Agilent Solutions for achieving 21 CFR 11 Compliance on Atomic Spectroscopy Instrumentation Page 1 Jon Talbott, PhD Application Scientist
Transcript

Agilent Solutions for achieving 21 CFR 11 Compliance on Atomic Spectroscopy Instrumentation

Page 1

Jon Talbott, PhDApplication Scientist

Spectroscopy Products currently supported with Compliance Solutions1. ICP-MSs (7800, 7900,

8900 and older units)

2. ICP-OES (5110)

3. GFAAs and Flame AAs(240Z, 480, 240FS)

Spectroscopy Products currently supported with Compliance Solutions - continued

4. Cary Win UV-Vis Spectrometers(8453 and Cary 60)

5. Dissolution InstrumentsUV DissolutionUV Fiber Optic

(Cary 60 w/ Multicell shown)

5. MicroLab

Agilent ICPMSs: 3 Features for « Pharma » Applications1) Matrix tolerance 2) Interferences removal 3) Wide Dynamic range

Ultra High Matrix Introduction System (UHMI)

Able to analyze high total dissolved solids (up to 25%)• Reduces sample prep time

and error• Better long-term stability

Octopole Reaction System (ORS3) – Helium Mode

He mode removes all polyatomic interferences• Easy to use - He collision

mode only • Effective for a wide range of

sample types

Wider Dynamic Range

11 orders of dynamic range• Simplifies method development• Easy sample prep. • Improves productivity

ICP-OES Agilent 5110: Benefits for USP <232> <233>

1) Matrix tolerance 2) Synchronous Axial and Radial Viewing

3) Interference removal

Vertical Torch

Able to analyze high total dissolved solids ( > 25%)

Background Correction and Interference Removal

FBC Fitted Background Correction FACT Spectral Deconvolution

Dichroic Spectral Combiner

Compliance for Elemental Impurities, Why Needed? USP Heavy Metal Chapters – Current and Planned

<2232>

Heavy Metals Limit Test

<233><232>

ElementalImpurities(Limits)

ElementalImpurities

(Procedures)

ElementalImpurities in

Dietary Supplements

Terminology changing:“Heavy Metals” “Elemental Impurities”

Is being replaced with:

Page 7September 26, 2017

Related method <2232> applies to dietary supplements only

<232> will eventually replace other “metals” General Chapters

<251>

<231>

Lead

<211>Arsenic

<261>Mercury

September 26, 2017

Confidentiality Label

9

ICH Class Element Symbol ICH (ug/day) Oral PDE

Class 1 Arsenic As 15Cadmium Cd 5Mercury Hg 30

Lead Pb 5Class 2A Cobalt Co 50

Nickle Ni 200Vanadium V 100

Class 2B Silver Ag 150Gold Au 100

Iridium Ir 100Osmium Os 100

Palladium Pd 100Platinum Pt 100Rhodium Rh 100

Ruthenium Ru 100Thalium Tl 8

Selenium Se 150Class 3 Barium Ba 1400

Chromium Cr 11000Copper Cu 3000Lithium Li 550

Antimony Sb 1200Tin Sn 6000

Molybdenum Mo 3000

• PDE’s apply to final products• Drug Substances• Excipients

• Does not apply to vaccines• Does not apply to veterinary

products• Does not apply to dietary

supplements• Testing is based on risk-based

assessment• Compliance is required with all

the limits throughout the life of the product

USP/ICH Oral PDE Limits

PDE Limits depend upon Route of Entry

What is FDAs 21 CFR Part 11 Compliance Chptabout? It is just 4 pages of the Code of the Federal Register!

Regulations for Pharma, Food, CROs, anyone sending Sx data to the FDA

What is FDA 21 CFR Part 11 Compliance Chptabout? Legalese to address …

Electronic Records (Sample DATA in electronic format)• Their management, storage,

authenticity, integrity, confidentiality and retrieval

ValidationofSystems (Instrument IQ/OQ/PQ)Limiting Access to systems to authorized individuals (User Access Control)

Protection of Records (Data)Secure Time Stamped Audit TrailsRecords shall not obscure preciously recorded information (version control)

Authorized IndividualsElectronically signed DataSOPs, Training

FDA 21 CFR Part 11 Compliance Chpt: Summary of Four Key Categories of Criteria1. Instrument Qualification Procedures and Documentation

– Instrument design qualification (DQ), lifecycle management, Installation and Operational Qualification (IQ/OQ) for analytical instruments + Software and Performance Qualification (PQ)

2. Access control of instrument and data processing limited to authorized individuals, eSignature with audit trail of user log-on/log-off and actions– User Access Control (UAC) on Agilent ICP-MSs – Spectroscopy Configuration Manager (SCM) on ICP-OESs

3. Electronic records management, secure data storage, version control, archival and retrieval of data records– SDA (Spectroscopy Database Administrator) – Data Store– OpenLab ECM (Enterprise Content Manager)

