What you don’t see… …CAN hurt you Sample Prep for Today’s Analytical World Christophe Deckers, M.Sc. Application Scientist, Sample Prep [email protected]
Transcript
Page 1: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

What you don’t see…

…CAN hurt you

Sample Prep for Today’s Analytical World

Christophe Deckers, M.Sc.Application Scientist, Sample Prep

[email protected]

Page 2: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Today’s Agenda

IntroductionAddressing difficult samples1. Polar analytes2. High fat samples3. How to get rid of interfering terpenesSummary and Conclusions

NACRW 2015

1March 17

Page 3: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Objectives of Sample Preparation

• Removal of interferences which would affect detection of analyte

• Removal of interferences that would affect instrument or column lifetime

• Concentration of an analyte to a detectable concentration

Page 4: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

With news instruments - more Sample Preparation Techniques can be used

4

More Specific ← Instrument Separation and Detection Specificity ← Less Specific

Less Specific → Sample Preparation Specificity → More Specific

Sample Prep Technique

Interference Removed

Dilute & Shoot Filtration Liquid/Liquid Extractions

Supported Liquid Extractions (SLE)

Dried Matrix Spotting Precipitation QuEChERS

Lipid Removal ‘Hybrid'

Filtration

Solid Phase Extraction

Lipids No No No Some No No Yes Yes Yes

Oligomeric Surfactants No No No No No No No Yes Yes

Particulates No Yes No Some No Yes Yes Yes Yes

Pigments No No No Some No No Yes No Yes

Polar Organic Acids No No Yes Yes No No Yes No

Proteins No No Yes Yes Yes Yes Yes Yes Yes

Salts No No Yes Yes No No No No Yes

Suggested Agilent Product

Agilent Autosampler

Vials

Captiva Syringe Filters Chem Elut Captiva ND Bond Elut

QuEChERS Captiva ND LIPIDSBond Elut Silica and Polymeric

SPE

Agilent Captiva Filtration Products are recommended for use with any LC or LC-MS method

Page 5: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Today’s Agenda

IntroductionAddressing difficult samples1. Polar analytes2. High fat samples3. How to get rid of interfering terpenesSummary and Conclusions

NACRW 2015

1March 17

Page 6: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Polar analytes often require advanced solid phase extraction (SPE)

- Analyte is too polar for acetonitrile extraction- Filtration is not clean enough- Ion exchange can be used to filter polar interferences away- We then often need an SPE phase that can bind a wide range of polar analytes that are basic, neutral and acidic

March 17

NACRW 2015

6

Page 7: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Bond Elut Packed Bed Phases – Which one!Non-polar

C18, C8, C2, C1C18 variations in carbon load and endcappingEnvirElutCH – cyclohexylCN-E – endcapped cyanoPH – phenylENV, LMS, PPL, Focus, Bond Elut Plexa

PolarPSA - primary and

secondary amineNH2 - aminopropylCN-U –unendcapped cyanoDEA - diethylaminopropylDiolSi - silica

Cation ExchangeSCX – benzenesulfonic acidPRS – propylsulfonic acidCBA – carboxylic acid

Anion ExchangeSAX – quaternary aminePSA – primary and

secondary amineNH2 – aminopropylDEA – diethylaminopropyl

Mixed mode IEX/NPCertify – SCX/C8Certify II – SAX/C8Plexa PCXPlexa PAX

Reversible CovalentPBA – phenylboronic acid

Alumina – aluminum oxideFlorisil – magnesium-silicaCarbonCarbon/NH2

Nexus

Specialty PhasesAccuCATAtrazineEtc..

Page 8: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Bond Elut Plexa - A Different Type of Polymeric Sorbent

Polar and non polar drugs bind in the SDVB hydrophobic end of the pore structure

• The hydroxylated ligands exist on the surface and in the outer regions of the pores. • Deep inside the pore is pure S/DVB. • There is no amide functionality (N-vinylpyrolidone) which might contribute to retaining matrix interferences

March 6, 2017Confidentiality Label

Page 9: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Plexa Recoveries Versus HLB – good results with a wide range of compounds

Analyte Plexa HLB pKa LogPAlbuterol 97.9 115.4 5.9 1.3

Atenolol 97.0 94.0 4.2 4.2

Loratadine 71.0 49.0 5.7 1.5

Metoprolol 92.0 74.0 5.7 1.5

Naltrexone 85.7 13.0 4.9 5.2

Pravastatin 85.0 59.0 4.9 5.2

Propranolol 55.0 35.0 4.9 5.2

Zolpidem 93.0 96.8 9.9 3.4

Plexa gives equivalent or better absolute recoveries than Oasis HLB for a wide range of pKa and logP compoundsConditions: Absolute Recovery from human plasma Basic load conditions, n=6, 200 ng/ml

Page 10: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Hydrophobic/ hydrophilic boundary

Sorbent Conditioning - Why Do It?

