Assessing the performance of SICCAS-type lead tungstate scintillators for EM
calorimetry in the CLAS12 Forward Tagger at Jefferson Lab
S.Fegan a,∗, E.Auffrayb, M.Battaglieria, E.Buchananc, B.Caiffia, A.Celentanoa, L.Colanerid, A.D’Angelod, R.De Vitaa,
V.Dormeneve, E.Fanchinia, L.Lanzad, R.W.Novotnye, F.Parodia, A.Rizzod, D.Sokhanc, I.Tarasovf, I.Zontad
aIstituto Nazionale di Fisica Nucleare, Sezione di Genova and Dipartimento di Fisica dell’Universitá, Via Dodecaneso 33, 16146 Genova, ItalybCERN, European Organisation for Nuclear Research, Geneva, Switzerland
cUniversity of Glasgow, Glasgow, G12 8QQ, United KingdomdIstituto Nazionale di Fisica Nucleare, Sezione Roma2 Tor Vergata and Università degli studi di Roma Tor Vergata, Via Scientifica 1, 00133, Roma, Italy
eII. Physikalisches Institut, Universität Gießen, 35392, Gießen, GermanyfGSI - Helmholtzzentrum fur Schwerionenforschung GmbH, Germany
Abstract
In this paper we describe a program of assessment carried out to determine the performance under irradiation of Lead Tungstate
crystals acquired from the Shanghai Institute of Ceramics, Chinese Academy of Sciences for use in the Forward Tagger device,
part of the CLAS12 detector in Hall B at Jefferson Lab. Obtaining crystals matching the demanding specifications of the Forward
Tagger has been a significant challenge, and the tests performed are intended to maximise the performance of the detector within
the practicalities of the crystal manufacturing process.
Results of light transmission, before and after gamma ray irradiation, are presented, and used to calculate dk, the induced radiation
absorption coefficient, at 420 nm, the peak of the Lead Tungstate emission spectrum. These results are compared with identical
measurements carried out on BTCP (Bogoriditsk Technical Chemical Plant) crystals, previously used in several calorimeter devices
but no longer commercially available.
Also presented are a series of tests performed to determine the feasibility of recovering radiation damage to the crystals using
illumination from an LED, with such illumination available in the Forward Tagger from a light monitoring system integral to the
detector.
Key words: Electromagnetic calorimetry, Lead tungstate scintillators, Radiation hardness, Optical transmission, Stimulated
recovery of radiation damage
PACS: 29.40.Vj, 61.80.Ed
1. Introduction1
Following the closure of the Bogoriditsk Technical Chemi-2
cal Plant (BTCP) in Russia, the Shanghai Institute of Ceramics,3
Chinese Academy of Sciences (SICCAS) is one of the most4
practical remaining facilities for the large-scale production of5
Lead Tungstate (PbWO4) crystals. This type of scintillator has6
been used for EM calorimetry in a variety of experimental fa-7
cilities [1, 2, 3, 4], including the Forward Tagger Calorimeter8
(FT-Cal), a subsystem of the Forward Tagger device, part of the9
CLAS12 facility being constructed in Hall B at Jefferson Lab10
[5].11
The Forward Tagger has been developed for meson spec-12
troscopy experiments in CLAS12 using the technique of low Q213
electron scattering. These electrons give rise to quasi-real pho-14
tons, which are reconstructed by detecting the scattered elec-15
tron in the Forward Tagger between polar angles of 2.5◦ and16
4.5◦. At such close proximity to the beamline, the FT-Cal scin-17
tillators will be subjected to radiation rates as high as 5 rad/h,18
∗Corresponding author
Email addresses: [email protected] (S.Fegan )
requiring a sufficiently radiation-hard material to be used. Lead19
Tungstate was chosen as it has been shown in many studies to20
be a material very resistant to radiation damage [6]. Combined21
with its fast decay time and small radiation length, PbWO4 is22
considered to be a good match to the demanding specifications23
of the FT-Cal [5].24
The FT-Cal comprises 332 PbWO4 crystals, each measuring25
200 × 15 × 15 mm, produced by SICCAS using the modified26
Bridgman method [7]. These crystals are read out with in-27
dividual Avalanche Photo Diodes (APD), whose gains can be28
matched, and monitored during run periods, via a light mon-29
itoring system, which can provide LED illumination tuned to30
the luminescence spectrum of the scintillator. The initial speci-31
