of 103
8/18/2019 61981264 Service Manual YP 90 Series
1/103
EDITION/REVISION B/0
NINGBO DAVIDR
SERVICE MANUAL
YP-90 YP-90A
YP-90B YP-90 AB
90 SERIESINFANT INCUBATOR
8/18/2019 61981264 Service Manual YP 90 Series
2/103
NINGBO DAVID MEDICAL DEVICE CO., LTD.
ADD: NO.158 DAQING ROAD, SHIPU, NINGBO, ZHEJIANG PROVINCE, CHINA
POST CODE: 315731 FAX: 0086-574-65962111
MARKETING CENTER: NO. 100, JINGHUA ROAD, HI-TECH INDUSTRIAL DEVELOPMENT ZONE,
NINGBO, CHINA
POST CODE: 315000
TEL: 0086-574-87800008, 87800009 FAX: 0086-574-87801111
E-MAIL: [email protected] [email protected]
WEBSITE: http ://www.nbdavid.com
EUROPEAN REPRESENTATIVE: SHANGHAI INTERNATIONAL HOLDING CORP. GMBH (EUROPE)
ADD: EIFFESTRASSE 80, 20537 HAMBURG, GERMANY
TEL: 0049-40-2513175 FAX: 0049-40-255726
EDITION/REVISION B/0
8/18/2019 61981264 Service Manual YP 90 Series
3/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
iI2
WARRANTY
The product being described in this manual is warranted against defects in materials or
workmanship for one year from the date of shipment, with the following exceptions.
1. All consumable and disposable products are guaranteed to be free from defects upon
shipment only.
2. Calibrations are considered normal maintenance and are not included in the 1-year warranty.
During the warranty period any defective parts other than those listed above will be replaced at no
charge to the customer.
This warranty is rendered void and our company cannot be held liable for conditions resultant
therefrom if:
1. Damage to the unit is incurred as a result of mishandling.
2. The customer fails to maintain the unit in a proper manner.
3. The customer uses any parts, accessories, or fittings not specified or sold by our company.
4. Sale or service is performed by the non-certified service/dealer agency.
This warranty is in lieu of all other warranties, expressed or implied, and our company shall in no
event be liable for incidental or consequential damages including loss of use, property damage, or
personal injury resulting from breach of warranty.
The Accreditation Manual for Hospitals requires each piece of equipment to be tested prior to initial
use and at least annually thereafter. To comply with this standard, we recommend that you participate inour accreditation Testing compliance Program during the warranty period. This service can be
performed through our company and authorized dealers.
SERVICE
For optimal performance, product service should be performed only by qualified service personnel.
Technical Services representatives can be reached for fixations and are dispatched for required
maintenance by calling 0086-574-87801003, 87800002. Customers outside China should contact their
local factory-authorized distributor for service.
8/18/2019 61981264 Service Manual YP 90 Series
4/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
iiI2
COMPLEMENTARY NOTICE
Since our company, conducts a continuous product improvement program, circuit and component
improvements are sometimes incorporated into equipment before they can be incorporated into the
printed manuals. This manual contains proprietary information, repairs and authorized modifications
should be performed only by qualified service personnel to maintain your warranty and to avoid creating
safety hazards. We cannot assume responsibility for any conditions affecting the proper operation of this
equipment which may result from unauthorized repair modification. Therefore, some parts used in your
equipment may be different than those which appear in the parts list of this manual. This sometimes
occurs due to difficulty in parts procurement, but does not alter the function of the equipment
The device, accessories and the packaging have to be disposed of waste correctly at the end of the
usage. Please follow Local Ordinances or Regulations for disposal.
TABLE OF DEFINITIONS AND SYMBOLS
TECHNICAL DEFINITIONS
SKIN TEMPERATURE SENSOR: A sensing device including the link with the equipment intended to
measure the infant’s skin temperature.
INCUBATOR TEMPERATURE: Air temperature at a point 10cm above and centered over the mattress
surface.
CONTROL TEMPERATURE: The temperature set at the temperature control.
AVERAGE INCUBATOR TEMPERATURE: The average of the maximum and minimum Incubator
temperatures achieved during Temperature Condition.
STEADY TEMPERATURE CONDITION: A condition which is reached when the temperature does not
vary by more than 1℃ over a period of 1 hour.
TEMPERATURE ALARM CHECKOUT STATE: The difference between real temperature and control
temperature is within ±0.5℃ and such state lasts for over 10 minutes. The equipment must stay in such
state when check up the alarm about temperature
TEMPERATURE UNIFORMITY: The amount by which the average temperature at each of four points
10cm above the mattress surface differs from the Average incubator Temperature at steady
Temperature Condition. The four points are the centers of four quadrants formed by lines that divide the
width and length of the mattress surface.
8/18/2019 61981264 Service Manual YP 90 Series
5/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
iiiI2
TEMPERATURE VARIABILITY: The variability of the Incubator Temperature that will be observed over
a one hour period after Incubator Temperature Equilibrium has been reached.
TEMPERATURE RISING TIME: The time required for the Incubator Temperature to rise 11℃, when the
Air Control Temperature is at least 12℃ above ambient.
NOTE, IMPORTANT, CAUTION AND WARNING
NOTE: A note is inserted in text to point out procedures or conditions, which may otherwise be
misinterpreted or overlooked. A note may also be used to clarify apparently contradictory or
confusing situations.
IMPORTANT: Similar to a Note but be used where greater emphasis is required.
CAUTION: A caution is inserted in text to call attention of a procedure which, It not followed
exactly, can lead to damage or destruction of the equipment.
WARNING: A warning is inserted in text to call attention to dangerous or hazardous conditions
inherent to the operation, cleaning, and maintenance of the equipment which may result in
personal Injury or death of the operator or patient.
8/18/2019 61981264 Service Manual YP 90 Series
6/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SYMBOLS
ivI2
>37
Attention, Consult Accompanying Documents.
Class I Equipment Type BF Applied Part
On (Connection to the mains) Off (Disconnection to the mains)
On (Only For A Part Of Equipment) Off (Only For A Part Of Equipment)
RS-232 Connector AC Power 220-230V, 50Hz
High temperature warning label Protective Earth (Ground)
Earth (Ground) Weight Limit
Heating Power Indicator Mattress Tilt Direction
Low Water Level High Water Level
Air Mode Key Baby Mode Key
Silence / Reset Key Keypad Key
Set Up Key Set Down Key
Lower device’s height Temperature override modeselection key
Raise device’s height Type T Fuse 400mAL/250V
Type F Fuse 5AH/250V Fuse Type F 6.3AH/250V
Type T Fuse 160mAL/250V Type T Fuse 800mAL/250V
Serial Number Date Of Manufacture
CE Marking
Auxiliary mains output, MAX:1.6A
CLASS I
RS-232 220-230V~/50Hz
T 400mAL/250V
F 6.3AH/250VF 5AH/250V
T 800mAL/250VT 160mAL/250V
SN
220-230V~/50Hz MAX:1.6A
8/18/2019 61981264 Service Manual YP 90 Series
7/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
vI2
SEASONAL SAFETY CHECK
1. Please clean the plug of power cord at least once a year. Too much dust on plug may cause the fire.
2. The following safety checks should be performed at least every 12 months by a qualified person who
has adequate training, knowledge, and practical experience to perform these tests. The data should be
recorded in an equipment log.
①. Inspect the equipment and accessories for mechanical and functional damage.
②. Inspect the safety relevant labels for legibility.
③. Inspect the fuse to verify compliance with rated current and breaking characteristics.
④. Verify that the device functions properly as described in the instructions for use.
⑤. Test the protection earth resistance according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit
0.1Ω.
⑥. Test the earth leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: NC
500μ A, SFC: 1000μ A.
⑦. Test the enclosure leakage current according to IEC 60601-1:1988 + A1:1991 + A2:1995: Limit:
NC 100μ A, SFC: 500μ A.
⑧. Test the patient leakage current according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: for
a.c.: 100μ A (BF), for d.c.: 10μ A (BF).
⑨. Test the patient leakage current under single fault condition with mains voltage on the applied
part according IEC 60601-1:1988 + A1:1991 + A2:1995: Limit: for a.c.:500μ A (BF), for d.c.: 50μ A (BF).
⑩. According to the test methods of IEC 60601-1:1988 + A1:1991 + A2:1995, the patient leakage
current (net voltage should be added on the applied part) of the testing device must less than 5000μ A.
○11 . Test the patient auxiliary leakage current according IEC 60601-1:1988 + A1:1991+ A2:1995:
Limit: NC for a.c.: 100μ A (BF), for d.c.: 10μ A (BF).SFC 500μ A (BF), for d.c.: 50μ A (BF).
