AUTOMATED
HIGH
RESISTANCE
MEASUREMENT
BRIDGE
Setup Guide & Operator’s Manual
2
Letter of Transmittal
University of Maryland
College Park, MD. 20742
April 21, 2015
Dear Scientists and Engineers,
We are pleased to present you the manual for an automated high resistance measurement
system. This manual was composed in partial fulfillment of the requirements for ENGL393
course during spring 2015 Semester. Specifically, this manual includes contributions of both
Jacob and myself. Being students of the University of Maryland majoring in Electrical
Engineering, Jacob and I gained the basic knowledge of electronic circuit theory and
applications. The technique described in this manual is based on the Wheatstone bridge
principle, which was introduced to us during our sophomore level circuit theory courses.
Furthermore, I have been employed at the National Institute of Standards and Technology
(NIST) since September 2014. During my first few months of employment, I was given the
opportunity to conduct research on advancing the existing high resistance measurement system
and to develop a new system that includes latest equipment and software. Similarly, Jacob has
gained a thorough understanding of the basic theory of the high resistance measurement system
and its applications from his course work. Based on our experience, we believe that performing
accurate, high resistance measurements are challenging and time consuming. Therefore, we are
taking this as an opportunity to present a detailed and simplified manual that describes the
equipment selection and the setup procedure for an automated high resistance measurement
system.
3
As we mentioned above, accurate high resistance measurements require a great effort. It
is important to consider the barriers ordinary scientists and engineers face while attaining
accurate high resistance measurements. Usually, engineers or scientists working on large scale
projects request measurement services at NIST. However, this requires time for shipping and
additional costs. Similarly, the second option would be to purchase a commercially available
measurement device. This option isn’t viable because it would cost more money and would
affect the accuracy of the measurement result. Therefore, from our experience, we believe that it
is more feasible to set up a measurement system in the lab if your project relies on accurate high
resistance measurements. The manual we completed focuses on engineers and scientists who
relies on high resistance measurements in a laboratory environment. Furthermore, the manual
will guide the reader through the proper equipment selection, assembly, software design, and
troubleshooting.
As emerging engineers, we recognize the importance of accuracy and the efficiency of
physical measurements for engineering projects. Therefore, we hope that this manual will ease
the complications linked with high resistance measurements and will make a positive impact on
your projects.
Sincerely,
Enclosure: Manual for an automated high resistance measurement system
4
Limitation of Liability
This manual was produced in partial fulfillment of the requirements for a University professional
writing course. The information contained in this user manual is provided "as is". The author(s),
though well-informed and experienced, are not professional or experts in the topic, so [they]
make[s] no representations or warranties of any kind, express or implied, about the completeness
of the information for a particular use, or merchantability. Nor do [they] make any
representations of freedom from infringement of any intellectual property or proprietary rights of
any third party. You use the information at your risk. You are advised to consult an appropriate
professional if you have any doubts related to the advice or procedures in this manual.
Identification of commercial equipment does not imply any endorsement or that the equipment is
the best available for the purpose.
5
Table of Contents
Section Title Page
1 Safety Information 8
Proper safety measures 9
Symbols used in this manual 10
First Aid for electric shock 11
2 Introduction 14
Specifications of the System 14
3 Background Information & Basic Theory 16
Wheatstone Bridge 17
Modified Wheatstone Bridge 18
4 Measurement Algorithm 20
5 Equipment Selection 22
Programmable High Precision Voltage Supply 22
Alternative Options for the Calibrator 23
Current Detector 24
Alternative Options for the Current Detector 25
Cables 25
Air Bath 27
Alternative Options for the Air Bath 28
Computer with GPIB support 28
6 Assembly Procedure 29
6
7 Creating the Automation Software 33
Visual Basic 33
LabVIEW 33
Developing the Automation Software 34
Overview of a sample Automation Software 36
8 Measurement Procedure 38
9 Testing the System for Accuracy 43
Using two standard resistors with adequate calibration data 43
Using a Hamon Transfer Standard 43
10 Troubleshoot Guide 45
11 Appendix 47
Glossary of Terms 47
Metric Prefixes 48
References 49
Interview Write-up 50
7
List of Figures
Figure Title Page
1.1 Warning Sign [1] 8
3.1 Conventional Wheatstone Bridge [14] 17
3.2 Modified Wheatstone Bridge with Guard Resistors [14] 18
4.1 Modified Wheatstone Bridge with Currents Labeled. 20
5.1 Overview of the Fluke 5730A Voltage Calibrator. 23
5.2 Overview of the Keithley 6430 Current Detector. 24
5.3 Polyethylene Cable and Teflon Cable. 25
5.4 Cable Connectors 26
5.5 Junction Boxes for the Voltage Calibrator 26
5.6 Node Box 26
5.7 Overview of the MI Model 9300 Air Bath 27
5.8 GPIB Cable 28
7.2 Overview of a sample automation software 36
9.1 Internal Diagram of a Hamon Transfer Standard [3] 43
8
1 Safety Information
WARNING
HIGH VOLTAGE is used during the operation of this measurement system.
High voltage is present during the operation, observe all safety precautions
mandated by your laboratory safety guide.
To prevent possible electric shock, when the measurement system is plugged in,
do not touch terminal connections. Do not attempt to change wires while the
system is running.
Figure 1.1 [1]
9
1.1 Proper Safety Measures
To prevent possible electrical shock, equipment damage, or personal injury:
Adhere to safety instructions mandated by your laboratory safety guide when assembling
and using the measurement system [7].
Carefully read all instructions.
Do not implement the measurement system around explosive gas, vapor, or in damp or wet
environments [8].
Do not use damaged cables.
Implement the system indoors only.
Do not let unauthorized personal use the measurement system [7].
When the measurement system is running, make sure to isolate the environment by either a
caution tape or a rope [7].
