Complex Surgical Instruments ANES 1502 ANESTHESIA TECHNOLOGY FUNDAMENTALS COLLEGE OF DUPAGE
Transcript
Page 1: ANES 1502 - M13 PPT: Complex Surgical Instruments

Complex Surgical Instruments

ANES 1502

ANESTHESIA TECHNOLOGY FUNDAMENTALS

COLLEGE OF DUPAGE

Page 2: ANES 1502 - M13 PPT: Complex Surgical Instruments

Objectives:

As a result of successfully completing this chapter, students will be able to:

Provide an overview of and discuss procedures to care for and effectively process powered surgical instruments.

Explain basic concerns important when handling and processing endoscopic instruments.

Discuss detailed information about flexible endoscopes.

Review general processing and inspection requirements for rigid and semi-rigid endoscopes and laparoscopic instruments.

Identify basic protocols important at each step in the loaner instrumentation process.

Page 3: ANES 1502 - M13 PPT: Complex Surgical Instruments

Powered Surgical Instruments (PSI’s)

Page 4: ANES 1502 - M13 PPT: Complex Surgical Instruments

What We See and Don’t See

Page 5: ANES 1502 - M13 PPT: Complex Surgical Instruments

Powered Surgical Instruments

Look Simple – Are Complex

Look Durable – Are Delicate

Look Plain – Are Expensive

Look Easy to Clean – Are Difficult to Clean

Page 6: ANES 1502 - M13 PPT: Complex Surgical Instruments

Electrically Powered Instruments

Cable attached to a motorized hand piece.

Page 7: ANES 1502 - M13 PPT: Complex Surgical Instruments

Electrical Powered Instruments

Require a cable that can be sterilized. One end attached to the instrument handpiece and the other to a motor or adapter that is connected to a 110 volt outlet.

Page 8: ANES 1502 - M13 PPT: Complex Surgical Instruments

Pneumatic Powered Instruments

Powered by Compressed Gas.

Page 9: ANES 1502 - M13 PPT: Complex Surgical Instruments

Pneumatic Powered Instruments

Powered by compressed gas. They require a hose that can be sterilized. One end attaches to the instrument handpiece and the other to a stand alone cylinder tank or a wall or column mounted regulator panel.

Page 10: ANES 1502 - M13 PPT: Complex Surgical Instruments

Battery Powered Instruments

Least cumbersome because there is no cord or cable. Requires batteries and a battery charger.

Page 11: ANES 1502 - M13 PPT: Complex Surgical Instruments

Battery Powered

Example of a Battery Powered PSI and Two Styles of Batteries

Page 12: ANES 1502 - M13 PPT: Complex Surgical Instruments

PSI Processing Concerns

PSIs are complex instruments that are difficult to clean and easy to damage.

Central Service Technicians should follow manufacturer’s processing instructions carefully.

Page 13: ANES 1502 - M13 PPT: Complex Surgical Instruments

Fluid Invasion

Occurs when water or cleaning solution is allowed to enter PSIs.

Fluid invasion can severely damage a PSI.

Page 14: ANES 1502 - M13 PPT: Complex Surgical Instruments

Decontamination Batteries and cords can protect PSIs from Fluid Invasion

Page 15: ANES 1502 - M13 PPT: Complex Surgical Instruments

Attachments are Complex Instruments Too!

Page 16: ANES 1502 - M13 PPT: Complex Surgical Instruments

PSI Accessories

ChucksKeysBurr GuardsCordsBatteries…and much, much more!

Page 17: ANES 1502 - M13 PPT: Complex Surgical Instruments

Common PSIs

Dermatomes/Dermabraiders – used to harvest skin grafts or reshape skin surfaces

Cebatomes – used to remove cement Sternal Saws – used to split the sternum for open heart surgery Dental Drills – used for teeth and jawbones Micro Drills – used for middle ear bones and to drive very small wires Wire Drivers, Drills and Saws – come in various sizes and are used

to work on small to large bones. Saws - designed to perform specific cutting actions such as

reciprocating or oscillating

Page 18: ANES 1502 - M13 PPT: Complex Surgical Instruments

PSI Instrument Cleaning

PSI’s must be processed manually.

