Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity measurement Choice of compliant turbidity sensors — EPA Method 180.1 — ISO Method 7027 Flow-through turbidity design — Fast response to process turbidity changes due to minimal sample residence time Automatic sensor cleaning — Reduces maintenance requirements — Optimizes performance sensitivity Secondary standards for calibration verification — Accurate, repeatable, simple, safe — Realize significant cost savings by reducing the use of primary standards Reliable and accurate measurement — Ultralow back scatter for accurate measurement of low turbidity values — Automatic bubble rejection to compensate for erroneous readings due to degassing
Transcript
Page 1: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

Data sheet DS/4690–EN Rev. A

4690 SeriesTurbidity systems

Accurate, reliable turbidity measurement

Choice of compliant turbidity sensors— EPA Method 180.1— ISO Method 7027

Flow-through turbidity design — Fast response to process turbidity changes due to minimal

sample residence time

Automatic sensor cleaning — Reduces maintenance requirements— Optimizes performance sensitivity

Secondary standards for calibration verification— Accurate, repeatable, simple, safe— Realize significant cost savings by reducing the use of

primary standards

Reliable and accurate measurement — Ultralow back scatter for accurate measurement of low

turbidity values— Automatic bubble rejection to compensate for erroneous

readings due to degassing

Page 2: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

2 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

ABB's 4690 range of turbidity systems

The measurement of turbidity is a critical measurement in drinking water as it is used to indicate water quality and filtration effectiveness (for example, whether disease-causing organisms are present).

ABB's 4690 Range of turbidity systems have been designed specifically for the measurement of turbidity throughout the potable water treatment process.

Whichever of the two measurement ranges you choose, low range (0 to 40 NTU) or higher range (0 to 400 NTU), you can benefit from reliable, accurate turbidity measurement with automatic optical cleaning and dry standard calibration verification.

Applications

Typical applications for the 4690 range of flow-through turbidity systems in potable water treatment include:— Raw water / source monitoring

Provides an indication of the solids loading entering the treatment plant to enable adjustment of the coagulant dose.

— Monitoring clarified waterClarifier supernatant turbidity measurement monitors clarifier efficiency to provide early warning of floc blanket break-up or incorrect coagulant dosing.

— Filter backwashing controlMonitors the filter backwash curve to minimize clean water usage.

— Monitoring of filtered waterEnsures correct operation of the filters and guards against turbidity breakthrough.

— Monitoring of final drinking water qualityProvides a final quality check after disinfection before distribution.

For higher turbidity applications or for turbidity measurement in open channels and tanks please refer to ABB's 4670 range of turbidity systems (data sheet DS/4670–EN).

Fig. 1: Typical turbidity monitoring application

Raw water20 to 200 FTU

Chemical dosing

Mixer

Retention tank

Clarifier

Raw waterSludge

Clarified10 to 200 FTU

Rapid gravity filter

Filter backwash water5 to 500 FTU Disinfectant applied

Contact tank

Filtered water0 to 5 FTU

Final water0 to 5 FTU

To distribution

Page 3: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 3

4690 turbidity system overview

Auto-clean system— Programmable automatic cleaning— Reduces operator maintenance— Ensures sensor accuracy and reliability

Sample Flow Cell— Flow-through nephelometric design— Minimal sample residence time provides fast

response to turbidity changes for improved process control

Emitter & Receiver Assemblies— EPA 180.1 or ISO 7027 compliant

Integral Sample Feed and Drain Valves— Simple installation— Quick and easy calibration

Dry Secondary Standards— Simplifies routine instrument verification— Eliminates chemical standard preparation

errors— Reduces cost-of-ownership

Wall- / Pipe- or panel-mount analyzers available

— Ultra-stable electronics— Flexible installation — In-built bubble rejection software— Easy-to-use

Page 4: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

4 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

Reliable measurement

ABB's 4690 range of turbidity systems have been designed for reliability, ease-of-use and maintenance simplicity. We offer a choice of US EPA* 180.1 and ISO** 7027 compliant sensing systems.

Measurement techniqueThe 4690 range of turbidity sensors utilize a flow-through nephelometric design. Water flows continually through the sensor body. Light directed by the emitter assembly passes through the sample where it is scattered by any suspended particles present in the sample. This scattered light is detected by the sensor receiver assembly that is placed at a 90° angle to the incident beam.

The ultralow back scatter allows for very accurate readings as only light scattering due to turbidity is measured. This is especially important when measuring low turbidity values.Due to the short sample residence time in the flow cell the system is able to respond much faster than many competing systems.

* United States Environmental Protection Agency** International Organization for Standardization

Automatic cell cleaning

The 4690 range of turbidity sensors feature an auto-clean system. This feature consists of a mechanical wiper assembly that physically wipes clean the optical cell at user-programmable intervals from every 15 minutes up to every 24 hours.

