+ All Categories
Home > Documents > Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005...

Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005...

Date post: 06-Aug-2020
Category:
Upload: others
View: 4 times
Download: 1 times
Share this document with a friend
32
Перевод с русского языка на английский Translated from Russian into English <Seal of Biopromin Limited Liability Company> Approved by Director of BIOPROMIN LTD <Signature> Sydora Volodymyr L. Report on clinical studies of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA” Technical Specifications of Ukraine ТУ У 33.1-22716816-001:2006 Code of Standard Industrial Classification of Economic Activities (HS code) 9018 19 90 00 manufactured by “Biopromin” Ltd. Title of the clinical studies: Monitoring of the device ANESA and software USPIH functioning. Improving of working protocol.Clinical study was carried out according to National Standards of Ukraine DSTU 4659-2:2006 (ISO 14155-2:2003) Carried out in cooperation with: The Institute of neurology, Psychiatry and Narcology of AMS of Ukraine (Kharkiv, Ukraine) Regional Clinical Psychiatry Hospital No.3, Section 8 (Kharkiv, Ukraine) “Alfa Labservice” LTD (Kharkiv, Ukraine) Description of model: Automatic Noninvasive Express Screening Analyzer ANESA” Technical Specifications of Ukraine ТУ У 33.1-22716816-001:2006 HS code 9018 19 90 00 Serial number of medical device: 38001515 Software version: USPIH 10.0 Kharkiv, 21-05-2013
Transcript
Page 1: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Перевод с русского языка на английский Translated from Russian into English

<Seal of Biopromin Limited Liability Company>

Approved by

Director of BIOPROMIN LTD

<Signature> Sydora Volodymyr L.

Report on clinical studies

of the medical device

“Automatic Noninvasive Express Screening Analyzer ANESA”

Technical Specifications of Ukraine ТУ У 33.1-22716816-001:2006

Code of Standard Industrial Classification of Economic Activities (HS code) 9018 19 90 00

manufactured by “Biopromin” Ltd.

Title of the clinical studies:

“Monitoring of the device ANESA and software USPIH functioning. Improving of working

protocol.”

Clinical study was carried out according to

National Standards of Ukraine DSTU 4659-2:2006 (ISO 14155-2:2003)

Carried out in cooperation with:

The Institute of neurology, Psychiatry and Narcology of AMS of Ukraine (Kharkiv, Ukraine)

Regional Clinical Psychiatry Hospital No.3, Section 8 (Kharkiv, Ukraine)

“Alfa Labservice” LTD (Kharkiv, Ukraine)

Description of model:

“Automatic Noninvasive Express Screening Analyzer ANESA”

Technical Specifications of Ukraine ТУ У 33.1-22716816-001:2006

HS code 9018 19 90 00

Serial number of medical device: 38001515

Software version: USPIH 10.0

Kharkiv, 21-05-2013

Page 2: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

LIST OF EXECUTORS

Executor

Name and Surname

Head of clinical tests

Malykhin Anatolii V., PhD

Academician of Russian Academy of

Natural Sciences

Leading scientist of the Institute of neurology, Psychiatry and Narcology

of AMS of Ukraine

Responsible executor Malykhin Anatolii, PhD

Researcher

Kucherenko Petro A., MD “Biopromin” LTD

Researcher

Kryvenko Sergii, PhD Leading engineer of “Biopromin”

LTD

Page 3: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

TABLES OF CONTENTS

SUMMARY ................................................................................................................................... 4 INTRODUCTION ......................................................................................................................... 5 PURPOSE OF STUDY ................................................................................................................ 6 OBJECTIVES OF STUDY ........................................................................................................... 6 MATERIALS AND METHODS OF STUDY ............................................................................. 6 RESULTS OF STUDY .............................................................................................................. 10 CONCLUSIONS ......................................................................................................................... 19 REFERENCES ............................................................................................................................ 20