Care MUST be taken to prevent Fluid Invasion.

Page 19: ANES 1502 - M13 PPT: Complex Surgical Instruments

Basic PSI Processing Procedure(You MUST Follow Manufacturer’s Specific Instructions)

Remove all bits, burrs and blades from the unit, disconnect from the power source.

Disassemble and remove debris at point of use, keep the instrument moist.

Separate devices with internal mechanisms from simple devices. Generally, simple components can be soaked.

Attach handpiece to a hose of specialized cleaning attachment to reduce the risk of fluid invasion during cleaning.

Clean the exterior with an approved germicidal detergent.

Pay special attention to recessed areas, moving levers, switches, etc.

Clean recessions and cannulas using a stiff bristle brush.

Page 20: ANES 1502 - M13 PPT: Complex Surgical Instruments

Basic PSI Processing Procedure(You MUST Follow Manufacturer’s Specific Instructions)

Clean, rinse and inspect the hose, cable or battery pack. Look for signs of damage or excessive wear.

Lubricate and test as required by the manufacturer.

Dry all components.

Package and sterilize as recommended by the manufacturer.

Page 21: ANES 1502 - M13 PPT: Complex Surgical Instruments

Endoscopy Instruments

Endoscopes greatly reduce patient trauma when physicians perform minimally invasive surgery.

Page 22: ANES 1502 - M13 PPT: Complex Surgical Instruments

Endoscopeman**with permission from Lighthouse Imaging Corporation

Page 23: ANES 1502 - M13 PPT: Complex Surgical Instruments

Rigid Endoscopes

Provide an optical view and in some cases minimal access through a rigid instrument. The scope is inserted through a sheath inserted via a small incision.

Page 24: ANES 1502 - M13 PPT: Complex Surgical Instruments

Flexible Fiberoptic Endoscope

Provides optical access using a flexible, maneuverable scope most commonly inserted through a mucous membrane opening.

Page 25: ANES 1502 - M13 PPT: Complex Surgical Instruments

Regulations and Guidelines

Several regulatory agencies and professional associations provide input into proper scope handling and processing practices.

For a list, please refer to pages 207-209 in the text.

Page 26: ANES 1502 - M13 PPT: Complex Surgical Instruments

Infection Control Concerns

Flexible endoscopes pose significant infection control challenges. Their configuration makes them difficult to clean and dry and they can harbor bacteria.

Page 27: ANES 1502 - M13 PPT: Complex Surgical Instruments

Infection Control Issues

Inadequate training of employees who clean and handle scopes.

Adequate time for thorough processing.

Commitment of employees (and policy writers) to safely reprocess scopes.

Failure to adequate inspect scopes prior to processing.

Failure to follow manufacturer’s instructions for reprocessing.

Page 28: ANES 1502 - M13 PPT: Complex Surgical Instruments

Infection Control Issues

Failure to follow label direction on processing chemicals.

Using the scope without a leak test.

Poor manual cleaning habits.

Failure to use automatic endoscope reprocessors in accordance with manufacturer’s instructions.

Improper drying and storing procedures.

Page 29: ANES 1502 - M13 PPT: Complex Surgical Instruments

Additional Concerns:

Failure to leak test correctly.

Failure to manually clean all channels.

Failure to flush all channels with disinfectant solution.

Failure to fully immerse.

Failure to adequately time the length of disinfectant contact.

Use of disinfectant solutions after their expiration date.

Page 30: ANES 1502 - M13 PPT: Complex Surgical Instruments

Additional Concerns:

Failure to process all scopes in the same manner (Standard Precautions).

Failure to sterilize biopsy forceps.

Inaccessible manufacturer’s instructions.

Variations in staff training.

Improper reprocessing of reusable cleaning supplies.

Inadequately trained personnel.

Page 31: ANES 1502 - M13 PPT: Complex Surgical Instruments

Additional Concerns:

Lack of competence reviews for scope processing and procedures.

Improper storage and transport.

