A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity. Thais Russomano*  Marlise A dos Santos*  João Castro*  John Whittle**  Simon Evetts**  John Ernsting**. *Microgravity Laboratory/IPCT-PUCRS, Brazil **Aerospace Medicine Group, King´s College London, UK. - PowerPoint PPT Presentation
Popular Tags:
50
A A ssessment of an Earlobe ssessment of an Earlobe Arterialized Arterialized Blood Collector for use in Blood Collector for use in Microgravity Microgravity Thais Russomano* Thais Russomano* Marlise A dos Santos* Marlise A dos Santos* João Castro* João Castro* John Whittle** John Whittle** Simon Evetts** Simon Evetts** John Ernsting** John Ernsting** *Microgravity Laboratory/IPCT-PUCRS, Brazil **Aerospace Medicine Group, King´s College London, UK
Transcript
Page 1: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

AAssessment of an Earlobe Arterialized ssessment of an Earlobe Arterialized Blood Collector for use in MicrogravityBlood Collector for use in Microgravity

Thais Russomano* Thais Russomano*  Marlise A dos Santos*  Marlise A dos Santos*

 João Castro*  João Castro*  John Whittle**  John Whittle**  Simon Evetts**  Simon Evetts**  John Ernsting**  John Ernsting**

*Microgravity Laboratory/IPCT-PUCRS, Brazil

**Aerospace Medicine Group, King´s College London, UK

Page 2: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

There is currently no method of directly measuring arterial blood gas tensions in space.

Background

Page 3: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

There is currently no method of directly measuring arterial blood gas tensions in space.

An alternative to direct arterial measurement is earlobe arterialized blood sampling, an accurate technique for measuring blood gas tensions, which has been in use in clinical medicine and physiology for more than 30 years (Lilienthal JL & Riley RL, 1944).

Background

Page 4: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

There is currently no method of directly measuring arterial blood gas tensions in space.

An alternative to direct arterial measurement is earlobe arterialized blood sampling, an accurate technique for measuring blood gas tensions, which has been in use in clinical medicine and physiology for more than 30 years (Lilienthal JL & Riley RL, 1944).

This technique has not yet been examined in the microgravity environment due to the risk of environmental contamination with blood products.

Background

Page 5: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

The aim of the project is to develop a device and associated procedures that enable an acceptable estimation of arterial blood gas tensions and pH to be conducted accurately and safely in microgravity.

Aim

Page 6: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe Blood Sampling – The Technique

Page 7: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.

Page 8: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.

The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.

Page 9: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.

The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.

Blood is collected anaerobically via capillary tubes.

Page 10: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.

The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.

Blood is collected anaerobically via capillary tubes.

Blood is then analyzed using a blood gas analyzer.

Page 11: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Introduction

Earlobe massage

Earlobe arterialized blood collection

Page 12: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Characteristic Radial Artery Sample

Hyperemic Earlobe Sample

Discomfort Painful Pain Free

Potential Complications

Hematoma Hemorrhage

Infection (systemic) Wrist pain

Arterial Spasm

Hemorrhage Infection (cutaneous)

Ease of Use Requires trained medical personnel

Performed by non-medical personnel.

Potential Usage Hospital based research

Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.

Radial Artery / Arterialised Earlobe Collection

Page 13: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Characteristic Radial Artery Sample

Hyperemic Earlobe Sample

Discomfort Painful Pain Free

Potential Complications

Hematoma Hemorrhage

Infection (systemic) Wrist pain

Arterial Spasm

Hemorrhage Infection (cutaneous)

Ease of Use Requires trained medical personnel

Performed by non-medical personnel.

Potential Usage Hospital based research

Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.

Page 14: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Characteristic Radial Artery Sample

Hyperemic Earlobe Sample

Discomfort Painful Pain Free

Potential Complications

Hematoma Hemorrhage

Infection (systemic) Wrist pain

Arterial Spasm

Hemorrhage Infection (cutaneous)

Ease of Use Requires trained medical personnel

Performed by non-medical personnel.

Potential Usage Hospital based research

Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.

Page 15: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Characteristic Radial Artery Sample

Hyperemic Earlobe Sample

Discomfort Painful Pain Free

Potential Complications

Hematoma Hemorrhage

Infection (systemic) Wrist pain

Arterial Spasm

Hemorrhage Infection (cutaneous)

Ease of Use Requires trained medical personnel

Performed by non-medical personnel.

Potential Usage Hospital based research

Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.

Page 16: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Characteristic Radial Artery Sample

Hyperemic Earlobe Sample

Discomfort Painful Pain Free

Potential Complications

Hematoma Hemorrhage

Infection (systemic) Wrist pain

Arterial Spasm

Hemorrhage Infection (cutaneous)

Ease of Use Requires trained medical personnel

Performed by non-medical personnel.

Potential Usage Hospital based research

Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.

