator OPTIMUS L 50
50kW 3phase generator for 1 tube.
Recommended standard tubes:RO 17 50 0.6 / 1.3 mm
R type stators only for low speed rotor control.
Compatible tube housings:ROT 350
Optional tubes: SRO 33 100 / ROT350 (for U.S.A. only) 9874 005 16022SRO 33 100 / ROT350 9874 005 09022SRO 25 50 / ROT350 9874 004 23022SRO 09 51 / ROT350 9890 000 64181”SRO” in combination with
High Speed Rotor Control 9848 600 01101
1.3. Mechanical data
1.3.1. Dimensions
Mechanical dimensions for wall stand left see Z--1.1
Mechanical dimensions for wall stand right see Z--1.2
Mechanical dimensions for OPTIMUS L 50 see Z--1.3
1.3.2. Movements
See Z--1.1 or Z--1.2
1.4. Electrical data
1.4.1. Table
Operating voltages: 230 V 10% --15% 50/60 Hz
Power consumption without tomography: appr. 420 WPower consumption with tomography: appr. 420 W
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--3 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1.4.2. X-ray generator OPTIMUS 50 / 65 / 80
OPTIMUS 50 OPTIMUS 65 OPTIMUS 80Power dataMains voltage 3 x 400 V ±10% ( 380 V --5%, 415 V +6%)
3 x 400/440/460 V ±10%3 x 480 V --10% +6%
3 x 190...343 V ±10% with external transformer, max. 50 kW (option)
Mains frequency 49 ... 61 HzMax. mains current-- at 400 V 145 A 190 A 230 A-- at 440 V 135 A 180 A 215 A-- at 460 V 125 A 170 A 210 A-- at 480 V 120 A 160 A 205 A-- at 190 V 300 A -- --
Fuse protection (slow–blow) 35 A 50 A 50 A100 A at ≤ 240 V
Mains resistance-- at 400 V ≤ 300 mΩ ≤ 200 mΩ ≤ 200 mΩ-- at 440/460 V ≤ 350 mΩ ≤ 240 mΩ ≤ 240 mΩ-- at 480 V ≤ 400 mΩ ≤ 300 mΩ ≤ 300 mΩ
-- at 480 V, valid for DOD only ≤ 300 mΩ ≤ 240 mΩ ≤ 180 mΩ
Unit supply 230 V / 400 V, max. 5 A
Max. output power 50 kW 65 kW 80 kW
Operating dataTube current 1.0 ... 650 mA 1.0 ... 900 mA 1.0 ... 1100 mA
Tube voltage 40 ... 150kV in steps of 1kV or %mAs product 0.5 ... 850mAsExposure time 1ms ... 6/16sExposure sequency ≤ 12 exp./sNoise level ≤ 40 dBAHeat dissipation 500WInterfacing option for.... door contact, external radiation warning indicator
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--4 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1.5. Power supply
1.5.1. Type of power supply
For 50 kW X-ray generator a three-phase system is needed.
The three-phase system must be balanced, i.e. the external conductor voltages against earth must be about equal,see sketch.
For reasons of operational safety and compliance with international rules, operation on three-phase mains that arefloating or unbalanced to earth (e.g. with one phase earthed instead of the neutral point) is impermissible unlessadequate adaptation measures have been taken! Check that the sequence of phases is correct corresponding tothe designations L1, L2, L3.
L1
L2
L3(N)
L1
L1
L2
L2
L3
L3
X-rayinstallation
X-rayinstallation
X-rayinstallation
3 phase WYE
3 phase DELTA,balanced earth or floating
3 phase DELTA,grounded
-- Working range 400 ... 480Vmeasured between phases:L1 -- L2 / L2 -- L3 / L1 -- L3
-- Working range 400 ... 480Vmeasured between phases:L1 -- L2 / L2 -- L3 / L1 -- L3
-- Working range 400 ... 480Vmeasured between phases:L1 -- L2 / L2 -- L3 / L1 -- L3
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--5 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1.5.2. Calculating the mains resistances
If possible, the sum of R1 , R2 and R3 should be smallerthan the RXG required.
With higher internal mains resistances the generatoroutput is reduced respectively.
Note that the cross section of lead l3 must not exceed25mm2.
-- R0 designates the mains resistance on the distributor transformer.
-- R1 is dependent upon the length of lead l1 between distributor transformer and main distributor and upon thecross section selected, so:
R1 = l1 x RCu RCu from table-- R2 is estimated to be 15mΩ
EMERGENCY--OFF switch 4mΩ Earth--leakage circuit breaker 5.5mΩ 15mΩFuse 5.5mΩ
-- R3 is dependent upon the length of lead l3 between main distributor and wall junction box and upon the crosssection selected, so:
R3 = l3 x RCu RCu from table
The resistances already consider the go and return line so that the calculation can be based on simple cablelengths.
Copper cross section [mm2 ] Resistance RCu [mΩ/m]16 2.1925 1.435 1.050 0.770 0.595 0.38
120 0.30150 0.24
1.6. Earth-leakage circuit breaker
To be provided between fuse and X-ray installation depending upon local regulations.Siemens earth-leakage circuit breaker FSJ 50 A
Rated fault current 30mA
Rated current 50A
Connection terminals for wire cross sections of up to 25mm2.
R0 + R2
L1L2L3
L1L2L3
l1 l3
357J94
MEX 100
+ R1 + R3 < RXG
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--6 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
1.7. Emergency-OFF device
To be provided depending upon local regulations.
There are 2 possibilities:
1. All the EMERGENCY-OFF buttons are connected in series and looped into the starting circuit (12V DC) of thegenerator.
2. The EMERGENCY-OFF circuit acts on an external main contactor which switches off the power before it is fedinto the generator.
1.8. Heat emission
Table: 530WX-ray generator: 500W (the generator is air cooled using natural convection)
2. Environmental conditions
2.1. Table
According to PMS standard UXW 13600 class C1 (indoor temperature controlled)
during regular work:
-- Ambient temperature: + 10 °C ... +40 °C-- Relative air humidity: 20% ... 80%
-- Air pressure: 700 ... 1100hPA
during storage / transport:
-- Ambient temperature: --25 °C ... +70 °C
-- Relative air humidity: 5% ... 100%-- Air pressure: 700 ... 1100 hPA
2.2. X-ray generator
according to PMS standard UXW 13600 class C1 (indoor temperature controlled)
during regular work:
-- Ambient temperature: +10 °C ... +40 °C
-- Relative humidity: 0% ... 90%, no condensation-- Air pressure: 700 ... 1100hPA
during storage / transport:-- Ambient temperature: --20 °C ... +50 °C
Note: Electrolytics require less than 40 °C for storage life.
-- Relative air humidity: 0% ... 95%, non--condensing
-- Air pressure: 700 ... 1100hPA
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--7 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
3. Measurements and weights
3.1. Pre-installed system
Table / column (standard)
Weights (kg) Dimensions (cm) (L/W/H)with shipping box without shipping box with shipping box without shipping box
Box 1Contents: manuals, table,table top, rails, SPS, set ofcable, covers, tube, controlhandle, column, set of tomodevices (option)
870 750 270 x 97 x 196 265 x 89 x 190
X-ray generator 50 kW
Weights (kg) Dimensions (cm) (L/W/H)with shipping box without shipping box with shipping box without shipping box
Box 1Contents: generator cabinet,
operating panel
226 178 210 x 82 x 84195,5 x 55 5243 x 33,5 x 6
Box 2Contents: H.V. generator
1-tube version
100 73 77 x 67 x 80
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--8 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
3.2. Transport data
3.2.1. Box dimensions
Weight
Gross 870kg
Loading 816kg
Net 750kg
Dimension (L x W x H)
Box 270cm x 97cm x 196cm
Pallet including loading 265cm x 89cm x 190cm
Sub-pallet 140cm x 89cmSub-pallet 125cm x 89cm
3.2.2. Roadway
1 Roadway for the gross palletCanting while moving min 170cm width of the floor
2 Roadway for the sub--pallets
Canting while moving min 115cm width of the floor
min 170cm width of roadwayfor gross pallet
min 115cm
min 170cm
870kg
1
2
min 115cm width of roadwayfor sub--pallet
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--9 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
3.2.3. Unpacking
The following steps should be done by four persons.
s Dismantle the box, remove the foil.
s 1 Remove the rails from the pallet.
s 2 Remove from the column the box with documentsand distance tool.
s 3 Erect the column.
-- 3a Move the column approximatly 15cm to footdirection of the column.
-- 3b Turn the column base about 90°.The column must be ratched in the cam.
-- 3c Lift up the column and place it.
Lay corrugated paper or somethingsimilar on the floor in the area wherethe column is to be moved.
Apply extreme care when handling theconnection field.
s Remove the wooden carrier where the column waspositioned by unscrewing the colored screws.
Unpacking0.5 h0.5 h0.5 h0.5 h
12
Documents anddistance tool
3a
3b3c
3
90_
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--10 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
s 4 Remove the box including cables from the table.
s 5 Remove the optional tomo devices from the pallet.
s 6 Remove the tube assembly from the pallet.
s Remove the wooden items by removing the coloredscrews.
NoteThe gross pallet can be splitted in two sub-pallets.
s 7 Remove the table from the pallet.Use the shoulder straps. Do not lift the table bymeans of the table top frame.
s 8 After unscrewing the sub–pallets from the pallet youget access to the table top. Remove the table top.
s Dispose of the package locally.
4. Scope of delivery
4.1. ”Fixed table”9848 600 00061 – bucky DIAGNOST TS with floating table top
-- Table, including column and X-ray tube carriage-- Set of floor fixing material incl. dowels for standard concrete floors
-- X-ray generator OPTIMUS L 50 including operating panel
-- Complete set of cables incl. plugs at both sides
-- Grid
-- X-ray tube-- Collimator
-- Control handle
Cables
4
5
6
74
8
7
Table top deliveredin the pallet bottom
The pallet is deviable
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--11 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
4.2. Height adjustable9848 600 00071 bucky DIAGNOST TS with floating table and elevator
-- Table, including column and X-ray tube carriage
-- Set of floor fixing material incl. dowels for standard concrete floors-- X-ray generator OPTIMUS L 50 including operating panel
-- Complete set of cables incl. plugs at both sides
-- Grid
-- X- tube-- Collimator
-- Control handle
4.3. Tomography9848 600 00081 bucky DIAGNOST TS with floating table and elevator and tomo extension
-- Table including tomography attachment, column and X-ray tube carriage
-- Set of floor fixing material incl. dowels for standard concrete floors
-- X-ray generator OPTIMUS L 50 including operating panel
-- Complete set of cables incl. plugs at both sides-- Grid
-- X-ray tube
-- Collimator
4.4. Options
4.4.1. bucky DIAGNOST TS
-- wall mounting of operating panel 9890 000 02451-- stand for operating panel 9890 000 02441
-- Amplimat measuring chamber 9890 000 01612
-- extension main rail 4512 590 30491
-- interface for medio CP / SZ3 4512 590 30351-- high speed rotor control 9848 600 01101
-- 3phase extension 9848 600 00122
-- SRO 33/100 / ROT 350 9874 005 09022
-- SRO 25/50 / ROT 350 9874 004 23022
-- SRO 09/51 / ROT 350 9890 000 64181-- SRO 33/100 / ROT 350/HEW (only for U.S.A.) 9874 005 16022
-- bucky unit non sensing left 9848 500 20303
-- bucky unit non sensing right 9848 500 35703
-- manual collimator T303 for TS 9848 600 00172-- manual collimator for TS (HHS) 9848 600 02092
-- APR 9848 600 00201
-- OPTIMUS 50 (standard) three phase
-- OPTIMUS 65 three phase
-- OPTIMUS 80 three phase-- OPTIMUS 30 single or three phase
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--12 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
-- OPTIMUS 40 single or three phase-- X-ray tube RO 17/50 9874 001 16002
-- grid 36/08 110F TS 9848 600 00821
-- grid 40/12 110F TS 9848 600 00831
-- H.V. cable, lengths = xx m 9806 402 6xx02
-- set of cables generator--column 14 m 9848 600 01871-- set of cables generator--table 20 m 9848 600 01901
-- set of cables table--column 20 m 9848 600 01911
-- set of cables table--column 25 m 9848 600 01921
-- TS control grip tomo and laser 9848 600 02081-- TS control grip standard and laser 9848 600 02071
-- table top Getalit PUR (2.4m) 9890 010 02194
-- table top Getalit (2m) 9890 010 81371
4.4.2. bucky DIAGNOST VE / VT
-- grid 36/8 110F 9860 834 90111
-- grid 36/8 140F 9860 834 90141-- grid 36/8 130F 9860 836 90131
-- grid 36/8 140F 9860 836 90141
-- grid 36/8 180F 9860 836 90181
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--13 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
5. Site planning
5.1. Planning data
The preliminary planning work is restricted to the selection of the suitable room. The outfit of the room has to beequipped with the conduits and pipes required and, in cases where the loading capacity of the bearing area plusweight of patient insufficient, the floor preparation for the table.
For detailed information about loading area see Z--5.
Where DIN specification are not applicable the minimum wall clearances should be determined in compliance withthe national regulations concerned.
5.2. Determining the site of installation
The choice of the suitable room should be based on the layout plan.
For detailed information see Z--3.1, Z--3.2, Z--4.1, Z--4.2
5.3. Floor fixing
The table is fixed by means of 4 bolts and the column rails are fixed by means of 4 + 2 bolts (ext. rail left or rightpos.) also.
Use e.g. Liebig safety dowels S12 on compound plaster floor concrete class Bn 150 or Upat--PSZ anchor dowelstype B or S12 on compound plaster floor concrete class Bn 250 or other national requirements.
The mounting points on mounting surface have to be suitable for tensile strength of F ≥ 2500N each.
In case of ”computer floors” (stilted floors) an appropriate understructure or staybolts have to be used.For detailed information see Z--5
5.4. Loading capacity of floor
Surface load table and column = 750kg plus 135kg weight of patient.
Surface load X--ray generator = 250kgFor detailed information see Z--5
5.5. Height of the room
For detailed information see Z--1.1 or Z--1.2
5.6. Electrical data
The bucky DIAGNOST TS is supplied incl. all necessary cables incl. plugs with required free length between themain components like X--ray generator and table.
For detailed information see Z--6.1 and Z--7, Z--8, Z--50
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS(99.0)1--14 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
6. Tools
-- Groof key set for column base and transversal tube carriage 4512 590 30362
-- 25mm aluminium for AMPLIMAT adjustment-- chip-ejector for service 2422 487 89772
-- Multimeter
-- Mirror (about 30 x 40 cm)
-- Spring balance 0 ... 200N-- Standard set of tools
-- Fork lift
-- ESD antistatic kit
-- Hammer drill with 12mm carbide drillbits.
-- Spirit level 1m-- Service-PC
-- IBM-compatible, 640 kB RAM, 3.5” floppy disk drive, with 1 serial port
-- Sub-D Adapter Kit 9/15/25 pole
-- OPTIMUS L 50 utility program for generator programming supplied on a floppy disk with the generator.-- PC-Hardkey to use the OPTIMUS L 50 utility program.
-- Mains resistance measuring instrument.
-- Collimator adjustment tool code No: 4522 980 31521
-- Tomo test tool Code No.: 8112 108 57929
-- PC-XRGSCOPE Cable Code No.: 4512 130 56931 for OPTIMUS generator-- Rail distance tool 4512 590 30501 (included in the delivery)
7. Installation materials required
To be ordered from the service department via PMS Hamburg:
-- Wall junction box for X--ray generator 4512 103 75385inclusive connection block (25 mm2) for mains supply and connection block (10 mm2) for unit supply.
-- Relay for radiation warning indicator 4512 100 452311 interface relay with a floating contact (230 V, 1 A) is included in the scope of delivery for the generator.
-- Grease – Molykote DX 4512 590 30321 (included with the delivery)
-- Set of three colours 4512 590 30031 (included with the delivery)
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS (99.0) 1--15 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_1_990
7.1. Cables
Cables supplied with the system. Cable length must be ordered via catalogue depending from outfit of the room.
H.V. cables 9806 402 6xx02
plugs: 75kVp federal standard female connectors O3/O3
length: 6 ... 20m in steps of 2 m
capacity: 155pF/m
diameter: 16.5mmThe cable length is indicated at the 9th and 10th digits of the code No.