Do not set up the measurement system where access to the main power cord is blocked.
Use only the mains power cord and connector approved for the voltage and plug
configuration in your country and rated for the each equipment of the measurement system
[9].
Always program the software so that the output current is limited by the voltage supply.
Replace the power cords if the insulation is damaged or if the insulation shows signs of
wear.
Make sure the ground conductor in the main power cord is connected to a protective earth
ground. Disruption of the protective earth could put voltage on the chassis that could cause
death [8].
Avoid using an extension cord or adapter plug.
Do not apply more than the rated voltage, between the terminals or between each terminal
and earth ground [8].
Use only cables with correct voltage ratings [9].
If it is possible use a caution light when the system is running to alert other personnel.
10
1.2 Symbols Used In This Manual
The following symbols are used throughout the manual to alert readers about safety concerns.
Table I: Symbols marked on this manual [10].
Symbol Definition
Danger is used to alert readers about an
immediate and serious hazard that will likely
be fatal.
Warning is used to alert readers about the
potential for serious injury, death, or serious
damage to equipment used.
Caution is used to alert readers about the
potential for anything from s moderate injury
to serious equipment damage or destruction.
Note: Note is used for a tip or suggestion to help
readers carry out the procedure successfully.
11
1.3 First Aid for Electric Shock
The following procedures must be followed in the case of an electric shock. Adapted from U.S.
National Library of Medicine [11].
Symptoms
Symptoms depend on several factors, including:
Type and strength of voltage
How long you were in contact with the electricity
How the electricity moved through your body
Your overall health
● Symptoms may include:
Changes in alertness (consciousness)
Broken bones
Heart attack (chest, arm, neck, jaw, or back pain)
Headache
Problems with swallowing, vision, or hearing
Irregular heartbeat
Muscle spasms and pain
Numbness or tingling
Breathing problems or lung failure
Seizures
Skin burns
12
First Aid
1. If you can do so safely, turn off the electrical current. Unplug the power cord, remove the fuse
from the fuse box, or turn off the circuit breakers. Simply turning off an appliance may NOT
stop the flow of electricity. Do NOT attempt to rescue a person near active high-voltage lines.
2. Call your local emergency number, such as 911.
3. If the current can't be turned off, use a non-conducting object, such as a broom, chair, rug, or
rubber doormat to push the person away from the source of the current. Do not use a wet or
metal object. If possible, stand on something dry and that doesn't conduct electricity, such as a
rubber mat or folded newspapers.
4. Once the person is away from the source of electricity, check the person's airway, breathing,
and pulse. If either has stopped or seems dangerously slow or shallow, start first aid.
5. If the person has a burn, remove any clothing that comes off easily, and rinse the burned area
in cool running water until the pain subsides. Give first aid for burns.
6. If the person is faint, pale, or shows other signs of shock, lay him or her down, with the head
slightly lower than the trunk of the body and the legs elevated, and cover him or her with a warm
blanket or a coat.
7. Stay with the person until medical help arrives.
8. Electrical injury is frequently associated with explosions or falls that can cause additional
severe injuries. You may not be able to notice all of them. Do not move the person's head or neck
if the spine may be injured.
13
DO NOT
● Do NOT touch the person with your bare hands if the body is still touching the source of
electricity
● Do NOT apply ice, butter, ointments, medications, fluffy cotton dressings, or adhesive
bandages to a burn
● Do NOT remove dead skin or break blisters if the person has been burned
● After the power is shut off, do NOT move the person unless there is a risk of fire or
explosion
When to Contact a Medical Professional
Call your local emergency number, such as 911 if a person has received an electrical burn.
14
2. Introduction
Precise electrical measurements have been a key ingredient in groundbreaking
engineering projects and research developments. Resistance measurements up to and beyond 100
Tera-Ohms requires expensive equipment and additional resources. However, resistance
measurements in the above mentioned scale can be performed in a laboratory environment with a
low uncertainty using an automated resistance measurement bridge. An automated high
resistance measurement bridge as described in this manual, consist of two resistors, two high
precision voltage supplies (Calibrators) and a current detector. This manual describes the best
equipment, the setup procedure and the measurement process of an automated high resistance
measurement bridge. Furthermore, this manual is primarily based on the automated resistance
measurement bridge that was implemented at the National Institute of Standards and Technology
(NIST). The measurement technique described in this manual is based on the Wheatstone bridge
principle. A Wheatstone bridge is a four resistor bridge that consists of three known resistors and
one unknown resistor [12]. However, large uncertainties and inaccessibility of standard resistors
makes the ordinary Wheatstone bridge unfeasible in high resistance measurements [7].
Therefore, a modified version of the Wheatstone bridge is used to precisely measure high
resistance values. A complete description of the Wheatstone bridge and the modified Wheatstone
bridge, is included in the following section of the manual.
2.1 Specifications
The basic specifications of the measurement system described in this manual are included in the
table below
Table II: Specifications of the measurement system described on this manual
Voltage Source Fluke 5730A Calibrator
Current Detector Keithley 6430 Detector
Air Bath Used Measurement International Model 9300
Connectivity Method Polyethylene Cables
Minimum Applied Voltage 0 V
Maximum Applied Voltage +1100 V or -1100 V
15
Voltage Source Uncertainty approximately ~2 ppm [8]
Current Limitation 3 mA (set by the measurement software)
Uncertainty in Resistance Range 1 Tera-Ohm ~10 ppm1
100 Tera-Ohm ~ 50 ppm
Settling Time Average 120 Seconds
Temperature Performance 0 °C to 50 °C
Relative Humidity 30% - 50%
Standard Interfaces GPIB (General Purpose Interface Bus)
Connectors to communicate with the
computer
Programming Platform Used LabVIEW Platform
Additional Options Available -Voltage Calibrator (voltage source)
Correction.
