ISO 9001 Manufacturer
p r e s s u r e
Specification SheetSS-IPI
Pressure rangeIPI100A 0to7barabsolute/100psiIPI030C Vacuumto2bar/30psiIPI100C Vacuumto7bar/100psiIPI300C Vacuumto21bar/300psiIPI500C Vacuumto35bar/500psiIPI015G 0to1bar/15psiIPI01KG 0to70bar/1,000psiIPI02KG 0to140bar/2,000psiIPI03KG 0to200bar/3,000psiIPI05KG 0to350bar/5,000psiIPI10KG 0to700bar/10,000psi
High accuracy±0.05% of F.S. for positive pressureVacuum ranges to 500 psi (35 bar)
High flexibilityEasily disassemble the test system and use the indicator for other tasks
True field indicatorLightweight and portable with full tempera-ture-compensation, long battery life and large display for easy visibility
Record min and max readingsCapture the min and max pressure readings for safety valve applications
Complete marine programPart of a complete program of marine approved temperature, pressure and signal calibrators; including temperature sensors
ATEX and CSA certificationThe IPI Indicator is ATEX and CSA certified and designed for use in potentially explosive environments
ATEX
JOFRAIPIindicatorsbringtogethertheeaseofananaloggaugewiththeaccuracyandeasy-to-readdisplayofadigitalcalibrator.
UsetheJOFRAIPIinapplicationsfromsafetyvalvecheckstosystempressureverification.
FurthermoretheIPIindicatorisATEXandCSAcertifiedforuseinpoten-tiallyexplosiveenvironmentssuchasoilrefineries,chemicalplantsandoffshoreplatform,wherethereisariskofinflammablegases.
Thisseriesisdesignedtomeetyourpressuremeasurementapplicationsandmakethetaskseasier.TheIPIoffers18differentpressureunits,longbatterylife,highaccuracy,andevenserialcommunications.TheaccuracyoftheIPIrivalsthatofapressurecalibratorandistemperaturecompensatedforshoporinprocessapplications.Thisversatileunitisavailableasastandaloneindicatororinacompletetestsystem.
YoucanperformacalibrationlocallywithoutreturningtheIPIunittothemanufacturer.Allyouneedisanaccuratepressurereference.Ifyoudorequirefactorycalibration,thepumpandindicatorareindependentandonlytheindicatorneedstobereturned.
TheJOFRAIPIisavailableasanindicatororinoneof6testreadysys-temsthatarecompleteandequippedtomeetyourpressuremeasuringortestingneeds.
Industrial Pressure Calibrator
IPI Indicators
CSA
Certified for use
in potentially explosive
environments
2 Industrial Pressure Calibrator • IPI
Clear dual line displayTheIPIdisplaydoesmorethan
justshowapressurereading.Theusercanchecktemperatureand
exactbatteryvoltagethroughtheuseofthekeypad.The
large,51/2digit,0.65in(1.65cm)characterdisplayislargeandeasy-to-read;evenfromadistance.Iconsarealsoincludedtoindicatebatterylife,engineeringunits,andthereisabargraphthatshowsthepercentageofscalereading.Allofthisinformationcanbereadinlowlightwiththeuseofthebacklight.
Pressure connections1/4”NPTmalelowerconnec-tionisstandard.A1/4”BSPmaleadapterisincluded.TheIPImaybeorderedwithabackmountingconfiguration.Thisdesignhasthe1/4in.NPTmaleconnec-tiononthebackofthecasevicethelowermanifoldconnection.Thisallowsforuseinpanelmountedappli-cationsorinapplicationswhereflatmountingisnecessarysuchasinanoverhead.Thisconfigurationisnotdeliveredwithaprotectiveboot.
MAX / MINAMIN/MAXfunctionreadsthemaximumandminimumpressurethathavebeenauto-maticallystored.Monitorasafetyvalveorlookforpressurespikesintheprocess.TheIPIalsohavetheabilitytochecktheminimumandmaximumreadingsduringtheransient.
ZEROKeepthehighaccuracy:ZEROtheinstrumentbeforeeverytest.
ON / OFFTheIPIisdeliveredwiththeauto-shutoffactiveandsetto30minutes:likemanyofourotherinstruments.Thisfeaturecanbereducedfrom30minutesdownto1minutein1minuteincrements.Incaseswhereaconstantreadingisneces-sary,thiscanalsobeturnedcompletelyofftoallowforcontinuousoperation.
CONFIGThisisthekeytoaccessalluser-settablefunctions.TheCONFIGkeyisusedinconjunctionwiththe▼and ▲ functionsontheZEROandMAX/MINkeystoselectandchangedifferentfunctions.
Youcansetoneof20engineeringunits,changetheauto-shutofffunctionsetting,displaytheactualbatteryvoltage,displaytheactualtemperature(in°Cor°F),turnthedamp-eningonoroff,changethesamplerate,andsettheTarevalue.Inadditiontothe20availableengineeringunitsontheIPI,youcancreateyourownunittomeetyourmeasurementneed.Forexample,ifyouneedtotakealevelmeasurement,simplydeterminethepressurethatequatestooneunitofthatmeasurementandsetthatasyourengineeringunit.Theapplicationmayrequiregallons,feet,inches,meters,liters,orsimilar;withtheIPIyoucantakethemeasurement.
ENTERSaveselectionsandturnthebacklightonandoff.