Page 4: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

SUMMARY Materials on clinical study of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) are provided. It is designed to collect data of a patient's clinical blood analysis values (hemoglobin, glucose, aspartate aminotransferase (AST), alanine aminotransferase (ALT), amylase, total protein, total bilirubin, direct bilirubin, ceruloplasmin, hematocrit, creatinine, leucocytes, lymphocyte, urea, monocytes, erythrocyte sedimentation rate (ESR), differential count – band neutrophils) without invasive blood drawing followed by data processing with “USPIH 10” software. Clinical study of volunteers was carried out by comparative analysis with using certified hematology analyzers and commonly accepted methods of laboratory examinations. Study results were processed by statistical methods. This study allowed determining that the test medical device ensures the necessary reproducibility with 95% of confidence interval. Parallel control by the standard (common) laboratory methods is recommended. The report contains 32 pages, 3 tables. Bibliography: 10 sources. Purposes of clinical studies

1. Comparative assessment of clinical blood analysis accuracy determined by the medical device by reference to the results obtained from laboratories.

2. Statistical processing of received data for objective comparative assessment of the device “Automatic Noninvasive Express Screening Analyzer ANESA”.

3. Detection of unexpected adverse reactions, risks and undesirable effect on a human. 4. Assessment of safety of the device “Automatic Noninvasive Express Screening Analyzer

ANESA” during usage, based on the subjective sensations of volunteers.

Inclusion criteria for study subjects:

Inclusion criteria: 1) Age: 18-80. 2) Body weight: 40-110 kg. 3) State of health: patients with alcohol abuse, neurological pathology, without allergic diseases 4) Ethnic status: irrelevant. 5) Gender: both. 6) Patient’s mental status: within normal. 7) Organ to be contacted: 5 areas on a body: cervical, axillary and abdominal artery bifurcation.

Commencement date of study: January 10, 2013 Completion date of study: May 20, 2013

Page 5: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

INTRODUCTION

In 2008, the main concept of activity of the World Health Organization (WHO) was primary care patients. The objective methods of patient examination, first of all blood analysis, are the base for diagnosing of diseases. Inflammatory and allergic processes, hematologic diseases, are the initial things for the subsequent comprehensive patient examination and identification of a disease without clinical manifestations. Blood analysis plays a leading role among the various comprehensive diagnostic methods. Correct and early diagnosing, reasonable treatment, true prognosis of a disease development are often impossible without morphological and biochemical blood analysis data. Blood picture analysis in the course of a disease allows judging of the process severity, patient status improvement or deterioration. Clinical blood analysis is of extremely great importance. General clinical blood analysis contains the results of examination of different parameters (including hemoglobin and glucose), influencing to a human life. “Automatic Noninvasive Express Screening Analyzer ANESA” Technical Specifications of Ukraine ТУ У 33.1-22716816-001:2006, HS code 9018 19 90 00 manufactured by “Biopromin” Ltd. is a device that allows measuring temperature on the “reference” biologically active points on a human body, and then determining (calculating) the values of the clinical and biochemical blood analysis, which is used for diagnostics of diseases in medical practice.

Page 6: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

MAIN PURPOSE OF STUDY The purpose of post-clinical study is to test capability of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) to determine the values of clinical and biochemical blood analysis automatically by results of temperature measurement on the “reference” points on a human body, without blood drawing and skin damage, as well as to assess safety of this device and detect the unforeseen side effects while in operation.

OBJECTIVES OF STUDY Clinical examination of the peripheral blood values using the “Automatic Noninvasive Express Screening Analyzer ANESA”, are carried out with the assistance of the Institute of neurology, Psychiatry and Narcology of AMS of Ukraine (Kharkiv, Ukraine), Regional Clinical Psychiatry Hospital No.3, Section 8 (Kharkiv, Ukraine) and “Alfa Labservice” LTD on 116 volunteers with the following diseases: Konovalov-Wilson’s disease, drug addiction, alcoholism   and endogenous depression, for the following:

• Comparative analysis of the peripheral blood values examination data using certified analyzers and commonly accepted methods of laboratory examinations.

• Detection of unexpected adverse reactions, risks and undesirable effect • Assessment of compliance of the clinical studies with the program developed

MATERIALS AND METHODS OF STUDY 1. Measurement conditions:

a. – room air temperature is not more than +27ºС or not less than +20°С; b. – room air humidity is not more than 80%, when the temperature is 25ºС; c. – no aggressive agents are kept in a room; d. – no excessive dust in a room; e. – no long-term exposure of direct sunlight and conditioned air flow; f. – no exposure of strong electric and magnetic fields; g. – atmosphere pressure should be in the bonds of 82 - 101.38 kPa.