-Null Current Measurement for offset
correction
-Acquiring resistor data from a standard
resistor data excel sheet
-Save measurement data onto a readable text
file
1 Parts per Million (ppm) is a value that represents the part of a whole number in units of 1/1000000.
1 PPM = 0.0001%
16
3. Background Information & Basic Theory
3.1 Background
High resistance measurements are an important part of the precision measurement
industry. Particularly, several commercial and government industries depend on accurate high
resistance measurements. For example, the department of energy and the nuclear power industry
uses Dosimeters to measure background radiation [7]. These Dosimeters uses a very small
current, usually in the Femto-Amp range. Therefore, in order to measure the small current, these
industries have to rely on precise high resistance standards. Similarly, high resistance
measurements are used to test the purity of materials. For example, in the textile industry, high
resistance measurements are used to judge the purity of the textile material and to determine the
use of the textile material [7].
Even though accuracy in these high resistance measurements is crucial for the above
mentioned applications, due to constraints such as longer settling times and inefficiencies in
manual operation, achieving accurate high resistance measurements has always been
challenging. There are several commercially available products that can be used to achieve high
resistance measurements with a higher uncertainty. The Guideline 6535 Automated High
Resistance Measurement System can produce high resistance measurements with approximately
50 parts per million (ppm) in the 1Tera Ohm range [13]. However, some industries require the
need of precise measurements up to 100 Tera ohms with an uncertainty of 5-10 ppm. As an
example, a common customer of the National Institute of Standards and Technology (NIST),
Sandia national labs, supports the nuclear weapons program at U.S. Army. According to an
interview with Mr. Dean Jarrett at NIST, Sandia national labs requested NIST to measure the
resistivity of a ceramic coating that was well above the 1Tera Ohm range. [7]. Therefore
measurement capabilities of the commercially available products are insufficient to serve the
needs of the industry.
A measurement system that can measure high resistance values with high precision can
be implemented using the Wheatstone bridge principle.
17
3.2 Wheatstone Bridge.
A conventional Wheatstone bridge is the most used method for high resistance measurements.
Circuit diagram of the conventional measurement bridge is shown below.
The conventional Wheatstone bridge used for high resistance measurements requires
three resistors to measure a single unknown resistor [12]. In the figure 3.1, Rx is the unknown
resistor that we are interested in. If the current of the detector is at zero, ratios between R1 and
R2 and Rx and Rs should be equal to each other based on Kirchhoff's Current laws. Therefore
the equation for the unknown resistor value can be derived as follows.
𝑅2
𝑅1=
𝑅𝑠
𝑅𝑋 → 𝑅𝑋 =
𝑅1
𝑅2∙ 𝑅𝑠
Please note that the complete derivation for the case of the modified Wheatstone bridge is
included in the following section. The derivation for the conventional Wheatstone bridge is
analogous to the modified Wheatstone bridge.
As shown on figure 3.1, resistors R1 and R2, typically in the range of 104 ohms and 107
ohms form the reference arms of the resistance measurement bridge [15]. When the bridge is
balanced, the ratio of these is equal to the ratio of the unknown resistor to the other resistor Rs
which form the other arms of the bridge as shown on figure 3.1. Wheatstone bridges used in the
measurement industry are often manually operated [15]. As an example, manually operated
Figure 3.1: Conventional Wheatstone bridge [14]
18
guarded Wheatstone bridge at the national institute of standards and technology required an
operator to manually change the resistance ratios while paying close attention to the current
detector [7].
Drawbacks of the Wheatstone bridge method are evidently as follows. Due to the nature
of the bridge, three resistors with precision and enough calibration data are required to perform a
measurement. Therefore, the resistance value of the unknown resistor will always depend on the
uncertainties of the other three resistors. Also, leakage in resistors is a possible source of error in
measurements [15]. Similarly, manual operation of the resistance bridge leads to human errors.
According to Mr. Jarrett at NIST, “patience of a manual operator is very limited. Therefore, we
always rely on two different individuals to operate the system in two different times” [7].
Therefore, the measurement industry showed the need of an improved resistance measurement
system.
3.3 Modified Wheatstone Bridge.
A modified Wheatstone bridge uses the same basic theory as the conventional
Wheatstone bridge. However, as shown in figure 3.2, modified Wheatstone bridge uses two
highly precise voltage supplies on the left side of the bridge replacing the R1 and R2 resistors in
the conventional Wheatstone bridge [14].
ID A B
IX
Is
Figure 3.2: Modified Wheatstone bridge with Guard Resistors [14]
19
The low output impedance of the high precision voltage sources minimizes the errors caused by
the leakage currents [14]. Also, by choosing programmable voltage sources, the system can be
automated to set the voltage precisely up to 2 ppm or lower [8]. Automating the system
eliminates the human factor from the measurement and helps to further refine the measurement.
Furthermore, automating the measurement process improves the repeatability and provides the
opportunity to perform measurements overnight without relying on a human operator.
A complete derivation of the equation for the unknown resistance value using Kirchhoff’s
current laws is as follows,
First, using the Kirchhoff’s current law at node ‘B’ on figure 3.2 results,
𝐼𝑋 – 𝐼𝑆 + 𝐼𝐷 = 0
Similarly, using Kirchhoff’s voltage laws at the top loop and the bottom loop,
(𝐼𝑋𝑅𝑋) − (𝐼𝐷𝑅𝐷) − 𝑉1 = 0
(𝐼𝑆𝑅𝑆) − 𝑉2 + (𝐼𝐷𝑅𝐷) = 0
When the bridge is balanced, 𝐼𝐷 = 0 since no current is flowing through the detector. Therefore
the above equations can be simplified as,
(𝐼𝑋𝑅𝑋) = 𝑉1
(𝐼𝑆𝑅𝑆) = 𝑉2
Rearranging, obtaining the ratios and solving for 𝑅𝑋 gives,
𝑅𝑋 = 𝑉1𝐼𝑠𝑅𝑠
𝑉2𝐼𝑋 When, 𝐼𝐷 = 0 𝐼𝑠 = 𝐼𝑋
Therefore, 𝑅𝑋 equals to,
𝑅𝑋 = 𝑉1
𝑉2𝑅𝑠
This equation can be used to calculate the unknown resistor value using the voltage of the
Calibrators and the reference resistor.
