Initial Clinical Experience with 3D Surface Image Guidance
Amanda Havnen-Smith, Ph.D.
Minneapolis Radiation Oncology
Ridges Radiation Therapy Center
Burnsville, MN
April 20th, 2012
OUTLINE
• Description of 3D Optical Tracking System and Purpose
• Use of the System and the Workflow
• Strengths and Weaknesses of the System
• TG 147 Recommended Quality Assurance
Non-funded
research
collaboration
Can be used for
patient set-up and
surveillance
3-D
Non-Invasive
Non-Ionizing
Why use 3D surface image guidance?
Purpose of Surface Image Guidance
Inter-Fraction
Give consistency with day-to-day patient setup
Intra-Fraction
Capability to pause treatment if patient moves out of
acceptable position
Capability to gate the beam for motion management
The goal is to reduce two types of uncertainty in external
beam RT
As EBRT
treatments
become more
conformal,
patient
positioning and
motion
management
become even
more critical!
Small errors can be “gross” errors!
Cameras project a red speckle pattern on the surface of the patient, and determine where the patient’s surface is in space.
Technology
Stereoscopic System
3D Surface Information from Image Pair
1) Known Feature Extraction
- Structured light projection (speckle pattern)
2) Locate corresponding feature in 2nd image
3) Compute 3D coords of features using triangulation
Camera Pod
Software calculates the position of each point along
the CT surface with respect to the isocenter
Compares the position of all points on the acquired
surface (in the Region of Interest) with respect to the
planned isocenter
Does not directly compare the acquired surface to the
planned surface
Does not directly account for deformation
Do have the option of adding additional ROIs, and
you may use centroid method for these
- This method compares the weighted centroid of the
actual surface to that of the reference surface
Strongly recommend using centroid option for objects
far from isocenter
-For example on a breast patient: chin ROI, arm ROI
Strongly recommend not to use the centroid option for
primary ROI set-up, system not optimized for this
Planning CT Surface (body contour)
Ideally our preference
Large amount of image data
Frame rate is slower
Surface can have features not present at treatment
Recorded surface in treatment position
Faster frame rate for real-time shifts, beam holds
Relationship to CT surface is not known
What do we compare the treatment position to?
Wires, Fiducials, Artifacts, etc.
Requires attention to the body contour?
WORKFLOW System isocenter “calibrated” by physics once a month
Constancy of iso checked daily by therapists
Dosimetrist/Physicist exports data and therapist
imports into AlignRT software:
Structure Data (dicom RTStructure file)
- From this only the Body contour is used
Treatment plan for each isocenter (dicom RTPlan
file)
- Only need to export a single plan if plans have
+ Same CT image set
+ Same isocenter
Imported Surface
& Isocenter
ROIs: Functionality largely dependent on ROIs
Manufacturer recommends:
Not too big – T.M.I.!
Not too small – Not enough info to localize!
Works best with well-defined features Does not work as well with fleshy mobile features (large breasts) or highly symmetric features (expanders)
It is recommended to include:
Breast tissue (if not too pendulous or symmetric)
Some axilla/arm
Some lower neck/SCLAV
Part of
lateral chest
wall
Patient Setup
Line up patient using marks, lasers, SSDs
Begin monitoring , use “real-time deltas” to move patient into correct
position
Real-time shifts shown for 3 translations,
3 rotations
• w/in 3 mm and 30 (default tolerance)
Then initiate treatment
Can record treatment surface
Can generate a report showing the offsets at any time
Treatment Monitoring
If patient moves significantly, could pause tx
and re-align to plan
- Can be helpful for non-compliant
patients, can see motion on camera,
but did they move back??
If bolus patient, we apply it, then capture new
image to monitor motion
Motion management tool
Deep Inspiration Breath Hold (DIBH)
Technique
Goal to reduce risk of cardiac toxicity by increasing
the distance between heart and breast.
Allows IMRT treatment of breast/chest wall
DIBH Workflow
Acquire free-breathing CT and DIBH CT w/o change
in position between scans
Treatment plan developed on DIBH CT (can plan a
back-up free breathing plan in case system goes down)
Both Body surfaces and DIBH plan exported Free-breathe surface is used for rough patient positioning
DIBH surface is used to guide patient into position for gated treatment
DIBH Workflow
Verification simulation performed on the linac prior
to initiating treatment Test patient compliance with gated technique
Verify ROIs drawn are appropriate for monitoring with gantry angles
Patient aligned using lasers, marks, SSDs then real-
time deltas using free-breathing surface
The surface is then changed to the DIBH surface,
monitoring of real-time deltas is initiated and the patient
is coached to take a deep breath in until all parameters
are within tolerance
DIBH Workflow
When the patient is at full inspiration Lateral position can be adjusted by moving the table
Vertical position is adjusted by asking the patient to breathe more/less
Longitudinal position typically also controlled by breath size
- Can fine-tune with couch if necessary
SSDs/Port films/KV images all acquired during
breath hold
There is a “Coaching” display option that is helpful
for therapists Displays only Vertical position parameter
Bar moves up and down like chest wall, turns from green to yellow
when all parameters within tolerance (gating window)
DIBH Technique
Coaching view with real-time deltas
Strengths and Weaknesses
Strengths:
Non-ionizing
Therapists find it to be very useful for setup
Faster
More efficient, fewer re-ports
Intra-fraction monitoring capability
Motion management option
Can capture charges for image guidance
Setup
Monitoring
Strengths and Weaknesses
Weaknesses:
ROI selection is very tricky, learning curve
Position readouts can become unreliable when
gantry interferes in camera line of sight
- Especially problematic in gating patients
Learning curve for developing troubleshooting
techniques
The system is very “Black Box”, requires
physicist to develop a thorough QA procedure
Strengths and Weaknesses
Black Box: Manufacturer’s calibrations and
constancy checks give only RMS error
output
TG 147 Report on QA for non-radiographic
radiotherapy localization and positioning
systems has detailed recommendations
TG 147 QA
Thorough guidance on commissioning
Daily QA – Safety & Static Localization
Phantom positioned at isocenter and can track
movement to isocenter from offset
Accuracy should be w/in 2 mm
Monthly QA –
Safety: Gating terminates, Couch motion
Static Localization: Localization test based on
radiographic analysis (i.e. hidden target test)
Dynamic localization: Motion table or manual
couch motion of monthly phantom by known
distances
TG 147 QA
Annual QA -
Safety: Camera mounting secure
Integrity: Check camera settings
Stability: Drift Measurement (min 1 hr),
reproducibility of localization
Static localization: Full end-to-end testing,
translation and rotation correct w/in clinical range
Dynamic (for gating): Use of motion phantom
Data transfer functionality
Quality assurance for nonradiographic radiotherapy
localization and positioning systems: Report of Task
Group 147 Medical Physics, Vol 39, Issue 4
Thank You!!
Any Questions??