8/9/2019 Aspirators surgical.PDF http://slidepdf.com/reader/full/aspirators-surgicalpdf 1/26  5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA  Tel +1 (610) 825-6000  Fax +1 (610) 834-1275  Web www.ecri.org  E-mail [email protected] UMDNS Information This Product Comparison covers the following device term and product code as listed in ECRI Institute’s Universal Medical Device Nomenclature System™ (UMDNS™):   Aspirators, Surgical [10-217] Aspirators, Surgical Scope of this Product Comparison This Product Comparison covers surgical aspirators capable of creating a vacuum pressure 500 mm Hg. Surgical waste disposal systems, which utilize a rover and docker system, and are capable of sterilizing and disposing of the waste, are not included in this report. Also excluded are centralized suction systems that connect to standard wall inlets. For information on other types of aspirators, see the following Product Comparisons:  Aspirators, Emergency  Aspirators, Thoracic  Aspirators, Uterine Purpose Most surgical procedures require suctioning to remove blood and irrigating fluids that accumulate in the operative field and obstruct the surgeon’s view. Hospitals generally use a central high -flow-rate vacuum pump that provides vacuum through pipelines to standard wall inlets throughout the hospital. However, portable or mobile vacuum systems can be used if there is no central vacuum system or if suctioning is required in areas that do not have vacuum inlets. Principles of operation The basic surgical aspirator consists of a line-powered vacuum pump, a vacuum regulator and gauge, a collection canister, and an optional  bacterial filter. Plastic tubing connects these components, completing an open-ended system that continuously draws tissue debris and fluid from the surgical field to the collection canister. The gauge allows the user to set a safe limit for suctioning, to assess the performance of the vacuum pump, and to detect leaks or blockages. Units are either portable or mounted on a stand or cart for mobility. Vacuum pumps A single- or double-vacuum pump creates suction within the system. Double-pump systems operate in synchrony, with both pumps contributing to the suctioning capacity. The three types of pumps used in surgical aspirators are rotary vane, diaphragm, and cylinder piston. In a rotary-vane pump, a rotor is mounted off-center in relation to the inside diameter of the pump housing (see Figure 1). As the rotor turns, sliding vanes that fit into notches in the rotor follow the contour Comprehensive of Biomedical Engineering Site (www.dezmed.com   W   W   W  .   D   E   Z   M   E   D  .   C   O   M
Transcript
Page 1: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 1/26

 

5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA  Tel +1 (610) 825-6000  Fax +1 (610) 834-1275  Web www.ecri.org  E-mail [email protected]

UMDNS Information

This Product Comparison covers the followingdevice term and product code as listed inECRI Institute’s Universal Medical DeviceNomenclature System™ (UMDNS™): 

  Aspirators, Surgical [10-217]

Aspirators, Surgical

Scope of this Product Comparison

This Product Comparison covers surgical aspirators capable of creating a vacuum pressure 500 mm Hg.

Surgical waste disposal systems, which utilize a rover and docker system, and are capable of sterilizing and

disposing of the waste, are not included in this report. Also excluded are centralized suction systems that connect

to standard wall inlets. For information on other types of aspirators, see the following Product Comparisons:

  Aspirators, Emergency 

  Aspirators, Thoracic 

  Aspirators, Uterine 

Purpose

Most surgical procedures require suctioning to remove blood and

irrigating fluids that accumulate in the operative field and obstruct the

surgeon’s view. Hospitals generally use a central high-flow-rate vacuum

pump that provides vacuum through pipelines to standard wall inlets

throughout the hospital. However, portable or mobile vacuum systems

can be used if there is no central vacuum system or if suctioning is

required in areas that do not have vacuum inlets.

Principles of operation

The basic surgical aspirator consists of a line-powered vacuum pump, a

vacuum regulator and gauge, a collection canister, and an optional

 bacterial filter. Plastic tubing connects these components, completing an

open-ended system that continuously draws tissue debris and fluid from the surgical field to the collection

canister. The gauge allows the user to set a safe limit for suctioning, to assess the performance of the vacuum

pump, and to detect leaks or blockages. Units are either portable or mounted on a stand or cart for mobility.

Vacuum pumps

A single- or double-vacuum pump creates suction within the

system. Double-pump systems operate in synchrony, with both

pumps contributing to the suctioning capacity. The three types of

pumps used in surgical aspirators are rotary vane, diaphragm, and

cylinder piston.

In a rotary-vane pump, a rotor is mounted off-center in relation to

the inside diameter of the pump housing (see Figure 1). As the rotor

turns, sliding vanes that fit into notches in the rotor follow the contour

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 2: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 2/26

Aspirators, Surgical

2 ©2007 ECRI Institute. All Rights Reserved.

Figure 1. Rotary-vane pump

Figure 2. Diaphragm pump

of the housing. Air that is trapped between a set of vanes

is compressed and expelled through the pressure outlet.

As rotation continues, the space between the vanes

expands, and the pressure of the air trapped between the

vanes drops, creating suction. Because of the friction

 between the vanes and the chamber wall and the need to

achieve airtight contact, several rotary-vane models use

self-lubricating graphite vanes. Some units require oil for

lubrication and sealing and have reservoirs that need

refilling after a specified number of hours of operation.

In a diaphragm pump, a mechanical drive flexes a

rubber diaphragm back and forth in a chamber, alternately

increasing and decreasing the chamber volume (see Figure

2). As the chamber volume decreases, air is compressed

and directed out through the one-way exhaust valve; as it

increases, a vacuum draws air into the one-way intake valve. Diaphragm pumps do not need a constant supply of

oil, and only the motor requires lubrication. Some diaphragm pumps require no lubrication.

Surgical aspirators using pistons to create suction operate similarly to diaphragm pumps; however, the pistons

slide in a reciprocating motion within the pump cylinder to alternately increase and decrease chamber volume

and pressure. One-way valves respond similarly to the pressure differences to draw air from the patient end of

the tubing.

Collection canisters

Aspirate passes through the tubing and into the collection canister. The typical collection canister is plastic

with a removable plastic lid that includes a vacuum port with a filter, a patient port to which a suction tube is

connected, and a shutoff valve to automatically close the vacuum port. Manufacturers supply reusable or

disposable canisters and plastic bag liners for reusable canisters. To save time and prevent cross-contamination

among patients and hospital personnel, ports of filled disposable canisters are capped, and the canister andcontents are discarded together.

When large volumes of tissue debris and fluid are removed, aspirate can quickly accumulate in the canister. As

a result, most surgical aspirators use two collection canisters connected in series. The second canister, called a

“trap,” collects the overflow from the first canister,

allowing the operator to postpone emptying it in response

to the rising level of aspirate.

Most surgical aspirators have an overflow-protection

assembly in the cap of the second canister that prevents

large amounts of fluid from overflowing into the pump

and valves. Some units feature a third canister to collect

overflow. Most surgical aspirators use a float mechanismas overflow protection. As aspirate collects in the second

canister, a ball or other mechanical device floats to the top,

covering the opening leading to the suction source and

cutting off suction, thereby alerting the user to the full

collection canister. This overflow mechanism cannot be

totally relied upon to prevent pump contamination; often,

some fluid passes through the tubing to the pump before

the float blocks the opening. As a result, most aspirators

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 3: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 3/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 3

also include a disposable in-line hydrophobic filter between the trap canister and the pump. This filter prevents

fluid or aerosol contaminants from flowing back into the vacuum pump; it can be reused several times, provided

that no overflow has occurred. Although most units have these overflow-protection features, the user should

check the collection canisters frequently and not allow them to fill above the caution line.

Reported problems

Factors affecting pump operation can reduce aspiration capability. Aspirate that overflows beyond the second

canister and filter can enter the pump and clog the valves; a clogged pump can jam and be difficult to restart. A

pump containing aspirated fluid can also be a source of contamination.

Purchase considerations

ECRI Institute recommendations

Included in the accompanying comparison chart are ECRI Institute’s recommendations for minimum

performance requirements for surgical aspirators.

Surgical aspirators should provide moderate to high vacuum and flow rates and suction of at least 500 mm Hgat maximum settings to remove tenacious materials but should also allow lower vacuum levels to help prevent

tissue damage. Vacuum gauges and vacuum-limiting devices (e.g., regulators) are required to aid clinicians in

applying appropriate and safe suction levels, and the former should be both accurate and easy to read. Collection

canisters—whether disposable or reusable—should have a total capacity of at least 2 liters (2,000 milliliters or

cubic centimeters). The aspirator should prevent overflow of aspirated material, which may clog the aspirator and

should be considered infectious. All controls and connections should be clearly labeled, and the device should be

quiet so that it does not interfere with communication among clinicians.

Other considerations

The aspirator should be easy and intuitive to turn on, adjust, and operate. It should provide an appropriate

vacuum limit and free airflow capability. Surgical aspirators typically have vacuum limits of 500 mm Hg andfree airflow capabilities of 25 to 40 L/min. (Free airflow has been defined as the maximum flow of air into the

aspirator with no restriction at the intake.)

