+ All Categories
Home > Documents > A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

Date post: 24-Jan-2022
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
89
1 A COMPARISON OF AUTOMATED AUDITORY BRAINSTEM RESPONSE (AABR) WITH INTEGRATED ELECTRODES AND OTOACOUSTIC EMISSIONS (OAEs) IN HIGH RISK NEWBORN HEARING SCREENING DR. MUHAMMAD NOOR AZMI BIN RAMLI DISSERTATION SUBMITTED IN PARTIAL FULFILMENT OF THE REQUIREMENTS FOR THE DEGREE OF MASTER OF MEDICINE (PAEDIATRIC) UNIVERSITI SAINS MALAYSIA 2017
Transcript
Page 1: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

1

A COMPARISON OF AUTOMATED AUDITORY

BRAINSTEM RESPONSE (AABR) WITH INTEGRATED

ELECTRODES AND OTOACOUSTIC EMISSIONS (OAEs)

IN HIGH RISK NEWBORN HEARING SCREENING

DR. MUHAMMAD NOOR AZMI BIN RAMLI

DISSERTATION SUBMITTED IN PARTIAL

FULFILMENT OF THE REQUIREMENTS FOR

THE DEGREE OF MASTER OF MEDICINE

(PAEDIATRIC)

UNIVERSITI SAINS MALAYSIA

2017

Page 2: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

2

ACKNOWLEDGEMENT

Alhamdulillah, praise be to ALLAH SWT for giving me the ability, passion and strength to

eventually complete this dissertation. The preparation of this important document would not

have been possible without the support, hard work and endless assistance of a large number of

individuals and institutions.

I would like to extend my heartiest gratitude to my supervisor Associate Professor Dr Salmi

binti Ab Razak, my co-supervisor Professor Dr Hans Rosternberghe, Dr Mohamad Normani

bin Zakaria and Dr Siti Azrin binti Ab Hamid for their advice, assistance and support. Many

thanks to all wonderful people who had involved and helped me in this study namely, all staff

in paediatric ward Nilam 2 and 1 Timur Belakang (1TB). Last but not least, to my dear mother,

Puan Hajjah Che Rohani binti Ibrahim, my wife Puan Sumaiyah binti Abdullah, my childrens,

Lu’ay and Alaa’ and all family members for their constant prayers, love and encouragement in

all my endeavours.

“Anything is possible to a willing heart.”

Page 3: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

3

TABLE OF CONTENTS

Page

Acknowledgement ii

Table of Contents iii

List of Tables vi

List of Figures vii

Abbreviations viii

Abstract x

Bahasa Malaysia x

English xiii

1 INTRODUCTION 01

2 LITERATURE REVIEW 05

2.1 Universal Neonatal Hearing Screening 06

2.2 Neonatal Hearing Screening Among High Risk Newborn 07

2.3 Screening Test and Equipment 08

2.3.1. Integrated Electrodes Otoacoustic Emissions 08

2.3.2. MB11 BERAphone 10

2.3.3. Auditory Brainstem Response 12

2.4 The Comparison Between OAEs and AABR 13

2.5 The Comparison Between OAEs and MB11 BERAphone 14

2.6 The Agreement Off OAEs And MB11 BERAphone 16

2.7 The Total Time Spent Between OAEs And MB11

BERAphone

17

Page 4: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

4

3

OBJECTIVES OF THE STUDY

3.1 General Objective 19

3.2 Specific Objectives 19

3.3 Research Hypothesis 20

4 METHODOLOGY 21

4.1 Research Design 22

4.2 Participants 22

4.2.1 Inclusion criteria

4.2.2 Exclusion criteria

22

23

4.3 Intervention 23

4.4 Outcomes 24

4.5 Sample size 25

4.6 Ethical approval 27

4.7 Statistical Analysis 27

4.8 Definitions 28

4.9 Study of Flow Chart 29

5

RESULTS

30

5.1 Sample Characteristic 31

5.2 Demographic Data 33

5.2.1 Demographic Characteristic 33

5.2.2 Ethnic Distribution 34

5.3 Risk Factors Among Newborn 35

Page 5: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

5

5.4 The Comparison Between OAEs and MB11 BERAphone 36

5.5 The “Refer Rates” of MB11 BERAphone and ABR 37

5.6 The “Refer Rates” of OAEs and ABR 38

5.7 The Agreement Between OAEs And MB11 BERAphone 39

5.8 The Total Time Spare Among Staff Between OAEs and

MB11 BERAphone

40

6

DISCUSSION

41

6.1 Demographic Characteristic and Risk Factor in Newborn 43

6.2 The Comparison Result Between Two Methods 44

6.3 The Refer Rate Of OAEs And MB11 BERAphone with

Screening Using ABR

46

6.4 The Agreement Between Two Methods 49

6.5 The Testing Time 50

6.6 Limitations Of Study 53

7 CONCLUSSION 54

8 RECOMMENDATIONS 56

9 REFERENCES 57

10 APPENDICES 62

Appendix 1 Conceptual Framework 63

Appendix 2 Data Collection Form 64

Appendix 3 Consent Form 65

Appendix 4 Ethics Approval Letter 73

Appendix 5 Agreement Letter for MB11 BERAphone 74

Appendix 6 Gantt Chart of Study 75

Page 6: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

6

LIST OF TABLES

TABLE 5.1 Demographic characteristic 32

TABLE 5.2 Risk factors among newborn 34

TABLE 5.3 The contingency of findings using MB11 BERAphone and OAEs 35

TABLE 5.4 The true and false positive rate for hearing loss using OAEs and

ABR 36

TABLE 5.5 The true and false positive rate for hearing loss using MB11

BERAphone and ABR 37

TABLE 5.6 The agreement between passing and referral rate between MB11

BERAphone and OAEs by using Kappa statistics 38

TABLE 5.7 The comparison of total staff time spent for screening programs

between Integrated Electrodes Otoacoustic Emissions (OAEs) and

MB11 BERAphone 39

Page 7: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

7

LIST OF FIGURES

FIGURE 2.1 Otoacoustic Emissions 9

FIGURE 2.2 Pass Result 10

FIGURE 2.3 Refer Result 11

FIGURE 5.1 Sample Characteristic 31

FIGURE 5.2 Ethnic Distribution 33

Page 8: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

8

ABBREVIATIONS

ABR Auditory Brainstem Response

AABR Automated Auditory Brainstem Response

AD Auditory Dyssynchrony

APGAR Appearance, Pulse, Grimace, Activity, Respiration

AN Auditory Neuropathy

DB Decibel

DP Distortion Product

DPOAEs Distortion Product Otoacoustic Emissions

ORL Otorhinolaryngology

G Gram

HREC Human Research Ethics Committee

HUSM Hospital Universiti Sains Malaysia

IQR Inter Quartile Range

JCIH Joint Committee on Infant Hearing

NHS Neonatal Hearing Screening

NICU Neonatal Intensive Care Units

OAEs Otoacoustic Emissions

Page 9: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

9

PASW Predictive Analytics Software

PC Personal Computer

UNHS Universal Neonatal Hearing Screening

SPSS Statistical Package of Social Science

SNHL Bilateral Sensorineural Hearing Loss (SNHL)

TEOAEs Transient Evoked Otoacoustic Emissions

TORCH Toxoplasma, Others (syphilis), Rubella, Cytomegalovirus, Herpes

Page 10: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

10

ABSTRAK

TAJUK

Perbandingan Antara “Automated Auditory Brainstem Response (AABR)” dan “Integrated

Electrodes and Otoacoustic Emissions (OAEs)” Dalam Saringan Pendengaran Di Kalangan

Bayi Yang Berisiko Tinggi di Hospital Universiti Sains Malaysia (HUSM)

OBJEKTIF

Untuk membandingkan keputusan saringan pendengaran menggunakan AABR ( melalui mesin

MB11 BERAphone) dan “Integrated Electrodes dan Otoacoustic Emissions (OAEs)” di

kalangan bayi yang berisiko tinggi dalam HUSM

TATACARA

Saringan pendengaran ini dilakukan dalam kajian keratan rentas. Seramai 195 bayi yang

berisiko tinggi memenuhi kriteria- kriteria yang telah ditetapkan terlibat di dalam kajian ini.

Bayi yang berisiko tinggi dalam masalah pendengaran diuji menggunakan mesin OAEs dan

diikuti dengan mesin MB11 BERAphone di wad yang sama dan dilakukan setelah pesakit yang

telah dirawat di wad disahkan sihat dan boleh didiscaj dari hospital. Kedua – dua mesin ini

akan menghasilkan signal “lulus” atau “ rujuk” sebagai keputusan. Keputusan yang telah

dikeluarkan tidak memerlukan sebarang kemahiran yang canggih untuk dianalisa.

Page 11: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

11

Keputusan yang dibuat oleh kedua – dua mesin ini sebagai “ rujuk” akan disaringkan sekali

lagi dengan menggunakan mesin ABR untuk menentukan keputusan muktamad. Masa akan

direkodkan semasa ujian saringan pendengaran oleh kedua – dua mesin tersebut bagi tujuan

analisa. Bagi mengurangkan ketirisan semasa penggunaan kedua -dua mesin tersebut, kami

menyediakan dua orang penganalisa yang akan membuat saringan ke atas pesakit.

KEPUTUSAN

Seramai 195 bayi (87, 44.6% adalah lelaki dan 108, 55.4% adalah perempuan) terlibat dalam

kajian ini. Risiko kesan sampingan ubat adalah yang paling ramai (51.8%) diiringi dengan

kesan penyakit kuning (51.3%) dan berat lahir < 1500g (27.2%). MB11 BERAphone

mempunyai markah tertinggi bagi lulus ujian saringan pendengaran (89.8%) berbanding

dengan OAEs sebanyak (85.2%). MB11 BERAphone mempunyai peratus yang sedikit bagi

ujian saringan yang “rujuk” iaitu 10.2% berbanding OAEs adalah 14.8%. Ini menunjukkan

perbezaan saringan pendengaran diantara kedua - dua ini adalah tercapai. Bagi “true negative”

untuk MB11 BERAphone adalah (29.4%) lebih tinggi berbanding OAEs iaitu (11.8%). “False

negative” MB11 BERAphone adalah (5.9%) dan OAEs adalah (0.0%). Daripada 195 bayi,

seramai 182 (93.4%) bayi menunjukkan kesepakatan di antara kedua - dua mesin saringan

pendengaran ini dan seramai 13 (6.6%) bayi tiada kesepakatan. Dari situ, sasaran kesepakatan

tercapai (kappa = 0.698, p<0.001). Ujian masa bagi saringan pendengaran yang telah dilakukan

oleh kedua - dua mesin menunjukkan MB11 BERAphone adalah lebih kurang 5 minit (IQR:

25th-75th) dan OAEs adalah lebih kurang 2 minit (IQR: 25th-75th). Ini menunjukkan perbezaan

masa di antara kedua – dua mesin tercapai (p = <0.001).

Page 12: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

12

KESIMPULAN

Kesimpulannya, saringan pendengaran menggunakan MB11 BERAphone adalah sesuai

digunakan semasa ujian saringan di kalangan bayi –bayi yang berisiko tinggi. Walaupun

terdapat keputusan “false negative” menggunakan MB11 BERAphone dan kemungkinan

keputusan tersebut tidak tepat disebabkan beberapa faktor yang mempengaruhi keputusan

tersebut. Terdapat kesepakatan di antara dua mesin tersebut. Walau bagaimanapun, masa yang

diambil oleh MB11 BERAphone untuk membuat ujian saringan pendengaran setiap bayi

mengambil masa yang agak lama berbanding dengan OAEs. Kami mencadangkan OAEs dan

MB11 BERAphone boleh digunakan bagi peringkat pertama ujian saringan pendengaran,

kemudian bayi boleh disaringkan sekali lagi melalui ABR sekiranya keputusan saringan

pendengaran adalah “rujuk” pada peringkat pertama.

Page 13: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

13

ABSTRACT

TITLE

A Comparison of Automated Auditory Brainstem Response (AABR) with Integrated Electrodes

and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing Screening in Hospital

Universiti Sains Malaysia.

OBJECTIVE

To compare the outcome between AABR (by using MB11 BERAphone method) and

Integrated Electrodes and Otoacoustic Emissions (OAEs) in high risk newborn in Hospital

Universiti Sains Malaysia.

METHODS

This is an observational study which is a cross-sectional study design involving two methods

in the same subject. A total of 195 high risk newborn and who have fulfilled the inclusion and

exclusion criteria will be participate in our study. These high risk babies were subjected to

OAEs and followed by an MB11 BERAphone screening test at the same setting as near to

discharge as possible or once the patient is stable enough to do a hearing screening. Both

instruments produced a “pass” or “refer” result and did not require any special skills for the

interpretation of results. The “refer” result from OAEs and MB11 BERAphone will be screened

using ABR to determine the false positive. The time will be measured by total staff time spent

on each instrument. To minimise the measurement bias, two testers will be used to handle both

screening methods.