Thermal contact cable
2-wire screened for one excess temperature switch 4512 100 6615110-wire screened for additional supervision like 0722 215 19005temperature alarm switch, buzzer, selection indicator.
Stator cable 3 x 1.31 mm2 , screened 0722 215 02054
AMPLIMAT cable with Sub-D (15pole)12 m 9890 000 0172116 m 9890 000 0173120 m 9890 000 0174124 m 9890 000 01751
AMPLIMAT cables 9803 507 0xx02 with 3 PLUS plugs at both ends can be connected in the generator via thefollowing adapter for each cable:
Adapter for AMPLIMAT cable 4512 108 09041
The generator includes five adapters, 3Plus--Sub-D (15pole)
Operating desk
cable set 10m 9890 000 0241120m 9890 000 0242130m 9890 000 02431
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Connection diagramStandard version
bucky DIAGNOST TS, VE
A3
00-
09-2
7 O
st./S
chr.
C 2000 Philips Medical SystemsALL RIGHTS RESERVED
(00.0)
temperature switch
H.V. cables
communication
32
3
XA 11 1
2
4TS2:5
XA 1
XF 7020
XA
SA
basic
X-ray tube
bucky examination unit
stator cable, screened
(bk)
(rd)
(wt)
UVW
supply J1
SA
bucky wall stand
amplimat chamber VPX4
7A2:J1 amplimat chamber PHX1
6J2
6J3B
6J3A
X23
X22
X23X22
X7X8
L1N(L2)
L2 (N)
mains fuses
earth lekagecircuit breaker
supply
O3 plugs9HT1
M9f
power supply SNX16
VPX1X3 release
D15f
D9f
D15f
D15m
D15m
X1 SZ1X1
M6
release
OPTIMUS 30 single phase
da
df
dc
dd
db
dd
df
tube housing
7A2:J2
M6f
M12f
D9m
C
PMA
L1/N
SZ1X11 M12
UB1X20 M12
dk control
VE VP
EPMA
XA
S
bucky DIAGNOST TS 1Z-6.3
0014
1z63
4TS2:4
4TS2:3
4TS2:2
4TS2:1
J3 D15m
AEC J57A2:J4D9f D9m
J2 D25f3TS1 223TS1 233TS1 243TS1 363TS1 373TS1 513TS1 523TS1 53
4TS2 44TS2 5 5
45352513736242322 interface
X13X14
switch controllerby emergency switch
emergency switch
L1
230V AC
7X11(door contact)
4
7
7X11810
Power ON bucky TS
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSA1/
A3
00-0
7-04
Kö
C 1997 Philips Medizin SystemeALL RIGHTS RESERVED
(00.0)
4,00 mm2SN
XA
S
bucky DIAGNOST TS 1Z-10
X16T1
Earthing diagram
4,00 mm2
4,00 mm2
4,00 mm24,00 mm2
4,00 mm2T261Z1
4,00 mm2
4,00 mm2
4,00 mm2
VASA (right)VA
6,00 mm2 UB6,00 mm2
4,00 mm2
4,00 mm22,50 mm2
2,50 mm2
U
UFUE XA
LA
Table
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 1999 Philips Medical Systems
ALL RIGHTS RESERVEDINH
2--0.1
INSTALLATION
TEXT
Contents 2--0.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Safety regulations 2--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. Transport data 2--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1. Box dimensions 2--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.2. Roadway 2--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Unpacking 2--3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.1. Removal of column base from column 2--4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3. Mounting the rails 2--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tomo option only 2--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Coarse Adjustment Tomo option only 2--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.1. Insertion of column to long rail 2--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.2. Installation of tube arm 2--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.3. Final alignment of the column (tomo only) 2--13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.4. Installation of bucky table 2--14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.5. Installation of Tomo pivot tower (tomo only) 2--16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.6. Installation of bucky-device extension (tomo only) 2--17. . . . . . . . . . . . . . . . . . . . . . . . . .3.7. Installation of position control switches (tomo only) 2--17. . . . . . . . . . . . . . . . . . . . . . . . .3.8. Installation of OPTIMUS 2--19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.8.1. Installation of bucky DIAGNOST TS
with tomo and Medio CP/CP-H 2--19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.8.2. Programming of APR data 2--20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.8.3. Programming tube select at control handle 2--20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.9. Installation of tabletop 2--21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.9.1. Installation of tabletop in small rooms 2--21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.10. Setting to work 2--22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11. Adjustment procedure for tomo 2--25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.1. Adjustment of the table level 2--25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.2. Correction for tomo layer height = fine adjustment under X-ray condition 2--25. . . . . . .3.11.3. Adjustment of tomo layer height indication 2--26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.4. Setting the display language 2--27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.5. Adjustment of the tomo axis 2--28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.6. Alignment for center line
= center of bucky 2--29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.11.7. Checking drive and position control switches 2--29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12. Tomo adjustment under exposure conditions (tomo only) 2--31. . . . . . . . . . . . . . . . . . . .3.12.1. Measurement of X-ray zone 2--31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12.2. Symmetry of tomo axis 2--32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12.3. Tomography resolution test 2--33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.12.4. Tomo layer height 2--34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.13. Final work 2--34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.14. Acceptance test 2--35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.15. Handing over to customer 2--35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.16. Installation for tomo bar parking frame 2--36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--1 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
Safety regulations
Only technically qualified personnel who has received the appropriate instructions and training forradiography equipment in medical applications and in particular for the installation and commissioningof the bucky Diagnost TS are allowed to install, maintain and service this equipment.
Therefore, instructions in this chapter are considered as a supplementary guideline for correctinstallation of the bucky Diagnost TS and are kept very briefly.
Furthermore, all personnel involved must be instructed on safety precautions with respect to radiographyand the related high voltage devices as well as to all other electrical low-voltage equipment. All relevantstandards and regulations must be complied with.
The local service organisation is responsible for the safety and training of personnel and prevention ofmaterial damage.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--2 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
1. Transport data
1.1. Box dimensions
Weight
Gross 870kg
Loading 816kg
Net 750kg
Dimension (L x W x H)
Box 270cm x 97cm x 196cm
Pallet including loading 265cm x 89cm x 190cm
Sub–pallet 140cm x 89cmSub–pallet 125cm x 89cm
1.2. Roadway
1 Roadway for the gross palletCrossing while moving min 170cm width of the floor
2 Roadway for the sub--palletsCrossing while moving min 115cm width of the floor
min 170cm width of roadwayfor gross pallet
min 115cm
min 170cm
870kg
1
2
min 115cm width of roadwayfor sub-pallet
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--3 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
2. Unpacking
The following steps should be done by four persons.
S Dismantle the box, remove the foil.
1. Remove the rails from the pallet.
NoteFor transportation in insufficient rooms only. See 2.1”Removal of column base from column”.
The column is supplied with an inserted counterweight.For transport, the counterweight and the transverse armare secured by a screw.
Do not remove these screws beforethe installation of the tube arm.
Loosen the transport screws and safety belts on thepallet.
2. Remove the box with documents and distancetool from the column.
3. Set up the column.-- 3a Move the column approximatly 15cm to foot
direction of the column.
-- 3b Turn the column base about 90°.The column must be ratched in the cam.
-- 3c Lift up the column and place it.
Lay corrugated paper or somethingsimilar on the floor in the area wherethe column is to be moved.
Apply extreme care when handlingthe connection field.
S Unscrew the colored screws from the wooden carrieron which the column was positioned and remove it.
Unpacking0.5 h 0.5 h 0.5 h0.5 h
12
Documents anddistance tool
3a
3b3c
90°
3
Unpacking0.5 h 0.5 h 0.5 h0.5 h
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--4 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
4. Remove the box including cables from the table.
5. Remove the optional tomo devices from thepallet.
6. Remove the tube assembly from the pallet.
Remove the wooden items by removing the coloredscrews.
NoteThe gross pallet can be splitted in two sub-pallets.
7. Remove the table from the pallet.
NoteUse the shoulder straps. Do not lift the table bymeans of the table top frame.
8. After unscrewing the sub-pallets from the palletyou get access to the table top. Remove the tabletop.
S Dispose the package locally.
2.1. Removal of column base from column
This section is needed if the column is too large to enterthe X-ray room.
Tools required
Groofkey for column base PMS Code No. 4512 590 30362
Spring scale 200 NImbus Key 5mm
Molykote DX PMS Code No. 4512 590 30321
Isotrophy alcohol
1. Dismantle base and column
1.1. Measure the breakaway force with a springscale when the column is in the the lock position.It should measure 110--150 N. Place the hook ofthe spring scale on the screw head.
1.2. Repeat point 1.1, only column must be in freeposition. Should measure ≤ 60 N.
Cables
4
5
6
74
8
7
Table top deliveredin the pallet bottom
The pallet is deviable
P1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--5 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
1.3. Release the pressure on the Beta grid (measurethe distance “X” between nut and screwhead.Dimensions needed by the adjustment ofreassembling).
1.4. Remove the 2 Imbus (M5) screws and plate(see P 3).
1.5. Loosen the safety screws (pos. 1) on the groovenut (see P6).
1.6. Remove groove nut and steelring.
1.7. Carefully remove column base from column:
S Remove the teflon ring.
S Clean it and grease it with Molykote DX.
P2
X
P3
M5
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--6 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
1.8. Clean running surface and thread with Isopropylalcohol (see P4).For transport cover running surface and thread.
NoteThe stop pin pos. 3 can be removed in case ofthe movement of the column is impaired. (Thisstop pin will not be needed any longer.)
2. Assembling of column base and column
2.1. Grease the running surface of the column andcolumn with Molykote(PMS No. 4512 590 30321).
2.2. Insert the teflon ring.
2.3. Assemble the column base onto the column.(The column must be laid down whenassembling the column base. Take extremecare of thread.)
2.4. Assemble the steelring and groove nut.Hand-tighten the groove nut.
2.5. Rotate the column base for functional check.
2.6. Tighten groove nut until the inner steelring(spring) is compressed. Tighten safety screwspos. 1 (see P 6).
2.7. Adjust the beta grid (see P 2 and step 1.3).
2.8. Repeat step 1.1 and step 1.2.The breakaway force should measure 110--150Nm and the free running force should measure≤ 60N.Adjust the force by loosening, resp. tighteningthe groove nut.
P4
3
2
P5
P6
1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--7 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3. Mounting the rails
Before starting, refer to yourplanning room layout! (see PRB.)
S Assemble main rail.
S Depending on the installation of the bucky wall stand(right or left), place the extension rail right or lefthandside to the main rail (see Z--1.1, Z--1.2 and Z--5).
S Insert the screw (supplied with system) with a washerand tighten with a spanner.
Tomo option only
S Loosen the 4 screws of the tomo switch rail (1) and theend switch cams (2) and remove them.
S Place the main rail parallel to the wall.
S Mark the 5 fixing points on the floor.
S Remove the rail.
S Drill the 8 mm holes approximately 75 mm deep in themiddle of the marked area.
S Insert the 5 anchor dowels (supplied with system) intothe holes.
S Move the rail back to the installation position. Levelthe rail using a spirit level. Use the spacer ifnecessary.
S Insert the screws (supplied with system) with thewashers and tighten with a spanner.
position of extension railfor bucky wall stand left
main rail
position ofextension railfor bucky wall stand righ
tpencil
2
1
extension rail
min 314 mm
longitudinal axis table
table position forbucky wall stand left
extension rail
min. 314 mm
longitudinal axis tabletable position forbucky wall stand right
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--8 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Place the short rail parallel to the long rail.
S Insert the tool (1 Rail distance tool) (supplied withsystem) to adjust the distance of the rails.
S Measure the distance of the rails using a rule.
S Mark the 2 fixing points on the floor.
S Remove the rail.S Drill the 8 mm holes approximately 75 mm deep in the
middle of the marked area.
S Insert the 2 anchor dowels (supplied with system) intothe holes.
S Move the rail back to the installation position.
S Level the rail using a spirit level. Use the spacer ifnecessary.
S Insert the screw only with washer (supplied withsystem) to the opposite side where the column is tobe inserted.
S Tighten with a spanner.
Coarse Adjustment Tomo optiononly
S Align front and back rails (2, 3) to same level using thespirit level. Use additional spacers if necessary.
For a good performance with thetomo option it is essential to alignfront and back rail very precisely.An alignment test will be madeafter installation of the column.
S Install the tomo switch rail (1).S Insert both end switch cams (2).
S Insert screws with washers and tighten with Allan key.
716 mm
1
shortlong
short
position forbucky wall stand left
extension rail
716 mm
1
shortlong
short
extension railposition forbucky wall stand right
2
3
2
1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--9 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.1. Insertion of column to long rail
S Remove buffer (1) from long rail.
S Insert the column into the rail system (2).
Apply extreme care when handlingthe ball bearings and be careful notto damage the microswitches andoptical sensor B1 positioned inarea 3.
S Install buffer (1) to the rail system.
S Check the leveling of the column in both directionsusing a spirit level.
S Use the spacers if necessary.
S Check whether the column moves easily andsmoothly.
S Insert the second screw with washers into the shortrail.
S Tighten with a spanner.
S Mount the longitudinal center switch UB S17 (pos.3)standard, or position control switch assembly fortomo.
3.2. Installation of tube arm
S Remove transport screw (2) from the counterweighton the front of the column.
Loosen transport screw (1) onlyand move the complete blockdownwards.
S Pull down the transverse arm and secure thecounterweight in its maximum top position byinserting transport screw (2).
1
2
2
3
1
2
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--10 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Route the cables from the top of the column to the leftand right of the column (1).
S Open top cover (2) of the transversal tube arm byremoving the 4 screws 1 (Allan key, 2,5 mm).
Do not remove the screws (2) ontop of the cover. These screws arefixing the brake rail!
S Place the pallet with transverse arm, including tubehousing and collimator, next to the column.
S Lift up the arm and insert it into column (3).
S Insert stop screw (1).S Remove the transport screw. The tube arm must be
easily movable in Z-direction.
S Connect the plug UFX38 (2).
1
2
3
2
11
11
1
52
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--11 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Fasten the cable (4) by using a Ty--Rap.
S Connect UFX38 (3)
S Install the cover of transverse arm (3).
S Install the front cover of transverse arm (5)(see previous page).
S Install the cable support into the top of the column.
S Install the wall junction box.
34
5
6
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--12 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Route the cables on the column as shown in thefigure 1.
S Replace top cover of the transversal tube arm.
S Connect X20 / X22 on stand control UB1 for standardversion.
S Connect X20 / X21 / 22 / 23 / 24 on stand control UB1for tomo version.
Figure 1
** All three items, wall box, column cable support undtube arm, should be located as shown in figure 2(design and function, especially cable lenghts).
*1 Location for ceiling suspension junction box. Thisbox is not included in the system. Please, order thisitem locally as an option.(PMS Artikel No. 9848 500 26201 ceiling suspensionbox for CS, this item is not included and must be orderedoptional).
NoteMax. room height with ceiling suspension box = 3.0 m
Figure 2
UB1
*1 = 0.6 meter
**
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--13 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.3. Final alignment of the column(tomo only)
S Move the column to the left end position.
S Check the alignment of the rails by moving thecolumn smoothly to the other end. The columnmust remain steadily upright over the whole length.If necessary, adjust the level of the front rail withspacers as described in Section “Mounting therails”.
The alignment can be done either with the spirit level orwith the following method:
Final adjustment
S Apply a vertical line to the wall on the right--hand side,e.g. with a colored adhesive tape, Check the verticalalignment while moving the column along the rail.
If corrections should be necessary:
S Add adjustment plates (0) under back (1) or front rail(2), or, for vertical correction (5),
S adjust the ball bearing (3). Three different holes areavailable. For easy access move the column to theleft or right end position.
If the column is not vertical (6), the correction can bemade with an eccentric ball bearing (4).
(6)(5)
(3)
(4)
(2)
(1)
(0)
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--14 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.4. Installation of bucky table
NoteFour (4) persons are necessary to lift the table.
The bucky table is supplied as a complete unit.
S Remove the wooden blocks from the longitudinal sideof the pallet and place them at the rear of the pallet.
S Remove the 4 transport screws and push them down.
Do not lift the table by means of thetabletop frame structure.
S Insert the shoulder straps into the loops (1).
