Application of a beryllium specific resin at the Y-12 National Security Complex. Darrin K. Mann, D.H. “Bo” Bowman, Thomas J. Oatts, and Vicki F. Belt Analytical Chemistry Organization Y-12 National Security Complex P.O. Box 2009, MS 8189 Oak Ridge, TN 37831-8189. YGG-01-0419. Outline. - PowerPoint PPT Presentation
Popular Tags:
22
Application of a beryllium specific resin at the Y-12 National Security Complex YGG-01-0419 Darrin K. Mann, D.H. “Bo” Bowman, Thomas J. Oatts, and Vicki F. Belt Analytical Chemistry Organization Y-12 National Security Complex P.O. Box 2009, MS 8189 Oak Ridge, TN 37831-8189
Transcript
Page 1: Application of a beryllium specific resin at the Y-12 National Security Complex

Application of a beryllium specific resin at the Y-12 National Security Complex

YGG-01-0419

Darrin K. Mann, D.H. “Bo” Bowman, Thomas J. Oatts, and Vicki F. Belt

Analytical Chemistry OrganizationY-12 National Security Complex

P.O. Box 2009, MS 8189Oak Ridge, TN 37831-8189

Page 2: Application of a beryllium specific resin at the Y-12 National Security Complex

Outline Short review of Be and Problems associated with use

Overview of Be program at Y-12 National Security Complex

Current Method– Problems with current method– Evolution of new resin– New method

Data from new method– Problems associated with new method– Scandium Problem– Post-Filter solution??

Conclusions/Future Work

Page 3: Application of a beryllium specific resin at the Y-12 National Security Complex

Be: Why we use it

Discovered in 1798

– Not widely used in Industry until 1940s and 50s

Lighter then Aluminum, Stiffer then Steel

– 2nd lightest metal

– 6 times stiffer then steel

High heat absorption

– One pound absorbs as much heat as 6 pounds of copper

Be Metals, Alloys, Salts and Oxides are used for a wide variety of Industries

– Structures in high-speed aircraft (space shuttle)

– Satellite mirrors and space telescopes

– Golf clubs and bicycle frames

– Neutron moderators or reflectors in nuclear reactors

Page 4: Application of a beryllium specific resin at the Y-12 National Security Complex

Problems Associated with Be

Physical Problems– Expensive

– Brittleness

• Increases toxicity

Health Hazard– Most Significant disadvantage for industrial use

– Causes Chronic Beryllium Disease (CBD)

• No known cure, can only be treated

– Produces scaring of lung tissue

• Chronic, may take years to develop

– Average latency period is 10-15 years

• 2-5 % of population Be sensitive

• Over 100 current and former DOE employees have CBD

Page 5: Application of a beryllium specific resin at the Y-12 National Security Complex

Be program at Y-12 National Security Complex

Controlled by US Dept. of Energy’s Chronic Beryllium Disease Prevention Program

– 10 CFR Part 850

– Promulgated in 1999 to protect DOE workers from CBD

– Requires Be surface and air monitoring to determine health risk

– Rule greatly increased the need for Be analysis in the DOE complex

– Current analytical methods include ICPOES and GFAA

Page 6: Application of a beryllium specific resin at the Y-12 National Security Complex

Overview of Y-12 Be Program

0

1000

2000

3000

4000

5000

6000

7000

Oct Feb Jun

FY04FY05

Over 50K Samples Analyzed in 04.

Average 53 +/- 79 samples/day

Average 2625 +/- 876 samples/month

Average turnaround time is 24 hours.