2. Requirements to the study subjects.

Before measurement procedure the following shall be observed:

– ordinary dietary regime for at least three days (carbohydrates content > 125-150 g a day) and habitual physical exercises for the patients shall be followed; – fasting test upon restraint from food during one night for 10-14 hours (smoking and alcohol use is prohibited); – lying position or stepped-lying position for a patient while drawing blood, no smoking, low temperature and physical exercises; – test is recommended upon or under stressed conditions; – no treatment procedures and drug administration (adrenaline, glucocorticoids, contraceptives, caffeine, thiazide diuretics, psychotropics and antidepressants, and other medical agents affecting hematic picture) before test;

3. Measurement procedure. Microprocessor sensors of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, were applied on certain “reference” biologically active points on a human body surface (after pre-treatment of skin and sensors with 96% ethyl alcohol for degreasing):

Page 7: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

- blue sensor – on bifurcation of the left carotid artery in the annular cartilage region; - green sensor – on bifurcation of the right carotid artery in the annular cartilage region; - yellow sensor – on the left axillary crease (thermometer position); - violet sensor – on the right axillary crease (thermometer position); - red sensor – on the abdominal area (inside the umbilical crease, if no umbilicus, on the area

where it was positioned before surgery). The measurement procedure, depending on the rate of human body temperature stabilization, took 180-720 seconds.

Values to be measured

While testing the medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, the following parameters were determined:

- hemoglobin, g/l - glucose, mmol/l - aspartate aminotransferase (AST), mmol/l - alanine aminotransferase (ALT), mmol/l - amylase, g/l*h - total protein, g/l - total bilirubin, mkmol/l - direct bilirubin, mkmol/l - ceruloplasmin, g/l - hematocrit, % - creatinine, mkmol/l - leucocytes, x10E9/l - lymphocyte, % - urea, mmol/l - monocytes, % - erythrocyte sedimentation rate (ESR), mm/h - differential count – band neutrophils, %

The results were analyzed with reference to the data obtained within measurement of the same venous blood parameters using the certified hematology analyzer and commonly accepted methods of laboratory tests in the above-mentioned laboratories. Venous blood was examined using the certified hematology analyzer, by venipuncture of the median cubital vein into the test tubes of Venosafe VF-053 SDK with EDTA. The composition of peripheral blood was determined using the certified analyzers according to the operating manual. Study results were delivered next day after blood drawing and measurement using a noninvasive hemogram analyzer.

The criteria for inclusion of volunteers into the study were: 1. Absence of allergic diseases. 2. Consent of a volunteer to take part in the study. Volunteers were excluded from the study in case of: 1. Existence of allergic diseases. 2. Volunteers take medications, which may affect to the blood parameters. 3. A volunteer refuses to take part in the research.

Page 8: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Variables to be measured In accordance with the National Standards of Ukraine, GOST ISO 5725 it is required to determine that the test medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, ensures necessary repeatability with the confidence interval of 95%. In this regard the study results were compared and analyzed as well as statistically processed.

Study subjects This study was carried out during the period from January 01, 2013 till May 20, 2013. The number of results is different fir different parameters. Thus the studied amount of people was 91 patients for examination of ALT, 88 patients – amylase, 94 patients – AST, 103 patients – total protein, 99 patients – total bilirubin and direct bilirubin, 6 patients – ceruloplasmin, 116 patients – hemoglobin, 36 patients – glucose, 9 patients – hematocrit, 113 patients – leukocytes, lymphocytes, monocytes, band neutrophils and ESR, 101 patients – urea. The patients were not grouped for the research by their pathology. The results of tests, made on certified hematology analyzer were considered as standard, with satisfied accuracy, certified and acceptable by the WHO (hereinafter standard method). The results of test, made on “Automatic Noninvasive Express Screening Analyzer ANESA”, were considered as the method, which should be approved (hereinafter new method).