S Lift up the table with the shoulder straps (1).
S Remove the pallet and lower the table carefully.S Place the table with rear end towards the short rail.
See room layout plan.
S Insert screws (1) into the short rail and fit the tableto the rail.
S Drill two 12mm holes approximately 75mm deep forscrews (2).
S Insert the 2 anchor dowels (supplied with system)into the holes.
S Use the spacer if necessary.
S Insert screws (2) supplied with system and washersand tighten with a spanner.
S Level the table in both directions using a spirit level.S Use the spacer if necessary.
S Insert screws (2) supplied with system andwashers and tighten with a spanner.
S Install the foot-switch assembly.
1
1
2
1
table position forbucky wall stand right
2
1
table position forbucky wall stand left
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--15 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Remove the red screws (2) of the transport securingdevice of the tabletop carrier on both sides.
S Remove the tie-wraps form the Bucky device (1)
i
Figure 1
S Remove the cassette tray from the bucky.
S Move the tabletop to its maximum front position (ifit is already mounted).
S Loosen 2 screws at the rear (1).
S Tilt the tabletop unit gently as shown in figure 2.
S Secure the tabletop by fixing the lever (2) via thescrew (3).
S Remove the tie-wraps (4) from the arm braces onboth sides (elevation table only).
Figure 2
Always remove cassette tray beforeopening the tabletop. Tilting thetabletop with inserted tray willcause damages to both table andcassette tray.
Tilting the table top beyond theupright position can causedamages to the cover of the tablebase.
1
2
2
1
2
3
4
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--16 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.5. Installation of Tomo pivot tower(tomo only)
S Position tomo pivot tower towards the correspondingmounting holes at the rear side of the table.
S Thread cable W12 (laser cable) through the centerholes of table base and pivot tower base.
S Insert 4 screws together with washers and toothlock washer (supplied with system) from the tableinside and mount the tomo pivot tower.
S Close the tabletop.
S Secure tabletop with the 2 screws at the rear.
S Reinstall the cassette tray.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--17 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.6. Installation of bucky-deviceextension (tomo only)
S Mount the bucky-device extension and insert the4 screws with washers and tooth lock washers(supplied with system).
Do not support or level the freehanging end of the bucky-deviceextension while tightening thescrews. It will self-align to the filmlevel in a slightly tilted positionwhen the screws are installed.
S Tighten the screws.
3.7. Installation of position controlswitches (tomo only)
Pull corner iron with position control switches out of thecolumn base (they are parked in the column base fortransportation)
S Insert screws with washers (supplied with system)into the column base and fit the position controlswitches as shown in (1).
S Tighten screws with Allan key.
23
14
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--18 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
Move column with extreme care inorder not to damage or shear off themicroswitches and the magneticsensor while passing the cams ofthe tomo switch rail.
S Adjust microswitches for center position andsafety end:
-- The microswitches (2) (center position) and (3)(safety limit switch) must trigger when passingtheir respective cams, as shown in the figurebelow.
S Adjust the proper alignment of the magneticmeasuring sensor: Slowly move the columnalong the tomo switch rail:
-- the magnetic sensor B1 (1) (angle position)must not touch any of the cams. The distancebetweenmagnetic sensor (1) and cam should be 1mm.
The functional test will be performed after power up (seeSection “Checking drive and position control switches”).The detection of any cam position regarding 40° / 20° /8° is indicated with the red LED located as shown in thefigure.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--19 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.8. Installation of OPTIMUS
The following items are different from the OPTIMUS generator:- Firmware TDL file- PCB low speed rotor control
For further information contact the Service Innovation Department in Hamburg.
S Place the X-ray generator cabinet in the position shown in the room layout.
S Place the keyboard in the position shown in the room layout.
For detailed information see manual “Installation of OPTIMUS generator”.
S Check cable connectors on all PCBs (Column Controller UB1, Table Controller SZ1).
S Route the cables from the wall connection box of the column to the X-ray generator cabinet.S Route the cables from the bucky table to the X-ray generator cabinet.
Location at generator site
HT cableStator cableExposure release 1 WAX 1Amplimat cable EZX 22 Table
EZX 31 BWS
Tomo release 1 WAX 11Thermo cable EZX 3Earth ConnectionPower suply 230V
3.8.1. Installation of bucky DIAGNOST TSwith tomo and Medio CP/CP-H
For bucky TS with tomo and Medio CP/CP-H, anInterface board SZ3 is needed. The SZ3 board must beordered separately (= option):
Code No.: 4512 590 30351 Medio CPExposure rel.
Bucky TS & Tomo
SZ3 SZ1
see Z1-4--2
X1X2X1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--20 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.8.2. Programming of APR data
The generator is pre-programmed at factory site.For RGDV programming see 2Z--2.0.If special APR assignements are needed, please, proceed as follows:
S Optimus / Program / Human interface / Predefined assignments
-- without vario focus:AR 46 ENG.TDL = EnglishAR 46 DEU.TDL = GermanAR 46 FRA.TDL = FrenchAR 46 ESP.TDL = Spanish
-- with vario focus for RO 1750:AR 46 V ENG.TDL = EnglishAR 46 V DEU.TDL = GermanAR 46 V FRA.TDL = FrenchAR 46 V ESP.TDL = Spanish
-- and select for Bucky TSRGDV 1 “Bucky GR”RGDV 2 “TOMO BTC” or “TOMO BTS”RGDV 3 “Wall Std GR”RGDV 4 “Free GR”
3.8.3. Programming tube select at control handle
S The tube select signal, e.g. for tube one is selected,is coming from generator as shown in diagram Z1--4with signal RGDV via release decade 1WAX1.
-- Exposure release for RGDV 2 is also coming overthe same 1WAX1 decade.
-- BWS is programmed via 1WAX3 for exposurerelease.
-- Free cassette is programmed via 1WAX4.
S Modification of tube one select for RGDV 3/4:Usually there is no tube select indication whenRGDV 3/4 are selected.If the tube select indication at the control handle isneeded, link X1:5 to X3:5 / X4:5 on 1WA.
In any case the tube 1 select will be present.
X1 X2 X3 X4
1WA
5
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--21 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.9. Installation of tabletop
In rooms with sufficient space on the longitudinal side ofthe table:
S Install the tabletop from the longitudinal side.
S Remove end cover (1) and stoppers from one sideof the tabletop.
S Push the tabletop along the transverse carriage.
Figure 1
Apply extreme care when handlingthe ball bearings.
S Test the easy and smooth movement of thetabletop, if not correct adjust the eccentric bearingbolts.
S Install the stoppers and end cover (1), see figure 1.
3.9.1. Installation of tabletop in small rooms
In rooms without sufficient space on the longitudinalside, the tabletop base unit should be rotated to installthe tabletop.
S Loosen screws (1).
S Remove the locking rings from bolts (2).
S Remove bolts (2).S Rotate the tabletop base unit until the tabletop can
be inserted.
S Remove end cover (1) and stoppers from one sideof the tabletop.
S Insert the tabletop along the transverse carriage.
S Rotate the tabletop unit as required.S Install bolts (2) and secure via locking rings.
S Install the stoppers and end cover (1).
S Fit screws (1).
1
1
2
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--22 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.10. Setting to work
The system is supplied fully prefabricated by the factory.
Before switching on the X-raygenerator:Switch off the auxiliary mainsbreaker for table.
For detailed information see manual installation ofthe OPTIMUS X-ray generator.
S Check the power supply for the table on the PCB SN.
Always remove cassette tray beforeyou open the tabletop. Tilting thetabletop with inserted tray willcause damages to both table andcassette tray.
Tilting the tabletop beyond theupright position can causedamages to the cover of the tablebase.
Figure 1
For access to the table control unit:
S Remove the cassette tray from the bucky.S Move the tabletop to its maximum front position.
S Loosen 2 screws at the rear (1).
S Tilt the tabletop unit as shown in figure 1.
S Secure the tabletop by fixing the lever (2) via thescrew (3).
The table power supply must be 230V, 50Hz or 60Hz.
NoteThere is no programming for 50 or 60Hz. The table canonly operate with 230V (and not with 110V).
1
2
3
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--23 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Follow the setting-to-work procedure described in the manual OPTIMUS X-ray generator XRG 90.
Programming
S Select Registration Devices-- RGDV1
( = registration device of the place for tables 1)
-- RGDV2( = registration device free exposure)
-- RGDV3( = registration device wall stand)
-- RGDV4( = registration device free exposure)
After finishing the described setting-to-work procedure of the OPTIMUS X-ray generator, check the functions belowon the table.
After switching on the auxiliary mains breaker for table
S Check all functions on the table:
-- Green LED on the power pack is on.-- The programmed language.
-- Function of foot-switch assembly.
-- Function of the rocker switches “table up/down”on both sides.
-- Function of end-switch upper position:Motor must stop before mechanical end positionis reached.
-- Function of end-switch lower position:Motor must stop before mechanical end positionis reached.
-- Function of safety switch:Apply about 20 kg of manpower against thetabletop during table down movement.
S Check the strength of upward table movement.S Check whether the range of movement is 270 mm.
S Check the movement of the bucky unit.
S Check the functions of the column:
-- Function of catches in all directions.
-- Function of light on the collimator.
1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--24 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Check the functions of the bucky unit:-- bucky exposure delay
S Check delay time of grid release with test exposuretime = 4 ... 6 msec.
There must not be visible any grid line.
Note:The X-ray tube has been adjusted at factory site.Do not ADAPT the X-ray tube.
In case the generator was not powered on during alonger period -- e.g. more than 2 months --, werecommend to recondition the X-ray tube. SeeOPTIMUS manual, section Installation, 10.1 Tubeconditioning.
S Check coincidence of X-ray beam and light.-- Make a test exposure.
S Check alignment of the stand.
S Check alignment of the bucky wall stand.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--25 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.11. Adjustment procedure for tomo
3.11.1. Adjustment of the table level
S Check whether the table-top is precisely levelled.
S Turn on light and laser beam.
S Adjust laser beam levelling:
-- Raise the table to a position somewhat belowmaximum level. The distance from the laser beamto front and to rear frame of the table-top must beequal, i.e. the laser beam must be plane-parallel tothe table-top.
-- If not, turn laser diode slightly with adjusting screw(1).
S Raise table-top precisely to laser beam levelaccording to figure 1. Table-top position must beslightly below the maximum level that can bereached. If not, adjust laser diode height (2).
If not correct:
S Open and secure the table-top. Set the upper lift limitSNV--S7 switch so that the center of the pivot pointmeets the 0-level of table-top.
3.11.2. Correction for tomo layer height = fineadjustment under X-ray condition
If a tomo test exposer (with a test phantom) deviatesfrom the tomo height, the correction can be done asfollows:
Solution 1 -- valid for table serial No. =500
If the 0mm layer height = the top of table-top differencesto the 0 point of the coupling element. The correctioncanbe done on the axis in both directions.
Solution 2 -- valid for table serial No. =499
Correct the tomo layer height = 0 mm via the table heightSNVS7 switch = table height for the most upper tableposition. After the SNVS7 correction follow up withsection 2.13.2 Adjustment of tomo layer height.
Figure 1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--26 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.11.3. Adjustment of tomo layer height indication
S Set the service switch (1) on the column control boardto service. The following message will appear on thecontrol panel display:
Service mode
Diagnosis: 0
S Press the <TOMO> key on the control panel (1).This will start the calibration menu. Now follow theinstructions on the display:
Service modeMove to h=0
S Raise the table-top up to laser beam level(corresponding to the minimum objective planelevel 0mm).
S Press the <Test> key on the control panel (2). Theposition is now stored in the PLC and the displayshows:
Service mode
Move to h=200
S Lower the table-top to 200 mm, use scale meter(corresponding to the maximum objective planelevel 200 mm).
S Press the <Test> key on the control panel (2). Theposition is now stored in the PLC and the displayshows:
Service modeDiagnosis: 0
You can always quit the calibrationmenu without saving the values bypressing the <TOMO> key again.
1
Figure 2
T
T
2
1
34
Figure 3
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--27 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Leave the service switch on the column control boardin position service to enter the display language.
3.11.4. Setting the display language
S The service switch should be in mode service.
Service modeDiagnosis: 0
S Press the <Tomo angle> key on the control panel (3).The following message will appear on the controlpanel display:
Service mode /V _._Language: english
( V_._ means Softfware--version )
S Press the <Tomo time> key (4) to select anotherlaguage. Repeat until the message appears in thedesired language.
S Press the <Tomo angle> key on the control panel (3)again to quit the menu.
S Set the service switch on the column control boardback to operation.
T
T
2
1
34
Figure 4
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--28 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.11.5. Adjustment of the tomo axis
S Prerequisite: Alignment of standard equipment hasbeen completed.
S Select tomo on control handle.
S Move column, table-top and bucky tray to the centerposition (the center position is reached if the 4LED-indicators on the control handle are lit).
S Move transverse arm to normal position.
S Insert the coupling bar into central and lower swivels(1, 2).
S Fit the coupling bar to the column (3). The locking bolt(4) will be pushed upwards (which will unlock theß-movement axis of the transverse arm).
1. Check that the tomo bar activates the tomo sensorswitch UB S15 when inserted.
2. Adjustment of ball bearing to tomo bar:Ball bearings must be engaged in direct contact tothe tomo coupling bar. So no mechanical clearanceis visible.
2
3
1
4
B
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--29 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Bucky device, light beam localiser and coupling barmust be aligned in a vertical plane.
S If necessary, align the system by adjusting centralswivel (1), lower swivel (2) and position control bar.
3.11.6. Alignment for center line= center of bucky
If the center line 3 is not in center of bucky proceed asfollows.
S The tomo coupling bar is y centered vertical (4).
S Open screws (5) on tomo coupling bar androtate/twist the tube support arm.
S The center line will still be in center of bucky tray (asshown in pos. 3).
3.11.7. Checking drive and position controlswitches
For parameter setting and control please refer to theoperating manual.
Procedure:S Move the column carefully from its center position
towards the left and right end positions and check thefollowing items:
-- the magnetic switch does not touch the positioncams.
-- the micro switches S17/S20 are not damaged orsheared off when passing a cam.
-- the magnetic switch is triggered at each angleposition cam and the end position cams. This canbe checked by observing the red LED on themagnetic switch B1.
2
3
1
5
UBS17
S20
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--30 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Test the safety limit switches: Move the columnmanually to the left and right limit switch positions.The limit switch must be triggered (I/O--LED 35 ontomo controller (PLC) lits) and the system beswitched off.
S Initiate a tomo test cycle on the control panel:Select <Tomo> mode with 40°/4s. Press the <Test>key until the cycle is completed and check thefollowing items:
-- the column moves first to the left (1).
-- the column moves at constant speed and withoutjerks (otherwise check: cable routing, railalignment, dirt on rail surface, spring pressure,driving axle).
-- the eccentric disk operates correctly (otherwisecheck coupling motor (M 3) and position switch UBS16 or both springs are ”fixed too tight”.Therefore, check the distance from screw head tothe nut as shown in the figure. The distance asshown must be 15 ± 0.5mm.
-- the driving wheels grip firmly (otherwise increasespring pressure).
-- the column reverses automatically at the endposition on each side.
-- the bucky device moves smoothly and withoutjerks (otherwise check alignment of the couplingbar).
-- the position control switches (magnetic switch andmicroswitches) operate correctly, see first step.
-- the position of the coupling bar is vertical aftercompletion of a cycle.
-- the distance between control switch B1 and tomoposition cams should be 1mm.
-- If the sequence is aborted, check the function ofthe position control switches (see first step).
-- If the sequence is aborted due to timeout (errorcode 30, see Chapter ”trouble shooting”), increasethe motor supply voltage by switching to the nexthigher transformer tap (see schematic diagramZ1--2).
S Repeat the tomo test cycle with 40°, 20° and 8°sweeps, with high and low speed settings. Again,check whether any anomalities occur during the test.
1
S16
15 ± 0,5mm
M2
M3
Pivot point
Coupling motor
Tomo motor
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--31 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.12. Tomo adjustment under exposureconditions (tomo only)
3.12.1. Measurement of X-ray zone
Two exposures are to be taken on the same film.
