A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services Organization 113C Union Valley Road, Oak Ridge, TN 37830, USA DISCLAIMER This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, make any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information , apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof. COPYRIGHT NOTICE The submitted manuscript has been authored by a subcontractor of the U. S. Government under contract DE‑AC05‑84OR21400. Accordingly, the U. S. Government retains a paid‑up, nonexclusive, irrevocable, worldwide license to publish or reproduce the published form of this contribution, prepare derivative works, distribute copies to the public, and perform publicly and display publicly, or allow others to do so, for U. S. Government
Transcript
Page 1: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

A Simple Method for the Analysis of Po210: An Initial Report

D.K. Mann and J.K. FranseY-12 National Security Complex

BWXT Y-12 LLCAnalytical Services Organization

113C Union Valley Road, Oak Ridge, TN 37830, USA

DISCLAIMER

This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, make any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information , apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

COPYRIGHT NOTICE

The submitted manuscript has been authored by a subcontractor of the U. S. Government under contract DE‑AC05‑84OR21400. Accordingly, the U. S. Government retains a paid‑up, nonexclusive, irrevocable, worldwide license to publish or reproduce the published form of this contribution, prepare derivative works, distribute copies to the public, and perform publicly and display publicly, or allow others to do so, for U. S. Government purposes.

Page 2: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

General Outline

Little bit of HistoryReview of Pb210 ProcedureProblems with Po210 and Bi210 SeparationDevelopment of Po210 ProcedureData and Final ProcedureBenefits and DrawbacksFuture Work

Page 3: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 4: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Introduction

Member of the Uranium Decay Series (found naturally)

Pb210---->Bi210---->Po210

Used extensively as Environmental process tracers (Atmospheric, Terrestrial, Oceanic)

Isotope ratio allows use as source locator

Page 5: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

U U-238 U-234 U-232 U-235

4.4680E+0999.84%

2.4450E+05 7.0380E+08

92 y P a234m y 72y y

P a1.17m

P a-231

0.16%

91 P a234 32760y

Th Th-234 6.7h Th-230 Th-232 Th-228 Th-231 Th-227

1.405E+10

90 24.1d 77000y y 1.9132y 25.52h 98.62% 18.718d

Ac Ac-228 Ac-227

89 6.13h 21.773y

Ra Ra-226 Ra-228 Ra-224 Ra2231.38%

88 1600y 5.75y 3.62d 99.994% 11.434d

Fr Fr-223

87 21.8m

Rn Rn-222 Rn-220 Rn-2190.006%

86 3.8235d 55.61s (3%) 3.96s

At At-218 At-219 At-215

85 0.02% (~2s) (0.9m) (0.00023%) (~1E-4s)

P o P o-218 P o-214 P o-210 P o-216 P o-212 P o-215 P o-211

1.637E-04 138.378 2.98E-07 (97%) 0.001778

84 3.05m 99.979% s ~100% d 0.146s 64.07% s s 0.273% 0.516s

Bi Bi-214 Bi-210 Bi212 Bi-215 Bi-211

99.98% 99.99977%

83 19.9m ~100% 5.013d 60.55m (7.4m) 2.13m

P b P b-214 P b-210 P b-206 P b-212 P b-208 P b-211 P b-207

0.021% (1.3E-4%) 35.93% 99.727%

82 26.8m 22.26y STABLE 10.643h STABLE 36.1m STABLE

Tl Tl-210 Tl-206 Tl-208 Tl-207(1.7E-6%)

81 1.3m (4.19m) 3.053m 4.77m

Hg Hg-206

80 (8.5m)

Figure 4-1. Chart showing the decay chain of the uranium and thorium series isotopes and the half-lives of each isotope. Alpha decays are shown

by the vertical arrows and beta decays by the diagonal arrows. From Broecker and Peng "Travers in the Sea" Lament-Doherty Geological Observatory,

New York 1982.