fications demanded of the FT-Cal crystals are outlined in Table32
1.33
Radiation hardness of the crystals is quantified by the radia-34
tion induced absorption coefficient, dk, given in equation 1,35
dk =1
lengthln
(
Tbe f
Tirr
)
(1)
where Tbe f is the light transmission at 420 nm, the peak of the36
PbWO4 emission spectrum, measured before irradiation, and37
Preprint submitted to Elsevier October 1, 2014
Property Value
Length (mm) 200.00 ±0.15
Width (mm) 15.00 ±0.15
Height (mm) 15.00 ±0.15
Longitudinal Transmission (360 nm) ≥25%
Longitudinal Transmission (420 nm) ≥60%
Longitudinal Transmission (620 nm) ≥70%
Light Yield (T = 18◦C) ≥13 phe/MeV
Radiation Hardness
(Induced absorption coefficient at
420 nm, dk)
≤1.0m−1
Table 1: Initial specifications requested from SICCAS for the PbWO4 crystals
of the FT-Cal
Tirr the light transmission at 420 nm after irradiation. Crystals38
exhibiting greater levels of radiation damage to light transmis-39
sion have higher values of dk.40
As a consequence of the practicalities of the manufacturing41
process, and the need to produce crystals in a timely and cost-42
effective manner, the crystal properties are often found to be43
outside the ranges specified in Table 1, and a further programme44
of quality control is required to assess whether their character-45
istics are within acceptable values.46
This paper describes the programme of evaluation under-47
taken to assess the SICCAS crystals acquired for the FT-Cal in48
terms of the light transmission and radiation hardness require-49
ments of operations within CLAS12. The program consists of50
three parts. The first is a study of the light transmission, verify-51
ing that, before irradiation, crystals possess light transmission52
values consistent with those specified in Table 1. The second53
part assesses the radiation hardness of the crystals by determin-54
ing the induced radiation absorption coefficient, dk, measuring55
the light transmission before and after irradiation and calculat-56
ing dk using equation 1. The third part is the assessment of the57
ability to recover radiation damage to the crystals by means of58
optical annealing using visible light illumination from an LED,59
exploiting the availability of such illumination from the light60
monitoring system of the FT-Cal.61
A total of 370 crystals, including spares, were initially ac-62
quired from SICCAS, and for each crystal the induced absorp-63
tion coefficient was measured. The results were used to reject64
51 crystals not suitable for deployment in the FT-Cal. These re-65
jected crystals were replaced by SICCAS, and subjected to the66
same irradiation procedure to find dk. The 332 most radiation-67
resistant crystals were then selected for installation, with the68
remainder to be held as spares. LED recovery measurements69
were performed for a smaller subset of these crystals, verifying70
the suitability of the light monitoring system for use in crystal71
annealing.72
The SICCAS crystals were also compared with sample73
BTCP crystals; a set of 8 crystals used for prototyping of the74
FT-Cal [5] and a single crystal of a slightly different geometry75
(160 mm long, tapering from 16× 16 mm at one end to 13× 1376
mm at the other), deployed in several previous calorimeter de-77
vices at Jefferson Lab, including the ECal subsystem of the78
Heavy Photon Search (HPS) experiment [8]. The BTCP exam-79
ples chosen represent a typical range of properties and response80
to irradiation of this type of crystal.81
2. Crystal irradiation tests82
The use of PbWO4 crystals as scintillators in EM calorime-83
ters has been extensively studied for a variety of experiments,84
including CMS and ALICE at CERN [1, 2], PANDA at FAIR85
[3], and the CLAS Inner Calorimeter (CLAS-IC) at Jefferson86
Lab [4]. As a result of these studies, a good understanding ex-87
ists of the mechanisms of radiation damage in PbWO4 crystals,88
and how this damage manifests in terms of crystal light yield89
[9]. Under irradiation, impurities and defects in the crystal90
structure, and traps for electrons and holes lead to the forma-91
tion of colour centres, the overall effect of which is a decrease92
in light transmission of the crystal at optical wavelengths. Af-93