First exposure = BuckyS Set the parameters for exposure 1 (without tomo):
-- Small focus
-- SID = tomo SID
-- X-ray area = 200 mm x 200 mm-- Exposure values = 40 kV / 6.3 mAs
Second exposure = Tomo
S Set the parameters for exposure 2 (with tomo):-- Small focus
-- SID = tomo SID
-- X-ray area = 150 mm x 200 mm
-- Tomo height = 130mm-- Exposure values = 40° / 50 kV / 12.5 mA
-- Tomo time = 2 s
S Take the exposure in tomo mode (tomo on) anddetermine the following values:
-- a = _____mm
-- b = _____mm
S Check it these values meet the following conditions:
-- ∆ a, b = ≤ 10 mm
-- Σ a, b = ≤ 80 mm
S If the above conditions are not fulfilled, check whetherthe free movement of the α--drive and bucky moduleis impaired.
S If under-voltage is indicated on the generator controlboard, the motor will supply voltage by switching tothe next higher transformer tap (see schematicdiagram Z1--2).
a b1
2
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--32 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.12.2. Symmetry of tomo axis
S Position the lead plate - 100mm above the X-Raycassette.
The X-ray cassette ist positioned on the table-top.
S Collimate the X-ray field (1) as far as the ”hole” will beexposed during the whole tomo sweep.
S Take an exposure in center position (bucky mode) =70 kV / 10 mAs
S Take an exposure in tomo mode. 70 kV 2 sec 5 mA40° .
The result should look like as shown in the figure:
-- c ≈ d, i.e. the generator should switch off shortlybefore the system.
-- No undervoltage indication should appear on thegenerator.
S Check whether these values meet the followingconditions:
-- |c|--|d| < 5mm.
S If the result is not symmetrical,proceed as follows:-- In case of overvoltage lower the tomo speed by
selecting a higher resistance with jumpers onboard SZ2 = fine adjustment, i.e. reduce the supplyvoltage of the drive or reduce the voltage attransformer T2 side for slow speed = coarse adj.17.5V / 15.5V / 13.5V ore available. For high speedprogr. 31V / 33V or 37V.
-- In case of unsymmetry (c < d) increase the tomospeed by selecting a lower resistance with jumperson board SZ2., i.e raise the supply voltage of thedrive
¯1
X-ray cassette
100mm
∅
Lead plate
Table top
dc
3 4
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--33 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
S Programming of symmetriy (please fill out the table)-- The following programming is selected:
U OnT2
R = OON SZ2
value for c and d in mm:
High speed 37V O
33V O
31V O
................ O
c = .....................mm
d =.......................mm
Low speed 17.5V O
15.5V O
13.5V O
................ O
c = .....................mm
d =.......................mm
Remarks: ⇒ = coarse adj. = fine adj. tested with 40° 4 sec and40° 2 sec
S If necessary, readjust the tomo axis (see Section3.13.4 Adjusting the tomo axis).
S Repeat the above test.
3.12.3. Tomography resolution test
S Take an exposure with the TOMO test phantom withfollowing parameter:
TOMO angle TOMO time TOMO height Actual valueResolution lp/mm
40° 4 sec 50 mm40° 2 sec 50 mm40° 4 sec 100 mm40° 2 sec 100 mm
-- Minimum resolution = 2.32 lp/mm with smallfocus.
+/-- 5 mm
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--34 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.12.4. Tomo layer height
(Use TOMO test phantom)
S Take the following test exposures with inclined buckytray.
Tomo angle Tomo time TOMO layer height --default= nominal value
Measured TOMOlayer height= actual value
Deviation(+/-- 3mm max.) =nom.--actual value
40° 4 seconds 45 mm mm mm
40° 4 seconds 130 mm mm mm
40° 4 seconds 200 mm mm mm
If the deviation is more than +/-- 3mm, please repeatethe table calibration procedure as follows:
1) Raise the table hight to theeffective (phantom measured)zero-level. In the display you willsee the amont of the deviation.
2) Follow the instruction part 3.13.2Calibration of the level sensor.Note that your table stays on levelzero, do not move the table to zeroas requested in the displayinstruction.
3) Adjust the laser beam.
3.13. Final work
S Install all rail covers.
S Connect the earth wiring to all covers.S Install all column covers.
S Close the X-ray generator cabinet.
S Clean all covers with a mild cleaner e.g. soapsuds.
S Check the labelling (See drawing Z--3).
11
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS INSTALLATION
bucky DIAGNOST TS (99.0) 2--35 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.14. Acceptance test
See national regulations as e.g. Röv / DOD / VA
3.15. Handing over to customer
The modified system is handed over
on ( date) ...................
to ( customer) ................................................
by (system engineer) ................................................
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSINSTALLATION
bucky DIAGNOST TS(99.0)2--36 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_2_990
3.16. Installation for tomo bar parking frame
Tomo bar
Wall mounting
65 cm ± 2 cm
10 cm ± 2 cm
Floor
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
2Z--2.0
OPTIMUS R. 2.x / 3.x
A4
00--12--04
Re
bdts_2z2_0_010
RGDV programming
E1999 Philips Medical Systems
ALL RIGHTS RESERVED
(01.0)OPTIMUS CDOPTIMUS Rbucky DIAGNOST TS
-Nam
e:
- DataSetA:
Desk:
RGDV1
RGDV2
RGDV3
RGDV4
RGDV5
RGDV6
RGDV7
RGDV8
--Room:
11
11
--Tube:
11
11
--Release
circuitnum
ber:
11
31
--Enablehandsw
itchatgeneratordesk:
yes
yes
yes
yes
--Syncm
asterpresent:
yes
yes
yes
no--Exposuresw
itchtype:
doublestep
doublestep
doublestep
doublestep
--Bucky
formatdensity
correction:
00
00
--Conedensity
correction:
00
00
--Dosemeasurementinput:
EZX21
none
EZX31
none
--Dosemeasurementsensor:
--Tomomovem
ent(Nobreakafterexposureend):
noyes
nono
--Release
delay:
enable
enable
enable
enable
--Mounted
radiographicalcontroller:
none
none
none
none
--Release
circuitadaptionunit:
1WA
1WA
1WA
none
--Mounted
tomoextension:
none
1WA
none
none
--MediumIIformatkV
correction(doseequiv.steps):
00
00
--MediumIIformatdensity
correction(6%steps):
00
00
--MediumIIformatmAscorrection(6%steps):
00
00
--SmallIIformatkV
correction(doseequiv.steps):
00
00
--SmallIIformatdensity
correction(6%steps):
00
00
--SmallIIformatmAscorrection(6%steps):
00
00
DataSetB:
--Usedfortomo:
nono
nono
--Disabletim
eoverride:
nono
nono
--Tube
powerfactor:
100
100
100
100
--kV
steps:
dose
equivalent
dose
equivalent
dose
equivalent
dose
equivalent
--mAssteps:
25%
25%
25%
25%
--mAsteps:
25%
25%
25%
25%
--tim
esteps:
25%
25%
25%
25%
--Densitysteps:
12%
12%
12%
12%
--Densitycorrection(6%steps):
00
00
--Underexposuredisplay:
yes
yes
yes
yes
--Tube
overload
protection:
onon
onon
Bucky/S
copo1W
B/D
ecadeBucky1(W
BX11):
Bucky/Tomo1W
A:DecadeBucky1/2
WAX11
WAX12
Bucky/Tomo1W
A\Tomotime[s]:
Tomomodesw
itch
enable
Tomotim
e1:
Tomotim
e5:
RGDV1[]RGDV2[]RGDV3[]
RGDV4[]
Bucky
RGDV--sw
itchrelated
RGDV1
Tomotim
e2:
Tomotim
e6:
Bucky
RGDV:
RGDV4[]
RGDV5[]RGDV6[]RGDV7[]
Bucky
RGDV
NONE
Tomotim
e3:
Tomotim
e7:
ucy
GRGDV5[]RGDV6[]RGDV7[]
RGDV8[]
Bucky
RGDV
NONE
Tomotim
e4:
Tomotim
e8:
RGDV8[]
TomoRGDV--sw
itchrelated
RGDV2
Timesetting
forinputatW
AX21:1......8
ForWBX11
:9--10
(ready)und1--2(formatsize
correction
contact)
Tomomodesw
itch:X
11:3SL_XG_TO/B
ucky
RGDV:X
11:1Format+:10Bucky
ready/Tom
oRGDV:X
11:1Format+:5Tomoready
switchrelatedX11:3---->
Bucky
--Tomoremotesw
itchoverRGDVs
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
Labellingbucky DIAGNOST TSA1
/A3
03-0
2-24
Sch
r.
4512 983 06151 C Philips Medical Systems
(b/03.0) 2Z-10
LASER RADIAT IONDO NOT STARE INT O BEAM
MAX. OUTPUT < 1 mWEMITTED WAVELENGTH 650 nmFDA CLASS II LASER PRODUCT
C A U T I O NLASER RADIATION
DO NOT STARE INT O BEAMCLASS 2 LASER PRODUCT
IEC 825-1: 1993MAX. OUTPUT < 1mW
EMITTED WAVELENGTH 650 nm
LASER RADIAT IONDO NOT STARE INTO BEAMCLASS 2 LASER PRODUCT
IEC 825-1: 1993MAX. OUTPUT < 1mW
EMITTED WAVELENGTH 650 nm
2 3
4a 4b
6d
6c7
9
10
8b, 9, 20
11
5
4a or 4b
4
9, 7, 6c or d
Remove cassette tray andmove tabletop i n the f ront posi tion before tilting the table
11
10
K
B
16; 17 (a, b, c, d or e)
18 19
1
15
14
14
1819
or
12a
12b
12a or b
13a 13b
13a or b5 Kg
or
2 or 3
115
P H I LI P S
type 9848 600 0264.s/n yy.00.nnnTS Family PEI
230V 50/60Hz 5A
PHILIPS
Class I - TYPE BIEC 60601-1
5
PH I LI PS
type 4512 201 0094.s/n yy.00.nnntube stand fixed arm
PHILIPS
8b
8a
Phil ips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
X-RAY TUBE HOUSING ASSEMBLY SRO 33100 ROT 350type 9874 005 16122 150kVs/n xxxxx-xxxxxx
X-RAY TUBE SRO 33100type 9874 206 70102 0,6 IEC 336s/n xxxxxx 1,2 IEC 336
PERMANENT FILTRATION MANUFACTURED 2,5 AL / 75
X-RAY TUBE HOUSING ASSEMBLY SRO 2550 ROT 350type 9874 004 23122 150kVs/n xxxxx-xxxxxx
X-RAY TUBE SRO 2550type 9874 206 11102 0,6 IEC 336s/n xxxxxx 1,0 IEC 336
PERMANENT FILTRATION MANUFACTURED 2,5 AL / 75
X-RAY TUBE HOUSING ASSEMBLY SRO 0951 ROT 350type 9890 000 63181 150kVs/n xxxxx-xxxxxx
X-RAY TUBE SRO 0951type 9890 000 64191 0,3 IEC 336s/n xxxxxx 1,0 IEC 336
PERMANENT FILTRATION MANUFACTURED 2,5 AL / 75
MANUFACTUREDPERMANENT FILTRATION2,5 AL / 75
X-RAY TUBE HOUSING ASSEMBLY RO 1750 ROT 350type 9890 000 85281 s/n xxxxx-xxxxxx
X-RAY TUBE RO 1750type 9800 000 85271s/n xxxxxx
0,6 IEC 3361,2 IEC 336
150 kV
MANUFACTUREDPERMANENT FILTRATION2,5 AL / 75
X-RAY TUBE HOUSING ASSEMBLY RO 1648 ROT 350type 9890 000 85301 s/n xxxxx-xxxxxx
X-RAY TUBE RO 1648type 9800 000 85291s/n xxxxxx
0,7 IEC 3361,5 IEC 336
150 kV17a
17b
17c
17d
17e
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Ha mburg/Germany
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
type 4512 201 0103.s/n yy.00.nnnbucky device left
type 4512 131 17424casset te tray left
type 9848 600 0209.s/n yy.00.nnn
This product complies with the DHHSrequirements of 21 CFR Sub-Chapter JManufactured: month yyyy
USC®
0,3 Al / 75
v ia Schiapparelli 27 / 33200035 Liss one (M i) Italy
type 9848 600 0017.s/n yy.00.nnnManufactured: month yyyy
USC®
0,3 Al / 75
v ia Schiapparelli 27 / 33200035 Lissone (Mi) Italy
P H ILI P S
type 9848 600 0006.s/n yy.00.nnnbucky table fi x
PHILIPS
PH I LI PS
type 9848 600 0023.s/n yy.00.nnntube stand
PHILIPS
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Ha mburg/Germany
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Ha mburg/Germany
P HILI P S
type 9848 600 0008.s/n yy.00.nnnTOMO Equipment
PHILIPS
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
This product complies withthe DHHS requirements of21 CFR Sub-Chapter JManufactured: month yyyy
Phil ips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
type 9848 600 0208.s/n yy.00.nnncontrol grip tomo/laser TS
type 9848 600 0207.s/n yy.00.nnncontrol grip laser TS
Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/GermanyThis product complies wi ththe DHHS requirements of21 CFR Sub-Chapter JManufactured: month yyyy
This product complies wi ththe DHHS requirements of21 CFR Sub-Chapter JManufactured: month yyyy
FP-64/2LF-Ö60-F450-K200-STs/n_ _ _ _ _ _ . _
PHILIPS
USC®
Philips Medical SystemsDMC GmbHRöntgenstr . 24D-22335 Hamburg/Germany
24V 350mA
~
0123
0051 0051
16Philips Medical SystemsDMC GmbHRöntgenstraße 24D-22335 Hamburg/Germany
® LR37323
C US
type 9890 010 8137.s/n yy.00.nnntabletop 2m
PHILIPSPhilips Medical SystemsDMC GmbHRöntgenstraß 24D-22335 Hamburg / GermanyThis product complies withthe DHHS requirements of21 CFR Sub-Chapter JManufactured: month yyyy
type 9890 010 0219.s/n yy.00.nnntabletop
PHILIPSPhilips Medical SystemsDMC GmbHRöntgenstraß 24D-22335 Hamburg / GermanyThis product complies withthe DHHS requirements of21 CFR Sub-Chapter JManufactured: month yyyy
LASER RADIAT IONDO NOT STARE INT O BEAM
MAX. OUTPUT < 1 mWEMITTED WAVELENGTH 650 nmFDA CLASS II LASER PRODUCT
C A U T I O N8a
P H ILI P S
P H ILI P S
type 9848 600 0007.s/n yy.00.nnnbucky table elevation
PHILIPS
Made in GermanyPHILIPSPHILIPS
PH I LI PS
PH I LI PS
STREU STRAH LEN RASTERGRILL E AN TI OFFUS ANTE
Lp/cm
fo(cm)
ratio
X-R AY GR ID
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULFINDING
bucky DIAGNOST TS (99.0) 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990_inh
3--0.1
FAULTFINDING
TEXT
Contents 3--0.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. Trouble shooting 3--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.1. Safety regulations 3--1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.2. Software and PLC faults 3--2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.3. Hardware and tomographic image problems 3--5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4. Specific instructions for fault elimination 3--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4.1. EPROM replacement 3--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4.2. Setting EPROM configuration jumper 3--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.4.3. Test exposure with inclined bucky tray 3--6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5. PLC input / output allocation 3--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5.1. Digital inputs 3--7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.5.2. Digital outputs 3--8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6. Indicator LED’s on table control board 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.1. Standard version 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.1.1. Signal exchange between generator and table control 3--9. . . . . . . . . . . . . . . . . . . . . .1.6.1.2. Power supply 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.2. Additional LED’s for Tomo version 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.2.1. Signals from PLC to table control 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.2.2. Signals from table control to PLC and generator 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . .1.6.2.3. Power supply 3--9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULTFINDING
bucky DIAGNOST TS (99.0) 3--1 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1. Trouble shooting
1.1. Safety regulations
Only technically qualified personnel who have received the appropriate instruction and training for radiographyequipment in medical applications and for the bucky Diagnost TS in particular are allowed to install, maintain andservice this equipment.
It is presumed that all personnel involved with trouble shooting are familiar with fault finding and repair of standardradiography equipment. Therefore only topics related to the process controller (PLC) and tomo specific problemsare treated in this chapter.
Furthermore, all personnel involved must be instructed on safety precautions with respect to radiography and therelated high voltage devices as well as to all other electric low voltage equipment. All relevant standards andregulations must be complied with.