Analytical procedures

These results were statistically processed in accordance with the relevant standards:

• National Standards of Ukraine 3514-97. Statistical methods of control and regulation. Terms and definitions. K.: Gosstandart of Ukraine. 1997. p. 52. • National Standards of Ukraine, GOST ISO 5725-1:2005 Validity (accuracy and precision) of measurement methods and results. Part 1. General principles and definitions (GOST ISO 5725-1-2003, IDT). • National Standards of Ukraine, GOST ISO 5725-2:2005 Validity (accuracy and precision) of measurement methods and results. Part 2. Basic method to determine repeatability and reproducibility of the standard measurement method (GOST ISO 5725-2-2003, IDT) • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate values of the standard measurement method precision (GOST ISO 5725-3-2003, IDT) • National Standards of Ukraine, GOST ISO 5725-4:2005 Validity (accuracy and precision) of measurement methods and results. Part 4. Basic method to determine standard measurement method accuracy (GOST ISO 5725-4-2003, IDT) • National Standards of Ukraine, GOST ISO 5725-5:2005 Validity (accuracy and precision) of measurement methods and results. Part 5. Alternative methods to determine standard measurement method precision (GOST ISO 5725-5-2003, IDT) • National Standards of Ukraine, GOST ISO 5725-6:2005 Validity (accuracy and precision) of measurement methods and results. Part 6. Use of accuracy values in practice (GOST ISO 5725-6-2003, IDT)

Statistical analysis methods

Statistical processing of results:

1) Determination of average value of each parameter and standard deviation (SD).

Page 9: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

2) Calculation of correlation factor between values of each hematologic parameter measured by the standard method and new method. Determination of statistical significance of correlation factor: hypothesis formulation of parameter measurement independence by the standard and new methods (НО: р = 0) and validation of this hypothesis. In the event that it is accepted, these results were deemed to be independent of one another and correlation between them was deemed to be statistically insignificant. In the event that it is rejected, these results were deemed to be dependent of one another and correlation between them was deemed to be statistically significant.

3) Interval estimate for each hematologic parameter and determination of parameter interval

overlap percent measured by the new and standard methods.

4) Hypothesis formulation of the same accuracy of the standard and new methods (НО: σ12 = σ2

2) and validation of this hypothesis. Further, hypothesis is formed with regard to possibility of correlation between measurements results obtained by the standard and new methods (НО: µ1

2 = µ2

2). At that, the standard method was deemed to be satisfactory and results obtained by this method were deemed to be accepted in medical practice, i.e., reference ones for new method. As a result of this analysis, decisions with regard to compliance of new method with the requirements of medical practice.

PATIENT EXAMINATION DESIGN

FUNCTIONAL AND ANTHROPOMETRIC MEASUREMENTS

Test Age Weight Heart rate, bpm Respiration rate, rpm

DATA OF AN EXAMINATION

Comparison between ANESA and laboratory results (units)

Patient’s ID

ANESA Lab Difference

List of detection of unexpected adverse reactions, risks and undesirable effect on a human

Date Type of unexpected reactions

Degree of severity Measures

RESULTS OF STUDY

Tables 1 and 2 provide anthropometric and physiological characteristics of volunteers as well as results of examination by each parameter, obtained by the different methods: standard and new.

Page 10: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Table 1

Anthropometric and physiological characteristics of volunteers

The study included males and females, normosthenic type, respiratory rate was within

physiological normal.

Comparative analysis of study results

The results of examination for each tested parameter, received from “Automatic Noninvasive Express Screening Analyzer ANESA” and from laboratory (using certified hematology analyzers and commonly accepted methods of laboratory examinations), as well as the differences between the data, which were received by two different methods, were generated by the template of the table 2:

Table 2. Comparative analysis of results by a parameter

(made on 16 pages)

Comparative analysis of results by a parameter (units) ID of a patient ANESA Lab Differences between the

results

Comparative analysis of results by the

parameter ceruloplasmin (g/l)

Comparative analysis of results by the

parameter haematocrit (%)

Parameter Number of subjects/patients

Average age, years

Average body weight, kg,

Average heart rate, bm

ALT 91 37,92±0,51 79,29±0,71 93,52±0,58 AST 94 37,77±0,49 79,01±0,68 92,74±0,58 Amylase 88 35,72±0,47 79,75±0,74 92,64±0,58 Total protein 103 37,07±0,46 77,64±0,62 92,27±0,56 Total bilirubin, Direct bilirubin