Page 7: Application of a beryllium specific resin at the Y-12 National Security Complex

Typical Workloads Associated with Be Program

FY 2005 Beryllium Filters Analyzed

3680

18511761

2444

3389

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

7500

Oct

Nov

Dec

Jan

Feb

Mar

Apr

May

Jun

Jul

Aug

Sep

Month

# of sa

mple

s

FY 05 Monthly Be samples FY 03 Monthly Be samples FY 01 Monthly Be samples FY 00 Monthly Be samples

FY 99 Monthly Be samples FY 02 Monthly Be samples FY 04 Monthly Be samples

Page 8: Application of a beryllium specific resin at the Y-12 National Security Complex

Breakdown of Be Smears at Y-12

Routine Large area wipes from special projects/areas (LAW5D) 5 day deadline

0.0%Rush Large area wipes from special

projects/areas (LAW24) 24 hr deadline11.6%

General housekeeping smears (HSKEEP) 3 day deadline3.7%

Movement of items from area to area (PMOVE) 24 hr deadline

0.7%

Non-routine smears (NRS) 5 day deadline16.1%

Be Site characterization smears (Be-Site) TBD deadlines

6.9%

Operations needed special attention (SPECIAL) 24 hr deadline

0.0%

Downposting of area smears, cleanup work (DNPOST) 24 hr deadline

0.2%

Respirator smears (RESP) 5 day deadline2.4%

Permanent Air filters (PA) 10 day deadline1.1%

QA Samples (daily)0.4%

Routine smears (RS) 10 day deadline11.6%

Non-routine next day smears (NRS-ND) 24 hr deadline34.2%

Others0.0%

Breathing Zone (BZ) 24 hr deadline11.2%

Page 9: Application of a beryllium specific resin at the Y-12 National Security Complex

Basic Flow Chart for Analysisof Be Smear Program

Sample group picked up by ACO sampling personnel (1500-

1600 hrs)

ACO Receiving personnel receive samples into ESLIMS

ACO Receiving personnel notify Be Lab of received

samples

Be Lab personnel accept chain of custody of samples from ACO

Receiving

Set samples aside and notify IH technician of

problems with samples

Group samples by type

Be Lab checks COC and samples for accuracy and

completeness

Batch samples into QA files based upon priority

Prepare all necessary paperwork for sample

preparation and analysis for

Be Lab personnel obtain QA file and

label test tubes with

Fold and stuff samples into test tubes

Add H2SO4 and H2O2 to digest filter and BeO

Place samples into microwave oven and heat

for 5 minutes

Remove samples and cool

Add H2O2 to each tube and heat in microwave for additional 10 minutes

Add 10% HNO3 to nominal 10 mL volume

Add 0.1 mL Sc internal standard

Standardize ICP instruments daily

Create autosampler file with sample and QC

ID's

Load autosampler with samples

Analyze samples

Be conc < linear range?

Dilute sample within linear range and

rerun

Store Be result

Import data to database and create analysis

documentation

Place batched QA files and samples

into Be Buffer Area

Upload data into ESLIMS

Review and approve data in ESLIMS

File datafileNO

NO

Yes

Yes

End of operations for the day

ICP interferents monitored: Cr (Cr267.7), Fe (273.9), Mo (Mo 204.5), Nb (Nb 316.3), Th (Th 401.9), Ti (Ti 337.2), U (U 367.0), V (V 292.4), Zr (Zr 339.1

We are using both the 313.042 nm and 313.107 nm lines to quantitate. Interferences checked on each

Page 10: Application of a beryllium specific resin at the Y-12 National Security Complex

The Problem? The Internal Standard (IS) works great correcting interferences to a

point:– High Concentrations of Interfering Elements

• Some elements are very spectral rich– Uranium

» Shift depends on enrichment – Some elements overlap spectrally

• Vanadium, Cesium and Zirconium are examples• Dilution not useful for these elements

Possible Solutions– Dilution

• May lose Be signal• Increase in MDL

– Run samples by ICPMS• Expensive (relative to an OES)• Not as rugged as OES (can’t handle 500 samples/day)

– Remove/Concentrate Be

Page 11: Application of a beryllium specific resin at the Y-12 National Security Complex

Method using Eichrom Be resin

Elegantly simplistic

Usually use 5 ml of sample left over from ICPOES analysis

Adjust sample pH to between 1-2 with 4 M Sodium Acetate

2% Crystal Violet used as indicator (3-4 drops)

Load sample onto Be cartridge (usually 10 mL) and pass through at 2 ml/min

Rinse cartridge with 10 ml of 0.2 M HNO3 at 2 ml/min

Elute Be with 10 ml of 4M HNO3 at < 1 ml/min

Sample can be re-run within a few hours.