External powerSometimesitismoreconvenienttotakeadvantageofanexistingpowersourceratherthanusingbatteries.TheIPImaybeorderedwitha24VDCpowersupplyoption.Theunithasterminalstoacceptthepositiveandnegativeleadstoallowforeasyconnectiontothepowersupply.Theseunitswillnotbesuppliedwithbatteriesandbatteriesshouldnotbeinstalledwhentheunitisconnectedto24VDCpower.
TheJOFRAIPIdigitalpressureindicatortakestheconceptofananalogtestgauge,andbringsittoanewlevel.TheIPIcombinestheaccuracyofdigitaltechnologywiththesimplicityofananaloggauge,andachievesperformance,ease-of-use,andafeaturesetunmatchedinthepressuremeasurementworld.SetupoftheIPIisfastandstraight-forward,throughamenu-drivendisplay,withminimaltext,andintuitivefunctions,thatissimpleenoughtoallowtheindicatortobeusedanywhereintheworld,withouttheneedformultilingualdisplays.
Protective bootStandardconfigurationIPIunitsandthosesuppliedwithsystemsaredeliveredwithaprotectiveboot.Unitswithbackmountingandthe24VDCpowersupplyarenotdeliv-eredwiththeprotectivebootduetodesignlimitations.
Serial communicationsTheIPIhasaserialportthatisaccessiblethoughthebackofthecase.ThisfeatureusesstandardASCIIcommandstoallowforextractionofdatafromtheIPIwhiletakingreadings.Thisrequirestheoptionalserialcable.
Hazard location information / approvalsAnexplosiveatmosphereisdefinedasa”mixturewithair,underatmosphericconditions,offlammablesubstancesintheformofgases,vapours,mistsordustsinwhich,afterignitionhasoccurred,combustionspreadstotheentireunburnedmixture”.
ThestandardintheEuropeanUnionhasbeensetwiththe9/94/ECDirective,commonlycalledATEX(“AtmosphèresExplosibles,”Frenchforexplosiveatmospheres).
TheJOFRAIPIIndustrialPressureIndicatorisATEXapprovedbyKEMAascomplyingwiththeEssentialHealthandSafetyRequirementsrelatedtothedesignandcon-structionofequipmentandprotectivesystemsintendedforuseinpotentiallyexplosiveatmospheresgiveninAnnexIIinthedirective,andwiththefollowingrating:II3GEExnAIIBT6(Ta=–10°C...+55°C).
TheJOFRAIPIIndustrialPressureIndicatorisalsocertifiedbyCSAasconformingtorelevantCanadianandUSAstandardswiththefollowingrating:Class1,Div.2,GroupsA-D.
SeethedefinitionsregardinghazardouslocationsinNFPA70,Article500orCSAC22.1Section18.NFPA70,Article500andCSAC22.1Section18.
Field recalibrationTheIPIdoesnotneedtobereturnedtothefactoryforcalibration.Ifyouhaveareliableandaccuratepressurereferenceoralocallaboratory,youmayre-calibratetheIPIlocally.Thisfeatureispasswordprotected.
Sampling rateTheIPIsamplingrateisuserselectable.Ifyouwanttocapturefastsystemtransients,theIPIcantakeareading10timespersecond.Conversely,youmaywanttocon-servebatterylifeandonlyneedperiodicsamples.Then,youcanselect1sampleperevery2seconds.Thisworkswellforinprocessandpanelmountedapplications.Youcanalsochoosetheacceptedinstrumentsamplingrateof3samplespersecondfornormaluseortoallowforfilter-ing,youcanselect1samplepersecond.
Damping adjustmentThedampingfunctioncanbeturnedonoroffontheIPI.Thisallowsforreadingstobeintegrated,whichaccountsformomentarychangessuchasthosefrompulsingsources.
TareBeyondzeroingtheIPI,youmayhavetoaccountforresidualpressure.TheTarefeatureallowsyoutotakecareofthaterrorandpreventthemanualcalculationofthedif-ference.Thiscanbeusedincombinationwiththecustomengineeringunitstomakethelevelmeasurementseasier.
Temperature display and compensation BecausetheIPIisdesignedforin-processtasks,temperaturecompensationisincludedtomakethejobeasier.ThisallowstheIPItomaintainaccuracyoverthemeasurementrange.Themeasuredtemperaturemaybecheckedfromthekeypad.
REQUIREMENTS JOFRACALMinimum hardware requirements: • Intel®Pentium®lI1.4GHzprocessor.
• 64MBRAM(128MBrecommended)
• 80MBfreediskspaceonharddisk(120MBrecommended)priortoinstallation
• StandardVGA(800x600,256colours).1024x768recommended.
• CD-ROMdriveforinstallationofprogram
• 1or2freeRS-232serialports,dependingonconfiguration
Minimum software equirements:• MicrosoftWindows®98,MicrosoftWindows®NT4.0,
MicrosoftWindows®2000,MicrosoftWindows®ME,MicrosoftWindows®XP,Vista.
• Systemfonts:MSSansSerifandArial
JOFRACAL CALIBRATION SOFTWARE
JOFRACALcalibrationsoftwareensureseasycalibrationofRTD´s,thermocouples,transmitters,thermoswithes,pressuregaugesandpressureswitches.JOFRACALcanbeusedwithJOFRADPC-500,APC,CPCandIPIpres-surecalibrators,allJOFRAtemperaturecalibrators,aswellasJOFRAAMC910,ASC300multisignalcalibratorandASM-800signalmultiscanner.WhenusedwithJOFRAASM-800signalmultiscanner,JOFRACALcanperformasimultaneoussemiautomaticcalibrationonupto24pres-sureand/ortemperaturedevicesundertestinanycombi-nation.
