AC CURRENTOSCILLOSCOPE PROBE SR661
E N G L I S H User Manual
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.
We guarantee that at the time of shipping your instrument has met its published specifications.
An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.
The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2113.49
Model #: SR661
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com
Table of Contents
1. INTRODUCTION ............................................................................... 21.1 International Electrical Symbols ................................................31.2 DefinitionofMeasurementCategories .....................................31.3 ReceivingYourShipment ..........................................................31.4 OrderingInformation .................................................................4
1.4.1 AccessoriesandReplacementParts ............................4
2. PRODUCT FEATURES ...................................................................... 52.1 Description ................................................................................52.2 Compatibility .............................................................................52.3 Features ....................................................................................6
3. SPECIFICATIONS............................................................................. 73.1 ElectricalSpecifications ............................................................73.2 MechanicalSpecifications ........................................................83.3 EnvironmentalSpecifications ....................................................93.4 SafetySpecifications ................................................................93.5 TypicalResponseCurves .......................................................10
4. OPERATION .................................................................................. 134.1 CurrentMeasurement .............................................................13
5. MAINTENANCE ............................................................................. 145.1 Maintenance ...........................................................................145.2 Cleaning ..................................................................................14
RepairandCalibration ...........................................................................15TechnicalandSalesAssistance ............................................................15Limited Warranty ...................................................................................16WarrantyRepairs ...................................................................................16
2 AC Current Oscilloscope Probe Model SR661
CHAPTER 1
INTRODUCTION
WARNING The safety warnings are provided to ensure the safety of personnel and proper operation of the instrument. Read the instruction completely.
• Usecautiononanycircuit:potentiallyhighvoltagesandcurrentsmaybepresentandmayposeashockhazard.
• Donotusetheprobe ifdamaged.Alwaysconnect thecurrentprobetothemeasuringdevicebeforeitisconnectedaroundtheconductor
• Donotuseonnoninsulatedconductorwithapotentialtogroundgreaterthan600VCATIIIpollution2.Useextremecautionwhenclampingaroundbareconductorsorbusbars.
• Beforeeachuse,inspecttheprobe;lookforcracksinhousingoroutputcableinsulation.
• Donotuseclampinwetenvironmentorinlocationsthathazard-ousgasesexist.
• Donotusetheprobeanywherebeyondthetactilebarrier.
AC Current Oscilloscope Probe Model SR661 3
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or rein-forced insulation.This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.This symbol refers to a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is per-mitted.Probe fitted with an electronic output limiter providing protection against voltage surges caused by the accidental opening of the probe secondary circuit to 30V max. peak.
In conformity with WEEE 2002/96/EC.
1.2 Definition of Measurement CategoriesCAT II: Formeasurements performed on circuits directly connected to
theelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.
1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsareconsistentwith thepacking list.Notify yourdistributor of anymissing items. If theequipmentappearstobedamaged,fileaclaimimmediatelywiththecar-rierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.
4 AC Current Oscilloscope Probe Model SR661
1.4 Ordering InformationACCurrentOscilloscopeProbeModelSR661 .................... Cat. #2113.49Includes a user manual.
1.4.1 Accessories and Replacement Parts
MaleBanana-FemaleBNCAdaptor(XF-SS) .................... Cat. #2111.32
AC Current Oscilloscope Probe Model SR661 5
CHAPTER 2
PRODUCT FEATURES
2.1 DescriptionTheACCurrentOscilloscopeProbeModelSR661expandsoscilloscopeapplicationsinindustrialorpowerenvironments,andisidealforanalysisandmeasurementofdistortedcurrentwaveformsandharmonics.
TheModelSR661permitsaccuratedisplayandmeasurementofcurrentsfrom100mAto1000Arms,1Hzto100kHz(withcurrentderating)withoutbreaking into the circuit.A passive filter eliminates noise, ring on rapidrising(di/dt)waveforms,andensuresaccuratescreendisplays.
Theprobeconnectsdirectlytoanoscilloscopethroughaninsulatedcoax-ialcablewithaninsulatedBNC.