10

Page 11: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

11

Drying a silica-based Cartridge (ex: C18)

Over drying / lack of conditioning before sample application can contribute to

• low recoveries• poor flow• irreproducibility (high RSDs)

Page 12: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Examples on how polymeric SPE perform

March 6, 2017Agilent Confidential

12

Page 13: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Bond Elut Plexa / Phenols from Drinking Water

Page 14: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Bond Elut PPL - EPA 528 Phenols from Drinking Water

Method 528 provides procedures for the determination of phenols in finished drinking water. The method may be applicable to untreated source waters but has not been evaluated for these uses. The method is applicable to a variety of phenols that are efficiently partitioned from the water sample into a modified PSDB SPE sorbent (Varian’s PPL) and is sufficiently volatile and thermally stable for gas chromatography.

Extraction is performed by passing a 1L water sample through a SPE cartridge with 500mg modified PSDB (PPL). The phenols are eluted with methylene chloride.

Page 15: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Determination of Organophosphates in Lake Water: Bond Elut PPL SI-02094Julia Regnery, Goethe-Universität Frankfurt, Institut für Atmosphäre und Umwelt, AG-Umweltanalytik, 60438 Frankfurt, Germany, Elisabeth Korte, Varian Deutschland GmbH

15

Chlorinated organophosphates are used as flame retardants in insulation foams, paints, coatings, plastics, and textiles while the non-chlorinated ones are mostly used as plasticizers. The compounds have been detected in air, sediment and soil, sewage sludge, streams and lakes. Particularly the chlorinated organophosphates are known to persist in the aquatic environment.

• Because of the low concentration of the analytes in the water sample an enrichment step is necessary.

Page 16: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

16

Cartridge: Bond Elut PPL, 100 mg sorbent in 1 mL cartridge

Condition: 1 mL methanol, 1 mL methanol / acetonitrile (1/1)

Apply 1.5 -2.5 L water sample

Dry the cartridge using nitrogen

Elution with 3 x 333 µL methanol / acetonitrile (1/1)

SPE Conditions

Page 17: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Results

17

The simple clean-up and enrichment with SPE has the advantage that up to 20 water samples can be extracted simultaneously without using complex apparatus. Bond Elut PPL has been proven to be a robust sorbent with high capacity for the extraction of polar and medium polar analytes.

Page 18: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Conclusion

Novel polymeric SPE phases are recommended:

1. They have double the binding capacity over silica2. They can bind a wide range of polar analytes3. They don’t decondition/dry out on the SPE manifold4. They provide more robust methods

March 6, 2017Confidentiality Label

18

Page 19: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Today’s Agenda

Introduction

Addressing difficult samples

1. Polar analytes

2. High fat samples3. How to get rid of interfering terpenes

Summary and Conclusions

NACRW 2015

1

March 17

Page 20: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Remove interferences…

• To acquire desired sensitivity/selectivity• To reduce contamination/carryover issues• Use of sensitive and expensive instruments: Protect your

investment!!!

Pesticides in Avocado with SPPesticides in Avocado without SP

March 17

NACRW 2015

20

Page 21: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Lipids

March 17

NACRW 2015

21

Page 22: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Matrices and Approximate Total Lipid Content<2% 4-12% >12%

Beef Liver (4%)

Canola Oil (100%)

Nut Butters (16%)Spinach (0%)Strawberry (0%)

Onion (0%)

Canned Pet Food (~10%)

Avocado (21%)

Chocolate (8%)

Soy Milk (4%)

Pork Liver (4%)

Soy Oil (100%)Avocado Oil (100%)

Coconut Oil (100%)

Paprika (1%)

Hops (2%)

Cumin (2%)

Salmon (27%)Corn (4%)

Rice (2%)