The highly efficient automatic cleaning process overcomes the problem of optical fouling and ensures that performance can be maintained for long periods (up to 6 months) without the need for manual intervention.

Automatic bubble rejection

Turbidity readings can be affected by short-term spikes, usually due to the effect of the sample degassing.The 4690 analyzer has an automatic bubble rejection feature that, when enabled, applies a digital filter to the measured results removing any spurious high reading.

Optional external debubblerAny bubbles present in the sample give false turbidity readings. For applications where bubbles are likely, it is recommended that the optional external debubbler assembly is used to supply a constant head of debubbled sample to the sensor.

Fig. 2: Nephelometric turbidity monitor

Detector

Lens

Lens

Light Source

Fig. 3: Auto-clean wiper unit

Page 5: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 5

Simple to calibrate

A key feature of the 4690 turbidity systems is the dry secondary calibration standard that simplifies routine instrument verification and removes the need for the use of chemical standards.

Both low- and high-range secondary standards are available so that the analyzer calibration can be verified at a level that is suited to the application. Each secondary standard is supplied factory-certified against a primary formazine standard.

The benefits of using ABB's Secondary Standards include:— Minimum analyzer downtime

Simple and fast procedure to verify analyzer performance.— Low cost of ownership

Reduce usage of consumable chemical standards and the time taken to prepare such standards.

— Minimise employee's exposure to Formazine Formazin is highly toxic and a suspected carcinogen.

— Repeatable and reliableRemoves any chemical standard preparation errors.

— Simple to useThe secondary standard is simply inserted into the sensor assembly optical light path allowing the pre-calibrated opto-mechanical filter to divert a fixed quantity of light to the detector that corresponds to the calibrated turbidity value. Rotation of the dry standard blocks the light path enabling a zero turbidity calibration to be made.

Fig. 4: Dry secondary standard

Fig. 5: Dry standard calibration

Page 6: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

6 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

Simple to maintain

The 4690 turbidity sensing systems are designed to be as maintenance-free as possible. The inherent product design and auto-clean feature minimize the amount of maintenance required to external cleaning of sample lines and periodic replacement of the wiper blade and light source. The sensor features ABB's patented light replacement system, a unique feature that enables bulb replacement in the field in seconds. Each bulb is supplied fully protected in a patented assembly with integral light guide so you never come into direct contact with the bulb. This not only protects the bulb, ensuring maximum lamp life, but also makes replacement very simple and fast as it just clicks into place.

1. Calibration Verification A procedure used to check whether or not the calibration of the analyzer is within certain limits.2. Secondary Standards Standards that the manufacturer (or an independent testing organization) has certified provide analyzer calibration results equivalent (within certain limits) to the results obtained when the instrument is calibrated with a primary standard.3. Calibration A procedure which checks or adjusts an analyzer’s accuracy by comparison with a defined standard or reference.4. Primary Standards Turbidity standards that are traceable and equivalent to the reference turbidity standard, within statistical errors. Formazine is the most commonly acceptable form of primary standard. Primary standards are used to calibrate a turbidity meter directly or to calibrate a secondary standard.

Fig. 6: Replacing the light source

TaskRecommended

frequency

Visual checks – sample flow, leaks

Calibration verification1 with secondary standard2

- as per regulatory guidelines

Calibration3 with primary standard4 –

as per regulatory guidelines

Replace wiper blade

Replace Incandescent Light Source (EPA 180.1)

Replace LED Light Source (ISO 7027)

Weekly

Monthly

Quarterly

Annually

Annually

Every 5 years

Page 7: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 7

4690 sensor specification

Range Low range 0 to 40 NTUHigh range 0 to 400 NTU

Measurement principle 90 ° scattered light measurement. Compliant to EPA 180.1 or ISO7027

Maximum linearityTypically <1.0 %

Accuracy1, 2

Low range version ±2 % of readingHigh range version ±5 % of reading or 0.3 NTU

Repeatability3 0 to 200 NTU: <1 % 200 to 400 NTU: 2 %

Limit of Detection4

Low range version: 0.003 NTUHigh range version: 0.3 NTU

Response timeT90 < 1 min at 1 l/min-1 (0.26 gall [US]/min-1)

Flow rate0.5 to 1.5 l/min (0.13 to 0.39 gall [US]/min)

Integral wiper cleaning systemProgrammable operational frequency every 0.25 hour, 0.5 hour, 0.75 hour or multiples of 1 hour up to 24 hours

Sample operating temperature 0 to 50 °C (32 to 122 °F)

Sample pressure Up to 3 bar (43.5 psi)

Ambient operating temperature 0 to 50 °C (32 to 122 °F)