Pressure from physicians to process scopes more quickly so they can perform more procedures on more patients.

Difficulties processing these complex instruments.

Space constraints.

Page 32: ANES 1502 - M13 PPT: Complex Surgical Instruments

Additional Concerns:

Absence of, or an inadequate quality control program.

Poor water quality.

Facility processing equipment.

In order for endoscope processing to be successful, each of these concerns must be addressed.

Page 33: ANES 1502 - M13 PPT: Complex Surgical Instruments

Liquid Chemical Processing Systems

Page 34: ANES 1502 - M13 PPT: Complex Surgical Instruments

Glutaraldehyde

High Level Disinfectant

Can be used manually or in automatic processors.

Must be tested.

Takes Time.

Employee Safety Concerns (ventilation a must!)

Page 35: ANES 1502 - M13 PPT: Complex Surgical Instruments

Ortho-Phthaladehyde (OPA)

High Level Disinfectant

May be used manually or in an automatic processor.

Must be tested.

Page 36: ANES 1502 - M13 PPT: Complex Surgical Instruments

Automatic Endoscope Reprocessors(AERs)

Automated equipment designed to clean, disinfect, and rinse flexible endoscopes.

Page 37: ANES 1502 - M13 PPT: Complex Surgical Instruments

Advantages of AERs

Process consistency Reduced staff exposure to chemicals Timed cleaning Consistent exposure to the cleaning

agent Timed contact with liquid

disinfectants An air flush cycle to remove excess

moisture Use of copious and consistent

amounts of rinse water

Page 38: ANES 1502 - M13 PPT: Complex Surgical Instruments

When using AERs:

Follow manufacturer’s instructions to connect the scope to the AER

Place removable parts in the AER if possible

Attach channel cleaning connectors to all channels

Follow manufacturer’s instructions for using disinfectants

Set the machine for the recommended time

Page 39: ANES 1502 - M13 PPT: Complex Surgical Instruments

Staff Education

All staff must be thoroughly educated in the proper cleaning, processing, and handling of endoscopes.

Page 215 in the text contains a sample competency checklist for endoscope training.

Page 40: ANES 1502 - M13 PPT: Complex Surgical Instruments

A Closer Look at Endoscopes

Although they look similar, endoscopes vary greatly in configuration.

Some scopes are only used to visualize and do not have channels, while others have internal channels.

Page 41: ANES 1502 - M13 PPT: Complex Surgical Instruments

Endoscope Channels

Instrument Channel

OpticalChannel

Air/WaterChannel*Suction

Channel

*Some Flexible Scopes have separate Air and Water Channels

Page 42: ANES 1502 - M13 PPT: Complex Surgical Instruments

Distal Tip

Page 43: ANES 1502 - M13 PPT: Complex Surgical Instruments

Flexible FiberopticEndoscope

Video Monitor

Water Bottle Light Source

Suction Canister

Page 44: ANES 1502 - M13 PPT: Complex Surgical Instruments

Common FFE’s

Page 45: ANES 1502 - M13 PPT: Complex Surgical Instruments

Bronchoscope

Used for the direct visualization of the tracheobronchial tree.

Page 46: ANES 1502 - M13 PPT: Complex Surgical Instruments

CystoscopeUreteroscope

Used to visualize the urethra and bladder (cystoscope) and to look for obstructions such as kidney stones (ureteroscope).

Page 47: ANES 1502 - M13 PPT: Complex Surgical Instruments

Gastroscope Esophagoscope

Used for the visual inspection of the upper GI tract (gastroscope) and for the direct visualization of the esophagus and the cardia of the stomach (esophagoscope).

Page 48: ANES 1502 - M13 PPT: Complex Surgical Instruments

Colonoscope Sigmoidoscope

Used for the visual inspection of the entire large intestine (colonoscope) and the visual inspection of the lower part of the large intestine (sigmoidoscope).