Page 14: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

14

RESULTS

There were 195 newborns (87, 44.6% boys and 108, 55.4% girls) who participated in this study.

Ototoxic medication was the most common risk factor (51.8%) followed by

hyperbilirubinaemia (51.3%), and birth weight <1500g (27.2%). MB11 BERAphone had a

higher passing rate (89.8%) as compared to OAEs (85.2%). MB11 BERAphone had a lower

refer rate (10.2%) compared to OAEs (14.8%). These differences are statistically significant.

The true negatives are MB11 BERAphone (29.4%) and OAEs (11.8%). False negative MB11

BERAphone (5.9%) and OAEs (0.0%). Out of 195 newborns examined, 182 (93.4%) showed

agreement between the two techniques, whereas in 13 (6.6%) there was no agreement. Inter-

observer agreement was good (kappa = 0.698, p=<0.001). The median test time that was done

for each newborn using MB11 BERAphone was 5 minutes (IQR: 25th-75th) and OAEs was 2

minutes (IQR: 25th-75th). The difference were statistically significant (p = <0.001).

CONCLUSION

The MB11 BERAphone is still a reliable device for auditory brainstem response among high

risk newborn hearing screening. In the presence of false negative in MB11 BERAphone, it

might not really be significant in this study due to a few factors affecting the result. Both

agreements were good. However, the duration of time for hearing screening for each newborn

took a significantly longer time compared to OAEs. Therefore, we recommend that both

methods can be used as a first screening, followed by a screening using ABR for those whose

result was “refer” from the first screening.

Page 15: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

15

1. INTRODUCTION

Page 16: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

16

CHAPTER 1

1. INTRODUCTION

Hearing is the ability to perceive sound by detecting vibrations. Hearing is performed

primarily by the auditory system in which mechanical waves, known as vibrations are detected

by the ear and transduced into nerve impulses that are perceived by the brain (primarily in the

temporal lobe). Hearing loss is a partial or total inability to hear. Hearing loss may occur in one

or both ears. In children hearing, problems can affect the ability to learn spoken language later

on when they grow. So it is important to detect earlier which of these babies are on the high

risk of hearing impairment.1

Early diagnosis and intervention are necessary for social and linguistic development in

children with congenital hearing loss. The Universal Newborn Hearing Screening (UNHS) has

proven beneficial in detecting hearing impairment shortly after birth and with adequate

habilitation, it gives the child a better chance of normal development. Many techniques are

used for the evaluation of hearing sensitivity and among them we have Integrated Electrodes

Otoacoustic Emissions (OAEs) and also automated auditory brainstem response (AABR).2

Otoacoustic emissions (OAEs) hearing screening is used widely in hospital-based

newborn hearing screening programs. Otoacoustic emissions (OAEs) screening can help to

detect sensorineural hearing loss occurring in the cochlea. It can also call attention to hearing

disorders affecting the pathway to the inner ear. The procedure is performed with a portable

handheld screening unit. A small probe is placed in the child's ear canal. This probe delivers a

low-volume sound stimulus into the ear. The cochlea responds by producing an otoacoustic

emission, sometimes described as an “echo,” that travels back through the middle ear to the ear

canal and is analysed by the screening unit. The otoacoustic emissions (OAEs) screening

Page 17: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

17

performed by White et al (1993) on 1,850 neonates showed a sensitivity of around 100% and

a specificity of 73%.1

The automated auditory brainstem response (AABR) screener is a dedicated hearing

screening device which provides information not only about the outer/middle ear and cochlea

but also about the auditory pathway up to the brainstem. AABR screening is highly sensitive.

The screening is based on the measurement of synchronous activity in the auditory nerve up to

the colliculus inferior in the brainstem as a reaction to click stimuli delivered to the ears. It is

useful in infants at risk of hearing impairment, including those admitted to a Neonatal Intensive

Care Unit (NICU).

MB11 BERAphone is one of the automated auditory brainstem response (AABR). The

MB11 BERAphone is a more recently developed hearing screening device. MB11

BERAPhone is the only automatic ABR screener without adhesive electrodes. The click

stimulus used simultaneously reaches cochlear areas generating a more robust and faster

auditory response compared to normal AABR clicks. The MB11 BERAphone test showed very

good specificity 96.8% and sensitivity 100%.3

Auditory Brainstem Response (ABR) is the gold standard test, which is essential to a

correct neonatal screening programme both in patients not passing the test with otoacoustic

emissions (OAEs) and automated auditory brainstem response (AABR). The auditory

brainstem response (ABR) reflects the function of the entire auditory pathway up to the

brainstem. While both ABR and the alternate screening technology of otoacoustic emissions

(OAEs) detect cochlear hearing loss, only ABR detects auditory neuropathy.

Page 18: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

18

However, it has some drawbacks such as high cost, difficult instrument transportation,

long execution time, and a need for qualified personnel to interpret the ABR. This study is

conducted due to there being only a few research studies done about the comparison of MB11

BERAphone with OAEs. So far, there is no research study done in Malaysia comparing both

of these instruments.

Page 19: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

19

2. LITERATURE REVIEW

Page 20: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

20

2. LITERATURE REVIEW

2.1 UNIVERSAL NEONATAL HEARING SCREENING

The ear is one of the important parts of the body in which to function as hearing.

Hearing is one indicator for a baby to grow as a normal child in terms of cognitive development,

as well as neurodevelopment. Early neonatal hearing screening was developed all over the

world to prevent late detection of hearing impairment. Thus, hearing screening was done during

the neonatal period. Hearing loss may have significant adverse effects on the development of

speech, language capabilities and social - emotional development, as well as leading to

worsening educational and occupational performance in adulthood. Regular physical

examinations cannot detect hearing loss, so neonatal hearing screening (NHS) is necessary.

Hearing loss may be sensorineural or conductive, permanent or transient, unilateral or bilateral

and of varying severity. Infants with moderate or worse (>40 dB) bilateral sensorineural

hearing loss (SNHL) have heightened risk of poor speech and language development outcomes

if hearing augmentation/intervention programs are not implemented promptly.4

The incidence of permanent hearing impairment in newborns ranges between 1.0% and

5.5% across regions and countries.5 Many studies have illustrated the validity and reliability of

Universal Neonatal Hearing Screening (UNHS) programs in the early detection of newborn

hearing impairment. More importantly, newborns with hearing loss that is detected early can

receive intervention before 6 months of age, which is critical in allowing them to develop

linguistic, cognitive and logical abilities on par with normal infants. Because of their feasibility

and effectiveness, Universal Neonatal Hearing Screening (UNHS) programs have been

implemented in many countries all over the world; its coverage rate is one of the most important

indicators to evaluate the impact of these programs.5

Page 21: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

21

2.2 NEONATAL HEARING SCREENING AMONG HIGH RISK NEWBORN

In most countries, newborn hearing screening programmes that screen only high-risk

infants have been in existence for more than 20 years. However, this group of infants with

hearing loss comprises only 50% of newborn population with hearing loss. Therefore, hearing

screening programs that screened only high-risk neonates missed out 50% of hearing impaired

newborns, who are from among infants without any risks factors. The prevalence of hearing

loss is estimated to be between 2.5% and 10% among high-risk infants.1

The risk factors of hearing loss in neonates were first documented in 1994 and then

revised in 2000 by the Joint Committee on Infant Hearing (JCIH). The newborn high risk group

included infants who had asphyxia (low APGAR score, 0–4 at 1 minutes or 0-6 at 5 minutes),

meningitis, congenital or perinatal infections, anatomic defects or stigmata, hyper

bilirubinemia, a family history of hearing loss, low birth weight, ototoxic medications,

syndromes known to be associated with hearing loss and neonatal illnesses requiring

mechanical ventilation.6’7 Other risk factors have been tested, such as maternal drug abuse,

persistent high pulmonary pressure, intra-ventricular haemorrhage, high C reactive protein

levels but were not proven to be significant.7

High risk newborns can be divided into congenital and acquired hearing loss. Mostly,

they have hearing impairment at birth and are potentially identifiable by newborn and infant

hearing screening. The risk factors for congenital hearing loss such as craniofacial anomalies,

syndromes related with hearing loss, a family history of hearing loss, premature baby and in

utero infection. However, some congenital hearing loss may not become evident until later in

childhood. Hearing impairment can also be acquired during neonatal or infancy period for

various reasons. Examples of babies who are at high risk of acquired hearing loss are those

Page 22: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

22

who suffer from such conditions such as meningitis, otitis media, mechanical ventilation, hyper

bilirubinaemia and ototoxic medications.8

2.3 SCREENING TEST AND EQUIPMENT

The ideal screening test would have a high sensitivity and a high specificity. A high

sensitivity is particularly important to enable catching up on all infants with significant hearing

loss without a delay in the diagnosis of hearing impairment. A high specificity is required, as

a false positive result could lead to much workload on the diagnostic services (over referrals)

and undue parental anxiety.6 Two objective methods were used in most universal hearing-

screening programs. They are automated otoacoustic emissions (OAEs) and automated ABR

(AABR). They are available as handheld portable equipment with a pass or fail criterion. In

this study, we are using MB11 BERAphone as an AABR and comparing it with otoacoustic

emissions (OAEs). The Auditory Brainstem Response (ABR) is still the gold standard test,

which is essential to correct neonatal screening programme both in patients not passing the test

with otoacoustic emissions (OAEs) and automated auditory brainstem response (AABR),

MB11 BERAphone.9

2.3.1 Integrated Electrodes and Otoacoustic Emissions (OAEs)

OAEs are used to assess cochlear integrity and are physiologic measurements of the

response of the outer hair cells to acoustic stimuli. They serve as a fast objective screening test

for normal preneural cochlear function through the use of probe in the ear canal.6 The presence

of normal responses in an OAEs test is a strong predictor of a full hearing function. The

procedure of OAEs suppression allows for a functional investigation of the efferent

Page 23: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

23

olivocochlear system, which plays an important role in auditory information processing.10 The

sensitivity and specificity of the OAEs were found to be 90% and 92.4% when compared to

ABR results and 90.9% and 91.1% when compared to the children’s hearing status,

respectively.11

Currently, two types of evoked OAEs measurements are used for newborn hearing

screening, transient evoked otoacoustic emissions (TEOAEs) and distortion product

otoacoustic emissions (DPOAEs). DPOAEs measurements are better suited to advanced

clinical investigation on adult patients, even though DPOAEs analysis is complex and

interpretation is difficult. The distortion product (DP) technique is more flexible and potentially

more powerful than TEOAEs analysis, having a wider useful frequency range. Waveform

based TEOAEs measurements, as originally used in universal newborn hearing screening

programmes, are also useful as a sensitive initial screen prior to full clinical examination.

TEOAEs are also more sensitive to cochlear status changes manifested in subtle changes in the

TEOAEs waveform. DPOAEs instruments can be used for screening with an appropriately low

stimulus level, but DPOAEs screening instruments are generally not flexible enough for

clinical applications.12

Page 24: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

24

Figure 2.1: (A) TEOAEs probe containing miniature sound source and microphone

transducers. The soft disposable tip carries sound ports for the stimulus and for the microphone.

DPOAEs probes have an additional stimulus port. In some probes, all ports feed a single sound

tube. (B) The probe needs to be deeply inserted in the ear canal for maximum OAEs capture

and noise exclusion, with the cable positioned so as to avoid noise production on movement.13

2.3.2 MBII BERAphone

As we know there are two methods commonly used in universal newborn hearing

screening: Otoacoustic emissions (OAEs) and automated auditory brainstem

response (AABR), both methods being automated. AABR is superior to OAEs, among the

screening tests, because the AABR will be able to identify retrocochlear hearing impairments,

such as auditory neuropathy which are missed out on OAEs screening. The AABR test uses a

series of click sounds at 35 dB hearing level and detects brainstem responses to these stimuli.

AABR has also been found to be time and cost effective, with a high sensitivity and a low

failure rate.14

A B

Page 25: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

25

The MB11 BERAphone is an AABR which has been found to be an effective neonatal

hearing screening tool. To identify the presence of an ABR response automatically, the device

uses a chirp stimulus and takes a maximum test time of 180 seconds based on an implemented

statistical test algorithm. This implemented test algorithm attempts to detect a response from

the auditory system. The screening test generates a “pass” result when a response is detected.

The screening test generates a “refer” result if there is no response within 180 seconds. The

device is set at a screening stimulus level of 35 dB HL for a pass.15

Based on a study by Andrea Melagrana et al, 2007, the MB11 BERAphone test showed

very good specificity 96.8% (95% CI 94.8—98.7%) and sensitivity 100% (95% CI 93.9—

100%) for diagnosis of hearing loss.3

Figure 2.2: Pass result16

Page 26: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

26

Figure 2.3: Refer result16

2.3.3 Auditory Brainstem Response (ABR)

Two methods for neonatal hearing screening based on different physiological

phenomena are available: otoacoustic emissions (OAEs) and auditory brainstem response

(ABR). For the neonatal period, conventional ABR is considered to be the most reliable method

for assessment of the hearing level. Conventional ABR has long been recognised as the most

sensitive method of assessing the auditory acuity of newborns.16 The conventional ABR

method is not widely used for screening because it is time consuming, tends to be too expensive

and needs a well-qualified technician and audiologist to perform the test and evaluate the

results.17 Now, a second-generation Automated ABR infant hearing screener is available for

screening purposes. So in this study, a high risk newborn baby who receives a ‘fail’ during

hearing screening via OAEs and AABR will be screened again using conventional ABR.