99 35,63±0,43 78,52±0,64 89,05±0,58

Ceruloplasmin 6 28,67±7,12 65,17±6,72 73,33±7,87 Haemoglobin 116 36,72±0,40 77,34±0,53 91,16±0,52 Glucose 36 31,31±0,53 77,50±1,59 92,72±1,15 Haematocrit 9 35,67±6,27 67,44±3,80 70,67±4,63 Creatinine 91 36,37±0,49 79,12±0,71 92,16±0,53 Leukocytes, Lymphocytes, Monocytes, Band neutrophils, ESR

113 36,78±0,41 77,60±0,54 91,27±0,53

Urea 101 35,64±0,43 78,35±0,63 90,50±0,56

Page 11: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter haemoglobin (g/l):

Page 12: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter glucose (mmol/l)

Page 13: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter AST (mmol/l)

Page 14: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter ALT (mmol/l)

Page 15: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter amylase (g/l*h)

Page 16: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter amylase (g/l)

Page 17: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter total bilirubin (mkmol/l)

Page 18: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter direct bilirubin (mkmol/l)

Page 19: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter creatinine (mkmol/l)

Page 20: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter leukocytes (x10E9/l)

Page 21: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter lymphocytes (%)

Page 22: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter urea (mmol/l)

Page 23: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter monocytes (%)

Page 24: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Comparative analysis of results by the parameter ESR (mm/h)

Page 25: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Statistical Analysis The received results were processed statistically in order to make the objective comparative assessment of the new method and standard one, using in medical practice (Table 3).

Table 3. Results of statistical processing of each parameter

Parameter

ANESA (n=91)

Lab (n=91)

ALT, MMOL/L Average value

SD Confidence interval

1,51 0,88 1,32÷1,69

1,61 1,04 1,39÷1,83

Zкр Zβ

1,996 0,716

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for ALT count measurement – accepted Parameter

ANESA (n=88)

Lab (n=88)

AMYLASE, g/l*h Average value

SD Confidence interval

25,00 15,51 21,71÷28,28

24,67 16,21 21,23÷28,10

Zкр Zβ

1,996 0,137

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for AMYLASE count measurement – accepted

Parameter

ANESA (n=94)

Lab (n=94)

AST, MMOL/L Average value

SD Confidence interval

0,92 0,48 0,82÷1,02

0,90 0,49 0,80÷1,00

Zкр Zβ

1,996 0,268

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for AST count measurement – accepted Parameter

ANESA (n=103)

Lab (n=103)

TOTAL PROTEIN, MMOL/L Average value

SD Confidence interval

69,84 10,51 67,79÷71,88

68,49 11,99 66,15÷70,82

Zкр Zβ

1,973 0,861

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for TOTAL PROTEIN count measurement – accepted

Page 26: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Table 3 (continued)

Parameter

ANESA (n=99)

Lab (n=99)

TOTAL BILIRUBIN, MMOL/L Average value

SD Confidence interval

20,83 10,25 18,79÷22,87

20,67 10,36 18,60÷22,74

Zкр Zβ

1,996 0,107

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for TOTAL BILIRUBIN count measurement – accepted

Parameter

ANESA (n=99)

Lab (n=99)

DIRECT BILIRUBIN, MMOL/L Average value

SD Confidence interval

5,50 3,16 4,87÷6,13

5,13 3,49 4,44÷5,83

Zкр Zβ

1,996 0,787

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for DIRECT BILIRUBIN count measurement – accepted

Parameter

ANESA (n=6)

Lab (n=6)

CERULOPLASMIN, G/L Average value

SD Confidence interval

41,40 2,75 38,65÷44,15

32,90 16,47 16,45÷49,35

Zкр Zβ

2,447 1,247

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for CERULOPLASMIN count measurement – accepted

Parameter

ANESA (n=116)

Lab (n=116)