Page 49: ANES 1502 - M13 PPT: Complex Surgical Instruments

Care & Handling

Page 50: ANES 1502 - M13 PPT: Complex Surgical Instruments

Steps in Endoscope Processing

1. Preclean

2. Leak Test

3. Clean

4. High-Level Disinfect or Sterilize

5. Dry

6. Store

Page 51: ANES 1502 - M13 PPT: Complex Surgical Instruments

Fluid Invasion

Fluid invasion can cause damage to flexible fiberoptic endoscopes.

Page 52: ANES 1502 - M13 PPT: Complex Surgical Instruments

Leak Testing

Detects leaks that can compromise the safety of the scope.

Scopes that fail a leak test must be removed from service and repaired.

Page 53: ANES 1502 - M13 PPT: Complex Surgical Instruments

Storage

Scopes must be dry when stored.

Do not kink or bend.

Page 54: ANES 1502 - M13 PPT: Complex Surgical Instruments

Endoscope Accessories

Diagnostic Accessories

Therapeutic Accessories

Follow manufacturer’s instructions for the proper processing of each type of accessory.

Page 55: ANES 1502 - M13 PPT: Complex Surgical Instruments

Rigid and Semi-Rigid Endoscopes

Page 56: ANES 1502 - M13 PPT: Complex Surgical Instruments

An Important Note:

The term “Rigid” is misleading! Rigid Endoscopes are VERY DELICATE and can be easily damaged.

Page 57: ANES 1502 - M13 PPT: Complex Surgical Instruments

Rigid Endoscope Guidelines

Always follow manufacturers’ recommendations.

Never soak instruments for cleaning or high level disinfection in a metal soak pan.

Instruments may need to be soaked in a vertical position.

Utilize flush ports when available to circulate (under pressure) enzymatic detergent through the channels.

Lenses must never be placed in an ultrasonic cleaner.

Page 58: ANES 1502 - M13 PPT: Complex Surgical Instruments

Rigid Scope Guidelines - continued

Gently brush clean the exterior and accessible lumens with the appropriate brush.

Thoroughly rinse with distilled water and rough dry.

Inspect instruments for cleanliness, missing parts, and damage.

Thoroughly air-dry instruments.

Remember that these instruments are fragile and must be handled carefully.

Page 59: ANES 1502 - M13 PPT: Complex Surgical Instruments

Clean carefully and check for missing parts.

Page 60: ANES 1502 - M13 PPT: Complex Surgical Instruments

Inspection

Check surfaces of the telescope for visible damage

Inspect for clarity using white paper with writing on it

Check the eyepiece seal for visible damage

Page 61: ANES 1502 - M13 PPT: Complex Surgical Instruments

Laparoscopic Instruments

Minimally Invasive Instrumentation

Page 62: ANES 1502 - M13 PPT: Complex Surgical Instruments

Inspect Insulation for Damage or Wear

Page 63: ANES 1502 - M13 PPT: Complex Surgical Instruments

Insulation Testers are available

Page 64: ANES 1502 - M13 PPT: Complex Surgical Instruments

Loaner Instrumentation

Owned by the vendor and brought in for a specific case.

Can have a significant impact on Central Service workload.

Page 65: ANES 1502 - M13 PPT: Complex Surgical Instruments

Loaner Instrument Processes

Log Receipt of Loaner Instruments and Implants. Include the following information: Date Time Signature of delivery person Initials of receiving person Doctor’s name Patient’s last name Number of trays Inventory check of tray(s) for completeness Inventory check of tray(s) for damaged items.

Page 66: ANES 1502 - M13 PPT: Complex Surgical Instruments

Loaner Instrument Processes

ALL Loaner instruments must be decontaminated by the receiving facility before use.

Page 67: ANES 1502 - M13 PPT: Complex Surgical Instruments

Loaner Instrument Processes

After cleaning and decontamination, inspect and assemble instruments for sterilization. Defective instruments should be documented.

Sterilize according to manufacturer’s instructions.

Page 68: ANES 1502 - M13 PPT: Complex Surgical Instruments

Loaner Instrument Processes

After sterilization, place in a low traffic area and handle as little as possible before use.

Page 69: ANES 1502 - M13 PPT: Complex Surgical Instruments

As technology advances, instruments (and the Central Service Technician’s job) become more complex.