ter exposure, these colour centres spontaneously relax, with a94
fast thermal component acting over a timescale of around 3095
minutes [11]. This fast recovery is demonstrated in figure 1, in96
the form of dk measurements at 420 nm performed on a sample97
SICCAS crystal immediately after irradiation. Following this98
fast recovery, crystals continue to recover on a slower timescale,99
and their light transmission is fully restored after several weeks100
at room temperature.101
Time since irradiation (minutes)0 5 10 15 20 25 30
)-1
dk (
m
1
1.01
1.02
1.03
1.04
1.05
1.06
1.07
1.08
1.09
1.1
Figure 1: Radiation induced absorption coefficient, dk, at 420 nm for a sample
SICCAS-type PbWO4 crystal, measured at three-minute intervals following ir-
radiation with 30 Gy. After around 30 minutes, the fast recovery component is
complete.
Although the processes of radiation damage in PbWO4 are102
well-known, the applications cited above used BTCP-type103
crystals, which are no-longer commercially available. The104
SICCAS-type crystals which will be used for the FT-Cal exhibit105
similar behaviour under irradiation to BTCP crystals, but there106
are appreciable differences with respect to the BTCP-type.107
To understand these differences, and any potential effects on108
the deployment of these crystals in the FT-Cal, a preliminary109
programme of irradiation studies was performed at CERN, us-110
ing the Automatic Crystal quality Control System (ACCOS)111
2
[12] on a subset of the crystals acquired from SICCAS. These112
studies showed significant variations of the crystal parameters,113
with several crystals failing to meet the radiadion hardness114
specifications shown in Table 1. For these reasons, a detailed115
assay of the performance under irradiation of all the crystals116
acquired for the FT-Cal was undertaken.117
To observe the effects of normal operation in CLAS12 on118
the FT-Cal crystals, a 30 Gy dose was chosen, equivalent to119
the dose expected in the region of the FT over one month of120
continuous operation in CLAS12 [5].121
2.1. The irradiation facility at Gießen University122
The biotechnical operating unit of the University of Gießen123
has an irradiation facility, the Strahlenzentrum, used by re-124
searchers in a variety of disciplines. The facility has 60Co125
sources, which can be used in gamma ray irradiation studies126
and have been utilised by the II Physikalisches Institut of the127
university in studies of PbWO4 crystals for the PANDA EM128
calorimeter.129
The irradiation station at the Strahlenzentrum used for crystal130
studies is equipped with a Cary 4000 spectrophotometer, featur-131
ing a customised enclosure into which crystals are loaded for132
measurements of the light transmission along their longitudi-133
nal axis. The spectrophotometer is controlled by a PC running134
Cary WinUV software that allows the viewing and saving of135
spectra and provides options for the export of data to various136
formats for offline analysis. This is located close to the irradia-137
tion chamber, into which the crystals are placed for irradiation.138
The irradiation chamber contains 6 volume sources, with super-139
position between the sources enabling irradiation of the entire140
crystal sample. Any position effects regarding dose imparted141
within the chamber are considered to be negligible, this was142
seen in multiple irradiation cycles of FT crystals, and in previ-143
ous work by the PANDA collaboration at the Strahlenzentrum.144
A 30 Gy dose is imparted to the crystals in approximately 15145
minutes, representing a rate of 120 Gy/hr.146
2.2. Crystal testing procedure147
Before irradiation, measurements were made of the longi-148
tudinal light transmission of the crystals. Crystals were then149
irradiated with a 30 Gy dose from the 60Co sources and stored150
for 30 minutes in a dark environment to allow the fast recov-151
ery component of the crystal to take effect without interference152
from optically-stimulated recovery effects. They are then mea-153
sured a second time, with the same alignment and orientation154
in the spectrophotometer as the initial measurement in order to155
minimise changes in the light transmission due to crystal posi-156