The local service organisation is responsible for the safety and training of personnel and prevention of materialdamage.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSFAULTFINDING
bucky DIAGNOST TS(99.0)3--2 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.2. Software and PLC faults
s In order to display software error messages, set the service switch on the column control board to service. Thefollowing message indicating an error code will appear on the control panel display:
Service modeDiagnosis: nn
s Follow the relevant suggestions in the table below to eliminate the disturbance.
s Clear error message by pressing . If the system is o.k., he following message will appear:Service modeDiagnosis: 0
s Contact the Help--Desk of your local service organization if the fault cannot be corrected.
Code Fault description Possible cause Suggested remedy Notes
-- Display not active -- RS232 serial link tocontrol panel dis-turbed
Check cable and plugs
-- LED ”supply” (yellow) not on Power supply missing 1. Check if power supply is on2. Check supply cable and
connection
-- LED ”Run” (yellow) not on -- Application programnot loaded
-- EPROM missing
1. Check EPROM2. Download application
program
3. see error code 104
0 no error message
SZ1 X1
X2
X3
X11
VA2X2
E1WA
X1
X11
SPStomo controller
bucky
UE2X30
coupling
M1 M2
tomomotor
tomo
UB1
Table controllerControlhandle
B1
40° 20° 8° 8° 20° 40°
SZ1
X1
X2 E1WA
X1
X11
Tablecontroller
Tomo mode
Tomo ready
Format contact
bucky ready
PO_26V_RE
GND
XG_RD_EX
12345678910
12345678910
RQ_XG_EX
RQ_XG_PR
RG_DV_1
1
1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULTFINDING
bucky DIAGNOST TS (99.0) 3--3 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
Code NotesSuggested remedyPossible causeFault description
1 Communication timeout: noanswer from PIC--proces-sors on table controller (de-centralized I/O)
-- Cable to table con-trol not correctlyconnnected
-- Interface module onPLC defective
-- RS485 bus nor cor-rectly connected
-- Jumpers on RS485interface board notset correctly
-- PIC--processors orinterface module de-fective
-- EPROM defective
1. Check cable and plugs2. Check position of jumpers
on RS485 interface board(must be set to CLOSED,see wiring diagram Z2--6)
3. PLC--reset:
s Turn PLC off
s Disconnect power supplycable of PLC (on thebottom, to the left of the 3LED)
s Switch PLC on
s Wait a few seconds
s Reconnect the powersupply cable
4. Check the hardware andreplace if necessary
2 Incorrect answer from PIC--processors on table control-ler
see code 1 see code 1
30 Tomo sequence timeout (se-quence time limit exceeded)
-- Tomo motor blocked-- Tomo motor is not or
deficiently engaged
-- Speed too low
-- Motor supply voltagetoo low
-- Angle and limit posi-tion control switchdisturbed
-- Center positionswitch disturbed
1. Check tomo drive (motor,driving wheel, eccentricdisk, see Section 4.12.4”Checking drive andposition control switches”)
2. Check motor supplyvoltage:voltager range: ....... If toolow, change tap of supplytransformer
3. 1. Check alignment ofposition control switches(see Section 4.12.4”Checking drive andposition control switches”)
31 Column did not stop in thecenter position after comple-tion of a tomo cycle, i.e. col-umn overran the center posi-tion
-- After running toohigh
-- Center positionswitch disturbed
1. Check function of columnbrake
2. Check center positionswitch (incl. Cable andconnections)and center position cam.
3. Check column stop control
100 PLC system RAM test failure -- PLC hardware fault 1. Replace PLC
101 PLC RAM test failure -- PLC hardware fault 1. Replace PLC
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSFAULTFINDING
bucky DIAGNOST TS(99.0)3--4 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
Code NotesSuggested remedyPossible causeFault description
104 PLC Program EPROM testfailure
-- EPROM defective-- EPROM not cor-
rectly installed
-- PLC configuration in-correct
1. Replace EPROM2. Check EPROM position
(see Section EPROMreplacement)
3. Check EPROMconfiguration (see SectionSetting EPROMconfiguration jumper)
120 PLC serial interface test fail-ure
-- Interface module de-fective
-- PLC hardware fault
1. Replace interface modulein PLC
2. Replace PLC
140 PLC system EPROM testfailure
-- PLC hardware fault 1. Replace PLC
252 Error flag is set -- Error in PLC applica-tion program
1. Check program release(must match PLC versionand generator type)
2. Replace EPROM
253 Index register overflow -- Error in PLC applica-tion program
1. Replace EPROM
254 Program timeout -- Error in PLC applica-tion program
-- PLC defective
1. Replace EPROM2. Replace PLC
255 Software watchdog error orpower supply error
-- Error in PLC applica-tion program
-- PLC defective
1. Replace EPROM2. Replace PLC
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULTFINDING
bucky DIAGNOST TS (99.0) 3--5 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.3. Hardware and tomographic image problems
Fault description Possible cause Suggested remedy Notes
Tomo sequence aborted -- Tomo speed too low-- see error code 30 in
Section Software andPLC faults
1. See error code 30 in SectionSoftware and PLC faults
Underexposure (indication ongenerator control desk)
Tomo speed too high 1. Check supply voltage andchange transformer tap, ifnecessary
Mains voltage too high 1. Install transductor or voltagestabilizer (contact serviceorganization to order)
Objective plane level not correct -- Calibration maladjust-ment
-- Laser misalignment
1. Check laser alignment (seeSection Adjusting table level)
2. Calibrate level sensor (seeSection Calibrating levelsensor)
3. Check level indicatorpotentiometer in table base:pinion axle must be rectangularto toothed belt
4. Replace level indicatorpotentiometer
Resolution too low with large ob-jective plane levels
-- Coupling bar has toomuch play
-- Uneven motion ofcoupling bar
-- Uneven motion ofbucky device
1. Check coupling bar(movement, fixations, swivels;see Section Adjusting the tomoaxis)
2. Check bucky device:
-- bearings
-- twisted rails
-- dirt on rails-- collision with cables
Generally low resolution, evenwith low objective plane levels
In addition to thepossibil-ities above:-- Column movement
not properly alignedwith bucky device
-- Column jerks
-- Column drive notproperly adjusted
-- Transverse arm yaws(spring pressure forb-axis lock too low)
-- Exposure time toolow or too high
1. Make test exposure withinclined bucky tray (seeSection Test exposure withinclined bucky tray) andevaluate result
2. Make test exposure withgrating evaluate result
3. Check tomo drive (motor,driving wheel, eccentric disk,see Section Checking driveand position control switches)
4. Increase spring pressure forβ-axis lock
5. Check speed adjustment(exposure time coordinationwith generator)
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSFAULTFINDING
bucky DIAGNOST TS(99.0)3--6 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.4. Specific instructions for fault elimination
1.4.1. EPROM replacement
s Switch off PLC power supply.
s Wait 1 minute, then remove the cover.
Protection against static electricity:an earth strap must be connected betweenthe wrist and the base of the column beforethe EPROM is removed. The EPROMcould otherwise be destroyed by staticelectricity.
s Replace EPROM, pins 1 and 2 of the socket(see Figure) must be left unused.
s Replace cover and switch power on.
1.4.2. Setting EPROM configuration jumper
s Switch off PLC power supply.
s Wait 1 minute, then remove the cover.
s Set jumper J1 to ”E”.
s Replace cover and switch power on.
1.4.3. Test exposure with inclined bucky tray
s Place bucky tray on table--top as shown in the figure.
s Adjust collimator aperture according to the figure.
s Make test exposures in tomo mode and interpret theresults.
70 Collimator aper
1 mm
max.
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULTFINDING
bucky DIAGNOST TS (99.0) 3--7 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.5. PLC input / output allocation
1.5.1. Digital inputs
Signification when LED ON
Operation mode Service mode
EA--0
¡ 0 Tomo on Calibration
¡ 1 Tomo test Confirmation of language/level
¡ 2 Set tomo angle Language select
¡ 3 Set tomo time Error reset
¡ 4 Raise table top Raise table top
¡ 5 Lower table top Lower table top
¡ 6 Light beam localiser ON
¡ 7 FFA ok
Operation mode / Service mode
EA--16
¡ 16 x-axis center lock (transverse arm extension)
¡ 17 y-axis center lock (column movement)
¡ 18 Service switch ON
¡ 19 spare
¡ 20 spare
¡ 21 spare
¡ 22 spare
¡ 23 spare
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSFAULTFINDING
bucky DIAGNOST TS(99.0)3--8 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.5.2. Digital outputs
Signification when LED ON
Tomo mode Standard mode
EA--32
¡ 32 Radiation ON command 1 (redundant) (permanently on)
¡ 33 Radiation ON command 2 (redundant) (permanently on)
¡ 34 Unlock β--movement axis
¡ 35 Activate brake y-axis (column movement)
¡ 36 Release brake y-axis (column movement)
¡ 37 Disable center position locks
¡ 38 spare
¡ 39 spare
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS FAULTFINDING
bucky DIAGNOST TS (99.0) 3--9 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_3_990
1.6. Indicator LED’s on table control board
1.6.1. Standard version
1.6.1.1. Signal exchange between generator and table control
H12 (yellow) Signal XG_RD_EX (Exposure) from generator (LED on = signal low)
H15 (yellow) Signal RG_DV (device active) from generator (LED on = signal low)
H16 (red) Signal RQ_XG_EX (Radiation on) to generator (LED on = signal low)
1.6.1.2. Power supply
H17 (green) Supply electronics 24VDC
H18 (green) Supply light beam localiser 24VAC
H20 (green) Supply motors 24VDC
1.6.2. Additional LED’s for Tomo version
1.6.2.1. Signals from PLC to table control
H1 (green) Speed: low motor speed
H2 (green) Dir: motor direction reverse
H3 (green) On1: motor1 on
H4 (green) On2: motor2 on
H5 (green) Laser: objective plane indication on
H6 (green) Bucky_contr: release bucky brake
H7 (green) Bucky_start: trigger bucky device
H8 (green) Tomo_ready: tomo start position reached
H9 (green) Tomo_mode: tomo mode active
H10 (green) Watchdog
1.6.2.2. Signals from table control to PLC and generator
H11 (yellow) Signal RQ_XG_PR (Preparation) from generator
H12 (yellow) Signal XG_RD_EX (Exposure) from generator
H13 (yellow) Coupling bar: coupling bar inserted to PLC
H15 (yellow) Signal RG_DV (device active) from generator
H16 (red) Signal RQ_XG_EX (Radiation on) to generator
1.6.2.3. Power supply
H17 (green) Supply electronics 24VDC
H18 (green) Supply light beam localiser 24VAC
H19 (green) Supply motors 12VDC
H20 (green) Supply motors 24VDC
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS (99.0) Z0--1 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_z01_990
Drawings
Block diagram standard version Z1--1
Block diagram tomo version Z1--1.1
Power supply standard version Z1--2
Power supply tomo version Z1--2.1
Table control standard version Z1--3
Table control tomo version Z1--3.1
Bucky control standard version Z1--4Bucky control tomo version Z1--4.1
Exposure Interface SZ3 Z1--4.2
Stand control standard version Z1--5
Stand control tomo version Z1--5.1
Bucky DIAGNOST VE with bucky unit non sensing Z1--6.1
Stand control tomo version Z1--7.1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Block diagramA3
00-
02-2
3 O
st.
bucky DIAGNOST TS C Philips Medical Systems
(99.0) Z1-1
See table control
See table control
ST
H.V. cable
temperature switch
EGH.V. cable
stator cable, screened
RS
232
PGU
EA0
X30
EA16/32X5XP1
RS 485
X21
X41
X23
X5
X10
X4
X7
X22X20 X31
X22X31 X24
X37
X38
X1X32
X36
X35
X40
X11
X8
X3
X1
X2
X6
X9
X5
X4
X1
X2
X4
X11WAX3
EZX31
SX..EZX21
X21WAX11
EZX3
EX1100
ENX2100
1WAX1
X16
X1
M
SNV
tomo driveUB
UELCD display
PLC
SA
SA
VA
PH
LA
SNSZ1
SC
UB1
UB
SNV
UF
UEXA
PLCtomo control
see Z1-7
see Z1-7
tomo height
table height
UB1stand control stand control
brakes
UBcenteredUB1
brakes
brakes
UF
UE2control handle
centered
brakes
UE
light
UE1
see OPTIMUS manual
stand bucky DIAGNOST TS
see Z1-2
table bucky DIAGNOST TS
PHamplimat chamber
VAbucky module
SNV
brakes
centered
SA
SCfootswitch
table heightand
SA
brakes
table height
power supplySN
bucky moduleVA
amplimat chamberPH
bucky DIAGNOST VEVP
wall connection box
amplimatchamber 2
release 2
amplimatchamber 1
bucky 1
formatcontacts
release 1
XA tube assembly
E generator
mains230V~
Laser (option)
LA manual collimator
field
table controlSZ1
X49
X32
X1
Upgrade kit Center Laser
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Block diagramA3
00-
02-2
3 O
st.
