Aesculap ® SterilContainer System Instructions for Use Steam, Ethylene Oxide and Vapor Hydrogen Peroxide Sterilization Modalities Instructions for Use Intended for US Only As of July 2019
Transcript
Page 1: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System

Instructions for Use

Steam, Ethylene Oxide and Vapor Hydrogen Peroxide Sterilization Modalities

Instructions for Use Intended for US Only

As of July 2019

Page 2: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 2 of 71

Table of Contents 1.0 PURPOSE OF INSTRUCTIONS FOR USE .......................................................................................................................... 4

2.0 STERILCONTAINER SYSTEM ............................................................................................................................................ 4

3.0 STERILCONTAINER SYSTEM REPAIRS ............................................................................................................................ 7

4.0 DECONTAMINATION AND CLEANING PROCESS ........................................................................................................... 8

5.0 INSPECTION PRIOR TO USE ........................................................................................................................................... 12

6.0 PREPARATION AND ASSEMBLY OF STERILCONTAINER SYSTEM ............................................................................. 15

7.0 STERILCONTAINER SYSTEM STERILIZER CYCLE PARAMETERS — STEAM AND ETO ............................................. 21

8.0 STERILCONTAINER SYSTEM STERILIZER CYCLE PARAMETERS — ASP STERRAD® ................................................. 29

9.0 STERILCONTAINER SYSTEM STERILIZER CYCLE PARAMETERS — STERIS® V-PRO® .............................................. 38

10.0 ASEPTIC PRESENTATION ............................................................................................................................................... 46

11.0 STERILE CONTAINER VALIDATION SUMMARY ........................................................................................................... 49

12.0 CUSTOMER VERIFICATION ............................................................................................................................................ 51

13.0 INDICATIONS FOR USE .................................................................................................................................................. 56

Page 3: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 3 of 71

List of Figures

List of Images

Page 4: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 4 of 71

1.0 Purpose of Instructions for Use

2.0 SterilContainer System

ONLY

Product Family SterilContainer SterilContainer S SterilContainer S2

Bottom Series

JK JN JM

JS

Lid Series3, 5 JK PrimeLine4 PrimeLine Pro4 JK PrimeLine4 PrimeLine Pro4 JM JS

PreVac Steam X X X X X X X X

PreVac IUSS X X X

Gravity X X X

EtO X X X

Low Temp

STERRAD1 X X

Low Temp STERIS2 X X

Figure 1: SterilContainer System

Page 5: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 5 of 71

Primary Name Which Includes May Also Be Referred to As

Figure 2: Sterilizations Modality Nomenclature and SterilContainer System Compatibility

JK Solid Bottom JS Perforated Bottom,

Identified by the Gold Handle & Latch

JN Perforated Bottom JK Solid Bottom JN Perforated Bottom JS Perforated Bottom

Identified by the Gold Handle & Latch

Page 6: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 6 of 71

Figure 3: SterilContainer System Basket Options

Basket Series JB JC JE JF

Figure 4: Container Basket Features

Image 1: Single Use Processing Supplies, Filters, Tamper Evident Locks, Indicator Cards

Page 7: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 7 of 71

Notes:

3.0 SterilContainer System Repairs

Page 8: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 8 of 71

ONLY Aesculap trained technicians are authorized to repair the Aesculap SterilContainer System. Using a

non-Aesculap repair technician to repair containers will void the Aesculap Warranty on the container and may

void any of the validation testing associated with Aesculap containers.

Notes:

4.0 Decontamination and Cleaning Process

DO NOT USE

Notes:

Page 9: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 9 of 71

4.1 Water Quality

4.1.1 Utility Water

4.1.2 Critical Water

4.2 Detergent Solutions

Notes:

DO NOT USE

4.3 Decontamination and Mechanical Cleaning

Page 10: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 10 of 71

Figure 5: PrimeLine Pro Lid Inspection Process

Notes:

DO NOT USE

DO NOT USE

Page 11: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 11 of 71

4.4 Decontamination and Manual Cleaning

Notes:

DO NOT USE

Page 12: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 12 of 71

DO NOT USE

DO NOT USE

5.0 Inspection Prior to Use PRIOR TO EVERY USE

DO NOT USE

Notes:

5.1 SterilContainer System Inspection Criteria

Page 13: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 13 of 71

Image 2: JK, JN and JM Lid Reusable Filter Inspection

Image 3: Filter Retention Plates and Silicone Gaskets Inspection Process

Page 14: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 14 of 71

Image 4: PrimeLine Lid Inspection Process

Notes:

Page 15: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 15 of 71

5.2 Basket, Tray and Accessory Inspection Criteria

6.0 Preparation and Assembly of SterilContainer System PRIOR TO EVERY USE

ONLY

Filter Type PreVac Steam PreVac IUSS Gravity EtO Low Temp

Paper Filter

w/ Indicator1

Paper Filter

w/o Indicator

Polypropylene Filter

w/o Indicator

Metal Retention Plate Lid

PTFE Reusable Filter

PrimeLine & PrimeLine Pro

PTFE Reusable Filter2

Figure 6: Filters for Perforated Bottoms and Lids

Page 16: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 16 of 71

Notes:

6.1 SterilContainer System Assembly ONLY USE

Image 5: Aluminum Lid with Metal Retention Plate and Reusable Filter

Image 6: Aluminum Lid with Metal Retention Plate and Single Use Filer Assembly

Page 17: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 17 of 71

Image 7: PrimeLine Lid Inspection Process

Notes:

6.2 Assembly of Surgical Instrumentation

6.3 Loading of Basket, Lifting Platform and Tray

Page 18: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 18 of 71

Notes:

Full-Size, Three-Quarter Size, Half-Size Wide-Body, Extra-Long Container

Extra-Long Mini-Size Container

Mini-Size and Quarter-Size Container

6.4 Internal Process Indicators

Notes:

6.5 External Process Indicators and Tamper Evident Seals

Page 19: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 19 of 71

Tamper Evident Locks PreVac Steam PreVac IUSS Gravity EtO Low Temp

Blue / US900

Green / US905

Orange / US906

Pink / US9102

Yellow / US399

The recommended post sterilization indicator color may not be evenly shaded but should be shades of the

proper processed color to indicate exposure to sterilant

Figure 7: Tamper Evident Locks

Installation Instructions for All Tamper Evident Locks

Page 20: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 20 of 71

Indicator &

Communication Cards PreVac Steam PreVac IUSS Gravity EtO Low Temp

MD334, MD335

w/ Indicator

MD346, MD876, US754

w/ Indicator

US963

w/o Indicator

MD399, MD345

w/ Indicator

The recommended post sterilization indicator color may not be evenly shaded but should be shades of the proper processed

color to indicate exposure to sterilant.

Figure 8: Indicator and Communication Cards

Notes:

amper Evident Locks US900, US905, US906 and US399

DO NOT USE

Tamper Evident Lock US910

DO NOT USE

Indicator & Communication Cards MD334, MD335

DO NOT USE

Page 21: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 21 of 71

6.6 Container Storage and Transportation

DO NOT

7.0 SterilContainer System Sterilizer Cycle Parameters — Steam and EtO

Steam and EtO modalities

ONLY

7.1 Steam and EtO Sterilization Modality Cycle Parameters

Primary Name Which Includes May Also Be Referred to As

Page 22: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 22 of 71

7.1.1 PreVac Steam Sterilization Cycle Parameters (1)

°

7.1.2 PreVac IUSS Sterilization Cycle Parameters (2)

Porous Instruments

°

Non-Porous Instruments

°

ONLY

7.1.3 Gravity Steam Sterilization Cycle Parameters (3)

7.1.4 EtO Cycle Parameters (4)

7.1.5 PreVac Steam Sterilization Cycle Parameters (5)

Page 23: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 23 of 71

Notes:

Detailed Size Description Abbreviation Size Description

Page 24: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 24 of 71

7.1.6 Steam and EtO – SterilContainer JK Series

JK Series

Bottom Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

Wide

XLL

1:1

3:4

1:2

Wide

XLL

Filter

Indicator

Card

Lock

Mini Mini

Filter

Indicator

Card

Lock

Page 25: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 25 of 71

7.1.7 Steam and EtO – SterilContainer JK Series with PrimeLine Pro / PrimeLine Lid

1:1

3:4

1:2

1:1

3:4

1:2

Filter, LidB

Filter, Bottom

Indicator

Card

Lock

JK Series

Bottom Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

1:1

3:4

1:2

Filter, LidB

Filter, Bottom

Indicator

Card

Lock

Page 26: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 26 of 71

7.1.8 Steam and EtO – SterilContainer JN Series

JN Series

Bottom

Aluminum

Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

XLL

Wide

1:1

3:4

1:2

XLL

Wide

Filter

Indicator

Card

Lock

Qtr

Mini

XLM

Qtr

Mini

XLM

Filter

Indicator

Card

Lock

Page 27: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 27 of 71

7.1.9 Steam and EtO – SterilContainer JN Series, PrimeLine Pro / PrimeLine Lid

JN Series

Bottom Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

1:1

3:4

1:2

Filter, LidB

Filter, Bottom

Indicator

Card

Lock

1:1

3:4

1:2

1:1

3:4

1:2

Filter, LidB

Filter, Bottom

Indicator

Card

Lock

Page 28: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 28 of 71

7.1.10 Steam and EtO – SterilContainer JM Series

JM Series

Bottom

Aluminum

Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

1:1

3:4

1:2

Filter

Indicator

Card

Lock

Mini

XLM

Mini

XLM

Filter

Indicator

Card

Lock

7.1.11 Steam and EtO – SterilContainer JS Series

JS Series

Bottom

Aluminum

Lid

Processing

Supplies

Validated and FDA 510(k) Cleared Sterilization Modalities

Max Total

WeightA

1:1

3:4

1:2

1:1

3:4

1:2

Filter

Indicator

Card

Lock

Page 29: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 29 of 71

8.0 SterilContainer System Sterilizer Cycle Parameters — ASP STERRAD®

Low Temperature modality when used with ASP STERRAD

sterilizers

ONLY

Notes:

NOT

NOT

NOT

Detailed Size Description Abbreviation Size Description

Page 30: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 30 of 71

8.1 STERRAD Sterilization Modality Cycle Parameters

Primary Name Which Includes May Also Be Referred to As

STERRAD

100S 200

NX1

Standard

NX1

Advanced

100NX1

Standard

100NX1

Flex

100NX1

Express

JM Series

JS Series

Page 31: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 31 of 71

8.1.1 STERRAD 100S Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERRAD 100S

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

Notes:

Page 32: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 32 of 71

8.1.2 STERRAD 100S Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERRAD 100S

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

8.1.3 STERRAD 200 Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERRAD 200

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

Page 33: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 33 of 71

8.1.4 STERRAD NX Standard Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERRAD NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

8.1.5 STERRAD NX Standard Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERRAD NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 34: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 34 of 71

8.1.6 STERRAD NX Advanced Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERRAD NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

8.1.7 STERRAD NX Advanced Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERRAD NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 35: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 35 of 71

8.1.8 STERRAD 100NX Standard Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERRAD 100NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

8.1.9 STERRAD 100NX Standard Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERRAD 100NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 36: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 36 of 71

8.1.10 STERRAD 100NX FLEX Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies STERRAD 100NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

8.1.11 STERRAD 100NX FLEX Cycle — SterilContainer S2 JS Series

Size Bottom Processing Supplies STERRAD 100NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Notes:

o o o

Page 37: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 37 of 71

8.1.12 STERRAD 100NX Express Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies STERRAD 100NX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Notes:

Page 38: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 38 of 71

9.0 SterilContainer System Sterilizer Cycle Parameters — STERIS® V-PRO®

Low Temperature modality when used with STERIS V-PRO

sterilizers

ONLY

Notes:

NOT

NOT

NOT

Detailed Size Description Abbreviation Size Description

Page 39: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 39 of 71

9.1 STERIS Sterilization Modality Cycle Parameters

Primary Name Which Includes May Also Be Referred to As

STERIS

V-PRO 60 maX1

Lumen Non-Lumen Flexible Flexible Lumen Non-Lumen

JM Series

JS Series

Page 40: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 40 of 71

9.1.1 STERIS V-PRO 60 Lumen Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

EfficacyB

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

Page 41: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 41 of 71

9.1.2 STERIS V-PRO 60 Lumen Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

EfficacyB

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 42: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 42 of 71

9.1.3 STERIS V-PRO 60 Non-Lumen Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Filter

Indicator

Card

Lock

9.1.4 STERIS V-PRO 60 Non-Lumen Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 43: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 43 of 71