tioning and the presence of dust, dirt, and imperfections on the157
crystal surface. All 370 SICCAS crystals, the 51 replacement158
crystals, and the BTCP crystals from both the HPS ECal and159
the FT-Cal prototype, were measured in this way at least once.160
2.3. Further studies of crystal properties161
In addition to the “once-through” irradiation procedure de-162
scribed above, several crystals were used for more detailed163
studies. This involved exposure to multiples of the 30 Gy dose,164
as well as repeat measurements of the “once-through” irradia-165
tion procedure after thermal annealing of the crystals, the latter166
being performed as a consistency check of the reproducibility of167
measurements. A separate facility at the University of Gießen168
was also used to study crystal light yield, in order to verify AC-169
COS results obtained during the preliminary crystal assessment170
procedure referred to earlier in this section. No significant vari-171
ation in the light yield measurements was seen in comparison172
to the ACCOS studies, or the quoted values from the manu-173
facturer, and all crystals tested in this work for radiation hard-174
ness are considered to have met the light yield requirements of175
the FT-Cal. Also performed were tests of recovery of radiation176
damage via optical annealing, discussed in the next section.177
3. Recovery of radiation damage by LED illumination178
Except in cases of prolonged high doses, for example in high-179
energy collider experiments such as CMS, radiation damage180
to PbWO4 crystals will typically recover on its own at room181
temperature over the characteristic timescale of several weeks.182
The colour centres formed around defects in the crystal release183
their captured electrons and/or holes, and again the crystal ex-184
hibits the same light transmission properties it did before irradi-185
ation. Crystals can also be recovered by thermal annealing, as186
the probability of releasing charge carriers from colour centres187
depends exponentially on temperature, allowing the recovery188
process to be performed in a period of several hours at temper-189
atures of around 200◦C.190
In addition to these processes, illumination from visible light191
is able to induce crystal recovery, by optically-stimulating the192
restoration of the colour centres in the crystals [10, 11]. Dur-193
ing the program of crystal tests performed in Gießen, irradiated194
examples of PbWO4 crystals of the SICCAS and BTCP types195
were exposed to varying degrees of illumination, both from196
LEDs planned for use in the gain monitoring system of the FT-197
Cal, and from readily-available sources of light in the Strahlen-198
zentrum, including fluorescent and incandescent lamps. The199
results of these studies using the LEDs from the FT-Cal mon-200
itoring system on the SICCAS and BTCP PbWO4 crystals is201
shown in section 4.2.202
4. Analysis and Results203
Light transmission measurements before and after irradi-204
ation, as well as all multiply-irradiated, annealed and re-205
irradiated crystals, and LED illumination studies were saved206
as .csv text files from the spectrophotometer control software207
and subjected to offline analysis. This analysis collated light208
transmission measurements at three wavelengths, 360 nm, 420209
nm and 620 nm, computed dk at 420 nm for all irradiated crys-210
tals, and examined the LED-stimulated recovery properties of211
the SICCAS and BTCP crystals.212
Also considered were possible systematic effects, and data213
normalisation issues. The spectrophotometer is automatically214
set up to subtract baseline measurements, a reference measure-215
ment taken with no crystal sample installed and no illumination216
3
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
20
40
60
80
After
Before
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
20
40
60
80Before
After
Figure 2: Longitudinal Light Transmission for SICCAS-type PbWO4 crystals,
before and after 30 Gy gamma irradiation. The irradiated spectra exhibits lower
light transmission, particularly around the 400 - 600 nm region. The plot on the
right represents good response of the SICCAS crystal to irradiation, while the
plot on the left is typical of a very poorly performing crystal.