bucky DIAGNOST TS C Philips Medical Systems
(99.0) Z1-1.1
H.V. cable
temperature switch
EG H.V. cable
stator cable, screened
PGU
EA0
X30
EA16/32X5XP1
RS 485
X21
X41
X23
X5
X10
X4
X7
X22X20 X31
X22X31 X24
X37
X38
X1
X1
X49
X32
X36
X35
X40
X11
X8
X3
X1
X2
X6
X9
X4
X1
X2
X4
X11WAX3
EZX31
SX..EZX21
X21WAX11
EZX3
EX1100
ENX2100
1WAX1
X16
X1
M
SNV
tomo driveUB
UELCD display
PLC
SA
SA
VA
PH
LA
SNSZ1
SC
UB1
UB
SNV
UF
UEXA
PLCtomo control
see Z1-7
see Z1-7
tomo height
table height
UB1stand control stand control
brakes
UBcenteredUB1
brakes
brakes
UF
UE2control handle
centered
brakes
UE
light
UE1
see OPTIMUS manual
stand bucky DIAGNOST TS
see Z1-2
table bucky DIAGNOST TS
PHamplimat chamber
VAbucky module
SNV
brakes
centered
SA
table heightand
SA
brakes
table height
power supplySN
bucky moduleVA
amplimat chamberPH
bucky DIAGNOST VEVP
wall connection box
amplimatchamber 2
release 2
amplimatchamber 1
bucky 1
formatcontacts
release 1
XA tube assembly
E generator
mains230V~
Laser (option)
LA hand collimator
field
table controlSZ1
footswitchSC
X5
tomo speedcontrol
SZ21/2
MEDIO CPrelease adapter (option)
SZ3
X1 X2
Upgrade kit Center Laser
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Power supplyA3
99-1
1-23
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-2
1.6A slow
F14
H.V. cable
temperature switch
EG H.V. cable
stator cable, screened
26VAC
28VAC
0VAC
230VAC
230VAC
0VAC
4A slow
X16:3
X16:4
X16:1.1
X16:2.1
EZX3
EX1100
X32:4
X32:3
X31:04
X31:07
X22:04
X22:07
X31:01
X31:12
X22:01
X22:12
+24V
30VAC
0VAC
24VAC
0VAC
X20:06
X20:08
+24VX20:01
X20:12
6.3A slow
6.3A slow
ENX2100:L1
ENX2100:N
5A slow
5A slow
X11:12
X8:5
X11:01
X8:4
X8:6
X8:8
L1.1 X11:06
X11:08
X8:2
X8:3
L1
N.1N
X16:1
X16:2
1L1
5N
24VAC
0VDC
5VDC
0VAC
+24V
24VAC
0VAC
X49:2
X49:1
X1:3
X1:4
Factory adjusted
F6
H17
H18
H20
XA tube assembly
E generator UE control handleUB stand controltable controlSZ
F2
F5
F12
F11
24VDC/5A
T1Z1
mainsfilter
mains230V~
switchingpower supply
G1
LAmanual collimator
SN power supply
X1:1
X1:2
centerlaser(optional)
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo versionPower supplyA
399
-11-
23O
st.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-2-1
0VDC
5VDC
X49:2
X49:1
X1:2
X1:1
F146.3A slow
H.V. cable
temperature switch
EG H.V. cable
stator cable, screened
+24VDC
GND
15.5VAC
13.5VAC
0VAC0VAC
X8:8
6.3A slow
34VAC
31VAC
26VAC
28VAC
0VAC
230VAC
230VAC
0VAC
4A slow
X16:3
X16:4
X16:1.1
X16:2.1
EZX3
EX1100
X32:4
X32:3
X31:04
X31:07
X22:04
X22:07
X31:01
X31:12
X22:01
X22:12
+24V
30VAC
0VAC
37VAC
17.5VAC
X20:06
X20:08
+24VX20:01
X20:12
6.3A slow
6.3A slow
ENX2100:L1
ENX2100:N
5A slow
5A slow
X11:12
X8:5
X11:01
X8:4
X8:6
X8:7
L1.1 X11:06
X11:08
X8:2
X8:3
L1
N.1N
X16:1
X16:2
1L1
5N
24VAC
0VAC
+24V
31VAC
15.5VAC
X1:3
X1:4
PLC tomo control
H19
F1
Factory adjusted
F6
H17
H18
H20
XA tube assembly
E generator UE control handleUB stand control
table controlSZ
F2
F5
F12
F11
T2
24VDC/5A
T1
mainsfilter
mains230V~
switchingpower supply
G1
LAhand collimator
SN power supply
centerlaser(optional)
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Table controlA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-3
k4 Offk4 ON
k1/k2 ONk1/k2 Off
Table downTable up
M-+
1
2
1a
1
1b
1b
1
1a
DOWN
ENABLE
+24V
&
& DOWN
ENABLE
+24V
DOWN
ENABLE
+24V
table motor
X6:10
X6:11
brake
+24V+24V
table on2
1
2 down
1
2 down
1K1
ENABLE
+24V
DOWN
ENABLE
ENABLE
+5V
B
A
1
41
2
B
A
2
1
3
4
footswitch
2
1
X10:3
X10:2
X4:07
X4:06
2
1
1
2
+24V
+5V
U2
+24V
JP1
1
4
1
2 2
1
F6
4A slow
buckycentered
bucky brakerelease
SWGND
X20:09
X20:01
X20:12
X20:06
X20:08
X11:09
0VAC
24VAC
24VAC
0VAC
F5X8:2
6.3A slow
+24V
X8:3
X8:5
X8:4+24VDC
0VAC
0VDC
24VACX11:06
X11:08
+24V
Q10
inactivefootswitch
buckycentered
U6+5V
X4:04
X4:05
X9:1
X9:2
D22
X4:03
X4:02
X7:6
X7:7
X7:5SAS5
+24V
X5:03
X5:02
X6:02
X6:06
X6:07
X6:04
X6:05
X6:03
safety
X7:2
X7:3
X7:4
X4:10
X4:12
X4:11
X11:01
X11:12
table height4on
5off
3
34
5
3
5
4
down
up
24VAC24VAC
0VAC
6.3A slowX8:6
X8:8
X6:08
X6:09
brake bucky module
2
1
X9:3
X9:4
X4:09
X4:08
transverse brakes
longitudinal brake
U7
U6
D5
D4
U7
U6
SNVY7
Q9Q7
Q9
Q3
K4
DK 1.2K2
U6
JP2B
Q11
DC/DCH17
H18
JP3
H20
from
pow
ersu
pply
see
Z1-
2
U1
Q6
D10SAH2
SAS3
D9
D13
SAS2
SCS4
SNVS9
SNVS8SNVS7
SAS22
SAS21
SNVM1K4
K1
0VAC
F2
GL1K2
SAH5
SAY1
SAY3
table control
SAY2b
SZ1
SAY2a
tost
and
see
Z1-
5
lowest range = 540mm
highest range = 810mmsafety switch ~ 200N without load!
Table up / down
SNVS7
SNVS8SNVS9
SAS21/22
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo versionTable controlA3
01
-09-
19 O
st./S
chr.
bucky DIAGNOST TS C Philips Medical Systems
(01.1) Z1-3.1k5 ON = fast
k4 Off
K5 Off = Slow
k4 ONk1/k2 ON
= Tomo StartK3 ON
k1/k2 OffTable down
+ TomoTomo Start
Table up
M
M
21
-
+
2
1
4
3
I =2A
slow
fast
1
1
1
1
1
0.5
0.5
1
1
1
1
1
Jumper
SNV H4
SNV S15
coupling barsensor
D
A
+
+5VDC
table potentiometer SNV P1
laserdiode
X5:05
X5:04
X5:06
X5:07
X5:08
500 EX5:09
X5:10
data +
data -X11:02
X11:03
polyswitchRUE700
RXE185polyswitch
X11:07
X11:10
34on
5off
slow
fast3
5
431VACX8:7 6.3A slow
31VAC
-+
1
2
1a
1
1b
1b
1
1a
DOWN
ENABLE
+24V
&
& DOWN
ENABLE
+24V
DOWN
ENABLE
+24V
table motor
X6:10
X6:11
brake
+24V+24V
table on2
1
2 down
1
2 down
1K1
ENABLE
+24V
DOWN
ENABLE
ENABLE
+5V
B
A
1
41
2
B
A
2
1
3
4
footswitch
2
1
X10:3
X10:2
X4:07
X4:06
2
1
1
2
+24V
+5V
U2
+24V
JP1
1
4
1
2 2
1
F6
4A slow
buckycentered
bucky brakerelease
SWGND
X20:09
X20:01
X20:12
X20:06
X20:08
X11:09
0VAC
24VAC
24VAC
0VAC
F5X8:2
6.3A slow
+24V
X8:3
X8:5
X8:4+24VDC
0VAC
0VDC
24VACX11:06
X11:08
+24V
Q10
inactivefootswitch
buckycentered
U6+5V
X4:04
X4:05
X9:1
X9:2
D22
X4:02
X7:6
X7:7
X7:5SAS5
+24V
X5:03
X5:02
up
down
X6:02
X6:06
X6:07
X6:04
X6:05
X6:03
safety
X7:2
X7:3
X7:4
X4:10
X4:12
X4:11
X11:01
X11:12
table height4on
5off
3
34
5
3
5
4
down
up
down
up
17.5VAC
17.5VAC
0VAC
6.3A slowX8:6
X8:8
X6:08
X6:09
brake bucky module
2
1
X9:3
X9:4
X4:09
X4:08
transverse brakes
longitudinal brake
SZ2
U6
Dk5 K5
F4
K6
PIC
-Con
trolle
r
RS485
(tom
o dr
ive)
to s
tand
see
Z1-
7(to
mo
driv
e)to
sta
nd s
ee Z
1-7
F3
K3
H19
K5F1
U6
D5
D4
U7
16
Dk3k3
U6
U6
Tomo Motor
M2
SNVY7
Q9 Q7
Q9
Q3
K4
Dk1.2
K2
U6
JP2B
Q11
DC/DCH17
H18
JP3
H20
from
pow
er s
uppl
y s
ee Z
1-2
U1
Q6
D10SAH2
SAS3
D9
D13
SAS2
SCS4
SNVS9
SNVS8SNVS7
SAS22
SAS21
SNVM1K4
K1
0VAC
F2
GL1 K2
SAH5
SAY1
SAY3
table control
SAY2b
SZ1
SAY2a
to s
tand
see
Z1-
5
lowest range = 540mmhighest range = 810mmsafety switch ~ 200N without load!Table up / down
SNVS7SNVS8SNVS9SAS21/22
1
1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
with bucky Module non sensingbucky DIAGNOST TS
standard versionA3
99-1
1-23
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-4
1) connector delivered with bucky DIAGNOST TS
1)
1)
JP4
JP5
WAX1:03RQ_XG_PR
X1:03
X1:02
X3:4
X3:3WAX1:02XG_RD_EX
&
+26V XG
+26V XG
+26V XG
+26V XG
ACTIVEX11:11
ACTIVE
bucky ready
tomo ready
tomo mode
format contact
SW BU 1
WAX11:09
WAX11:10
SW SF CFWAX11:01
WAX11:02
SW TO 1WAX11:05
WAX11:06
WAX11:04
SL XG T0WAX11:03
+5V
ACTIVE
+5V
RQ_XG_PR
XG_RD_EX
+5V
+26V XG
+26V XG
XG
XG
+26V XG
WAX1:05RG_DV
X1:04
X1:07
X1:05
X1:01
X1:02
X1:03
X1:05
X1:06
X1:04
FD R CH
FD C CH
PO 15/40V
NG 15V
REL CH
FD L CH X1:01
X1:08
X1:09
X1:10
X1:11
0V PW
screen
RF 0V CH
PO 400V
SIGN CH X1:07
SX11:11
SX11:03
SX11:04
SX11:12
SX11:05
SX11:07
SX11:06
EZX21:05
EZX21:07
EZX21:06
EZX21:11
EZX21:03
EZX21:04
EZX21:12
SX11:13
SX11:01
SX11:08
housing
EZX21:13
housing
EZX21:01
EZX21:08
WAX1:04
WAX1:07GND XG
+26V XG
WAX1:01RQ_XG_EX
24
1
2
1
X23
X28
X21
X20
+24V5
CASS_PR
4
3
X30
+15V
2
1 SYNC
+24V
RESET
X22+15V
321
X24
X25
X26
X27
+24V
X29+15V
+15V
START
X31
X32
X33
2
+24V
1 CASS_PR
+24V
+15V
X2:1
X2:5
X2:6
X2:4
3
5
SYNC
7
6FREE EXPOSURE
X2:3
X2:9
X2:2
X2:7
4
8
X3:1
X3:5
X3:6
X3:2
H16
to stand
bucky controlVA1
see Z1-5
bucky 1
H15
L5
H11
L4
H12
L7
amplimat measuring chamberPH
"bucky"release 1
VA2 sensing
tilt switch
S1
Y1
VA2 sensing
RY2
D1
T2
RY1IC1B
~20stimer
IC1A
timer
P2P1
R4
C7
W1
R2
delay new
D5
T3R13
RY2D6
DC/DC
R12
OC1R13
R11
C8
D2
D3
D4
RY1
RY2
R8
generatorE/WA
table controlSZ1
VAbucky module
Bucky control
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo versionBucky control
A3
99-1
1-23
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-4-1
1) connector delivered with bucky DIAGNOST TS
JP4
JP5
WAX1:03RQ_XG_PR
X1:03
X1:02
X3:4
X3:3WAX1:02XG_RD_EX
&
Rad on2X11:05
Rad on1
WAX11:08 X2:08
X2:10
X2:09
X2:05
X2:06
X2:04
X2:03
X2:02
X2:01
+26V XG
+26V XG
+26V XG
+26V XG
ACTIVEX11:11
ACTIVE
bucky ready
tomo ready
tomo mode
format contact
SW BU 1
WAX11:09
WAX11:10
SW SF CFWAX11:01
WAX11:02
SW TO 1WAX11:05
WAX11:06
WAX11:04
SL XG T0WAX11:03
+5V
ACTIVE
+5V
RQ_XG_PR
XG_RD_EX
+5V
+26V XG
+26V XG
XG
+26V XG
WAX1:05RG_DV
X1:04
X1:07
X1:05
X1:01
X1:02
X1:03
X1:05
X1:06
X1:04
FD R CH
FD C CH
PO 15/40V
NG 15V
REL CH
FD L CH X1:01
X1:08
X1:09
X1:10
X1:11
0V PW
Screen
RF 0V CH
PO 400V
SIGN CH X1:07
SX11:11
SX11:03
SX11:04
SX11:12
SX11:05
SX11:07
SX11:06
EZX21:05
EZX21:07
EZX21:06
EZX21:11
EZX21:03
EZX21:04
EZX21:12
SX11:13
SX11:01
SX11:08
housing
EZX21:13
housing
EZX21:01
EZX21:08
WAX1:04
WAX1:07GND XG
+26V XG
WAX1:01RQ_XG_EX
24
1
X23
X28
X21
X20
+24V5
CASS_PR
4
3
X30
+15V
2
1 SYNC
+24V
RESET
X22+15V
321
X24
X25
X26
X27
+24V
X29+15V
+15V
START
X31
X32
X33
2
+24V
1 CASS_PR
+24V
+15V
X2:1
X2:5
X2:6
X2:4
3
5
SYNC
7
6FREE EXPOSURE
X2:3
X2:9
X2:2
X2:7
4
8
X3:1
X3:5
X3:6
X3:2
from stand (PLC)
see Z1-7
cont
rolle
r
H16
to stand
bucky controlVA1
see Z1-7
bucky 1
H15
H11
L4
H12
L7
amplimat measuring chamberPH
"bucky"release 1
VA2 sensing
tilt switch
S1
Y1
VA2 sensing
RY2
D1
T2
RY1
delay old
IC1B
~20stimer
IC1A
timer
P2P1
R4
C7
W1
R2
delay new
D5
T3
timer~2s
IC2
R13
RY2D6
DC/DC
R12
OC1R13
R11
C8
D2
D3
D4
RY1
R8
generatorE/WA
table controlSZ1
VAbucky module
with bucky Module non sensing
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
and Medio CP /CP-H or SCPbucky DIAGNOST TS /Tomo
Exposure interface SZ3A3
00-0
3-01
Sch
r.
bucky DIAGNOST TS C Philips Medical Systems
(00.0) Z1-4.2
Interface
1µF40V
1µF40V
1µF40V
1N5060
1N5060
1N5060
EZAX1/2/3/4EMX2/5/12/22EZX1/2MEDIO CP-HMEDIO CPSCP
X1:1
X1:2
X1:3
X1:5
X1:7
X1:4
SZ3
SZ1
bucky TStable
100n 100n
10R
0V
-(24)V
RDGV
PREP
EXPX2:1
X2:2
X2:3
X2:5
X2:7
X2:4
EXP
PREP
RDGV
-(24)V
0V
X1:1
X1:2
X1:3
X1:5
X1:7
X1:4
7924
K4 K4
K3
K2
K1
K3
K2
K1
C2C1m100
63Vm100
63V
10R
10R
10R
10R
Generator
1
2
3
5
7
4
1
2
3
5
7
4
1
2
3
5
7
4
SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
and Medio CP /CP-H or SCPbucky DIAGNOST TS /Tomo
Exposure interface SZ3A3
00-0
3-02
Sch
r.
bucky DIAGNOST TS C Philips Medical Systems
(a/00.0) Z1-4.2
Interface
EZAX1/2/3/4EMX2/5/12/22EZX1/2MEDIO CP-HMEDIO CPSCP
X1:1
X1:2
X1:3
X1:5
X1:7
X1:4
SZ3
SZ1
bucky TStable
100n 100n
0V
-(24)V
RDGV
PREP
EXP
X2:1
X2:2
X2:3
X2:5
X2:7
X2:4
EXP
PREP
RDGV
-(24)V
0V
X1:1
X1:2
X1:3
X1:5
X1:7
X1:4
K4
K3
K2
K1K1
C3C1m100
63Vm100
63VGenerator
SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41
1
2
3
5
7
4
1
2
3
5
7
4
1
2
3
5
7
4
K3
K2
R1MP1
C5m100
63V 63V
C2
MP2
V1
V2
V8
A1
V7
C4
63V R10
H5
MP4
MP5RGDV
-24V DC
21
3Vin Vout
GND
PREP
EXPout
EXPin
MP6
MP7
MP8
V3
R2R6
H1
H3
H2 V5
R4
V4
R3R8
R7
C8
C7
V10
V9
K4R9 R5
V6C6H4
0V
-24V AC
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSstandard version
Stand controlA3
00-
02-2
3 O
st.
bucky DIAGNOST TS C Philips Medical Systems
(99.0) Z1-5
b
y
a
z x
UE control handle
* Optional
column rot.
