38
Adventurer Pro Grain Test Scale Instruction Manual
Transcript
Page 1: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleInstruction Manual

1 Front cover english.pmd 7/27/2006, 3:11 PM1

Page 2: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Note: This equipment has been tested and found to comply with the limits for a Class A digitaldevice, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonableprotection against harmful interference when the equipment is operated in a commercial environment.This equipment generates, uses, and can radiate radio frequency energy and, if not installed andused in accordance with the instruction manual, may cause harmful interference to radiocommunications. Operation of this equipment in a residential area is likely to cause harmfulinterference in which case the user will be required to correct the interference at his own expense.

This Class A digital apparatus complies with Canadian ICES-003.Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.

ISO 9001 RegistrationIn 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 byBureau Veritus Quality International (BVQI), confirming that the Ohaus quality managementsystem is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, OhausCorporation, USA, was re-registered to the ISO 9001:2000 standard.

Mark Standard

Compliance to the following standards is indicated by the corresponding marking on the product.

CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1

1 Front cover english.pmd 7/27/2006, 3:11 PM3

Page 3: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-1Adventurer Pro Grain Test Scale

1. INTRODUCTION ........................................................................................ EN-31.1 Description .............................................................................................. EN-31.2 Features .................................................................................................. EN-31.3 Safety Precautions ................................................................................... EN-3

2. INSTALLATION ......................................................................................... EN-42.1 Unpacking .............................................................................................. EN-42.2 Installing Components ............................................................................. EN-4

2.2.1 Assembly .................................................................................... EN-42.3. Selecting the Location .............................................................................. EN-52.4 Leveling the Scale .................................................................................... EN-52.5 Connecting Power .................................................................................... EN-5

2.5.1 AC Adapter ................................................................................... EN-52.5.2 Turning Power On and Off ............................................................. EN-5

2.6 Initial Calibration ...................................................................................... EN-5

3. OPERATION .................................................................................................. EN-73.1 Overview of Controls ................................................................................ EN-7

3.1.1 Model AVD4102N ......................................................................... EN-73.2 Overview of Display Functions .................................................................. EN-83.3 Button Control Functions .......................................................................... EN-93.4 Using the Button Control Functions .......................................................... EN-10

3.4.1 Setting the Scale to Zero ............................................................ EN-103.4.2 Taring ....................................................................................... EN-103.4.3 Changing Units of Measure ......................................................... EN-103.4.4 Printing Data .............................................................................. EN-10

3.5 Menu ................................................................................................ EN-113.5.1 Menu Structure ........................................................................... EN-113.5.2 Menu Navigation ........................................................................ EN-123.5.3 Changing Settings ...................................................................... EN-12

3.6 Application Modes ................................................................................. EN-133.6.1 Weighing ................................................................................... EN-133.6.2 Test Weight Mode ....................................................................... EN-133.6.3 Dockage Mode .......................................................................... EN-14

3.7 Scale Settings ........................................................................................ EN-163.7.1 Calibration ................................................................................. EN-163.7.2 Calibration Masses ..................................................................... EN-183.7.3 Setup ........................................................................................ EN-193.7.4 Readout .................................................................................... EN-193.7.5 Mode ........................................................................................ EN-193.7.6 Unit ........................................................................................... EN-193.7.7 Print-1 and Print-2 ...................................................................... EN-203.7.8 RS232-1 and RS232-2 ............................................................... EN-21

TABLE OF CONTENTS

2 toc.pmd 7/27/2006, 3:11 PM1

Page 4: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-2

3.7 Scale Settings (Cont.)3.7.9 GLP Data ................................................................................... EN-213.7.10 GLP Print ................................................................................... EN-223.7.11 Reset ......................................................................................... EN-223.7.12 Lockout ..................................................................................... EN-233.7.13 End ........................................................................................... EN-24

3.8 Legal for Trade (LFT) ............................................................................. EN-243.8.1 Setting the Scale ......................................................................... EN-243.8.2 Sealing the Scale ....................................................................... EN-24

3.9 Printing Data ......................................................................................... EN-25

4. MAINTENANCE .............................................................................................. EN-274.1 Calibration ............................................................................................ EN-274.2 Cleaning ............................................................................................... EN-274.3 Troubleshooting .................................................................................... EN-274.4 Service Information ................................................................................ EN-284.5 Parts ................................................................................................ EN-284.6 Accessories .......................................................................................... EN-29

5. TECHNICAL DATA ........................................................................................... EN-315.1 Outline Drawing ..................................................................................... EN-315.2 Specifications ........................................................................................ EN-325.3 Capacity x Readability Table ................................................................... EN-325.4 Communication ..................................................................................... EN-33

5.4.1 Commands ............................................................................... EN-335.4.2 Connections .............................................................................. EN-34

Limited Warranty ............................................................................................... EN-35

TABLE OF CONTENTS (Cont.)

2 toc.pmd 7/27/2006, 3:11 PM2

Page 5: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-3

1 INTRODUCTIONThis manual contains installation, operation and maintenance instructions for the Ohaus AdventurerPro Grain Test Scale. Please read the manual completely before using the scale.

Certificate of Conformance 05-118 has been issued by the National Conference on Weightsand Measures, confirming that the Adventurer Pro models listed in this manual conform to therequirements in NCWM Publication 14 and NIST Handbook 44.Before this scale can be used in commercial trade, it must be inspected by local weights andmeasure authorities. It is the responsibility of the purchaser to ensure that all pertinent legalrequirements are met. Please contact your local weights and measures office for further details.

1.1 DescriptionThe Adventurer Pro Grain Test Scale is a precision weighing instrument that will provide youwith years of service if properly cared for. The Adventurer Pro Grain Test Scale has a capacityof 4100 grams.

1.2 FeaturesThe Adventurer Pro Grain Test Scale includes many standard features. They are:

• AC Adapter operation (AC Adapter included)• Weighing, Test Weight, Dockage application modes.• Protective In-use cover• RS232 interface• Integral security bracket

The Adventurer Pro Grain Test Scale also offers optional features• USB interface• Second RS232 Interface

1.3 Safety PrecautionsPlease follow these safety precautions:

• Verify that the input voltage printed on the AC Adapter matches the local AC powersupply.

• Use the scale only in dry locations.• Do not operate the scale in hostile environments.• Do not drop loads on the weighing pan.• Service should be performed only by authorized personnel.