of the photodetector within the device. Changes in this baseline217
are primarily due to thermal fluctuations, and are mitigated by218
maintaining a stable ambient temperature in the Strahlenzen-219
trum and regularly updating the baseline spectrum used, par-220
ticularly after longer periods where the spectrophotometer was221
not in use. The reproducibility of crystal positioning is the other222
major systematic effect in these measurements, with even small223
misalignments of the crystal having an appreciable effect on the224
measured values of light transmission. All crystals had a small225
identifying mark on one surface of the crystal, and this was used226
to ensure that every crystal had the same orientation each time it227
was measured in the spectrophotometer. Several crystals were228
measured multiple times and their spectra compared, in order229
to ensure that crystal positioning was sufficiently reproducible230
and would have minimal effect on the spectra measured. The231
total systematic error in these measurements is estimated to be232
1-2%, resulting in an error in dk between 0.1 and 0.2 m−1. For233
this reason, the requirement for dk was increased to accept crys-234
tals with dk ≤ 1.3 m−1.235
4.1. SICCAS PbWO4 crystal results236
In figure 2, the light transmission spectra before and after ir-237
radiation is shown for two examples of the SICCAS-type crys-238
tal, demonstrating the different behaviours under irradiation.239
Damage due to the irradiation is seen across the visible spec-240
trum for both crystals, with the right hand plot representing a241
fairly resistant example, accepted for use in the FT-Cal and the242
left hand plot typical of a badly-damaged example, which was243
rejected.244
Figures 3-5 show the light transmission measurements at 360245
nm, 420 nm, and 720 nm, before and after irradiation for all the246
SICCAS crystals tested, demonstrating the spread in physical247
properties of the crystals supplied.248
Crystal ID0 100 200 300 400 500
360n
mLT
5
10
15
20
25
30
35
40
Figure 3: Light Transmission measurements at 360 nm for all SICCAS-type
PbWO4 crystals tested in Gießen. Filled squares are measurements before ir-
radiation and open squares from measurements made after irradiation with 30
Gy. “Crystal ID” is an internal numbering scheme used to track the crystals.
Crystal ID0 100 200 300 400 500
420n
mLT
10
20
30
40
50
60
70
Figure 4: Light Transmission measurements at 420 nm for all SICCAS-type
PbWO4 crystals tested in Gießen. Filled squares are measurements before ir-
radiation and open squares from measurements made after irradiation with 30
Gy. “Crystal ID” is an internal numbering scheme used to track the crystals.
The collected statistics for all crystals with respect to dk at249
420 nm are shown in figure 6. Crystals with a dk higher than250
1.3 m−1 were rejected.251
Two crystals, referred to here as A and B, were subjected252
to multiple exposures to the 30 Gy irradiation, totalling 1, 2,253
4, and 51 times a single dose, allowed 30 minutes in a dark254
environment for fast recovery, then re-measured in the spec-255
trophotometer. The results of these studies are shown in figure256
7. The two crystals show different responses to the irradiation,257
with crystal B experiencing less reduction in light transmission258
as a result of radiation damage. However, after the 51× dose,259
the degradation in light transmission is severe for both crystals,260
indicating that under sufficient gamma irradiation, all SICCAS-261
type crystals will suffer appreciable levels of light transmission262
damage.263
4
Crystal ID0 100 200 300 400 500
620n
mLT
40
45
50
55
60
65
70
75
80
85
90
Figure 5: Light Transmission measurements at 620 nm for all SICCAS-type
PbWO4 crystals tested in Gießen. Filled squares are measurements before ir-
radiation and open squares from measurements made after irradiation with 30
Gy. “Crystal ID” is an internal numbering scheme used to track the crystals.