If reusable canisters are used, emptying and exchanging the canisters should be uncomplicated and without

significant risk of instrument or operator contamination. Reusable containers are less common because of the risk

of cross-contamination; reusable containers sometimes include disposable liners that minimize this risk.

Discarding disposable canisters and their contents should be easy, with minimal risk of spills or contamination.

The cost of disposables should be weighed against the cost, time, and infection risks associated with the

disposal of infectious waste and disinfection of reusables. Most units with disposable containers can accept

alternate manufacturers’ containers. These standard containers may be more convenient to order and offer the

user a choice of suppliers. However, installed filters supplied with the canisters will differ; some may severely

limit airflow rates.

Cost containment

Because surgical aspirators entail ongoing operational costs, the initial acquisition price does not accurately

reflect the total cost of ownership. Therefore, a purchase decision should be based on issues such as life-cycle cost

(LCC), discount rates and non-price-related benefits offered by the supplier, and standardization with existing

equipment in the department or hospital (i.e., purchasing all aspirators and/or collection canisters from one

supplier).

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 4: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 4/26

Aspirators, Surgical

4 ©2007 ECRI Institute. All Rights Reserved.

An LCC analysis can be used to compare high-cost alternatives and/or to determine the positive or negative

economic value of a single alternative. Because it examines the cash-flow impact of initial acquisition costs and

operating costs over a period of time, LCC analysis is most useful for comparing alternatives with different cash

flows and for revealing the total costs of equipment ownership. One LCC technique—present value (PV)

analysis—is especially useful because it accounts for inflation and for the time value of money (i.e., money

received today is worth more than money received at a later date). Conducting a PV/LCC analysis often

demonstrates that the cost of ownership includes more than just the initial acquisition cost and that a small

increase in initial acquisition cost may produce significant savings in long-term operating costs. The PV is

calculated using the annual cash outflow, the dollar discount factor (the cost of capital), and the lifetime of the

equipment (in years) in a mathematical equation.

The following represents a sample seven-year PV/LCC analysis for a surgical aspirator.

Present Value/Li fe-Cycle Cost An alysis

 Assumptions

  Operating costs are considered for years 1 through 7.

  Dollar discount factor is 6.25%.

  730 procedures are performed per year with single-use disposable collection containers and PVC

(polyvinyl chloride) tubing.

Capital Costs 

  Aspirator unit = $1,200

Total Capital Costs = $1,200

Operating Costs

  Disposable collection containers = $3,000/year

  Disposable tubing = $2,500/yearTotal Operating Costs = $5,500/year

PV = ($36,573)

Other costs not included in the above analysis that should be considered for budgetary planning include those

associated with the following:

  Other disposables and accessories, such as bacterial filters and optional accessories

  Contributions to overhead

As illustrated by the above sample PV/LCC analysis, the initial acquisition cost is only a fraction of the total

cost of operation over seven years. Therefore, rather than making a purchase decision based solely on the

acquisition cost of a surgical aspirator, buyers should consider operating costs over the lifetime of the equipment.

Prices for disposable tubing and catheters vary greatly depending on a number of factors, including type, size,material, manufacturer, and bulk-rate discounts. Many of these disposables can be used with surgical aspirators

from a number of manufacturers; therefore, disposables are usually more relevant to an LCC analysis for an

aspirator than the device itself.

For customized analyses and purchase-decision support, readers should contact ECRI Institute’s SELECT™

Group.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 5: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 5/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 5

Stage of development

Surgical aspiration is a mature and well-accepted technology, although some new innovations have been

made. Surgical waste disposal systems have been developed that do not utilize canisters at all, making waste

collection easier, safer for healthcare workers, and more cost-effective than previous systems. These systems

consist of a rover (a mobile collection tank) and a docking station (a wall unit). After a surgical procedure, the

rover is transported and connected to the docker, which is capable of sterilizing and disposing of the waste(Williamson 1999).

Bibliography

ECRI Institute. Aspirators. Health Devices and Inspection and Preventive Maintenance System. Procedure no. 433.

Williamson J. Purchasing analyst’s cost cutting creates new product. Healthcare Purch News 1999 Jun;23(6):4-5.

Supplier information

ACOMA

Acoma Medical Industry Co Ltd [152410]2-14-14 Hongo Bunkyo-ku

Tokyo 113-0033

 Japan

Phone: 81 (3) 38166911 Fax: 81 (3) 38143845

Internet: http://www.acoma.com 

E-mail: [email protected]  

ANAND

Anand Medicaids Pvt Ltd [287761]

16 Road No 33 Punjabi Bagh (Extension)

New Delhi 110 026

India

Phone: 91 (11) 5160025 Fax: 91 (11) 5451062Internet: http://www.anandind.com  

E-mail: [email protected] 

ATMOS

ATMOS Medica SL [439522]

Calle Torrent d’en Pregaria 27 

Mataro (Barcelona) E-08304

Spain

Phone: 34 (90) 2193580 Fax: 34 (93) 7573283

Internet: http://www.atmosmed.es  

E-mail: [email protected] 

ATMOS Medical AG [439532]

via San Gottardo 81

Lugano/Massagno CH-6900

Switzerland

Phone: 41 (91) 9618040 Fax: 41 (91) 9618049

Internet: http://www.atmosmed.ch  

E-mail: [email protected]  

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 6: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 6/26

Aspirators, Surgical

6 ©2007 ECRI Institute. All Rights Reserved.

ATMOS MedizinTechnik GmbH & Co KG [152463]

Ludwig-Kegel-Strasse 16

Lenzkirch D-79853

Germany

Phone: 49 (7653) 6890 Fax: 49 (7653) 689190

Internet: http://www.atmosmed.de  

E-mail: [email protected] 

ATMOS MedizinTechnik Asia Sdn Bhd [439530]

Menar KLH Suite 2-4-12 4/Fl Business Centre No 2 Jalan Kasipillai

Kuala Lumpur 51200

Malaysia

Phone: 60 (3) 23813885 Fax: 60 (3) 23813889

Internet: http://www.atmosmalaysia.com  

E-mail: [email protected]  

CA-MI

CA-MI snc [309629]

via Ugo La Malfa 31

Pilastro PR I-43010Italy

Phone: 39 (0521) 637133 Fax: 39 (0521) 639041

Internet: http://www.ca-mi.it 

E-mail: [email protected]  

CLEMENTS MEDICAL

Clements Medical Equipment Pty Ltd [305625]

5 Cal Close

Somersby 2250

Australia

Phone: 61 (2) 43402484 Fax: 61 (2) 43405121

Internet: http://www.clemequip.com.au  E-mail: [email protected]  

ESCHMANN EQUIPMENT

Eschmann Equipment [157749]

Peter Road

Lancing BN15 8TJ

England

Phone: 44 (1903) 753322 Fax: 44 (1903) 766793

Internet: http://www.eschmann.co.uk  

E-mail: [email protected]  

GABLER

Gabler Medical (UK) Ltd [192820]Unit 5 Blackwood Hall North Duffield

Selby Y08 5DD

England

Phone: 44 (1757) 289546 Fax: 44 (1757) 282338

Internet: http://www.gablermedical.com  

E-mail: [email protected]  

IMPACT/SORENSON

Impact Instrumentation Inc [101885]

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 7: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 7/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 7

27 Fairfield Pl PO Box 508

West Caldwell, NJ 07006-0508

Phone: (973) 882-1212, (800) 969-0750 Fax: (973) 882-4993

Internet: http://www.impactii.com  

E-mail: [email protected]  

LAMIDEY NOURY

Lamidey Noury Medical [434224]

Zone d’Activite des Godets 3 rue des Petits Ruisseaux 

Verrieres le Buisson F-91370

France

Phone: 33 (1) 69203021 Fax: 33 (1) 60139747

E-mail: [email protected]  

MAQUET

Maquet GmbH & Co KG [305482]

Kehlerstrasse 31

Rastatt D-76437

Germany

Phone: 49 (7222) 9320 Fax: 49 (7222) 932828Internet: http://www.maquet.com 

E-mail: [email protected] 

Maquet (Shanghai) International Trading Co Ltd [439188]

Room 905 Wai Gao Qiao Building No 6 Jilong Road

Shanghai 200336

People’s Republic of China 

Phone: 86 (21) 52081122 Fax: 86 (21) 52081310

Internet: http://www.maquet.com 

E-mail: [email protected]  

MEDELA

Medela Espana [451801]

Calle Manual Fernandez Marquez 49

Badalona E-08918

Spain

Phone: 34 (93) 3205969 Fax: 34 (93) 3205531

Internet: http://www.medela.se  

E-mail: [email protected] 

Medela Inc [103640]

1101 Corporate Dr PO Box 660

McHenry, IL 60051-0660

Phone: (815) 363-1166, (877) 735-1626 Fax: (815) 363-2487, (800) 995-7867

Internet: http://www.medela.com  E-mail: [email protected] 

Medela OOO [451802]

Fersmana Street 5A

Moskva 117312

Russia

Phone: 7 (095) 9806194 Fax: 7 (095) 7181700

Internet: http://www.medela-russia.ru 

E-mail: [email protected] 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 8: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 8/26

Aspirators, Surgical

8 ©2007 ECRI Institute. All Rights Reserved.