3 page 3 to 6.pmd 7/27/2006, 3:11 PM3

Page 6: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-4

2 INSTALLATION2.1 UnpackingCarefully remove your Adventurer Pro Grain Test Scale and each of its components from thepackage. Save the packaging to ensure safe storage and transport.

2.2 Installing ComponentsUse the illustration and instructions below to identify and assemble your Adventurer Pro GrainTest Scale with its components. All components must be installed before using the scale.

2.2.1 Assembly1. Insert the Sub-platform on the Mounting Cone located in the center of

the scale. Align the Sub-platform so that it is fully seated on the cone.

2. Place the Weighing Pan on the Sub-platform.

Sub-platform

Mounting Cone

Weighing Pan

3 page 3 to 6.pmd 7/27/2006, 3:11 PM4

Page 7: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-5

The scale performs a segment check. The scale thendisplays the last selected application mode.

To turn power off, press and hold the Off button until OFFis displayed, then release the button.

2.5.2 Turning Power On and OffPress the On O/T button to turn power on.

2.6 Initial CalibrationWhen the scale is first installed, and when it is moved to another location, it must be calibratedto ensure accurate weighing results. Have the appropriate calibration masses available beforebeginning calibration. Refer to Section 3.7.1and 3.7.2 for masses and calibration procedure.

2.3 Selecting the LocationLocate the scale on a firm, steady surface. Avoid locations with excessive air current, vibrations,heat sources, or rapid temperature changes.

2.4 Leveling the ScaleBefore the scale is used, the feet should be adjusted so that the scale is level with the surface.This will enable accurate weighing. You will see a level bubble indicator in a small roundwindow on the front of the scale. Level the scale by adjusting the leveling feet so the bubble iscentered in the circle, as seen below.

Balance Level

2.5 Connecting Power2.5.1 AC AdapterConnect the AC Adapter to the wall outlet. Connect the plug into the receptacle on the rear ofthe scale.

For use with CSA Certified (or equivalent approved) power source, which musthave a limited circuit output.

3 page 3 to 6.pmd 7/27/2006, 3:11 PM5

Page 8: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-6

3 page 3 to 6.pmd 7/27/2006, 3:11 PM6

Page 9: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-7

3 OPERATION3.1 Overview of Controls3.1.1 Model AVD4102N

2 3

1

8

5

7

1411

10

4

6

Rear View

12

Top View Bottom View

15

13

TABLE 3-1. ADVENTURER PRO GRAIN TEST SCALE CONTROLS.

No. Description1 ON / O/T / Off / Yes Button2 UNIT / Print / No Button3 Level Bubble4 % / Recall / Back button5 TEST WEIGHT / Menu-Cal / Exit Button6 Display7 Pan8 Type label on side of housing9 Data label10 Leveling feet11 Power Connector12 COM 2 Connector (AV...R and AV...U Models only)13 Security Loop14 Lock Switch15 COM 1 Connector

9

4 page 7 to 8.pmd 7/27/2006, 3:11 PM7

Page 10: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-8

3

Digital Display Window

3.2 Overview of Display Functions

TABLE 3-2. ADVENTURER PRO GRAIN TEST SCALE DISPLAY FUNCTIONS.

1 2

45

678910

12

1314

15

11

No. Description 1 Primary (7 segment) display 2 Brackets 3 Units of Measurement 4 Memory Indicator 5 Battery Strength Indicator (not used) 6 Pieces Indicator (not used) 7 Secondary (14 Segment) Display 8 Gross Indicator (not used) 9 Brutto Indicator (not used)10 Tare Indicator (not used)11 Preset Tare Indicator (not used)12 Net Indicator (not used)13 Stability Indicator14 Negative Indicator15 Center of Zero Indicator

4 page 7 to 8.pmd 7/27/2006, 3:11 PM8

Page 11: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-9

Back

%•Store the reference weight.

Recall•Recalls reference

weight frommemory.

Back•Reverts back to

previous menuitem.

•Decrements avalue beingentered.

Primary Function

Secondary Function

Menu Function

3.3 Button Control FunctionsFour multifunction buttons are used to to operate the scale and navigate the menus. Thefunction of each button is shown below.

Press and Release

Menu Function

On O/T• If scale is off, turns scale on.

• If scale is on,sets zero.

Off•Turns scale off.

Yes•Accepts the

current (blinking)setting on thedisplay.

UNIT•Change

weighing unit

Print•Sends current

display value tothe serialinterface.

No•Rejects the

current setting(blinking) on thedisplay.

• Increments avalue beingentered.

TEST WEIGHT• Activates selected

Test Weight unit.

Menu-Cal• Enter the User

menu.Calibration is thefirst sub-menu.

Exit• Immediately

exits menu mode.

• Aborts acalibration inprogress.

Press and Hold

Yes No Back Exit

5 page 9 to 15.pmd 7/27/2006, 3:11 PM9

Page 12: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-10

3.4 Using the Button Control Functions3.4.1 Setting the Scale to ZeroRemove the load from the pan and press the ON O/T button to set the display to zero.The Center of Zero indicator turns on when the measurement is within + 1/4 d of the zerosetting.

3.4.2 TaringTaring refers to the action of adjusting out the weight of a container so that only the weight ofobjects held in the container (net weight) is displayed.

To TarePlace the empty container on the pan and press the ON O/T button.Add material to the container. The net weight of the material is displayed.To clear the Tare value, remove the container from the pan and press the ON O/T button.

Preset TareUse the xT command in the command table to enter a preset tare value through a computer.To clear the Tare value, enter a value of 0.0. See Section 5.4.1.

3.4.3 Changing Units of MeasureThe Adventurer Pro Grain Test Scale can be configured to measure in a variety of units, includingcustom units. The Unit Sub-menu is used to enable or disable a specific unit.To select a unit of measure:Momentarily press the UNIT button, then release it when the desired unit is displayed.Note: If the desired unit is not displayed, it must be turned on in the Unit menu (See Section3.7.6).

3.4.4 Printing DataPress and hold the Print button to send the displayed value to the COM port (See Section3.7.7).Note: The port must have Auto Print set to OFF.