Ambient operating humidity Up to 95 % RH

Wetted parts – materials usedCell body unit— Black POM (Polyoxymethylene) Copolymer— Spectrosil 2000 fused silica— Nitrile (O-ring)— Epoxy preform (cured): Uni-forms 5034-00— Polyamide 6— Nickel plated brass— Teflon

Wiper unit— Black Polycarbonate, 10% Glass Fibre filled – Lexan 500R— Stainless Steel (SS 316 S13/S11) w/ Chemical Black – MIL-

C13924 class 4— Silicone grease (WRC Approved): Unisilkon L 250 L— 2-part Epoxy Adhesive (cured): DELO AD894— EPDM (ethylene propylene diene Monomer) black

1Maximum measured error across full measurement range (typical, limited by uncertainty in Formazine standards).2Tested in accordance with IEC 61298 Parts 1-4: Edition 2.0 2008-10.3Tested in accordance with MCERTS: Performance Standards and Test Procedures for Continuous Water Monitoring Equipment. Version 3.1: Environment Agency 2010.4Tested in accordance with BS ISO 15839: 2003.

Page 8: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

8 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

4690 analyzer specification

GeneralMeasured value 5-digit x 7-segment backlit LCD Information 16-character, single line, dot matrix, backlit LCD

Units of measurement All models: NTU and FNUmg/l and ppm for high range models

Accuracy ±0.2 % of reading, ±1 digit

Linearity ±0.1 % FSD Auto-clean timing (7998011, 7998012)Programmable 15 min, 30 min, 45 min or 1 hour up to24 hours in 1 hour increments

Environmental Data Operating temperature limits –20...55 °C (–4...131 °F) Storage temperature limits –25...55 °C (–13...131 °F) Operating humidity limits Up to 95 % RH non-condensing

Power Supply Voltage requirements 100...130 V, 200...260 V, 50/60 Hz Power consumption < 6 VA AC Error due to power supply variation Less than 0.1 % for +6 % –20 % variation from nominal supply

Insulation Mains to earth (line to ground) 2 kV RMS

Relay Outputs and Set Points No. of relays Two Relay contacts Single pole changeover Rating 250 V AC 250 V DC max. 3 A AC, 3 A DC max. Loading (non-inductive) 750 VA 30 W max. (inductive) 750 VA 3 W max. Insulation 2 kV RMS contacts to earth (ground) No. of set points Two Set point adjustment Programmable

Set point hysteresis ±1 % fixed Local set point annunciation Red LED

Page 9: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 9

Retransmission No. of retransmission signals One fully isolated programmable 0...10 mA, 0...20 mA or 4...20 mA Optional second current output Accuracy ±0.25 % FSD ±0.5 % reading Resolution 0.1 % at 10 A, 0.05 % at 20 mA Max. load resistance 750 (20 A max.)

Mechanical Data Model 4690 Wall- / Pipe-mount transmitter

Model 4695 Panel-mount transmitter

Wall-mounting

Protection IP66 / NEMA4X

Dimensions 160 mm (6.30 in.) wide x 214 mm (8.43 in.) high x

68 mm (2.68 in.) deep

Weight 2 kg (4½ lb)

Panel-mounting (¼ DIN)

Protection IP66 / NEMA4X front

Dimensions 96 mm (3.78 in.) wide x 96 mm (3.78 in.) high x

191 mm (7.52 in.) deep

Weight 1.5 kg (3¼ lb)

Panel cut-out: mm x mm

( in. x in.)

92+0.80– 92+0.8

0–

3.62+0.030– 3.62+0.03

0–

Page 10: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

10 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

Overall dimensions

4690 turbidity sensor (with optional wiper unit)

Dimensions in mm (in.)

6 x Ø7 (0.3) Mounting Holes

Sample Drain Connector(12 mm [0.5] internal diameter tube)

Sample Inlet Connector(12 [0.5] internal

diameter tube)

403.0(17.0)

298.0 (11.7)

230.0 (9.0)

118.0(4.65)

155.5 (6.12)46.5 (1.83)

Page 11: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 11

4690 turbidity sensor (without optional wiper unit)

Dimensions in mm (in.)

6 x Ø7 (0.3) Mounting Holes

Sample Drain Connector(12 mm [0.5] internal diameter tube)

Sample Inlet Connector(12 [0.5] internal

diameter tube)

298.0 (11.7)

230.0 (9.0)

282.0(11.1)

118.0(4.65)

155.5 (6.12)46.5 (1.83)

Page 12: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

12 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

4690 wall- / pipe-mount analyzer

4695 panel-mount analyzer

Optional de-bubbler assembly

Dimensions in mm (in.)

Dimensions in mm (in.)