Medela AG [156407]

Laettichstrasse 4

Baar/Schweiz CH-6341

Switzerland

Phone: 41 (41) 7695151 Fax: 41 (41) 7695100

Internet: http://www.medela.ch  

E-mail: [email protected] 

MGE

MG Electric (Colchester) Ltd [151092]

Wyncolls Road

Colchester CO4 4HT

England

Phone: 44 (1206) 842244 Fax: 44 (1206) 845849

Internet: http://www.mgeworldwide.com  

E-mail: [email protected]  

NOUVAG

Nouvag AG [170072]

St Gallerstrasse 23-25Goldach CH-9403

Switzerland

Phone: 41 (71) 8466600 Fax: 41 (71) 8453536

Internet: http://www.nouvag.com 

E-mail: [email protected] 

Electronic Services Nouvag USA [226127]

18058 Albyn Ct

Lake Hughes, CA 93532

Phone: (661) 724-0217, (800) 673-7427 Fax: (661) 724-1590

Internet: http://www.nouvag.com 

E-mail: [email protected] ORDISI

Ordisi SA [163027]

Progreso 3 y 5 L’hopitalet de Llobregat

Barcelona E-08904

Spain

Phone: 34 (93) 3340112 Fax: 34 (93) 4402564

Internet: http://www.ordisi.com 

E-mail: [email protected] 

OXYLITRE

Oxylitre Ltd [187944]

Morton House Skerton Road Old TraffordManchester M16 0WJ

England

Phone: 44 (161) 8726322 Fax: 44 (161) 8487914

Internet: http://www.oxylitre.co.uk  

E-mail: [email protected] 

PHIPPS BIRD

Phipps & Bird Inc [101986]

1519 Summit Ave

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 9: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 9/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 9

Richmond, VA 23230-4511

Phone: (804) 254-2737, (800) 955-7621 Fax: (804) 254-2955

Internet: http://www.phippsbird.com  

E-mail: [email protected] 

SAMED

Samed Elettromedicali srl [187040]strada Provinciale 181 N 1/B

Merlino LO I-26833

Italy

Phone: 39 (02) 90658787 Fax: 39 (02) 90658795

Internet: http://www.samedelettromedicali.com  

E-mail: [email protected] 

SIARE

Siare Engineering International Group Srl [152520]

via Giulio Pastore 18

Crespellano BO I-40056

Italy

Phone: 39 (051) 969802 Fax: 39 (051) 969366Internet: http://www.siare.it 

E-mail: [email protected]

SISMATEC

Sismatec Ind Com Equips Hospitalares Ltda [152450]

Rua Omilio Monteiro Soares 1614 Vila Fanny

Curitiba-PR 81030-001

Brazil

Phone: 55 (41) 332 3735 Fax: 55 (41) 332 3735

Internet: http://www.sismatec.com.br  

E-mail: [email protected]  

WELLS JOHNSON

Wells Johnson Co [101872]

8000 S Kolb Rd

Tucson, AZ 85706-9104

Phone: (520) 298-6069, (800) 528-1597 Fax: (520) 885-1189

Internet: http://www.wellsgrp.com  

E-mail: [email protected] 

Note:  The data in the charts derive from suppliers’ specifications and have not been verified through

independent testing by ECRI Institute or any other agency. Because test methods vary, different products’

specifications are not always comparable. Moreover, products and specifications are subject to frequent changes.

ECRI Institute is not responsible for the quality or validity of the information presented or for any adverse

consequences of acting on such information.

When reading the charts, keep in mind that, unless otherwise noted, the list price does not reflect supplier

discounts. And although we try to indicate which features and characteristics are standard and which are not,

some may be optional, at additional cost.

For those models whose prices were supplied to us in currencies other than U.S. dollars, we have also listed the

conversion to U.S. dollars to facilitate comparison among models. However, keep in mind that exchange rates change

often.

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 10: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 10/26

Aspirators, Surgical

10 ©2007 ECRI Institute. All Rights Reserved.

Need to know more?

For further information about the contents of this Product Comparison, contact the HPCS Hotline at +1 (610)

825-6000, ext. 5265; +1 (610) 834-1275 (fax); or [email protected]  (e-mail).

Last updated May 2007

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 11: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 11/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 11

Policy Statement

The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization.

HPCS provides comprehensive information to help healthcare professionals select and purchase diagnostic and

therapeutic capital equipment more effectively in support of improved patient care.

The information in Product Comparisons comes from a number of sources: medical and biomedical

engineering literature, correspondence and discussion with manufacturers and distributors, specifications from

product literature, and ECRI Institute’s Problem Reporting System. While these data are reviewed by qualified

health professionals, they have not been tested by ECRI Institute’s clinical and engineering personnel and are

largely unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the

quality or validity of information derived from outside sources or for any adverse consequences of acting on such

information.

The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison

System does not constitute the endorsement or approval of the product’s quality, performance, or value, or of

claims made for it by the manufacturer. The information and photographs published in Product Comparisons

appear at no charge to manufacturers.

Many of the words or model descriptions appearing in the Healthcare Product Comparison System are

proprietary names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any

name without designation as proprietary should not be regarded as a representation that is not the subject of

proprietary rights.

ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare

Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI

Institute and its employees accept no royalties, gifts, finder’s fees, or commissions from the medical device

industry, nor do they own stock in medical device companies. Employees engage in no private consulting work

for the medical device industry.

About ECRI Institute

ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research

in healthcare to uncover the best approaches to improving patient care. As pioneers in this science for nearly 40

years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.

More than 5,000 healthcare organizations worldwide rely on ECRI Institute’s expertise in patient safety

improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.

ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World

Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and

Quality. For more information, visit http://www.ecri.org. 

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 12: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 12/26

Aspirators, Surgical

12 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL ECRI INSTITUTE'SRECOMMENDEDSPECIFICATIONS1 

ACOMA ACOMA ACOMA

Surgical Aspirators SB-10mkII SB-5 SB-6III

WHERE MARKETED Not specified Not specified Not specifiedFDA CLEARANCE No No NoCE MARK (MDD) No No NoCONFIGURATION Mobile Optional portable MobilePUMP TYPE Piston Rotary vane Piston

Number of pumps 1 1 1VACUUM LIMIT, mm Hg >500 675 (0.09 MPa) 500 675 (0.09 MPa)

Free airflow, L/min >25 40 13.5 40Gauge units  Accurate within 10% MPa mm Hg MPaGauge diam, cm (in) 5.5 (2.2) 5.5 (2.2) 5.5 (2.2)

VACUUM CONTROLMETHOD

Required; regulatorpreferred

Regulating valve Regulating valve Regulating valve

SOUND LEVEL, dBA Not specified 40 @ 1 m (3.3 ft) Not specifiedCOLLECTION CANISTERS Reusable Reusable Reusable

Material Glass Glass GlassNumber 2 1 1

Capacity limit line, cc >2,000 (2 L) 3,000 1,500 3,000Graduations, cc 100 100 100Connector labels Not specified Not specified Not specifiedOverflow protection Required; hydrophobic filter

preferredYes Yes Yes

COMPATIBLE TUBINGMaterial PVC PVC PVCInternal diam, mm (in) 4 (0.16) 4 (0.16) 4 (0.16)Length, m (ft) Not specified Not specified Not specified

FILTER TYPE Not specified Not specified Not specifiedSTANDARDACCESSORIES

None specified None specified None specified

SWITCH TYPE Double switch Not specified Double switchPOWER REQUIREMENTS

VAC 100-240, 50/60 Hz 100/240 100-240, 50/60 HzW or VA 570 VA 170 VA 570 VA

H x W x D, cm (in) 87 x 40 x 35 (34.8 x 16 x14)

30 x 39 x 37 (11.8 x 15.4 x14.6)

89 x 44.5 x 34 (35 x 17.5 x13.4)

WEIGHT, kg (lb) 34 (75) 12 (26) 28 (62)LIST PRICE, Unit Not specified Not specified Not specified

Canisters Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 yearOTHER SPECIFICATIONS None specified. None specified. None specified.LAST UPDATED May 2007 May 2007 May 2007Supplier Footnotes 1These recommendations

are the opinions of ECRIInstitute's technologyexperts. ECRI Instituteassumes no liability fordecisions made based onthis data.