Property Mean rms
LT Before Irradiation (360 nm, %) 27.05 27.40
LT After Irradiation (360 nm, %) 23.65 24.06
LT Before Irradiation (420 nm, %) 61.48 61.56
LT After Irradiation (420 nm, %) 50.77 51.01
LT Before Irradiation (620 nm, %) 74.72 74.77
LT After Irradiation (620 nm, %) 68.29 68.39
Table 2: Average and rms light transmission values before and after irradiation
of the SICCAS PbWO4 crystals acquired for the FT-Cal
4.1.1. Comparison with BTCP crystal from HPS calorimeter264
Spectra for the BTCP crystal from the HPS ECal and an ex-265
ample FT-Cal SICCAS crystal, both before and after irradia-266
tion, are compared in figure 8. The BTCP crystal is noteable for267
having a greater overall light transmission at all wavelengths,268
and a less pronounced “wiggle” feature which is clearly seen269
around 600 nm in the SICCAS spectra. The BTCP crystal also270
displays a far lower degree of radiation damage across the vis-271
ible spectrum, resulting in a lower value of dk = 0.6 m−1 being272
computed from the light transmission measurements at 420 nm.273
4.1.2. Comparison with BTCP crystals from FT-Cal prototype274
The BTCP crystals show a narrower range of values of dk275
than is seen in the SICCAS crystals, and all have dk less than276
1.0 m−1, the originally-specified value for the FT-Cal. The in-277
duced radiation absorption coefficients for the 8 BTCP sample278
crystals are shown in figure 9, alongside the SICCAS crystals279
which have dk less than 1.0 m−1. Together with the greater light280
transmission at visible wavelengths when compared to the SIC-281
CAS crystals, the superior performance of the BTCP crystals282
has resulted in the available samples being considered for use283
in the regions of the FT-Cal expected to receive the highest rates284
of radiation.285
)-1dk (m0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
No.
of C
ryst
als
0
5
10
15
20
25
30
35
40
Figure 6: Histogram of the radiation induced absorption coefficient, dk, for
all SICCAS-type PbWO4 crystals tested in Gießen. Crystals with a radiation
induced absorption coefficient greater than 1.3 m−1 were rejected.
4.2. LED recovery results286
As shown in the previous section, and in figure 7, even the287
most resistant of the SICCAS-type crystals will suffer degrada-288
tion to their light transmission under sufficient irradiation. For289
the deployment of these crystals in the FT-Cal, thermal anneal-290
ing will not be a practical option for recovering this damage,291
as it would require either the dismounting of all 332 crystals,292
or the capability to heat the calorimeter to a high enough tem-293
perature to significantly speed up the recovery process without294
causing damage to the surrounding components of the FT, or295
indeed to CLAS12.296
One alternative is to use LED illumination, available in the297
FT-Cal from the light monitoring system, to mitigate the effects298
of radiation damage to the crystals. This use of LED illumina-299
tion to recover radiation damage was studied for both the SIC-300
CAS and BTCP type crystals. Irradiated samples of the two301
crystal types were illuminated with blue light from a single-302
colour NICHIA NSBP500AS LED [13], and the light transmis-303
sion recovery effect observed.304
4.2.1. Results for SICCAS crystals305
Three SICCAS crystals, A and B, from section 4.1, and a306
third crystal, referred to here as crystal C, were all used for307
LED recovery studies. All three crystals were subjected to the308
“once-through” irradiation procedure described in section 2.2.309
Additionally, crystals A and B were subjected to multiple irra-310
diations, already discussed in section 4.1. Following their re-311
spective maximum irradiations, 1× 30 Gy dose for crystal C,312
51× 30 Gy dose for crystals A and B, the crystals were then ex-313
posed to LED illumination, and the recovery effects observed.314
For the single 30 Gy dose irradiation, just 30 seconds of LED315
illumination is sufficient to begin to observe recovery in light316
transmission, with significant recovery observed after several317
minutes of exposure to illumination from the LED. Figure 10318
shows the light transmission curves for SICCAS crystal C after319
a series of 30 second illuminations from LED light, demonstrat-320
ing the fast rate of recovery possible via LED illumination.321
5
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80
Before
1530 Gy
60 Gy30 Gy
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80
Before
30 Gy60 Gy
1530 Gy
Figure 7: Light Transmission measurements for multiply irradiated SICCAS
crystals, referred to as A (left) and B (right). Each crystal was subjected to
sequential doses of 30 Gy totalling 1 (30 Gy), 2 (60 Gy), and 51 (1530 Gy)
times the 30 Gy dose representing one month of expected FT-Cal operational
dose, with greater reductions in light transmission seen as the dose received
increases. Labels show intersections of these curves with a line at 440 nm.
