A Mechanical Installation Prototype for the DCLS Radiator Section L. X. Yin, L. Wang, S. Sun, W. Zhang, X. Hu, J. Chen, G. Q. Lin Shanghai Institute of Applied Physics, Shanghai, P. R. China Abstract The design of the mechanical system for DCLS radiator section was completed. The key components in each system, including the undulator, phase shift, magnets, vacuum chambers, CBPM, profile, and their supports, were fabricated and tested, respectively. In order to check the overall design and confirm some components design, a radiator section prototype was installed by using these components. The dynamic properties of the mechanical assembly were tested in this prototype. Most of the design was confirmed during the installation, but still some problems were found. The modification for the overall design and some components design, as well as the selection from several design for some components, have been made before their mass production. Dalian Coherent Light Source (DCLS) is designed to be working on the HGHG FEL principle. DCLS could be lasing at extreme ultra-violet wavelength with full tunable regime of 50 to 150nm. The construction started at 2014. Three key techniques 1. Elliptical pipe stretching 2. Long chamber brazing 3. Inner surface polishing Undulator Vacuum chamber OFHC chamber 15mm×6mm×δ0.75mm×3.3m Al alloy chamber 12mm×5mm×δ0.5mm×3.3m AFP process: R A <300 nm Pmax < 1×10-7 Torr OFHC Chamber Support dcls Installation AL Chamber Granite+3D adjustment 水平方向在±15mm范围内的调节定位精度和重复性误差 -4 -3 -2 -1 0 1 2 3 4 -20 -10 0 10 20 光栅尺的读数(mm) 误差=干涉仪读数-光栅尺读 数(um) 0mm运动到15mm 15mm运动到-15mm 误差上限 误差下限 -15mm运动到0处 >100μJ FEL pulse energy 50-150 nm FEL wavelength 0.3-1.6 Radiator parameter 30 mm Radiator period length 1 ps Seed laser width (FWHM) 240-360 nm Seed laser wavelength 300 A Peak current 1-2 mm.mrad Normalized emittance 1×10 -4 Relative energy spread 300 MeV Beam energy <3 m Length of each undulator <5 R.M.S phase error <20 μm Peak-to-peak trajectory deviation <3000 Gs-cm2 Second field integral <50 Gs-cm First field integral 0.2~0.8 T Peak field <5 m Resolution of Gap 9~25 mm Gap range Quadrupole, CBPM, profile, phase shifter, corrector are supported on the girder. Mechanical stability is the key issue in the design. Quadrupole requires a adjustment range of ±0.5mm with ±3μm precision for BBA. Granite Concrete Granite+3D adjustment Confirm the components mechanical design and fabrication. Confirm the overall design and interface of hardware. Confirm the installation procedure. Platform for some performance test. Performance test Radiator section prototype Installation prototype design
Transcript
Page 1: A Mechanical Installation Prototype for the DCLS Radiator Section · 2017. 3. 30. · A Mechanical Installation Prototype for the DCLS Radiator Section L. X. Yin, L. Wang, S. Sun,

A Mechanical Installation Prototypefor the DCLS Radiator Section

L. X. Yin, L. Wang, S. Sun, W. Zhang, X. Hu, J. Chen, G. Q. LinShanghai Institute of Applied Physics, Shanghai, P. R. ChinaAbstract

The design of the mechanical system for DCLS radiator section was completed. The key components in each system, including the undulator, phase shift, magnets, vacuum chambers, CBPM, profile, and their supports, were fabricated and tested, respectively. In order to check the overall design and confirm some components design, a radiator section prototype was installed by using these components. The dynamic properties of the mechanical assembly were tested in this prototype. Most of the design was confirmed during the installation, but still some problems were found. The modification for the overall design and some components design, as well as the selection from several design for some components, have been made before their mass production.

Dalian Coherent Light Source (DCLS) is designed to be working on the HGHG FEL principle. DCLS could be lasing at extreme ultra-violet wavelength with full tunable regime of 50 to 150nm. The construction started at 2014.

Three key techniques1. Elliptical pipe stretching

2. Long chamber brazing

3. Inner surface polishing

Undulator Vacuum chamber• OFHC chamber

15mm×6mm×δ0.75mm×3.3m• Al alloy chamber

12mm×5mm×δ0.5mm×3.3m• AFP process: RA <300 nm• Pmax < 1×10-7 Torr

OFHC Chamber

Support

dcls

Installation

AL Chamber

Granite+3D adjustment

水平方向在±15mm范围内的调节定位精度和重复性误差

-4

-3

-2

-1

0

1

2

3

4

-20 -10 0 10 20

光栅尺的读数(mm)

误差

=干

涉仪

读数

-光

栅尺

读数

(um)

0mm运动到15mm

15mm运动到-15mm

误差上限

误差下限

-15mm运动到0处

>100μJFEL pulse energy50-150 nmFEL wavelength

0.3-1.6Radiator parameter30 mmRadiator period length

1 psSeed laser width (FWHM)240-360 nmSeed laser wavelength

300 APeak current1-2 mm.mradNormalized emittance

1×10-4Relative energy spread300 MeVBeam energy

<3 mLength of each undulator<5 R.M.S phase error

<20 μmPeak-to-peak trajectory deviation

<3000 Gs-cm2Second field integral<50 Gs-cmFirst field integral0.2~0.8 TPeak field

<5 mResolution of Gap

9~25 mmGap range

• Quadrupole, CBPM, profile, phase shifter, corrector are supported on the girder. Mechanical stability is the key issue in the design.

• Quadrupole requires a adjustment range of ±0.5mm with ±3μm precision for BBA.

Granite Concrete Granite+3D adjustment

• Confirm the components mechanical design and fabrication.

• Confirm the overall design and interface of hardware.

• Confirm the installation procedure.• Platform for some performance test.