5 page 9 to 15.pmd 7/27/2006, 3:11 PM10

Page 13: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-11

READOUT (3.8.4)Auto Zero Off, .5d, 1d,

2d,5d

Filter Low, Med, HiEnd Read

MODE (3.8.5)Weigh On, Off

Test Weight lb/bu (qt),lb/bu (pt),kg/hL (L),kg/hL (.5L),

Dockage Pos,NegEnd Mode

UNIT (3.8.6)g On,Offkg On,Offoz On,Offlb On,OffGN On,OffCustom On,OffEnd Unit

RESET (3.8.11)Setup Reset ?Readout Reset ?Mode Reset ?Unit Reset ?Print-1 Reset ?Print-2 Reset ?RS232-1 Reset ?RS232-2 Reset ?GLP Data Reset ?GLP Print Reset ?Lockout Reset ?Global Reset ?End Reset

LOCKOUT (3.8.12)Cal On,OffSetup On,OffReadout On,OffMode On,OffUnit On,OffPrint-1 On,OffPrint-2 On,OffRS232-1 On,OffRS232-2 On,OffGLP Data On,OffGLP Print On,OffReset On,OffLock Set On,OffEnd Lockout

END (3.8.13)

RS232-1 (3.8.8)Baud 600--2400--19,200Parity 7Even, 7Odd, 7 No

par, 8 No parHandshake Off, XONXOFFEnd RS1

RS232-2 (Optional)Baud 600--2400--19,200Parity 7Even, 7Odd, 7 No

par, 8 No parHandshake Off, XONXOFFEnd RS1

GLP DATA (3.8.9)User ID SetProj ID SetTime Type, Set, AdjDate Type, SetEnd GLP Data

GLP PRINT (3.8.10)Time On, OffBalance ID On, OffUser ID On, OffProject ID On, OffDifference On, OffName On, OffEnd GLP Prt

PRINT-2 (Optional)Output When Stable

GLP TareAuto Print Off, Cont, Int, StabContent Num Only Header,

Weight, Ref, Result, GLP

Layout Line Format 4LF Form Feed

ListEnd Print-2

3.5.1 Menu Structure

CALIBRATION (3.8.1)SpanLinearityCalibration Test (CalTest)End Cal

PRINT-1 (3.8.7)Output When Stable

GLP TareAuto Print Off, Cont, Int, StabContent Num Only Header,

Weight, Ref, Result,GLP

Layout Line Format4LFForm Feed

ListEnd Print-1

SETUP (3.8.3)Legal Trade Off, Comm,

GIPSAAuto Off Off, 1,2,5End Setup

The Adventurer Pro Grain Test Scale menu structure is illustrated below.

3.5 MENU

5 page 9 to 15.pmd 7/27/2006, 3:11 PM11

Page 14: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-12

3.5.2 Menu NavigationThe Adventurer Pro Grain Test menu structure consists of three levels. The top level is the mainmenu which consists of a number of sub-menus.

CALIBRATION SETUP READOUT MODE UNIT PRINT-1 PRINT-2

END LOCKOUT RESET GLP PRINT GLP DATA RS232-2 RS232-1

TOP LEVEL

END RS232 POWER

BAUDHANDSHAKE

PARITY

7 EVEN8 NOP

7 ODD7 NOP

SECOND LEVEL

BOTTOM LEVEL

The second level of the menu structure is thesub-menu level. Each sub-menu is a loop thatconsists of a number of menu items.

The bottom level of the menu structure is themenu item level. Each menu item level consistsof two or more settings

Typical Sub-menu

Typical Menu Item

Typical Setting

3.5.3 Changing SettingsTo change a menu setting, perform the following steps:ENTER THE MENUPress and hold the Menu button until MENU appears on the primary (7-Segment) display.Release the button and the first sub-menu appears on the secondary (14-Segment) display(blinking).

Note: When the secondary display is blinking, press Yes to accept.Press No or Back or view another selection.

SELECT THE SUB-MENUPress the No button to select the next sub-menu or the Back button to select the previous sub-menuin the loop. Press the Yes button to select the sub-menu that appears on the secondary display(blinking). The sub-menu name now appears on the primary display and the first menu item in thesub-menu now appears on the secondary display (blinking).

SELECT THE MENU ITEMPress the No button to select the next setting or the Back button to select the previous setting in theloop. Press the Yes button to select and save the setting on the secondary display (blinking) as thenew menu item setting.

EXIT THE MENUPress the Exit button at any time to quickly exit the menu or select END sub-menu and press the Yesbutton.

Enter

5 page 9 to 15.pmd 7/27/2006, 3:11 PM12

Page 15: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-13

WeighPress the UNIT button repeatedly (short presses) until thedisplay shows 0.00g.Press ON O/T button to zero the balance.Place objects to be weighed on the pan to display the weight.The example displays a 200 gram weight.

3.6 Application ModesThe Adventurer Pro Grain Test Scale incorporates Weighing, Test Weight and Dockage applicationmodes. The default setting has the Weighing mode turned on, Test Weight to lb/bu (qt) andDockage set to positive.

3.6.1 WeighingUse this mode to determine the weight of items in the selected unit of measure. The AdventurerPro Grain Test Scale is shipped with grams enabled. Before using other units of measure,these units must be turned on in the Unit menu ( See Section 3.7.6 ). Grams is the defaultsetting.

3.6.2 Test Weight ModeThe Test Weight Mode allows the measurement of the density of a sample using one of fourstandard container sizes. Test Weight values can be displayed as follows:

lb/bu (qt) = pounds per bushel based on a quart container.lb/bu (pt) = pounds per bushel based on a pint container.kg/hL (L) = kilograms per hectaliter based on a liter container.kg/hL (.SL) = kilograms per hectaliter based on a half liter container.

Before starting, enable the desired test weight mode in the Mode Menu (See Section 3.7.5).

Press the TEST WEI GHT button. The scale enters the TestWeight mode. the selected test weight unit is indicated onthe second line of the display.

Place the empty container on the scale pan and press theON O/T button to tare the container.

Remove the container from the scale and fill it with the testsample.

Place the container with sample on the pan. The primarydisplay shows the actual test weight (density) with the unitshown on the secondary display.

To clear scale entry, press the ON O/T button.

5 page 9 to 15.pmd 7/27/2006, 3:11 PM13

Page 16: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-14

3.6.3 Dockage ModeDockage is material other than the predominant grain that can be easily removed with sievesand cleaning devices. The dockage mode is used to determine the percent of waste materialcontained in the sample. Before starting, select the test weight container and dockage type inthe Mode Menu (See Section 3.7.5).