9.1.5 STERIS V-PRO 60 Flexible Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

9.1.6 STERIS V-PRO 60 Flexible Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO 60

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Notes:

o o

Page 44: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 44 of 71

9.1.7 Amsco STERIS V-PRO maX Flexible Cycle — SterilContainer S JM Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO maX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

9.1.8 Amsco STERIS V-PRO maX Flexible Cycle — SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO maX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 45: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 45 of 71

9.1.9 Amsco STERIS V-PRO maX Lumen Cycle— SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO Max

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

9.1.10 Amsco STERIS V-PRO maX Non-Lumen Cycle— SterilContainer S2 JS Series

Size Bottom Lid Processing Supplies

STERIS

V-PRO maX

Efficacy

Max Total

WeightA

Filter

Indicator

Card

Lock

Page 46: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 46 of 71

10.0 Aseptic Presentation

Notes:

Page 47: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 47 of 71

For Reference Only

Filter Type PreVac Steam PreVac IUSS Gravity EtO Low Temp

Paper Filter

w/ Indicator1

Paper Filter

w/o Indicator

Polypropylene Filter

w/o Indicator

Metal Retention Plate

PTFE Reusable Filter

PrimeLine & PrimeLine Pro

PTFE Reusable Filter

Tamper Evident Locks PreVac Steam PreVac IUSS Gravity EtO Low Temp

Blue / US900

Green / US905

Orange / US906

Pink / US9102

Yellow / US399

The recommended post sterilization indicator color may not be evenly shaded but should be shades of the

proper processed color to indicate exposure to sterilant

Notes:

Page 48: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 48 of 71

For Reference Only

Indicator &

Communication Cards PreVac Steam PreVac IUSS Gravity EtO Low Temp

MD334, MD335

w/ Indicator

MD346, MD876, US754

w/ Indicator

US963

w/o Indicator

MD399, MD345

w/ Indicator

The recommended post sterilization indicator color may not be evenly shaded but should be shades of the

proper processed color to indicate exposure to sterilant.

10.1 SterilContainer System Reference Guidelines

Figure 9: Easy Reference Handout

10.2 SterilContainer System Transportation to Decontamination

Page 49: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 49 of 71

11.0 Sterile Container Validation Summary

11.1 Validation Testing

11.1.1 Reprocessing Validation and Verification

11.1.2 Sterilization Efficacy

Page 50: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 50 of 71

11.1.3 Sterility Maintenance

Event Related Storage Study

Aerosol Challenge Test

Notes:

11.2 Aesculap SterilContainer System and FDA Clearances

Page 51: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 51 of 71

Container System DescriptionFDA 510(k)

Clearance

References:

12.0 Customer Verification validate

verify

Page 52: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 52 of 71

Placement of Chemical and Biological Indicators for Product Testing

Page 53: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 53 of 71

AAMI ST79 Guideline Aesculap Product Placement

Example of External Chemical Indicators

Page 54: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 54 of 71

Example placement of Chemical Indicator for Routine Load Release

3. Periodic Product Quality Assurance Testing

Page 55: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 55 of 71

AAMI ST79 Reference

Biological indicators (BIs):

o

Chemical indicators (CIs):

o

Master product:

o o

Product family:

o

Notes:

Page 56: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 56 of 71

13.0 Indications for Use

13.1 SterilContainer and SterilContainer S — Steam and EtO Sterilization

13.2 SterilContainer — PreVac IUSS Sterilization

13.3 SterilContainer — JK / JN744 PreVac Steam, IUSS and EtO Sterilization

Page 57: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 57 of 71

Validated Sterilization Cycle Parameters

Sterilization Cycle Parameters Max No. of Lumens/Lumen Configuration*

≥≤≥ ≤

≥≤≥ ≤

≥≤≥ ≤

Accessories

PreVac

Steam

PreVac

IUSS

EtO

13.4 SterilContainer with PrimeLine Lid

13.5 SterilContainer with PrimeLine Pro Lid

Page 58: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 58 of 71

Steam and IUSS Compatible SterilContainer with PrimeLine Pro Lid A combined maximum load validated for all container configurations is 25lbs.