Wheat (3%)

Catfish (12%)

Sea Bass (3%)

Carp (12%)

Trout (8%)

Tilapia (2%)

Pork (8%)

Plasma or whole blood(10 – 20 %)

Cow’s Milk (5%)

Low Med High

March 17

NACRW 2015

22

Page 23: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Current Procedures for Lipid Removal

Method How Lipids are removed Weakness

Dilute and Shoot No lipid removal, only dilution No lipid removal

Protein Precipitation PPT followed by centrifugation Insufficient lipid removal

PPT followed by filtration with or without sorbent

Insufficient lipid removal; low analyterecovery

QuEChERS PSA/C18 sorbent (dSPE) Not selective; insufficient lipid removal; analyte loss

Zr-containing sorbent Low total lipid capacity; analyte recovery

Freeze sample Time needed; loss of analyte

SPE/SLE Load and elute Time needed; solvent usage; extensive method development

SEC/GPC Chromatographic separation Uses copious amounts of solvent and time; capital expense

March 17

NACRW 2015

23

Page 24: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR-Lipid Enhanced Matrix Removal

EMR: As easy to use as QuEChERS; as clean as SPE

March 17

NACRW 2015

24

Page 25: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR fits into current sample preparation workflows

EMR-Lipid (p/n 5982-1010)

EMR-Polish (p/n 5982-0102)Extraction Tube

EMR Product offering

March 17

NACRW 2015

25

Page 26: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR Sorbent - What is it?

When “activated” by water… - The materials selective

hydrophobic interactions increase.

- Suspension of nano particles (high surface area).

- Rapidly interacts with straight chain, “lipid-like” functional groups.

Centrifugation preferably used to separate precipitate from solution (not filtration).

EMR-Lipid Mechanism – Size exclusion and hydrophobic interaction.

March 17

NACRW 2015

26

1.0 g EMR in 15 mL tube

Page 27: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

… and what does it do?

EMR sorbent removes LipidsWhat are Lipids?

A class of naturally occurring hydrocarbon containing compounds commonly known as fats and oils

Free Fatty Acids

Triglycerides

Cholesterol Phospholipids

March 17

NACRW 2015

27

Page 28: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

What Does EMR NOT Interact With?

EMR does NOT remove analytes of interestExceptions?

Compounds containing long aliphatic functional groups (e.g. prostaglandins)

Fluoroquinolones Tetracyclines

Vitamin DPAHsOrganochlorine Pesticides

Imidazole pesticides

March 17

NACRW 2015

28

Page 29: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR Fits into Existing Workflows

QuEChERS (Quick, Easy, Cheap, Effective, Rugged, Safe)• Easy-to-use sample preparation for food testing, solid

samples (e.g. vegetables, fruits, meat, seafood, etc.)EMR Applications: - Pesticide Residues in Avocado,

Modified Liquid Extraction (Protein Precipitation)• Proteins are removed by a “crash” step prior to injection

or cleanup (e.g. milk, meat, seafood, etc.)EMR Applications: - PAHs in Salmon,

- Veterinary Drugs in Bovine Liver

March 17

NACRW 2015

29

Page 30: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Pros• Fast and inexpensive• Takes minimal experience• Doesn’t require special equipment• Accommodates multiple matrices• Accommodates large analyte groupsCons• Large amount of coextractives

Improving dSPE in QuEChERS1. Extraction 2. Dispersive SPE

Pros• Same as extractionCons• Minimal cleanup provided• Can remove analytes• Lipids are challenging to remove

selectively

March 17

NACRW 2015

30

Page 31: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR – Lipid – dSPE Cleanup

A B C D

E

A. QuEChERS or Liquid ExtractB. Add H2O to EMR tube (“activation”)C. Transfer extractD. Vortex and centrifugeE. Supernatant (1:1; extract: H2O)

March 17

NACRW 2015

31

Page 32: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR – Polish – ACN/H2O Phase Separation

A B C D

E

A. Add supernatant to EMR – Polish tubeB. Vortex immediatelyC. Phase separation after centrifugeD. Transfer upper layer for analysisE. Final samples split for GC and LC analysisF. Extra dry step recommended prior to GC

