Arsine Detection using Chemcassette ® Technology
Transcript
Page 1: Arsine Detection using Chemcassette ® Technology.

Arsine Detection using Chemcassette® Technology

Page 2: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

2

History

• In 1998 ACGIH proposed changing the TLV for AsH3 from 50 ppb to 2 ppb

• In 2007 ACGIH changed the TLV for AsH3 from 50 ppb to 5 ppb

• Honeywell Analytics has already developed a higher sensitivity and longer lasting Chemcassette®

• Study data supported the ability for HA to meet the proposed lower TLV level for AsH3

Page 3: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

3

Development

• The increased sensitivity of the new formulation allowed Honeywell Analytics to achieve better lower detectable limits for all equipment, as shown

Page 4: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

4

Gas Monitors’ Low Detection Limit Improvement

  AsH3 PH3 SiH4 B2H6

Monitor XP TapeStandard

TapeXP Tape

StandardTape

XP TapeStandard

TapeXP Tape

StandardTape

Vertex/CM4 0.3 ppb 5 ppb 5 ppb 5 ppb 0.25 ppb 0.5 ppb 10 ppb 15 ppb

System 16 0.7 ppb 7 ppb 8 ppb 10 ppb 0.4 ppb 1.33 ppb 13 ppb 21 ppb

SPM 0.7 ppb 15 ppb 20 ppb 30 ppb 0.25 ppb 0.5 ppb 10 ppb 30 ppb

TLD 0.7 ppb 15 ppb 20 ppb 30 ppb 0.25 ppb 0.5 ppb 10 ppb 30 ppb

Page 5: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

5

Sensitivity Aging Results

• Honeywell Analytics also tested for Chemcassette® sensitivity. Would the formula remain consistent over three months?

Day 1 Day 14 Day 42 Day 77 Day 122

Page 6: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

6

Gas Generation Methodology• Apparatus A: gas source with 2 certified AsH3 gas cylinders

• Apparatus B: Kin-Tech gas generation with permeation device

• Concentrations determined by theoretical calculations, and confirmed by 3 standard methods

First Stage Gas

Mixing Chamber

Dilution AirConcentrated AsH3

AsH3 Gas(from cylinder orpermeation tube

standard gasgenerator)

FilteredCompressed

Air

Samplerfor quantitative

analysis

Mass Flow Controller0-500 ml/min

Vertex Gas Monitor

CM4Gas Monitor

SPMGas Monitor

System16 Gas Monitor

Mass Flow Controller1-20 l/min

Page 7: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

7

Validation Methods

• Honeywell verified standard gas concentrations using 3 independent methods

– FTIR (64 scans with a 10 m cell)– Analytical method (Visible spectrophotometry based on NIOSH method S332)– Kin-Tech Standard Gas Generator (Permeation Device Method)

Page 8: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

8

Gas Cylinder Consistency Test

Page 9: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

9

Signal to Noise Ratio

• To validate instrument response, Honeywell Analytics did extensive “noise level” studies both in the laboratory and in the field

• A new LDL (lower detection limit) was determined by finding an acceptable signal to noise ratio

• Ratio was set at >3:1 signal to noise

Page 10: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

10

CM4® Low Concentration AsH3 Signal and Noise

Page 11: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

11

System16 Low Concentration AsH3 Signal and Noise

Page 12: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

12

SPM Low Concentration AsH3 Signal and Noise

Page 13: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

13

Ambient RH vs. Noise Test

• Normal RH setting in semiconductor fabs is about 40%• Extensive studies were done on noise levels at 5% RH, at

40% RH and at 90% RH versus the signal generated by 0.5 ppb AsH3 (one tenth of the proposed TLV level)

• Maximum noise drift caused by shifting RH from 5 to 40% is 10 cts.

• This is less than 1/10th the signal generated by 0.5 ppb of AsH3

Page 14: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

14

CM4® Ambient RH Noise vs. 0.5 ppb AsH3 Signal

Page 15: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

17

CM4® Low Level Calibration Accuracy Verification II

  Test 1 Test 2 Test 3 Test 4 Test 5 Test 6 Test 7 Test 8

Oven Temperature (°C) 30 30 30 50 50 50 60 60

Room RH (%) 50 48 48 51 50 52 50 49

Gas RH (%) 40 40 40 40 40 40 40 40

Total Output Flow* (ml/min) 3273 3273 3273 3273 3273 3273 3273 3273

Cylinder Pressure (psi) 50 50 50 55 55 55 55 55

Permeation Oven Temperature (°C) 30 30 30 50 50 50 60 60

Permeation Rate (nl/min) 140.84 140.84 140.84 283.39 283.39 283.39 390.36 390.36

KT Gas Concentration (ppb) 43.03 43.03 43.03 86.58 86.58 86.58 119.27 119.27

KT-Gas Flow (ml/min) 230 110 80 245 560 950 3273 3273

Dilution Air Flow (L/min) 10.05 10.05 10.05 10.07 10.07 7.6 16.2 4.5

Standard Gas Concentration (ppb) 0.96 0.47 0.34 2.06 4.56 9.62 20.05 50.22

CM4 Response (ppb) 0.98 0.5 0.32 2.3 5.2 10.1 21.5 52

CM4 Response/Standard Gas (ppb) 102.08% 107.32% 94.16% 11.84% 114.00% 104.98% 107.25% 103.55%

Standard Gas Source: Standard Gas Generator with Permeation Device

Page 16: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

18

CM4® Low Level Calibration Accuracy Verification IIStandard Gas Source: Standard Gas Generator with Permeation Device

Std Gas Conc (ppb)

CM4 (ppb) Accuracy

0.96 0.98 102.08%

0.47 0.5 107.32%

0.34 0.32 94.16%

2.06 2.3 111.84%

4.56 5.2 114.00%

9.62 10.1 104.98%

20.05 21.5 107.25%

50.22 52 103.55%

Page 17: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

19

CM4® XP Result Consistency Over Time

• Before placing an XP CM4® at a Beta site, the CM4® was tested for consistency over time in laboratory condition

Page 18: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

20

Beta Testing Data

• Beta Customer ran sample lines from the test CM4® to existing monitoring locations at the tool, also in the fab utility area, and to VMBs located in the sub-fab

• The CM4® was run “in situ” for over three months• Data was collected by the Customer and transmitted to

Honeywell Analytics

Page 19: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

21

CM4® AsH3 Low Detection Limit Feasibility Field Study

Beta Testing Site A, Test Result I

Page 20: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

22

CM4® AsH3 Low Detection Limit Feasibility Field Study

Beta Testing Site A, Test Result II

Page 21: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

23

CM4® AsH3 Low Detection Limit Feasibility Field Study

Beta Testing Site A, Test Result III

Page 22: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

24

CM4® AsH3 Low Detection Limit Feasibility Field Study

Beta Testing Site A, Test Result IV

Page 23: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

25

CM4® AsH3 Low Detection Limit Feasibility Field Study

Beta Testing Site A, Test Result V

Page 24: Arsine Detection using Chemcassette ® Technology.

Honeywell.com

26

Conclusion

• Based on Lab work, proof of methodology, and field trials, Chemcassette Technology can successfully monitor for AsH3 at the newly established level of 5 ppb without risk of false positives.