Model FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 13: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 13/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 13

Product Comparison Chart

MODEL ANAND ANAND ATMOS ATMOSEurovac Eurovac-Elite ATMOS C 401 ATMOS S 351

WHERE MARKETED Worldwide Worldwide Worldwide, except USA Worldwide, except USAFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Portable Mobile, stand mounted Mobile, portable, tabletop Mobile, portable, tabletopPUMP TYPE Single diaphragm Double diaphragm 1-cylinder diaphragm 1-cylinder diaphragm

Number of pumps 1 2 1 1VACUUM LIMIT, mm Hg 500 700 682.5 675

Free airflow, L/min 22 35 40 35Gauge units mm Hg, kPa mm Hg, kPa mm Hg, mbar mm Hg, kPa, mbarGauge diam, cm (in) 6.4 (2.5) 6.4 (2.5) 5.3 (2) 4.1 x 6 (1.6 x 2.3) LCD

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve Electronic control

SOUND LEVEL, dBA 50 50 <50 @ 1 m (3.3 ft) <43.9 @ 1 m (3.3 ft)COLLECTION CANISTERS Reusable, collection pouch Reusable Reusable collection pouch Reusable collection pouch

Material Polycarbonate Polycarbonate Polycarbonate, plastic Polycarbonate, plasticNumber 1; 2 optional 2 1-2 1-2Capacity limit line, cc 2,000 2,000 1,000, 1,500, 2,000, 3,000,

5,000

1,500, 2,000, 3,000, 5,000

Graduations, cc 50 50 50 50Connector labels Yes Yes Optional OptionalOverflow protection Yes Yes Optional Optional

COMPATIBLE TUBINGMaterial Silicone, PVC Silicone, PVC Silicone SiliconeInternal diam, mm (in) 8 (0.3) 10 (0.4) 6 (0.25) or 10 (0.39) 6 (0.25) or 10 (0.39)Length, m (ft) 2 (6.6) 2 (6.6) ≥2 (≥6.6)  ≥2 (≥6.6) 

FILTER TYPE Bacterial Bacterial Bacterial, hydrophobic Inline bacterial, hydrophobicbacterial

STANDARDACCESSORIES

Canister, patient tube 2 canisters, patient tube Canister, footswitch, patienttubing, filters, cannulas,tissue collectors, trolley

Canister, footswitch, patienttubing, filters, cannulas,tissue collectors, trolley

SWITCH TYPE Indicator cum switch Push button Pump panel, footswitch Pump panel, footswitchPOWER REQUIREMENTS

VAC 230, 50 Hz 230, 50 Hz 100/115/127/230 115/127/230

W or VA 70 W 150 W 100 W 100 WH x W x D, cm (in) 20 x 42 x 36 (7.9 x 16.5 x

14.2)39 x 55 x 119 (15.3 x 21.6 x46.8)

33 x 24 x 36 (13 x 9.4 x14.1) with secretion jar; 101x 36 x 44 (39.7 x 14.1 x17.3) with trolley

30 x 33 x 20 (11.8 x 13 x7.8) without trolley; 84 x 49x 52 (33 x 19.3 x 20.4) withtrolley

WEIGHT, kg (lb) 7.5 (16.5) 18 (39.7) 6.3 (13.9) without secretion jar

10.2 (22.5) without secretion jar/trolley; 24.7 (55) withtrolley

LIST PRICE, Unit $130 $230 Not specified Not specifiedCanisters $14 $15 Not specified Not specified

WARRANTY 1 year 1 year 2 years 2 yearsOTHER SPECIFICATIONS Meets requirements of ISO

13485:2003.Meets requirements of ISO13485:2003.

Protection category: IPX1;protection class (IEC 601): I;degree of protection: typeBF; class: IIa (Appendix IX,directive 93/42/EEC).

Protection category: IPX0;protection class: I; degree ofprotection: type B; class: IIb(Appendix IX, directive93/42/EEC).

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 14: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 14/26

Aspirators, Surgical

14 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL ATMOS CA-MI CA-MI CA-MIRecord 55 NEW ASKIR 20 NEW ASKIR 230/12V BR NEW ASKIR 30

WHERE MARKETED Worldwide, except USA Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Mobile Portable, tabletop Portable, tabletop Portable, tabletopPUMP TYPE 2-cylinder diaphragm 1-cylinder piston 1-cylinder piston 1-cylinder piston

Number of pumps 1 1 1 1VACUUM LIMIT, mm Hg 735 563 (75 kPa) 564 (75 kPa) 600 (80 kPa)

Free airflow, L/min 55 16 16 40Gauge units mm Hg, mbar kPa, bar kPa, bar kPa, barGauge diam, cm (in) 4.3 (1.7) 5 (2) 5 (2) 5 (2)

VACUUM CONTROLMETHOD

Regulating valve Yes Yes Yes

SOUND LEVEL, dBA <46 @ 1 m (3.3 ft) Not specified Not specified Not specifiedCOLLECTION CANISTERS Reusable collection pouch Reusable, disposable Reusable, disposable Reusable, disposable

Material Polycarbonate, glass,plastic

Plastic Plastic Plastic

Number 1-2 1 1 1

Capacity limit line, cc 1,000, 1,500, 2,000, 3,000,5,000

1,200-2,200 1,200-2,200 1,200-2,200

Graduations, cc 50 100 100 100Connector labels Optional Yes Yes YesOverflow protection Optional Yes Yes Yes

COMPATIBLE TUBINGMaterial Silicone Silicone Silicone SiliconeInternal diam, mm (in) 6 (0.25) or 10 (0.39) 6 (0.24) 6 (0.24) 6 (0.24)Length, m (ft) ≥2 (≥6.6)  1.5 (4.9) 1.5 (4.9) 1.5 (4.9)

FILTER TYPE Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobicSTANDARDACCESSORIES

Canister, footswitch, patienttubing, filters, cannulas,tissue collectors, trolley

Canister, patient tube withfilter, conical connector andprobe

Canister, patient tube withfilter, conical connector andprobe

Canister, patient tube withfilter, conical connector andprobe

SWITCH TYPE Pump panel, footswitch Pump panel Pump panel Pump panelPOWER REQUIREMENTS

VAC 230, optional 100-127 110, 220-240, 50/60 Hz 110, 220-240, 50/60 Hz 110, 220-240, 50/60 Hz

W or VA 100 W 184 VA 50 VA 107 VAH x W x D, cm (in) 94 x 50 x 39 (37 x 19.7 x

15.4)35 x 21 x18 (13.7 x 8.3 x 7) 35 x 21 x 18 (13.7 x 8.3 x 7) 35 x 21 x 18 (13.7 x 8.3 x 7)

WEIGHT, kg (lb) 36 (79) 2.5 (5.5) 4.9 (10.7) 3.6 (7.9)LIST PRICE, Unit  €2,500 (US$3,401)  Not specified Not specified Not specified

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 2 years 2 years 2 years 2 yearsOTHER SPECIFICATIONS Protection category: IPX1;

protection class: I; degree ofprotection: type BF; class:IIa (Appendix IX, directive93/42/EEC).

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

LAST UPDATED May 2007 May 2007 May 2007 May 2007

Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 15: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 15/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 15

Product Comparison Chart

MODEL CA-MI CA-MI CA-MI CA-MINEW ASKIR 30 12V NEW ASKIR C30 NEW HOSPIVAC 350 NEW HOSPIVAC 400

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Portable, tabletop Mobile on trolley with

castorsMobile on castors Mobile on castors

PUMP TYPE Cylinder piston 1-cylinder piston Cylinder diaphragm 2-cylinder diaphragmNumber of pumps 2 1 1 1

VACUUM LIMIT, mm Hg 565 (75 kPa) 600 (80 kPa) 675 (90 kPa) 676 (90 kPa)Free airflow, L/min 25 40 60 90Gauge units kPa, bar kPa, bar kPa, bar kPa, barGauge diam, cm (in) 5 (2) 5 (2) 5 (2) 6 (2.4)

VACUUM CONTROLMETHOD

Yes Yes Yes Yes

SOUND LEVEL, dBA Not specified Not specified Not specified Not specifiedCOLLECTION CANISTERS Reusable, disposable Reusable, disposable Reusable, disposable Reusable, disposable

Material Plastic Plastic Plastic PlasticNumber 1 2 2 2

Capacity limit line, cc 1,200-2,200 2,200 2,200-4,000 2,200-4,000Graduations, cc 100 100 100-200 (for 4,000 capacity) 100-200 (for 4,000 capacity)Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Silicone Silicone Silicone SiliconeInternal diam, mm (in) 6 (0.24) 8 (0.3) 8 (0.3) 8 (0.3)Length, m (ft) 1.5 (4.9) 1.5 (4.9) 1.5 (4.9) 1.5 (4.9)

FILTER TYPE Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobicSTANDARDACCESSORIES

Canister, patient tube withfilter, conical connector andprobe

2 canisters, patient tubewith filter, conical connectorand probe

2 canisters, patient tubewith filter, conical connectorand probe

2 canisters, patient tubewith filter, conical connectorand probe

SWITCH TYPE Pump panel Optional pump panel,footswitch

Optional pump panel,footswitch

Optional pump panel,footswitch

POWER REQUIREMENTSVAC 110, 220-240, 50/60 Hz 110, 220-240, 50/60 Hz 220-240, 50/60 Hz 220-240, 50/60 Hz

W or VA 107 VA 110 VA 230 VA 300 VAH x W x D, cm (in) 35 x 21 x 18 (13.7 x 8.3 x 7) 32 x 99.5 x 30.5 (12.6 x

39.1 x 12) with trolley46 x 60 x 42 (18.1 x 23.6 x16.5)

47 x 47 x 43 (18.5 x 29.1 x16.9)

WEIGHT, kg (lb) 2.5 (5.5) 6.2 (13.6) with trolley 13 (28.5) 20 (43.9)LIST PRICE, Unit Not specified Not specified Not specified Not specified

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 2 years 2 years 2 years 2 yearsOTHER SPECIFICATIONS Disposable collection

systems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

Disposable collectionsystems availableseparately. Meetsrequirements of MDD93/42/EC, ISO 13485:2003,ISO 9001:2001, EN60601/1, EN 60601/1/2, andEN 10079/6.