JOFRACALsoftwarecontrolsthecompletecalibrationprocedure,storestheresultsandprovidesacalibrationaudittrailthroughhard-copycertificates.Allcalibrationdataarestoredforeachsensortomonitordriftandopti-miserecalibrationintervals.Aschedulerfeatureallowsplanningoffuturecalibrations.
4 Industrial Pressure Calibrator • IPI
IPI
111 mm36 mm
38 mm
25,4 mm
99,3 mm
125 mm
Pressure; gauge / compound ranges
psi...........................................vacuumto30,100,300,500bar....................................................vacuumto2,7,21,35psi.................... 0to15,1,000,2,000,3,000,5,000,10,000bar........................................ 0to1,70,140,200,350,700
Pressure; absolute ranges
psi............................................................................0to100bar...............................................................................0to7
Engineering units
Userdefined................................... Oneuser-definableunitUserselectable........................................................ 20units
(PSI,Bar,kg/cm2,inH2O(4°C,20°Cor60°F),ftH2O(4°C,20°Cor60°F),mmH2O(4°Cand20°C),cmH2O(4°Cand20°C),mH2O(4°Cand20°C),KPa,mBAR,inHg,mmHg,Torr)
Notallunitsareavailibaleinallranges.
Pressure accuracy
Pressure........................................................... ±0.05%F.S.Fulltemperaturecompensation......0to50°C/32to122°FVacuum(100,300,500psiindicator/7,21,35barindicator)...................... ±0.25%F.S.Vacuum(30psiindicator/2barindicator).......... ±0.1%F.S.
Includingnon-linearity,hysteresis,repeatability,andtemperatureeffect.F.S.(fullscale)=maximumgaugepressureinpsi
Serial communication interface
Connector.......................................................... StereojackSerial......... 0-3VDC,9600baud,8data,noparity,1stopProtocol..................................... ASCllcommandlanguage
Media compatibility
Liquidsandgassescompatiblewith316stainlesssteel.
Environmetal
Storagetemperature....................-20to70°C/-4to158°FOperatingtemperature...............-10to55°C/-14to131°F
Pressure connection
Allranges.....................................................1/4”NPTmale
Adaptersto1/4”BSPmaleareincludedasstandard
Pressure overload
1to35bar/15to500psi……..............................3Xrange70to350bar/1,000to5,000psi........................2Xrange700bar/10,000psi…………….........................1.5XrangeOverloadalarm…………………………………OLondisplayOverloadalarmrange……………………………1.2Xrange
Display
Display.......................................................... Backlight,blueDisplayresolution.......................... 5digitfloatingdecimalBargraph........................................ 20segment,0to100%Displayindicators..............................Engineeringunitsicon...................................................Lowbatteryindicationicon.............................. Batterylifeindicationutilizingbargraph.............................................MeasuredtemperaturedisplayDisplayupdate......................................... twicepersecond
Power supply
Battery.......................................................... (3)AAAlkalineBatterylife......... 1500operationalhourswithoutbacklightBatterylife......... 2000operationalhourswithoutbacklight..............................................................atlowsamplingrateBatterylife.................150operationalhourswithbacklightLowbatteryindicator.............................................at3VDCOptionalinputport..................... 24VDCconnection1)2)3)
1)Unitsdeliveredwith24VDCpowersupplyconnectionarenotsuppliedwithbatteries
2)Batteriesshouldnotbeinstalledwhenusing24VDCpowersupplyasdamagetotheIPImayoccur
3)Unitswith24VDCpowersupplyconnectioncannotbeorderedwiththeprotectivebootoption
Instrument dimensions
IndicatorLxWxH..... 125x111x38mm/4.9x4.4x1.5inIndicatorweight(includingbattery)................1.6lb/0.7kgInputport..........................1/4”NPTMalelowerconnectionOptionalinputport........1/4”NPTMalebackconnection1)
3)Unitswithbackconnectioncannotbeorderedwiththeprotec-tivebootoption
Shipping dimensions
IndicatorLxWxH... 250x160x100mm/9.8x6.3x3.9inIndicatorweight..............................................1.8lb/0.8kgInstrument case
Rating..........................................................NEMA4/IP651)
1)Doesnotapplyiftheserialconnectionisinuseoriftheexternal24VDCpoweroptionisused.