4. Physical access control (e.g. lab access), standard operating procedures– Making the lab “compliant” is the responsibility of the user, using the

tools and software the vendor provides

Application software+ 4 Qs

Graphical Representation of Key Categories of Agilent Solutions for 21 CFR Part 11 Compliance

Data Management

Control the instrument for data acquisition and processing

ICPExpert or MassHunter

Administration of databases and data, records integrity, security and retention

SDA, OpenLab ECM

+ +

Access Control

Manage groups, projects, users and user profiles, e-sigs and Audit Trail

SCM or UAC

2nd Category 3rd Category1st Category

Agilent Solution for First Category of 21 CFR 11- 4 Qs: Design Qualification, Installation and Operational Qualification (IQ/OQ)

PQ is customer’s responsibility• Agilent’s compliance products and services includes Installation

Qualification and Operational Qualification for Hardware and Software,• Documentation of IQ/OQ

IQ/OQ (Installed correctly/ Work as expected)Hardware & Software

IQ checks if the system is installed correctly prior to initial utilization:• Operating environment• Physical connection• Safety• Functional parameters

OQ defines functional tests and compares them against Agilent specifications.

IQ/OQ performed jointly with customer representative

Example OQ Tests for ICP-MS

Example OQ Tests for ICP-OES

Agilent Solution for First Category of 21 CFR 11- Design Qualification, Installation and Operational Qualification (IQ/OQ)

• Agilent’s compliance products and services includes Installation Qualification and Operational Qualification for Hardware and Software,

• Documentation of IQ/OQ

Application software

21 CFR Part 11 – 2nd Category of CriteriaUser Access Control & Electronic Signatures

Data Management

Control the instrument for data acquisition and processing

ICPExpert or MassHunter

Administration of databases and data, records integrity, security and retention

SDA, DS or OpenLab

+ +

Access Control

Manage groups, projects, users and user profiles, esigs and Audit Trail

UAC or SCM

User Access Control SW on ICP-MSs for 21CFR11 Compliance - Access Control of Instrument/SW, eSignature and Audit Trail

MassHunter User Access Control (UAC) functionality provides:• Password protection for user access to ICP-MS MassHunter• Flexible, multi-level configurable access to MassHunter functions (defined by user

levels, Admin, ECM Mgr, SDAAdmin, SDAApprover, Mgr and Operator)• Audit trail - Record of user(s) logon/logoff to operating system (supported OS

function) and MassHunter logon; audit trail also records user actions as defined in ATM (Audit Trail Map)

ICP-MS MassHunter User Access Control for Audit Trail and some of eSignature

ICP-MS MassHunter Audit Trail

Page 22

User Access Control with OpenLab ECM on ICPMSs – A Second level of UAC exists Systems with OpenLab ECM

ECM is the most comprehensive solution for protecting, storing and accessing your dataNot only is there MassHunter User Access Control (UAC) Users and their access roles/privileges are also set up in OpenLab ECM

• Users must log on to ECM AND UAC foraccess to ICP-MS MassHunter

• Different User roles/Privileges/Access rights are also set up in ECM

Note Network Domain

Application software

21 CFR Part 11 - 2nd Category for OESs, UV-VisUser Access Control & Electronic Signatures

Data Management

Control the instrument for data acquisition and processing

ICPExpert or MassHunter

Administration of databases and data, records integrity, security and retention

SDA, DS or OpenLab

+ +

Access Control

Manage groups, projects, users and user profiles, esigs and Audit Trail

UAC or SCM

25

Spectroscopy Configuration Manager What is it?

1. User Access Control Manager for all other Spectroscopy instrumentation (OES, AA, UV, Dissolution)

click

2. Allows set up Groups, Programs, Users, User Profiles (access rights) and even Stations

3. Network capable/scalable

4. Audit trail contained within

Potential SCM Configurations Network configurable and Scalable

There are 5 Scenarios (configurations) for Compliance SCM software. • Configuration depends upon customer wishes at installation.

Scenario 5Scenario 1

ApplicationSDASCM

Scenario 2

Application

SDA

SCM

Scenario 3

SCM

Application

SDA

Application

Scenario 4

SDA

SCM SDAApplication

SCM

In contrast, UAC SW for ICPMS is not configurable as a network (Client/Server) and is not scalable

Spectroscopy Configuration Manager (SCM)User Profile set up and are similar to Audit Trail Map access rights (Actions) in UACProfile capabilities different depending upon the Instrument

• Select the applications which a User can access by adding ‘x’ next to each application.

• Note: For Cary WinUVsoftware the Color, Fabric Protection and Sunglasses Applications are not 21 CFR 11 compliant.