LAST UPDATED May 2007 May 2007 May 2007 May 2007

Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 16: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 16/26

Aspirators, Surgical

16 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL CLEMENTS MEDICAL CLEMENTS MEDICAL CLEMENTS MEDICAL ESCHMANN EQUIPMENTCompact High-SuctionPump

HandiVac High-SuctionPump

High-Suction Pumps Hi Flo :Hi Vac

VP25

WHERE MARKETED  Africa, Asia, Australasia,Middle East

 Africa, Asia, Australasia,Middle East

 Africa, Asia, Australasia,Middle East

Worldwide

FDA CLEARANCE No No No NoCE MARK (MDD) No No No YesCONFIGURATION Portable Portable Mobile PortablePUMP TYPE Oil-filled rotary vane Diaphragm Twin piston Rotary vane

Number of pumps 1 1 1 1VACUUM LIMIT, mm Hg 720 540 699 : 735 600 (-80 kPa)

Free airflow, L/min 32 30 55 : 32 25Gauge units mm Hg, kPa mm Hg, kPa mm Hg, kPa mm Hg, kPaGauge diam, cm (in) 6 (2.5) 6 (2.5) 6 (2.5) 5 (2)

VACUUM CONTROLMETHOD

Needle valve Needle valve Needle valve Regulating valve

SOUND LEVEL, dBA Not specified Not specified Not specified Not specifiedCOLLECTION CANISTERS Reusable Reusable Reusable Reusable and disposable

with liner system

Material Shatterproof polycarbonate Shatterproof plastic,autoclavable

Shatterproof polycarbonate Disposable liner

Number 1 1 2 1Capacity limit line, cc 2,000 1,000 2,000 2,000Graduations, cc 100 100 100 200Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Silicone, PVC Silicone, PVC Silicone, PVC Antistatic neopreneInternal diam, mm (in) 8 (0.3) 8 (0.3) 8 (0.3) 6 (0.25)Length, m (ft) 2 (6.6) 2 (6.6) 2 (6.6) Not specified

FILTER TYPE Bacterial Bacterial Bacterial Sealed disposablehydrophobic bacterial

STANDARDACCESSORIES

1 canister, tubing, filter,disposable handpiece,optional footswitch

1 canister, tubing, filter,disposable handpiece

2 canisters, tubing, optionalfootswitch

Canister and filter

SWITCH TYPE Not specified Illuminated Illuminated IlluminatedPOWER REQUIREMENTS

VAC 240, 50 Hz 220/240, 50 Hz 220/240, 50 Hz 115/230W or VA 385 W Not specified 360 W Not specified

H x W x D, cm (in) 40 x 23 x 46 (15.7 x 9.1 x18.1)

26 x 18 x 42 (10.2 x 7.1 x16.5)

82 x 46 x 30 (32.3 x 18.1 x11.8)

32 x 32 x 34 (12.6 x 12.6 x13.4)

WEIGHT, kg (lb) 15 (33) 5.5 (12) 20 (45) 9 (19.8)LIST PRICE, Unit Not specified Not specified Not specified Not specified

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 2 years 1 yearOTHER SPECIFICATIONS Optional obstetric cups,

footswitch, handpieces,cannulae, adapters fordisposable canisters.Manufactured to IEC 60601requirements.

Optional obstetric cups,footswitch, handpieces,cannulae, adapters fordisposable canisters.Manufactured to IEC 60601requirements.

Optional obstetric cups,footswitch, handpieces,cannulae, adapters fordisposable canisters.Manufactured to IEC 60601requirements.

None specified.

LAST UPDATED June 2005 June 2005 June 2005 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 17: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 17/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 17

Product Comparison Chart

MODEL ESCHMANN EQUIPMENT ESCHMANN EQUIPMENT GABLER GABLERVP35 VP45 FSE500-110019 FSE550-110019 : FSE570-

110021

WHERE MARKETED Worldwide Worldwide Worldwide, except USA Worldwide, except USAFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Mobile Mobile Portable Portable or mobilePUMP TYPE Rotary vane Rotary vane Piston Piston

Number of pumps 1 1 1 1VACUUM LIMIT, mm Hg 600 (-80 kPa) 700 (-93 kPa) 625 625

Free airflow, L/min 35 40 26 26Gauge units mm Hg, kPa mm Hg, kPa mm Hg, kPa mm Hg, kPaGauge diam, cm (in) 5 (2) 5 (2) 6.3 (2.5) 6.3 (2.5)

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Air-leak valve Air-leak valve

SOUND LEVEL, dBA Not specified Not specified 44 44COLLECTION CANISTERS Reusable and disposable

liner systemReusable and disposableliner system

Reusable Reusable

Material Disposable liner Disposable liner Polysulfone Polysulfone

Number 2 1, 2, or 4 1 2Capacity limit line, cc 2,000 2,000 1,800 1,800Graduations, cc 200 200 100 100Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial  Antistatic neoprene Antistatic neoprene Silicone SiliconeInternal diam, mm (in) 6 (0.25) 6 (0.25) 7 (0.28) 7 (0.28)Length, m (ft) Not specified Not specified Not specified Not specified

FILTER TYPE Sealed disposablehydrophobic bacterial

Sealed disposablehydrophobic bacterial

Bacterial, hydrophobic Bacterial, hydrophobic

STANDARDACCESSORIES

Canister and filter Canister and filter None None

SWITCH TYPE Illuminated Illuminated Illuminated IlluminatedPOWER REQUIREMENTS

VAC 115/230 115/230 220/240, 50 Hz, others

optional

220/240, 50 Hz, others

optionalW or VA Not specified Not specified 75 W 75 W

H x W x D, cm (in) 67 x 44 x 41 (26.4 x 17.3 x16.1)

83 x 41 x 32 (32.7 x 16.1 x12.6)

44 x 22 x 24 (17.3 x 8.7 x9.4)

44 x 29 x 24 (17.3 x 11.4 x9.4)

WEIGHT, kg (lb) 13 (28.6) 32 (70.5) 6.4 (14.1) 7.3 (16.1) : 12 (26.5),including trolley

LIST PRICE, Unit Not specified Not specified Not specified Not specifiedCanisters Not specified Not specified Not specified Not specified

WARRANTY 1 year 1 year 1 year 1 yearOTHER SPECIFICATIONS None specified. None specified. Any liner system can be

fitted as an option; 2 L and4 L suction jars available.

 Any liner system can befitted as an option; 2 L and4 L suction jars available.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 18: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 18/26

Aspirators, Surgical

18 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL GABLER IMPACT/SORENSON IMPACT/SORENSON IMPACT/SORENSONFSE700-110027 1120 Series 2110 Series 2120 Series

WHERE MARKETED Worldwide, except USA Worldwide Worldwide WorldwideFDA CLEARANCE No Yes Yes YesCE MARK (MDD) Yes No No NoCONFIGURATION Mobile Portable Mobile, stand mounted PortablePUMP TYPE Diaphragm Diaphragm Diaphragm Diaphragm

Number of pumps 1 1 1 1VACUUM LIMIT, mm Hg 632 550 550 550

Free airflow, L/min 45 33 (23 with filter) 33 (11 with filter) 33 (23 with filter)Gauge units mm Hg, kPa mm Hg mm Hg mm HgGauge diam, cm (in) 6.3 (2.5) 5 (2) 5 (2) 5 (2)

VACUUM CONTROLMETHOD

 Air-leak valve Regulating valve Regulating valve Regulating valve

SOUND LEVEL, dBA 60 66 @ 1.5 m (5 ft) 62 @ 1.5 m (5 ft) 66 @ 1.5 m (5 ft)COLLECTION CANISTERS Reusable Reusable Reusable Reusable

Material Polysulfone Glass Glass GlassNumber 2 1 1 1Capacity limit line, cc 1,800 950 or 1,900 950, 1,900, or 3,800 950 or 1,900

Graduations, cc 100 50 or 100 50 or 100 50 or 100Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Silicone PVC PVC PVCInternal diam, mm (in) 7 (0.28) 6 (0.25) 6 (0.25) 6 (0.25)Length, m (ft) Not specified 2 (6.6) 2 (6.6) 2 (6.6)

FILTER TYPE Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobicSTANDARDACCESSORIES

None Canister, filter, tubing Canister, filter, tubing Canister, filter, tubing

SWITCH TYPE Illuminated Pump mounted Panel mounted Pump mountedPOWER REQUIREMENTS

VAC 220/240, 50 Hz, othersoptional

115/220-240, 50/60 Hz 115/220-240, 50/60 Hz 115/220-240, 50/60 Hz

W or VA 130 W 483/418 VA 212/330 VA 483/418 VAH x W x D, cm (in) 90 x 45 x 36 (35.4 x 17.7 x

14.1)

36 x 22 x 31 (14 x 19 x 12) 41 x 36 x 86 (16 x 14 x 34) 36 x 22 x 31 (14 x 19 x 12)

WEIGHT, kg (lb) 18.5 (40.8) 9 (19) 10.5 (23) 15.9 (35)LIST PRICE, Unit Not specified $910-1,035 $1,150-1,270 $1,330-1,450

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearOTHER SPECIFICATIONS  Any liner system can be

fitted as an option; 2 L and4 L suction jars available.

