International Workshop on Aging Phenomena in Gaseous Detectors October 2-5, 2001 at DESY in Hamburg A Longterm Aging Study of the Honeycomb Drift Tubes of HERA-B Using a Circulated and Purified CF -Gas Mixture 4 M. Cape ns , K. Dehmelt , M. Hohlmann , B. Schmidt a c b b a b c CERN (Geneva) DESY (Hamburg) University of Mainz á Wed, Oct 03, 01 Klaus Dehmelt
Transcript
Page 1: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

A Longterm Aging Study of the Honeycomb Drift

Tubes of HERA-B Using a Circulated and Purified

CF -Gas Mixture4

M. Cape ns , K. Dehmelt , M. Hohlmann , B. Schmidta c b b

a

b

c

CERN (Geneva)DESY (Hamburg)

University of Mainz

á

Wed, Oct 03, 01

Klaus Dehmelt

Page 2: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01

Motivation for agingstudy:

gas-systemwith CF -gas mixture for

the

Need to test purifiers

4

OTR

Monitoring of pollutants

Longterm-behaviour ofHoneycomb Drift Tubes forthe OTR

?

?

?

?

Recirculating

YSimulation of

the OTR and itsgas-system

Klaus Dehmelt et al.

Page 3: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01

Honeycomb Drift Tube(HDC)

of the OTR

?

?

Pocalon, 75 mCoated with

Cu & Au

Signal wires,

Au/W, 25 m

m

m

Klaus Dehmelt et al.

Endpiece

Signal Wires

G10-strips

Single Layer Double Layer

Pocalon Foil

5mm / 10mm

Page 4: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

4050

6070

8090

100

1.7 1.8 1.9 2 2.1

Hochspannung [kV]

η Sp

ur

[

%]

65% / 35% / 0%

65% / 32% / 3%

65% / 30% / 5%

65% / 28% / 7%

Gas: Ar / CF

4 / CO

2

R.P

ern

ack

U

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01

Counting Gas forthe OTR

?

?

?

?

?

Interaction rate:

needs

Gas mixture:Ar:CF :CO ( )

due to

40MHz

65:30:54 2

Drift velocity:

(4-6kV/cm)140 m / nsm

Price of CF :

Gas volume of OTR

4

~50 � / kg

22 m

:

~3 Y

Recirculatinggas system

4050

6070

8090

100

1.7 1.8 1.9 2 2.1

Hochspannung [kV]

η Hi

t

[%

]

65% / 35% / 0%

65% / 32% / 3%

65% / 30% / 5%

65% / 28% / 7%

Gas: Ar / CF

4 / CO

2

98 %

R.P

ern

ack

U

Klaus Dehmelt et al.

Page 5: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01

? Recirculating gas-system:

Air (leaks)Plasma chemistryproducts ?(Processes inavalanche,materialoutgassing,crackproducts)

ACCUMULATION !Needs purification.

Y

Klaus Dehmelt et al.

The Purifiersfor the OTR

6

6

R3-11G Catalyst

Molsieve 3A

:

Removal of oxygen46% Cu (highly dispersive)

Regenerative

:

Removal of H O

(and other pollutants?)“Inverse sieve”

2Cu + O 2CuO2 B

2

Page 6: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Klaus Dehmelt et al. Wed, Oct 03, 01

Experimental Setup forAging Test

Honeycomb :

drift tubes, each of length

:

Cu-anode, running atBeam spot size ~ diameter

Customized :

Operation in ‘ ’ & ‘ ’Regulated chamber pressure (no bubbler)Monitoring main & trace gas components

Removal of oxygen & water

Electronic :

FADC, ADC, Currentmeter

16 40cm

20kV12cm

open-loop closed loop

test module

X-ray source

gas system

readout system

Page 7: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

Frontend-electronics

Gas inlet

Gas outlet

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

EquipmentTest Modules & DAQ

HDC-test-modules:2 x 16

twochannels,

coated pocalon foils

DAQ

Page 8: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Equipment - Gas Supply

PID-flowregulated,

gas system,chambercovered

withMylar-foil

very clean

Page 9: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

EquipmentGas Monitoring

Gas ChromatographMass spectrometer

Electron Capture Detector

oxygen meter

dewpoint sensor

HP5890,+

with individual chromatography columns

Quality of Counter gasPollutants (if any)

Commercial

Commercial

C

Page 10: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Aging Test Procedures

?