Equation 3.1: Unknown resistance based on other parameters
20
4. Measurement Algorithm
The measurement algorithm for the measurement system is based in the modified Wheatstone
bridge principle explained on section 3 of this manual.
First, the voltage on V1 and V2 will be set according to the ratio of the nominal values of
the two resistors. For example, if the RS value is 100Giga ohm and the RX value is 10 Giga
ohms, the Ratio will be 10:1. Then, V1 and V2 will be set using the software based on the ratio.
After the voltage is being set, the detector current will be measured based on the settling time of
the resistor. Settle down time of the resistors, also known as the RC time constant, will depend
on the resistor value. Therefore, the software will have the functionality to manually enter the
settle time before starting the measurement. After the settling time, current detector will acquire
200 data points of the current (ID). The average value of the above mentioned current will be
used in the calculations shown below.
Using Kirchhoff’s current and voltage laws, derivation of the value for the V2 voltage at the
balance point is as follows.
Applying Kirchhoff’s current law at node B as shown
in figure 4.1 results,
𝐼𝑋 = 𝐼𝑆 + 𝐼𝐷
Converting the above equations to voltages;
𝑉1
𝑅𝑋 =
𝑉2
𝑅𝑆 + 𝐼𝐷 − (1)
At the Balance point, current 𝐼𝐷 = 0; Therefore,
𝑉1
𝑅𝑋 =
𝑉2′
𝑅𝑆 − (2)
Where 𝑉2′ is the voltage of the calibrator (V2) at the balance point.
Figure 4.1: Modified Wheatstone bridge with labeled currents
21
Subtracting (2) from (1) and solving gives the 𝑉2′ value to be set, in order to balance the bridge.
𝑉2′ = 𝑉2 + 𝐼𝐷𝑅𝑆
The above calculated voltage will be set on the V2 detector and the current is measured again
after the settling time. Furthermore, in order to make a precise measurement, the whole process
will loop two times. Therefore, a finer value of 𝑉2′ voltage can be calculated.
Using the 𝑉1and 𝑉2′ at the balance point, unknown resistance value can be calculated using the
equation 3.1 derived in section 3 as shown below.
𝑅𝑋 (𝑢𝑛𝑘𝑛𝑜𝑤𝑛) =𝑉1
𝑉2′ × 𝑅𝑆
Equation 4.1: Recalculating V2 voltage at balance point.
Equation 4.2: Unknown resistance based on V1 voltage corrected V2 voltage and the reference resistor Rs.
22
5. Equipment Selection
Choosing the right equipment is crucial to achieve the highest possible accuracy of an
automated high resistance measurement system. Specifically, certain evaluation factors are being
used to determine the best equipment for the measurement system. The selection process of the
equipment is based on the user’s budget, level of uncertainty required in the measurement, and
the amount of measurement time. For example, if the measurement system is being developed in
a laboratory where a high resistance measurement system is used for testing the effectiveness of
an insulator, a higher accuracy may not be required. Therefore, the necessity of the equipment is
based on the user's needs.
Resistance laboratory at the National Institute of Standards and technology is constantly
in the forefront of research in high accuracy resistance measurements. Therefore, the equipment
described below are based on the research that has been conducted at NIST for the highest level
of accuracy [7]. According to Mr. Jarrett at NIST, an accuracy of 5 ppm was achieved using the
equipment described below [7]. However, based on the user’s expectations, different equipment
with the same functionality can be used. A brief description of alternative options is included in
the end of each section.
5.1 Programmable High Precision Voltage Supply (Voltage Calibrator).
The elimination of the R1 and R2 resistors of the conventional Wheatstone bridge as
described in section 3 of this manual, resulted the system to rely on high precision voltage
supplies. Similarly, as described in section 3, using a voltage source instead of two resistors
eliminates the possible leakage current that could occur between the resistors [15]. Furthermore,
choosing a programmable voltage source will provide the opportunity to control the supplied
voltages using an automation software such as LabVIEW or Microsoft Visual basic.
Examining these reasons, it is evident that the voltage sources are a crucial part of the
measurement system. Accuracy on these voltage supplies will ultimately affect the accuracy of
the final measurement. There are several commercially available voltage supplies on the market
that can produce accurate constant voltages. Namely, industrial testing equipment manufacturer
Fluke Corporation, manufactures several programmable voltage supplies that can produce a high
precision voltage. From Fluke’s collection, Fluke 5730A is capable of producing a voltage
output with an uncertainty of 1.5 ppm [8]. Therefore, the best possible accuracy can be achieved
using the Fluke 5730A voltage source.
23
Furthermore, the Fluke 5730A can be programmed using a computer software through
the GPIB interface. This functionality is extremely useful to eliminate the human operator during
the balancing stage of the bridge.
Overview of the front panel of the voltage calibrator is shown below.
Figure 5.1: Overview of the Fluke 5730A Voltage Calibrator
(1) Standby / Operate indicator
(4) Power Switch
(2) Output Terminals
(5) Number Keys for User Input
(3) Information Display
(6) Operate Key
5.2 Alternative options for the voltage calibrator
Based on the user’s budget and the expected uncertainty level, a different voltage source
than the above mentioned fluke 5730A can be used as the two voltage source for the
measurement bridge. Based on his experience, Mr. Jarrett recommends buying a used voltage
source from the used equipment market [7]. However, he also mentioned that old equipment may
no longer be supported by the manufacturer and would not be repaired [7]. Furthermore, if an
absolute precision is not necessary, any programmable voltage source can be used as the two
voltage supplies for the measurement system.