March 17

NACRW 2015

32

Page 33: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

EMR Workflow

Extract dSPE Polish Dry

In acetonitrileOr acetone

3/6/2017

33

Page 34: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Fullscan Avocado

Avocado – Untreated

March 17

NACRW 2015

34

Page 35: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Fullscan Avocado

Avocado – UntreatedAvocado – EMR Treated

March 17

NACRW 2015

35

Page 36: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Fullscan Pet Food

Pet Food – Untreated

March 17

NACRW 2015

36

Page 37: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Fullscan Pet Food

Pet Food – UntreatedPet Food – EMR Treated

March 17

NACRW 2015

37

Page 38: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Full Scan– Avocado Oil

Avocado Oil ExtractEMR – Lipid Extract

March 17

NACRW 2015

38

Page 39: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Spinach Fullscan

March 17

NACRW 2015

39

Spinach matrix blank w/ EMR cleanup

Spinach matrix blank w/o cleanup

Spinach matrix blank w/ Q-Pigment dSPE cleanup

Spinach matrix blank w/ Q-Universal dSPE cleanup

5.00 10.00 15.00 20.00 25.00 30.00 35.000

500000

1000000

1500000

2000000

2500000

3000000

3500000

4000000

4500000

5000000

Time-->

Abundance

Page 40: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

GC-MS Red Pepper Fullscan

March 17

NACRW 2015

40

5.00 10.00 15.00 20.00 25.00 30.00 35.000

500000

1000000

1500000

2000000

2500000

3000000

3500000

4000000

4500000

5000000

5500000

Red Pepper matrix blank w/ EMR cleanup

Red Pepper matrix blank w/o cleanup

Red Pepper matrix blank w/ Q-General dSPE cleanup

Red Pepper matrix blank w/ Q-Universal dSPE cleanup

Page 41: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Benefits to Instrumental Flowpath

March 17

NACRW 2015

41

Page 42: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Comparison of Analytes Response Consistency over Multiple Avocado Sample Injections

TPP

Commonly used IS

RT: 18.3 min

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

20000

5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100

TPP

Peak

Are

a

Injections of Avocado Samples on GC-QQQ

TPP Peak Area over 100 Injections

EMR cleanup

Z-Sep+ cleanup

C18 dSPE cleanup

Linear (EMR cleanup)

Linear (Z-Sep+ cleanup)

Linear (C18 dSPE cleanup)

Response drops ~ 50% after 100 avocado sample injections

Response increases ~ 35% after 100 avocado sample injections

Response drops ~ 4% after 100 avocado sample injections

March 17

NACRW 2015

42

Page 43: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Analytes Responses Reproducibility on GC/MS/MS over 100 Injections of Avocado Samples

March 17

NACRW 2015

43

Page 44: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

MS Source Critical Tuning Parameters

0

200

400

600

800

1000

1200

1400

1600

0

10

20

30

40

50

60

70

80

90

100

110

120

130

0 500 1000 1500 2000 2500

EM V

olat

age

Rela

tive

Abun

danc

e fo

r Tun

ing

Ion

(%)

Injections of Avocado Samples with EMR cleanup on GC/MS

MS (GC/MS) Performance Monitoring over Multiple Avocado Sample Injections

Rel Abundance for Ion 219, Minimum of 70 needed

Rel Abundance for Ion 502, minimum of 5 needed

EM voltage, higher than 1400 indicated needs to replace

March 17

NACRW 2015

44

Page 45: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

67 Pesticides Analysis in Avocado by LC and GC-QQQ

March 17

NACRW 2015

45

Page 46: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Accurately weigh 15 g comminuted Avocado sample in 50 mL centrifuge tube

Centrifuge @ 5000 rpm for 5 min

Transfer 5 mL of upper ACN extract and 5 mL of water to EMR dSPE 15 mL tubes

Add 15 mL of 1% Acetic Acid in Acetonitrile, and AOAC QuEChERS extraction kit

Cap and shake vigorously on mechanical shaker for 2 min

Vortex and Centrifuge

QuEChERS-EMR Protocol for Multi-residue Analysis of Pesticides in Avocado

Combine 200 µL of ACN extract and 800 µL water, vortex, and centrifuge if needed

Transfer 5 mL supernatant to empty tube, add content of Polish Pouch

Vortex, centrifuge, and transfer the upper ACN layer to another vial

Samples are ready for GC-QQQ analysis after dry step Samples are ready for LC-QQQ analysis