None specified. None specified. None specified.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 19: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 19/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 19

Product Comparison Chart

MODEL IMPACT/SORENSON LAMIDEY NOURY MAQUET MAQUET2135 Series AS64R MEDAP BORA UP 2080 MEDAP BORA UP 2080 OP

WHERE MARKETED Worldwide Worldwide Worldwide, except USA Worldwide, except USAFDA CLEARANCE Yes No No NoCE MARK (MDD) No Yes Yes YesCONFIGURATION Mobile, stand mounted Mobile Portable PortablePUMP TYPE Rotary vane Diaphragm Diaphragm Diaphragm

Number of pumps 1 Not specified 1 1VACUUM LIMIT, mm Hg 550 608 675 675

Free airflow, L/min 32 (19 with filter) 60 42 42Gauge units mm Hg mbar mm Hg, kPa mm Hg, kPaGauge diam, cm (in) 5 (2) Not specified 6.4 (2.5) 6.4 (2.5)

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve,membrane type

Regulating valve,membrane type

SOUND LEVEL, dBA 67 @ 1.5 m (5 ft) Not specified 50.5 50.5COLLECTION CANISTERS Reusable Optional disposable Reusable, disposable Reusable, disposable

Material Glass Not specified Glass, polysulfone,disposable liner

Glass, polysulfone,disposable liner

Number 1 2 Up to 4 Up to 4

Capacity limit line, cc 950, 1,900, or 3,800 2,000 1,000-5,000 1,000-5,000Graduations, cc 50 or 100 Not specified 50/100 50/100Connector labels Yes Not specified Yes YesOverflow protection Yes Automatic cutoff, 250 mL

trap bottleDisposable, optionalreusable

Disposable, optionalreusable

COMPATIBLE TUBINGMaterial PVC Not specified Silicone SiliconeInternal diam, mm (in) 6 (0.25) 7 (0.28) 6 (0.24) 6 (0.24)Length, m (ft) 2 (6.6) 2.5 (8.3) 1 (3.3) 1 (3.3)

FILTER TYPE Bacterial, hydrophobic Not specified Bacterial BacterialSTANDARDACCESSORIES

Canister, filter, tubing None specified Canister, overflowprotection, connection tube

Canister, overflowprotection, footswitch,connection tube

SWITCH TYPE Panel mounted Pneumatic pedal Pump panel Pump panel, footswitchPOWER REQUIREMENTS

VAC 115/220-240, 50/60 Hz 220 230 V, 50-60 Hz; 115 V, 60

Hz

230 V, 50-60 Hz; 115 V, 60

HzW or VA 483/528 VA Not specified 160 W 160 W

H x W x D, cm (in) 41 x 36 x 86 (16 x 14 x 34) 43 x 36 x 35 (16.7 x 14.2 x13.8)

31 x 36 x 31 (12.2 x 14.1 x12.2)

31 x 36 x 31 (12.2 x 14.1 x12.2)

WEIGHT, kg (lb) 15.9 (35) 16 (35.3) 11 (24) 11 (24)LIST PRICE, Unit $1,575-1,700  €2,576 (US$3,504)  Not specified Not specified

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 2 years 2 yearsOTHER SPECIFICATIONS None specified. None specified. Meets requirements of ISO

9001, ISO 13485:2003, IEC60601, ISO 10079-1, andCE 0123.

Meets requirements of ISO9001, ISO 13485:2003, IEC60601, ISO 10079-1, andCE 0123.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 20: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 20/26

Aspirators, Surgical

20 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL MAQUET MEDELA MEDELA MGEMEDAP TWISTA SP 1070 Basic 30 Dominant 50 Series SAM 12 : SAM 14

WHERE MARKETED Worldwide, except USA Worldwide Worldwide Worldwide, except North America

FDA CLEARANCE No Yes Yes NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Mobile Mobile or pedestal Mobile Mobile or staticPUMP TYPE Rotary vane 2-cylinder, piston/cylinder,

self-lube2-cylinder, piston/cylinder,self-lube

Diaphragm

Number of pumps 1 1 1 1VACUUM LIMIT, mm Hg 675 638 (85 kPa) 675 (90 kPa) 600

Free airflow, L/min 55 (50 Hz), 65 (60 Hz) 30 50 25Gauge units mm Hg, kPa mm Hg, kPa mm Hg, kPa mm Hg, kPaGauge diam, cm (in) 6.4 (2.5) 6.35 (2.25) 6.35 (2.25) Not specified

VACUUM CONTROLMETHOD

Regulating valve,membrane type

Regulator, membrane type Regulator, membrane type Regulating valve

SOUND LEVEL, dBA 52-54 37 44 <65COLLECTION CANISTERS Reusable, disposable Reusable, disposable Reusable, disposable Reusable, disposable

Material Glass, polysulfone,

disposable liner

Polysulfone, disposable bag

liner

Polystyrene, disposable bag

liner

Plastic, other

Number Up to 4 1 autoclavable 1 autoclavable 1 : 2Capacity limit line, cc 1,000-5,000 1,000-5,000 1,000-5,000 2,000Graduations, cc 50/100 100 100 100Connector labels Yes Yes Yes YesOverflow protection Disposable, optional

reusableYes Yes Yes

COMPATIBLE TUBINGMaterial Silicone Silicone Silicone PVCInternal diam, mm (in) 6 (0.24) 12 x 7 (0.4 x 0.3) 12 x 7 (0.4 x 0.3) 8 (0.3)Length, m (ft) 1 (3.3) 2 (6.6) 2 (6.6) 2 (6.6)

FILTER TYPE Bacterial Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobicSTANDARDACCESSORIES

Canister, overflowprotection, footswitch,connection tube

Canister, tubing Canister, tubing, filters (5) Canisters, footswitch, cart

SWITCH TYPE Pump panel, footswitch Not specified Not specified On/off

POWER REQUIREMENTSVAC 230 V, 50-60 Hz; 100-

110/100-120 V, 50/60 Hz;127 V, 60 Hz

120/240, 50/60 Hz 120/240, 50/60 Hz 220/250 or 110/120, 50/60Hz, single phase

W or VA 100 W Not specified Not specified Not specifiedH x W x D, cm (in) 100 x 56 x 50 (39.3 x 22 x

19.6)97 x 51 x 48 (38 x 20 x 19) 95 x 50 x 59 (37 x 20 x 23) 38 x 45 x 20 (15 x 17.7 x

7.9) : 38 x 65 x 20 (15 x25.6 x 7.9)

WEIGHT, kg (lb) 26 (57) 16 (35) 11.5 (25.4) 10 (22.1) : 11 (24.3)LIST PRICE, Unit Not specified $1,575 portable, $1,995

mobile$2,095 portable; $2,495mobile

£529 (US$1,020) : £645(US$1,243)

Canisters Not specified Not specified Not specified £154.80 (US$298)/pack of 4WARRANTY 2 years 5 years 5 years 1 yearOTHER SPECIFICATIONS Meets requirements of ISO

9001, ISO 13485:2003, IEC60601, ISO 10079-1, andCE 0123.