1
2
3
5
4
6
1
24
5.3 Current Detector (Current Meter)
The current detector is another crucial part of the automated high resistance measurement
system. The basic use of the current detector on the measurement bridge is to measure the
current at the balance point. Specifically, after the voltages V1 and V2 are set based on the two
resistor ratio, current (ID) as shown in figure 3.2 is measured using the current detector.
Therefore, the accuracy of the current detector is vastly important to produce accurate
measurement results from the measurement bridge.
There are several current detectors available on the commercial measurement equipment
market that can produce high precision measurement results. Specifically, the Keithley 6430
current detector can measure current less than a femto-Amp with an accuracy of 1% of the
nominal value [16]. Therefore, for the maximum possible accuracy of the measurement, the
Keithley 6430 current detector is the best option available on the market at the time this manual
was composed. Furthermore, Keithley 6430 can be programmed remotely. Remote programming
functionality is useful to automate the measurement process and eliminate human errors.
Overview of the front panel of the current detector is shown below.
Figure 5.2: Overview of the Keithley 6430 Current Detector
(1) Measured Current Display
(3) Measurement ON/OFF Key
(2) Main Power Switch
(4) Device Status Indicator
1
2
4
3
25
5.4 Alternative options for the current detector.
Analogous to the case of the voltage calibrator, based on the user’s intended budget and
the level of uncertainty, different equipment can be used. However, using a current detector that
can measure current up to nano-Amp range is recommended because, at the balance point of the
measurement, the detector current (ID) as shown in figure 3.2 is significantly lower. Also, a
current detector with the remote programing functionality is recommended in order to automate
the measurement system.
5.5 Cables
For the measurement system described in this manual, connection cables are used to
connect the unknown resistor (Rx), the reference resistor (RS), voltage sources and the current
detector. Choosing the correct type of connection cable is important to minimize noise during
measurement and improve the measurement time. In the measurement industry, Teflon and
Polyethylene are the two types of connection cable that are used for resistance measurements [7].
At NIST, Teflon cables are widely used for resistance measurements due to its high
insulation characteristics. However, experiments conducted at NIST prove that Teflon cables
retains charge at high resistance levels [7]. This charge retention during a measurement results in
a noisy measurement result. Therefore, for high resistance measurements, Polyethylene cables
are used to reduce errors in the measurement result. Polyethylene cables are specially made to
reduce the charge retention and to operate with low noise [7]. Additionally, Polyethylene cables
have same insulation characteristics as Teflon cables. Therefore, for high resistance
measurements, Polyethylene is the most suitable type of wire for highest precision of the
measurement. Additionally, junction boxes and node boxes, as shown in figure 5.5 and figure
5.6, are used to connect cables to the calibrator and to wire the bridge circuit.
Figure 5.3: Polyethylene Cable shown left and Teflon Cable shown right.
26
Figure 5.6: Node Box
Figure 5.4: Cable Connectors
Figure 5.5: Junction Boxes for the voltage Calibrator.
27
5.6 Air Bath
Standard resistors are highly susceptible to variations in temperature [7]. Therefore, a
temperature controlled enclosure is used to minimize errors from temperature fluctuations. In the
resistance measurement industry, standard resistors are either stored in oil or air baths to prevent
temperature fluctuations. According to Mr. Jarrett at NIST, Oil Baths provide better temperature
control than Air baths. However, Mr. Jarrett also mentions that, for high resistance values,
especially resistors that are higher than 10 Mega Ohm, using Oil Baths could create leakage
problems due to the lower insulation of the Oil [7]. Therefore, for high resistance standards, a
high precision temperature controlled Air Bath should be used.
For the measurement system described in this manual, Measurement International (MI)
Model 9300 Air Bath will be used for this purpose. Specifically, the MI Model 9300 Air Bath is
capable of controlling its temperature with an accuracy of ± 50mK for ± 1°C change in ambient
temperature [5].
Figure 5.7 Overview of the MI Model 9300 Air Bath
Temperature
Display
Temperature
Controller
Resistor
Storage Area
28
5.7 Alternatives options for the Air Bath.
Similar to the case of the Voltage Calibrator or the Current Detector, a different Air Bath
can be used based on the user’s budget and expected level of uncertainty. Furthermore, there are
several different temperature controlled Air Baths available in the industry that are relatively
lower in price with lower performance and functionality. Another alternative for the above
mentioned Air Bath is to build your own Air Bath. However, similar to any build your own
project, this process would require additional resources such as time, money and knowledge.
5.8 Computer with GPIB Support.
The measurement system described on this manual can be automated using a
programming platform with the aid of a GPIB interface. General Purpose Information Bus, also
known as GPIB, is a widely used communication interface in the measurement industry.
Moreover, GPIB connectors provide the ability to attach more than one component onto the
same cable providing the ability to control every equipment on the measurement system using a
single computer.
Figure 5.8: GPIB Cable [4]
Furthermore, in order to automate the measurement system, a computer that meets the
minimum system requirements for the LabVIEW software or the Visual Basic software, is
required.
29
6. Assembly Procedure
The following instructions elaborate the hardware setup procedure before operating the
measurement system.
The calibrators are capable of producing lethal voltages up to 1100V. Do not
touch output terminals when the device operates.
Step 1: Power off the Calibrator.
Before attempting to
make any connections
ensure the device is
powered off.
Figure 6.1 illustrates the front layout of
the Fluke 5730A voltage calibrator.