Application software

21 CFR Part 11 – 3rd Category of CriteriaElectronic Records Mgmt, Storage & Security

Data Management

Control the instrument for data acquisition and processing

ICPExpert or MassHunter

Administration of databases and data, records integrity, security, retention and retrieval

SDA, DS or OpenLab

+ +

Access Control

Manage groups, projects, users and user profiles, esigs and Audit Trail

UAC or SCM

Agilent Compliance Products for ICPOES & ICP-MS

Large/global Pharma

MidRegulated(CRO)

Small Regulated

All MS DB options need UAC

OpenLab Data Store

SDA

OpenLab ECM

ICP-MS MassHunter Compliance SolutionsPage 29

OpenLAB ECM 3.5• Supports 7800, 7900 ICP-MS, and 8900

ICP-QQQ• Accepts data files from 5110 with

Scheduler but still not compliant (versioning missing)

OpenLAB Data Store A.01.03• Supports 7800, 7900 ICP-MS, and 8900

ICP-QQQSpectroscopy Database Administrator A.01.02• Supports 7800, 7900 ICP-MS, and 8900

ICP-QQQ• Supports 5110 ICP-OES (different SDA)

Recommended for:

Database provides secure storage of DataSDA and ECM Provide Database Management functions:• Creation/ Attaching / Detaching / Archiving / Retrieval of a database

• An “attached” database is “online” and • a “detached” database is “offline” and cannot be accessed by users

• Version Control on file save/upload• Includes Database Viewer• Audit Trail for databases• Batch approval in SDA• Database backup/upload• Admin functions

• Control of database access rights• SDA log

•SDA Database Management pane

Page 30

Agilent Solution for 3rd Category of 21CFR11 Compliance – Electronic Records Management, eSig, Version, Archive

SDA Database Viewer for more eSignature

ICP-MS MassHunter with SDA includes a Database viewer for data management. Batch data is secured using ssizip checksum protected file

ICP-MS SDA Database Viewer for batch approval, upload, etc.. Version numbers can be seen

Batch Properties, Audit and Approval logs are available

April 11, 2013Page 32

SDA Database Viewer showing Versioning, Secure Data Storage

ICP-MS SDA Database Viewer for batch approval, upload, etc.. 13 Versions of the DA Method Post processed shown here

Page 33

Sitting on this batch

OpenLAB ECM Provides Secure Content Management for all files from all InstrumentsECM maintains all electronic data regardless of source:

• Analytical data from mulitple techniques and suppliers

• Chromatographic result data• Master methods, and

sequence and report templates

• Pictures and graphic images • PDF documents und

formulars• Email• Microsoft Office files• Electronic reports

Search function is really cool

OpenLAB Data Store Compliance FunctionsServer-based solution for ICP–MS Compliance

Efficient Data Storage and Retrieval Centralized Storage for ICP–MS MassHunter (rev B.01.03) Efficient search and retrieval of files Local Language Support (English, Chinese & Japanese)

Security Comprehensive data security Access for registered users with roles Integration to OpenLAB Shared Services/Control Panel

Data Integrity and Traceability Maintain file revision and history Comprehensive Audit Trails

Trail of who has what item Trail of when an item was received

21 CFR Part 11 Compliant Built to support FDA regulations

Data Integrity and Traceability Electronic Signatures Archival Capabilities

ServerOpenLAB Data StoreOpenLAB Shared Services2 MS instrument connections

Agilent Compliance Products for ICP-MS• Functional Comparison

OpenLAB ECM Data Store for MS SDA for ICP - MSGeneral Specs International Language support (From A.01.02)

Multi – Vendor Support

Integrated Search & Retrieve

# of Instruments supported No limit Up to 15 1Server Required (Enhanced Security)

Automated Uploads

Integrated (free) Database

Compliance e-Signature

PDF Signature

Audit Trails

File Revisions

Integration OpenLAB CDS (From A.02.01)

MassHunter (Before A.02.01,

only ICP–MS) (ICP–MS)

OpenLAB ELN

Third-party LIMS and ELN

ECM SDA

Summary of 21 CFR Part 11 Compliance Solutions

1. Software and hardware design validation – Agilent excels in IQ/OQ for both hardware and software – Agilent consistently ranked #1 for compliance services

2. User Access Control – ICP-MS add-on– User Access Control for ICP-MS MassHunter OR SCM for all other Atomic line

and Spectroscopy products provides fully flexible solution, with configurable access levels

3. Electronic records Mgmt – 3 DB options: SDA, Data Store or OpenLabECM (Enterprise Content Manager)

4. ICP-MS MassHunter User Access Control add-on functionality and SDA/Data Store/OpenLAB ECM software provides full compliance with 21 CFR Part 11

– Electronic records and version management done by SDA/OpenLAB ECM– Data security is ensured through use of secure (encrypted, non-editable) files– Databatch information (raw data and reports) is secured and versioned– All back-up and archival functions are performed in SDA/OpenLAB ECM

Page 40


Recommended