APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
1-ANTITRYPSIN SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME AAT PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 20.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 20.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 6.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 160 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test AAT
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information AAT/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
1- ACID GLYCOPROTEIN SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME AGP PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test AGP
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information AGP/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STL-001 Protein Standard Low, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
COMPLEMENT C3 SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME C3C PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test C3C
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information C3C/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
CERULOPLASMIN SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME CER PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 15.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test CER
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information CER/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
HAPTOGLOBIN SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME HAP PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 6 VOLUME 8.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 7 Conversion Factor 1.000000 VOLUME 15.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test HAP
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information HAP/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 17.10.2008
APPLICATION NOTE PENTRA 400
IgA KIT 2
nd GENERATION
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgA2 PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 20.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 20.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 12.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgA2
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information 104C002 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 06.04.2009
APPLICATION NOTE PENTRA 400
IgG KIT 2nd
Generation
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgG PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 20.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 20.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 7.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 165 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 15.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgG
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information 105C003 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High , 1 mL MPS/STS-5X1 Protein Standard Set, 5 x 1 mL
139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 17.10.2008
APPLICATION NOTE PENTRA 400
IgM KIT 2
nd GENERATION
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgM2 PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 15.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 15.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 30.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgM2
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information 106C002 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
PREALBUMIN SD (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME PAL PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10.0 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 15.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 150 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 15.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test PAL
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information PAL/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
TRF SD (SEL-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME TRF PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 10.0 POST DILUTION 5.0 Factor 2 2.0 POST CONCENTRATION 2.0 Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name Saline Factor 6 0.0 Factor 10 Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 4.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 250 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 10.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test TRF
Main Standard ** ** Standard value multiplied by dilution factor
3. Ordering information TRF/SEL-000 1 x 5 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein standard high, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23.04.2007 APPROVED APPLICATION
APPLICATION NOTE PENTRA 400
1-ANTITRYPSIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use. Reagent 1: Buffer, ready for use. Reagent 2: Antiserum, ready for use. Calibrator: Ready for use.
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME AAT PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 1.0 POST DILUTION 2.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 0.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 350 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 70.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test AAT
Main Standard ** ** Standard value
3. Ordering information AAT/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23.04.2007 APPROVED APPLICATION
APPLICATION NOTE PENTRA 400
1- ACID GLYCOPROTEIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use. Reagent 1: Buffer, ready for use. Reagent 2: Antiserum, ready for use. Calibrator: Ready for use.
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME AGP PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 0.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) - Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 290 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 40.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test AGP
Main Standard ** ** Standard value
3. Ordering information AGP/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
ASL(O) N-DIL (AUT-KIT) LATEX
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Latex, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME ASL PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 3.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 6 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 140 µL Unit IU/mL REAGENT R2 CYCLE 7 Conversion Factor 1.000000 VOLUME 30.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 600 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test ASL
Main Standard ** ** Standard value
3. Ordering information ASL/AUT-000 1 x 10 mL Latex 5 x 25 mL Buffer ASL/STH-001 ASL Standard High, 1 mL ASL/CON-001 ASL Control, 1mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
