+ All Categories
Home > Documents > A Tribological Study of a Series of Retrieved

A Tribological Study of a Series of Retrieved

Date post: 03-Jun-2018
Category:
Upload: amber-williams
View: 221 times
Download: 0 times
Share this document with a friend

of 6

Transcript
  • 8/11/2019 A Tribological Study of a Series of Retrieved

    1/6

    A tribological study of a series of retrieved

    Accord knee explants

    e.EIII;;iez*, J. Fisher*, D. Dow son*, S.A. Sampath, R. Johnson* and

    .

    *Department of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK;

    University Departm ent of Orthopaedics and Accident Surgery, U niversity of Liverpool,

    Liverpool, UK; *Knee Research U nit, Arrowepark Hospital, U pton, Wirral, Merseyside,

    L49 5PE, UK;

    * *DePuy International Ltd., St Anthony s Road, Leeds LSll SDT, UK

    ABSTRACT

    A tribological s y of a series of 27 retriev ed Accord meniscal bearing knee joints has been carried out. Ihe roughness of

    the articulating sufaces of the men al and the femoral componen ts was investigated, and the penetration into the

    polyethylene meniscal comp onent was determined. There was generally little damage to the highly polished metallic

    femoral components. A general polishing and smoothing of the ultra high molecular weight polyethylene meniscal

    compon ents was observed, although there was some subsequent deterioration of

    the

    olyethylene surfaces in certain cases.

    h e penetration rate of thefemoral componen ts into the menisci was low, and was found to be similar to that reportedfor

    other m eniscal k nee joints.

    Keywords: Tribology, retrieval, knee prostheses

    Med. Eng. Phys., 1994, Vol. 16, 223-228, May

    INTRODUCTION

    The natural kn ee offers a complex range of move-

    ments with 150 of motion in flexion-extension,

    rotation, and translation in the anterior-posterior

    direction. In the more widely used total condylar

    artificial knees this range of movem ents is achieved

    by having relatively low conforming tibia1 and

    femoral components . This produces high stresses in

    the contact. In certain knee designs the high cyclic

    contact stresses can cause sub-surface fatigue and

    structural failure o f the ultra high molecular weight

    polyethylene (U HM WPE) compo nent3. In some

    cases inadequate

    mechanical properties and

    poor quality UHM WPE have been acknowledged

    as accelerating factorsk7. Sub-surface fatigue can

    initiate several millimetres into the UHM WPE at the

    position of the maxim um shear stress, with cracks

    propagating to the surface causing com plete failure

    and break up of the UHMW PE component. This

    type of failure has been widely described as delamin-

    ation in the literature8~g. In addition structural failure

    can also initiate from stress concentrations arising

    from the metal tibia1 support or tray, which can lead

    to the complete breaking up of the polyethylene

    compo nent. It is important to emphasize that this

    mode of failure and wear is not primarily a surface

    wear process, as it is not caused by relative motion of

    the surface asperities, and it initiates in the bulk of the

    material. In some total condylar knees it has been

    possible to reduce the levels of contact stress that

    cause structural failure, by increasing the conformity

    Correspondence and reprint requests: Dr J.L. Hailey

    0 1994 Butterworth-Heinemann for BES

    1350-4533/94/03223-06

    of the compon ents, but this can restrict the range of

    movem ents an d impose biomechanical constraints.

    The meniscal bearing knee offers an alternative

    approach to overcoming the desi

    ated with total condylar knees

    p problems associ-

    .

    Memscal bearing

    knees achieve a wide range of motion by allowing

    articulation at both the femoro-meniscal and the

    tibio-meniscal surfaces. Figure 7 shows an Accord

    meniscal knee. The femoro-meniscal bearing surfaces

    are conforming and cylindrical in shape which gives

    rise to a large area of contact on both cond yles in a

    correctly aligned knee. The tibio-meniscal surfaces

    are conical in shape and allow rotation of the

    polyethylene meniscal comp onent with respect to the

    tibial tra

    .

    In most components the translation of the

    menisca ly omp onent is constrained by a pin through

    the apex of the truncated conical surface. Thu s the

    Accord meniscal knee permits flexion-extension at

    the femoro-meniscal bearing surface, and rotation at

    the tibio-meniscal surfaces. This allows more highly

    conforming bearing surfaces to be used and it has

    been postulated that reduced contact stresses may

    reduce the incidence of early structural failure and

    perhaps reduce the volume of wear debris generated

    in the long term. The purpose of this study was to

    investigate the tribological character istics of a series of

    retrieved Accord meniscal knees, and in particular

    focus on the condition of the UHM WPE and the

    femoro-meniscal wear surfaces.