Page 27: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

27

2.4 THE COMPARISON BETWEEN OAEs AND AABR

The goal of newborn auditory screening is to identify accurately infants with significant

hearing impairment in the most rapid and cost effective way. Currently, the two recommended

methods for screening include evoked otoacoustic emissions (OAEs) and auditory brainstem

response (ABR). Both of these methods do not assess hearing. Rather, they are an objective

means for estimating the likelihood of adequate hearing function in newborns. Several

automated commercial devices have been developed to meet the increasing demand for

newborn hearing screening tests. They developed Automated Auditory Brainstem Response

(AABR), one of advance and modified branch of ABR. They provide a final pass or fail score

without the need for a subjective interpretation of the data. These devices are convenient in that

they may be utilized quickly by personnel who are typically not trained to perform a clinical

assessment of hearing.18

Based on a study by J.I. Benito-Orejas et al, from 2001 to 2003, all the newborns in

their health area (2454 infants) were evaluated with TEOAEs (ILO92, otodynamics) and all

those born from 2004 to 2006 (3117) were evaluated with AABR (AccuScreen, Fischer-

Zoth).18 The population studied included all healthy newborns and those admitted to neonatal

intensive care units (NICU). The results from each study group were compared and analysed

for significant differences. TEOAEs screening yielded 10.2% fail results from the first

screening step; AABR gave 2.6%. In the second screening step, 2.0% of the newborns screened

with TEOAEs were referred, whereas 0.32% of those screened with AABR were referred.

These differences are statistically significant.18

Based on a study by S. Meier et al, they did a study on a comparison of OAEs and

AABR by using Echoscreen-TDA from Fischer-Zoth, and two AABR screeners, the Algo 3

from Natus and the MB11 BERAphone from Maico.19 Transiently evoked otoacoustic

Page 28: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

28

emissions (TEOAEs) were measured in one ear of 150 healthy newborns using the Echoscreen-

TDA. Three groups of 50 subjects each were tested additionally for AABR recordings either

with Echoscreen-TDA, Algo 3 or MB11 BERAphone. Their conclusion was that pass rates

were highest with 98% for AABR recordings using the Algo 3 and lowest with 92% for AABR

recordings using the MB11 BERAphone, but these differences were not statistically

significant.19

On the other hand, Bolajoko O. Olusanya et al, their study was about determining the

perinatal predictors of discordant screening outcomes based on a two-stage screening protocol

with transient-evoked otoacoustic emissions (TEOAEs) and automated auditory brainstem

response (AABR) under a hospital-based universal newborn hearing screening program in

Lagos, Nigeria. Their results were that 4718 of the infants enrolled under the program, 1745

(36.9%) completed both TEOAEs and AABR. Of this group, 1060 (60.7%) passed both

TEOAEs and AABR (‘‘true-negatives’’); 92 (5.3%) failed both TEOAEs and AABR (‘‘true-

positive’’), 571 (32.7%) failed TEOAEs but passed AABR (‘‘false-positives’’) while 22

(1.3%) passed TEOAEs but failed AABR (‘‘false-negatives’’).20

2.5 THE COMPARISON BETWEEN OAEs AND MBII BERAPHONE

The goal of newborn hearing screening (NHS) is to identify infants with significant

hearing impairment. A low false-positive rate is central to the success of a NHS program.

Therefore, minimizing false-positive results is critical in developing a more reliable newborn

hearing screening program. The otoacoustic Emissions (OAEs) and/or automated auditory

brainstem response (AABR) are most often used to detect sensory or conductive hearing loss.

Based on a study by Luca Guastini et al, they tested the validity of this automated ABR

screening method (using MB11 BERAphone) and OAEs in a multi-stage newborns hearing

Page 29: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

29

screening (NHS). Among the 8,671 newborns tested (males 3,889; females 4,782), only 42

newborns were lost to follow-up and the final false positive rate was of 0.03%.16

Another study conducted by S. Meier et al, was of the comparison of currently available

devices designed for newborn hearing screening, done by using OAEs combining with 3

methods of AABR (the Echoscreen-TDA from Fischer-Zoth, and two AABR screeners, the

Algo 3 from Natus and the MB11 BERAphone from Maico).19 They used transiently evoked

otoacoustic emissions (TEOAEs) and distortion product otoacoustic emissions (DPOAEs),

measured in one ear of 150 healthy newborns and three groups of 50 subjects each and were

tested additionally for AABR recordings either with Echoscreen-TDA, Algo 3 or MB11

BERAphone. They concluded that the device was the easiest for EOAEs measurements with

the Echoscreen-TDA, followed by AABR recordings with the Algo 3 and Echoscreen-TDA

and were most difficult with the MB11 BERAphone. Pass rates were highest with 98% for

AABR recordings using the Algo 3 and lowest with 92% for AABR recordings using the MB11

BERAphone, but these differences were not statistically significant. Connecting the subject

was the most difficult with the MB11 BERAphone, but there were no disposable supply costs.19

Based on a study by Ozlem Konukseven et al, they used a combination in initial stage

of hearing screening by using OAEs and AABR (MB11 BERAphone) for their study to see the

accuracy of the screening. One thousand and nine hundred seventeen babies were tested with

both TEOAEs and AABR in the first day of their life. Out of 1917 neonates, 202 (10.53%)

failed the initial TEOAEs, and 37 (1.97%) failed the initial AABR. Hundred and fifty eight of

the 202 neonates who failed the TEOAEs and all neonates that failed AABR were subjected to

the second test. Four (2.5%) neonates failed the second TEOAEs, and three (1.89%) failed the

second AABR.21

Page 30: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

30

A recent study by M. van Dyk et all, 2015 was done at a hospital in South Africa, which

is a comparison of TEOAEs and AABR using MB11 BERAphone. This study evaluated the

outcome of NHS within the first 48 hours. NHS was performed on 150 healthy newborns (300

ears) with TEOAEs and AABR techniques before being discharged from the hospital. Over the

three-stage screen, the AABR had a significantly lower refer rate of 16.7% (24/144 subjects)

compared to the TEOAEs 37.9% (55/145 subjects). Screening refer rate showed a progressive

decrease with increasing age.22

2.6 THE AGREEMENT OF OAEs AND MBII BERAPHONE

There were no specific journal done for the agreement among OAEs and AABR. Based

on Andrea Melagrana et al, the comparison of MB11 BERAphone and Auditory Brainstem

Response in newborns is at an audiologic risk. They had done a sample of newborns evaluated

after the second month of life from October 2002 to February 2005. The study was performed

on full-term newborns who presented with altered otoacoustic emissions and in newborns

considered at audiologic risk admitted to the Otorhinolaryngology (ORL) Unit of Giannina

Gaslini Institute. The sample included 201 children (104 males and 97 females) who underwent

on the same day, an audiologic test using MB11 BERAphone and standard ABR test. The

results were, out of the 388 ears examined, 378 (97.4%) showed agreement between the two

techniques, whereas in 10 (2.6%) there was no agreement. Inter-observer agreement was

excellent (kappa = 0.92, p = 0.0001).3 There is currently no known study by anyone comparing

the OAEs methods with the AABR about the agreement of neonatal hearing screening.

Page 31: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

31

2.7 THE TOTAL TIME SPENT BETWEEN OAEs AND MBII BERAPHONE

Although TEOAEs hearing screening was easier and more comfortable to perform than

that of AABR, the differences are becoming less and less important. As new generations of

AABR devices are developed, administration time is decreasing. It is true that screening test

time varies depending on the child’s hearing, the presence of partial obstructions in the external

auditory canal or the middle ear and on test conditions. Based on a study by J.I. Benito-Orejas

et al, which was conducted from 2001 to 2003 in 14 hospitals in Spain with all the newborns

in their health area (2454 infants) being evaluated. They have noted that the average time

needed to carry out automated ABR testing ranges from 8 to 15 minutes and conventional

TEOAEs tests take from 2 to 13 minutes. They calculated, without exact measurements, that

each test took an average of 15 minutes using the specified AABR equipment and about 10

minutes using that of TEOAEs.18

Another study by S. Meier et al, compared the OAEs with 3 other methods of AABR.

They used transiently evoked otoacoustic emissions (TEOAEs) and distortion product

otoacoustic emissions (DPOAEs) and measured in one ear of 150 healthy newborns using the

Echoscreen-TDA. Three groups of 50 subjects each were tested additionally for AABR

recordings either with Echoscreen-TDA, Algo 3 or MB11 BERAphone. From the study, they

noted that the median test time on one ear was less than 30 seconds for OAEs measurements

and 4 to 5 minutes for AABR recordings. 19

Page 32: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

32

3. OBJECTIVES

Page 33: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

33

3. OBJECTIVES

3.1 The general objective of this study:

To compare MB11 BERAphone and Integrated Electrodes and Otoacoustic Emissions

(OAEs) as a screening tool for hearing loss in high risk newborn

3.2 The specific objectives are:

1. To determine the association of passing and referral rate between MB11 BERAphone

and OAEs in high risk newborn

2. To determine true and false positive rate for hearing loss using MB11 BERAphone and

OAEs with ABR as a gold standard

3. To determine the agreement passing and referral rate between MB11 BERAphone and

OAEs in high risk newborn

4. To compare the total staff time spent for hearing screening program with MB11

BERAphone versus OAEs in high risk newborn

Page 34: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

34

3.3 RESEARCH HYPOTHESIS

1. Hyperbilirubinaemia and ototoxic medications are the commonest the risk factor

between MB11 BERAphone and OAEs in high risk newborn

2. There is significant difference of passing and referral rate between MB11 BERAphone

and OAEs in high risk newborn

3. There is high agreement of referral rate between MB11 BERAphone and OAEs in high

risk newborn

4. There is significant different of mean of total staff time spent between MB11

BERAphone and OAEs as screening program in high risk newborn

Page 35: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

35

4. METHODOLOGY

Page 36: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

36

4. METHODOLOGY

4.1 RESEARCH DESIGN

This was an observational study (a cross sectional study design). The study was conducted at

neonatal unit, Hospital Universiti Sains Malaysia (HUSM), Kubang Kerian, Kelantan from

April 2014 until December 2014.

4.2 PARTICIPANTS

All high risk newborns who were admitted to neonatal unit Hospital Universiti Sains Malaysia

(HUSM) and who had fulfill the inclusion criteria and exclusion criteria.

4.2.1 INCLUSION CRITERIA

The inclusion criteria for the study:

1. Meningitis

2. Birth weight <1500g

3. Craniofacial anomalies

4. Mechanical ventilation > 5 days

5. Low APGAR scores (0–4 at 1 min or 0-6 at 5 min)

6. Family history of hereditary childhood sensory neural hearing loss

7. Hyperbilirubinemia (severe jaundice- total bilirubin >300umol/L, early onset jaundice,

<24hours, rapidly rising jaundice >8.5umol/L/hr) or requiring exchange transfusion.

8. In utero infections (TORCH—toxoplasmosis, others (e.g. syphilis), rubella,

cytomegalovirus, herpes)

Page 37: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

37

9. Stigmata or other findings associated with a syndrome known to include sensory

neural and/or conductive hearing loss (eg: Down syndrome)

10. Ototoxic medications including but not limited to aminoglycosides used in multiple

courses or in combination with loop diuretics

4.2.2 EXCLUSION CRITERIA

The exclusion criteria for the study:

1. All healthy new born without a risk factor

2. Lethal congenital malformation

3. Severe neurological deficit since birth

4.3 INTERVENTIONS

This is an observational study which is a cross-sectional study designed to involve two

methods with the same subject. All newborns who were admitted to the neonatal unit in

Malaysia are made as a reference of this study. The source of population are all high risk

newborns who were admitted to the neonatal unit of Hospital Universiti Sains Malaysia

(HUSM), Kelantan and those who have fulfilled the inclusion and exclusion criteria will

participate in our study.

The researcher had explained about the study to the parents and caretakers, and the

inform consent was taken. The high risk babies who have fulfilled the inclusion and exclusion

criteria in the neonatal unit were subjected to OAEs and followed by MB11 BERAphone

screening test at the same setting as near to discharge as possible or once the patient is stable

enough to do a hearing screening. The test was done in their bassinets in the neonatal unit itself.

This is a non-invasive procedure with no sedation given to the babies. To minimise the risk of

Page 38: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

38

infection, alcohol swab were used to sterilise the instruments after using them, and before

screening another baby. The de-identification of patients will be ensured.