8/18/2019 61981264 Service Manual YP 90 Series
8/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
viI2
TABLE OF CONTENTS
SECTION PAGE
1. GENERAL INFORMATION…………………………………………………………………………………1-1
1.1 INTRODUCTION……………………………………………………………………………………………1-1
1.2 DESCRIPTION……………………………………………………………………………………………1-1
1.3 ACCESSORIES……………………………………………………………………………………………1-1
2. INSTALLATION………………………………………………………………………………………………2-1
2.1 UNPACKING………………………………………………………………………………………………2-1
2.2 ASSEMBLING ………………………………………………………………………………………………2-1
2.3 OPERTION CHECKOUT PROCEDURES ………………………………………………………………2-6
3 TECHNICAL INFORMATION………………………………………………………………………………3-1
3.1 SPECIFICATION……………………………………………………………………………………………3-1
3.2 WORKING PRINCIPLE……………………………………………………………………………………3-3
4. PREVENTIVE MAINTENANCE……………………………………………………………………………4-1
4.1 GENERAL……………………………………………………………………………………………………4-1
4.2 CLEANING…………………………………………………………………………………………………4-1
4.3 STERILIZATION……………………………………………………………………………………………4-8
4.4 MAINTAINANCE……………………………………………………………………………………………4-8
5 SERVICE………………………………………………………………………………………………………5-1
5.1 GENERAL……………………………………………………………………………………………………5-1
5.2 REPLACEMENT OF ELDMENT…………………………………………………………………………5-1
5.3 SYSTEM SETTING PROCEDURES……………………………………………………………………5-2
5.4 TROUBLESHOOTING PROCEDURES…………………………………………………………………5-6
6. REPLACEMENT PARTS……………………………………………………………………………………6-1
6.1 GENERAL……………………………………………………………………………………………………6-1
7. DIAGRAMS…………………………………………………………………………………………………7-1
7.1 GENERAL……………………………………………………………………………………………………7-1
8/18/2019 61981264 Service Manual YP 90 Series
9/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SECTION 1
GENERAL INFORMATION
1.1 INTRODUCTION
This manual provides instructions for installation, maintenance and repair the David InfantIncubators, Model YP-90,YP-90A, YP-90B, YP-90AB. Among these models, YP-90AB is with all
functions other models have.
This manual is intended for use only by trained, qualified service personnel. Instructions for
the operator of the equipment are provided in a separate operator’s manual.
1.2 DESCRIPTION
Based on YP-90, the Infant Incubator YP-90A added with baby mode , cabinet; the Infant
Incubator YP-90B added the Neonate Bilirubin Phototherapy Equipment; YP-90AB added with
baby mode and the Neonate Bilirubin Phototherapy Equipment, cabinet.
YP-90, YP-90B just have one kind of control mode-air mode, YP-90A, YP-90AB have 2
kinds of control mode- air mode and Baby mode. Optional part :>37℃ Temperature Override
Mode, Neonate Bilirubin Phototherapy Equipment , and Vertical Height Adjustment cabinet (VHA
cabinet) . If you donot purchase the optional part, please omit it.
1.3 ACCESSORIES
The following diagram shows the main parts of the Infant Incubator:
Neonate Bilirubin Phototherapy Equipment
I.V. Pole
Baby compartmentShelf
1-1I2
Controller
Base
Mobile cabinet
8/18/2019 61981264 Service Manual YP 90 Series
10/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
1-2 I2
DESCRIPTION OF PART EXPLANATION
I.V. Pole A kind of bearing part, which is used for hanging the infusion bottle.Max. Load: 2Kg
Controller
The core part with two kinds of temperature control modes: airmode, baby mode, and the optional function :>37°TemperatureOverride Mode, and it is used for automatic controlling of heatoutput, please refer to section 4 on the user's manual.
Shelf A kind of bearing part, which is used for putting some small objects.Max. Load: 3.5Kg
Baby compartment
It is used for place the infant inside, including the Acrylic hood,bassinet, and so on. The bassinet can be tilted at ±12°at the requestof clinical needs. The baby scale is optional part, and the Max. Loadof bassinet is 10 Kg.
Size of mattress: 630mm × 355mm
Base An important part of Infant Incubator, and it is mainly composed ofthe Aluminium tank, humidity chamber, air filter, and so on.
Mobile cabinet
A bearing part for the incubator, namely stand, cabinet or VHAcabinet of the incubator. The standard equipment for YP-90 and YP-90B is stand, for YP-90A and YP-90AB is cabinet. VHA cabinet isequipped as an optional part.
Neonate BilirubinPhototherapy
Equipment (for YP-90B,YP-90AB)
It is intended for treating the bilirubin of patient.
This part is mounted on top of Infant Incubator , and its light sourcehas two kinds: fluorescent lamp and LED lamp, the user can choosethis.for the Neonate Bilirubin Phototherapy Equipment with fluorescentlamp, please refer to the user’s manual XHZ-90, while for theNeonate Bilirubin Phototherapy Equipment with LED lamp, pleaserefer to the user’s manual XHZ-90L.
NOTE: Size of Infant Incubator: W600×H1550×D1150mm(YP-90、YP-90B)
W650×H1560×D1440mm(YP-90A、YP-90AB)
The distance from the bassinet to the floor: 950mm
Weight of Infant Incubator :65Kg(YP-90)
103Kg(YP-90A)
75Kg(YP-90B)
116Kg(YP-90AB)
8/18/2019 61981264 Service Manual YP 90 Series
11/103
8/18/2019 61981264 Service Manual YP 90 Series
12/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
b. Install the main body onto the fixed stand
1). See figure 2.2, lock the casters with brake, and put the main body onto the fixed, and connect it
with the lock mechanism on two sides of stand.
Main body of Incubator
2-2
I2
Lock mechanism
II
Fixed Stand
I
(Enlarged figure I) (Enlarged figure II)
1
2
wheel unlocked wheel locked hang on the lock close the lock spanner
NOTE: See Enlarged figure I, lock the casters as the arrow 1 ind icates, and unlock the casters as
the arrow 2 indicates.
See Enlarged figure II, hang on the lock , and close the lock spanner.FIGURE 2.2
8/18/2019 61981264 Service Manual YP 90 Series
13/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
B. The installation of main body onto the cabinet
a. As figure 2.3 indicates, put on the spring washer, flat washer on the casters, and then connect with
cabinet firmly. Note: The caster with brake must be installed in front of the cabinet.
FIGURE 2.3
b. As figure 2.2 indicates, lock the casters with brake, see figure 2.4, and put the guard rail and main
body on the cabinet, and then fix them with the lock mechanism on two sides of cabinet.
Main body of incubator
Guard rail
Lock mechanism
Cabinet
FIGURE 2.4
2-3I2
8/18/2019 61981264 Service Manual YP 90 Series
14/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
C. Install the main body onto the Vertical Height Adjustment Stand
a. The method of installation casters is same as part B.
b. The method of installation main body, guard rail on the Vertical Height Adjustment cabinet same as
part B.
c. After installation, insert the power cord on the right side of Vertical Height Adjustment cabinet into
the power socket in figure 4.2.
2-4
I2
Main body of incubator
Power cord
Guard rail
Lock mechanism
The VHA Stand
FIGURE 2.5
The guard rail ensure grounding th rough bo lt therefore, fixing the guard
railon the cabinet with bolts.
WARNING
Firm the main body onto the fixed cabinet or Vertical Height Adjustment cabinet with lock
mechanism. If not, it will cause the main body to deviate from the fixed cabinet or Vertical
Height Ad justment cabinet when the cabinet tilts, especially when the hood is open.
2. Install the NEONATE BILIRUBIN PHOTOTHERAPY EQUIPMENT, I.V. Pole and shelf
A. Install the NEONATE BILIRUBIN PHOTOTHERAPY EQUIPMENT
Please refer to the method of installation for incubator mounted PHOTOTHERAPY unit XHZ-90 and
XHZ-90L in the accompanying documents.
B. Install I.V. Pole
1) Install the I.V. Pole beside the incubator as figure 2.6 indicates, and then adjust the hook to a
proper height, fix it with the turnbuckle.
8/18/2019 61981264 Service Manual YP 90 Series
15/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Hexagon bolt
2-5I2
Fixed seat
Infusion rod
Step 1: loose the hexagon bolts on the fixed seat with spanner, insert the I.V. Pole to the bottom, and fix
the bolt to fasten the infusion rod.
Hook For Infusion Rod
Hexagon Bol t
Step 2: Tight the hexagon bolts on the
bottom of infusion rod to fix the rod.
Step 3: Rotate the hook into the I.V. Pole,
and then tighten the hexagon bolts.
FIGURE 2.6
C. Install shelfUnscrew the hexagon bolts on the fixed seat for shelf, See figure 2.7, and put on the fixed seat on the
infusion rod, see Enlarged figure A, to make sure the top of fixed seat and the position limit mark on the
rod is on same level. And then reinstall the hexagon bolt on the fixed seat.
ENLARGED FIGURE A
Infusion Rod position limit mark
AHexagon Bolt
Fixed Seat
NOTE: when installing the shelf, the side with note label should face the front of incubator.
FIGURE 2.7
3. Insert the power cord
Insert the power cord into the socket of general power supply on the right side of main body; For the
VHA stand, Insert the power cord into the socket of general power supply on the VHA stand.
Check the WARMER according to the instruction in section 2.3.