LED_ b
bLED_
Q1/Q2
X31:9
X22:09X31:09
+24V
+24V
+24V
JP1
U5
+5VX40:12X1:12
+5V
U7
U7
U7
U7
U7
X20:09
X11:11
X11:06
X11:08
X11:01
X11:12
X11:09
SZ1 table control
X20:11
X20:06
X20:08
24VAC
0VAC
ACTIVE
X20:01
X20:12
+24V
see Z1-2/Z1-3
41
2
1
1 2
1 4
1 4
1 4
2
1
2
1
2
1
UF tube arm
stand controlUB1
UB1 stand control
see Z1-2
+24V
ACTIVEX31:05X22:05
U7
ACTIVE = Tube select
X31:09X22:09
D1driver
D1driver
D1driver
D1driver
driverD1
P1time
field
H124VAC
20VAC
1
K1
timerM4
5V DC
X1:2
X1:1
X1:5
X49:1Laser *
LIGHT ON
M3X1:6
M5X1:4
X49:2
X32:4
M6X1:3X32:3
X31:04
X31:07
X31:01+24V
X22:04
X22:07
X22:01
LA manual collimator
X31:12X22:12
a
b
clock
U6
generator
U3
U6
X31:11
X31:02
X31:10
X24:5
X24:6
X24:4
X24:1
X24:3
X24:2
U6
U3
x
U3
U6
U6a
U5
U5
U5
U5
U5
U3
+5V
+5V
+5V
+5V
+5V SID 110 cm
+24V
+24VUBS19X22:11
UBY8
UBS17
X22:02
+24V
X22:10
X37:3
X37:2
X36:7
X36:6
X36:3
X36:2
X37:4
X36:8
X36:4
S13
S14
X37:1+24V
X36:5+24V
X36:1
X38:8
X38:7
X38:6
X38:5
X38:3
X38:4
X38:2a
z
x
y
UE2
UE1
active
DC/DC +5V+24V
U8
+24V
X40:21X1:21+5V+5V
X40:24X1:24
+5V
X1:08 LIGHT ONX40:14S8
LED_ACTIVEX40:07H7
+5V
X1:07
LED_X40:03H3+5V
X1:03
LED_X40:05H5
+5V
X1:05
LED_ZX40:04H4+5V
X1:04
LED_XX40:01H1+5V
X1:01
H2LED_YX40:02X1:02
+5V
UES26
S25
UES25
X35:3
X35:1
X35:7
S22
S23
S20
S12
U4
Y5
X35:6
UES24
X35:4
UES22
X35:5
UES23
X35:2
UES20
UES21Y4
Y6
X38:1+5V
16-b
it m
icro
con
tolle
r
b
X32:5
X32:6
0V
b
b
a
column centered
tube support
column
tube support
Upgrade kit Center Laser
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DAIGNOST TStomo versionStand controlA
3 0
0-02
-23
Ost
.
bucky DIAGNOST TS C Philips Medical Systems
(99.0) Z1-5.1
0VX1:2
X1:1Laser *
b
y
a
z x
PLC
UE control handle
UE1
display
LCD
PGU:1PGU:9PGU:8PGU:3EA16:0EAO:7EAO:6EAO:5EAO:4EAO:3EAO:2EAO:1EAO:0
5VRTS
RX
lock
X
FFA
ok
light
on
tabl
e do
wn
tabl
e up
tom
o tim
e
tom
o an
gle
tom
o te
st
tom
o on
tom
o on
tomo control handle
(2x10) X34
MAX 232signal converter
X30:15X30:11X30:10
X41:03
X41:06
X41:07
X41:02X41:05X41:04
X30:01X30:09 X30:02 X30:03 X30:12 UE2X30:07 X30:08X30:06X30:05X30:04
EPLD
X41:13 LED_a
X41:11 LED_Z
X41:15 LED_X
bLED_b
LED_b
X31:9
X22:09X31:09
+24V
+24V
+24V
JP1 D1
U5
VCLX41:01
X20:09
X11:11
X11:06
X11:08
X11:01
X11:12
X11:09
SZ1 table control
X20:11
X20:06
X20:08
24VAC
0VAC
ACTIVE
X20:01
X20:12
+24V
see Z1-2/Z1-3
41
2
1
1 4
1 4
1 4
2
1
2
1
2
1
UF tube arm
stand controlUB1
UB1 stand control
see Z1-2
+24V
ACTIVEX31:05X22:05
ACTIVE = Tube select
X31:09X22:09
P1
laser (option)Centering
time
field
H124VAC
20VAC
1
K1
timer
M2
M1M4X1:5
X49:1
LIGHT ON
M3X1:6
M5X1:4
X49:2
X32:4
M6X1:3X32:3
X31:04
X31:07
X31:01+24V
X22:04
X22:07
X22:01
LA hand collimator
X31:12X22:12
clock
U6
generator
U3
U6
X31:11
X31:02
X31:10
X24:5
X24:6
X24:4
X24:1
X24:3
X24:2
U6
U3
x
U3
U6
U6a
U5
U5
U5
U5
U5
U3
+5V
+5V
+5V
+5V
+5V SID 110 cm
+24V
+24VUBS19
X22:11
UBY8
UBS17
X22:02
+24V
X22:10
X37:3
X37:2
X36:7
X36:6
X36:3
X36:2
X37:4
X36:8
X36:4
S13
S14
X37:1+24V
X36:5+24V
X36:1
X38:8
X38:7
X38:6
X38:5
X38:3
X38:4
X38:2 a
z
x
y
DC/DC +5V+24V
U8
+24V
X41:09
LIGHT ONX41:08
LED_ACTIVEX41:13
LED_bX41:12
LED_YX41:14
UES26
S25
UES25
X35:3
X35:1
X35:7
S22
S24
S23
S20
S21
S12
U4
Y5
X35:6
UES24
X35:4
UES22
X35:5
UES23
X35:2
UES20
UES21
Y6
X38:1+5V
b
X32:5
X32:6
DC
Upgrade kit Center Laser
* Optional
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DAIGNOST TSFix tube arm and tomo version
Stand controlA3
00-
03-2
9 O
st./S
chr.
bucky DIAGNOST TS C Philips Medical Systems
(00.0) Z1-5.1.1
0VX1:2
X1:1Laser *
b
y
a
z x
PLC
UE control handle
UE1
display
LCD
PGU:1PGU:9PGU:8PGU:3EA16:0EAO:7EAO:6EAO:5EAO:4EAO:3EAO:2EAO:1EAO:0
5VRTS
RX
lock
X
FFA
ok
light
on
tabl
e do
wn
tabl
e up
tom
o tim
e
tom
o an
gle
tom
o te
st
tom
o on
tom
o on
tomo control handle
(2x10) X34
MAX 232signal converter
X30:15X30:11X30:10
X41:03
X41:06
X41:07
X41:02X41:05X41:04
X30:01X30:09 X30:02 X30:03 X30:12 UE2X30:07 X30:08X30:06X30:05X30:04
EPLD
X41:13 LED_a
X41:11 LED_Z
X41:15 LED_X
bLED_b
LED_b
X36:9
X22:09X31:09
+24V
+24V
+24V
JP1 D1
U5
VCLX41:01
X20:09
X11:11
X11:06
X11:08
X11:01
X11:12
X11:09
SZ1 table control
X20:11
X20:06
X20:08
24VAC
0VAC
ACTIVE
X20:01
X20:12
+24V
see Z1-2/Z1-3
41
2
1
1 4
1 4
2
1
2
1
UF tube arm
stand controlUB1
UB1 stand control
see Z1-2
+24V
ACTIVEX31:05X22:05
ACTIVE = Tube select
X31:09X22:09
P1
laser (option)Centering
time
field
H124VAC
20VAC
1
K1
timer
M2
M1M4X1:5
X49:1
LIGHT ON
M3X1:6
M5X1:4
X49:2
X32:4
M6X1:3X32:3
X31:04
X31:07
X31:01+24V
X22:04
X22:07
X22:01
LA hand collimator
X31:12X22:12
clock
U6
generator
U3
U6
X31:11
X31:02
X31:10
X24:5
X24:6
X24:4
X24:1
X24:3
X24:2
U6
U3
U3
U6
U6a
U5
U5
U5
U5
U5
U3
+5V
+5V
+5V
+5V
+5V SID 110 cm
+24V
+24VUBS19
X22:11
UBY8
UBS17
X22:02
+24V
X22:10
X37:3
X37:2
X36:7
X36:6
X36:3
X36:2
X37:4
X36:8
X36:4
S14
X37:1+24V
X36:5+24V
X36:1
X38:8
X38:7
X38:6
X38:5
X38:3
X38:4
X38:2 a
z
y
DC/DC +5V+24V
U8
+24V
X41:09
LIGHT ONX41:08
LED_ACTIVEX41:13
LED_bX41:12
LED_YX41:14
UES26
S25
UES25
X35:3
X35:1
X35:7
S22
S24
S23
S20
S21
S12
U4
X35:6
UES24
X35:4
UES22
X35:5
UES23
X35:2
UES20
UES21
Y6
X38:1+5V
b
X32:5
X32:6
DC
Upgrade kit Center Laser
* Optional
UE3 Brake amplifier
R2V5
V2
V4R1
V1
X36A:9
X36A:7
X36A:6
X36A:3
X36A:2
X36A:8
X36A:4
X36A:5
X36A:1
X36:9
X36:7
X36:6
X36:3
X36:2
X36:8
X36:4
X36:5
X36:1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
with bucky unit non sensingbucky DIAGNSOT VE
A3
99-1
1-22
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z1-6.1
X1:02
X1:04
X1:03
X1:01
X1:06
X1:08
X1:07
X1:11
X1:10
X1:09
X1:05
X4:12
X4:04
X4:03
X4:11
X4:06
X4:08
X4:07
X4:01
housing
X4:13
X4:05
REL CH
NG 15V
PO 400V
screen
0V PW
PO 15/40V
FD L CH
FD C CH
FD R CH
SIGN CH
RF 0V CH
EZX22:12
EZX22:04
EZX22:03
EZX22:11
EZX22:06
EZX22:08
EZX22:07
EZX22:01
housing
EZX22:13
EZX22:05
VA2 sensing
VA2 sensing
tilt switchdelay new delay old
D62
1
2
1
78
6
54
3
54
3
2
1
24
1
W1 32
+15V
X31
X29
Y1
S1
VA1bucky control
X28+24V
X30
X21
X22
RY2CASS_PR
X23
X20
X33
X32
+24V
RY2CASS_PR
C7
1
P2
R4
P1
+15V
R2
RY2FREE EXPOSURE
X2:9
X2:7
RESET
X2:3
X2:2
+24V
D1
T2
IC1A
timer
IC1B
timer~20s
X24
X25
+15V
R8
START
~2s
&
D5
T3
X26
X27
+24V
R13
R13
R12
C8
R11
D4
D3
IC2
D2
X2:4
DC/DC
OC1
SYNC
SYNC
timer
+15V
RY1
+15V
+24VX2:1
X2:5
X2:6
RY1
boxconnection
"Inalfa"
not used
RQ_XG_EX
XG_RD_EX
GND XG
+26V XG
"BWS"release 2
amplimat measuring chamberPH
generatorE/WA
wallVP
WAX2:04
WAX2:07
WAX2:02
WAX2:01
X1:1
X1:5
X1:6
X1:4
X1:3
X1:2
X1:9
X1:7
VAbucky unit
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo versionStand controlA3
01
-09-
19 O
st./S
chr.
bucky DIAGNOST TS C Philips Medical Systems
(a/01.0) Z1-7.1
Nikkai/SS-12 SDP2service switch
K2 finder 4061.9024
JP1Jumper
(1W)R3
(DO-41)PS 102RD1
K1
1.85Apolyswitch
F1
100V1.5UC1
100V1.5UC2
(1W)1RR2
(1W)1RR1
100V1.5UC3
100V1.5UC4
GL1
(1W)R4
beta-locksensorS19:
column axisbrake YY8:
lock YsensorS17:
angle sensorB1:
safety end-switchS20:
motortomoM2:
limit switchS16:
motorcouplingM3:
MM
shielding
U finder 1
U finder 2
X31:04
X31:01
X31:07
GNDX31:12
X31:06
24V
locks off
spare
device active
brake Y
X31:08
X31:02
X31:03
X31:05
sensor Y
sensor beta
X31:10
X31:11
X31:09
release brakes
X lock
FFA ok
light on
table down
table up
tomo time
tomo angle
tomo test
tomo on
RS232
RX
RTS
5V
X30:11
X30:15
X30:10
X30:09
X30:03
X30:12
X30:02
X30:04
X30:06
X30:05
X30:08
X30:07
EA16:8
EA16:3
EA16:1
EA16:0
EA16:9
EA0:9
EA0:8
EA0:4
EA0:5
EA0:7
EA0:6
EA0:2
EA0:3
EA0:1
EA0:0
PGU:9
PGU:8
PGU:3
brake Y
release clutch
interrupt2
interrupt1
24VDC
GND
GND
24VDC
Rad on2
Rad on1
data +
data -
EA32:8
EA32:9
XS:11
EA32:0
XS:12
EA32:1
XPI:+
XPI:-
XPI:INB1
XPI:INB2
EA32:5
EA32:4
EA32:3
EA32:2
-+
~~
X22:07
X22:04
X22:01
X22:12
X22:06
X22:08
X22:02
X22:03
X22:05
X22:11
X22:10
X22:09
X21:13
X21:14
X21:01
X21:12
X21:11
X21:10
X21:09
X21:08
X21:02
X21:15
X21:07
X21:03
X21:04
X21:05
X21:06
X23:12X23:11X23:10X23:09X23:07X23:08X23:06X23:05X23:04X23:03X23:02
X20:11device active
X11:11
X20:10Umot 2-
X11:10
X20:07Umt 2+
X11:07
X20:12GND
X11:12
X20:0124VDC
X11:01
X20:08U finder 2
X11:08
X20:06U finder 1
X11:06
X20:05Rad on2
X11:05
X20:04Rad on1
X11:04
X20:03RS485 data +
X11:03
X20:02
X20:09X11:01
RS485 data -
shieldingX11:02
X23:01
W31
W30
UEcontrol handle
W29UE
control handle
PLCtomo control
W27
W27
UB1stand control
W28
W28
S1
rectifier
24V
GND
W20
W23W22W21W24
W25W26
2
1
4
1magnetic sensor
1
2
SZ1table control
1
1R 1R
PGU:1 X30:01shielding
sensor
UB
X24A:1 X24A:2 X24A:5 X24A:6X24A:4X24A:3
X24:1 X24:2 X24:5 X24:6X24:4X24:3
X24:1 X24:2 X24:5 X24:6X24:4X24:3
V1R1
V4Brakeamplifier
UB2
grey
blue
pink
4 4
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS (99.0) Z0--2 1999 Philips Medical Systems
ALL RIGHTS RESERVEDbdts_z02_990
Drawings
Survey of components tomo version Z2--1
Wiring diagram table standard version Z2--2.1
Wiring diagram table tomo version Z2--2.2
Wiring diagram stand standard version Z2--3.1Wiring diagram stand tomo version Z2--3.2
Wiring diagram stand standard version Z2--3.3
Wiring diagram control handle tomo version Z2--4.1
Wiring diagram control handle standard version Z2--4.2
Eathing diagram Z2--5
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Survey of componentsA4
00-
11-1
0 O
st./S
chr.