Page 12: Application of a beryllium specific resin at the Y-12 National Security Complex

Vacuum System with Be cartridges and 10 ml Reservoirs

Page 13: Application of a beryllium specific resin at the Y-12 National Security Complex

How clean are the Samples?

Samples after digestion

Samples after Be Column

Page 14: Application of a beryllium specific resin at the Y-12 National Security Complex

Comparison of Be data using Be Resin and ICPMS Analysis

Sample Type Be Conc. ICP Total Be Original ICP Result ICP-MS Be Diff ICPMS Diff ICPMS Rec. ICPMS Adj ICP Rec** Removal of U(3130 line) (Column) ug (No Column) ug (No Column) ug Conc (ug) %(Percent) %(Percent) % (Percent) % (Percent)

Sample 1 0.00148 0.0296 -0.003 0.058 0.0284 48.97% 51.03% 69.72% 99.96%Sample 2 0.00198 0.0396 -0.00803 0.064 0.0244 38.13% 61.88% 84.53% 99.98%Sample 3 0.00824 0.165 0.01619 0.214 0.049 22.90% 77.10% 105.33% 99.98%Sample 4 0.00082 0.00164 -0.00234 0.025 0.02336 93.44% 6.56% 8.96% 100.00%Sample 5 0.00522 0.1044 -0.00222 0.138 0.0336 24.35% 75.65% 103.35% 99.99%Sample 6 0.00769 0.1538 0.01155 0.186 0.0322 17.31% 82.69% 112.96% 99.98%

** Adjusted recovery using ratio of 10 ppm Be std. recovery and 4 M HNO3 (7.32)

Page 15: Application of a beryllium specific resin at the Y-12 National Security Complex

Original ColumnBe9(LR) 1.42 1.27Rb85(LR) 21.01 0.43Sr88(LR) 259.61 98.86Y89(LR) 56.22 0.59Zr90(LR) 1364.75 6.24Nb93(LR) 13.46 0.29Mo95(LR) 710.93 2.13Ru99(LR) 1.16 0.29Rh103(LR) 0.92 0.89Pd105(LR) 9.17 1.36Ag107(LR) 1037.85 13.62Cd114(LR) 227.05 244.87In115(LR) 32.44 0.51Sn118(LR) 2321.91 -0.82Sn119(LR) 2253.31 0.45Sb121(LR) 86.66 13.01Cs133(LR) 1.53 0.68Ba137(LR) 140.47 287.90La139(LR) 53.98 0.82Ce140(LR) 38.28 0.59

Original ColumnNd146(LR) 11.42 0.22Sm149(LR) 5.76 0.14Eu153(LR) 1.08 0.46Gd155(LR) 30.62 1.53Tb159(LR) 0.61 0.72Dy163(LR) 1.55 0.02Er166(LR) 39.37 0.36Tm169(LR) 0.35 0.61Yb172(LR) 0.87 0.44Lu175(LR) 0.40 0.72Hf178(LR) 22.99 0.71Ta181(LR) 1.68 0.29W182(LR) 263.55 1.64Os189(LR) 0.46 0.86Ir191(LR) 0.05 0.10Ir193(LR) 0.25 0.11Pt194(LR) 1.55 0.45Pt195(LR) 1.18 0.66Au197(LR) -0.37 -3.94Hg200(LR) 267.41 10.60