Power off the calibrator by switching
the power button ‘off’ as shown in figure
6.1
Step 2: Attaching the Junction Boxes to the Calibrators
Figure 6.1 POWER SWITCH
OUTPUT TERMINALS
BACK TERMINALS
FRONT TERMINALS
Figure 6.2
30
Figure 6.2 illustrates the front terminals and the back terminals of the junction boxes. The
back terminals of the junction boxes should align with the output terminals of the
calibrator as shown in figure 6.3. Similarly, the front terminals are used to connect other
components to the calibrator using cables.
Align the back terminals of the junction box with
the output terminals of the calibrator as shown in
figure 6.4.
Ensure that connectors are properly aligned when
attaching the junction box to the
calibrator. Improper alignment will
damage the terminals of the
junction box and the calibrator.
Repeat the above process for the second calibrator.
Figure 6.3
Figure 6.4
31
Step 3: Attaching front cables to the Calibrator
As shown on Figure 6.5, connect the
Positive wire to the top terminal and the Ground
wire to the second terminal.
Ensure that the open ends of
the wires are not connected to
any equipment while
performing this step.
Repeat the above step for the second calibrator.
Step 4: Connecting the Node Boxes
Figure 6.6 Illustrates a 3 terminal Node Box.
Connect the two ground wires from the two
calibrators (step 3), to any two of the three
terminals of the Node Box.
Connect a third wire to the remaining terminal
of the Node box. Figure 6.7 illustrates the final
result.
Note: the three terminals of the Node Box are
internally shorted together. Therefore, the wires
can be connected to any of the terminals.
Figure 6.5
(+) POSITIVE
GROUND
Figure 6.6
32
After completing the steps above the wiring
assembly should look similar to Figure 6.7.
Step 5: Current Detector
The current detector does not require additional
wire connections because, the Remote PreAmp as
shown in Figure 6.9, comes pre-assembled with
the current detector.
The Remote PreAmp, as shown in Figure 6.9, is
used to measure the current of the bridge during
runtime.
Note: instructions on attaching resistors to the
measurement system are included in the
measurement procedure section of the manual.
Figure 6.7
Figure 6.8
Figure 6.9
33
7. Creating the Automation Software
The implemented measurement system can be automated using a programming platform
to perform the balancing algorithm with minimum human interaction. Voltage calibrator and the
current detector used in this system provides the functionality to remote control over a GPIB
interface. Therefore, using the GPIB interface as the main communication protocol for the
measurement system makes it feasible to automate the measurement process. There are a wide
variety of programming platforms available in the industry. A comparison between two major
programming platforms used in laboratory environments is included below.
7.1 Visual Basic
Microsoft Visual Basic is widely used in laboratory environments due to its ease of use
and compatibility. According to Microsoft, Visual Basic is engineered for productively building
type-safe and object-oriented applications. Furthermore, Visual Basic enables developers to
target Windows, Web, and mobile devices [6]. Main programming style of Visual Basic is text
based. Therefore, creating a software requires fundamental knowledge in programming and
syntaxes. Furthermore, Visual Basic is vastly compatible with measurement equipment and
components. However, due to its code based nature, upgrading an equipment or an operating
system could create compatibility issues.
7.2 LabVIEW
LabVIEW is also widely used in laboratory environments and measurement industry due
to its simplicity and ease of use. LabVIEW program platform consists of a graphics based
programming structure. Therefore, in order to make a software program, a user can select several
graphical components from a LabVIEW library and attach using graphical wires. According to
National Instruments, LabVIEW is a platform that helps engineers scale from design to test and
from small to large systems. It offers unprecedented integration with existing legacy software,
IP, and hardware while capitalizing on the latest computing technologies [2]. Therefore, in
comparison, LabVIEW programming platform is more suitable for this measurement system.
34
7.3 Developing the automation software.
The voltage Calibrators are capable of producing lethal voltages up to
1100V. While testing the software, keep away from the output terminals.
Key aspects of the automation software
1. Safe operation of the system.
The Calibrators used on this measurement system are capable of producing
voltages up to 1100 volts. In order to ensure safe operation, output current of the
voltage Calibrators must be remotely limited to 3 mA using the software [7].
2. Controlling the voltage sources/calibrators
The software will control the outputs of the voltage Calibrators based on the value
of the resistor and the balance point calculations.
3. Controlling the current source/detector
The software will control the functionality of the current detector based on the
amount of settling time and the number of data points of the detector current
required.
4. Acquiring resistor data from a resistor database
When the user selects the resistors on the measurement software, the software will
search for the relevant data and use the search results during the measurement.
5. Applying voltage corrections
Corrections for the voltage supplies based on the calibration data will be applied
during the runtime.
6. Measurement algorithm and Data acquisition
The software will automate the measurement algorithm based on the input
parameters of the user. Furthermore, the software will acquire measurement data
and save them accordingly.
7. Performing calculations
Calculations such as the voltage at the balance point and the unknown resistor
value will be calculated using the measurement software.
35
The following flowchart shows a simplified procedure to be used on the measurement software.
Figure 7.1: Flow chart of the measurement procedure [15]
36
7.4 Overview of a sample automation software.
The following program was created at NIST using the LabVIEW programming platform.
(1) GPIB address for the V1 Calibrator
(7) Directory path for the measurement
results
(2) GPIB address for the V2 Calibrator
(8) Measurement voltage (V1 calibrator
voltage)
(3) GPIB address for the Current Detector
(9) Number of tests to be performed.
(4) Directory path for the resistor database
file
(10) Button to test the resistor in reverse
polarity
(5) Resistor to be tested (RX)
(11) Settle time for the test
(6) Reference resistor (RS)
(12) Test control Start/Stop.
1 2 3
4
7
5 6
8 9 10 11
13 14 15
16
19
18 17
Figure 7.2: Overview of a sample automation software.