UL 544 listed. UL 544 listed. Autoclavable collectionvessel supplied, automaticshutoff, overflow controlvalves and internal filter.Meets requirements of BSI,DEMOK, ENISO 9001,GS/TUV, IEC 601-1,NEMKO, SEMKO.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 21: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 21/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 21

Product Comparison Chart

MODEL MGE MGE NOUVAG ORDISISAM 35 SAM 420 Vacuson 60C A-70

WHERE MARKETED Worldwide, except North America

Worldwide, except North America

Worldwide Worldwide

FDA CLEARANCE No No Submitted NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Mobile Portable Portable, stand mounted Mobile standPUMP TYPE Diaphragm Diaphragm Piston, oil free Monobloc, rotating,

submerged in oilNumber of pumps 1 1 1 Not specified

VACUUM LIMIT, mm Hg 600 600 686 (0.9 bar) 720Free airflow, L/min 45 25 60 50Gauge units mm Hg, kPa mm Hg, kPa Bar, mm Hg Not specifiedGauge diam, cm (in) Not specified Not specified Not specified 5 (2)

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve, suctioncontrol valve

Yes

SOUND LEVEL, dBA 55 Not specified 55 50COLLECTION CANISTERS Reusable, disposable Reusable Not specified Reusable

Material Plastic, other Plastic, other Polysulfone bottle, monokit

disposable

Glass

Number 2 1 2 1Capacity limit line, cc 2,000 or 4,000 1,000 5,000 5,000Graduations, cc 100 100 50 1,000Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial PVC PVC Silicone SiliconeInternal diam, mm (in) 8 (0.3) Not specified 8 x 14 (0.3 x 0.6) 8 (0.3)Length, m (ft) 2 (6.6) Not specified 2 (6.6) Not specified

FILTER TYPE Bacterial, hydrophobic Bacterial, hydrophobic Disposable Bacterial, hydrophobicSTANDARDACCESSORIES

Canisters, footswitch, cart None specified Canisters, tubing, stand Switch

SWITCH TYPE On/off On/off Footswitch, on/off electric,foot-controlled vacuum

Footswitch

POWER REQUIREMENTS

VAC 220/240 or 110/120, 50/60Hz, single phase

100-240, 12 VDC, 50/60 Hz 115/230, 50/60 Hz 120/220

W or VA Not specified Not specified 215 W 500 VAH x W x D, cm (in) 84 x [39 or 43] x 33 (33 x

[15.4 or 17] x 13)Not specified 29 x 28 x 36 (11.4 x 11 x

14.2)46 x 111 (18 x 43.7)

WEIGHT, kg (lb) 25 (55.1) 5.5 (12.1) 12 (26.5) 31 (68.4)LIST PRICE, Unit £1,113 (US$2,146) £597 (US$1,151) $2,495 $1,700

Canisters Not specified Not specified $130 for 2 L, $160 for 5 L Not specifiedWARRANTY 1 year 1 year 1 year 1 yearOTHER SPECIFICATIONS  Autoclavable collection

vessel supplied, automaticshutoff, overflow controlvalves and internal filter.Meets requirements of BSI,DEMOK, ENISO 9001,GS/TUV, IEC 601-1,NEMKO, SEMKO.

 Autoclavable collectionvessel supplied, automaticshutoff, overflow controlvalves. Meets requirementsof BSI, DEMOK, ENISO9001, GS/TUV, IEC 601-1,NEMKO, SEMKO.

None specified. Meets requirements of CE0318 and ISO 9002.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 22: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 22/26

Aspirators, Surgical

22 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL ORDISI ORDISI ORDISI ORDISI AP-92 Aspit Futurvac FV-LIP

WHERE MARKETED Worldwide Worldwide Worldwide WorldwideFDA CLEARANCE No No No NoCE MARK (MDD) Yes Yes Yes YesCONFIGURATION Portable Portable Mobile stand Mobile standPUMP TYPE Membrane Membrane Monobloc, rotating,

submerged in oilMonobloc, rotating,submerged in oil

Number of pumps 1 2 Not specified Not specifiedVACUUM LIMIT, mm Hg 620 600 720 758

Free airflow, L/min 10 13.5 50 166Gauge units Not specified Not specified Not specified Not specifiedGauge diam, cm (in) 5 (2) 5 (2) 7 (2.8) 7 (2.8)

VACUUM CONTROLMETHOD

Yes Yes Yes Yes

SOUND LEVEL, dBA 30 30 50 50COLLECTION CANISTERS Reusable Reusable Reusable Reusable

Material Plastic Glass Plastic PlasticNumber 1 1 2-3 2-3

Capacity limit line, cc 1,000 1,600 1,000-3,000 1,000-3,000Graduations, cc 100 200 500 500Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Silicone Silicone Silicone SiliconeInternal diam, mm (in) 8 (0.3) 8 (0.3) 8 (0.3) 8 (0.3)Length, m (ft) Not specified Not specified Not specified Not specified

FILTER TYPE Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobic Bacterial, hydrophobicSTANDARDACCESSORIES

Switch Switch Switch Switch

SWITCH TYPE Footswitch Footswitch Footswitch FootswitchPOWER REQUIREMENTS

VAC 120/220 120/220 120/220 120/220W or VA 80 VA 115 VA 500 VA 650 VA

H x W x D, cm (in) 34 x 15 x 30 (13.4 x 5.9 x

11.8)

34 x 15 x 30 (13.4 x 5.9 x

11.8)

50 x 43 x 80 (19.7 x 16.9 x

31.5)

50 x 43 x 80 (19.7 x 16.9 x

31.5)WEIGHT, kg (lb) 1.4 (3.5) 4.6 (10.2) 22.1 (48.7) 30 (66.2)LIST PRICE, Unit $465 $595 $1,600-1,700 $2,490

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearOTHER SPECIFICATIONS Meets requirements of CE

0318, EN 60601, and ISO9002.

Meets requirements of CE0318, EN 60601, and ISO9002.

Meets requirements of CE0318 and ISO 9002.

Meets requirements of CE0318 and ISO 9002.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 23: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 23/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 23

Product Comparison Chart

MODEL OXYLITRE OXYLITRE OXYLITRE PHIPPS & BIRD1 elite HIGH ESS100 : eliteHIGH ESS200

elite LOW ESS050 elite THEATRE ESS5200 AUL Aspirator

WHERE MARKETED Worldwide, except USA Worldwide, except USA Worldwide, except USA USAFDA CLEARANCE No No No YesCE MARK (MDD) Yes Yes Yes NoCONFIGURATION Mobile Mobile Mobile Portable, mobile cart

availablePUMP TYPE Diaphragm Diaphragm Diaphragm Cylinder piston/diaphragm

Number of pumps 1 1 2 1VACUUM LIMIT, mm Hg 620 620 620 560

Free airflow, L/min 25 25 55 Not specifiedGauge units mm Hg, kPa mm Hg, kPa mm Hg, kPa cm HgGauge diam, cm (in) 6.3 (2.5) 6.3 (2.5) 6.3 (2.5) 5 (2)

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve Regulating valve

SOUND LEVEL, dBA Not specified Not specified Not specified Not specifiedCOLLECTION CANISTERS Reusable, disposable Reusable, disposable Reusable, disposable Reusable; accepts most

brands of disposable

suction canistersMaterial Plastic Plastic Plastic GlassNumber 1 : 2 1 2 1Capacity limit line, cc 2,000 2,000 2,000 900Graduations, cc 200 200 200 250Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Polyethylene (FR) Polyethylene (FR) Polyethylene (FR) Vinyl supplied; compatible

with other typesInternal diam, mm (in) 8 (0.3) 8 (0.3) 8 (0.3) 4.8 (0.19)Length, m (ft) 1 (3.3) 1 (3.3) 1 (3.3) ~1 (3.3)

FILTER TYPE Sealed disposablehydrophobic

Sealed disposablehydrophobic

Sealed disposablehydrophobic

Hydrophobic

STANDARDACCESSORIES

Tubing Tubing Tubing None specified

SWITCH TYPE Illuminated Illuminated Illuminated TogglePOWER REQUIREMENTS

VAC 220-240 220-240 220-240 120 (1/10 Hp motor)W or VA 127 W 127 W 170 W Not specified

H x W x D, cm (in) 84 x 44 x 34 (33 x 17 x 13) 84 x 44 x 34 (33 x 17 x 13) 84 x 44 x 34 (33 x 17 x 13) 27.9 x 20.3 x 35.6 (11 x 8 x14)

WEIGHT, kg (lb) 13 (28.7) 13 (28.7) 15.5 (33) 7.2 (16)LIST PRICE, Unit Not specified Not specified Not specified Not specified

Canisters Not specified Not specified Not specified Not specifiedWARRANTY 1 year 1 year 1 year 1 yearOTHER SPECIFICATIONS Meets requirements of BS

EN 46001 and 93/42/EEC,BS EN ISO 9001, and ISO10079-1.

Meets requirements of BSEN 46001 and 93/42/EEC,BS EN ISO 9001, and ISO10079-1.

Meets requirements of BSEN 46001 and 93/42/EEC,BS EN ISO 9001, and ISO10079-1.

Hospital-grade plug; ratedfor 1 year of continuousoperation. UL listed.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier Footnotes 1 All models sold under a

variety of private labelagreements.Model FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 24: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 24/26

Aspirators, Surgical

24 ©2007 ECRI Institute. All Rights Reserved.