For the crystals receiving the higher doses, much longer illu-322
mination times, of order several hours, are required to see any323
significant recovery effect, demonstrated in figure 11, which324
shows the light transmission recovery after 4 hours illumina-325
tion.326
It is expected that the light monitoring system of the FT-Cal327
will be used to regularly optically anneal the crystals, usually328
between experiments. The doses accumulated over the course329
of an experiment are not expected to cause significant damage330
to the crystals, and a few minutes illumination should be suffi-331
cient for recovery, although the possibility exists to illuminate332
crystals between runs during an experiment if necessary, pro-333
vided the opportunity to do so arises through beam operations334
or configuration changes in the experimental hall.335
4.2.2. Results for HPS BTCP crystal336
In a similar programme of tests to the SICCAS crystals, the337
BTCP crystal from the HPS calorimeter was also subjected to338
LED illumination. Like SICCAS crystal C, a “once-through”339
irradiation procedure was performed on the crystal, and expo-340
sure to varying degrees of LED illumination used to realise light341
transmission recovery. After 35 minutes of illumination, almost342
complete recovery of the light transmission is observed, shown343
in figure 12.344
5. Conclusions and Outlook345
The demanding specifications of the Forward Tagger346
calorimeter has proven a challenge with respect to acquiring347
lead tungstate crystals with sufficient performance for deploy-348
ment in the device, of particular concern was the degradation349
in light transmission under irradiation. The program of crys-350
tal assay undertaken at the University of Gießen has enabled351
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
20
40
60
80
Before
After
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
20
40
60
80
After
Before
Figure 8: Light Transmission measurements for a single SICCAS PbWO4crystal, before and after irradiation (left), alongside similar measurements per-
formed for a BTCP crystal (right). The irradiated spectra exhibit lower light
transmission, particularly around the 400 - 600 nm region, however the BTCP-
type crystal exhibits a greater resistance to light transmission damage under
irradiation.
an informed selection to be made of those crystals requiring352
replacement, and the positioning of the selected crystals within353
the calorimeter according to their radiation hardness. Studies of354
possible recovery of radiation damage to the crystals by means355
of LED illumination were also performed, leading to greater356
confidence that the LED-based light monitoring system of the357
FT-Cal, designed for preliminary gain equalisation, can also be358
applied to the optically-stimulated recovery of radiation dam-359
age to the crystals, rather than the time-consuming alternative360
of dismounting the calorimeter for thermal annealing.361
Acknowledgements362
The authors would like to acknowledge the assistance and363
support provided in this work by the members of the PANDA364
group at the University of Gießen and the staff of the Strahlen-365
zentrum, in addition to that of the technical staff and electron-366
ics workshop of the INFN Genova section, in particular Fabio367
Barisone and Paolo Pozzo.368
The proposed application of the Forward Tagger light moni-369
toring system to perform stimulated recovery of radiation dam-370
age via external light on a quasi-online basis (i.e. without dis-371
mantling the calorimeter for annealing between experiments)372
was discovered and developed as a result of research and de-373
velopment on PbWO4 crystals for PANDA. This application is374
described in reference [11], and can be performed with infrared375
or visible light. A European patent for a device providing ex-376
ternal light for this purpose is held by the University of Gießen.377
References378
[1] The CERN Large Hadron Collider: Accelerator and Experiments, vol 1-2,379
CERN, Geneva (2009)380
[2] A.A. Annenkov, et. al., Nucl. Instr. and Meth. A 490 (2001) 30381
6
Crystal ID0 100 200 300 400 500
Gie
ssen
)-1
dK(m
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
Figure 9: Induced absorption coefficient, dk, for the BTCP sample crystals from
FT-Cal prototypes (squares), shown alongside all FT-Cal SICCAS crystals with
dk less than 1.0 m−1 (crosses), the originally-specified value for the FT-Cal.