Store Dockage Reference WeightPlace the container on the pan and tare by pressing theON O/T button. The tare weight is stored in memory.

Fill the container with sample material and press the % button.If the current weight is less than the minimum allowedreference weight (10 grams), the secondary displayalternates between “MIN WT ERR” and “NOT STORED” twicewhile returning to current mode.The scale then returns to the previous mode if the minimumweight is not accepted as shown.

If the sample weight is greater than or equal to 10 grams, the sample weight is established asthe Dockage Reference weight and is stored in memory as the 100% reference weight for thedockage calculation. Additionally, the tare weight is stored in memory as the dockage tareweight. With the sample still on the pan, the primary display shows 100.00 if Positive Dockageis selected in the Dockage menu. If Negative Dockage is selected, the primary display shows0.00. The user can process the sample to remove any undesirable material. After processing,the user will place the container back on the pan, and the display will read the dockage value.

The % display icon indicates percent and the MEM icon indicates a dockage reference weightis stored in memory. A display icon also annunciates stability. The secondary display shows+DOCKAGE or -DOCKAGE.

Remove container from the scale. Scale maynow be used to weigh by pressing the UNITbutton. The MEM icon indicates that aDockage Reference Weight is stored.

5 page 9 to 15.pmd 7/27/2006, 3:11 PM14

Page 17: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-15

Recall Dockage Reference WeightIf a Dockage Reference Weight is stored in memory (MEM icon is visible), it can be recalledfrom memory by a long press of the Recall button. This may become necessary if the scale isused for normal weighing or test weighing while the sample is being processed.

To recall stored dockage reference weight, the container withthe processed sample is placed on the pan and the referenceweight is recalled from memory by a long press of the Recallbutton. The primary display shows the percent dockage ofthe sample based on the previously established dockagereference weight and dockage tare weight.

Percent Dockage DisplayThe percent dockage display capacity is fixed at 410.09%.The percent dockage display readability is fixed at 0.01%

To Clear Dockage Reference WeightWhile in the Dockage mode, press the ON O/T button. Thiswill clear the existing reference weight.

5 page 9 to 15.pmd 7/27/2006, 3:11 PM15

Page 18: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-16

3.7 Scale SettingsRefer to Section 3.5.2 to enter and navigate the menus.

3.7.1 Calibration

Adventurer Pro Grain Test Scale offer a choice of three calibration methods: Span Calibration,Linearity Calibration and Calibration TestTM.

• SpanSpanSpanSpanSpan - Span calibration ensures that the scale reads correctly withinspecifications using two weight values: zero and a weight valuebetween 25% and 100% of the scale’s capacity.

• LinearityLinearityLinearityLinearityLinearity - Linearity calibration minimizes deviation between actual and displayedweights within the scale’s weighing range. Three weight values areused: zero, a weight value at midpoint of the scale's weighing range,and a weight value at or near the scale’s specified capacity.

• Cal Test Cal Test Cal Test Cal Test Cal Test - Calibration test allows the stored calibration data to be tested against thecurrent mass being used for the test.

Notes: Calibration may be locked out to prevent unauthorized personnel from changingcalibration.

Before beginning calibration, have masses available. Masses required for calibrationare listed in the table in section 3.7.2. Default Span Calibration Points are shown inBold print.

6 page 16 to 18.pmd 7/27/2006, 3:11 PM16

Page 19: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test Scale EN-17

With the scale on, press and hold the Menu-Cal buttonuntil MENU is displayed, then release the button. CALIBRATE(blinking) is displayed.

Press the Yes button to enter span calibration.

Press the Yes button to initiate span calibration.

First the zero reading is taken. Next the display shows thedefault span calibration weight value.

Press No to change to an alternate calibration weight. PressYes when the desired calibration weight is displayed.

Place the specified calibration weight on the pan. Aftercalibration, remove the weight from the pan.

To abort calibration at any time, press the Exit button.

Span Calibration

NOTICE:Linearity Calibration is required only if the linearity error exceeds tolerance inspecification table.

Linearity Calibration

With the scale on, press and hold the Menu-Cal buttonuntil MENU CALIBRATE is displayed, then release the button.CALIBRATE (blinking).

Press the Yes button to enter calibration.

Press the Yes button and follow the screen instructions.Use the weight values shown on the primary display.

6 page 16 to 18.pmd 7/27/2006, 3:11 PM17

Page 20: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-18

First the zero reading is taken. Next the display shows thefirst calibration weight value. PUT WEIGHT (blinking) isdisplayed. Place the specified calibration weight on thepan.

After a few seconds, place the second specified calibrationweight on the pan. PUT WEIGHT (blinking) is displayed.After a few seconds, the calibration weight is displayed.

LINEAR DONE is momentarily displayed when calibration iscompleted.

After calibration, remove weight from pan. To leave themenu, press Exit.

To abort calibration at any time, press the Exit button.

3.7.2 Calibration Masses

Model Linearity Span Weight Class Calibration Points Calibration Points ASTM OIML

AVD4102N 2000g / 4000g 2000g, 3000g, 4000g 1 E2

Calibration TestCalibration test allows a check of a known calibration massagainst the last stored calibration information in the scale.

Press and hold the Menu-Cal button until MENU CALIBRATEis displayed.

Press Yes to enter calibration.Press No until CAL TEST is displayed.

Press Yes and follow the screen instructions.

Place the specified weight on the pan. After a short period,the display indicates the difference in weight from the lastcalibration, and then will display the calibration weight onthe pan. After the test, remove the calibration weight from thepan. To leave the menu, press Exit.

Linearity Calibration (Cont.)

6 page 16 to 18.pmd 7/27/2006, 3:11 PM18

Page 21: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-19Adventurer Pro Grain Test Scale

3.7.3 Setup

Legal TradeWhen Comm is selected, the scale is suitablefor commercial grain inspection. When GIPSAis selected, the scale is suitable for officialGIPSA grain inspection.

Auto OffWhen set to 1, 2 or 5 minutes, the scale willturn off in the selected time interval if there hasbeen no activity. Set to OFF for continuousoperation.

End SetupPress the Yes button, to advance to the Readoutsubmenu. Press the No button to return to theLEGAL TRADE menu item.

Auto ZeroAllows setting the scale auto zero level setting:.5, 1, 2 or 5 divisions. Scale maintains zerountil threshold is exceeded.