8/18/2019 61981264 Service Manual YP 90 Series
16/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
2-6
I2
2.3 OPERATION CHECKOUT PROCEDURE
WARNING
1. Please stop using this model once some function lost or the spare parts for fron t
access panel is loose.
2. set temperature must be high 3 than ambient temperature. And then you canproceed this checkout procedure.
Please proceed the following checkout procedure each time before operation.
A. Check caster
Check whether the caster can drop when lifting the radiant warmer 2cm above the ground. The drop
of the caster will cause danger during transporting.
IMPORTANT: At least two persons are required, one lifts the warmers, and the other has to check
the casters. It needs Min. 140Kg to l ight the whole unit . Please support the stand withblock to avoid the warmer inclining if the casters drop.
B. START-UP THE CONTROLLER
Turn on the switch of the main power supply and controller, the controller gives sound “Ding”, at the
same time, the Set Temperature indicator does not indicate, while the Ai r Temperature indicator and
Skin Temperature indicator indicates“--.-”, it means the controller is doing self-test automatically for 5s.
After then, the Ai r Temperature indicator will displays air temperature, the Skin Temperature indicator
displays the baby temperature, and Set Temperature indicator displays the former set value, and the
controller can enter into the control mode automatically before turning off, and the relevant mode light will
be on.
C. CHECK POWER FAILURE ALARM
Pull off the power cord of whole unit, the Power failure alarm should active, and the “Power” light will
flash with continuous audible sound.
This operation is used for checking if the power failure is normal or not. Insert the power cord again
after finishing checking.
IMPORTANT: Make sure that the rechargeable battery is fu ll before usage. If not fu ll, it may causethe power failure without the alarming indication. If full, and there is no any
indication after disconnecting the main power supply, please refer to the authorized
and qualified service personnel.
D. CHECK HEATER
Control the environment temperature at 21℃ ~ 26℃, and choose the Ai r Mode, and set the
temperature at 33.0 ℃, and all heat power indicator are on, and heater will output heat completely.
NOTE: When the incubator works under the set state, and continue the following operation
procedure.
8/18/2019 61981264 Service Manual YP 90 Series
17/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
E. CHECK THE ACCURACY OF TEMPERATURE CONTROL
Select the Air Mode, and set the air temperature at 36℃, after the air temperature enters into the
STEADY TEMPERATURE CONDITION, put the calibrated temperature measuring device on the position
above 10cm from the center of mattress to measure the air temperature, compared with the indicated air
temperature to check whether the deviation between them is within 0.8℃.
F. CHECK BABY MODE
Opening
2-7I2
Arrow Sensor socket
The arrow sign on the plug should aim to the opening on the sensor socket so that the sensor is
inserted correctly.
1. Insert or pull out the skin sensor, you must hold the plug of skin sensor,
pulling the leads is forbidden.
2. Please do not bend the connection of sensor.
When the Ai r Mode works stably, change it into the Baby Mode, set temperature at 37℃. Put the
skin sensor on the center position, and the incubator will give alarm after a while. The “Sensor” alarm
light will flash with continuous audible sound, and the Set Temperature indicator will show alarm code
E0.8.
Pull off the skin temperature sensor, the sensor failure alarm should active. The “Sensor”
alarm light will flash with continuous audible sound, and the Set Temperature indicator will show alarm
code E0.5.
G. CHECK PRECISION OF SKIN TEMPERATURE SENSOR
Put the skin temperature sensor with the mercury thermograph for the accuracy with ±0.1ºC into the
water cup with the temperature as 30ºC±5ºC. Make the probe of the skin temperature sensor and the
mercury ball as closely as possible and stir enough, then read the value of mercury thermograph.
Compare the value of the skin temperature sensor and the mercury thermograph, and the deviation must
be within 0.5ºC.
NOTE: Please check again if the accuracy of the skin temperature sensor exceeds the
permissibility deviation. Please let the authorized and qualified service personnel service
machine if the accuracy the skin temperature sensor exceeds the permissibility of
deviation again.
8/18/2019 61981264 Service Manual YP 90 Series
18/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
H. CHECK FAN MOTOR ALARM
Block the inlet port and outlet port on the left and right side of main deck, after several minutes, the
device gives alarms, the “Fan” alarm light will flash with continuous audible sound, and the Set
Temperature indicator will show alarm code E0.7.
Turn off the power supply of controller, and pull off the power cord, and take out the controller.Reconnect the power cord, and hold the fan slightly, and turn on the switch of controller, and the fan
motor alarm occurs, the “Fan” alarm light will flash with continuous audible sound, and the Set
Temperature indicator will show alarm code E0.9.
WARNING: The temperature of heater inside controller is too high, and you can
remove the controller or touch the heater after the unit is switched o ff at
least 45min to avoid the burnt.
I. CHECK OVER-TEMPERATURE ALARM
In Air Mode, enter into the working state, press Up, Down and Silence/Reset key, there is no
indication on the Set Temperature indicator, and the heat power indicator is on in full, and the controller
has entered into the overtemp checkout state, and when the incubator temperature is less than 38.0 ,℃
the over-temperature alarm should active. The “Over” light should flash with continuous audible sound,
and the Set Temperature indicator should show alarm code E0.6. Under the state of 37
Override
Mode, enter into the working state, press Up, Down and Silence/Reset key, there is no indication on the
Set Temperature indicator, and the heat power indicator is on in full, and the controller has entered into
the overtemp checkout state, and when the incubator temperature is less than 40.0℃, the
over-temperature alarm should active. The “Over” light should flash with continuous audible sound, and
the Set Temperature indicator should show alarm code E0.6.
In Baby Mode, press Up, Down and Silence/Reset key, there is no indication on the Set
Temperature indicator, and the heat power indicator is on in full, and the controller has entered into the
overtemp checkout state, and when the incubator temperature is less than 40.0 , the℃ over-temperature
alarm should active. The “Over” light should flash with continuous audible sound, and the Set
Temperature indicator should show alarm code E0.6.
J. CHECK DEVIATION ALARM
In Ai r Mode, close all door and panels, set the temperature at 32 ºC. Enter TEMPERATURE ALARM
CHECKOUT STATE, fan into the hot air inside the hood. When the Ai r Temperature indicator indicates
35.1ºC, the HIGH deviation alarm will active. The “High” light should flash with continuous audible sound,
and the Set Temperature indicator should show alarm code E1.0; Set the temperature at 35℃, after the
device enters into the TEMPERATURE ALARM CHECKOUT STATE, open the front access panel, when
the Ai r Temperature indicator indicates 31.9 ºC, the low deviation alarm should active. The “Low” light
should flash with continuous audible sound, and the Set Temperature indicator should show alarm code
E1.1.
2-8
I2
8/18/2019 61981264 Service Manual YP 90 Series
19/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
NOTE: If the system can not enter into the TEMPERATURE ALARM CHECKOUT STATE or the air
temperature does vary within ±3
than the sett ing temperature, the deviation alarm
can not occur.
In Baby Mode, set the temperature at 35 ºC. Enter TEMPERATURE ALARM CHECKOUT STATE,
put the skin sensor into the water cup at 37 ºC. When the skin temperature indicates 36.1 ºC, the high
deviation alarm will active. The “High” light should flash with continuous audible sound, and the Set
Temperature indicator should show alarm code E1.0; Set the temperature at 35℃, after the device
enters into the TEMPERATURE ALARM CHECKOUT STATE, put the skin sensor into the water cup at
temperature 33℃, when the skin temperature indicates to 33.9ºC, the low deviation alarm should active.
The “ Low” light should flash with continuous audible sound, and the Set Temperature indicator should
show alarm code E1.1.
NOTE: If the system can not enter into the TEMPERATURE ALARM CHECKOUT STATE or the
baby temperature does vary within ±1 than the sett ing temperature, the deviation
alarm can not o ccur.
K. CHECK THE HOOD OPERATION
See the following figure, raise the hood until you hear the sound “kata”, and the hood is in the lock
state. Press the lock downward as the arrow indicates so that the hood can return the original position.
CAUTION: Before raise the hood, ensure all accessories is disconnected to avoid interfere
operating.
2-9I2
Disconnect the skin sensor, slowly tilt the pressing the lock as the arrow indicates, and close
Hood back until the Hood locks in place. the hood slowly.
8/18/2019 61981264 Service Manual YP 90 Series
20/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
L. CHECK FRONT ACCESS PANEL
The Pawl Latch
LOCKED UNLOCK
Rotate the panel Latches and open the Access Pane to the full open position (rotate downward).
Close the Access Panel and rotate both latches until they are fully locked. Both latches must be fully
locked to avoid accidental opening of the panel.
M. CHECK IRIS ENTRY PORTS
Rotate the outer ring of each entry Port, the port should be open and close as rotation is continued
through 360 degree.
N. CHECK THE LATCH AND GASKET OF ACCESS DOOR
Press on-off of each access door to open it automatically and check the seal of gaskets.