Both crystal types shown have the same 200 × 15 × 15 mm geometry, allowing
these BTCP samples to be deployed alongside SICCAS crystals in the Forward
Tagger.
[3] PANDA Electromagnetic Calorimeter (EMC), Technical Design Report,382
http://www-panda.gsi.de/archive/public/panda tdr EMC.pdf383
[4] Deeply Virtual Compton Scattering with CLAS at 6 GeV, Experimental384
Proposal http://www.jlab.org/exp prog/proposals/01/PR01-113.pdf (2001)385
[5] M. Battaglieri, et. al. (CLAS Collaboration), The CLAS12 For-386
ward Tagger, Technical Design Report, http://www.ge.infn.it/˜batta/jlab/ft-387
tdr.2.0.pdf (2012)388
[6] P. Adzic, et. al., J. Inst. 5 (2010) 03010389
[7] X. Qu, et. al., Nucl. Instr. and Meth. A 480 (2002) 470390
[8] Heavy Photon Search - A Proposal to Search for Mas-391
sive Photons at Jefferson Laboratory, experimental Proposal,392
http://www.jlab.org/exp prog/proposals/11/PR12-11-006.pdf (2011)393
[9] P. Lecoq, et. al., Nucl. Instr. and Meth. A 365 (1995) 291394
[10] R.Y. Zhu, et. al., Nucl. Instr. and Meth. A 376 (1996) 319395
[11] V. Dormenev, et. al., Nucl. Instr. and Meth. A 623 (2010) 1082396
[12] E. Auffray, et. al., Nucl. Instr. and Meth. A 456 (2001) 325397
[13] NICHIA Corportation, Technical specifications for NICHIA blue LED398
NSPB500AS399
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80 Before
After
30 Seconds60 Seconds
90 Seconds
Figure 10: Light Transmission measurements for SICCAS-type PbWO4 crystal
C, irradiated with 30 Gy (‘After’ curve) and recovered using LED illumination.
Initial recovery of light transmission can be seen after just 30 seconds illumi-
nation (green curve). Successive 30-second illuminations show that significant
recovery of light transmission is possible after a few minutes (further curves
shown representing 60 seconds accumulated illumination, and 90 seconds). La-
bels show intersections of these curves with a line at 440 nm.
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80
Before
After
4 Hours LED
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80
Before
4 Hours LED
After
Figure 11: Light Transmission measurements for multiply irradiated SICCAS
crystals A (left) and B (right), showing pre-irradiated measurements, irradiated
spectra for a 51 times 30 Gy dose, and the effect of 4 hours LED illumination on
the recovery of light transmission. After such heavy doses, where light trans-
mission has been severely decreased, crystal recovery via LED illumination is
more difficult.
7
Wavelength (nm)300 400 500 600 700 800 900
LT (
%)
0
10
20
30
40
50
60
70
80
90
Before
30 Gy
35 Min LED
Figure 12: Light Transmission measurements for BTCP-type PbWO4 crystal,
before and after irradiation with 30 Gy, and recovered using LED illumination
totalling 35 minutes. After this illumination, almost complete recovery of the
light transmission is observed.
8