The neonates were tested in a sleeping or quiet state. The parents are allowed to help

calm their baby during the procedure. Both instruments produced a “pass” or “refer” result and

did not require any special skills for the interpretation of results. The test was repeated two

times in cases with “refer” result to confirm that it is not due to external factors. The “refer”

result from OAEs and MB11 BERAphone will be screened by using ABR to determine the

false positive. The time will be measured by the total time the staff spent on each instrument.

To minimize the measurement bias, two testers will be used to handle both screening methods.

All data were recorded on a data collection sheet and entered into IBM Statistical

Package for Social Sciences (SPSS) Statistics 21 by the main researcher. All documentation

was locked in a secure cabinet and was kept confidential for the length of this study.

4.4 OUTCOMES

The Neonatal Hearing Screening (NHS) programs have the purpose to help identify

hearing loss early, reducing the age of audiological diagnosis and therapeutic intervention. Two

physiological procedures are currently recommended: the evoked Otoacoustic Emissions

(OAEs) and the Automated Auditory Brainstem Response (AABR). However, the Auditory

Brainstem Response (ABR) is still a gold standard for neonatal hearing screening. In this study,

by using both methods, we were screening newborns with high risk factors and comparing the

effectiveness of both methods in term of passing rate, total time spent using both methods, the

agreement of screening methods and referral rate by using ABR as a second screening.

Page 39: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

39

4.5 SAMPLE SIZE

Sample size calculation:

1. Based on objective 1 are Mc Nemar Test and kappa agreement.

A. For Mc Nemar Test :

The sample size was calculated using PS software (Dupont & Plummer, 2009) and

online calculation at http://sampsize.sourceforge.net/iface. Based on studies by Meier

et al. (2004) and Abdul Wahid et al. (2012) and by using PS software the minimal

sample size is 65 babies.

Based on studies by Meier et al. (2004) and Abdul Wahid et al. (2012) and by using PS

software:

α 0.05

Power (1-β) 0.9

Failure rate in control (p0) 0.08

Failure rate in tested group (p1) 0.3

Group ratio (m) 1.0

Minimum sample size 65

Page 40: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

40

B. For Kappa (concordant observation) and the sample size was calculated using Stata

software (sskdlg). The minimal sample size is about 195 babies.

The sample size was calculated using Stata software (sskdlg):

Expected kappa 0.80

Proportion positive by rater 1 0.92

Proportion positive by rater 2 0.90

Precision 0.15

Minimum sample size 195

2. Based on Objective 3 using Paired T- Test and the sample size cannot be calculated as there

are no data on mean and standard deviation in the previous studies (effect size was unable

to be determined).

The sampling methods are using non-probability sampling in our study. Based on calculation

using Kappa (concordant observation), we decided that our sample is 195 newborns and no

drop outs in our study. In this study the newborns were chosen after being admitted to our

neonatal ward as they had fulfilled the inclusion criteria and were screened via two hearing

screening methods. Data was entered and analysed using SPSS (Statistical Package of Social

Science) software version 21.0.

Page 41: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

41

4.6 ETHICAL APPROVAL

The study was approved by the Research and Ethics Committee, School of Medical Sciences,

Universiti Sains Malaysia. A inform consent will be obtained by investigators from parents

before conducting the procedure. The consent form will be uploaded from the Human Research

Ethics Committee (HREC), Universiti Sains Malaysia (USM).

4.7 STATISTICAL ANALYSIS

Data entry and analysis were performed using the IBM Statistical Package for Social Sciences

(SPSS) Statistics version 21.0 software. These data were analysed using the Predictive

Analytics Software (PASW) for Windows version 21.0. Numerical data were presented in

mean (standard deviation), unless data was skewed, then it will be presented in median (IQR =

interquartile range). Categorical data were presented in frequency (percentage). Normal

distribution was checked using histogram with normal distribution curve and Whisker and

boxplots. For comparison for both methods, we used the Mc Nemar test and Measurement of

Kappa for agreement of both methods. A two tailed p value of <0.05 was considered

statistically significant.

Page 42: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

42

4.8 DEFINITIONS:

PASS RATES

A ‘‘pass’’ was defined as follows:

(a) Testing provided a ‘‘pass’’ after the first measurement or

(b) The ear probe, coupler earphone or the applicator were removed and repositioned if

testing did not give a pass, and measurements were repeated once to achieve a ‘‘pass’’

result.

REFER RATES

The test was rated as “refer”, if both measurement fail.

TOTAL TESTING TIMES

The total test time include the time for the unit startup, the preparation of the subject, the

measurement, and the unit shut down. The time for the unit startup include the time for starting

up the PC (Personal Computer) or the device only, if no PC was required. Preparation time

include fitting the probe, coupler earphones or the applicator in/around the ear and attaching

the electrodes to the infant. These time segments were measured with a stopwatch by the

observer.19

Page 43: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

43

4.9 FLOW CHART

Comparison of Automated Auditory Brainstem Response (AABR) with Integrated Electrodes

and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing Screening

Source of Population:

All high risk newborns who admitted to neonatal ward, HUSM,

Kelantan

Inclusion criteria

Exclusion criteria

Sample, n = 195

OAEs, n:195

Total Screening

Duration (minutes)

MB 11 BERAphone,

n:195

Total Screening

Duration (minutes)

Data entry, Analysis and Interpretation

using SPSS 21.0

Write up and preparation for thesis

Thesis Submission

Pass Refer

ABR

Page 44: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

44

5.RESULTS

Page 45: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

45

5. RESULTS:

5.1 SAMPLE CHARACTERISTICS

During the period of study, a total of 195 newborns with high risk were screened in the neonatal

unit, Hospital Universiti Sains Malaysia. All the 195 babies had fulfilled the inclusion and

exclusion criteria and their parents/caretakers have consented. The babies were screened via

Integrated Electrodes Otoacoustic Emissions (OAEs) and Automated Auditory Brainstem

Response (using MB11 BERAphone) on both ears. Each machine was set for the timing and

the result who were “refer” to one or both ears of both these machines were subjected to

auditory brainstem response (ABR) within 3 months after discharge. There were newborns who

were missing in our follow-up for ABR screen. There were a few factors causing them to count

as a loss in our follow-up, which will be elaborated further in the next result and discussion.

Page 46: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

46

The flow diagram of the study as shown in figure 5.1.

Source of Population:

All high risk newborns who admitted to neonatal ward, HUSM,

Kelantan

Inclusion criteria

Exclusion criteria

Sample, n = 195

OAEs, n:195

Total Screening

Duration (minutes)

MB 11 BERAphone,

n:195

Total Screening

Duration (minutes)

Data entry, Analysis and Interpretation

using SPSS 21.0

Pass, n:

OAEs: 166

MB11: 175

Refer, n:

OAEs: 29

MB11: 20

ABR, n: 17

Total Refer,

n: 31

Loss of

follow up,

n:14 *One patient may have “refer”/ “pass” both methods

Total Pass,

n: 164

Page 47: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

47

5.2 DEMOGRAPHIC DATA

5.2.1 DEMOGRAPHIC CHARACTERISTICS

As shown in table 5.1, the total newborn babies who were involved in this study were 195

babies. Among them, 87 babies are boy and 108 babies are girl. The newborn babies were

recruited from neonatal unit, Hospital Universiti Sains Malaysia and the hearing screening was

done as near to discharge as possible or once the patient is stable enough to do a hearing

screening The distribution as shown in table 5.1. The median age during this study was 7 days

old and base on inter-quartile range (histogram skewed to the right) were between 3 – 14 days

old.

Table 5.1: Demographic profile of newborn baby undergone hearing assessment using

Integrated Electrodes Otoacoustic Emissions (OAEs) and MB11 BERAphone (n=195)

Variables

n

(%)

Gender

Boy 87 44.6

Girl 108 55.4

Ethnic

Malay 191 98.0

Chinese 3 1.5

Indian 0 0.0

Other 1 0.5

Age (days)

Median (IQR) 7 (25th, 75th)

* IQR = interquartile range (25th – 75th )

* Ethnic; other: siamese

Page 48: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

48

5.2.2 ETHNIC DISTRIBUTION

A hundred ninety one of these babies were Malay. Three were Chinese and only one was

Siamese. This is graphically shown in Figure 5.2.

Figure 5.2: Ethnic distributions of newborn baby who had undergone hearing assessment

using Integrated Electrodes Otoacoustic Emissions (OAEs) and MB11 BERAphone (n=195)

191

3 1

Ethnic

Malay Chinese Others

Page 49: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

49

5.3 RISK FACTORS AMONG NEWBORN

Risk factors were identified in 195 newborns. The most often risk factors were:

administration of ototoxic medications, hyper bilirubinaemia and newborn with a birth

weight of less than 1500 grams. As the list stated below, the newborns have at least one

risk factor.

Tables 5.2: Risk factors of hearing loss among newborns in the neonatal unit, Hospital

Universiti Sains Malaysia (HUSM),( n= 195 )

Risk Factors n (%)

Family history

6

3.1

In utero infection 11 5.6

Syndromes 4 2.1

Birth weight < 1500 gram 53 27.2

Hyper bilirubinaemia 100 51.3

Ototoxic medications 101 51.8

Meningitis 6 3.1

Low Apgar score 5 2.6

Mechanical ventilation > 5d 7 3.6

Craniofacial anomalies 6 3.1

* One patient may have multiple risk factors

Page 50: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

50

5.4 THE COMPARISON BETWEEN OAEs AND MBII BERAPHONE

During the screening of 195 newborns with both OAEs and AABR, we found that 164 (84.2%)

newborns were ‘pass’ and 18 (9.2%) newborns results were ‘failed’ with hearing screening

using both methods. There was an association between the results of hearing screening between

two methods. The P value was significant (p= 0.022). The data is as shown in the Table 5.3:

Table 5.3: The contingency table of findings using Integrated Electrodes Otoacoustic

Emissions (OAEs) and MB11 BERAphone

MB11

Pass Refer P valuea

Pass

164

2

(84.2%) (1.0%)

OAEs

Refer 11 18 0.022

(5.6%) (9.2%)

a Mc Nemar Test

*p value <0.05 indicate significant

Page 51: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

51

5.5 THE “REFER RATES” OF OAEs AND ABR

Among these 195 high risk newborn, there were 31 newborns having the result “refer” during

screening with both methods. Out of 31 newborns, only 17 newborns went through a second

screening via the gold standard, ABR. Out of 17 newborns, 15 newborns had the result “refer”

via OAEs method whether from one or both ears. The table below showed the result

“refer” from OAEs and the second screening with ABR.

Table 5.4: The true and false positive rate for hearing loss using OAEs and a gold standard

Auditory Brainstem Response (ABR), n = 17

OAEs

Pass Refer

Pass

2 (11.8%)

12 (70.6%)

ABR

Fail 0 (0.0%) 3 (17.6%)

*One patient may have “refer rate” from both methods

Page 52: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

52

5.6 THE “REFER RATES” OF MBII BERAPHONE AND ABR

Only 31 newborns had the result of “refer” from both methods. Out of these 31, 17 newborns

need to undergo a second screening via ABR. From that, 11 newborns had the result “refer”

via MB11 BERAphone method whether on one or both ears. The table below shows the result

“refer” from MB11 BERAphone and second screening with ABR:

Table 5.5: The true and false positive rate for hearing loss using MB11 BERAphone and the

gold standard Auditory Brainstem Response (ABR)

MB11

Pass Refer

Pass

5 (29.4%)

9 (52.9%)

ABR

Fail 1 (5.9%) 2 (11.8%)

*One patient may have “refer rate” from both methods

Page 53: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

53

5.7 THE AGREEMENT BETWEEN OAEs AND MBII BERAPHONE

As the table below shows, the Kappa statistic is 0.698, although there are 182 cases which

agree between two tests out of the total of 195 cases. There was a substantial agreement

between passing and referral rate of Integrated Electrodes Otoacoustic Emissions (OAEs) and

MB11 BERAphone by using the Kappa statistic.

Table 5.6: The agreement between passing and referral rate between MB11 BERAphone and

OAEs by using Kappa statistics, (n=195).

MB11

Pass Refer P value

Pass

164 (84.2%)

2 (1.0%)

OAEs

Refer 11 (5.6%) 18 (9.2%)

Measure of agreement Kappa

(value)

0.698 <0.001

*If Kappa is “zero”, agreement is only at the level expected by chance.

*If Kappa is “negative, the level of agreement is less than by chance.

*If Kappa: 0-0.2=slight; 0.2-0.4=fair; 0.4-0.6=moderate; 0.6-0.8=substantial; >0.8=almost

perfect

Page 54: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

54

5.8 THE TOTAL TIME SPENT AMONG STAFF BETWEEN USING OAEs AND

MBII BERAPHONE

There were two machines used for this hearing screening. The total staff time spent was

recorded by using OAEs machine and MB11 BERAphone. The median for total time using the

OAEs machine was about 2 minutes and the MB11 BERAphone was about 5 minutes duration

during the hearing screening. The table is as shown below:

Table 5.7: The comparison of total staff time spent in minutes for screening programs

between Integrated Electrodes Otoacoustic Emissions (OAEs) and MB11 BERAphone.