2-10
I2
8/18/2019 61981264 Service Manual YP 90 Series
21/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
2-11I2
O. CHECK MATTRESS TILT MECHANISM
The mattress tilt mechanism to adjust it.
P. CHECK BASSINET
Open the Access Panel and slide it out to the fully extended position indicated by the arrow. Lean on
Mattress Tray to make sure it is properly supported to provide a firm infant platform. Return the mattress
and close Access Panel.
Q. CHECK THE AIR INTAKE FILTER
Loosen the two thumb screws of the Air Intake Filter Cover and remove the cover and inspect the
filter, if dirty, it should be cleaned. Put the filter cover back.
8/18/2019 61981264 Service Manual YP 90 Series
22/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
WARNING: A dirty Air Intake filter may affect oxygen concentration and/or cause Carbon
Dioxide build-up. The filter must be checked on a routine basis and changed at
least every two months.
R. CHECK THE HIGHT ADJUSTMENT OF THE CABINET
Step the Up and Down button with your feet to adjust the height of whole unit.
Down foot button
2-12
I2
Up foot button
S. CHECK OXYGEN INPUT VALVE
Connect the oxygen tube on the monitor with the oxygen input valve on back of incubator (see
following figure), and make sure its seal. Input the 8lpm of oxygen, monitor it with a calibrated oxygen
analyzer inside hood, check whether the indicated value of oxygen analyzer reach the predicted level as
indicated on the oxygen concentration table.
T. CHECKOUT PROCEDURE ENDS
8/18/2019 61981264 Service Manual YP 90 Series
23/103
8/18/2019 61981264 Service Manual YP 90 Series
24/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-2
I2
TABLE 1.1 SPECIFICATIONS (continued)
Deviation of air and isolated sensor failure…………………………………………… Alarm code E0.3
Ai r f low temperature sensor fai lure……………………………………………………… Alarm code E0.4
Skin temperature sensor failure………………………………………………………… Alarm code E0.5
Over-temp failure…………………………………………………………………………… Alarm code E0.6
Ai r cycle fai lure …………………………………………………………………………… Alarm code E0.7
Wrong position of skin sensor failure…………………………………………………… Alarm code E0.8
Fan motor failure…………………………………………………………………………… Alarm code E0.9
High deviation alarm……………………………………………………………………… Alarm code E1.0
Low deviat ion alarm………………………………………………………………………… Alarm code E1.1
Internal system failure……………………………………………………….. Alarm code H0.1~H1.3
ENVIRONMENT TEMPERATURE
Operating Range…………………………………………………………………………………+20~+30℃
Storage Range……………………………………………………………………………………-40~+55℃
ENVIRONMENT HUMIDITY
Operating Range……………………………………………………………………………………30-75%RH
Storage Range………………………………………………………………………………………< 93%RH
ATMOSPHERIC PRESSUR
Shipment and Store atmospheric pressure range……………………………………500hPa-1060hPa
Operating atmospheric pressure range………………………………………………700hPa-1060hPa
AIR VELOCITY
Ambient Air Veloci ty…………………………………………………………………………………< 0.3m/s
OTHER SPECIFICATION
Noise level within Hood Environment…………………………………………………………< 50dB(A)
Related ambient noise should be
8/18/2019 61981264 Service Manual YP 90 Series
25/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-3
I2
3.2 WORKING PRINCIPLE
3.2.1 General
This section contains a functional description and detailed theory of operation of the equipment. A
system block diagram of the Controller is shown in Figure 3.2 and 3.3. The control of temperature is achieved by means of air circulation system as shown in Figure 3.1.
A controlled amount of room air is drawn through the air intake filter by means of the motor-driven
impeller on the Controller and then drawn back to motor though air return filter. Circulated air flows
through temperature sensing sensor. Temperature Controller automatically modulates the temperature
inside Hood according to the temperature sensed by the air temperature sensing sensor.
Heat output control under air mode: control the output of heater according to the air temperature
automatically;
Heat output control under baby mode: control the output of heater according to the skin
temperature automatically.
FIGURE 3.1
8/18/2019 61981264 Service Manual YP 90 Series
26/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-4
I2
A A
B
FIGURE 3.2 CONTROL BLOCK DIAGRAM
MAIN
MCU
Keypad control
Indicator
I/O control
Fan motor checking
Temperature heatercontrol
ADC
Air Temperature sensor
AUXILIARYMCU
Skin temperature sensor
Air flow temperature sensor
ADC Isolated temperature sensor
Temperature heater control
8/18/2019 61981264 Service Manual YP 90 Series
27/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Start
INT1
Initialization
Real timer
Data displayReturn
3-5
I2
INT0
Timer interruption
Series portinterruption
FIGURE 3.3 PROCEDURE BLOCK DIAGRAM
Key input
Data handling
Heating control
Trouble checking
Alarmidentification
YES
NO
Data display
Key handling
YES
NO
ADC Sampling
Return
Timer
Return
Data transfer
Return
8/18/2019 61981264 Service Manual YP 90 Series
28/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-6
I2
3.2.2 Temperature Control
In baby Mode, the controller will control the incubator temperature automaticlly to keep the air
temperature close to the set value. Comparing the air temperature and the air set temperature, the
controller will adjust the heat output proportion to keep the heat balance inside hood. The other probe
inside of air sensor is regarded as the isolated temperature sensor, when set temperature is < 37ºC
incubator temperature must be≤38.0ºC (while when set temperature is > 37ºC incubator temperature
must be≤40.0ºC ), the heater is cut off automatically.
In Air Mode, the controller will control the incubator temperature automaticlly to keep the skin
temperature close to the set value. Skin sensor is attached to the skin of patient. Comparing the skin
temperature and the skin set temperature, the controller will adjust the heat output proportion to keep
the heat balance inside hood. In the whole baby mode, the isolated temperature sensor of air sensor is
always monitor the inside temperature, when the incubator temperature is ≤40ºC, the heater is cut off
automatically.
When the baby mode is working, if the skin temperature sensor drops from the socket, the skin
temperature indicator displays “--.-”, the sensor alarm actives and the heater is cut off automatically.
3.2.3 Alarm
Alarm is used for monitoring the temperature control system state. The system will check the state
by the alarm checkout procedure. When the checked data is abnormal, it will alarm. Table 3.2 listed
alarm condition.
TABLE 3.2 ALARM
ALARM MESSAGE DESCRIPTION
Power Failure indicator is onIf the mains power supply fails or no power supply, the “ POWER” alarm with audible and visual will active.
Code
E0.1
In air mode, If short-circuit or open-circuit or bad connection occurs on the air temperature sensor, at the same time, heater stops
working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.
CodeE0.2
If short-circuit or open-circuit or bad connection occurs on theisolated temperature sensor, at the same time, heater stopsworking, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
Sensor Alarmindicator is on
CodeE0.3
If the deviation between the air temperature and isolatedtemperature sensor is more than 0.8℃, at the same time, heaterstops working, press Silence/Reset Key to clear alarming for 4min,
please refer to the service part.
8/18/2019 61981264 Service Manual YP 90 Series
29/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-7
I2
TABLE 3.2 ALARM
ALARM MESSAGE DESCRIPTION
CodeE0.4
If short-circuit or open-circuit or bad connection occurs on the air flow
temperature sensor, at the same time, heater stops working, pressSilence/Reset Key to clear alarming for 4min, please refer to theservice part.Sensor Alarm
indicator is on
CodeE0.5
In baby mode, If short-circuit or open-circuit or bad connection occurson the skin temperature sensor, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
In the air mode, when the incubator temperature is less than 38℃ (set temperature <37℃) or less than 40℃(set temperature >37℃), atthe same time, heater stops working, press Silence/Reset Key toclear alarming for 4min, it can not reset after clearing the failure,
press Silence/Reset Key to clear alarming
Over-Temp
Alarm indicator ison
CodeE0.6
In the Baby mode, when the incubator temperature is less than 40℃,at the same time, heater stops working, press Silence/Reset Key toclear alarming for 4min, it can not reset after clearing the failure,press Silence/Reset Key to clear alarming
Fan alarm indicatoris on
CodeE0.7
The cycle air flow way of incubator is blocked. at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min, itcan reset when the failure does not exist.
Sensor Alarmindicator is on
CodeE0.8
In baby mode, If the temperature of skin temperature sensor is always lower 2℃ than the set temperature or more, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can resetwhen the failure does not exist.
When the fan motor stops working, at the same time, heater stopsworking, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
Fan alarm indicatoris on
codeE0.9
When the wind speed is lower than 1000rpm, at the same time, heater stops working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.
In air mode, If the air temperature is higher 3℃ than the set
temperature, heater stops working, press Silence/Reset Key to clearalarming for 4min, it can reset when the failure does not exist.HIGH Deviation
Alarm indicatoris on
CodeE1.0
In baby mode, If the air temperature is higher 1℃ than the settemperature, heater stops working, press Silence/Reset Key to clear alarming for 4min, it can reset when the failure does not exist.
LOW Deviation Alarm indicator is
on
CodeE1.1
In air mode, If the air temperature is lower 3℃ than the settemperature. At this time, the heater keeps working, pressSilence/Reset Key to clear alarming for 4min, it can reset when thefailure does not exist.