HAEMOGLOBIN, G/L Average value

SD Confidence interval

136,84 18,25 133,49÷140,19

136,54 18,08 133,22÷139,86

Zкр Zβ

1,973 0,125

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for HAEMOGLOBIN count measurement – accepted

Page 27: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Table 3 (continued) Parameter

ANESA (n=36)

Lab (n=36)

GLUCOSE, MMOL/L Average value

SD Confidence interval

4,82 0,98 4,49÷5,16

5,06 1,05 4,71÷5,42

Zкр Zβ

2,028 0,986

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for GLUCOSE count measurement – accepted

Parameter

ANESA (n=9)

Lab (n=9)

HAEMATOCRIT, % Average value

SD Confidence interval

40,44 2,75 38,36÷42,52

41,01 3,53 38,35÷43,68

Zкр Zβ

2,262 0,383

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for HAEMATOCRIT count measurement – accepted

Parameter

ANESA (n=91)

Lab (n=91)

CREATININE, MKMOL/L Average value

SD Confidence interval

68,63 11,53 66,23÷71,03

69,23 12,63 66,61÷71,86

Zкр Zβ

1,996 0,338

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for CREATININE count measurement – accepted

Parameter

ANESA (n=113)

Lab (n=113)

LEUKOCYTES, x10E9/L Average value

SD Confidence interval

10,26 6,40 9,07÷11,45

10,26 6,45 9,06÷11,46

Zкр Zβ

1,973 0,009

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for LEUKOCYTES count measurement – accepted

Page 28: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Table 3 (continued) Parameter

ANESA (n=113)

Lab (n=113)

LYMPHOCYTES, % Average value

SD Confidence interval

18,95 8,25 17,42÷20,49

18,47 8,07 16,97÷19,97

Zкр Zβ

1,973 0,444

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for LYMPHOCYTES count measurement – accepted

Parameter

ANESA (n=101)

Lab (n=101)

UREA, MMOL/L Average value

SD Confidence interval

5,38 1,21 5,14÷5,62

5,18 1,37 4,91÷5,44

Zкр Zβ

1,973 1,111

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for UREA count measurement – accepted

Parameter

ANESA (n=113)

Lab (n=113)

MONOCYTES, % Average value

SD Confidence interval

7,84 2,69 7,34÷8,34

6,87 2,35 6,43÷7,31

Zкр Zβ

1,973 2,886

Zβ> Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for MONOCYTES count measurement – not accepted

Parameter

ANESA (n=113)

Lab (n=113)

BAND NEUTROPHILS, % Average value

SD Confidence interval

6,46 4,92 5,55÷7,38

5,76 5,71 4,70÷6,82

Zкр Zβ

1,973 0,991

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for BAND NEUTROPHILS count measurement – accepted

Page 29: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Table 3 (continued) Parameter

ANESA (n=113)

Lab (n=113)

ESR, mm/h Average value

SD Confidence interval

20,74 17,69 17,45÷24,04

19,41 19,63 15,75÷23,06

Zкр Zβ

1,973 0,538

Zβ< Zкр accepted

Hypothesis of the same accuracy of the standard and new methods for ESR count measurement – accepted

Average level of ALT in volunteers measuring by the standard method totaled 1.61 ± 1.04 mmol/l, and by a new method totaled 1.51 ± 0.88 mmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of ALT is accepted”. Hypothesis acceptance level Zβ (0.716) < Zкp (1.996) is deemed to be sufficient. Average level of amylase in volunteers measuring by the standard method totaled 24.67 ± 16.21 g/l*h, and by a new method totaled 25.00 ± 15.51 mmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of amylase is accepted”. Hypothesis acceptance level Zβ (0.137) < Zкp (1.996) is deemed to be sufficient. Average level of AST in volunteers measuring by the standard method totaled 0.90 ± 0.49 mmol/l, and by a new method totaled 0.92± 0.48 mmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of AST is accepted”. Hypothesis acceptance level Zβ (0.268) < Zкp (1.996) is deemed to be sufficient. Average level of total protein in volunteers while measuring by the standard method totaled 68.49 ± 11.99 g/l, and by a new method totaled 69.84 ± 10.51 g/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of total protein count is accepted”. Hypothesis acceptance level Zβ (0.861) < Zкp (1.973) is deemed to be sufficient. Average level of total bilirubin in volunteers measuring by the standard method totaled 20.67 ± 10.36 mkmol/l, and by a new method totaled 20.83± 10.25 mkmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of total bilirubin is accepted”. Hypothesis acceptance level Zβ (0.787) < Zкp (1.996) is deemed to be sufficient. Average level of direct bilirubin in volunteers measuring by the standard method totaled 5.13 ± 3.49 mkmol/l, and by a new method totaled 5.50± 3.16 mkmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of