    MATERIALS AND METHODS

    A total of 768 Accord knees Figure

    7)

    have been

    implanted at Arrowepa rk Hospital since a clinical

  • 8/11/2019 A Tribological Study of a Series of Retrieved

    2/6

    Tribological tudy

    o

    retrieve knee expiaat s:J.L. Hai lq et al.

    Figure Photograph of the Accord meniscal bearing knee

    Figure

    2 Photograph of an explant which suffered from metal-to-

    metal contact during articulation

    trial of the knee began in 1982, of which 73 have now

    been revised. The majority of these were revised for

    failure of cementless fixation and infection. None of

    the explants were revised for structural failure of the

    UHM WPE meniscus. Twenty-seven paired meniscal

    and femoral com ponents were available for examina-

    tion in this study.

    An initial visual examination and the clinical

    history of the explants were used to determine

    the modes of failure of the compo nents. Certain

    mechanical modes of failure were then eliminated

    from the tribological study. These included joints

    where the meniscus had dislocated and led to the two

    metal compo nents, the femoral compon ent and the

    tibia1 tray, contacting. During articulation these two

    metal compo nents maintained contact leading to the

    generation of large amounts of metal w ear debris. A n

    example of this type of failure is shown in Figure 2.

    Sixteen of the 27 paired explants had failed in this

    manner and were thus discarded from the rest of the

    study, which has focused on the articulating surfaces.

    The full clinical history, p atient details and regular

    follow-up information

    of the eleven remaining

    explants (which were evaluated for wear) were

    available from hospital records, and are summ arized

    in Table 1. The average age of the patients was 60.0

    years a nd their average weight was 71.1 kg. The

    period of implantation for these compon ents ranged

    from less than 1 year to over 5 ears with an average

    lifetime of 33.4 months. In a dy ition to the revised

    explants, four unwo rn p rostheses taken from clinical

    stock prior to implantation were also evaluated as

    control samples.

    The articulating surfaces of the

    menisci a nd femoral comp onents were characterized

    by R, surface roughness measurem ents using a

    Talysurf 5 machine (Rank Taylor Hobson , Leicester).

    The R, was measured perpendicular to the direction

    of sliding using a 0.8 mm cut off. Three readings were

    obtained from each cond yle, giving a total of six

    measurem ents per comp onent. The mean of these

    readings was then calculated and used as the average

    R, value for that particular compo nent.

    The UHMW PE meniscal components were wider

    than the femoral compon ents, and due to a close

    fitting central locating step (see Figure 4b), the edges

    of the menisci were not subjected to wear. This

    resulted in the formation of a lip at the edges of

    Table 1 Summ ary of clinical data, patient details and follow-up information

    Case

    Wear life

    number (months)

    Activity

    capability

    Age

    (years)

    Weight

    (kgs)

    Sex

    Operation

    side

    Initial

    diagnosis

    Reason for

    revision

    I 8 A 73

    72 F I. OA

    I,, ST

    2 9 - 62

    66 F R RA

    I

    3 IO A 60

    71 F L OA

    I

    4 I6 A 61

    64 M L O A

    I,

    5 23 P 61

    54 F R RA

    I

    6 26 G 67

    79 M I* Other

    I, L

    7 43 G 46

    73 F R RA

    L, ST

    8 45 G 63

    61 F R RA

    I, U

    9 59 A 62

    88 M L O A

    I,, ST

    IO 64 G 62 77 M R O A L

    II 64 A 43

    77 F L OA

    L

    G, good mobility; A, average mobility; P, poor mobility; OA, osteo arthritis; RA, rheumatoid arthritis; I, infection; L, loosening; ST, subsidence of

    tibia; U, unstable

    224 Med. Eng. Phys. 1994, Vol. 16, May

  • 8/11/2019 A Tribological Study of a Series of Retrieved

    3/6

    the menisci. A Talycontor machine (Rank Taylor

    Hobson, Leicester) was used to trace across the

    surface o f the menisci (perpendicular to the sliding

    direction) from one lip, over the worn region to the

    worn area and unworn lip on the op osite side of the

    comp onent. A rotating levelling ta 1 le was used to

    ensure that the two unw orn lips were at the same

    height, and that the menisci were orientated at

    exactly 90 to the direction of sliding. This proce-

    dure was repeated twice for each sample and the

    average difference between the height of the unwo rn

    lip and the height of the worn region was evaluated to

    give the penetration depth for each polyethylene

    comp onent. It should be pointed out that the

    measured penetration into the menisci occurred as a

    result of both creep and wear of the polyethylene.