8/18/2019 61981264 Service Manual YP 90 Series
30/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-8
I2
TABLE 3.1 ALARM (CONTINUE)
ALARM MESSAGE DESCRIPTION
CodeH0.1
ROM inside of main MCU fails. At this time, the heater will stop working ,
press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
CodeH0.2
Internal system inside of main MCU fails. At this time, the heater will stopworking, press Silence/Reset Key to clear alarming for 4min, please referto the service part.
CodeH0.3
Communication of main and auxiliary MCU A fails, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.
Code
H0.4
ROM inside of auxiliary MCU fails, At this time, the heater will stopworking, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
CodeH0.5
Internal system inside of auxiliary MCU fails, At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
CodeH0.6
E2ROM inside main MCU fails, At this time, the heater will stop working ,press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
CodeH0.7
SRAM failure (including system cord failure). At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
CodeH0.8
Main ADC failure(e.g., auxiliary system TLV2548, including wrong sample,chip damage, no response and so on). At this time, the heater will stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
CodeH0.9
Auxil iary ADC failure(e.g., auxil iary system TLV2544, including wrongsample, chip damage, no response and so on). At this time, the heaterwill stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service part.
CodeH1.0
Heat circuit failure(e.g., heater, solid relay, mechanical relay and allcontrol circuit fails). At this time, the heater will stop working, pressSilence/Reset Key to clear alarming for 4min, please refer to the service
part.
Code
H1.1
The timer RTC inside the system fails or is not accurate, At this time, the heater will stop working, press Silence/Reset Key to clear alarming for4min, please refer to the service part.
Code
H1.2
If the keyboard circuit is broken, At this time, the heater will stop working,press Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
The system
indicator is on
Code
H1.3
If the battery inside of the controller is error, At this time, the heater willstop working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
8/18/2019 61981264 Service Manual YP 90 Series
31/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SECTION 4CLEANING AND MAINTENANCE
4.1 GENERAL
The section provides cleaning and maintenance instructions.
WARNING
1.Disconnect all the connections with oxygen feeding device before cleaning and
maintenance. Cleaning or maintenance in the environment full with oxygen will cause
fire or explosion.
2. The incubator with cabinet adopts the connector of power cord as the isolation
mains power supply, therefore, to make sure the safety, and pull off all power cord
before cleaning.
4.2 CLEANING
This device must be cleaned and sterilized for the first time for initial use, or after used it for
one week.
4.2.1 DISASSEMBLY BEFORE CLEANING
NOTE: Please do not assemble the spare parts for regular cleaning, and it is unnecessary
to depart the base and the cabinet. If need to disassemble, please refer to the
section about assemble.
A. Remove the controller
Disconnect the Power Cord and Sensors from the side of the Incubator. Release the latch on
each side of the Controller as shown in Figure 4.1 to take out the controller.
Latch
Take out the Temperature Controller asindicated by the arrow.
WARNING: The controller can be taken out from the incubator, when the
Spanner pull out to >90° direction.
4-1I2
WARNING: The Controller heater can be sufficiently hot to cause burns;
avoid removing the Controller or touching the heater until theunit has been switched off for at least 45 minutes.
FIGURE 4.1
8/18/2019 61981264 Service Manual YP 90 Series
32/103
8/18/2019 61981264 Service Manual YP 90 Series
33/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
D. Remove Air Intake Tube Connector
See Figure 4.4, Move the Air Intake Tube out in direction of arrow 2 after disconnecting Air
Intake Tube from Air Tube Connector in direction of arrow 1 .
4-3I2
FIGURE 4.4
E. See figure 4.6, take out the Humidity baffle. See step P in section 2.3, remove 2pcs bolts
on the air filter cover, and take out the air filter.
F. Remove the Access door gasket and Access door cuff, and Iris port sleeve from hood.
G. Remove the soft tubing from hood.
4.2.2 CLEANING PROCEDURE
CAUTION: Some chemical cleaning agents may be conductive and/or leave a residue
which may permit a build-up of dust or dirt which may be conductive. Do not
permit cleaning agents to contact electrical components. Do not spray
cleaning solu tions onto any of those surfaces.
A. Clean the skin temperature sensor
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or
paper towel.
CAUTION: Do not steep Sensor in to d isinfectant-detergent.
B. Clean the Soft tubing, Main Deck Seal, Gaskets, Iris Port Sleeves, Air Intake Tube
Allow parts to soak as recommended by the cleaning solution’s manufacturer, then remove
them and dry completely with a clean cloth or paper tower.
C. Clean the Hood
Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations,
Access Doors and Iris Ports then dry them with a clean cloth or paper towel.
CAUTION: Alcohol can cause crazing of the clear Acrylic panel. Do not use alcohol,
acetone, or any organic solvents for cleaning. Do not expose the panel
assembly to direct ultraviolet radiation.
8/18/2019 61981264 Service Manual YP 90 Series
34/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
D. Clean the controller
Use detergent to clean the all surfaces thoroughly, then be especially careful when cleaning
fans, heater and surface of air temperature sensor. Then dry them with clean cloth.
CAUTION: Pay particular attention to avoid liquid seeping into the Controller Hood.
clean the surface of controller after the heater becomes cold .
WARNING: Failure to c lean the Heater and Fan Impeller could result in sufficient lint build -
up to reduce airflow, which will affect temperature control and cause high CO2
concentration.
E. Clean the bassinet, Main Deck, shelf
Use a disinfectant-detergent to thoroughly clean all surfaces then dry them with a clean cloth
or paper towel.
CAUTION: Do not lubricate the Mattress Tilt Mechanism with oil or other potentially
F. Clean Mattress Cover
Use a disinfectant-detergent to thoroughly clean it and sun it after washing it with clean water.
G. Clean Air filter
Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations,
Access Doors and Iris Ports then dry them with a clean cloth or paper towel. If the air filter is dirt
or is used for more than 2 months, please replace it.
H. Clean Water tank and Humidity chamber fill spout
Fill the humidity chamber with a disinfectant-detergent, completely cleaning the water
chamber and cup inside and outside. (see figure 5.5)Drain the humidity chamber, scrub it
thoroughly, including all indentations, then try the chamber and water cup with a clean cloth or
paper towel. After scrubbing, continuous rotating water cup per counter-clockwise until it take out
for separate scrubbing.
FIGURE 4.5
4-4I2
8/18/2019 61981264 Service Manual YP 90 Series
35/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
I. Clean NEONATE BILIRUBIN OHOTOTHERAPY EQUIPMENT
See the user’s manual for XHZ-90 and XZH-90L.
J. Clean surface of device
Use a disinfectant-detergent to thoroughly clean all surfaces, including all holes, indentations,
Access Doors and Iris Ports then dry them with a clean cloth or paper towel.
4.2.3 REASSEMBLY AFTER CLEANING
NOTE: Before reassembling parts into Incubator, check carefully if there is any crack or
damage. Some disinfectant-detergent may cause crazing of plastic parts in
Incubator.
A. Install Buffer, Humidity chamber fill spout, Air Filter
Place the Buffer into Trough as described in Figure 4.6 in correct direction. Humidity
chamber fill spout is rotated clockwise to its original position and placed into the Air Filter Cover.
All the screws should be tightened.
4-5I2
Correct Position Wrong Position
FIGURE 4.6
B. The Air Intake Tube should be installed by reversing the procedure , Installed Air Intake
Tube must comply to Figure 4.7.
NOTE: Air Intake Tube must be placed inside the Air Intake Tube Connector and
completely sealed.
FIGURE 4.7
8/18/2019 61981264 Service Manual YP 90 Series
36/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
C. Place Main Deck back and install the seal on it.
WARNING: Be sure the two thumb screws that hold the bassinet are tigh tened. The
right side and the wrong side of seal must be placed in the right
position, and make sure the airproof to avoid the inepuality of heating
towards the bed.
D. Install Controller back according the oposite order , see figure 4.1.
Controller is the key unit of Incubator and must be installed carefully.
Insert Controller into Base Assembly. After the whole part is inserted into Base Assembly,
depress the snaps at both sides of Controller until they are parallel to the side then pull the
Controller outward. If the Controller is not loosened, it is already locked. (See Figure 4.8)
CORRECT INSTALLATION STATE
FIGURE 4.8
E. Rotating the iris entry sleeves , see figure 4.9.
Step 1: Install the elastic band over the inner ring of the port housing. Fold back and slip
elastic band over the outer ring of the port housing.
4-6I2
8/18/2019 61981264 Service Manual YP 90 Series
37/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Step 2: Rotate outer ring to close. If properly installed, the sleeve will open again if rotation is
reversed.
FIGURE 4.9
F. See figure 4.10, install the Access door gasket, and then install the Access door cuff.
access door gasket, inner wall of hood, Access door wall, outside wall of hood, installation groove
of Access door cuff.