Median (IQR)

(minutes)

Z- Statistic

P valuea

OAEs

2 (25th, 75th)

MB11

5 (25th, 75th)

-10.851 <0.001

a Wilcoxon Signed Rank Test

* IQR = interquartile range (25th – 75th)

Page 55: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

55

6. DISCUSSION

Page 56: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

56

6. DISCUSSION

The incidence of permanent hearing impairment in newborn infants ranges between

1.0% and 5.5 % across regions and countries.5 Hearing loss may have significant adverse

effects on the development of speech, language capabilities and social emotional development,

as well as leading to worsening educational and occupational performance in adulthood.

Regular infant physical examinations cannot detect hearing loss, so neonatal hearing screening

(NHS) is necessary. Many studies have illustrated the validity and reliability of universal

neonatal hearing screening (UNHS) programs in the early detection of newborn hearing

impairment.5

This increase in newborn hearing screening resulted from several factors. Firstly,

largely as a result of the pioneering efforts of Dr Marion Downs, the importance of newborn

hearing screening became recognised as early as 1969, when the Joint Committee on Infant

Hearing was established. Secondly, the identification of hearing loss in the neonatal period

became possible by the late 1980s as a result of the revolutionary work of Dr David Kemp in

the development of the otoacoustic emissions (OAEs) technology that could be applied to the

screening of hearing in infants and the development of automated procedures and equipment

for doing auditory brainstem response (ABR) testing. With the availability of this technology,

Surgeon General C. Everett Koop issued an ambitious challenge in 1988 that by the year 2000,

90% of all infants who were born with significant hearing loss would be identified by 12

months of age.23

There were many studies regarding hearing screening in newborns and some research

comparing methods of hearing screening in which mostly were using otoacoustic emissions

(OAEs) and automated brainstem response (AABR). There were a few studies done by

comparison of methods in high risk newborns. In this study, the neonatal hearing screening

Page 57: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

57

was done in Hospital Universiti Sains Malaysia (HUSM), Kubang Kerian, Kelantan by

comparing two methods, OAEs with MB11 BERAphone in high risk newborns.

6.1 Demographic Characteristics and Risk Factors in Newborns

In our study, the median age for high risk newborns during the hearing screening was

7 days old with girls more than boys in the gender distribution. During the study period, we

identified that most of the newborns are Malays, compared to other races. It could be because

in Kelantan, the majority of the population are Malays. We have one patient who is Siamese,

possibly due to Kelantan being a state close to neighbouring Thailand. The most common risk

factors in our study were hyper bilirubinaemia and ototoxic medications. Almost 50% of our

high risk newborns who were screened suffer from both risk factors. The birth weight is less

than 1500 grams, comprising also about 27% among risk factors. In my study, each newborn

screening may have at least one risk factor.

In a study by Wroblewska-Seniuk et al, they aim to evaluate the results of newborn

hearing screening by means of TEOAEs. From their study, the risk factors were identified in

739 newborns and the most often found risk factors were: use of ototoxic drugs, low apgar

score and prematurity. The ototoxic medications were observed as an important risk factor in

most cases were not a single risk factor.13

In the study performed by A. Zamani et al, the incidence of hearing loss in neonates

who were in the NICU for more than 48 hours was about 2-4%. Patients hospitalised in the

NICU must be screened for hearing loss, because these neonates probably have more problems

and multiple risk factors, such as prematurity, low birth weight, use of ototoxic drugs and

mechanical ventilation.24

Page 58: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

58

There is one study done in Hospital Universiti Sains Malaysia, Kubang Kerian under

Hidayah S.N et al, about a comparison of OAEs and AABR which involved 73 newborns. In

the study, it showed that ototoxic medication was the most common risk factor followed by

hyperbilirubinaemia and low birth weight.2

The least risk factor in our study was syndromic baby, about 2%. In our study in high

risk newborns, the median of age was 7 days, however, the age of hearing screening ranged

from 2 to 100 days because some preterm infants required a longer stay due to few reasons

such as awaiting weight gain, establishing feeding and etc. In Tersia de Kock et all, the mean

age at first stage screen was 6.1 days (SD 8.1) and the age of hearing screening ranged from 0

to 189.25

6.2 The Comparison Result between Two Methods

The acceptable methods for universal newborn hearing screening include both OAEs

and AABR, either alone or in combination. The OAEs method measures sound waves

generated in the inner ear (cochlea) by miniature microphones placed in the external ear canals.

Although OAEs screening is quicker and easier to perform than AABR, OAEs is affected by

external ear wax or fluid.26 The AABR measures the electroencephalographic waves generated

in response to a click sound by three electrodes pasted to the infant’s scalp.26

While newborn hearing screening provides early detection of babies with hearing loss,

it also helps to differentiate auditory neuropathy (AN) and auditory dyssynchrony (AD) cases

when the screening is performed with both AABR and otoacoustic emissions (OAEs) tests.

Thus, the routine combined the use of AABR and OAEs tests in UNHS programs, especially

for high-risk infants, and is thought to provide better detection of newborns with AN/AD.27

Page 59: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

59

In this study, we are comparing both methods by using OAEs and MB11 BERAphone

in high risk newborns. The result in our study by using a comparison between OAEs and MB11

BERAphone showed both methods agree that 164 (84.2%) out of 195 newborn passed the test

during hearing screening. A total of 31 (15.4%) newborns ‘failed’ during this hearing screening

using both methods. From 31 newborns who failed this hearing screening, only 18 (9.2%) were

showed fail results by both machines. By using MB11 BERAphone only 20 newborns were

refer rate compared to using the OAEs, about 29 were refer rate. This shows that using MB11

BERAphone (AABR) has a lower refer rate compared to OAEs. However, there was an

association between the results of hearing loss examination between two methods. The P value

was significant (p= 0.022).

M. van Dyk et al in their study, over the three-stage screen AABR (MB11 BERAphone)

had a significantly lower refer rate of 16.7% (24/144 subjects) compared to TEOAEs 37.9%

(55/145 subjects). For both TEOAEs and AABR, refer rate per ear screened 24 hours post birth

was significantly lower than for those screened before 24 hours. For infants screened before 12

hours post birth, the AABR refer rate per ear (51.1%) was significantly lower than the TEOAE

refer rate (68.9%).22 Based on a study by S.Noor et al, by comparison, OAEs and AABR

showed significant differences. TEOAEs screening yielded 10.2% fail results from the first

screening step, while AABR gave 2.6%. In the second screening step, 2.0% of the newborns

screened with TEOAEs were referred, whereas 0.32% of those screened with AABR were

referred.2 In NICU babies, the different aetiologies have different referral rates. The presence

of OAEs provides direct evidence of the existence of an active mechanism in the cochlea.

However, OAEs will be absent if there is a significant external and/or middle ear problems.

Besides that, the OAEs test must be done in an environment where it is quiet, the infants are in

a calm condition and the probe tip is snugly fitted in order to get reliable results. 2

Page 60: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

60

The AABR test measures the electrical activity beyond the cochlea until the lower

brainstem level and in contrast to the OAEs test it is less affected by the outer, middle and/or

inner ear disorders. Therefore, the utilisation of AABR technology rather than OAEs will

further improve the values of passing rate and false positive. However, similarly to OAEs, the

results of AABR test is also subjected to the influence of factors such as the environmental

noise, child condition and equipment condition during the testing. 2

6.3 The Refer Rate of OAEs and MBII BERAphone with Second Screening Using ABR

Auditory Brainstem Response (ABR) is the gold standard test, which is essential to a

correct neonatal screening programme both in patients not passing the test with otoacoustic

emission and in patients coming from NICU. However, it has some drawbacks such as high

cost, difficult instrument transportation, long execution time, and a need for qualified personnel

to interpret the ABR.3 Because of that, they modified it to provide an easier way in which it

developed automated auditory brainstem response. In this study, we were using OAEs and

MB11 BERAphone as the first screening. The resulting “refer” will be screened again by using

this afore-mentioned gold standard as the second screening.

In our study from 195 high risk newborns, 31 newborns had the result “refer” during

first screening using both methods: OAEs and MB11 BERAphone. From that we gave them an

appointment for the second screening, using ABR within 3 months. Out of 31 high risk

newborns whose results were “refer” with both methods, only 17 newborns had undergone the

second screening within 3 months and 14 newborns were loss from the follow up. It was about

45% of our study by using second stage of screening was lost to this loss of follow up.

Internationally, loss to follow-up is one of the greatest challenges experienced in newborn

Page 61: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

61

hearing screening programmes, with follow-up rates in the region of 50% often being

reported.25

In Koenraad Smets et al, they showed that 25% from their study have loss follow up

due to the babies passing away, multiple congenital anomalies required first line ABR and

some have no apparent reason for not being screened.28 In this study, 14 of our patients who

were loss to our follow up, and they have a few reasons to explain it. The reasons were: loss of

their contact numbers, staying far away from our place of study, difficulty in transportation,

infant death and a few patients just refused to come for further screening due uncertain reasons.

Based on our study by using ABR for the second screening, the true positive result is

(OAEs: 17.6%, MB11: 11.8%), true negative (OAEs: 11.8%, MB11: 29.4%), and false positive

(OAEs: 70.6%, MB11: 52.9%). However, the false negative in OAEs was 0.0% compared to

MB11 BERAphone which was 5.9%. From this study, MB11 BERAphone had a better result

compared to OAEs. Unfortunately, the MB11 BERAphone had false negative result compared

to the OAEs in which this gave significant impact of effectiveness of the MB11 BERAphone

when screening the babies. This produced a warning that the chance of missed patients with

hearing impairment is higher in MB11 BERAphone compared to usage of the OAEs. In our

study, the MB11 BERAphone also has a lower false positive rate compared with OAEs.

In a study by Luca Guastini et al, they have noted that by using MB11 BERAphone

there was lower false positive rate compared to OAEs. But the conventional auditory brainstem

response is still the most reliable method for assessing the hearing level and minimising the

false-positive rate. They have noted that AABR is easy to use, fast and with a good compliance,

but the device is unable to provide accurate and certain diagnosis on the degree of hearing loss

to allow a proper treatment.16

Page 62: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

62

Adrea Melagrana et al, were using MB11 BERAphone screening with ABR on

audiologic risk newborns and showed that the MB11 BERAphone test yielded no false

negatives and 10 false positives which resulted as normal at ABR. The MB11 BERAphone test

showed very good specificity 96.8% and sensitivity 100%, positive predictive value 88.2% and

negative predictive value 100% for diagnosis of hearing loss.3

In our study, it showed that the referral rate in MB11 BERAphone was lower compared

to OAEs. A few studies also showed the same as results as ours did. The research by Ozlem

Konukseven et all, showed that a first stage screening using the AABR method, it has the lowest

false positive rate, referral rate and high specificity.21

Bolajoko O. Olusanya et al, stated that the 4718 infants enrolled under the program

1745 (36.9%) completed both TEOAEs and AABR. Of this group, 22 (1.3%) passed TEOAEs

but failed AABR (‘‘false-negatives’’).20 Similar to our study, one patient passed during

screening via MB11 BERAphone, however the result showed false negative via ABR. There

are possibly a few reasons why the MB11 BERAphone is showing a false negative result in

our study. First, it is possible due to lack of experience and technique by our testers who were

using this machine especially when it took more time to screen the baby to maintain the

effectiveness of this machine. Second, there is a possibility that due to the movements of the

babies or them not really being calm during screenings caused some effects on the results.

Third, it is possible due to uncertain medical conditions that will be affecting their hearing

function later on. Some available evidence also suggests that some false-negatives are likely to

be associated with both pathological and medical conditions in which the infant cannot avoid

during screening.20 For example, while the prevalence of AN/AD varies widely across studies,

rates of up to 15% have been reported among infants with permanent hearing loss or less than

1% among ‘‘at risk’’ population in developed countries.27

Page 63: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

63

Other potential contributors to false-negatives based on evidence from the literature

include progressive or delayed-onset hearing loss possibly from asymptomatic

cytomegalovirus (CMV) as well as acquired hearing loss following the use of ototoxic drugs

after TEOAEs screening.20 In this study, it showed that the false negative rates is lower

compared with false positive rates and supported with data from UNHS. However, both

outcomes are sources of concern in any screening program particularly for ‘‘irreversible’’

conditions like sensorineural hearing loss that is also associated with social stigma.