Approvals - IPI system
CEConformity..............EN61326:1998,EN60079-0:2006,................................................................. EN60079-15:2005
TheJOFRAIPIsystemsaretypeapprovedbyDetNorskeVeritas.Findthecertificateatwww.jofra.comApproval,Certificateno..........................................A-10549
Ex approvals - IPI indicator only
CSA......................................Class1,Div.2,GroupsA-D1)
ATEX................ II3GEExnAIIBT6(Ta=–10°C...+55°C)1)
1)The24voltversionisnotCSAorATEXapproved
FUNCTIONAL SPECIFICATIONS
JOFRA IPI PRESSURE RANGES
Resolution obtained by the IPI indicator
IPI015G0 to 15 psi0 to 1 bar
IPI01KG0 to 1,000 psi0 to 70 bar
IPI02KG0 to 2,000 psi0 to 140 bar
IPI03KG0 to 3,000 psi0 to 200 bar
IPI05KG0 to 5,000 psi0 to 350 bar
IPI10KG0 to 10,000 psi 0 to 700 bar
Imperial ranges
psi 15.000 1000.0 2000.0 3000.0 5000.0 10000
inH2O@4°C 415.21 27681 55361 83042 N/A N/A
inH2O@20°C 415.95 27730 55459 83189 N/A N/A
inH2O@60°F 415.62 27708 55416 83124 N/A N/A
ftH2O@4°C 34.601 2306.7 4613.5 6920.2 11534 23067
ftH2O@20°C 34.662 2310.8 4621.6 6932.4 11554 23108
ftH2O@60°C 34.635 2309.0 4618.0 6927.0 11545 23090
inHg@0°C 30.540 2036.0 4072.1 6108.1 10180 20360
Torr 775.73 51715 N/A N/A N/A N/A
Metric ranges
bar 1.0000 70.000 140.00 200.00 350.00 700.00
mbar 1000.0 70000 N/A N/A N/A N/A
kPa 100.00 7000.0 14000 20000 35000 70000
kg/cm2 1.0197 71.380 142.76 203.94 356.90 713.80
cmH2O@4°C 1019.7 71382 N/A N/A N/A N/A
cmH2O@20°C 1021.5 71508 N/A N/A N/A N/A
mH2O@4°C 10.197 713.82 1427.6 2039.5 3569.1 7138.2
mH2O@20°C 10.215 715.08 1430.2 2043.1 3575.4 7150.8
mmHg@0°C 750.06 52504 N/A N/A N/A N/A
Resolution obtained by the IPI indicator
IPI30CVacuum to 30 psiVacuum to 2 bar
IPI100A / IPI100CVacuum to 100 psiVacuum to 7 bar
IPI300CVacuum to 300 psiVacuum to 21 bar
IPI500CVacuum to 500 psiVacuum to 35 bar
Imperial ranges
psi -12.000 30.000 -12.000 100.00 -12.000 300.00 -12.00 500.00
inH2O@4°C -332.17 830.42 -332.2 2768.1 -332.2 8304.2 -332 13840
inH2O@20°C -332.76 831.89 -332.7 2773.0 -332.7 8318.9 -333 13865
inH2O@60°F -332.50 831.24 -332.5 2770.8 -332.5 8312.4 -332 13854
ftH2O@4°C -27.681 69.202 -27.68 230.67 -27.68 692.02 -27.7 1153.4
ftH2O@20°C -27.730 69.324 -27.73 231.08 -27.73 693.24 -27.7 1155.4
ftH2O@60°C -27.708 69.270 -27.71 230.90 -27.71 692.70 -27.7 1154.5
inHg@0°C -24.432 61.081 -24.43 203.60 -24.43 610.81 -24.4 1018.0
Torr -620.6 1551.5 -620.6 5171.5 -620 15514 -621 25858
Metric ranges
bar -0.8300 2.0000 -0.8300 7.0000 -0.8300 21.000 -0.8300 35.000
mbar -830.0 2000.0 -830.0 7000.0 -830.00 21000 -830 35000
kPa -83.00 200.00 -83.00 700.00 -83.0 2100.0 -83.0 3500.0
kg/cm2 -0.8464 2.0394 -0.8464 7.1380 -0.846 21.414 -0.846 35.690
cmH2O@4°C -846.4 2039.5 -846.4 7138.2 -846 21415 -846 35691
cmH2O@20°C -847.9 2043.1 -847.9 7150.8 -847 21452 -848 35754
mH2O@4°C -8.464 20.395 -8.464 71.382 -8.46 214.15 -8.46 356.91
mH2O@20°C -8.479 20.431 -8.479 71.508 -8.48 214.52 -8.48 357.54
mmHg@0°C -622.6 1500.1 -622.6 5250.4 -622 15751 -623 26252
ThistableshowstheresolutionsthatcanbeobtainedbytheIPIindicatorsthroughoutallengineeringunits.
6 Industrial Pressure Calibrator • IPI
•
AccessoriesPart No. Description75P014 ServicekitforT-960pump75P016 ServicekitforT-970pumpT-733-2 PressurehoseforT-960/970,0.6meterstd.T-733-3 PressurehoseforT-960/970,0.5meterT-733-4 PressurehoseforT-960/970,1.0meterT-733-5 PressurehoseforT-960/970,2.0meterT-733-6 PressurehoseforT-960/970,5.0meter65R191 SetofBSPfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x3/8”,1x1/2”65R192 SetofNPTfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x1/4”,1x3/8”1x1/2”60R191 1setofpackingandsealsfor65R191and65R192123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACSPK-HPC-005 QuickconnectorsetforIPIsystemA/B
Ordering information System AOrder no. Description Pressure indicator IPI IndustrialPressureIndicator
Indicator pressure range 015G 0to1bar(15psi) 100A 0to7barabsolut(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number AXX IPISystemA(pumpT-960) AHX IPISystemA(pumpT-970)
Certification G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI030CBXXAXXG Sample order numberJOFRAIPI2bar(30psi)indicatorwithbootandinpres-sureSystemAwithpneumaticpressurehandpumpandNISTtraceablecalibrationcertificate.
• 0to1bar(15psi)• Vacuumto2bar(30psi)• Vacuumto7bar(100psi)• Vacuumto21bar(300psi)
ThissystemincludestheJOFRAIPIcalibratortogetherwithoneofthefollowingpneu-matichandpumps:T-960andT-970,dependingonpressurerange.SystemAisaneasy-to-usesinglehandoperatedcalibrationsystem.Noneedforaflatsurfacetooperatethesystem-justholditinyourhand.