Product Comparison Chart

MODEL PHIPPS & BIRD1  PHIPPS & BIRD1  SAMED SAMEDCAULCompressor/Aspirator

DCA Aspirator/Compressor MOD CH1 MOD CH2

WHERE MARKETED USA USA Worldwide WorldwideFDA CLEARANCE Yes Yes No NoCE MARK (MDD) No No Yes YesCONFIGURATION Portable, mobile cart

availablePortable Mobile on 4 antistatic

castersMobile on 4 antistaticcasters

PUMP TYPE Cylinder piston/diaphragm Cylinder piston/diaphragm 1-cylinder piston 2-cylinder pistonNumber of pumps 1 1 1 1

VACUUM LIMIT, mm Hg 560 560 650 650Free airflow, L/min Not specified Not specified 50 75Gauge units cm Hg cm Hg cm Hg cm HgGauge diam, cm (in) 5 (2) 5 (2) 48 x 48 (18.9 x 18.9),

external48 x 48 (18.9 x 18.9),external

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve Regulating valve

SOUND LEVEL, dBA Not specified Not specified <50 @ 1 m (3.3 ft) <50 @ 1 m (3.3 ft)COLLECTION CANISTERS Reusable; accepts most

brands of disposablesuction canisters

Disposable Reusable Reusable

Material Glass Containers sold separately Polycarbonate(autoclavable)

Polycarbonate(autoclavable)

Number 1 Not specified 1 2Capacity limit line, cc 900 Not specified 2,000; optional 4,000 2,000; optional 4,000Graduations, cc 250 Not specified 50 50Connector labels Yes Yes Yes YesOverflow protection Yes Yes Yes Yes

COMPATIBLE TUBINGMaterial Vinyl supplied; compatible

with other typesVinyl supplied; compatiblewith other types

Silicone Silicone

Internal diam, mm (in) 4.8 (0.19) 4.8 (0.19) 8 (0.3) 8 (0.3)Length, m (ft) ~1 (3.3) ~1 (3.3) 1.8 (5.9) 1.8 (5.9)

FILTER TYPE Hydrophobic Hydrophobic Bacterial BacterialSTANDARD

ACCESSORIES

0-40 psi pressure gauge,

air-filter regulation valve

0-38 psi pressure gauge,

air-filter regulation valve

Canister, tubing, footswitch,

cable, bacterial filter

2 canisters, tubing,

footswitch, cable, 2 bacterialfiltersSWITCH TYPE Toggle Toggle Manual/pneumatic Manual/pneumaticPOWER REQUIREMENTS

VAC 120 (1/10 Hp motor) 12 VDC 220-240 220-240W or VA Not specified Not specified 155 W (engine), 185 W

(suction)370 VA (engine), 400 VA(suction)

H x W x D, cm (in) 27.9 x 20.3 x 35.6 (11 x 8 x14)

17.8 x 22.9 x 35.6 (7 x 9 x14)

93 x 38 x 30 (36.6 x 15 x11.8)

93 x 50 x 30 (36.6 x 19.7 x11.8)

WEIGHT, kg (lb) 8.2 (18) 3.2 (7) 21 (46) 28 (62)LIST PRICE, Unit Not specified Not specified  €1,150 (US$1,384) : €1,250

(US$1,504) with two 4 LMakrolon jars

 €1,400 (US$1,684) : €1,550(US$1,865) with two 4 LMakrolon jars

Canisters Not specified Not specified  €48 (US$58) 2 L canister    €48 (US$58) 2 L canister  WARRANTY 1 year 1 year 1 year, parts except for

batteries and consumables1 year, parts except forbatteries and consumables

OTHER SPECIFICATIONS Hospital-grade plug; ratedfor 1 year of continuousoperation. UL listed.

Optional battery pack andbattery charger.

None specified. None specified.

LAST UPDATED May 2007 May 2007 May 2007 May 2007Supplier Footnotes 1 All models sold under a

variety of private labelagreements.

1 All models sold under avariety of private labelagreements.

Model FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 25: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 25/26

  Aspirators, Surgical

©2007 ECRI Institute. All Rights Reserved 25

Product Comparison Chart

MODEL SAMED SIARE SISMATEC WELLS JOHNSONPORTABLE UNIT DC 1000/C ASPIROTEC II HERCULES

WHERE MARKETED Worldwide Worldwide Brazil, South America WorldwideFDA CLEARANCE No No No YesCE MARK (MDD) Yes Yes No NoCONFIGURATION Portable, carrying case Mobile Mobile MobilePUMP TYPE 1-cylinder piston Rotary Piston 2-cylinder piston

Number of pumps 1 1 1 3VACUUM LIMIT, mm Hg 580 760 635 760 @ sea level

Free airflow, L/min 13 50 35 192.6Gauge units cm Hg cm Hg mm Hg, pol Hg Not specifiedGauge diam, cm (in) 48 x 48 (18.9 x 18.9),

externalNot specified 6.4 (2.5) Not specified

VACUUM CONTROLMETHOD

Regulating valve Regulating valve Regulating valve Not specified

SOUND LEVEL, dBA <50 @ 1 m (3.3 ft) 40 maximum 51 53 ±3 @ 1.8 m (6 ft)COLLECTION CANISTERS Reusable Reusable or disposable Reusable, autoclavable Reusable, disposable

Material Polycarbonate(autoclavable)

Plastic Glass Not specified

Number 1 2 1, 2, 3 Not specifiedCapacity limit line, cc 1,000 4,000, 2,000 3,000, 4,000, 5,000 Not specifiedGraduations, cc 50 100 250, 500, 1,000, 2,000,

3,000, 4,000, 5,000Not specified

Connector labels Yes Yes No Not specifiedOverflow protection Yes Yes Yes Trap

COMPATIBLE TUBINGMaterial Silicone Silicone Silicone SiliconeInternal diam, mm (in) 8 (0.3) 7 (0.28) 8 (0.3) Not specifiedLength, m (ft) 1.8 (5.9) 1.5 (4.9) 1.5 (4.96) 4.25 (9)

FILTER TYPE Bacterial Bacterial Optional bacterial Bacterial, HEPASTANDARDACCESSORIES

Canister, bacterial filter,cable

Canisters Glass jar, silicone tubing,optional suction cannulae

None specified

SWITCH TYPE Continuous operation Optional foot remote control On/off, knob, optionalfootswitch

Footswitch

POWER REQUIREMENTS

VAC 220/230, battery 12 VDC 230; 115 optional 110/220, 50/60 Hz 110/220W or VA 25 W 200 VA 250 W Not specified

H x W x D, cm (in) 29 x 40 x 14 (11.4 x 15.7 x5.5)

85 x 57 x 54 (33.4 x 22.4 x21.2)

32 x 35 x 91 (12.5 x 13.7 x37.4)

47 x 45.7 x 38.1 (18.5 x 18x 15)

WEIGHT, kg (lb) 5 (11) 20 (44.1) 15 (40) 33 (73)LIST PRICE, Unit  €480 (US$578)  Not specified $928 Not specified

Canisters  €41 (US$49) 1 L canister   Not specified $15 Not specifiedWARRANTY 1 year, parts except for

batteries and consumables1 year 1 year 5 years, parts and labor

OTHER SPECIFICATIONS Includes carrying case. 100 mL safety canister. Meets requirements of ISO9001:2000.

None specified.

LAST UPDATED May 2007 May 2007 May 2007 April 2000Supplier FootnotesModel FootnotesData Footnotes

Comprehensive of Biomedical Engineering Site (www.dezmed.com

  W

  W  W .  D

  E  Z  M  E  D

 .  C  O  M

Page 26: Aspirators surgical.PDF

8/9/2019 Aspirators surgical.PDF

http://slidepdf.com/reader/full/aspirators-surgicalpdf 26/26

Aspirators, Surgical

Product Comparison Chart

MODEL WELLS JOHNSONWHISPERATOR

WHERE MARKETED WorldwideFDA CLEARANCE YesCE MARK (MDD) NoCONFIGURATION MobilePUMP TYPE 2-cylinder piston

Number of pumps 2VACUUM LIMIT, mm Hg 760 @ sea level

Free airflow, L/min 147.2Gauge units Not specifiedGauge diam, cm (in) Not specified

VACUUM CONTROLMETHOD

Not specified

SOUND LEVEL, dBA 53 ±3 @ 1.8 m (6 ft)COLLECTION CANISTERS Reusable, disposable

Material Not specifiedNumber Not specifiedCapacity limit line, cc Not specified

Graduations, cc Not specifiedConnector labels Not specifiedOverflow protection Trap

COMPATIBLE TUBINGMaterial SiliconeInternal diam, mm (in) Not specifiedLength, m (ft) 4.25 (9)

FILTER TYPE Bacterial, HEPASTANDARDACCESSORIES

None specified

SWITCH TYPE FootswitchPOWER REQUIREMENTS

VAC 110/220W or VA Not specified

H x W x D, cm (in) 47 x 45.7 x 38.1 (18.5 x 18x 15)

WEIGHT, kg (lb) 29 (64)LIST PRICE, Unit Not specified

Canisters Not specifiedWARRANTY 5 years, parts and laborOTHER SPECIFICATIONS None specified.LAST UPDATED  April 2000Supplier FootnotesModel FootnotesData Footnotes