6.4 The Agreement between two methods

The first important finding was that, though patients’ age ranged from 2 to 100 days,

the results obtained with the two tests were very similar. In our study, out of the 195 high risk

newborns examined, 182 (93.3%) showed agreement between the two techniques, whereas in

13 (6.6%) there was no agreement. The inter-observer agreement was good (kappa = 0.698, p

= <0.001). The inter-observer agreement was described by using Cohen’s k statistics. Results

were expressed as k values and could be classified according to the scale of Landis and Koch

as follows: slight (0.0—0.20), moderate (0.41—0.60), good (0.61—0.80), and excellent

(0.81—1.00).29 In our study, it showed a p value less than <0.001 and was considered

statistically significant, and all p values were based upon two tailed tests.

There was a study which showed the kappa agreement in which in Andrea Melagrana

et al conducted a hearing screening using MB11 BERAphone with ABR in audiologic risk, and

their agreement showed an excellent and statistically significant result (kappa = 0.92 , p =

0.0001).3 Another study done by E. van den Berg et al, showed good kappa agreement (kappa

= 0.709) between MB11 BERAphone and AABR in a pilot study.14

Page 64: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

64

6.5 The Testing Time

For the neonatal period, conventional ABR is considered to be the most reliable method

for assessment of hearing level. The conventional ABR method is not widely used for screening

because it is time consuming and it needs a well qualified technician and audiologist to perform

the test and evaluate the results. Now, a second-generation Automated ABR infant hearing

screener is available for screening purposes. Most of the babies were screened within neonatal

period, which is before hospital discharge in nearly all cases. The time necessary for screening

is supposed to be 4 to 7 minutes for second-generation devices in an inpatient setting.17 In our

study, we were testing the total time spent by staff required to screen the ears each baby using

OAEs and MB11 BERAphone. We are trying to compare the time required between two

methods.

In HLM van Straaten et al, they had done a study about ALGO AABR in neonatal

hearing screening in which they used the second generation AABR. In this study, they noted

that the mean screening time in the NICU setting seems to be somewhat longer than in universal

neonatal hearing screening. Since the introduction of the second-generation AABR screening

devices the mean screening time has been reduced from 26 to 13 minutes, mainly as a result of

a reduction in preparation time.17

In our result, it showed that the total time spent by staff during screening of each high

risk newborn baby were different between OAEs and MB11 BERAphone. The total test time

which included the time for the unit start-up, the preparation of the subject, the measurement,

and the unit shut down. By using OAEs during screening, the median total test time was 2

minutes (IQR: 25th – 75th) as compared to the MB11 BERAphone AABR test with a median of

5 minutes (IQR: 25th – 75th). There were differences of time duration while screening these

high risk newborns using both methods. In this study, we noted that by using OAEs, it required

Page 65: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

65

less time compared to the MB11 BERAphone. Statistically, the comparing of time between

two methods was significant, with p value <0.001. This is important to know to minimise the

time spent by staff for the screening per day.

However, in a study M. Hahn et al, by comparing ALGO AABR and OAEs, they found

that Automated ABR was to be less time-consuming than TEOAEs screening even in the Quick

Screen mode. The median total test time, preparation and follow-up time included, was 9

minutes as compared to TEOAEs screening, with the median of 12 minutes.30

Over the years with the experience of technicians and audiologists, the duration became

less for the hearing screening. A study by Luca Guastini et al evaluated an automated auditory

brainstem response in a multi-stage hearing screening and it shows several advantages of

screening by the AABR with MB11 BERAphone (non-invasive), acoustic stimulus not related

to the external ear canal condition, easy, quick the testing time per infant ranges from 1 to 5

minutes.16 M. Cebulla et all, conducted a study on mean detection time for MB11 BERAphone

but the screening does not include the time required for preparation or documentation. A

maximum test time of 180 seconds was possible for the recording of each test. The test times

showed a median value of 28 seconds and a range of 15 to 112 seconds (5 – 95th percentile).31

Even though the MB11 BERAphone took more time compared to the OAEs, it took

less time duration compared to other AABR and ABR. In Meier et al’s study, they compared

the testing times between OAEs, AABR and also ABR. Based on their study, it showed that

the time to start up the unit was longer for Algo 3 and MB11 BERAphone, both of which also

required a computer, than for Echoscreen. Median preparation and measuring time was shorter

for OAEs measurements compared to AABR recordings. The preparation time for AABR

recordings with the MB11 BERAphone was shorter than the preparation time for AABR with

the Echoscreen and Algo 3, because the MB11 BERAphone does not use electrodes which

Page 66: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

66

need to be fixed separately. The measuring time of AABR with the Algo 3 was longer compared

to the other AABR units. The median time for measurement without preparation was 7 to 13

seconds for the Echoscreen- TDA using OAEs, 68 to 77 seconds using ABR, 50 seconds for

the MB11 BERAphone and 167 seconds for the Algo 3 device.19

It is supported with the study by E. van den Berg et al, the mean MB11 BERAphone

test time was shorter 11.4 minutes (SD 6.6) than with the ALGOTM portable 13.9 minutes (SD

8.1) eventhough it was not significant (p < 0.08) with a maximum difference of 18 minutes.32

However sometimes the testing time depends on a few reasons:

1) It takes a longer duration because the newborns may be aroused and become noisy, so

that either the test became interrupted and had to be repeated later, or much more time

will be needed in order to complete the test.

2) The technician in charge: whether it was performed by otolaryngologists, which is not

a common practice in most countries when testing is performed by individuals with

minimal qualifications.33

3) The most commercially available equipment the recorded parameters are continuously

evident and refreshed on screen, and may be easily estimated in order to stop the

examination manually.33 This would result in significant time saving and improving the

overall efficiency of the screening program.

So we believe that if the test is performed by a professional as soon as possible with modern

methods, the average recording time is further reduced, because the newborns are less likely to

be aroused and become noisy, interfering with the test procedure.

Page 67: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

67

6.6 LIMITATION OF STUDY

1) The sample of study may not be big enough to get a better result.

2) Almost 50% of patients who yield the result “refer” during hearing screening using

OAEs and MB11 BERAphone were loss to follow up for second screening by using

gold standard ABR. There were a few reasons:

a. Loss of their contact number. We gave their appointments within 3 months, but

during that time the parent already changed their phone number.

b. One patient passed away in the ward before discharge.

c. The mothers were in confinement for about 1 or 2 months in Kelantan, then

after completion, they returned to their homes in other states out of Kelantan

and continue follow up in other hospital.

d. Some parents refused to let their baby undergo a second hearing screening due

to uncertain reasons.

3) The total testing time spent using both methods can be biased if the person in charge of

the screening is not familiar with both methods, especially the MBII BERAphone.

Page 68: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

68

7. CONCLUSION

Page 69: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

69

7. CONCLUSION

There were 195 newborns (87, 44.6% boy and 108, 55.4% girl) who participated in this

study. Ototoxic medication was the most common risk factor (51.8%) followed by

hyperbilirubinaemia (51.3%) and low birth weight (27.2%). Usage of the MB11 BERAphone

had a higher passing rate (89.8%) as compared with the OAEs (85.2%). Usage of the MB11

BERAphone had a lower refer rate (10.2%) compared to the OAEs (14.8%). These differences

are statistically significant. The true negatives are MB11 BERAphone (29.4%) and OAEs

(11.8%). In the presence of false negative in MB11 BERAphone, this will give provide

significant impact to this machine concerning the effectiveness for hearing screening. Out of

195 newborns examined, 182 (93.4%) showed agreement between the two techniques, whereas

13 (6.6%) there was no agreement. Inter-observer agreement was good (kappa = 0.698,

p=<0.001). The median test time conducted on each newborn using MB11 BERAphone was 5

minutes (IQR: 3-5 minutes) and OAE was 2 minutes (IQR: 2-3 minutes). The differences were

statistically significant (p = <0.001).

The MB11 BERAphone is a rather reliable device for auditory brainstem response

among high risk newborns hearing screening. Both agreements were good. However, the

duration for hearing screening for each newborn significantly took a longer time compared

OAEs. Therefore, we recommend that both methods are to be used as a first screening, followed

by a screening via ABR, to those whose result was “refer” from first screening.

Page 70: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

70

8. RECOMMENDATIONS

1) Suggestion of a larger sample population of high risk newborns for comparison of

OAEs and MB11 BERAphone to obtain a better result.

2) To get equal distributions of sample among risk factors newborns to reduce bias of the

result.

3) Suggestion of conducting a test for sensitivity and specificity between OAEs and MB11

BERAphone via using the gold standard ABR.

4) To reduce limitation of study during second hearing screening by using ABR as a gold

standard, my suggestion is to get a more detailed demographic data of the patient, with

contact numbers of both parents and to provide the appointment during the mother’s

confinement period to minimize loss of follow up patient.

Page 71: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

71

8.REFERENCES

Page 72: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

72

REFFERENCES

1. Abdullah, A., Hazim, M.Y.S., Almyzan, A., Jamilah, A.G., Roslin, S., Ann, M.T.,

Borhan, L., Sani, A., Saim, L., Boo, N.Y. (2006). Newborn hearing screening :

experience in a Malaysian hospital. Singapore Med J 47, 60–64.

2. Noor, S.W., M.Khairi M.D., Sidek, D., A.Rahman, N., Mansord, S., M.Normani, Z.

(2012). The performance of distortion product otoacoustic emissions and automated

auditory brainstem response in the same ear of the babies in neonatal unit. International

Journal of Pediatric Otorhinolaryngology 76, 1366–1369.

3. Melagrana, A., Casale, S. & Grazia, M. (2007). MB11 BERAphone and Auditory

Brainstem Response in newborns at audiologic risk : Comparison of results.

International Journal of Pediatric Otorhinolaryngology 71, 1175–1180.

4. Calcutt, T. L., Dornan, D., Beswick, R. & Tudehope, D. I (2016). Newborn hearing

screening in Queensland 2009 – 2011 : Comparison of hearing screening and diagnostic

audiological assessment between term and preterm infants. Journal of Pediatrics and

Child Health 52, 995–1003.

5. Fang, X., Li, X., Zhang, Q., Wan, J., Sun, M., Chang, F., Lu, J., Chen, G. (2016).

Universal neonatal hearing screening program in Shanghai , China : An inter-regional

and international comparison. International Journal of Pediatric Otorhinolaryngology

90, 77–85.

6. Kamal, N. (2013). Newborn hearing screening : Opportunities and challenges. Egyptian

Journal of Ear, Nose, Throat and Allied Sciences 14, 55–58.

7. Chobaut, J., Tavernier, L., Ohl, C. & Dornier, L. (2009). Newborn hearing screening on

infants at risk. International Journal of Pediatric Otorhinolaryngology 73, 1691–1695.

8. Michael Cunningham, M. E. O. C. (2003). Hearing Assessment in Infants and Children:

Recommendations Beyond Neonatal Screening. Official Journal Of The American

Academy Of Pediatrics 111, 436–440.

Page 73: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

73

9. Village, E. G. (2007). Year 2007 Position Statement : Principles and Guidelines for

Early Hearing Detection and Intervention. Official Journal Of The American Academy

Of Pediatrics 898–920.

10. R. Mota., M. Carvalloa., S.Gabriela, G. Sanchesb., S. Maria, I., J, Costa, S., A. Spada,

D. (2015). Efferent inhibition of otoacoustic emissions in preterm neonates ଝ. Brazilian

Journal of Otorhinolaryngology 81, 491–497.

11. Apostolopoulos, N. K., Psarommatis, I. M. & Tsakanikos, M. D. (1999). Otoacoustic

emission-based hearing screening of a Greek NICU population. International Journal

of Pediatric Otorhinolaryngology 47, 41–48.

12. Smyth, V., Mc Pherson, B., Kei, J., Tudehope, D., Maurer, M., Rankin, G. (1999).

Otoacoustic emission criteria for neonatal hearing screening. International Journal of

Pediatric Otorhinolaryngology 48, 9–15.

13. Wroblewska-seniuk, K., Chojnacka, K., Pucher, B., Szczapa, J., Gadzinowski,

Grzegorowski, M. (2005). The results of newborn hearing screening by means of

transient evoked otoacoustic emissions. International Journal of Pediatric

Otorhinolaryngology 69, 1351–1357.

14. Berg, E. Van Den, Deiman, C. & Straaten, H. L. M. Van. (2010). MB11 BERAphone 1

hearing screening compared to ALGO TM portable in a Dutch NICU : A pilot study.

International Journal of Pediatric Otorhinolaryngology 74, 1189–1192.

15. Rajkumar, C., Augustine, A. M., Lepcha, A. & Balraj, A. (2016). Automated auditory

brainstem response : Its efficacy as a screening tool for neonatal hearing screening in the

postnatal ward. 22, 237–242.

16. Guastini, L., Mora, R., Dellepiane, M., Santomauro, V., Mora, M., Rocca, A., Salami,

A. (2010). Evaluation of an automated auditory brainstem response in a multi-stage

infant hearing screening. Eur Arch Otorhinolaryngol 267, 1199–1205.

Page 74: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

74

17. HLM van Straaten Isala. (1999). Automated auditory brainstem response in neonatal

hearing screening. 432, 76–79.