March 17

NACRW 2015

46

Page 47: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

CleanupAmount of co-

extractives(mg, n= 3)

Amount of co-extractives removed by

cleanup (mg, n = 3)

% of matrix co-extractives removed by

cleanup

No further cleanup 14.7 --

C18/PSA Cleanup 9.5 5.2 35.4

EMR-Lipid Cleanup 4.2 10.5 71.4

Zirconia sorbent Cleanup 7.0 7.7 52.4

% Matrix Co-extractives Removed by Cleanup = 𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝐴𝐴𝑜𝑜 𝐶𝐶𝐴𝐴−𝑒𝑒𝑒𝑒𝐴𝐴𝑒𝑒𝑒𝑒𝑒𝑒𝐴𝐴𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒 𝑅𝑅𝑒𝑒𝐴𝐴𝐴𝐴𝑒𝑒𝑒𝑒𝑅𝑅 𝑒𝑒𝑜𝑜𝐴𝐴𝑒𝑒𝑒𝑒 𝐶𝐶𝐶𝐶𝑒𝑒𝑒𝑒𝐴𝐴𝐴𝐴𝐶𝐶𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝐴𝐴𝑜𝑜 𝐶𝐶𝐴𝐴−𝑒𝑒𝑒𝑒𝐴𝐴𝑒𝑒𝑒𝑒𝑒𝑒𝐴𝐴𝑒𝑒𝑒𝑒𝑒𝑒𝑒𝑒 𝑤𝑤𝑒𝑒𝐴𝐴𝑤𝐴𝐴𝐴𝐴𝐴𝐴 𝐶𝐶𝐶𝐶𝑒𝑒𝑒𝑒𝐴𝐴𝐴𝐴𝐶𝐶

× 100%

The use of EMR material cleanup removes extra 20-30% of Avocado co-extractives in comparison to traditional QuEChERS and/or competitor’s cleanup

Comparison of Avocado Co-extractives by Weight

March 17

NACRW 2015

47

Page 48: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Comparison of GC/MS Full-scan Chromatogram for Matrix Background

The use of EMR material cleanup provides significantly cleanup chromatographic sample background.

March 17

NACRW 2015

48

Page 49: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Comparison of GC/MS/MS MRM Chromatogram for Matrix Background

March 17

NACRW 2015

49

6x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

Counts vs. Acquisition Time (min)3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

6x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

Matrix blank w/ C18/PSA cleanup

Matrix blank w/ zirconia sorbent cleanup

Matrix blank w/ EMR cleanup

Page 50: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Chromatographic Benefits of Matrix Removal Provided by EMR Cleanup

ExampleEPN in Avocado on LC-QQQ

+ MRM (151.0 -> 79.1) AV MBPI-a-CD C2 spike 50ppb-1.D

Acquisition Time (m

12.7 12.8 12.9 13 13.1 13.2 13.3 13.4 13.5 13.6

Coun

ts 4x10

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

2.2

2.4

2.6

13.163 min.

+ MRM (151.0 -> 79.1) AV Z-Sep+ C5 spike 50ppb-1.D

Acquisition Time (m

12.7 12.8 12.9 13 13.1 13.2 13.3 13.4 13.5 13.6

Coun

ts 5x10

-0.1

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

13.163 min.

+ MRM (183.1 -> 168.1) AV MBPI-a-CD C2 spike 50ppb-1.D

Acquisition Time (m

20.4 20.5 20.6 20.7 20.8 20.9 21 21.1 21.2

Coun

ts 4x10

-0.25

0

0.25

0.5

0.75

1

1.25

1.5

1.75

2

2.25

2.5

2.75

3

3.25

3.5

3.75

4

4.25

4.5

4.75

520.788 min.

+ MRM (183.1 -> 168.1) AV Fatty dSPE spike 50ppb-1R.D

Acquisition Time (m

20.4 20.5 20.6 20.7 20.8 20.9 21 21.1 21.2

Coun

ts 4x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

20.833 min.

+ MRM (183.1 -> 168.1) AV Z-Sep+ C5 spike 50ppb-1.D

Acquisition Time (m

20.4 20.5 20.6 20.7 20.8 20.9 21 21.1 21.2

Coun

ts 4x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

20.819 min.