Page 6: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Methods of Determination

Gas Proportional CountingAlpha Spectrometry (Silver

Autoplate)PERALS (Polex extraction)LSC LSC Alpha/Beta Proportional

Page 7: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

The Project and Problems

Po210 as source of high alpha in Hi-Vol Ambient Air monitors.

Needed quick and inexpensive method (very little funding).

No funds for Tracer.Minimal Sample Size.Startup time was minimal.

Page 8: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Carrying of Po206 on Te

0.3 M HCL 1.0 HCL 2.0 HCL 3.0 HCL 6.0 HCL

Te 57,470 58,458 57,024 55,866 54,611

Supn’t 570 673 511 930 2,624

Total 58,040 59,131 57,535 56,796 57,235

Te Precipitated by SnCl2

Taken from NAS-NS Nuclear Science Series “The Radiochemistry of Polonium”, pp.56-57, 1961

Page 9: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Dependency of Po210 Recovery on Te Concentration

0

10

20

30

40

50

60

70

0.2mg

0.4mg

0.6mg

0.8mg

1.0mg

Po-210Recovered

Page 10: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Dependence of Resolution on Te concentration

0

20

40

60

80

100

120

140

160

0.2

mg

0.4

mg

0.6

mg

0.8

mg

1.0

mg

Resolution (FWHM)

Page 11: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 12: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 13: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Dependency of Po210 recovery with varying Acid Concentrations

Plot shows Recovery of Po210 with varying acid concentrations. Average for 0.1M HCL is 66.3 +/- 1.2 with a RSD of 1.8%. When using 4M HCL, the recovery is slightly less, averaging 62.0 +/- 1.3 with a RSD of 2.0 %.

0

20

40

60

80

100

0.6mg

0.8mg

1.0mg

0.1 M HCL4 M HCL

Page 14: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Dependency of Po210 Resolution (FWHM) with varying Acid Concentrations

Plot shows Resolution of Po210 spectrum with varying acid concentrations. Average for 0.1M HCL is 144.7 +/- 13.6 keV with an RSD of 9.4%. When using 4M HCL, the overall resolution is better, averaging 98.8 +/- 19.5 keV with a RSD of 19.5 %. 0

20406080

100120140160180

0.6mg

0.8mg

1.0mg

0.1 M HCL4 M HCL

Page 15: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 16: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 17: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Recovery of Po209 vs. Po210

Recovery of Po210

averaged 61.2 +/- 2.3% while the recovery of Po209 was slightly higher, at 70.6 +/- 5.1. Excluding 1st point, average drops to 68.2 +/- 1.5

0

10

20

30

40

50

60

70

80

90

Sam #1 Sam #2 Sam #3 Sam #4

% R

ecov

ery

Po-210Po-209

Page 18: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 19: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.
Page 20: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Outline of Po210 Procedure

M ount fi lter on Lexan diskCount sam ple by Alpha Spec

I m m ediately fi lter the ppt through0.1 um M em brane Filter (used of N dF)

W hile still w arm , add 100 u l of0 .5 M SnC l2 . A Black ppt form s

Add 0 .6 m g Te M etal andHeat in Steam bath to 65 -70 Degrees

Convert to 4 M HCL(10-20 m l)

Elute Bi-210 and Po-210 in 10 m l 1 M HN O 3and 0 .1M HN O 3 from EichroM Pb R esin

Page 21: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Benefits of Po210 Method

Relatively InexpensiveRequires no special equipmentAble to use existing Eichrom Pb210

procedureRapid (1 hour from elution to counting)Can use a Tracer

Page 22: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Drawbacks of Po210 Method

Recovery not quantitative (60-70%).Resolution fairly large (70-100 keV).Not as good as PERALS or Silver Disk

method.May not be practical for labs who

Electroplate.

Page 23: A Simple Method for the Analysis of Po 210 : An Initial Report D.K. Mann and J.K. Franse Y-12 National Security Complex BWXT Y-12 LLC Analytical Services.

Future Work

Increase recovery (better Flocculation)Decrease FWHM (activity on surface)Try method with High Uranium SamplesWhat is the effect of heatingTry different Filter types (larger/smaller

pore size)


Recommended