18. J.I. Benito-Orejas, B. R., ´rez, D. M., A. Almaraz, J. L. F. & ´ndez-Calvo. (2008).

Comparison of two-step transient evoked otoacoustic emissions ( TEOAE ) and

automated auditory brainstem response ( AABR ) for universal newborn hearing

screening programs. International Journal of Pediatric Otorhinolaryngology 72,

1193—120.

19. Meier, S., Narabayashi, O., Probst, R. & Schmuziger, N.(2004). Comparison of

currently available devices designed for newborn hearing screening using automated

auditory brainstem and / or otoacoustic emission measurements. International Journal

of Pediatric Otorhinolaryngology 68, 927–934.

20. Olusanya, B. O. & Bamigboye, B. A. (2010). Is discordance in TEOAE and AABR

outcomes predictable in newborns ? International Journal of Pediatric

Otorhinolaryngology 74, 1303–1309.

21. Konukseven, O., Genc, A., Muderris, T., Kayikci, K, M., Turkyilmaz, D., Ozturk, B.,

Gunduz, B. (2010). Can Automated Auditory Brainstem Response be Used as an Initial

Stage Screening Test in Newborn Hearing Screening Programs. 231–238.

22. Dyk, M. Van, Swanepoel, D. W. & Iii, J. W. H. (2015). Outcomes With OAE And

AABR Screening In The First 48 Hours–Implications For Newborn Hearing Screening

In Developing Countries. International Journal of Pediatric Otorhinolaryngology 1–25.

23. Johnson, J. L., White, K. R., Widen, J. E., Gravel, J. S., James, M., Kennally, Maxon,

A. B., Spivak, L., Mahoney, M. S., Vohr, B. R., Weirather, Y., Holstrum, J., Johnson, J.

(2013). A Multicenter Evaluation of How Many Infants With Permanent Hearing 663-

672

24. A. Zamani, K. Daneshjou, A. A. and J. T. (2002). Estimating The Incidence Of Neonatal

Hearing Loss In High Risk Neonates. 42, 176–180.

Page 75: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

75

25. Kock, T. De, Swanepoel, D. & Hall, J. W. (2016). International Journal of Pediatric

Otorhinolaryngology Newborn hearing screening at a community-based obstetric unit :

Screening and diagnostic outcomes. International Journal of Pediatric

Otorhinolaryngology 84, 124–131.

26. Iwasaki, S., Hayashi, Y., Seki, A. & Nagura, M. (2003). A model of two-stage newborn

hearing screening with automated auditory brainstem response. International Journal of

Pediatric Otorhinolaryngology 67, 1099–1104.

27. Kirkim, G., Serbetcioglu, B. & Erdag, T. K. (2008). The frequency of auditory

neuropathy detected by universal newborn hearing screening program. 72, 1461–1469.

28. Smets, K., Verrue, N. & Dhooge, I. (2012). Implementation and results of bedside

hearing screening with automated auditory brainstem response in the neonatal intensive

care unit. Acta Pædiatrica ISSN 0803–5253 101, 392–398.

29. Landis, J. R. & Koch, G. G. (1977). The Measurement of Observer Agreement for

Categorical Data Data for Categorical of Observer Agreement The Measurement. 33,

159–174.

30. Hahn, M., Lamprecht-dinnesen, A., Heinecke, A., Hartmann, S. & Bu, S. (1999).

Hearing screening in healthy newborns : feasibility of different methods with regard to

test time. International Journal of Pediatric Otorhinolaryngology 51, 83–89.

31. Cebulla, M. & Shehata-dieler, W. (2012). ABR-based newborn hearing screening with

MB11 BERAphone 1 using an optimized chirp for acoustical stimulation. International

Journal of Pediatric Otorhinolaryngology 76, 536–543.

32. E.van den Berg, C.Deiman, H. L. . van S. (2010). MBII BERAphone hearing screening

compared to ALGO portable in a Dutch NICU: A Pilot study. 1189–1192.

33. S.G. Korres, D.G. Balatsouras, P. Kanellos, A., Georgiou, V. K. & E. F. (2004).

Decreasing test time in newborn hearing screening. 29, 219–225.

Page 76: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

76

10. APPENDICES

Page 77: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

77

Appendix 1

A Comparison of Automated Auditory Brainstem Response (AABR) with Integrated

Electrodes and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing Screening

Source population: (All high risk newborn in neonatal unit

HUSM)

Sample, n: 195 (Inclusion and exclusion criteria)

Both OAEs and MB11 BERAphone

Refer one or both ears Pass both ears

ABR Total time of screening

Data collection and analysis

True positive False positive

Page 78: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

78

Appendix 2 DATA COLLECTION

CODE:

Date:

Time:

Place:

Risk Factors:

Family history In utero infection

Birth weight <1500g Apgar score

Ototoxic medication Mechanical ventilation >5days

Hyperbilirubinemia Craniofacial anomalies

Syndromic baby Meningitis

Instrument:

1) Otoacoustic Emission (OAEs)

Time:

Time start:

Time end:

o Total testing time :

Result:

Pass Refer:

2) MB11 BERAphone

Time:

Time start:

Time end:

o Total testing time :

Result:

Pass Refer:

ABR:

Normal Positive

Page 79: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

79

Appendix 3

JAWATANKUASA ETIKA PENYELIDIKAN (MANUSIA) - JEPeM

HUMAN RESEARCH ETHICS COMMITTEE - HREC

________________________________________________________________________________

Consent Form

LAMPIRAN A

MAKLUMAT KAJIAN

Tajuk Kajian: Comparison of Automated Auditory Brainstem Response (AABR) with Integrated

Electrodes and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing

Screening

Nama Penyelidik:

Profesor Madya Dr Salmi Abd Razak

Profesor Dr Hans Van Rostenberghe

Dr Muhammad Noor Azmi Ramli

No. Pendaftaran MMC : 48601

PENGENALAN

Anda dipelawa untuk menyertai satu kajian penyelidikan secara sukarela yang melibatkan dua jenis

alat kajian: Automated Auditory Brainstem Response (AABR) – BERAphone dan Otoacoustic

Emissions (OAEs). Kedua – dua peralatan ini adalah di gunakan untuk mengenal pasti masalah

pendengaran di kedua – dua belah telinga pesakit. Ia di gunakan bagi membolehkan masalah

pendengaran dapat di kenal pasti terlebih awal. Kedua dua peralatan ini bagus di gunakan untuk

saringan pendengaran supaya pesakit mendapat saringan pendengaran dengan lebih efektif. Sebelum

anda bersetuju untuk menyertai kajian penyelidikan ini, adalah penting anda membaca dan memahami

borang ini. Sekiranya anda menyertai kajian ini, anda akan menerima satu salinan borang ini untuk

disimpan sebagai rekod anda.

Penyertaan anda di dalam kajian ini dijangka mengambil masa sehingga 12 minggu. Seramai 195

pesakit akan menyertai kajian ini.

TUJUAN KAJIAN

Kajian ini bertujuan adalah untuk menentukan sama ada, semasa tempoh kajian selama 12 minggu,

saringan pendengaran yang menggunakan peralatan BERAphone dan OAEs terhadap pesakit adalah

supaya:

Pesakit mendapat saringan pendengaran dengan lebih efektif lagi supaya tiada sebarang

kesulitan berlaku

Bagi mengelakkan pesakit yang mengalami masalah pendengaran tidak terlepas dari

saringan

BORANG ETIKA - 02

Page 80: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

80

Terdapat kemungkinan maklumat yang dikumpulkan semasa kajian ini akan dianalisa oleh pihak

penyelidik pada masa depan untuk menilai BERAphone dan OAEs untuk kegunaan lain yang mungkin

atau untuk tujuan perubatan atau saintifik lain yang selain dari yang kini dicadangkan.

KELAYAKAN PENYERTAAN

Doktor yang bertanggungjawab dalam kajian ini atau salah seorang kakitangan kajian telah

membincangkan kelayakan untuk menyertai kajian ini dengan anda. Adalah penting anda berterus

terang dengan doktor dan kakitangan tersebut tentang sejarah kesihatan anak anda. Anak anda tidak

seharusnya menyertai kajian ini sekiranya anak anda tidak memenuhi semua syarat kelayakan.

Beberapa keperluan untuk menyertai kajian ini adalah –

Anak yang baru lahir

Anak anda adalah seorang pesakit yang di rawat di unit bayi “noenatal unit”

Anak anda mempunyai risiko masalah pendengaran seperti:

o Masalah kekuningan “jaundice”

o Lahir tidak cukup matang

o Sejarah masalah pendengaran dalam keluarga

o Risiko jangkitan kuman dalam kandungan

o Risiko kesan daripada ubat yang menyebabkan masalah pendengaran

o Jangkitan di selaput otak

o Lahir dengan rendah “Apgar Score”

o Terdapat masalah struktur muka yang tidak normal

o Memerlukan bantuan mesin pernafasan lebih dari 5 hari

o Terdapat struktur yang tidak normal sejak lahir yang mempunyai risiko untuk masalah

pendengaran

Anak anda tidak boleh menyertai kajian ini sekiranya –

Sekiranya anak anda lahir dengan sihat dan cukup bulan yang tiada risiko untuk mendapat

masalah pendengaran

PROSEDUR-PROSEDUR KAJIAN

Kajian dengan menggunakan peralatan BERAphone dan OAEs akan dilakukan bila pesakit mempunyai

risiko untuk mendapat masalah pendengaran yang di masukkan ke dalam wad bayi. Saringan

pendengaran akan di lakukan setelah pesakit yang telah di rawat di wad di sahkan sihat dan boleh di

discaj dari hospital. Pada hari pesakit di discaj dari hospital saringan pendengaran akan di lakukan oleh

staf – staf yang berpengalaman.

RISIKO

Kajian ini sangat mudah dan senang di lakukan dan tidak mempunyai banyak masalah komplikasi ke

atas pesakit. Peralatan yang di buat saringan ke atas pesakit akan di bersihkan terlebih dahulu sebelum

saringan di lakukan bagi mengelakkan sebarang jangkitan luar berlaku.

Jika apa-apa maklumat penting yang baru dijumpai semasa kajian ini yang mungkin mengubah

persetujuan anda untuk terus menyertai kajian ini, anda akan diberitahu secepat mungkin.

Page 81: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

81

MELAPORKAN PENGALAMAN KESIHATAN

Jika anak anda mengalami apa-apa kecederaan, kesan buruk, atau apa-apa pengalaman kesihatan

yang luar biasa semasa kajian ini, pastikan anda memberitahu Dr Muhammad Noor Azmi Ramli (No.

Pendaftaran Penuh Majlis Perubatan Malaysia: 48601) di talian 017-6849513 secepat mungkin. Anda

boleh membuat panggilan pada bila-bila masa, siang atau malam, untuk melaporkan pengalaman

sedemikian.

PENYERTAAN DALAM KAJIAN

Penyertaan anak anda dalam kajian ini adalah secara sukarela. Anda berhak menolak untuk menyertai

kajian ini atau anda boleh menamatkan penyertaan anda pada bila-bila masa, tanpa sebarang hukuman

atau kehilangan manfaat yang sepatutnya anda perolehi.

Penyertaan anak anda juga mungkin boleh diberhentikan oleh doktor yang terlibat dalam kajian ini

tanpa persetujuan anda. Sekiranya anak anda berhenti menyertai kajin ini, doktor yang terlibat di dalam

kajian ini atau salah seorang kakitangan akan berbincang dengan anda mengenai apa-apa isu

perubatan berkenaan dengan pemberhentian penyertaan anak anda.

MANFAAT YANG MUNGKIN [Manfaat terhadap Individu, Masyarakat, Universiti]

Prosedur kajian ini akan diberikan kepada anak anda tanpa kos. Anda mungkin menerima maklumat

tentang kesihatan anak anda daripada pemeriksaan fizikal dan ujian makmal yang dilakukan dalam

kajian ini. Hasil atau maklumat kajian ini diharapkan, dapat memberi manfaat kepada pesakit-pesakit

pada masa hadapan. Anak anda tidak akan menerima sebarang pampasan kerana menyertai kajian

ini. Namun sebarang keperluan perjalanan berkaitan dengan penyertaan ini akan diberi.

PERSOALAN

Sekiranya anda mempunyai sebarang soalan mengenai prosedur kajian ini atau hak-hak anda, sila

hubungi;

<Dr Muhammad Noor Azmi Ramli > & <No.MMC: 48601>

<Jabatan Pediatrik>

<Pusat Pengajian Sains Perubatan>

<USM Kampus Kesihatan>

<No Tel: 09- 7673000><017-6849513>

Sekiranya anda mempunyai sebarang soalan berkaitan kelulusan Etika atau sebarang pertanyaan dan

masalah berkaitan kajian ini, sila hubungi;

Puan Mazlita Zainal Abidin

Setiausaha Jawatankuasa Etika Penyelidikan (Manusia) USM

Pusat Inisiatif Penyelidikan -Sains Klinikal & Kesihatan

USM Kampus Kesihatan.