    Therefore the derivation of wear rates from penetra-

    tion rates should be treated with caution, particularly

    when penetration depths are small.

    RESULTS

    The surface roughnesses of the femoral compo nents

    are shown in the graph in F igure 3. Th e number of

    months that the joint was im lamed for is shown on

    the x-axis, and the R, (in pm

    P

    It was found that general1

    is shown on the y-axis.

    were in very good cr

    the femoral compo nents

    con ition and had R, values

    similar to the unw orn control. The av erage R, value

    for the worn femoral com ponents was 0.055~m with

    a standard deviation of 0.075p.m (average R= of

    0.032pm if the one anomaly, 59 months, was not

    included). This corn

    P

    ares with an Ra of 0.02p.m for

    the unwo rn contra compo nent. There was one

    sample (59 months) which ha d a surface roughn ess

    an order of magnitude higher than the rest of the

    samples. This comp onent had been badly scored

    g

    robably as a result of the entrapment of cement and

    one debris. One other compo nent (64 months)

    showed a slightly higher R, and a slight deterioration

    of the surface.

    Figure 4a shows the surface roughn esses of the

    meniscal compon ents. The implant lifetime is shown

    on the x-axis, and the R, is shown on the y-axis. From

    this graph it can be seen that there was an initial

    Lifetime months)

    Figure 3 Histogram of the femoral surface roughness (R,) and

    implant life

    Tri bologicnl tudy of retri eved nee explant s:JL. Hai lq, et al.

    a

    Lifetime months)

    Figure 4 a Histogram of the meniscal surface roughness ( ) and

    implant life. 4b, Photograph of a worn (right) and an unworn (left)

    meniscus

    smoothing of the surfaces followed by some deteriora-

    tion in certain cases. The average R, for the worn

    menisci was 1.15 pm with a standard deviation of

    1.03pm (using the eight polished com ponents the

    average Ra was 0.60pm, using the three bad1

    damaged compon ents the average R, was

    2.63pm .

    r

    This compares with an avera e

    (standard deviation 0.23 pm

    Y

    value of 1.04 pm

    for the unworn

    samples. The Ra values for the worn co mpon ents

    ranged from 0.28 pm up to 3.27pm, whereas those

    for the unwo rn con trols ranged from 0.76pm up to

    1.33pm. Initially the polymer surface was quite

    rough due to the machining marks. Initial wear and

    creep produced a smoothing or polishing of the

    UHM WPE meniscal compo nents. The polishing of

    the worn menisci, compa red to an unwo rn control,

    can be clearly seen in the

    4b, where the sample on tK

    hotograph shown in F igu re

    e right is the worn sample.

    Subsequently in certain cases the polymer surface

    became damaged and the surface roughness

    increased.

    F igu re 5 shows the articulating surface of Case 10

    (the first of the two samples implanted for 64 months

    on the graph in

    F igure 4a .

    This sample showed severe

    surface dama

    explants, an B

    e compa red with that seen in the other

    failure of the surface can be seen in

    F igure 5. This was the worst case of the 27 explants

    which were studied. The Talycontor profile showed

    Med. Eng. Phys. 1994, Vol. 16, May 225

  • 8/11/2019 A Tribological Study of a Series of Retrieved

    4/6

    Tri bological tudy of etr ieved nee explunt xJ.1. Hai lq et al.

    Figure 5 Photograph of the meniscus from Case 10 showing surface

    micro-delamination

    that this surface dam age and wear occurred within

    0.3 mm of the meniscal bearing surface. Similar wear

    features have been seen on acetabular cups from

    simulator studies which we have performed at Leeds

    University *. Howev er, there was also evidence of

    gross misalignment and dislocation of the poly-

    ethylene meniscal compo nent, and this surface

    damage could have been caused by contact with the

    edge of the femoral compo nent when grossly mis-

    aligned. Local surface damage at the edge of the wear

    scar was noted on several of the older meniscal

    compo nents and this is shown in Figure 6. This is

    believed to have occurred when the femoral com-

    ponent and the meniscus became slightly misaligned.