FilterAllows setting the scale filter level: Low,Medium, High. Compensates for vibration orexcessive air currents.

READOUTAuto Zero Off, .5d, 1d, 2d, 5dFilter Low, Medium, HighEnd Read

3.7.4 Readout

3.7.5 ModeThe Mode submenu is used to set Applicationmodes parameters.

MODEWeigh On/OffTest Weight lb/bu (qt)

lb/bu (pt)kg/hL (L)kg/hL (.5L)

Dockage Pos, NegEnd Mode

WeighUse to set the Weighing Mode On or Off.

Test WeightUse to set the Test Weight to: lb/bu (qt),lb/bu (pt),kg/hL (L) or kg/hL (.5L)

DockageUse to set Dockage Mode to: Positive ornegative.

End ModePress the Yes button to advance to the Unitsubmenu. Press No to return to Weigh.

3.7.6 UnitThe Unit submenu is used to turn measuringunits On or Off.

End ReadPress the Yes button to advance to the Modesubmenu. Press No to return to Auto Zero.

UNITg On/Offkg On/Offoz On/Offlb On/OffGN On/OffCustom On/OffEnd Unit

SETUPLegal Trade Off, Comm,

GIPSAAuto Off Off, 1,2,5End Setup

7 page 19 to 30.pmd 7/27/2006, 3:11 PM19

Page 22: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-20

Defining Custom UnitSet Custom to ON in the Unit menu to enableand define the Custom Unit. The custom unit isdefined by a conversion factor and a leastSignificant Digit (LSD). The conversion factoris used by the balance to convert grams to thecustom weighing unit and is defined by enteringa factor and an exponent. The factor is a valuebetween 0.1000000 and 1.999999 inclusive.The exponent moves the decimal point of thefactor to the right for positive values or to the leftfor negative values.

Custom Unit = Conversion Factor x Grams.

The LSD is the value by which the displayedweight is incremented or decremented.

LSD Result.5 Adds one decimal place

Display counts by 5 1 Display counts by 1 2 Display counts by 2 5 Display counts by 5 10 Display counts by 10100 Display counts by 100

End UnitPress the Yes button to advance to the Print-1submenu. Press No to return to the first availableunit.

3.7.7 Print-1 and Print-2The Print-1 submenu is used to set printingparameters for an external printer or computer.

Print-2 contains an identical menu and is usedwhen the optional COM2 port is installed.

Enter the Factor, following the instructions forEditing Limits in Section 3.5.4. Enter theExponent and LSD using the Yes No buttons.

Factor Exponent Conversion (+3 to -3) Factor

.1234 3 123.4

.1234 2 12.34

.1234 1 1.234

.1234 0 .1234

.1234 -1 .01234

.1234 -2 .001234

.1234 -3 .0001234

PRINT-1Output When Stable On/Off

GLP Tare On/OffAuto Print Off, Cont., Interval, When Stable

1 -3600 seconds Stable- Load, Load & Zero

Content Num Only-Off, Header, Weight,Ref-Off, Result-On,GLP-Off- all othersOff

Layout Line Format -Multi, Single4LF-Yes/NoForm Feed-Yes/No

List Yes/NoEnd Print-1

OutputSet When Stable to On to print only stable values.Set When Stable to Off to print stable or unstablevalues.Set GLP Tare to On to print GLP data once aftera tare operation. Set GLP Tare to Off to disablethis feature.

Auto PrintWhen set to Continuous, the displayed value isprinted continuously. When set to Interval, thedisplayed value is printed at the user specifiedtime interval (1 to 3600 seconds).If set to When Stable, the scale willautomatically print the displayed value whenstability is achieved. An additional settingmust be made to determine if only stable non-zero values will be printed (load setting) or ifstable zero and non-zero values will be printed(Load & Zero). When set to Off, the Auto printfeature is disabled.

7 page 19 to 30.pmd 7/27/2006, 3:11 PM20

Page 23: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-21Adventurer Pro Grain Test Scale

3.7.7 Print-1 and Print-2 (Cont.)

LayoutDetermines the format of data output to a printeror computer. If Line Format is set to Multi, amulti-line printout is generated. If it is set toSingle, a single line printout is generated. If 4LF is set to yes, 4 line feeds are appended to theprintout. If Form feed is set to Yes, a form feed isappended to the printout. This is useful forprinting to page printers.

ListWhen Yes is selected, a printout of scalesettings is generated.

End Print-1Press the Yes button to advance to the nextmenu, Print-2 if COM2 is installed or RS232-1 if COM2 is not installed. Pressing No returnsto Output menu item.

3.7.8 RS232-1 and RS232-2The RS232-1 submenu is used to setcommunication parameters for an external printeror computer.

RS232-2 contains an identical menu and isavailable when the optional COM2 port isinstalled.

RS232-1Baud 600..2400...19200Parity 7 Even, 7 Odd, 7 No

Par, 8 No ParityHandshake Off, XONXOFF

HardwareEnd RS1

ContentAll of these features can be set On or Off.Numeric data only, Header, Weight, Reference,Result, GLP. (See sample printout 3.9)

GLP DATAUser ID Set...Proj ID Set...Time Type-12hr, 24hr,

Set..., Adj-60, +60Date Type mdy,...dym

Set...End GLP Data

User IDUsed to enter a User ID up to 10 characters.

Project IDUsed to enter a Project ID up to 10 characters.

BaudBaud rates of 600, 1200, 2400, 4800, 9600and 19,200 are available.

ParityParity settings of 7 even, 7 Odd, 7 No Parityand 8 No Parity are available.

HandshakeSettings of Off, XONXOFF and (for RS232-1only) Hardware are available.

End RS1 (RS2)Pressing the Yes button will advance to theRS232-2 (or GLP Data) submenu. PressingNo returns to the Power (or Baud) menu item.

3.7.9 GLP Data

7 page 19 to 30.pmd 7/27/2006, 3:11 PM21

Page 24: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-22

Entering a User ID or Project IDEnter the user or Project ID one character at atime. The character to be entered is highlightedby a blinking cursor (underscore). Press the Nobutton repeatedly to scroll through the list ofavailable characters: (space), -, 0 through 9, Athrough Z. Press the Yes button to select thedisplayed character and move the cursor oneposition to the right. After the tenth character isentered, the ten character ID will blink. Pressthe No button to change the displayed ID or theYes button to accept the ID and proceed to thenext menu item.