Page 17: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Development and Evaluation of a Earlobe Arterialized Blood Collector

Page 18: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Prototype of the Earlobe Arterialized Blood (EAB) Collector

Page 19: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Current Prototype of the EAB Collector

Blade module

Capillary tube module

Capillary tube

Body (module housing)

Ophthalmic blade

Page 20: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.

Method

Page 21: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.

Six healthy subjects breathed a gas mixture of 12.8% O2 in N2 (equivalent to breathing air at 12,000 feet) during 15 min of head-down tilt.

Method

Page 22: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.

Six healthy subjects breathed a gas mixture of 12.8% O2 in N2 (equivalent to breathing air at 12,000 feet) during 15 min of head-down tilt.

The blood samples were analyzed immediately.

Method

Page 23: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

MethodHead Down Tilt to partially replicate the

effects of microgravity

Page 24: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples  

  Radial arteryMean SD (range)

Arterialized EarlobeMean SD (range)

pH(pH Unit)

7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)

PO2

(mmHg)

42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)

PCO2

(mmHg)

34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)

SaO2

(%)

79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)

Page 25: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples  

  Radial arteryMean SD (range)

Arterialized EarlobeMean SD (range)

pH(pH Unit)

7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)

PO2

(mmHg)

42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)

PCO2

(mmHg)

34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)

SaO2

(%)

79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)

Page 26: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples  

  Radial arteryMean SD (range)

Arterialized EarlobeMean SD (range)

pH(pH Unit)

7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)

PO2

(mmHg)

42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)

PCO2

(mmHg)

34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)

SaO2

(%)

79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)

Page 27: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples  

  Radial arteryMean SD (range)

Arterialized EarlobeMean SD (range)

pH(pH Unit)

7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)

PO2

(mmHg)

42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)

PCO2

(mmHg)

34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)

SaO2

(%)

79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)

Page 28: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples  

  Radial arteryMean SD (range)

Arterialized EarlobeMean SD (range)

pH(pH Unit)

7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)

PO2

(mmHg)

42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)

PCO2

(mmHg)

34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)

SaO2

(%)

79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)

Page 29: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.

Page 30: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.

Comparison of PCO2 of the arterialized earlobe and arterial samples showed a mean difference of -1 ( ± 0.75) mmHg.

Page 31: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.

Comparison of PCO2 of the arterialized earlobe and arterial samples showed a mean difference of -1 ( ± 0.75) mmHg.

There was no difference between the pH values of the arterialized earlobe and arterial samples.

Page 32: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Correlations between Arterial and Earlobe Blood Sample Measurements

r = 0.93 r = 0.94

r = 0.97 r = 0.92

Page 33: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11) 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 34: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11) 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 35: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11) 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 36: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11) 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 37: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11)

 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 38: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

 

 

 

Arterilaized Capillary- Arterial Mean ( SD)

PO2

mmHg

PCO2

mmHg  Langlands & Wallace

(1965) (n=16)

 

0.62(4.1)

 

 

1.05(1.6)

 

Godfrey et al. (1971) (n=8) 

2.09(2.48)

 

 

0.65(1.2)

 

Spiro & Dowdeswell (1971) (n=11) 

-0.72(1.67)

 

 

1.0(1.91)

 

Dar et al. (1995) (n=55) 

 

0.675(4.43)

 

 

0.75(2.25)

 

Present study (n=6) 

 

0.25(1.25)

 

-1.0(0.75)

Study Author

Page 39: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

Points of note.

Page 40: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

Points of note.

• Arterialized blood collection from the earlobe is virtually pain free.

Page 41: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

Points of note.

• Arterialized blood collection from the earlobe is virtually pain free.

• Anaesthetic is not required.

Page 42: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

Points of note.

• Arterialized blood collection from the earlobe is virtually pain free.

• Anaesthetic is not required.

• Anaerobic blood collection is possible.

Page 43: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Discussion

Points of note.

• Arterialized blood collection from the earlobe is virtually pain free.

• Anaesthetic is not required.

• Anaerobic blood collection is possible.

• EAB Collector prevents environmental contamination.

Page 44: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Conclusion

This limited study suggests that:

Page 45: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Conclusion

This limited study suggests that:

1. Arterialized blood sampled from the earlobe provides accurate estimations of arterial blood measurements of PO2, PCO2 and pH during microgravity simulation and in hypoxia.

Page 46: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Conclusion

This limited study suggests that:

1. Arterialized blood sampled from the earlobe provides accurate estimations of arterial blood measurements of PO2, PCO2 and pH during microgravity simulation and in hypoxia.

2. Blood collection will be possible in microgravity without environmental contamination.

Page 47: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Further Studies

1. To validate the EAB Collector during normoxia in other body positions (sitting and supine);

2. To test the efficiency of the EAB Collector in microgravity (parabolic flight).*

* Protocol under evaluation by ESA.

Page 48: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Thank you for listening

Questions?

Page 49: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Professor Thais Russomano

[email protected]

[email protected]

Page 50: A ssessment of an Earlobe Arterialized  Blood Collector for use in Microgravity

Results

Individual Data