2.2 CompatibilityTheModelSR661iscompatiblewithanyanalogordigitaloscilloscopeorothervoltage-measuringinstrumentwhichhasthefollowingfeatures:
• BNCinputconnector
• Rangecapableofdisplaying0.2to0.5Vperdivision,or2Vrange
• Minimuminputimpedanceof1MΩ
6 AC Current Oscilloscope Probe Model SR661
2.3 Features
4.37" (111mm)
3.98" (101mm)
2.25" (57mm)1.22" (31mm)
8.50" (216mm
)
1.71"(43.5 mm)
1.38"(35 mm)
1.77"(45 mm)
3.90" (99 mm)
2.13" (54.1mm)
1
2
3
5
4
6
1. Jawopeningof2.25"
2. Maximumdimensionofconductis2.13"(54mm)
3. TacticalBarrier.Alwayskeephandsbelowthisbarrieratalltimes.
4. Arrow indicatesdirectionofcurrentflow.Currentflows inapositivedirectionwhenittravelsfromsupplytoload.
5. Threepositionswitch
6. 6.5ft(2m)outputcable
AC Current Oscilloscope Probe Model SR661 7
CHAPTER 3
SPECIFICATIONS
3.1 Electrical Specifications*Referenceconditions:23°C±3°K,20to75%RH,48to65Hz,externalmagneticfield<40A/m,noDCcomponent,noexternalcurrentcarryingconductor,testsamplecentered.
Operating Range: 0.1to2000Apeak
Measurement Range: 100mAto1200Arms(2000Apeak)
Output Signal:mVoutputsignal(2Vpeakmax)100mV/A: 10mAto20Apeak10mV/A: 0.1Ato200Apeak1mV/A: 1Ato2000Apeak
1mV/A, (1V at 1000A)Range Accuracy Phase Shift
1 to 50A
≤1% ± 1mV
N/A50 to 200A ≤3°
200 to 1000A ≤2°1000 to 1200A ≤1°
Overload: 1200Afor40mnON,20mnOFF
10 mV/A, (1V at 100A)Range Accuracy Phase Shift
0.1 to 5A
≤2% ± 5mV
N/A5 to 20A ≤15°
20 to 100A ≤10°100 to 120A ≤5°
Overload:120Acontinuous
100mV/A, (1V at 10 Arms)Range Accuracy Phase Shift
0.1 to 0.5A
≤3% ± 10mV
N/A0.5 to 2A N/A2 to 10A ≤15°10 to 12A
Overload: 12Acontinuous
8 AC Current Oscilloscope Probe Model SR661
Frequency Range (with current derating):1Hzto100kHz(@-3dB)Seetypicalresponsecurves(pages10-11)
Crest Factor:≤6Load Impedance: ≥1MΩ@≤47pF
Rise/Fall Time: <40µS
Working Voltage:600VCATIIICommon Mode Voltage:600VCATIIIInfluence of Adjacent Conductor: <0.2mA/AACInfluence of Conductor Position in Jaw Opening:0.02%ofReadingunder400Hz
Influence of Frequency:Range1mV/A: From10to1000Hz:<1%ofReading From1to10kHz:<2%ofReading From10to50kHz:<10%ofReading From50to100kHz:-3dB
Range10mV/A: From10to1000Hz:<5%ofReading From1to10kHz:<3%ofReading From10to50kHz:<20%ofReading From50to100kHz:-3dB
Range100mV/A: From10to1000Hz:<10%ofReading From1to10kHz:<5%ofReading From10to50kHz:<20%ofReading From50to100kHz:-3dB
3.2 Mechanical SpecificationsDimensions: 4.4x8.5x1.8"(111x216x45mm)
Weight:1.21lbs(550g)Jaw Opening:2.25"(57mm)max
Maximum Cable Diameter:2.13"(54mm)Maximum Conductor Size: Cable:2.05"(52mm)BusBar:1.95x.19"(50x5mm)
Case Protection:IP40(IEC529)
AC Current Oscilloscope Probe Model SR661 9
Drop Test:1m(IEC68-2-32)Mechanical Shock:100g(IEC68-2-27)Vibration: 5to15Hz,0.15mm(IEC68-2-6) 15to25Hz,1mm 25to55Hz,0.25mm
Output: InsulatedleadwithinsulatedBNCConnector,6ft(2m)
3.3 Environmental SpecificationsOperating Temperature/RH:14°to122°F(-10°to50°C)Storage Temperature/RH:-4°to158°F(-20to70°C)
Altitude:Non-operating:0to12,000m Operating:0to2000m
Operating Relative Humidity:0to85%@35°C
Influence of Temperature: ≤0.15%/10K
Influence of Humidity:10to90%:0.1%
3.4 Safety Specifications
E192383
Electrical:DoubleinsulationorreinforcedinsulationbetweentheprimaryorsecondaryandtheoutercaseofthehandleperEN61010-2-032.