Captan in Avocado on GC-QQQ

Permethrin in Avocado on GC-QQQ

EMR-Lipidcleanup

Zirconia sorbent cleanup

C18/PSAcleanup

Benefit Reduced matrix suppression Improved S/N ratio Less interferences for

accurate integration

March 17

NACRW 2015

50

Page 51: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Selected Problematic Pesticides for Recovery Comparison

March 17

NACRW 2015

51

■ Zirconia cleanup

0

20

40

60

80

100

120

Methamidophos Thiophanatemethyl

Bentazone 2,4-D Acid Dichlorprop Penconazole TPP EPN Captafol

Anal

ytes

Rec

over

y %

(n=6

) in

50pp

b pe

stic

ides

fort

ified

Avo

cado

sam

ples

Selected Pesticides Recovery Comparison from Multi-residue Pesticides Analysis in Avocado by LC/QQQ or GC/QQQ

EMR cleanup C18/PSA cleanup Zirconia sorbent cleanup

Page 52: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Statistical Recovery Results Comparison

March 17

NACRW 2015

Page 52

Page 53: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Method Accuracy and Precision

March 17

NACRW 2015

53

Page 54: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Summary and Conclusion

• EMR-Lipid provides the most complete lipid removal of any sorbent on the market.

• Achieve SPE cleanliness with dSPE simplicity.• EMR is a one size fits all sorbent for a variety of sample types and

analytes.

• Key applications were validated with EMR and demonstrate better recovery, better precision, and decreased matrix impact to the instrument and results.

March 17

NACRW 2015

54

Page 55: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Today’s Agenda

IntroductionAddressing difficult samples1. Polar analytes2. High fat samples3. How to get rid of interfering terpenesSummary and Conclusions

NACRW 2015

1March 17

Page 56: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Constituents of Cannabis and Hop: Complex• Nitrogenous compounds (27 known)

• Amino acids (18),

• Proteins (3)

• Glycoproteins (6)

• Enzymes (2)

• Sugars and related compounds (34)

• Hydrocarbons (50)

• Simple alcohols (7)

• Aldehydes (13)

• Ketones (13)

• Simple acids (21)

• Fatty acids (22)

• Simple esters (12)

• Lactones (1)

• Steroids (11)

• Terpenes (120)

• Non-cannabinoid phenols (25)

• Cannabinoids (66)

• Flavonoids (21)

• Vitamins (1) [Vitamin A]

• Pigments (2)

• Elements (9).

56

Page 57: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

9x10

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

1.65

1.7

1.75

1.81 1

Counts vs. Acquisition Time (min)4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32

QuEChERS AOAC Extraction/Partitioning with Custom dSPE versus Universal dSPE: Hops

57

GC/MSMS MS1 Scan

Universal dSPE

Custom B dSPECustom C dSPECustom D dSPE

Custom F dSPE

Page 58: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

QuEChERS AOAC Extraction/Partitioning with Custom F dSPE versus Universal dSPE: Hops

58

Universal dSPE

Custom F

9x10

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

1.65

1.7

1.75

1.81 1

Counts vs. Acquisition Time (min)4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32

GC/MSMS MS1 Scan

Page 59: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

59

QuEChERS AOAC Extraction/Partitioning with Custom F dSPE versus Universal dSPE: Hops

8x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5

2.6

2.7

2.8

2.9

3 1 1

Counts vs. Acquisition Time (min)4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32

GC/MSMS MS1 Scan

Page 60: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

0

10

20

30

40

50

60

70

80

90

100

110

120

130

140

150

Re

co

ve

ry

Pesticide

QuEChERS AOAC with Custom F dSPE versus Universal dSPE

AOAC with Custom FAOAC with Universal dSPE

60

Pesticide Recovery after QuEChERS AOAC with Custom F dSPE or Universal dSPE: Hops

Daminozide is not extracted into the ACN layer after QuEChERS AOAC salt partitioning Concentration of spiked pesticides: 100 ppb

Page 61: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Conclusion

Addressing difficult samples1. Polar analytes : Explore new polymeric SPE phases2. High fat samples: Consider EMR Lipids as a dSPE3. How to get rid of interfering terpenes: stay tuned

NACRW 2015

1March 17

Page 62: Analytical World …CAN hurt youir4.rutgers.edu/Other/NEC 2017/2017 NEC Agilent sample prep.pdf · Method 528 provides procedures for the determination of phenols in finished drinking

Thank You!