Original ColumnTl203(LR) 0.75 0.34Tl205(LR) 0.83 0.26Pb206(LR) 1412.97 1562.90Pb208(LR) 1413.20 1583.78Bi209(LR) 260.22 2.31Th232(LR) 3.52 0.39U238(LR) 81.63 0.58Al27(MR) 2654.82 1720.80Ti47(MR) 57.17 7.01V51(MR) 2.68 1.43Cr52(MR) 59.34 30.56Mn55(MR) 37.50 942.14Fe56(MR) 2614.15 -222.28Co59(MR) 7.58 1.52Ga69(MR) 0.23 -0.13Ga71(MR) 0.44 0.45K39(HR) 536.48 103.76Ge72(HR) 0.93 2.20As75(HR) 4.21 -0.43Se77(HR) 1.38 7.72

Multi-element analysis of Be resinBefore and After Column

Units of ppb

89.4% Recovery of Be using the Resin

Page 16: Application of a beryllium specific resin at the Y-12 National Security Complex

Be recovery Conc. Before and After Be cartridgeICP Analysis Represents “Normal” Analysis

0

5

10

15

20

25

30

35

40

Con

cen

trat

ion

(p

pb

)

Sw

ipe

1

Sw

ipe

2

Sw

ipe

3

Sw

ipe

4

Sw

ipe

5

Sw

ipe

6

Sw

ipe

7

Sw

ipe

8

Sw

ipe

9

Be9(LR) B

Be9(LR) A

Be (ICP)

Page 17: Application of a beryllium specific resin at the Y-12 National Security Complex

Percent Recovery of Be vs. Initial ConcentrationICPMS and ICPOES Results

0

10

20

30

40

50

60

70

80

90

100

% R

ecov

ered

Sw

ipe

1

Sw

ipe

2

Sw

ipe

3

Sw

ipe

4

Sw

ipe

5

Sw

ipe

6

Sw

ipe

7

Sw

ipe

8

Sw

ipe

9

Be9(LR) ICPMSBe (ICP)

Page 18: Application of a beryllium specific resin at the Y-12 National Security Complex

So Where is the Be Going?

ICPMS data appears acceptable, so Be is there.

Recovery of Be in ICV and CCV is fine

Colored samples may cause pH problem

– Yellow Samples are a particular problem

Answer may lie in looking at IS Scandium

– Signal is being suppressed

– RSD is very poor (5-10%)

Page 19: Application of a beryllium specific resin at the Y-12 National Security Complex

Recovery of IS Scandium2 mg/L standard added

0

20

40

60

80

100

120IC

V

ICB

RB

LC

S S1

S2

S3

S4

S5

S6

CC

V

CC

B

Be Samples

Sc

Rec

over

y (%

)

Page 20: Application of a beryllium specific resin at the Y-12 National Security Complex

Why is IS Signal being Suppressed?

Doesn’t seem to correct in same ratio as Be

Very large suppression at times, up to 75% of signal.

Only source of suppression can come from column

Doesn’t seem to be acid concentration based

An organic from the resin may be complexing with Sc and pulling it out of solution

Need to remove organic after it passes through initial column.

Page 21: Application of a beryllium specific resin at the Y-12 National Security Complex

Recovery of Sc IS using Post Organic Resin Cartridge

2 ug/L Sc added

0

20

40

60

80

100

120S

c R

ecov

ery

(%)

ICV

ICB P1

P2

P3

P4

P5

NP

1

NP

2

NP

3

NP

4

0.2M

RB

4MR

B

Be Samples

Page 22: Application of a beryllium specific resin at the Y-12 National Security Complex

Conclusion/Future Work

Eichrom’s Be resin seems to be a fast and reliable method to remove spectral interferences from samples when analyzing for Beryllium

Need to better understand the effects of residual organics on the recovery of Internal Standards such as Sc.

– A post organic filter seems to solve the problem

Filter could be used to Lower MDL– ACGIH has recently issued a Notice of Intended Change (NIC) to lower

the Threshold Limit Value (TLV) for airborne beryllium to 0.02 micrograms per cubic meter, or one-tenth of the current DOE action level.

– Concentrate sample onto column, elate with smaller volume

Need to quicken the process– Reduce number of sample