Page 30: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

the standard and new methods for measurement of direct bilirubin is accepted”. Hypothesis acceptance level Zβ (0.107) < Zкp (1.996) is deemed to be sufficient. Average level of ceruloplasmin in volunteers while measuring by the standard method totaled 32.90 ± 16.47 g/l, and by a new method totaled 41.40 ± 2.75 g/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of ceruloplasmin count is accepted”. Hypothesis acceptance level Zβ (1.247) < Zкp (2.447) is deemed to be sufficient. Average level of hemoglobin in volunteers while measuring by the standard method totaled 136.54 ± 18.08 g/l, and by a new method totaled 138.84 ± 18.25 g/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of hemoglobin count is accepted”. Hypothesis acceptance level Zβ (0.125) < Zкp (1.973) is deemed to be sufficient. Average level of glucose in volunteers measuring by the standard method totaled 5.06 ± 1.05 mmol/l, and by a new method totaled 4.82± 0.98 mmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of glucose is accepted”. Hypothesis acceptance level Zβ (0.986) < Zкp (2.028) is deemed to be sufficient. Average level of hematocrit in volunteers measuring by the standard method totaled 41.01 ± 3.53 %, and by a new method totaled 40.44± 2.75 %, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of hematocrit is accepted”. Hypothesis acceptance level Zβ (0.383) < Zкp (2.262) is deemed to be sufficient. Average level of creatinine in volunteers measuring by the standard method totaled 69.23 ± 12.63 mkmol/l, and by a new method totaled 68.63± 11.53 mkmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of creatinine is accepted”. Hypothesis acceptance level Zβ (0.338) < Zкp (1.996) is deemed to be sufficient. Average level of leukocytes in volunteers measuring by the standard method totaled 10.26 ± 6.45 X10E9/l, and by a new method totaled 10.26± 6.40 X10E9/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of leukocytes is accepted”. Hypothesis acceptance level Zβ (0.009) < Zкp (1.973) is deemed to be sufficient. Average level of lymphocytes in volunteers while measuring by the standard method totaled 18.47 ± 8.07%, and by a new method totaled 18.95 ± 8.25%, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of lymphocyte count is accepted”. Hypothesis acceptance level Zβ (0.444) < Zкp (1.973) is deemed to be sufficient.

Page 31: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

Average level of urea in volunteers measuring by the standard method totaled 5.18 ± 1.37 mmol/l, and by a new method totaled 5.38± 1.21 mmol/l, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of glucose is accepted”. Hypothesis acceptance level Zβ (1.111) < Zкp (1.973) is deemed to be sufficient. Average level of monocytes in volunteers while measuring by the standard method totaled 6.87 ± 2.35%, and by a new method totaled 7.84 ± 2.69%, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of monocyte count is not accepted”. Hypothesis acceptance level Zβ (2.886) < Zкp (1.973) is deemed to be sufficient. Average level of band neutrophiles in volunteers while measuring by the standard method totaled 5.76 ± 5.71%, and by a new method totaled 6.46 ± 4.92%, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of band neutrophile count is accepted”. Hypothesis acceptance level Zβ (0.991) < Zкp (1.973) is deemed to be sufficient. Average level of ESR in volunteers while measuring by the standard method totaled 19.41 ± 19.63 mm/h, and by a new method totaled 20.74 ± 17.69 mm/h, that is an uncertain difference (р > 0.05). When testing the hypothesis of equal variances with confidence interval of α = 0.05, the particular indicators of statistics (Zβ) allowed drawing a following conclusion “Hypothesis of the same accuracy of the standard and new methods for measurement of ESR is accepted”. Hypothesis acceptance level Zβ (0.538) < Zкp (1.973) is deemed to be sufficient. Thus, when statistically processing the blood values in volunteers (hemoglobin, AST, ALT, amylase, total protein, total bilirubin, direct bilirubin, ceruloplasmin, hematocrit, creatinine, leukocytes, lymphocytes, urea, ESR and band neutrophils), the hypothesis of the same accuracy of standard hematology analyzer and medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) was accepted. It is recommended to exercise concurrent control of monocyte by the standard (common) laboratory methods.