    The very small radii on the femoral compo nents led

    to the development of high contact stresses a t the

    edge contact with the meniscus. This in turn resulted

    in local stru ctural failures, in the form of local severe

    damage to the surface of the polyethylene.

    The penetration de ths into the menisci are shown

    in

    Figure 7. The

    imp ant life is shown on the x-axis

    and the penetration (in pm) is shown on the y-axis.

    There was a considerable variation in the penetration

    depths of the explants, and no correlation could be

    found with either wear life, patient weight, or patient

    activity. It can also be seen from Figure 7 that four of

    the menisci showed no detectable penetration, even

    after implantation for over 5 years in one case (64

    months on the graph in Figure 7). Two of the samples

    were too badly deformed to obtain penetration

    Figure 6 Photograph showing local structural failure of a meniscus

    due to misalignment of the femoral component

    i 23 26 45 59 64

    Lifetime months)

    Figure 7 Histogram of the meniscal penetration depth and implant

    life

    depths from. This was due to severe misalignment of

    the meniscal compo nent and gross damage of the

    polyethylene, and hence these samples were dis-

    regarded from this part of the study. The average

    penetration rate from the data shown in Figure 7 w as

    found to be 40pmyea r-. In this study the

    P

    enetra-

    tion depths an d the Ra values for the latera side of

    both compon ents were found to be similar to those

    obtained for the medial side.

    DIS USSION

    Recent reports13 of structural fatigue and early failure

    in highly stressed nonconform ing total condylar

    knees have caused considerable concern in the

    surgical commu nity. This has promp ted a renewed

    interest in the meniscal bearing knee w here it is

    possible to increase contact area and reduce contact

    stresses to levels m ore generally found in hip

    prostheses. It is important to recognize that changes

    in the design an d tribological cond itions in artificial

    joints can cause marked changes in the wear and

    failure processes in the UHM WPE compo nents. In all

    polyethylene acetabular cups in hip prostheses, where

    contact stresses are generally lower than in total

    condylar knees, structural failure is extremely rare,

    and wear particles are generated by surface wear

    processes produced by asperity interactions14. In

    Argenson and OConn ors long term study of

    Oxford meniscal knees, surface wear and pentration

    rates were reported, but no evidence of structural

    failure was recorded, demonstrating the potential of

    the meniscal bearing concept. In this study the

    majority of the implants retrieved showed metal-to-

    metal contact due to dislocation of the meniscal

    compo nents. As this would have produced consider-

    able metallic debris which could subsequently

    damage the articulating surfaces, these prostheses

    were excluded from the study.

    It was shown that, in general, the surface roughness

    of the femoral comp onents did not change signifi-

    cantly with implantation time. Some fine scratches

    running parallel to the direction of sliding w ere

    visible on the surfaces of the femoral compo nents,

    although these did not appear to have a significant

    226 Med. Eng. Phys. 1994, Vol. 16, May

  • 8/11/2019 A Tribological Study of a Series of Retrieved

    5/6

    Tri boh@al study of etri eved nee explunt s:JL. Ha y et al.

    reduced stresses have been demonstrated in this

    study. However,

    it is necessary to

    em hasise

    biomechanical difficulties of ensuring sta

    the

    meniscal knee design.

    ility in the

    influence on the R, values. Since the scratches were

    orientate d in the direction of sliding, it is unlikely that

    they would hav e had a significant effect on the wear

    of the polyethylene, compa red to scratches which

    were perpendicular to the sliding direction. In hip

    joints this type of scratching has been related to

    increased polyethylene wear as in the hip the

    scratches tend to be more random ly orientated due to

    the more complex movem ents generated.

    The pattern of wear and changes in the surface

    to ograph of the meniscal compo nent were gener-

    al y ip ly similar to those found in acetabular

    cr

    where surface wear processes remove

    EYsethylene particles within 10pm of the surface,

    an dr severe plastic deformation of the polyethylene

    has been noted up to a depth of 40pm below the

    surface. T he initial roughness of the pol

    marks was smoothed by creep

    an Jrn

    er machine

    counterface as

    microscopic

    t

    erity wear (Figure 4). Subsequent1 in

    certain cases

    e polymer surface deteriorated. Jh is

    may be due to abrasive action by third body bone

    cement particles, or alternatively the onset of macro-

    scopic

    1

    olymer w ear processes which have been

    reporte in explanted acetabular cups15. One

    meniscal compo nent (Case 10) showed severe surface

    damage, and the depth of this severe surface damage

    was up to 300pm from the articulating surface.