TimeTypeSet the type to 12 hours or 24 hours.

SetSet the current time using the format selectedin the type menu item.

AdjustEnter an adjustment value from -60 to + 60seconds per day.

DateSet the date type: M/D/Y, D/M/Y, Y/M/D,M/Y/D, Y/D/M, D/Y/M and the actual date.

End GLP DataPress the Yes button to advance to the GLPPrint submenu. Press the No button to return toUser ID.

3.7.10 GLP PrintSelect GLP items to be printed by setting themto On.

GLP PRINTTime On/OffBalance ID On/OffUser ID On/OffProject ID On/OffDifference On/OffName On/OffEnd GLP Prt

RESETSetup RESET?Readout RESET?Mode RESET?Unit RESET?Print-1 RESET?Print-2 RESET?RS232-1 RESET?RS232-2 RESET?GLP Data RESET?GLP Print RESET?Lockout RESET?GlobalRESET?End Reset

SetupSelect Yes to return all Setup menu items to theirfactory settings.

ReadoutSelect Yes to return all Readout menu items totheir factory settings.

ModeSelect Yes to return all Mode menu items to theirfactory settings.

3.7.11 Reset

End GLP PrintPress the Yes button to advance to the resetsubmenu. Press the No button to return to Time.

7 page 19 to 30.pmd 7/27/2006, 3:11 PM22

Page 25: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-23Adventurer Pro Grain Test Scale

UnitSelect Yes to return all Unit menu items to theirfactory settings.

Print-1Select Yes to return all Print-1 menu items totheir factory settings.

Print-2 (If COM2 is installed)Select Yes to return all print-2 menu items totheir factory settings.

RS232-1Select Yes to return all RS232-1 menu items totheir factory settings.

RS232-2 (If COM2 is installed)Select Yes to return all RS232-2 menu items totheir factory settings.

GLP DataSelect Yes to return all GLP Data menu items totheir factory settings.

GLP PrintSelect Yes to return all GLP Print menu items totheir factory settings.

LockoutSelect Yes to return all Lockout menu items totheir factory settings.

GlobalSelect Yes to return the menu items in all sub-menus to their factory settings.

End ResetPress the Yes button to advance to the Lockoutmenu. Press the No button to return to Setup.

3.7.12 LockoutUse this submenu to prevent unauthorizedchanges to menu settings. If a submenu islocked, its menu item settings can be viewedbut not changed.

LOCKOUTCal On/OffSetup On/OffReadout On/OffMode On/OffUnit On/OffPrint-1 On/OffPrint-2 On/OffRS232-1 On/OffRS232-2 On/OffGLP Data On/OffGLP Print On/OffReset On/OffLock Set On/OffEnd Lockout

CalSet On to lock and hide the Calibration Menu.

SetupSet On to lock the Setup Menu.

ReadoutSet On to lock the Readout Menu.

ModeSet On to lock the Mode Menu.

UnitSet On to lock the Unit Menu.

Print-1/2Set On to lock the Print-1/2 Menu.

RS232-1/2Set On to lock the RS232-1/2 Menu.

GLP DataSet On to lock the GLP Data Menu.

GLP PrintSet On to lock the GLP Print Menu.

7 page 19 to 30.pmd 7/27/2006, 3:11 PM23

Page 26: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-24

ResetSet On to lock the Reset Menu.

Lock SetSet On to lock all submenu lock settings. Whenset Off, the lock settings for all menus areaccessible. To turn Lock Set off, refer to section

End LockoutPress the Yes button to advance to the Endmenu. Press the No button to return to the Calmenu.

3.7.13 EndThe End menu is used to exit the menus andreturn to previous application

3.8 Legal for Trade (LFT)The Adventurer Pro Grain Test Scale complieswith NTEP regulations and is suitable forcommercial and GIPSA grain testingapplications.

3.8.1 Setting the Scale

• The Calibrate Sub-menu is hidden andnot available.

• In the Setup Sub-menu, the Legal Trademenu item is locked.

• In the Setup Sub-menu, the Auto Zero(AZSM) menu item is set to 0.5d andlocked.

• In the Readout Sub-menu, the GrossIndicator menu item setting is locked.

• The Mode Sub-menu is locked.• In the Unit Sub-menu, the Custom Unit

menu item is hidden and not available.• In the Print Sub-menu, the Output When

Stable menu item is set to ON and locked.• In the Print Sub-menu, the Continuous

setting in the Auto Print menu item ishidden and not available.

• In the Reset Sub-menu, the Reset Readout,Reset Mode, Reset Unit and Reset Globalmenu items are locked.

• In the Lockout Sub-menu, the Calibrationmenu item is locked.

To regain access to the locked Menu settings,turn the scale off. Turn the scale on and pressthe Lock Switch while OHAUS is displayedduring power up.

Press and hold the Menu-Cal button until MENUis displayed. Move to the Setup Sub-menuusing the No button. When SETUP isdisplayed, press the Yes button to enter theSetup Sub-menu. When LEGAL TRADE isdisplayed, press the Yes button to view thesetting OFF. Press the Yes button to changethe setting to Comm or GIPSA. Press the Exitbutton to store the setting and return to thecurrent weighing mode.Note: Before setting the scale, enter the Menuto perform a calibration and make any desiredbalance setting adjustments. Once the scaleis set, the following conditions apply:

3.8.2 Sealing the ScaleTo seal the scale, cover the Lock Switchaccess hole by installing the Security Coverand sliding it to the right. Then install a wirethrough the holes in the tabs of the Securitycover and Top Housing.

Wire Seal Security Cover

7 page 19 to 30.pmd 7/27/2006, 3:11 PM24

Page 27: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-25Adventurer Pro Grain Test Scale

3.9 Printing DataPrinting data to an external computer or printer requires that the communication parameters inthe RS232-1 and/or RS232-2 submenus be set to match external device communicationparameters.

Weigh ModeOHAUS // Header Line 1Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed

// If Print -x- > Content GLP is ON04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONUSER ID ABCDEFGHIJ // If GLP PRINT> User ID is ONPROJ ID 1234567890 // If GLP PRINT > Proj ID is ONName....................... // If GLP PRINT > Name is ON

// Weight is printed if selected in the definitiion string.0.0200kg // Result ON

0.0200kg

Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed

// If Print -x- > Content GLP is ON04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONUSER ID ABCDEFGHIJ // If GLP PRINT > User ID is ONPROJ ID 1234567890 // If GLP PRINT > Proj ID is ONName....................... // If GLP PRINT > Name is ON0.0200kg // Weight is printed if selected in the definition string.