TheSystemAcomesinanaluminiumcarryingcasewithcutoutsforfittings,hose,andthecompleteassembledcali-brationunit-notimeisrequiredtoassembletheuniteverytimeyouneedit.Thecalibrator(indicator)canatanytimebeusedseparatelyforotherpressuretesttasks.Aspecialquickconnectorbetweenthepumpandtheindi-catormakesitpossibletoseperatethesysteminsecondsandtoswiveltheindicatorforeasyviewing.
TheJOFRAIPIdeliveryalwaysincludesaprotectivebootfortheindicator.AllIPIindicatorsusedinSystemAaredeliveredcalibratedinbothpressureandvacuum.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
JOFRA IPI System A
T-960 and T-970 pumps for system AModelT-960pneumaticpressurepumpfeaturesanextendedrangevernieradjustmentforpreciselycontrollingpumppressureandcomfortablepistolgriphandle.
ModelT-970pneumaticpressurepumpissimilartotheT-960,butgeneratepressureupto40bar(580psi).
Pressurerange,T-960.....................0to2bar/0to30psiPressurerange,T-970..................0to40bar/0to580psiType..................................................................... PheumaticTestmedium.....................................................................AirOperation.................................................................. Scissor“O”-rings..................................................................Buna-NWettedparts......................Aluminium,brass,stainlesssteel,...................................................................... nylon,NylatronGSConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize.........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb
Ordering Part No. DescriptionT-960 Pump:0to2bar(30psi)T-970 Pump:0to40bar(580psi)
Included with System A delivery• Standarddelivery(seepage12)• HandpumpT-960orT-970• 3/8”BSPfemaleadapterwith1/4”BSPandNPT
femaleterminations
Shipping InformationSize...................570x180x350mm/22x7,1x13,8inchWeight........................................................5,5kg/12.13lb
• 0.6m/24inhosewith1/4”BSPandNPTfemaleterminations
• 1rollofTeflontape• Protectivecarryingcase• QuickConnectorbetweenpumpandcalibrator
AccessoriesPart No. Description75P017 ServicekitforT-975pumpT-733-2 PressurehoseforT-975,0.6meterstd.T-733-3 PressurehoseforT-975,0.5meterT-733-4 PressurehoseforT-975,1.0meterT-733-5 PressurehoseforT-975,2.0meterT-733-6 PressurehoseforT-975,5.0meter65R191 SetofBSPfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x3/8”,1x1/2”65R192 SetofNPTfemalefittings,teflontape,packings *Fittingsize:1x1/8”,1x1/4”,1x3/8”1x1/2”60R191 1setofpackingandsealsfor65R191and65R192123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACSPK-HPC-005 QuickconnectorsetforIPIsystemA/B
Included with System B delivery• Standarddelivery(seepage12)• HandpumpT-975forallranges• 3/8”BSPfemaleadapterwith1/4”BSPandNPT
femaleterminations• 0.6m/24inhosewith1/4”BSPandNPTfemale
terminations• 1rollofTeflontape• Protectivecarryingcase• QuickConnectorbetweenpumpandcalibrator
Ordering information System BOrder No. Description Pressure indicator IPI IndustrialPressureIndicator
Indicator pressure range 100A 0to7barabsolute(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) 500C Vacuumto35bar(500psi) NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number BXX IPISystemB(pumpT-975)
Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI300CBXXBXXG Sample order number JOFRAIPI21bar(300psi)indicationwithbootandinpressureSystemBkitwithpneumaticcombinationhand-pumpandNISTtraceablecalibrationcertificate.
T-975 hand pump for system BModelT-975pneumaticpressurepumpfeaturesbothvacuumandpressuregeneration.Abutton(valve)makesiteasytoswitchfromvacuumtopressuremeasurements.Built-inreleasevalve,verniervalveforfineadjustmentanddualpressureoutputenablesafeandsimpleoperation.Theunitoffersbothmetricandimperialthreadsontheref-erenceconnectionandthehose.
Pressurerange...........-0.91to40bar/-27inHgto580psiType..................................................................... PheumaticTestmedium.....................................................................AirOperation.................................................................. Scissor“O”-rings..................................................................Buna-NWettedparts.......Anodizedaluminium,nickel-platedbrass,............................................................ stainlesssteel,NylonConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize.........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb
OrderingPart No. DescriptionT-975 Pump:-0.91to40bar/-27inHgto580psi
• Vacuumto2bar(30psi)• Vacuumto7bar(100psi)• Vacuumto21bar(300psi)• Vacuumto35bar(500psi)
Thissystemincludesthedualfunctionpneumatichandpump.Withthispumpyoucancali-bratebothvacuumandpres-sureapplications.Ittakesjustafewsecondstoswitchbetweenvacuumandpressure,justpushonebutton(valve)ontheT-975handpump.
TheSystemBcomesinanaluminiumcarryingcasewithcutoutsforfittings,hoseandthecompleteassembledcali-brationunit-notimerequiredtoassembletheuniteverytime.