600V,CATIII,PollutionDegree:2
Dielectric Strength: 5550V,50/60Hzbetweenprimary, secondary,andtheoutercaseofthehandle
Electromagnetic Compatibility: EN 50081-1 Class B EN50082-2Electrostaticdischarge IEC 1000-4-2 RadiatedfieldIEC1000-4-3 Fast transients IEC 1000-4-4 Magneticfieldat50/60HzIEC1000-4-8
*Specifications are subject to change without notice
10 AC Current Oscilloscope Probe Model SR661
3.5 Typical Response Curves
1mV/A Range:
10 100 1000 10,000 100,000
-4.00
-5.00
1.00
0.00
-1.00
-2.00
-3.00
10 100 1000 10,000 100,000
-35.00
-40.00
5.00
-15.00
-20.00
-25.00
-30.00
-10.00
-5.00
0.00
Accura
cy in %
Phase in °
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency
@ 10 A
@ 50 A
@ 10 A
@ 50 A
AC Current Oscilloscope Probe Model SR661 11
10mV/A Range:
10 100 1000 10,000 100,000
-25.00
-30.00
0.00
-5.00
-10.00
-15.00
-20.00
10 100 1000 10,000 100,000
-60.00
-70.00
10.00
-20.00
-30.00
-40.00
-50.00
-10.00
0.00
5.00
Accura
cy in %
Phase in °
@ 1 A
@ 5 A
@ 1 A
@ 5 A
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency
12 AC Current Oscilloscope Probe Model SR661
100mV/A Range:
10 100 1000 10,000 100,000
10 100 1000 10,000 100,000
-40.00
-45.00
-15.00
-20.00
-25.00
-30.00
-35.00
5.00
-10.00
-5.00
0.00
-40.00
-50.00
10.00
0.00
-10.00
-20.00
-30.00
20.00
30.00
40.00
@ 0.1 A
@ 0.5 A
@ 0.1 A
@ 0.5 A
Accura
cy in %
Phase in °
Frequency (Hz)
Frequency (Hz)
Accuracy vs. Frequency
Phase Shift vs. Frequency
AC Current Oscilloscope Probe Model SR661 13
CHAPTER 4
OPERATION
4.1 Current Measurement
WARNING: Always connect the probe to the instrument before clamping onto the circuit under test.
• Connect the Model SR661 to the proper input channel on theoscilloscope.
• Beginwiththeleastsensitiverangeonthecurrentprobe(1mV/A)
• Selectthe0.5V/Divisionrangeontheoscilloscope.
• Clamptheprobeontheconductortobemeasuredandreadthecurrentflowingdirectlyonyouroscilloscope.
(Remember to unclamp the probe from the conductor before discon-necting it from your meter or instrument.)
You may also use your oscilloscope to amplify the signal while usingthe1mV/Aproberange(whichoffersthebestaccuracyandleastphaseshift).
NOTE: It is possible to change the range on the current probe without remov-ing the probe from the current carrying conductor, but it is important to remember not to exceed the permissible peak ratings of 2000mV peak or 4000 mV peak to peak maximum. The peak ratings by range are: 20A peak @ 100mV/A, 200A peak @ 10mV/A and 2000A peak @ 1mV/A.
14 AC Current Oscilloscope Probe Model SR661
CHAPTER 5
MAINTENANCE
5.1 Maintenance
WARNING• Toensureoptimumperformance,itisimportanttokeeptheprobe
jawmatingsurfacescleanatalltimes.Failuretodosomayresultinerrorinreadings.
• Formaintenanceuseonlyspecifiedfactoryreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservicingunlessyouarequalifiedtodoso.
• Toavoidelectricalshockand/ordamagetotheinstrument,donotallowwaterorotherforeignsubstancesintothecase.
• Disconnecttheunitfromallcircuitsandtestcablesbeforeopen-ingthecase.
• Usecautionwithmetallictoolsthatmayshortbatterypacks,powersupplies,etc.
5.2 Cleaning• Tocleantheprobebody,useasoftclothdampenedinasolution
ofmilddetergentandwater.Tocleanthecore,openthejawandcleantheexposedcoresurfaceswithacottonswabdampenedwith isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol orethanol).Lubricatethejawsmatingsurfaceswithalightoil.
• Do not use chemicals containing benzine, benzene, toluene,xylene,acetone,orsimilarsolvents.
• Donotimmersetheprobeinliquidsoruseabrasivecleaners.
AC Current Oscilloscope Probe Model SR661 15
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.
For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]
(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:
ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail:[email protected] www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.
16 AC Current Oscilloscope Probe Model SR661
Limited Warranty
TheACCurrentOscilloscopeProbeModelSR661iswarrantedtotheownerforaperiodoftwoyearsfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC® Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.
Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do:Ifamalfunctionoccurswithinthewarranty period,youmayreturntheinstrumenttous forrepair,providedwehaveyourwarrantyregistration informationonfile or a proof of purchase.AEMC® Instrumentswill, at its option, repair orreplacethefaultymaterial.
REGISTER ONLINE AT: www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments15FaradayDrive•Dover,NH03820USAPhone:(800)945-2362(Ext.360)
(603)749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]
Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.
AC Current Oscilloscope Probe Model SR661 17
Notes:
10/17
99-MAN100130v13
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346