bucky DIAGNOST TS C Philips Medical Systems
(00.0) Z2-1
SB
SB
UB
LA
XA
SNV SC SZ1
UE
UF
UB1
UE
VA
SN
SASA PH
SASBSCSNSNVSZ1SZ2SZ3UBUB1UFPH
------------
Transverse carriage brakesPivot towerFootswitchPower supplyMotor assy & Motor control switchesTable controlTomo speed controlTomo Medio CP interfaceColumn basementColumn control & Tomo controlColumn and transversal tube support armAmplimat chamber
VA - Bucky carriage brakeUELAXA
---
Control handleCollimatorX-ray tube
SZ3 SZ2
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Wiring diagram tableA3
99-1
1-24
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-2.1
1
8 94 53
6 72
14
2
ywgnbnwt
blpkgy
BN
WH
GNBNWH
BUPKGY
PK
GY
blk
rd
WH
BN
GN
gn
YE
yw
wt
bn
10A slow
10A slow
4A slow
6.3A slow
SNVP1:2SNVP1:1SBH 4SBH 4SBS 15SBS 15SNVS15:1
1WAX2
F1SNT2:17.5VAC
SN
1X8:
3
SN
1X8:
2
230V
AC
SN
1X8:
80V
AC
230V
AC
115V
AC
X16
:4
230V
AC
SN
1X8:
6
+26V XG
H15 = ACTIVEH12 = XG_RD_EXH11 = RQ_XG_PR
+24V 24VAC
S3SZ1X9:2
SZ1X9:1
2
1
412
SAY1
A
1a
1b1
1
B2
121212
4B3
1b11a
Y11
2SZ1X6:11
SZ1X6:10
SNVY7:2SNVY7:1
screenSCS4:1SCS4:2
SAS2:1SAS2:2
screenscreen
SAS5:4
SAS5:3
screen
SAY2a:2SAY2a:1
X5:01X5:02
X5:04X5:05
X5:07X5:08
X5:06
X5:03
X5:10X5:11
X5:13X5:14
X5:12
X5:15
X5:09
X4:14X4:13
X4:15
X4:02X4:01
X4:08X4:07X4:06X4:05X4:04X4:03
X4:10
X4:12X4:11
X4:09
X7:1
X7:3X7:4X7:5X7:6X7:7
X7:2
X7:9X7:8
X6:12X6:11X6:10X6:09X6:08X6:07X6:06X6:05X6:04X6:03X6:02X6:01
X6
X7
X5X4
X9
X10
14
98
1211107
2 3
5 6
151411 12
8 9710
13
65
314
2
151411 12
8 9710
13
65
314
2
4 521
63
7 98
421 3
421 3
H17 H18
H16
H15
H12
H11
JP6
JP3
JP1
JP5
230V
AC
24VAC
F2
F6
JP2B
8
7
6
5
4
3
2
from generator
Z1:
L1
X16
:
+24
V1N
G1: G
1:N
G1:
L1
T2
: T2
:0V
T2
:230
V
T1
:0V
T1
:230
V
X16
Z1:
N
T1:
N
4 3 2.1 1.1
4 3 2 1
L
F11
F12
N.1 L1.1
LOAD
N L1LINE
Z1
L1
0V5
0VA
C
0VA
C
115V
AC
0VA
C
1
T2
X16
:3
X16
:3
X16
:4
SN
1X8:
5
G1
SN
1X8:
4
X16
:3
X16
:4
power supply
T1
SNT1:24VAC
SNT2:0VAC
SNT1:0VAC
SNT2:37VAC
SNG1:+24V
SNG1:0V
S4
UB1X20
1WAX1
SZ1X4:05SZ1X4:04
SZ1X4:10
SZ1X4:12SZ1X4:11
SZ1X4:02SZ1X4:03
SAS21
SAS2/H2
SZ1X7:7
SAS5/H5
SZ1X7:6
1
10
09
08
06
05
04
03
07
01
02
10
09
08
07
06
05
04
03
02
01
X11
JP4
JP2A
F5
X8
X3
X1
SAS22:1b
SAS22:1aSAS22:1
SAS21:1SAS21:1b
SAS21:1a
SNVM1:+SNVM1:-
SNVS8:1SNVS8:2SNVS9:1SNVS9:4
SNVS7:1SNVS7:2
SAY3:1SAY3:2
SZ1X7:3SZ1X7:2
SAS22
SAY3SAY2aSAY2b
SAH5:B
SAS5:3 SAH2:ASAH2:B
VAX2
X1
PHVA
bucky module
VAX2
SZ1X3
generator
SN
-
+SZ1X6:08
SZ1X6:09M1
S8
S7
S9
SZ1X6:04
SZ1X6:05
SZ1X6:02
SZ1X6:03
SZ1X6:07
SZ1X6:06
1
4
2
1
4
2
2
1
4
SNVmotor
SA
Identitication of LED´s see chapter 4 "Trouble shooting"
foot switch
table control
SCSZ1X5:03
SZ1X5:02
SZ1
X2
SZ1X7:5
SZ1X7:4
transverse carriage
SZ
1X10
:3
SZ
1X10
:2
SZ
1X4:
09
SZ
1X4:
08
SZ
1X4:
07
SZ
1X4:
06
H20
SA
Y2a
:1
SA
Y2a
:2
SA
S3
:1
SA
S3
:4
SA
Y1
:2
SA
Y1
:1
Z1:
N.1
Z1:
L1.1
X16
:4
X16
:3
X16
:2.1
X16
:
X16
:1.1
SZ
1X9:
3
SZ
1 X9:
4
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Wiring diagram tableA3
99-1
1-25
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-2.2
SB
1
4
2
2
1
3
1
8 94 53
6 72
14
2
ywgnbnwt
blpkgy
bn
wt
gnbnwt
blpkgy
pk
gy
blk
rd
wt
wt
bn
bn
gn
gn
gn
yw
yw
gn
yw
wt
bn
wt
bl
10A slow
10A slow
4A slow
6.3A slow
H19
H2
H1 H9
H8
H5
H6H4
H7H3 H10 H14
H13
P1
H4
S15
+
-
SZ1X5:06
SZ1X5:07
SZ1X5:05
SZ1X5:04
SZ1X5:10
SZ1X5:09
SZ1X5:08
SNVP1:2SNVP1:1SBH 4SBH 4SBS 15SBS 15SNVS15:1
1WAX2
F1SNT2:17.5VAC
SN
1X8:
3
SN
1X8:
2
230V
AC
SN
1X8:
80V
AC
230V
AC
115V
AC
X16
:4
230V
AC
SN
1X8:
6
+26V XG
H15 = ACTIVEH12 = XG_RD_EXH11 = RQ_XG_PR
+24V 24VAC
S3SZ1X9:2
SZ1X9:1
2
1
412
SAY1
A
1a
1b1
1
B2
121212
4B3
1b11a
Y11
2SZ1X6:11
SZ1X6:10
SNVY7:2SNVY7:1
screenSCS4:1SCS4:2
SAS2:1SAS2:2
screenscreen
SAS5:4
SAS5:3
screen
SAY2a:2SAY2a:1
X5:01X5:02
X5:04X5:05
X5:07X5:08
X5:06
X5:03
X5:10X5:11
X5:13X5:14
X5:12
X5:15
X5:09
X4:14X4:13
X4:15
X4:02X4:01
X4:08X4:07X4:06X4:05X4:04X4:03
X4:10
X4:12X4:11
X4:09
X7:1
X7:3X7:4X7:5X7:6X7:7
X7:2
X7:9X7:8
X6:12X6:11X6:10X6:09X6:08X6:07X6:06X6:05X6:04X6:03X6:02X6:01
X6
X7
X5X4
X9
X10
14
98
1211107
2 3
5 6
151411 12
8 9710
13
65
314
2
151411 12
8 9710
13
65
314
2
4 521
63
7 98
421 3
421 3
H17 H18
H16
H15
H12
H11
JP6
JP3
JP1
JP5
230V
AC
24VAC
F2
F6
JP2B
8
7
6
5
4
3
2
from generator
Z1:
L1
X16
:
+24
V1N
G1:
G1:
N
G1:
L1
T2
:
T2:
0V
T2
:230
V
T1
:0V
T1
:230
V
X16
Z1:
N
T1:
NZ1:L1.1Z1:N.1
4 3 2.1 1.1
4 3 2 1
L
F11
F12
N.1 L1.1
LOAD
N L1LINE
Z1
L1
0V5
0VA
C
0VA
C
115V
AC
0VA
C
1
T2
X16
:3
X16
:3
X16
:4
SN
1X8:
5
G1
SN
1X8:
4
X16
:3
X16
:4
power supply
T1
SNT1:24VAC
SNT2:0VAC
SNT1:0VAC
SNT2:37VAC
SNG1:+24V
SNG1:0V
S4
UB1X20
1WAX1
SZ1X4:05SZ1X4:04
SZ1X4:10
SZ1X4:12SZ1X4:11
SZ1X4:02SZ1X4:03
SAS21
SAS2/H2
SZ1X7:7
SAS5/H5
SZ1X7:6
1
10
09
08
06
05
04
03
07
01
02
10
09
08
07
06
05
04
03
02
01
X11
JP4
H20
JP2A
F5
X8
X3
X1
SAS22:1b
SAS22:1aSAS22:1
SAS21:1SAS21:1b
SAS21:1a
SNVM1:+SNVM1:-
SNVS8:1SNVS8:2SNVS9:1SNVS9:4
SNVS7:1SNVS7:2
SAY3:1SAY3:2
SZ1X7:3SZ1X7:2
SAS22
SAY3SAY2aSAY2b
SAH5:B
SAS5:3 SAH2:ASAH2:B
VAX2
X1
PHVA
bucky module
VAX2
SZ1X3
generator
SN
-
+SZ1X6:08
SZ1X6:09 M1
S8
S7
S9
SZ1X6:04
SZ1X6:05
SZ1X6:02
SZ1X6:03
SZ1X6:07
SZ1X6:06
1
4
2
1
4
2
2
1
4
SNVmotor
SA
Identitication of LED´s see chapter 4 "Trouble shooting"
foot switch
table control
SCSZ1X5:03
SZ1X5:02
SZ1
X2
SZ1X7:5
SZ1X7:4
transverse carriage
SZ
1X4:
07
SZ
1X4:
06
SZ
1X10
:3
SZ
1X10
:2
SZ
1X4:
09
SZ
1X4:
08
SZ
1X9:
3
SZ
1X9:
4
SA
Y2a
:1
SA
Y2a
:2
SA
S3:
1
SA
S3:
4
SA
Y1
:2
SA
Y1
:1
X16
:3
X16
:4
X16
:1.1
X16
:2.1
X16
:
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Wiring diagram standA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-3.1
WHBN
wtbn
GNYE
wtbn
bnwt
UB
S19
:2
UB
S19
:1
UB
Y8:
2
UB
Y8:
1
UB
S17
:1
UB
S17
:4
UB1X24:2UB1X24:1
UB
1X24
:4
UB
1X24
:3
UB1X24:5UB1X24:6
X38:4X38:3
UE2X37:4UE2X37:3
X38
:6
X38
:5
X38:8X38:7
S14:4
S14:1
Y6:2
Y6:1
S12:4
S12:1
Y4:2
Y4:1
UE
2X37
:1
UE
2X37
:2
X38
:2
X38
:1
98
1211107
2 3
5 6
14
98
1211107
2 3
5 6
14
7 98
4 521
63
1 4 2 1 4 2
1 4 2
1 4 2
1 4 2
8
7
6
5
4
3
2
1
21
21
21
21
see Z2-6
see Z2-4
UEcontrol handle
UFtube arm
JP1
X20
y b
y
a
a
z
xx
UB UBS17
UBY8
S19
S12
S13
Y6
S14
UB1stand control
UE2X31
X22
SZ1X11
X24
UE2X36
X38
Y5
Y4
UFtube arm
center position
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Wiring diagram standA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-3.2
WHBN
wtbn
GNYE
wtbn
bnwt
UB
S20
:2
INB
1
EA
16:8
XP
I:-E
A32
:5
EA
32:3
XS
:12
EA
32:4
INB
2E
A16
:3
EA
32:1
XP
I:+
EA
32:0
EA
32:2
XS
:11
EA
16:1
UB
M3:
a
UB
S16
:2
UB
B1
UB
B1
UB
B1
UB
S20
:1U
BM
2:b
UB
M2:
aU
BS
16:4
UB
M3:
bU
BS
16:1
UB
S19
:2
UB
S19
:1
UB
Y8:
2
UB
Y8:
1
UB
S17
:1
UB
S17
:4
UB1X24:2UB1X24:1
UB
1X24
:4
UB
1X24
:3
UB1X24:5UB1X24:6
X38:4X38:3
UE2X37:4UE2X37:3
X38
:6
X38
:5
X38:8X38:7
S14:4
S14:1
Y6:2
Y6:1
S12:4
S12:1
Y4:2
Y4:1
UE
2X37
:1
UE
2X37
:2
X38
:2
X38
:1
98
1211107
2 3
5 6
14
98
1211107
2 3
5 6
14
7 98
4 521
63
1 4 2 1 4 2
1 4 2
1 4 2
1 4 2
14
15
13
11
12
52
63
9
8
1 4 7 10
14
15
13
9
8
12
11
107
2
3
5
6
1 4
8
7
6
5
4
3
2
1
21
21
21
21
see Z2-6
X21
X23
see Z2-4
UEcontrol handle
UFtube arm
JP1
X20
y b
y
a
a
z
xx
UB UBS17
UBY8
S19
S12
S13
Y6
S14
UB1stand control
UE2X31
X22
SZ1X11
X24
UE2X36
X38
Y5
Y4
UFtube arm
center position
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Wiring diagram standA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-3.3
UB S20 UB S16UB B1UB M3
M3
UB M2
XPI
EA32
X5
EA16
EA16
EPROM
PLC
PGUD-SUB 9P.
M2
RS485
241 241
4
+
89
5 13 2
- INB
1
INB
2
1
2
3
45
6
7
0
9
8
11 12 9
7
8
5
6
4
2
3
1
0
9
8
7
5
6
4
3
2
1
0
7 6CLOSED
OPEN
Interface Board
Jumpers (bottom side)
UB1X23:9UB1X23:7
UB1X23:3UB1X23:5UB1X23:4
UB
1X23
:10
UB
1X23
:11
UB
1X23
:12
UB
1X23
:01
UB
1X23
:02
a b+ -
a b+
UB
1X23
:06
-
UB
1X23
:08
UB1X21:02
UB1X21:15
UB1X21:07
UB1X21:08
UB1X21:05
UB1X21:06
UB
1X21
:3
UB
1X21
:4
UEX30:15
UB1X21:11
UB1X21:12
UB1X21:13
UEX30:04
UEX30:05
UEX30:06
UEX30:07
UEX30:08
UEX30:09
UEX30:10
UEX30:11
UB
X30
:01
UB
X30
:03
UB
X30
:02
UB
X30
:12
wtbn
GYPKBU
gn yw
rd gn RD
wt
bnG
N
Limit position
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TStomo version
Wiring diagram control handleA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-4.1
S24
control handle LED's
UE1
UE2control handle basic
S23S22
S21 S20
S26 S25
X31
X32
X35
X41
X40
X1
S8
H1
H2
H6
H3
H4
H5
UB1X22
UE1X1
UE2X40
X36
X30
LAX1
UFX38
X37
24 124 1
24 1
24 124 1
24 1 24 1
35 4
9 8
12
7 6
3
1
2
4
UE
2X35
:6
S23
:4S
26:4
S24
:4S
22:4
UE
2X35
:5
S23
:4S
21:4
UE
2X35
:4
S22
:4S
20:4
UE
2X35
:3
S21
:4U
E2X
35:1
UE
2X35
:2
S24
:4S
25:4
S25
:1
S26
:4
UE
2X35
:7
1
1
S23
:1
S21
:1
S22
:1
S20
:4
S20
:1S
25:1
S24
:1
1
S26
:1
1
UFS13:4
UFS13:1
UFY5:2
UFY5:1
LAX1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSStandard version
Wiring diagram control handleA3
99-1
1-26
Ost
.
bucky DIAGNOST TSC Philips Medical Systems
(99.0) Z2-4.2
S24
control handle LED's
UE1
UE2
UE1 x 1
control handle basic
S23S22
S21 S20
S26 S25
X31
X32
X35
X41
X40
X1
S8
H1
H2
H6
H3
H4
H5
UB1X22
UE1X1
UE2X40
X36
Tomo control handle
Display
X34
X30
LAX1
UFX38
X37
24 124 1
24 1
24 124 1
24 1 24 1
35 4
9 8
12
7 6
3
1
2
4
UE
2X35
:6
S23
:4S
26:4
S24
:4S
22:4
UE
2X35
:5
S23
:4S
21:4
UE
2X35
:4
S22
:4S
20:4
UE
2X35
:3
S21
:4U
E2X
35:1
UE
2X35
:2
S24
:4S
25:4
S25
:1
S26
:4
UE
2X35
:7
1
1
S23
:1
S21
:1
S22
:1
S20
:4
S20
:1S
25:1
S24
:1
1
S26
:1
1
UFS13:4
UFS13:1
UFY5:2
UFY5:1
LAX1
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty
Philips
bucky DIAGNOST TSEarthing diagram
A3
00-0
3-28
Ost
./Sch
r.
bucky DIAGNOST TSC Philips Medical Systems
(a/00.0) Z2-5
SA
T2
SN
G1T1
Z1
X16
M1SNV
SA
UE
LA
XA
UB
UF
U
to generator8.25 mm2
6 mm2
4 mm2
4 mm2
6 mm2
2.5 mm2
2.5 mm2
6 mm2
22.5 mm
22.5 mm
Philips
Phili
ps M
edic
al S
yste
ms:
D
EC
LA
SS
IFIE
D C
usto
mer
Ser
vice
Inte
llect
ual P
rope
rty