232(9.13)

214(8.43)

250(9.84)

160 (6.3) 68 (2.68)

69 (2.72)

92(3.62 )

+0.8–0

+0.003–0

92 (3.62 )96 (3.78)

96(3.78)

12 (0.47) 191 (7.52)

Panel cut-out

+0.8–0

+0.003–0

Dimensions in mm (in.)

Disconnect Fittings – 12 (0.5) ID Tube (rotatable through

360°)

Removable Fitting

12 (0.5) ID Tubing

150 (5.9) CRS

Drain Outlet

Sample Outlet

Sample Inlet

540 (21.3)

Ø 6.5(0.25)

88(3.5)

91(3.6)

Ø 6.5(0.25)

Page 13: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 13

Electrical connections

Panel-mount analyzer connections

Wall- / Pipe-mount analyzer connections

Note.1. A second retransmission output is available if the RS485 serial communications facility is not used.2. If 'Test Cleaner' is selected during analyzer configuration, Relay 2 becomes 'Failed Wiper Alarm' relay.

Relay 2(see Note 2)

Earth (Ground) Stud

Earth (Ground) Stud on Case

Relay 1

RetransmissionOutput

NCC

NONC

CNO

EarthNeutral

LinePower Supply

RS485 Serial Interface (If fitted)

Turbidity Sensor

Link to Earth (Ground) Stud on Case

2nd RetransmissionOutput(see Note 1)

0VRx–Rx+Tx–TX+Blue

BlackGreenRedYellowWhite

Braid

Turbidity Sensor7998 Series

SerialRS485

RetransmissionOutput

Relay 1Relay 2

Power Supply

Earth (Ground) Stud

Earth (Ground) Stud on Case

Earth (Ground) Stud on Case

1 – White2 – Yellow3 – Red4 – Green5 – Black6 –

Blue7 –

Braid

SerialRetransmission

Relay 1Relay 2

PowerSupply

1 –2 – Rx+3 – Rx–4 – Tx+5 – Tx–6 – 0V(Optional)

RetransmissionOutput

2nd RetransmissionOutput (see Note 1)

Relay 1

Relay 2(see Note 2)

N – NeutralL – Line

– Earth

1 – NC2 – C3 – NO4 – NC5 – C6 – NO

Page 14: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 SeriesTurbidity systems

14 DS/4690–EN Rev. A | 4690 Series | Turbidity systems

Typical system installation schematic

With optional de-bubbler

Dimensions in mm (in)

Sample In

Drain Valve

Sample Outlet ConnectorTundish

500 (19.7) Minimum

Drain Outlet

De-bubbler

4690 Analyzer

4695 Analyzer

Inlet / Isolating Valve

150 (5.9) Maximum

Sample Out

7998SeriesSensor

150 (5.9) Maximum

Tundish

Optional

Page 15: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

4690 Series | Turbidity systems | DS/4690–EN Rev. A 15

Ordering information

Accessories

Turbidity system 4690/ X X X X X X X X

Transmitter type

Wall-/ Pipe-mount (gland fittings)

Wall- / Pipe-mount (conduit)

Panel-mount

1

2

5

Voltage

115 V AC

230 V AC

1

2

Communications and IO

Standard (1 analog output)

Additional analog output

Modbus

1

2

3

Sensor type

Flow-through system (ISO 7027 compliant) 1

Flow-through system (EPA 180.1 compliant) 2

Sensor range

0 … 40 NTU (without auto-clean)

0 … 40 NTU

0 … 400 NTU

1

2

3

Secondary dry calibration

Not included

Low range (<5 NTU)

High range (50 … 100 NTU)

0

2

3

Sensor cable length

No sensor cable supplied

1 m (3.3 ft)

5 m (16.4 ft)

10 m (32.8 ft)

0

1

2

3

Manual

English 1

Description Part number

De-bubbler assembly 7997 500

Dry standard HIGH (for ISO infrared LED version) 7998 048

Dry standard LOW (for ISO infrared LED version) 7998 047

Dry standard HIGH (for EPA white light version) 7998 036

Dry standard LOW (for EPA white light version) 7998 035

Page 16: 4690 Series Turbidity systems Accurate, reliable turbidity measurement€¦ ·  · 2013-07-31Data sheet DS/4690–EN Rev. A 4690 Series Turbidity systems Accurate, reliable turbidity

Contact us

DS

/469

0–E

N R

ev. A

07.2

012ABB Limited

Process AutomationOldends LaneStonehouseGloucestershire GL10 3TAUKTel: +44 1453 826 661Fax: +44 1453 829 671

ABB Inc.Process Automation125 E. County Line RoadWarminsterPA 18974USATel: +1 215 674 6000Fax: +1 215 674 7183

www.abb.com

NoteWe reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.