Theindicatorandthepumparefittedwithaspecialquickconnectorthatmakesitpossibletoseparatethesysteminsecondswithoutanytool.Thisquickconnectoralsomakesitpossibletousetheassembledsystemwiththeindicatortwistedtoanyangle.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
JOFRA IPI System B
Shipping InformationSize...................570x180x350mm/22x7,1x13,8inchWeight........................................................5,5kg/12.13lb
8 Industrial Pressure Calibrator • IPI
AccessoriesPart No. DescriptionT-656 ServicekitforT-620andT-620HpumpsT-649 PressurehoseforT-620/T-620Hpumps65P175 1mhosewith1/4”BSPfemaletermination Max.pressure350bar(5,000psi)65P180 2mhosewith1/4”BSPfemaletermination Max.pressure350bar(5,000psi)123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACMGAAA/GL OilforT-620&T-1pumpAAAOILin 1-GALLONCANMGAAA/QT OilforT-620&T-1pumpAAAOILin1- QUARTCAN
Shipping InformationSize...................560x188x350mm/22x7,4x13,8inchWeight........................................................5,5kg/12.13lb
• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)
ThissystemconsistsofanIPIindicatortogetherwithahydraulic,highpressurehandpumpT-620orT-620H,whichfeaturesanoilreservoirtoprimethesystem.SystemCisaneasy-to-usesingle-handoperatedtestsystem.Noneedforaflatsurface(table)tooperatethesystem.Thesystemincludesreleasevalveandverniervalveforfineadjustmenttoenablesafeandsimpleoperationofthepump.
TheSystemCcomesinaprotectivecarryingcasewithcutoutsforfittings,hose,andthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressureindicatorseparatelyforotherpressuretesttasks.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
Included with System C delivery• Standarddelivery(seepage12)• HandpumpT-620orT-620H• 0.6m(24in)hosewith1/4”NPTfemaleand1/4”BSP
terminations• 1rollofTeflontape• Protectivecarryingcase
AMETEK hydraulic handpumpThiseasy-to-usehandpumpisspeciallydesignedforhighpressureapplications.Thepumphasabuilt-inreservoir,ventvalve,andverniervalveforfineadjustment.Thepumphasadualpressureoutput-oneforthetestobjectandonefortheIPIindicator.
Pressurerange,T-620.................... 0to200bar/3,000psiPressurerange,T-620H................. 0to350bar/5,000psiType....................................................................... HydraulicTestmedium.....................................................................OilReservoircapacity.......................................200cl/0.5pintOperation..............................................................Pistol-grip“O”-rings..................................................................Buna-NWettedparts..........Aluminum,brass,stainlesssteel,LexanConnectiontotestobject....................... Hose0.6m/24in........................with1/4”BSPandNPTfemaleterminationsSize........................21.6x12.1x6.2cm/8.5x4.8x2.4inWeight............................................................1.4kg/3.0lb
OrderingPart No. DescriptionT-620 Pump:0to200bar(3,000psi)T-620H Pump:0to350bar(5,000psi)
Ordering information IPI System COrder No. Description Pressure indicatorIPI IndustrialPressureIndicator
Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar3,000psi) 05KG 0to350bar5,000psi) NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number CXX IPISystemC(pumpT-620/T-620H)
Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI01KGBXXCXXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootandinpressureSystemCwithhydraulicpressurehandpumpandNISTtraceablecalibrationcertificate.
JOFRA IPI System C
AccessoriesPart No. Description60P013 Verniervalvefineadjustmentfor65-P016/01760P016 Packingsetforscrewpumpoilpartno.65-P01660P017 Packingsetforscrewpumpwaterpartno.65-P01765-R975 Reservoirsetwithseals,valveandbondedseal60R135 Valveforoilreservoir65P175 1m.Pressurehose,(Max.350bar/5,000psi) *1/4”BSPfemaleto1/4”BSPmale65P180 2m.Pressurehose,(Max.350bar/5,000psi) *1/4”BSPfemaleto1/4”BSPmale60I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale65N000 Extensiontube,120mm(Max.350bar/5.000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VAC50-REP615 Quickconnectorset,female1/4”BSP to1/4”BSPSPK-HPC-004BanjoConnectorSetforpump
Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight...........................................................12kg/26.5lb
• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)
ThissystemconsistsofanIPIindicatortogetherwitharug-ged,hydraulic,highpressurescrewpump.SystemDisaneasy-to-usetestsystem.Placethepumponaflatsurfaceandturnthehandletogeneratepressure.Youcanadjusttheindicatortofitthebestviewingangle.Thesystemincludesamanifoldforconnectionofthetestdevice,afineadjustmentverniervalve(optional),andareservoirforextraoilorwater(optional).
TheSystemDcomesinaprotectivecarryingcasewithcutoutsforfittingsandthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
AMETEK screwpumpThe65-P016and65-P017screwpumpsaredesignedforveryeasygenerationofhighpressure.Thescrew-pumpshasa4connectionmanifold.Fineadjustment,verniervalve,liquidreservoir,andfittingsareoptional.TheAMETEKscrewpumpmaybedeliveredfor2differenttestmedia:hydraulicoilorwater.
Pressurerange............................... 350bar/0to5,000psiType....................................................................... HydraulicTestmedia............................................. Hydraulicoil,waterOperation..........................................................Screwpump“O”-rings..................................................................Buna-NWettedparts..................... Aluminum,brass,stainlesssteelConnectionstotestobjects................ 4x1/4”BSPfemaleSize....................36.6x19.5x10.5cm/14.4x7.7x4.1inWeight............................................................3.7kg/8.2lb
OrderingPart No. Description65-P016 Screwpumpoil:0to350bar(5000psi)1)65-P017 Screwpumpwater:0to350bar(5000psi)1)
1)25ml(0.85oz)reservoir
Included with System D delivery• Standarddelivery(seepage12)• Screwpump65-P016forhydraulicoilor65-P017for
water• 1rollofTeflontape• 10bondedseals• 6mmAllenkey• Protectivecarryingcase• BanjoConnectorSet
Ordering information IPI System DOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator
Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number DOX IPISystemD(pump65-P016) DWX IPISystemD(pump65-P017)
Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI01KGBXXDOXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootandinpressureSystemDwithhydraulicoilscrewpumpandNISTtraceablecalibrationcertificate.