// Reference On// Result On

TEST WT UNIT: Lb/bu (qt)XXXX lb/bu (qt)

Test Weight Mode

OHAUS // Header Line 1Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed

// If Print -x- > Content GLP is ON04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONUSER ID ABCDEFGHIJ // If GLP PRINT> User ID is ONPROJ ID 1234567890 // If GLP PRINT > Proj ID is ONName....................... // If GLP PRINT > Name is ON

// Weight is printed if selected in the definitiion string.0.0200kg // Reference On

// Result OnDOCKAGE REF WT: 0.8000 kg100.0 % ?

Dockage Mode

7 page 19 to 30.pmd 7/27/2006, 3:11 PM25

Page 28: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-26

OHAUS // Header Line 1Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed-----------SPAN CAL-----------04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONCal: 4000.00gOld Cal: 3999.96g // Add zero??Diff: 0.04gWt. ID_______________USER ID:ABCDEFGHIJ // iF GLP PRINT > User ID is ONPROJ ID 1234567890 // If GLP PRINT > User ID is ONName....................... // If GLP PRINT > Name is ON-----------END---------------

Span Calbration Printout

OHAUS // Header Line 1Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed-----------LIN CAL-----------04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONCal: 4000.00gOld Cal: 3999.96g // Add Mid Point and Zero??Diff: 0.04gWt. ID_______________USER ID:ABCDEFGHIJ // iF GLP PRINT > User ID is ONPROJ ID 1234567890 // If GLP PRINT > User ID is ONName....................... // If GLP PRINT > Name is ON-----------END---------------

OHAUS // Header Line 1Line 2 // Header Line 2Line 3 // Up to 5 Header Lines are possibleLine 4 // If none of the 5 header lines are enteredLine 5 // (using the H x command) no header is printed-----------CAL TEST-----------04/01/06 12:30:00 PM // If GLP PRINT > Time is ONBal ID 1234567 // If GLP PRINT > Balance ID is ONCal: 4000.00gOld Cal: 3999.96gDiff: 0.04gWt. ID_______________USER ID:ABCDEFGHIJ // iF GLP PRINT > User ID is ONPROJ ID 1234567890 // If GLP PRINT > User ID is ONName....................... // If GLP PRINT > Name is ON-----------END---------------

Lin Calbration Printout

Cal Test Calbration Printout

7 page 19 to 30.pmd 7/27/2006, 3:11 PM26

Page 29: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-27Adventurer Pro Grain Test Scale

Symptom Possible Cause Remedy

Scale will not turn on • AC power not connected • Connect AC adapter

Scale does not display • Improper calibration • Performaccurately • Unstable environment calibration

• Move scale to asuitable location

Cannot calibrate the scale • Calibration menu locked • Turn Calibration menulock off

• LFT set to on • Turn LFT off• Unstable weight reading • Eliminate vibration and

drafts.Cannot change Menu settings • Submenu locked • Unlock submenu

• LFT set to on • Turn LFT offError 7.0 • Unstable weight reading • Eliminate vibration and

when defining the drafts.reference weight.

Error 8.1 • Weight reading exceeds • Clear pan. Power On Zero limit.

Error 8.2 • Weight reading below • Put pan on scale.Power On Zero limit.

Error 8.3 • Weight reading exceeds • Clear pan.overload limit.

Error 8.4 • Weight reading below • Put pan onunderload limit. scale.

Error 9.0 • Internal fault. • Return scale for service.

4. MAINTENANCE4.1 CalibrationPeriodically verify calibration by placing an accurate weight on the scale and if calibration isrequired, refer to section 3.8.1.

4.2 CleaningIf cleaning is required, use a soft cloth dampened with water and a mild detergent. Do not allowliquids to enter the scale. Do not use harsh chemicals to clean the scale, as the finish may bedamaged.

4.3 TroubleshootingThe following table lists common problems, possible causes and remedies. If the problem persists,contact Ohaus or your authorized Ohaus dealer.

7 page 19 to 30.pmd 7/27/2006, 3:11 PM27

Page 30: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-28

Symptom Possible Cause Remedy

Error 9.5 • Production calibration • Return scale for service.not present.

Error 9.8 • User calibration data not • Calibrate scale.present. (Required forLFT ON only)

Error 53 • EEPROM checksum error. • Cycle power on, off. I fscale fails to operate,return for service.

LOW REF WT • Average piece weight too • See section 3.6.3.small. (Warning)

REF WT Err • Reference weight too small • Increase sample size.The weight on the pan istoo small to define a valid reference weight.

------ • Busy (tare, zero, printing) • Wait until completion.

4.3 Troubleshooting (Cont.)

4.4 Service InformationIf the Troubleshooting section does not resolve or describe your problem, contact your authorizedOhaus Service Agent. For Service assistance or technical support in the United States, call toll-free1-800-526-0659 between 8:00 AM and 5:00 PM EST. An Ohaus product service specialist willbe available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com tolocate the office nearest you.

4.5 PartsDescription Ohaus Part Number

AC adaptersOutput: 12 VAC 500 mA

• US 120V/60Hz 12102320

7 page 19 to 30.pmd 7/27/2006, 3:11 PM28

Page 31: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-29Adventurer Pro Grain Test Scale

4.6 Accessories

Security device 76288-01

RS232 Cables• Cable, DB9M-DB9F 80500525• Cable, DB9M-DB25F 80500524• Cable, DB9M-SF42 Printer 80500571• Cable, DB9M-Apple 80500562

SF42 Printer SF42

4.5 Parts (Cont.)Description Ohaus Part NumberReplacement Pan

• Rectangular pan (7.6 x 8 in. / 19.3 x 20.3 cm) 12103880

7 page 19 to 30.pmd 7/27/2006, 3:11 PM29

Page 32: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-30

7 page 19 to 30.pmd 7/27/2006, 3:11 PM30

Page 33: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-31Adventurer Pro Grain Test Scale

5. TECHNICAL DATAAmbient conditionsThe technical data are valid under the following ambient conditions:

• Ambient temperature: 10 °C to 30 °C• Relative humidity: 15 % to 80 % at 31 °C non-condensing, decreasing linearly to 50

% at 40 °C• Height above sea level: Up to 4000 m

Operability is assured at ambient temperatures between 5 and 40 °C

Power• AC adapter - To national standard as per list in Section 4.6. Scale power input

6-14.5 VAC, 50/60Hz 4VA or 7-20VDC, 4W

Materials• Housing base: plastic (ABS/PC)• Top Housing: plastic (ABS/PC)• Platform: 18/10 stainless steel

Protection• Protected against dust and water• Pollution degree: 2• Installation category: Class II• EMC: see declaration of conformity

Overall Dimensions

C

B

A

5.1 Outline Drawing

A B COVERALL OVERALL OVERALLHEIGHT WIDTH DEPTH 8.5 cm 22 cm 30 cm 3.3 in. 8.7 in. 11.8 in.