Access door wall
Inner wall of hood
4-7I2
Access door gasket
Outside wall of hood
Installation groove of Access door cuff
FIGURE 4.10
G. Install the soft tubing in figure 4.11.
NOTE: The opening of soft tubing should face outward.
FIGURE 4.11
8/18/2019 61981264 Service Manual YP 90 Series
38/103
4-8I2
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
H. Install the Mattress cover with the mattress, and put the bassinet inside hood according to
the oposite order. (see Figure 4.2)
WARNING: Place the bassinet as the label ind icates on the infant bed.
4.3 STERILIZATION
NOTE: Please do not steam autoclave.
Please sterilize under low temperature, e.g., gas sterilization.
Clean the incubator before gas sterilization.
WARNING: Please remove the control ler when do gas s terilization, or else, it wi ll cause
danger.
After gas sterilization, there will be ventilation of 16 ~ 24hrs, and then put back the controller.
During the whole ventilation, the incubator shoule work under the dry environment of 32℃~ 35℃.
After ventilation, please cover the dustproof cover if do not use the infant incubator.
CAUTION: The temperature of sterilization should not exceed 54.5 .
IMPORTANT: Please check all parts prior to use.
4.4 MAINTENANCE
Please check the condition of the build-in rechargeable battery before the first use of device
or in the alternation of device using.
A. Operate the unit for a period of 12 to 24 hours.
B. Trigger a power failure alarm by disconnecting the AC power cord.
C. The power failure alarm should activate and continue to alarm for at least 10 minutes.
D. Reconnect the unit to the AC line and recharge the battery.
If the power failure alarm cannot last more than 10 minutes, please replace the rechargeable
battery. For the replacement of battery, please refer to section 5.
8/18/2019 61981264 Service Manual YP 90 Series
39/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SECTION 5SERVICE
5.1 GENERAL
This section provides system setting procedure and the troubleshooting procedure of these parts:
replacement of Fuse, Internal charging Battery and Controller.
5.2 REPLACEMENT OF ELEMENT
5.2.1 REPLACEMENT OF FUSE
Open Temperature Controller, replace the same type fuse with same specification accorind to the
label on the circuit board.
NOTE: When replacing the fuse, switch off the power supp ly and disconnect the power cord.
5.2.2 REPLACEMENT OF INTERNAL BATTERY
See figure 5.1, Open Controller, and disonnect the battery fastener and the internal charing battery,
and take out the baffle to replace the battery.
5-1I2
Battery Fastene Internal Charing Battery Baffle
FIGURE 5.1
NOTICE: 1.During replacing the internal battery, the plastic washer should be placed the orig inal
position, and keep correct connection between the anode/ cathode of battery and its
fastener;the faulty connection will cause the damage to the battery and no hont for
power failure alarm.
2.Please replace theb internal battery after using it for three years.
3. Battery will be automatically charged when the temperature control ler works,
charging the battery outside of the incubator is fo rbidden.
4.The charging battery inside the wartmer is GP17R8H Ni-MH8.4V/170mAh, to ensure
safety, do not d isassemble battery. Connect the battery correctly, do not use other type
battery.
5.It will not influence battery charging and the battery capacity when the battery is full.
(Switch on controller)
6. Please do not th row the old battery into f ire or water to avoid explos ion or l eakage.7.Please dispose the Old battery accoring to local law, and please do not throw them
away randomly.
8/18/2019 61981264 Service Manual YP 90 Series
40/103
8/18/2019 61981264 Service Manual YP 90 Series
41/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.3.3 PROCEDURE
At first, connected power supply, press and hold Set key for 2s as soon as switch on until Ai r
temprature indicator displays “ --.-“, and no display on Skin and Set temprature indicator. And then,
press Up or Down until Ai r temprature indicator displays “78.6”, after then, press Set key to enter into
the system set screen, or press Silence/Reset key to exit.
After entering into the set screen, Set temprature indicator displays set code, Ai r temprature
indicator displays the specific set data.
Turn on the power switch
5-3I2
Press and hold the for 3s
Enter password
Press key, EXIT Press or key to enter password 78.6,
Then press key to confirm
Error Password check
YES
System setting screen
Press key or Key, set temperature indicator displays
the set code
Press or Key can change
current set value
END
Press key Press and together
SAVE and EXIT Return to original set
Brightness set of indicator
Figure 5.2
8/18/2019 61981264 Service Manual YP 90 Series
42/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-4I2
CODE 00.1 LED Lightness set
Indicate the lightness , and set range is 00.1~01.5. The larger the lighter.
CODE 00.2 Air temperature sensor compensation
Set value of temperature compensation for air temperature sensor , and set range is -5.0~05.0 ,the first figure of set value means the increasing and decreasing value. For Example, 00.2℃ means
that the air temperature adds 0.2 as℃ compensation. -1.3℃ means that the air temperature reduces
1.3 as℃ compensation.
CODE 00.3 Air flow temperature sensor compensation
temperature compensation for air flow temperature sensor , and set range is -5.0~05.0 , the
first figure of set value means the increasing and decreasing value. For Example, 00.2℃ means that
the air flow temperature adds 0.2 as℃ compensation. -1.3℃ means that the air flow temperature
reduces 1.3 as℃ compensation.
CODE 00.4 Skin temperature sensor compensation
Set value of temperature compensation for skin temperature sensor , and set range is
-5.0~05.0 , the first figure of set value means the increasing and decreasing value. For Example,
00.2℃ means that the skin temperature adds 0.2 as℃ compensation. -1.3℃ means that the skin
temperature reduces 1.3 as℃ compensation.
CODE 00.5 Machine Number
Set method of machine number, and set range is 00.0~02.0. For example, 00.1 measn No. 1
machine, 01.5 means No. 15.
CODE 00.6 Motor speed alarm value
Set method of fan speed, and set range is 00.0~25.5. Every 0.1 means 20 rpm. For example,
05.0 means 1000 rpm. Alarm will be active if the fan speed is lower than this value.
CODE 00.7 Set value of Low deviation for air temperature
Set method of low deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .℃
CODE 00.8 Set value of High deviation for air temperature
Set method of high deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .℃
CODE 00.9 Set value of low deviation for skin temperature
Set method of low deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .℃
CODE 01.0 Set value of high deviation for skin temperature
Set method of high deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .℃
8/18/2019 61981264 Service Manual YP 90 Series
43/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-5I2
CODE 01.1 Minimum set value of air temperature
Set method of Minimum set value of air temperature, and set range is 20.0~34.0. For example,
25.0 means Minimum set value 25.0 .℃
CODE 01. 2 Maximum set value of air temperature
Set method of Maximum set value of air temperature in mode of>37.0℃ , and set range is
37.0~39.0. For example, 39.0 means Maximum set value 39.0 .℃
CODE 01.3 Minimum set value of skin temperature
Set method of Minimum set value of skin temperature, and set range is 30.0~34.0. For example,
34.0 means Minimum set value 25.0 .℃
CODE 01.4 Maximum set value of skin temperature
Set method of Maximum set value of skin temperature in mode of>37.0℃ , and set range is
37.0~39.0. For example, 39.0 means Maximum set value 39.0 .℃
CODE 01.5 Set value of Over temperature alarm A (<37℃)
Set method of Over temperature alarm in air mode (set temperatuer <37℃), and set range is
37.5~39.0. For example, 38.0 means over-temp alarm value 38.0 .℃
CODE 01.6 Set value of Over temperature alarm B (>37℃)
Set method of Over temperature alarm in air mode (set temperatuer >37℃), and set range is
38.5~40.0. For example, 40.0 means over-temp alarm value 40.0 .℃
CODE 01.7 Set value of air flow Over temperature alarm A (<37℃)
Set method of alarm value for air flow blocked (set temperatuer <37℃), and set range is
38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0℃
CODE 01.8 Set value of air flow Over temperature alarm B (>37℃)
Set method of alarm value for air flow blocked (set temperatuer >37℃), and set range is 40.5~52.0.
For example, 45.0 means alarm value for air flow blocked 45.0℃
8/18/2019 61981264 Service Manual YP 90 Series
44/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.4 TROUBLESHOOTING PROCEDURES
5.4.1 GENERAL
Troubleshooting guides for the Controller are provided in Paragraph 5.4.3 and Paragraph 5.4.4.
Paragraph 5.4.3 provides the alarm code when the controller alarms. Paragraph 5.4.4 provides
troubleshooting in the form of flowcharts.
It is confi rmed that the system settings are correct before troubleshooting.
5.4.2 TEST EQUIPMENT REQUIRED
The test equipment listed below is required for troubleshooting the Controller. Equivalent test
equipment may be substituted.
Digital VOM, FLUKE 8060A
31.0℃ Air Temperature Simulation
5-6I2
Resistor
Resistor R=23166Ω/0.25W/0.5%
31.0℃ Air flow Temperature Simulation
Resistor
Resistor R=23166Ω/0.25W/0.5%
31.0℃ skin Temperature Simulation
Resistor R=23166Ω/0.25W/0.5%
5.4.3 TROUBLESHOOTING
All alarms have the relevant alarm code except for the power ailure alarm, please refer to Table 5.1.