12
37
(13) R1 (RX) Nominal value display
(17) V1 Voltage real-time graph
(14) R2 resistor information display
(18) V2 Voltage real-time graph
(15) Voltage ratio of the bridge display
(19) Detector Current real-time graph
(16) Real-time information of the ongoing test
38
8. Measurement Procedure
The following instructions will guide you to measure a resistor using the measurement system.
Additional Things You Will Need:
Resistors.
o One Standard Resistor with accurate calibration data.
o Unknown Resistor. (Resistor you would like to measure.)
Figure 8.1 illustrates a 10 Tera-Ohm Standard
Resistor used at NIST.
The calibrators are capable of producing lethal voltages up to 1100V.
Do not touch output terminals when the device operates. Use a caution
tape to isolate the system during operation.
Figure 8.1
39
Step 1: Attaching Resistors
Referring to the set-up procedure described in section 6 of this manual, connect the
positive terminal wire 1 from the first calibrator to one of the terminals of the standard
resistor as shown in Figure 8.2 and 8.3.
Similarly, connect the positive terminal wire 2 to one of the terminals of the unknown
resistor.
Step 2: Connecting the Current Detector
Connect the two open ends of the two resistors to
the Remote PreAmp of the current detector using
two wires.
Figure 8.4 illustrates the connection terminals of
the Remote PreAmp of the current detector.
Figure 8.2 Figure 8.3
POSITIVE TERMINAL WIRE 1
POSITIVE TERMINAL WIRE 2
Figure 8.4
40
Step 3: Storing in the Air Bath
Resistors must be kept in an Air Bath during the
measurement to minimize errors from temperature
fluctuations.
Place the two resistors and the Remote PreAmp of
the current detector as shown in figure 8.5.
Set the desired temperature on the Air Bath
controller.
Wait at least 2 hours before measurement.
Note: approximately 2 hours of settling time is required
before the measurement to ensure temperature
equilibrium inside the Air Bath.
Step 4: Setting up the software and making a measurement.
Before running the software, make sure that all the cables are attached.
Use a caution tape to warn or catch the attention of passersby of the area
Figure 8.5 TEMPERATURE CONTROLLER
41
Follow steps 1 through 12 to set up the software for the measurement (refer to figure 8.6).
(1) Select the GPIB address for the V1
Calibrator
(7) Set the directory path for the
measurement results
(2) Select the GPIB address for the V2
Calibrator
(8) Set the measurement voltage (V1
calibrator voltage)
(3) Select the GPIB address for the Current
Detector
(9) Set the number of tests to be performed.
(4) Set the directory path for the resistor
database file
(10) Select to test the resistor in reverse
polarity
(5) Select the resistor to be tested (RX)
(11) Set the settle time for the test
(6) Select the reference resistor (RS)
(12) Press ‘Start’ to begin the measurement
and press ‘Stop’ to abort.
1 2 3
4
7
5 6
8 9 10 11
13 14 15
16
19
18 17
Figure 8.6
12
42
Progress of the measurement can be monitored using 13 through 19 (refer to figure 8.6).
(13) R1 (RX) Nominal value display
(17) V1 Voltage real-time graph
(14) R2 resistor information display
(18) V2 Voltage real-time graph
(15) Voltage ratio of the bridge display
(19) Detector Current real-time graph
(16) Real-time information of the ongoing test
43
9. Testing the System for Accuracy
After the implementation, it is crucial to test the system for accuracy before using the
system for resistance measurements. Testing of the system for accuracy can be conducted within
the laboratory environment using two methods.
9.1 Using two standard resistors with adequate calibration data.
In order to verify the accuracy of the measurement system, two standard resistors with
adequate calibration data can be used as the known resistor and the unknown resistor of
the measurement system. For this test, a standard resistor with adequate calibration data,
such as, the measured resistance value, temperature coefficient and the resistance drift,
must be used to judge the accuracy of the measurement system. Procedure for the test is
analogous to the resistance measurement procedure described in section 8 of this manual.
After conducting the test using the resistors with adequate calibration data, produced
results can be compared with the previously measured data of the resistor to verify the
accuracy of the system.
9.2 Using a Hamon Transfer Standard.
If a resistor with adequate calibration data is unavailable, a Hamon transfer standard can
be used to verify the accuracy of the measurement system. A Hamon transfer standard is
a resistor that contains 10 or more resistors in series. Specifically, the Hamon transfer
standard provides the functionality to measure resistance in all 10 of these resistors in
series or in parallel [3].
Series
Connectors
Parallel
Connectors
Figure 9.1: Internal Diagram of a Hamon Transfer Standard [3]
44
By switching the relevant terminals as shown in figure 9.1, Hamon transfer standard can
be used either as 10 resistors in series or 10 resistors in parallel.
In order to test the system, a ratio measurement using the Hamon transfer standard can be
conducted. Specifically, two measurements must be made in all the resistors in series and
resistors in parallel. Afterwards, the two measurements can be compared with each other
to verify the accuracy of the system.
45
10. Troubleshoot Guide
To avoid electrical shock, personal injury or damage to equipment:
Do not perform any troubleshooting described below unless you are authorized
and qualified to do so.
Please follow electrical safety precautions carefully while performing these
procedures.
Problem Probable Cause Solution
No voltage output on the
Voltage Calibrator.
Calibrator in ‘standby’ mode
when the voltage is set
beyond 22V.
No input power to the
calibrator
Disable the automatic
‘standby’ of the voltage
calibrator using the
automation software. (refer to
the user’s manual of the
calibrator)
Ensure the calibrator is
plugged in to the main power
and verify the main power is
good.
Voltage calibrator shows
‘Calibration due’ message on
the display.
Fluke calibrator must be
calibrated periodically to
ensure an accurate voltage
output.
Refer to calibrator’s user
manual to perform the
calibration.
Voltage Calibrator won’t
power on.
Blown fuse of the voltage
calibrator.