JOFRA IPI System D
10 Industrial Pressure Calibrator • IPI
AMETEK high pressure hydraulic oil pump The65-P014hydraulicpumpisdesignedforhighpres-sureapplications.Thepumphasa4connectionoutputmanifold.Fineadjustmentverniervalveandfittingsareoptional.
Pressurerange............................. 0to700bar/10,000psiType....................................................................... HydraulicTestmedia........................................................................OilOperation............................................................ Jackpump“O”-rings..................................................................Buna-NWettedparts..................... Aluminum,brass,stainlesssteelConnectionstotestobjects................ 4x1/4”BSPfemaleSize....................36.5x20.0x14.2cm/14.4x7.9x5,6inWeight........................................................ 5.3kg/11.6lbs
Ordering Part No. Description65-P014 Hydraulicpumpforoil:0to700bar(10,000psi)1)
1)450ml(15.22oz)reservoir
• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)• 0to700bar(10,000psi)
ThissystemconsistsofanIPIindicatortogetherwitharug-ged,hydraulic,highpressurepump.
SystemEisaneasy-to-usetestsystem.Thehydraulicpumpmakesitveryeasytoprimethesystem.Youcanadjusttheindicatortofitthebestviewingangle.Thesys-temincludesamanifoldforconnectionoftwotestdevicesandthefineadjustmentverniervalveisstandard.
TheSystemEcomesinaprotectivecarryingcasewithcutoutsforfittingsandthecompleteassembledtestunit-notimeaddertoassembletheuniteverytime.Easyandfastconnectionbetweenpumpandindicatormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
AccessoriesPart No. Description60P013 Verniervalveforfineadjustment60P014 PackingsetforJackpumppartno.65-P01460I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VAC50-REP615 Quickconnectorset,female1/4”BSP to1/4”BSP
Included with System E delivery• Standarddelivery(seepage12)• Hydraulicpump65-P014forhydraulicoil• Verniervalve• 1rollofTeflontape• 10bondedseals• 6mmAllenkey• Protectivecarryingcase
Ordering information IPI System EOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator
Indicator pressure range01KG 0to70bar(1,000psi)02KG 0to140bar(2,000psi)03KG 0to200bar(3,000psi)05KG 0to350bar(5,000psi)10KG 0to700bar(10,000psi)NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number EXX IPISystemE(pump65-P014)
Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI01KGBXXEXXG Sample order number JOFRAIPI70bar(1,000psi)indicatorwithbootaninpressuresystemSystemEwithhighpressurepumpforhydraulicoilandNISTtraceablecalibrationcertificate.
JOFRA IPI System E
Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight...........................................................13,5kg/30lb
Included with System F delivery• Standarddelivery(seepage12)• TypeTHydraulicpump• 1rollofTeflontape• Protectivecarryingcase
• 0to35bar(500psi)• 0to70bar(1,000psi)• 0to140bar(2,000psi)• 0to200bar(3,000psi)• 0to350bar(5,000psi)• 0to700bar(10,000psi)
TheSystemFhydraulicpumpmakesitveryeasytoprimethesystem.Itcontains75in3/1.23lofliquid.Thesystemincludesamanifoldforcon-nectionoftwotestdevicesandthefineadjustmentverniervalveisstandard.TheSystemFcomesinaprotectivecar-ryingcasewithcutoutsforfittings.Easyandfastconnec-tionbetweenpumpandcalibratormakesiteasytousethepressurecalibratorseparatelyforotherpressuretesttasks.
OnlytheIPIindicatorisapprovedforuseinpotentiallyexplosiveatmospheres.
AccessoriesPart No. DescriptionT-250 ServicekitforTpumpwithBunaNO-ringsT-250 ServicekitforTpumpwithVITONNO-ringsT-250 ServicekitforTpumpwithEPTNO-rings60I156 1.5m.Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale60I157 5meter,Pressurehose,(Max.700bar/10,000psi) *1/4”BSPfemaleto1/4”BSPmale123958 RS232cablewithstereoJackconnector,2m/6ft124716 4x1,5Voltrechargeablebatteries124718 Chargerforrechargeablebatteries-115/230VACMGAAA/GL OilforT-620&T-1pumpAAAOILin 1-GALLONCANMGAAA/QT OilforT-620&T-1pumpAAAOILin1- QUARTCAN
Ordering information IPI System FOrder No. Description Pressure indicatorIPI IndustrialPressureIndicator
Indicator pressure range 500C Vacuumto35bar(500psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) 10KG 0to700bar(10,000psi) NONE Pumpsystemonly(nocalibrator)
Boot B Withboot(standardforindicatorwith batterypowerandlowermounting– andonallsystemkits)
Power supply X Batteries(standard)
Mounting X Lowermount(standard)
Base model number FWB IPISystemF(pumpT-1) FWV IPISystemF(pumpT-1/VITON) FWE IPISystemF(pumpT-1/EPT) FOB IPISystemF(pumpT-1/OIL) FOV IPISystemF(pumpT-1/OIL/VITON) FOE IPISystemF(pumpT-1/OIL/EPT)
Certificate G NISTtraceablecertificate(standard) H Accreditedcertificate (optionalquotationbasis)
IPI10KGBXXFOEG Sample order number JOFRAIPI700bar(10,000psi)indicatorwithbootandinpressureSystemFwithTypeTpumpforoil,EPTsealsandNISTtraceablecalibrationcertificate.