TABLE 5.1 ERROR CODES
ALARM CODE TYPEEXPLANATION AND
CORRECTIVE ACTION
E0.1 Air temperature sensor failure alarm Refer to Flow chart 5.1
E0.2 Isolated temperature sensor alarm Refer to Flow chart 5.2
E0.3 Air temperature deviation failure alarm Refer to Flow chart 5.3
Resistor
8/18/2019 61981264 Service Manual YP 90 Series
45/103
SERVICE MANUAL FOR I FANT INCUBATOR EDITION/REVISION B/0 N
5-7I2
TABLE 5.1 ERROR CODES (Continued)
ALARM CODE TYPEEXPLANATION AND
CORRECTIVE ACTION
E0.4 Air flow sensor alarm Refer to Flow chart 5.4
E0.5 Skin temperature sensor failure alarm Refer to Flow chart 5.5
E0.6 Over temperature failure alarm Refer to Flow chart 5.6
E0.7 Air circulation failure alarm Refer to Flow chart 5.7
E0.8 Wrong position of skin sensor alarm Check the position of skin sensor
E0.9 Fan motor stop alarm Refer to Flow chart 5.8
E1.0 High deviation temperature failure alarmCheck whether there is heat sourcenearby, if yes, please leave far away
from the heat source.
E1.1 Low deviation temperature failure alarmCheck the close condition of all doorsand panels
H0.1 ROM failure alarm inside of Main MCU Replace Main MCU
H0.2Internal system failure alarm inside ofMain MCU
Replace Main MCU
H0.3 communication failure alarm
1. Check the connection of circuitbetween the main MUC and auxiliaryMCU2. Replace main MUC and auxiliaryMCU
H0.4 ROM failure alarm inside auxiliary MCU Replace auxiliary MCU
H0.5Internal system failure alarm insideauxiliary MCU
Replace auxiliary MCU
H0.6 E2PROM failure alarm Replace Main MCU
H0.7 SRAM failure alarm Replace U13 on mian board
H0.8 Main ADC failure alarm Refer to Flow chart 5.9
H0.9 Auxiliary ADC failure alarm Refer to Flow chart 5.10
H1.0 Heat system failure alarm Refer to Flow chart 5.11
H1.1 Real timer RTC failure alarm Replace U20 on mian board
H1.2 Keyboard fault alarmWhen the key board of the control panelon controller short-circuit alarms, pleasereplace the control panel.
H1.3 Battery failure alarm
Inner storage battery failure alarm:1. Check the wire connection betweenthe inner storage battery and the powerboard.
2. Replace the inner storage battery.
Power failure alarm Refer to Flow chart 5.12
8/18/2019 61981264 Service Manual YP 90 Series
46/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.4.4 TROUBLESHOOTING FLOW CHART
Before troubleshooting, forst, please switch off, and disconnect the power cable, and then open the
top cover of warmer module and the back cover of column. And then connect the power cable and turn
onthe switch, at the same time, make sure that power +5V, +12V of controller, element, and all cables
are in good condition.In working state, press Up Key and Air mode Key , you can find the isolated
temperature and air flow temperarture in the air and set temperature indicator respectively. So it is easy
for service man to find the cause of trouble.
AIR TEMPERATURE SENSOR FAILURE ALARM E0.1
``
no
Good conection? Check connection of air temperature sensor,
and the wiring of all circuit board
The 31.0℃±0.2℃ Air Temperature Simulation (decade
Equal to 23166Ω±182Ω)is instead of the air
temperature sensor, turn on power supply
the temperature display is
31.0℃ and add the no
compensation temperature alarm E0.1? Replace the air temperature sensor
of system set procedure
yes
on the main board U11-6 is 2.3V? no check the connect to R21, U1-6
Yes
CHECK U11
FLOW CHART 5.1 AIR TEMPERATURE SENSOR FAILURE
5-8I2
8/18/2019 61981264 Service Manual YP 90 Series
47/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
ISOLATED TEMPERATURE SENSOR FAILURE ALARM E0.2
``
no
Good conection? Check connection of isolated temperature
sensor, and the wiring of all circuit board
The 31.0℃±0.2℃ Air Temperature Simulation (decade
Equal to 23166Ω±182Ω)is instead of the isolated
temperature sensor, turn on power supply
See 5.4.4, the isolated no
temperature is 31.0℃ alarm E0.2? Replace the isolated temperature sensor
yes
no
on the main board U21-6 is 2.3V? check the connect to R16, U21-6
Yes
Check U21
FLOW CHART 5.2 ISOLATED TEMPERATURE SENSOR FAILURE
5-9I2
8/18/2019 61981264 Service Manual YP 90 Series
48/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Air temperature deviat ion failure alarm E0.3
``
no
Good conection? Check connection of isolated temperature
sensor, and the wiring of all circuit board
The 31.0℃±0.2℃ Air Temperature Simulation (decade
Equal to 23166Ω±182Ω)is instead of the air and isolated
temperature sensor, turn on power supply
See 5.4.4, check whether
The air temeprature and no
isolated tempertire is alarm E0.3? Replace sensor
added the temperature
compensation based on 31.0℃
yes
U11-6, no
on the main board U21-6 is 2.3V? check the connect to R16 and U21-6,
R21 and U11-6
Yes
Replace U11 or U21
FLOW CHART 5.3 AIR TEMPERATURE DEVIATION FAILURE
5-10I2
8/18/2019 61981264 Service Manual YP 90 Series
49/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Air f low sensor fai lure alarm E0.4
``
no
Good conection? Check connection of sensor,and the wiring of all
circuit board
The 31.0℃±0.2℃ Air Temperature Simulation (decade
Equal to 23166Ω±182Ω)is instead of air flow
temperature sensor, turn on power supply
See 5.4.4, check whether
The air temeprature and no
isolated tempertire is alarm E0.4? Replace air flow sensor
added the temperature
compensation based on 31.0℃
yes
no
on the main board U11-8 is 2.3V? check the connect to R23 and U11-8
Replace U11
FLOW CHART 5.4 AIR FLOW SENSOR FAILURE
5-11I2
8/18/2019 61981264 Service Manual YP 90 Series
50/103
8/18/2019 61981264 Service Manual YP 90 Series
51/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Over-temp failure E0.6
yes
Is there heat source nearby? Far away from the heat source
5-13I2
no
yes
Air and isolated temperature Check air and isolated temperature sensor
Sensor is OK?
no
Replace air and isolated temperature sensor
FLOW CHART 5.6 OVER TEMPERATURE FAILURE
Check whether the inlet
Or outlet is blocked
Air c irculat ion failure alarm E0.7
Check the air flow condition of inlet and outlet
FLOW CHART 5.7 AIR CIRCULAATION FAILURE
8/18/2019 61981264 Service Manual YP 90 Series
52/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Fan Motor Failure Alarm E0.9
Motor rotating no Motor input terminal no Check the power cord
Is AC power?
yes
yes
on the power Pulse at no Check the Capacitor no
supply board U5-2? fan detected CM is good Replace the capacitor CM
board
yes yes
5-14I2
no Replace the Motor
on the power Pulse at Check U5
supply board U5-3
yes
on the main no
board Pulse U12-16? check the connection between the main board
and the power supply board
yes
Check U12
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM( SHEET 1 OF 2)
8/18/2019 61981264 Service Manual YP 90 Series
53/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Slow Motor Speed Alarm E0.9
yes
Motor input no
Voltage is ok? Check the power input
yes
no
Capacitor Replace the capacitor CM
CM is good
yes
Replace the motor
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM ( SHEET 2 OF 2)
5-15I2
8/18/2019 61981264 Service Manual YP 90 Series
54/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Connection of U11
and U12 is OK?
Check the jumper wire of
U11-13, R5, R6
noPin of U11-13 is
+2.5V?
no
Check R5, R6Crosspoint of R5
and R6 is +2.5V?
Main ADC failure alarm H0.8
On main board
On main board
Check the connecting wireOn main board
Replace U11
FLOW CHART 5.9 MAIN ADC FAILURE ALARM
5-16I2
8/18/2019 61981264 Service Manual YP 90 Series
55/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-17I2
Check the connecting wireConnection of U21
and U18 is OK?
Check the jumper wire of
U21-9, R14, R15
noPin of U21-9 is
+2.5V?
no
Check R14, R15Crosspoint of R14
and R15 is +2.5V?
Auxi liary ADC fai lure alarm H0.9
On main board
On main board
On main board
Replace U21
FLOW CHART 5.10 AUXILIARY ADC FAILURE ALARM
8/18/2019 61981264 Service Manual YP 90 Series
56/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Heat circuit failure alarm H1.0
Disconnet power supply, measure resistance of two
terminals for heater
5-18I2
yes
Replace heater
no
Heater is open circuit or
short circuit?
Conneat power board and main board
Element number from here is on
power board
Connect pin 1 (ssr-ctr) and pin 3
(relay-ctr) on J8 to VCC power make two
conrol signals high level
no
Base of triode Q2, Q3
is >0.7V?