    Although similar surface wear patterns have been

    seen in hip joint simulator studies in the laboratory,

    there was also evidence of gross misalignment of this

    compo nent and this severe surface damage ma

    have

    been caused by edge contact of the femora ly

    com-

    ponent. In the explants which d id not dislocate, or

    show evidence of gross misalignment, gross structural

    failure (characterized by delamination originating

    from a depth of greater than 0.5mm into the

    UHM WPE) was not found. However, the follow u

    of 5 years is quite sh ort and direct comparison

    wi

    t

    longer term follow u p studies of total condylar knees

    should be approached with caution. W hen misalign-

    ment of the Accord knee occurred, local stress

    concentrations were formed at the edge of the

    femoral co mpon ent, and local structural failure

    resulted due to the abnormally high contact stresses.

    The average penetration rate was also evaluated

    from our data and compared with that obtained by

    other workers. We calculated a penetration rate of

    0.040 mm year- (4Opm ) for the Accord knee, w hich

    compares well with the figure of 0.047 mm

    year

    obtained by Argenson and OConnor for the

    Oxford m eniscal knee. In comparison Dow son et

    a1.16 obtained significantly higher penetration rates

    for conventional Geomedic and Freeman-Swan son

    knees - 0.22 and 0.26mmyea r respectively. The

    penetration rate in the meniscal knee was comparable

    with that found with hip joints as shown by Dowling17

    who obtained a penetration rate of 0.06 mm year- for

    post mortem acetabular cups. How ever, care must be

    taken when comparing penetration rates in different

    types of knees and hips, as the different geometries

    involved will produce different v olume

    than es for

    the same penetration rate. Also these volume c a anges

    can include different contributions from creep and

    wear. Des

    mental tri

    K

    ite this, the authors feel that the funda-

    ological

    advantages of a meniscal knee

    design, which include larger contact areas and

    CONCLUSIONS

    In conclusion, the tribological advantages of the

    Accord meniscal bearing knee joint are as follows.

    There was generally little damage of the articulating

    surfaces. The wear and penetration rates were very

    low. There was no evidence of any structural failures,

    except in cases where the femoral corn

    onent and the

    meniscus were misaligned, and lot

    ap

    damage was

    formed at the edge of the contact area. The p attern of

    surface w ear in this knee was similar to that seen in

    hips with comparable penetration rates for the two

    types of joint.

    In the 27 retrieved knees reported in this study

    failure of the cementless fixation was the most

    frequen t cause of revision. This led to instability of the

    joint and edge contacts between the femoral com-

    ponent and the meniscus, which due to the small radii

    on the femoral compo nent led to the generation of

    high contact stresses and local structural failure.

    There were a number of cases where dislocation of

    the meniscus which led to metal-on-metal contact

    during articuation, and the potential for generation of

    large quantities of metal wear debris.

    AC KNOWLEDGFBIENTS

    This work was funded by the Arthritis and Rheum at-

    ism Council (ARC).

    REFERENCES

    1.

    2.

    3.

    4.

    5.

    6.

    7.

    Walker PS. Bearing surface design in total knee replace-

    ment. Eng Med 1988; 17: 149-56.

    Bartel DL, Bicknell VL, W right TM. The effect of

    conformity, thickness and material on stress in ultra-high

    molecular weight components for total joint replacement.

    JBoneJoint Surg 1986; 68-A: 1041-51.

    Wright TM, Rimnac CM, Stulberg SD, Mints L, Tsao AK,

    Klein RW, McCrae C. Wear of polyethylene in total joint

    replacements. Clin Orthop Rel Res 1992; 276: 126-34.

    Kilgus DJ, Moreland JR, Finerman GAM , Funahashi IT,

    Tipton JS. Catastrophic wear of tibial polyethylene inserts.

    Clin Orthop Rel Res 1992; 273: 223-31.

    Mayor MB, W rona M, Collier JP, Jensen RE. The role of

    polyethylene quality in the failure of tibial knee com-

    ponents. Transactions of 39th Annual Meeting of Orthopaedic

    Research Society, Feb. 15-18, San Francisco, 1993; 292.