Sterilization Cycle Parameters PrimeLine Pro

Container Lid with

JP050 – Lid Size

Solid Base to be used

with Lid

Max No. of Lumens

Lumen Configuration

Immediate Use – Non-porous

Immediate Use – porous

PreVacuum Dry Time Study

Immediate Use – Non-

Porous

Immediate Use – Porous

Prevacuum Dry Time Study

Table 2: Steam and IUSS Cycle Compatible Accessories Accessories Steam and IUSS

Stainless Steel baskets, basket lids, and dividers

Yes

Instrument Organization System (Silicone and Stainless Steel racks, brackets, holders, and clamps)

Yes

Silicone mats Yes

Stainless Steel racks, trays, holders, clamps, brackets, and platforms Yes

Page 59: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 59 of 71

13.6 SterilContainer with Aluminum Lid and Metal Retention Plate — Reusable Filter

13.7 SterilContainer — JS Series

Table 1. SterilContainer S2 Validated Load Configurations

Sterilization Cycle Container Size Validated Load Configuration

≥ ≤

Page 60: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 60 of 71

≥ ≤≥ ≤

≥ ≤

≥ ≤≥ ≤

≥ ≤

Table 2. SterilContainer S2 System Configurations – Prevac Steam and Gravity Steam

Sterilization Method Container Size

Container

Bottom Part #

Container

Lid Part #

Total Loaded

Container Weight*

Page 61: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 61 of 71

Table 3. Sterilization Cycle Compatible Accessories - Prevac Steam and Gravity Steam

Accessories

Compatible with

Prevac Steam Gravity Steam

Table 4. SterilContainer S2 System Configurations – STERRAD Sterilization Systems

Sterilization

Method Container Size

Container

Bottom

Part #

Container

Lid Part # Total Loaded Container Weight

Page 62: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 62 of 71

Table 5. Sterilization Cycle Compatible Accessories – STERRAD Sterilization Systems

Accessories

Compatible with STERRAD

100S

NX

Standard

NX

Advanced

100NX

Standard

100NX

Flex

Page 63: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 63 of 71

Table 6. SterilContainer S2 System Configurations – STERIS Sterilization Systems

Sterilization

Method Container Size

Container

Bottom

Part #

Container

Lid Part # Total Loaded Container Weight

Page 64: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 64 of 71

Page 65: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 65 of 71

Table 7. Sterilization Cycle Compatible Accessories – STERIS Sterilization Systems

Accessories

Compatible with STERIS V-PRO

60

Lumen

60

Non-Lumen

60

Flex

maX

Lumen

maX

Non-Lumen

maX

Flex

13.8 SterilContainer S — STERRAD® 100S

Page 66: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 66 of 71

13.9 SterilContainer S — STERRAD® 200 System, NX™ System, and 100NX System

≥ ≤ ≥ ≤ ≥ ≤ ≥ ≤ ≥ ≤

13.10 SterilContainer S — STERRAD 100NX EXPRESS Cycle

Page 67: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 67 of 71

Container Configuration Intended Load

Accessories STERRAD 100NX Express Cycle

13.11 STERIS V-PRO 60 — SterilContainer S with Aluminum Lid

Lumen Cycle

Validated Container Load

o ≥ ≤

o ≥ ≤ ≥ ≤ ≥ ≤

Page 68: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 68 of 71

Non-Lumen Cycle

Validated Container Load

Flexible Cycle

Validated Container Load

≥ ≤

Accessories V-PRO 60

13.12 STERIS V-PRO® maX Flexible Cycle — SterilContainer S with Aluminum Lid

Validated V-PRO maX Sterilizer Flexible Cycle Load Configurations

Load Configuration 1

Page 69: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 69 of 71

Load Configuration 2

V-PRO maX Sterilizer Flexible Cycle Compatible SterilContainer S Container Systems

Accessories V-PRO maX Flexible Cycle

13.13 STERIS V-PRO® 1 and V-PRO® 1 Plus— SterilContainer S with Aluminum Lid

Processing STERIS V-PRO Low Temperature Sterilization Systems Lumen and Non-Lumen Cycles

Suggested Sterilizer Cycle Parameters for SterilContainer S products:

Page 70: Aesculap SterilContainer System Instructions for Useweb.orthofix.com/Products/Products/IFU/AesculapSterilContainerSystem.pdf · 8.1 STERRAD Sterilization Modality Cycle Parameters

Aesculap® SterilContainer™ System Instructions for Use (IFU)

SOP-AIC-5001592 Rev. 04 Intended for US Only

As of July 2019 Page 70 of 71

o o o

o o

Notes:


Recommended