Check triode Q2, Q3
and resistor R8,R9
yes
nothere +12V on solid
relay and control
terminal of relay K1
Replace triodeQ2 or Q3
yes
To sheet 2
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 1 OF 4)
8/18/2019 61981264 Service Manual YP 90 Series
57/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 1
Output terminal of solid
relay and relay K1 can
work?
no Replace solid relay
or relay K1
yes
Voltage of U1-2 is
> 5mV
yes
no
Check inductor T1
Base of triode Q1
>0V?
yes
noBase of triode Q1
>0.7V?
yes
no
Voltage of U1-1
is >2V?
Check U1,R1 ,C4
Check R2,R6,C5
no
Repalce triode Q1
Connet pin 1 (ssr-ctr) and pin3 (relay-ctr) on J8 to GND
make two control signals high level
yes
To sheet 3
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 2 OF 4)
5-19I2
8/18/2019 61981264 Service Manual YP 90 Series
58/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 2
Base of triode Q1,Q3is 0V?
yes
Check triode Q2,Q3
and resistor R8, R9
no
Base of triode Q1
is 5V?
yes
noBase of triode Q1
is 0V?
Yes
no
Voltage U1-1 is
0V?
Yes
no
Voltage U1-2 is
0V?
no
yesIs there +12V on
control terminal of solid
relay and relay K1
Replace triode
Q2 or Q3
Check inductor T1
Check U1,R1,C4
Check R2, R6, C5
no
Replace triode Q1
yes
To sheet 4
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 3 OF 4)
5-20I2
8/18/2019 61981264 Service Manual YP 90 Series
59/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 3
The pins on porer
board J8-1, 3, 4 and
J2-1, 3, 4 on main
board works?
Switch off during measuring
no
Replace connecting wire
yes
Connect power board and main board
Connect pins U8-2, 11 to VCC power
Element number from here is on
mainboard
Yes
Pin of U8-9, 18 is
5V?
no
Check U8
Measure the voltage of U15-2 for self-checking
yes
no
Replace U15
yes yes
Replace main board
Still alarming after
replacing U15
Voltage is changingfrom low level to highlevel?
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 4 OF 4)
5-21I2
8/18/2019 61981264 Service Manual YP 90 Series
60/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Power Failure Alarm
5-22I2
Is there AC power on
input socket?Check outside power input
No
Yse
Initial trans for mer is
AC power?
No Check the connecting wire of
trans for mer and ower cord,
and Fuse F1, F2, F3 on AC
entry boardYes
No
Pin 3 on U2, U3 is +5V?
No
Cross point of R17,R18 is +2.5V?
No
Check U2, U3 and prefix circuit
Yes
On power board
Pin4 of U4 is
+12V?
Check U4 and prefix circuitOn power board
Yes
Check R17, R18
Yes
Secondary output of transformer is 9V, 15V? Replace transformer
On power board
Yes
To sheet 2
FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 1 OF 2)
8/18/2019 61981264 Service Manual YP 90 Series
61/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 1
NoCrosspoint of
R19, R20 is 2.5V?
On power board Replace R19, R20
Yes
Crosspoint of
R21, R22 is
2.1V?
No
Replace R21, R22On power board
5-23I2
Yes
Pins7, 1 of U6
is high level?
No
On power board Replace U6
Yes
NoPin5 of U8 is low
level?Check the connection of power board
and main board
On main board
Yes
Pin4 of U8 islow level?
No
Replace U8On main board
Yes
Replace main board
FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 2 OF 2)
8/18/2019 61981264 Service Manual YP 90 Series
62/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SECTION 6REPLACEMENT PARTS
6.1 GENERAL
This section provides parts lists for infant incubator. Part numbers of accessories are provided
below:
6-1I2
05
03
04
02
01
FIGURE 6.1 PARTS LOCATION DIAGRAM OF WHOLE UNIT
TABLE 6.1 REPLACEMENT PARTS AND ACCESSORIES LIST
ITEM No. DESCRIPTION PART NUMBER
1 Basinet 01 000 00
2 Controller 02 000 00
3 Hood 03 000 00
4 Base 04 000 00
5 I. V. Pole Groupware 05 000 00
8/18/2019 61981264 Service Manual YP 90 Series
63/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
FIGURE 6.2 PARTS LOCATION DIAGRAM, HOOD (SHEET 1 OF 2)
6-2I2
8/18/2019 61981264 Service Manual YP 90 Series
64/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-3I2
FIGURE 6.2 PARTS LOCATION DIAGRAM, HOOD (SHEET 2 OF 2)
8/18/2019 61981264 Service Manual YP 90 Series
65/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-4I2
TABLE 6.2 REPLACEMENT PARTS AND ACCESSORIES LIST
ITEM No. DESCRIPTION PART NUMBER
1 Outlet port 03 037 00
2 Bumper 03 013 00
3 Hood 03 038 00
4 Direction label 03 012 00
5 Access door 03 010 00
6 Access panel 03 039 00
7 Latch pivot 03 040 00
8 Cross groove pan head bolt M3×25
See No. of
standard Parts
9 Long chain 03 041 00
10 split baffle ring 4See No. of
standard Parts
11 Flat washer 6See No. of
standard Parts
12 Long hinge for access door 03 042 00
13 Short hinge for access door 03 043 00
14 Soft tubing port 03 011 00
15 Right hinge 03 044 00
16 Cover plate 03 006 00
17 Cross groove countersunk head bolt M5×16See No. of
standard Parts
18 Compound nut M5See No. of
standard Parts
19 No.7 stoppert 03 045 00
20 Inlet port 03 046 00
21 Left hinge 03 047 00
22 Fixing hook 03 001 00
23 Cross groove pan head bolt M 6×12See No. of
standard Parts
24 Spring washer 6See No. of
standard Parts
25 Cross groove oblate head bolt M4×12See No. of
standard Parts
26 Pawl latch for access door 03 030 00
27 Sheath for pawl latch 03 031 00
8/18/2019 61981264 Service Manual YP 90 Series
66/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-5I2
TABLE 6.2 REPLACEMENT PARTS AND ACCESSORIES LIST
ITEM No. DESCRIPTION PART NUMBER
28 Spring(IV) 03 034 00
29 Hinge pivot 03 020 00
30 Cross groove pan head bolt M3×10See No. of
standard Parts
31 Bolt cover M3 03 051 00
32 Bolt cover M5 03 052 00
33 Torsion spring 03 021 00
34 Access door gasket 03 032 00
35 Access door cuff, reusable 03 033 00
36 O-ring Φ8.5×1.8See No. of
standard Parts
37 Top pin 03 026 00
38 Rolling lock bush 03 028 00
39 Spring (I) 03 025 00
40 Latch for access door 03 024 00
41 Column pin 3×30See No. of
standard Parts
42 Cross groove pan head bolt M 5×10See No. of
standard Parts
43 Flat washer 5See No. of
standard Parts
44 Spring washer 5See No. of
standard Parts
45 Iris port sleeve, reusable 03 027 00
46 Ring asembly 03 017 00
47 Soft retaining ring 03 018 00
48 Rotatable ring 03 019 00
49 Cross groove countersunk head bolt M6×20See No. of
standard Parts
50 Washer 03 003 00
51 Cross groove pan head tapping bolt ST2.9×9.5See No. of
standard Parts
52 White elastic part for Pawl Latch 03 035 00
8/18/2019 61981264 Service Manual YP 90 Series
67/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-6I2
FIGURE 6.3 PARTS LOCATION DIAGRAM, BASE (SHEET 1 OF 2)
8/18/2019 61981264 Service Manual YP 90 Series
68/103
8/18/2019 61981264 Service Manual YP 90 Series
69/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-8I2
TABLE 6.3 REPLACEMENT PARTS AND ACCESSORIES LIST
ITEM No. DESCRIPTION PART NUMBER
1 Base outer shell 04 001 00
2 Lock base 04 002 00
3 Cross groove countersunk head bolt M3×8See No. of
standard Parts
4 Humidity chamber fill spout 04 004 00
5 Label for Humidity chamber fill spout 04 003 00
6 O-ring Φ16×2.65
See No. of
standard Parts
7 Cap of fill spout for Humidity chamber 04 005 00
8 Cross groove pan head bolt M4×12See No. of
standard Parts
9 Flat washer 4See No. of
standard Parts
10 Small pad 04 006 00
11 Limit panel 04 007 00
12 Fixing bolt(II) 04 008 00
13 Bush (I) 04 009 00
14 Seal 04 010 00
15 Humidity baffle 04 011 00
16 Humidity chamber 04 012 00
17 Bassinet 04 018 00
18 Pull handle 04 014 00
19 Washer for pull handle 04 015 00
20 Flat washer 8See No. of
standard Parts
21 Spring washer 8See No. of
standard Parts
22 Hexagon nut M8See No. of
standard Parts
8/18/2019 61981264 Service Manual YP 90 Series
70/103
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
6-9I2
TABLE 6.3 REPLACEMENT PARTS AND ACCESSORIES LIST
ITEM No. DESCRIPTION PART N