8 page 31 to 35.pmd 7/27/2006, 3:11 PM31

Page 34: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-32

Item Number AVD4102N Capacity (g) 4100 Readability (g) 0.01 Repeatability (Std.dev.)(g) 0.01 Linearity (g ± 0.02 Weighing Units Gram, Kilogram, Ounce, Pound, Grain, Custom Application Modes Weighing, Test Weight, Dockage Tare Range To Capacity by Subtraction Stabilization Time (s) 3 Power Requirements AC Adapter (Included) Calibration Digital with External Weight Display Type 2-Line LCD w/Backlight Display Size (in/cm) 4 x 1 / 10 x 2.5 Platform Size (in/cm) 6.6 x 7.1 / 16.8 x 18 Dimensions 8.7 x 3.3 x 11.8 / W x H x D (in/mm) 22 x 8.5 x 30 Net Weight (lb/kg) 6.1 / 2.8

5.2 Specifications5. TECHNICAL DATA

5.3 Capacity x Readability Table

Unit LPT OFF COMMERCIAL GIPSAgrain 63272.8 x 0.2 63272 x 2 63272 x 2gram 4100.00 x 0.01 4100.0 x 0.1 4100.0{0} x 0.0{1}kilogram 4.10000 x 0.00001 4.1000 x 0.0001 4.1000{0} x 0.0000{1}ounce 144.6230 x 0.0005 144.625 x 0.005 144.625 x 0.005pound 9.03895 x 0.00005 9.0390 x 0.0005 9.0390 x 0.0005

8 page 31 to 35.pmd 7/27/2006, 3:11 PM32

Page 35: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-33Adventurer Pro Grain Test Scale

5.4 CommunicationThe scale is equipped with an RS232 interface (COM1). Some models are also equipped witha second RS232 interface (COM2). Connecting the scale to a computer enables you to operatethe scale from the computer, as well as receive data such as displayed weight.

5.4.1 CommandsCommands listed in the following table will be acknowledged by the scale. The scale willreturn “ES” for invalid commands.

Command FunctionIP Immediate Print of displayed weight (stable or unstable).P Print displayed weight (stable or unstable).CP Continuous Print. 0P ends Continous Print.SP Print displayed stable weight.SLP Auto Print stable non-zero displayed weight.SLZP Auto Print stable non-zero weight and stable zero reading.xP Interval Print x = Print Interval (1-3600 seconds) 0P ends Interval Print.H Enter Print Header LinesZ Same as pressing Zero KeyT Same as pressing Tare Key.PM Print current mode (weighing mode).M Scroll to the next enabled mode.PU Print current weighing unit.U Scroll to the next enabled unit.OFF Turns scale OFF.ON Turns scale on.PSN Print Serial Number.PV Print Version: name, software revision and LFT ON (if LFT is set ON).PTIME Print current time.PDATE Print current date.

8 page 31 to 35.pmd 7/27/2006, 3:11 PM33

Page 36: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Adventurer Pro Grain Test ScaleEN-34

5.4.2 ConnectionsRS232 InterfaceOn the rear of the scale, the 9-pin femalesubminiature “D” connector COM 1, is providedfor interfacing to other devices. The pinconnections are shown in the illustration below.

COM 1 Connector

COM1 Pin Connections1 – Remote Tare2 – TxD3 – RxD4 – DSR5 – Ground

6 – DTR7 – CTS8 – RTS9 – Remote Print

COM 2 MINI DIN Connector

TARE*

PRINT*

* External PRINT and/or TARE switches may beinstalled as shown in the diagram. Momentarycontact switches must be used. To enable thisfeature, contact Ohaus Aftermarket.

A 6-pin Mini DIN connector is provided whenthe optional second RS232 interface isinstalled.

12345

6789

COM2 Pin Connections1 – TxD2 – RxD3 – Ground4 – Vout5 – reserved for future use6 – No connection

COM 2 USB Type B Connector

USB InterfaceA USB Type B connector is provided when theoptional USB interface is installed.

8 page 31 to 35.pmd 7/27/2006, 3:11 PM34

Page 37: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

EN-35Adventurer Pro Grain Test Scale

LIMITED WARRANTYOhaus products are warranted against defects in materials and workmanship from the date ofdelivery through the duration of the warranty period. During the warranty period Ohaus willrepair, or, at its option, replace any component(s) that proves to be defective at no charge,provided that the product is returned, freight prepaid, to Ohaus.

This warranty does not apply if the product has been damaged by accident or misuse, exposedto radioactive or corrosive materials, has foreign material penetrating to the inside of theproduct, or as a result of service or modification by other than Ohaus. In lieu of a properlyreturned warranty registration card, the warranty period shall begin on the date of shipment tothe authorized dealer. No other express or implied warranty is given by Ohaus Corporation.Ohaus Corporation shall not be liable for any consequential damages.

As warranty legislation differs from state to state and country to country, please contact Ohausor your local Ohaus dealer for further details.

8 page 31 to 35.pmd 7/27/2006, 3:11 PM35

Page 38: Adventurer Pro Grain Test Scale Instruction Manual€¦ · Adventurer Pro Grain Test Scale Instruction Manual ... Ohaus Corporation, ... EN-10 Adventurer Pro Grain Test Scale

Ohaus Corporation19A Chapin RoadP.O. Box 2033Pine Brook, NJ 07058, USATel: (973) 377-9000Fax: (973) 593-0359

*80103489*

With offices worldwidewww.ohaus.com

© Ohaus Corporation 2006, all rights reserved

Printed in China

P/N 80103489

9 Back cover.pmd 7/27/2006, 3:11 PM1