JOFRA IPI System F
1)Customermustverifycompatibilityofsealsandpumpmaterialswithpressuremedium.Vitonisaregisteredtrademark.2)BunaNarethestandard”O”ringmaterialforthehydraulictesters.Theyareusedwithwater/alcoholornoncorrosiveoils.3)Viton”O”ringmaterialisusedforallcorrosiveoilsotherthanSkydrol.4)EPT”O”ringmaterialisspecifi-callyforSkydrolOilusedforhydraulicsystemsinairplanes,itisverycorrosive.
AMETEK Type T hydraulic oil or water pumpThetypeTpumpisdesignedforhighpressureapplica-tionsupto15,000psi(1,000bar).Thepumpandsystemmaybeorderedwitheitheroilorawater/alcoholmixtureasthepressuremedium.Thepumpisconstructedof300seriesstainlesssteelandMonelallowingfortheuseofotherhydraulicmedia.Therearethreeavailablesealpackagesforthesystem:Buna-N,Viton®,andEPT.1)Thepumpfeaturesadualpressureoutputmanifold,fineadjust-mentverniervalve,reliefvalve,anddualvolumecontrolforrapidpressureincreaseatlowerpressuresandeasierpumpingathigherpressures.
Pressurerange.......................... 0to1,000bar/15,000psiType....................................................................... HydraulicTestmedia............................................. Hydraulicoil,waterReservoircapacity...........................................1.23l/75in3
Operation............................................................ Jackpump“O”-rings......... Buna-N(standard)orEPT/Viton(optional)Wettedpart.......................................Stainlesssteel,MonelConnectionstotestobjects................... 1/4”and1/2”BSP............ terminations;1/4”and1/2”PTfemaleterminationsSize................................ 79.4x22.9x50.8cm/31.3x9x20inWeight........................................................... 9.1kg/20lbs
OrderingPart No. Description Seals T-1 0to1,000bar(15,000psi)water-BunaN2)
T-1/VITON 0to1,000bar(15,000psi)water-Viton3)
T-1/EPT 0to1,000bar(15,000psi)water-EPT4)
T-1/OIL 0to1,000bar(15,000psi)oil-BunaN2)
T-1/OIL/VITON 0to1,000bar(15,000psi)oil-Viton3)
T-1/OIL/EPT 0to1,000bar(15,000psi)oil-EPT4)
Shipping InformationSize...................730x220x430mm/28,7x8,7x17inchWeight..............................................................18kg/40lb
www.jofra.comInformation in this document is subject to change without notice. ©2009, by AMETEK, Inc., www.ametek.com. All rights reserved.
Pub code SS-IPI Issue 0912
AMETEK Calibration Instrumentsis one of the world’s leading manufacturers and developers of calibration instruments for temperature, pressure and process signals as well as for temperature sensors both from a commercial and a technological point of view.
JOFRA Temperature InstrumentsPortable precision thermometers. Dry-block and liquid bath calibrators: 5 series, with more than 25 models and temperature ranges from -90° to 1205°C / -130° to 2200°F. All featuring speed, portability, accuracy and advanced documenting functions with JOFRACAL calibration software.
JOFRA Pressure InstrumentsConvenient electronic systems ranging from -25 mbar to 1000 bar (0.4 to 15,000 psi) - multiple choices of pressure ranges, pumps and accuracies, fully temperature-compensated for problem-free and accurate field use.
JOFRA Signal InstrumentsProcess signal measurement and simulation for easy control loop calibration and measurement tasks - from handheld field instruments to laboratory reference level bench top instruments.
JOFRA / JF Marine InstrumentsA complete range of calibration equipment for temperature, pressure and signal, approved for marine use.
FP Temperature SensorsA complete range of temperature sensors for industrial and marine use.
M&G Pressure TestersPneumatic floating-ball or hydraulic piston dead weight testers with accuracies to 0.015% of reading. M&G PumpsPressure generators from small pneumatic “bicycle” style pumps to hydraulic pumps generating up to 1,000 bar (15,000 psi).
...because calibration is a matter of confidence
Order No. Description Pressure indicatorIPI IndustrialPressureIndicator
Indicator pressure range 100A 0to7batabsolute(100psi) 030C Vacuumto2bar(30psi) 100C Vacuumto7bar(100psi) 300C Vacuumto21bar(300psi) 500C Vacuumto35bar(500psi) 015G 0to1bar(15psi) 01KG 0to70bar(1,000psi) 02KG 0to140bar(2,000psi) 03KG 0to200bar(3,000psi) 05KG 0to350bar(5,000psi) 10KG 0to700bar10,000psi()
Boot B Withboot(standardforindicatorwithbatterypower andlowermounting-andonallsystemkits) X Noboot(standardforindicatorswithbackmounting and/or24VDCpowerssupplyconnection)
Power supply X Batteries(standard) E Connectionforexternal24VDCpowersupply1)
Mounting X Lowermount(standard) B Backmount
Base model number IND IPIindicatorwithstandardaccessoriesforconfiguration
Certification G NISTtraceablecertificate(standard) H Accreditedcertificate(optional–onquotationbasis)
IPI030CBXXINDG Sample order numberJOFRAIPI2bar(30psi)indicatorwithboot,standardaccessories,andNISTtraceablecalibrationcertificate.
ORDERING INFORMATION - JOFRA IPI INDICATOR
• IPIindicator
• CalibrationcertificationperformancetraceabletoNIST
• (3)AAbatteries
• Adapterto1/4”BSPmale
• Usermanual
• Protectiveboot
STANDARD DELIVERY
1)The24voltversionisnotCSAorATEXapproved