    Landy MM , Walker PS. Wear of UHM WPE in 90

    retrieved knee prostheses.

    J Arthrq ty,

    1988; Oct.

    Suppl. 73.

    Blunn GW, Joshi AB, Lilley PA, Engelbrecht E, Ryd L,

    Lidgren L, Hardinge K, Nieder E, Walker PS. Poly-

    ethylene wear in unicondylar knee prostheses: 106

    retrieved Marm or, PCA, and St Georg tibia1 components

    compared. Acta Orthop and, 1992; 63: 247.

    8. Collier JP, Mayor M B, McNamara JL, Surprenant VA,

    Jensen RE. Analysis of the failure of 122 polyethylene

    inserts from uncemen ted tibia1 knee components. Cl in

    Orthop Rel Res

    1991; 273: 232-42.

    9. Engh GA, Dwyer KA, Hanes CK. Polyethylene wear of

    Med. Eng. Phys. 1994 Vol. 16 May 227

  • 8/11/2019 A Tribological Study of a Series of Retrieved

    6/6

    Tribologicaltudy f et ri eved nee xpkmt s:J.L. Ha y

    et al

    metal-backed tibia1 components in total and unicompart-

    mental knee prostheses. JBone Joi nt Surg 1992; 74-B: 9- 17.

    10. Argenson J-N, OConnor JJ. Polyethylene wear in

    meniscal knee replacement. J BoneJoi nt Surg 1992; 74-B:

    228-32.

    11. Johnson R, Walker CR, Harvey IA, Barry GK, Elloy MA.

    Five to eight year results of the Johnson-Elloy Accord)

    total knee arthroplasty. JAr t hrop lasty , 1993; 8: 27-32.

    12. Dowson D, Jobbins B. Design and development of a

    versatile hip joint simulator and a preliminary assessment

    of wear and creep in Charnley total replacement hip

    joints. Eng Med 1988; 17: 111-17.

    13. Goodfellow J. Editorial, Knee prostheses - one step

    forward, two steps back. JBoneJoi nt Surg 1992; 74-B: l-2.

    14. Cooper JR, Dowson D, Fisher J. Macroscopic and

    microscopic wear mechanisms in ultra-high molecular

    weight polyethylene. Wear 1993; 162-164: 378-84.

    15. Cooper JR, Dowson D, Fisher J. Birefringent studies of

    polyethylene wear specimens and acetabular cups. Wear

    1991; 151: 391-402.

    16. Dowson D, McCullagh PJ, Wright V. An assesssment of

    the relative importance of wear and creep in the overall

    performance of load-bearing total replacement knee

    joints. In: Willert H-G, Buchhom GH, Eyerer P, eds.

    Ul tr a-H h Mok cular Weight Polyethyl ene as Biomat erial

    in Or thopaedic Surgeery. Toronto: Hogrefe and Huber

    Publishers 1991; 32-40.

    17. Dowling JM. Wear analysis of retrieved prostheses. In:

    Szycher M, ed. Bi ocompati ble Pol ymers, M et als, and Com-

    posi t es. Pennsylvania: Technomic 1983; 407-425.

    1994 IEEE

    ULTRASONICSSYMPOSIUM

    NOVEMBER~ST-~TH

    1994

    CANNES

    RANCE

    Sponso red by the IEEE Ultrasonics, Ferroelectrics Frequency Control Society

    There will be a large range of topics for discussion, including:

    Bio-effects and Biophysics

    Hyperthermia

    Medical imaging

    Physical acoustics

    Quantitative laser ultrasonics

    Tissue characterization

    Tissue motion and compliance

    Ultrasonic in surgery

    Abstract submission deadline: Friday April 29th, 1994

    Abstrac ts to be sent to: Ultrasonics Sym posium, c/o LRW A ssociates, 1218

    Balfour Drive, A rnold, M D 21012-2150, USA .

    Tel: +410 647 1591 or Fax: +410 647 5136.

    For further information please contact: Bernhard R. Titmann, The Pennsylvania

    State University, 227 Hamm ond Building, University Park, PA 16802, USA .

    Tel: +814 865 7827 or Fax: +814 863 7967.

    228

    Med. Eng. Phys. 1994 Vol. 16 May


Recommended