COMPLEMENT C3 N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME C3C PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION 2.0 Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 200 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 30.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test C3C
Main Standard ** ** Standard value
3. Ordering information C3C/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
COMPLEMENT C4 N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME C4C PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 3.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 165 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test C4C
Main Standard ** ** Standard value
3. Ordering information C4C/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
CERULOPLASMIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME CER PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 200 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 30.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test CER
Main Standard ** ** Standard value
3. Ordering information CER/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
CRP N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME CRP PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 15 VOLUME 14.0 µL Last Reading Cycle 50 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 200 µL Unit mg/dL REAGENT R2 CYCLE 16 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test CRP
Main Standard ** ** Standard value
3. Ordering information CRP/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer CRP/STH-001 CRP Standard High, 1 mL CRP/COL-001 CRP Control Low, 1 mL CRP/COH-001 CRP Control High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
HAPTOGLOBIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME HAP PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 8 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 280 µL Unit mg/dL REAGENT R2 CYCLE 9 Conversion Factor 1.000000 VOLUME 40.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test HAP
Main Standard ** ** Standard value
3. Ordering information HAP/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 17.10.2008
APPLICATION NOTE PENTRA 400
IgA KIT 2
nd GENERATION
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgA2 PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 1.0 POST DILUTION 5.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 0.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 350 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 60.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgA2
Main Standard ** ** Standard value
3. Ordering information 104C002 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
IgG N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgG PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 5.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 330 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 75.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgG
Main Standard ** ** Standard value
3. Ordering information IGG/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 17.10.2008
APPLICATION NOTE PENTRA 400
IgM KIT 2
nd GENERATION
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME IgM2 PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main Indirect NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 1.0 POST DILUTION 5.0 Factor 2 2.0 POST CONCENTRATION - Factor 3 2.0 Factor 4 2.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 0.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 330 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 40.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test IgM
Main Standard ** ** Standard value
3. Ordering information 106C002 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
MAL N-DIL (AUT-KIT) 2
nd generation
1. Reagent preparation Sample: Centrifuged urine, ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME MAL 2nd PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 3.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 12.0 µL Last Reading Cycle 30 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 170 µL Unit mg/L REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 30.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test MAL 2nd
Main Standard ** ** Standard value
3. Ordering information 102C002 1 x 10 mL Antiserum 5 x 25 mL Buffer MAL/STD-001 Microalbumin Standard, 1 mL MAL/CON-001 Microalbumin Control, 1 mL MAL/CON-005 Microalbumin Control, 5 mL
APTEC DIAGNOSTICS 23-04-2007
APPLICATION NOTE PENTRA 400
PREALBUMIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: Ready for use
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME PAL PRE-DILUTION SAMPLE TYPE Serum/Plasma Type Main direct NUMBER OF REAGENTS Reagent 2 Calibrator Diluent Saline Factor 1 150.0 POST DILUTION 2.0 Factor 2 16.0 POST CONCENTRATION - Factor 3 8.0 Factor 4 4.0 PRE-DILUTION Factor 5 2.0 Diluent Name - Factor 6 1.0 Factor - Factor 7 - Incubation time (in cycles) 1 Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 180 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test PAL
Main Standard ** ** Standard value
3. Ordering information PAL/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer MPS/STH-001 Protein Standard High, 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL
APTEC DIAGNOSTICS 23.04.2007 APPROVED APPLICATION
APPLICATION NOTE PENTRA 400
TRANSFERRIN N-DIL (AUT-KIT)
1. Reagent preparation Sample: Ready for use Reagent 1: Buffer, ready for use Reagent 2: Antiserum, ready for use Calibrator: TRF standard set. Use saline 9g/L as zero point.
2. Instrument setting
GENERAL PARAMETERS CALIBRATION PARAMETERS
TEST NAME TRF PRE-DILUTION SAMPLE TYPE Serum/Plasma Type - NUMBER OF REAGENTS Reagent 2 Calibrator Diluent - Factor 1 - POST DILUTION - Factor 2 - POST CONCENTRATION - Factor 3 - Factor 4 - PRE-DILUTION Factor 5 - Diluent Name - Factor 6 - Factor - Factor 7 - Incubation time (in cycles) - Factor 8 -
CORRELATION CALIBRATION
Slope 1.00000 Calibration Mode Logit/Log5
Intercept 0.00000 Level 6 Calibration Factor - Run (s) 2 VALIDITY On Request
ANALYSIS PARAMETERS CALCULATION PARAMETERS
ANALYSIS SEQUENCE CALCULATION TYPE End Point SAMPLE CYCLE 1 First Reading Cycle 9 VOLUME 2.0 µL Last Reading Cycle 40 DILUENT 1 NAME H2O VOLUME 10.0 µL
REAGENT R1 CYCLE 1 UNITS PARAMETERS
VOLUME 220 µL Unit mg/dL REAGENT R2 CYCLE 10 Conversion Factor 1.000000 VOLUME 20.0 µL DILUENT R2 NAME H2O VOLUME 10.0 µL WAVELENGTH Primary Wavelength 340 Secondary Wavelength - MIXING SPEED 40
CALIBRATION
TARGET VALUE ENTRY
Test TRF
Main Standard ** ** Standard value
3. Ordering information TRF/AUT-000 1 x 10 mL Antiserum 5 x 25 mL Buffer TRF/STS-5X1 Transferrin Standard Set, 5 x 1 mL 139F003 Immunology Control Low, 1mL 139F002 Immunology Control High, 1mL