?

?

?

?

?

?

Examination of module functionality

Irradiation of half circle area ( 12cm)

Readout of 3 wires (PH-spectra)

Current monitoring

Monitoring of T, p

Monitoring of gas quality

HERA-B OTR gas-conditions:RecirculationGas purificationFresh gas ~ 2% / h

øy

g

Monitoring ofPH-spectra

Page 11: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Aging test: Gas GainDevelopment I

Relativeraw gas gaindevelopmentand current

for all periods

Bre

ak

Break

1.) No Purification,open loop

2.) Molsieve,closed loop

3.) Molsieve + R3,closed loop

!

!

3.)1.) 2.)

Page 12: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Aging test: Gas GainDevelopment II

Not saturated molsieve

gas-compositionchanges

0,6

0,7

0,8

0,9

1

1,1

Aug/3 Aug/4 Aug/5 Aug/6 Aug/7 Aug/8

0,6

0,7

0,8

0,9

1

1,1

1,2

1,3

1,4

Jul/23/1999 Jul/25/1999 Jul/27/1999 Jul/29/1999 Jul/31/1999

Date

SWPC ---

Wire No. 12 ---

Rel

. G

as

Gain

Molsieve

in

loop

Molsieve

removed

SWPC

DateR

el. G

as

Gain

Run without purification:

Correlation of HDCand SWPC

a

Page 13: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

0,6

0,7

0,8

0,9

1

1,1

1,2

1,3

1,4

0 48 96 144 192 240 288 336 384

Corrected to amb. pressure & temp.

High

irr.

Low

irr.

Module201-Wire07

Time in hrs

Aging test:very high irradiation

g

Irradiation performed with

~1 A / cmclosed loop without

purification

m?

Observation ofdark pulses (kHz)

Observation ofpollutants:CH Cl

C H F Si

Disappearedwhen switching

to open loop

2 2

2 6 2

Disappearedwith purification

(closed loop)

Page 14: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Aging test: Gas GainDevelopment III

?

?

?

~2.3C / cmAccumulated

Different periodsof operation:

w/opurificationMolsieveR3-11

~0.2 A / cmm

Corrected to p & T

G=G x(( )xP) x( /T)` 1/Pi i i i

F1 F2T

Page 15: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

The open module

Gas outlet coloredG10-strip shows sparktraces

Irradiation area visibleNo wire was ruptured

Page 16: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

Summary and conclusion

?

?

?

?

Total accumulated charge ~2.3 C / cm within ~ 3000 hours withinclosed loop and purification

No significant losses in gain observed

Deviations in gain explainable due to ambient conditions

Equivalent to more than three years of nominal HERA-Bconditions in the hottest region

REMARK: after continuing of irradiationchamber became unusable. Reason ??

Under investigation.

Page 17: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

EquipmentGas Chromatograph

dire

ctio

n o

f drif t

time

sta

rt

Basic idea:of

gas componentsseparation

Agilent

Page 18: A Longterm Aging Study of the Honeycomb Drift Tubes of ...agingw/trans/ps/dehmelt.pdf · Frontend-electronics Gas inlet Gas outlet International Workshop on Aging Phenomena in Gaseous

International Workshop onAging Phenomena in Gaseous DetectorsOctober 2-5, 2001 at DESY in Hamburg

Wed, Oct 03, 01Klaus Dehmelt et al.

EquipmentGas Chromatograph

Detectors

Electron Capture Detector (ECD):

in detector volume

electronegative species I const

b IconstYKY

Mass spectrometer:

Identification ofcomponents

Quadrupolemass filter

H O2

H O2

Library spectrumof CF4

Ar/CF /CO4 2

CF4

CO2 Unknown species


Recommended