Assessment of unexpected adverse reactions, risks and undesirable effects

No unexpected adverse reactions, risks and undesirable effects were reported within this study.

Assessment of safety During the study, no adverse effect of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA” on the volunteers’ health was recorded.

CONCLUSIONS

1. Based on the results of comparative analysis of the medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) it was determined that the test medical device ensures necessary repeatability of test blood values (hemoglobin, AST,

Page 32: Approved by Director of BIOPROMIN LTD · • National Standards of Ukraine, GOST ISO 5725-3:2005 Validity (accuracy and precision) of measurement methods and results. Part 3. Intermediate

ALT, amylase, total protein, total bilirubin, direct bilirubin, ceruloplasmin, hematocrit, creatinine, leukocytes, lymphocytes, urea, ESR and band neutrophils) with the confidence interval of 95% during studies in volunteer in the neurological and alcoholic patient groups. At the same time, it was determined that the test medical device doesn’t ensure necessary repeatability of monocytes values with the confidence interval of 95% .

2. Taking into account a sufficiently restricted statistical sampling, it is recommended to use the

following medical device “Automatic Noninvasive Express Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) for needs of medical practice in line with the generally accepted methods of blood analyses.

3. As far as no unexpected adverse reactions, risks and undesirable effect on human were reported, it

can be concluded that operation of the device both for a patient and an operator providing observance of the operating manual is deemed to be safe.

4. Study results show no adverse effect of the medical device ““Automatic Noninvasive Express

Screening Analyzer ANESA”, manufactured by “Biopromin” Ltd. (Ukraine) on the subject health status.

REFERENCES 1. L.A. Danilova. Blood and urine analyses. – Saint-Petersburg. 1999. – 128 p. 2. Clinical laboratory diagnostics: study methods: Educational guide for students specialized in

Pharmacy. “Clinical Pharmacy”, “Laboratory diagnostics” for Higher Educational Institutions/I.A. Zupanets, S.V. Misyureva., V.V. Propisnova et al.: under the editorship of I.A. Zupanets. – 3rd revised and enlarged edition. – Kharkov city: Publishing House NFaU: Golden pages. 2005. –200 p.

3. G.I. Kozinets. Interpretation of blood and urine analyses. – Saint-Petersburg, 1997. – 128 p. 4. L.V. Kozlovskaya, M.A. Martynova. Educational guide in clinical laboratory study methods

(programming principles): under the editorship of the Academician E.M. Tareev, Professor A.V. Sumarokov. – Moscow: Medicine, 1975. – 352 p.

5. Laboratory study methods in a clinic: Guidance/V.V. Menshikov, L.N. Delektorskaya, R.P. Zolotitskaya et al.: under the editorship of V.V. Menshikov. – Moscow: Medicine, 1987. – 368 p.

6. Norm in medical practice: Reference book. – Moscow: “MEDpress” Ltd., 1998. – 144 p. 7. Guide in clinical laboratory diagnostics: Educational guide. 4.1-2./M.A. Bazarnova, A.I. Vorobyov,

Z.S. Narkagan et al.: under the editorship of M.A. Bazarnova, A.I. Vorobyov. – К.: Vashcha shk\. 1991. – 615 p.

8. F.J. Shiffman. Pathophysiology of blood. Translation from English. – Moscow-Saint-Petersburg: “Publishing House BINOM” – “Nevskiy Dialekt”, 2000. – 448 p.

9. Henry J.В. Clinical diagnosis and management by Laboratory Methods. – Philadelphia. PA: Saunders. 1991. 682 p.

10. Winter M.E. et al. Basic Clinical Pharmacokinetics. Applied Therapeutics. – Vancouver, 1994. – 93 p.


Recommended