No. Tel: 09-767 2355 / 09-767 2352

Email : [email protected]

Page 82: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

82

KERAHSIAAN

Maklumat perubatan anda akan dirahsiakan oleh doktor dan kakitangan kajian. Ianya tidak akan

dedahkan secara umum melainkan jika ia dikehendaki oleh undang-undang.

Data yang diperolehi dari kajian yang tidak mengenalpasti anda secara perseorangan mungkin akan

diterbitkan untuk tujuan memberi pengetahuan baru.

Rekod perubatan anda yang asal mungkin akan dilihat oleh pihak penyelidik, Lembaga Etika kajian ini

dan pihak berkuasa regulatori untuk tujuan mengesahkan prosedur dan/atau data kajian klinikal.

Maklumat perubatan anda mungkin akan disimpan dalam komputer dan diproses dengannya.

Dengan menandatangani borang persetujuan ini, anda membenarkan penelitian rekod, penyimpanan

maklumat dan pemindahan data seperti yang dihuraikan di atas.

TANDATANGAN

Untuk dimasukkan ke dalam kajian ini, anda atau wakil sah anda mesti menandatangani serta

mencatatkan tarikh halaman tandatangan (Borang Keizinan Pesakit di LAMPIRAN S dan LAMPIRAN

P).

Page 83: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

83

LAMPIRAN S

Borang Keizinan Pesakit/ Subjek

(Halaman Tandatangan)

Tajuk Kajian: Comparison of Automated Auditory Brainstem Response (AABR) with Integrated

Electrodes and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing

Screening

Nama Penyelidik: Profesor Madya Dr Salmi Abd Razak (No Pendaftaran Penuh:30493) Profesor Dr Hans Van Rostenberghe (No Pendaftaran Penuh:31042) Dr Muhammad Noor Azmi Ramli (No Pendaftaran Penuh: 48601) Untuk menyertai kajian ini, anda atau wakil sah anda mesti menandatangani mukasurat ini. Dengan

menandatangani mukasurat ini, saya mengesahkan yang berikut:

Sa ya te lah m em baca sem ua m ak lum at da lam Borang Mak lumat dan Ke iz inan Pesak i t i n i t e rm asuk apa-apa m ak lum at berka i tan r is iko yang ada da lam ka j ian dan saya te lah pun d iber i m asa yang m encukup i un tuk m em per t im bangkan mak lum at te rsebut .

Sem ua soa lan -soa lan saya te lah d i jawab dengan m em uaskan. Sa ya, secara sukare la , berse tu ju m enyer ta i ka j ian pen ye l id ikan in i ,

m em atuh i sega la p rosedur ka j ian dan m em ber i m ak lum at yang d iper lukan kepada dok tor , para ju rurawat dan juga kak i tangan la in yang berka i tan apab i la d im in ta .

Sa ya bo leh m enam atkan penyer taan sa ya da lam ka j ian in i pada b i la -b i l a m asa.

Sa ya te lah pun m ener im a sa tu sa l inan Borang Mak lum at dan Ke iz inan Pesak i t un tuk s im panan per ibad i saya .

Nama Pesakit Nama Singkatan & No. Pesakit

No. Kad Pengenalan Pesakit (Baru) No. K/P (Lama)

Tandatangan Pesakit atau Wakil Sah Tarikh ( )

(Masa jika perlu)

Page 84: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

84

Nama & Tandatangan Individu yang Mengendalikan Tarikh ( )

Perbincangan Keizinan

Nama Saksi dan Tandatangan Tarikh (dd/MM/yy)

Nota: i) Semua subjek/pesakit yang mengambil bahagian dalam projek penyelidikan ini tidak dilindungi insuran.

Page 85: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

85

LAMPIRAN P

Borang Keizinan bagi Penerbitan Bahan yang berkaitan dengan Pesakit/ Subjek

(Halaman Tandatangan)

Tajuk Kajian: Comparison of Automated Auditory Brainstem Response (AABR) with Integrated

Electrodes and Otoacoustic Emissions (OAEs) in High Risk Newborn Hearing

Screening

Nama Penyelidik: Profesor Madya Dr Salmi Abd Razak (No Pendaftaran Penuh:30493) Profesor Dr Hans Van Rostenberghe (No Pendaftaran Penuh:31042) Dr Muhammad Noor Azmi Ramli (No Pendaftaran Penuh: 48601) Untuk menyertai kajian ini, anda atau wakil sah anda mesti menandatangani mukasurat ini.

Dengan menandatangani mukasurat ini, saya memahami yang berikut:

Bahan yang akan d i t e rb i tkan tanpa d i lam pi rkan dengan nam a saya dan se t iap percubaan yang akan d ibuat un tuk m em ast ikan ke tanpanam aan saya. Sa ya m em aham i , wa laubaga im anapun, ke tanpanam aan yang sem purna t idak dapat d i j am in . Kem ungk inan ses iapa yang m enjaga saya d i hosp i ta l a tau saudara dapat m engena l i saya.

Bahan yang akan d i te rb i tkan da lam penerb i tan m ingguan/bu lanan/dw ibu lanan/suku tahunan/dwi tahunan m erupakan sa tu pen yebaran yang luas dan te rsebar ke se luruh dun ia . Kebanyakan penerb i tan in i akan te rsebar kepada dok tor -dok tor dan juga bukan dok tor te rm asuk ah l i sa ins dan ah l i j u rna l .

Bahan te rsebut j uga akan d i lam pi rkan pada lam an web ju rna l d i se lu ruh dun ia . Sesetengah lam an web in i bebas d ikun jung i o leh sem ua orang.

Bahan te rsebut j uga akan d igunakan sebaga i penerb i tan te m patan dan d isam paikan o leh ramai dok tor dan ah l i sa ins d i se luruh dun ia .

Bahan te rsebut j uga akan d igunakan sebaga i penerb i tan buku o leh penerb i t j u rna l .

Bahan te rsebu t t idak akan d igunakan untuk peng ik lanan a taupun bahan untuk m em bungkus .

Sa ya juga m em ber i ke iz inan bahawa bahan te rsebut bo leh d igunakan sebaga i

penerb i tan la in yang d im in ta o leh penerb i t dengan k r i te r ia ber iku t :

Bahan te rsebu t t idak akan d igunakan untuk peng ik lanan a tau bahan untuk m em bungkus .

Bahan te rsebut t idak akan d igunakan d i luar kon teks – contohn ya: Gam bar t idak akan d igunakan untuk m enggam barkan sesuatu ar t ike l yang t idak berka i tan dengan sub jek da lam fo to te rsebut .

Page 86: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

86

Nama Pesakit Nama Singkatan atau No. Pesakit

No. Kad Pengenalan Pesakit T/tangan Pesakit Tarikh ( )

Nama & Tandatangan Individu yang Mengendalikan Tarikh ( )

Perbincangan Keizinan

Nota: i) Semua subjek/pesakit yang mengambil bahagian dalam projek penyelidikan ini tidak dilindungi insuran.

Page 87: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

87

Appendix 3

Jw\•atankul\•• l:ltlka l'eny~Jldlkan Manutla USM (J£PtM) tlunuon lte- &aroh Ethic, Comnoittoo USM (lllUtC)

Our,l\el. : Oot•

IJSM/JEPo~/278.3.(5) t)~' Mor~h ~0~4

Or. Muhammad Noor Alml R1mll Oeportmem of Pedlwles School of Medlc_ol S~~I\CQS Unwe($jtl S~lns Ma!IIV$la l6~~0 (ul>\lng Kerian, ~eiant;tn.

Unhersld &al ... ~~· K•mf'lu• K'"11¥~an. 101.50 K\fbin((>K&-.,.n. K•lbun. M•tayab. '1"1 Ci09 • 'fti1 $900 lllMh. ~5" /t!S~

...-. ~·1(!7 ·~·· &)!l,.m . .,[email protected] www.rcte•rch.wm.my

Tllc H\Jm&n Rtsearch EthiO$ Committee, Uollver$itl Salns Malayiia (FWA Reg. No: 00007718; IRB ll.es. No: 00004494) h•s •pproved In principle the study mention-ed below:

_j eo,n..,~son of Automat<# ~ Btairut~ Response (AABft) will\ ln"'8'•ted EliWO<Ie$ and I OtQiiCOU\tio Em~J!OAEslln H1811 R~Si(New'born ~·~f'ISCteetWlB- _I

ProUKol No I Prlndplo tnvestlptor J Or. M""""mad Noor .WOI Ra!nr. j Date. of approval Mc«01r~~~ ltcvoe\\'e'd :rf ~~mn•I!CW. Jl.tttlv~Rrtell11f<l P(~

n"' Mo•d,J 2014 1~~r1011 ,..-J~!f%014 II"Mmh'-OJ4

~~~~::.:.~nvenR~ J Or. Moh<j NOfll)OnlbiW!a Or. Npuarwarry~ad

-- --Hooplt.>J v • .,...,,d Malll)'1la.

SOln$ 1 OllO'OhtudySialt I Mat<I\2014-August20_~_· ___ __.1

I-~-M_nd_•_'~-port---~---------'-N-•_m_b•_~_sa_m_~_~l~b~------------J' Tl\e following Item Vl have been reoeivtland reviewed:·

( .1') EthiCal Appro"'! Application Form ("') R"st>arch Proposal (v') Parents/Guardian Information Slfeet and Consent Form ('/) Data Collection Eorm

lnvestigaror(s) ar<~required to: a) follow instructions, guidelines and rcquireme11ts of tlte Human Researdl Ethics Comndttl!li;·'tl\11!

Salns Malaysia (JEReM) b) report any protocql deviations/violations to ·Humall Research Ethics Committee (JEPeM) c) comply with International Conference on Harmonization- Guidelines forGood:CIIrlleai·Pt\Jctl*'

and the eclaratlon ofHelslnkl d) notetha Humar. R6~earcl\ [thlcs Committee (JEP.eM) may a_udit the approved study.

PROFESSOR DR. MOHD$HUKRI OTHMAN Deputy Chairma.o ttuman Research EthiC$ Committee

Page 88: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

88

Appendix 4

4 Jun 2014

UNIVEHSm SAINS MALAYSIA

Pharmanlaga Logistics Sdn. Bhd. No. 7, Lorong Kelull18 Kawasan Perlndustrian Bukit Raja Selatan P.O Box 2030 Pusat Blsnes Buklt Raja 40000 Shah Alam Selangor (UIP: Enclk Masri Muhammad Vusof)

'fuan

I Jaba~n Pediatrik Depatiment ol Paedlatda

I jab;u:ut Pcdiatrik

Kampus Kesihatan UniveBiti Sains Malaysia 16150 Kubang Kerillro. l(elantan. Malay!ia. Tel:6097673000 EJ<t !6563 www.me<li<wsm.my/-paed .. ttiesl

Surat Pengesahan Plnjaman Peralatan MAICO Beraphone MB11 AABR Newbol:n Hearing Screening

Perkara di alas adalah dirujuk.

2. Unluk makluman tuan, Dr. Muhammad Noor Azmi bin Ram8 (No.J<P: 03-5967) merupakan pelajar Sarjana Perubatan (Pediatrlk) Unjverslli Kampus Kesihatan Kelanfan S:edang menjalankan penyelldikan Comparison of Automated Auditory Bra.instein Response !AJ>.Bf~J

Electrodes And Otocoustic Emissions In High Risk Newborn Hearing S~=~t~ ini sedang menggunakan Peralatan MAICO Beraphone MB11 AA8R 1

Screening yang dipinjamkan oleh pioak Pharmaniaga selama 6 bulan mullli·lllii hingga September 2014.

Sekian untuk makluman plhak tuan.

Terlma kasih.

"BERKHIDMAT UNTUK NEGARA" Memastikan Kelestarian Hari Esok,

{;· 'c PROF. MADYA D~.J~RIZAN H A MAJID Pemangku Ketua Jaliatan Pedialrik Pusat Pengajjan Sains Perubatan Universiti Sains Malaysia Kampus Kesihatan

Page 89: A COMPARISON OF AUTOMATED AUDITORY DR. MUHAMMAD …

89

Appendix 5

Gantt Chart of Research Activities: Comparison of Automated Auditory Brainstem Response

(AABR) with Integrated Electrodes and Otoacoustic Emissions (OAEs) in High Risk Newborn

Hearing Screening

PROJECT ACTIVITIES

2014

2017

Research Activities J F M A M J J A S O N D J F M A M J J A S O N D

Staff training

Participants

recruitment and

intervention

Data Analysis /

Interpretation

Writing final report

Presentation

Writing and submit

journal papper

Milestone of Research Activities:

1. February 2014 untill Mac 2014: To train the staff using MB11 BERAphone and OAEs before

data collection

2. April 2014 untill December 2014: Completion of Patient Recruitment and Data Collection

3. January 2017 untill April 2017: Completion of Data Analysis and interpretation

4. May 2017 untill November 2017: Completion of writing and submit journal paper, presentation

and final report


Recommended