No input power to the
calibrator.
Refer to the user’s manual of
the calibrator and check the
fuse. If the fuse is blown,
replace it with an appropriate
replacement fuse.
Ensure the calibrator is
plugged in to the main power
and verify the main power is
good.
46
Current detector shows no
output during run-time
Remote PreAmp of the
current detector is not
properly plugged in.
Verify the connections of the
Remote PreAmp of the
current detector and the
resistor.
Current detector won’t power
on.
Blown fuse of the current
detector.
No input power to the current
detector.
Refer to the user’s manual of
the current detector and check
the fuse. If the fuse is blown,
replace it with an appropriate
replacement fuse.
Ensure the current detector is
plugged in to the main power
and verify the main power is
good.
Signal of the current detector
on the real-time display of the
automation software appears
to be noisy.
Inadequate settling time for
the measurement.
Improper temperature control
of the Air Bath.
Input voltage of the calibrator
is too low.
Ensure that the settle time for
the measurement is adequate
based on the magnitude of the
resistor.
Check the Air Bath is
properly closed and the
temperature is set.
If the voltage of the calibrator
is too low, the current of the
detector could be noisy. For
higher resistance values, use
higher voltages.
Resulting resistor value is
significantly different than
the theoretical predictions.
Incorrect nominal value of
the unknown resistor.
Incorrect corrected value of
the standard resistor
(reference resistor).
Improper cable attachments.
Ensure that the nominal value
of the unknown resistor is
valid.
Ensure that the corrected
value of the standard resistor
is calculated to today’s date
and used during the
measurement.
Verify the cable attachments.
Refer to setup guide and the
measurement guide sections
of this manual.
47
11. Appendix
I. Glossary of Terms
Air Bath
High precise temperature controlled enclosure
filled with air that can be used to minimize
the effects of temperature fluctuations during
measurements.
Calibrator
Electronic device that can produce a precise
voltage supply that can be used to calibrate
other voltage source. During this manual, the
Calibrator is used as a voltage supply.
Current
The electric current. Current is a flow of
electric charge. SI unit of current is Ampere
and symbolized as ‘A’
Current Detector
Current detector, also known as the Amp-
Meter, is a device that detects electrical
current.
Current Leakage
Current leakage is the loss of useful electric
current in a circuit. For example, if the circuit
has a bad insulator, the current could flow in a
direction that is unintended by the user.
Dosimeters
Dosimeter is a device that can measure
radiation absorbed by an object.
GPIB Interface
GPIB (General Purpose Interface Bus), also
known as IEEE-488 Interface, is a
communication interface often used on
measurement instruments.
Hamon Transfer Standard
Hamon Transfer Standard is a resistor that
contains 10 or more resistors in series within
the resistor. Using different terminals, the
value of the resistor can be changed based on
user’s intention.
Kirchhoff’s Laws
Kirchhoff’s laws in current: at any node of a
circuit, sum of current in and out should be
equal to zero.
Kirchhoff’s voltage law: sum of the voltages
around a closed loop should be zero.
Nominal Value
The value stated on the face of the object. For
example, a resistor nominal value is the value
the resistor is built for. The actual value may
differ from the nominal value.
48
Oil Bath
High precise temperature controlled enclosure
filled with oil that can be used to minimize
the effects of temperature fluctuations during
measurements.
Parts Per Million (PPM)
Parts per Million (ppm) is a value that
represents the part of a whole number in units
of 1/1000000.
1 PPM = 0.0001%
Resistor
A resistor is a device that is designed to resist
the current flow on a circuit. The SI unit for
resistance is Ohms.
Signal Noise
Random fluctuations of a signal due to
external factors.
Standard Resistor
Standard resistor is a resistor that is used for
precise calibrations. Usually standard resistors
are very stable and contains adequate data
about its performance.
Voltage
Voltage, also known as electromotive force, is
the potential difference between two points of
a circuit.
II. Metric Prefixes
Table I I I
METRIC PREFIXES – SOURCE: U.S. DEPARTMENT OF COMMERCE, NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, THE
INTERNATIONAL SYSTEM OF UNITS (SI), NIST SPECIAL PUBLICATION 330, 1991 EDITION (WASHINGTON, DC, AUGUST 1991), P.10.
49
III. References
[1] K. Barnett, 'Electrician’s assistant sues over alleged electric shock received at New Orleans
Riverwalk project | Louisiana Record', Louisianarecord.com, 2015. [Online]. [Accessed: 04-
May- 2015].
[2] Ni.com, 'NI LabVIEW - Improving the Productivity of Engineers and Scientists - National
Instruments', 2015. [Online]. Available: http://www.ni.com/labview/. [Accessed: 04- May-
2015].
[3] D. Jarrett, 'Evaluation of guarded high-resistance Hamon transfer standards', IEEE Trans.
Instrum. Meas., vol. 48, no. 2, pp. 324-328, 1999.
[4] National Instruments, 'Shielded GPIB Cables', 2015. [Online]. Available:
http://sine.ni.com/nips/cds/view/p/lang/en/nid/1281. [Accessed: 04- May- 2015].
[5] MI Model 9300 Temperature Controlled Air Bath, 1st ed. Ontario, Canada: Measurement
International, 2015, p. 2.
[6] Msdn.microsoft.com, 'Visual Basic', 2015. [Online]. Available:
https://msdn.microsoft.com/en-us/library/2x7h1hfk.aspx. [Accessed: 04- May- 2015].
[7] D. Jarrett, 'ENGL393 - Expert Interview on high resistance measurement bridge', Panera
Bread, Gaithersburg, MD., 2015. [Summary attached].
[8] FLUKE 5730A Multifunction Calibrator Operators Manual, 1st ed. Fluke Coperation, 2013,
pp. 5-7, 13-25.
[9] P. Sutherland, Principles of Electrical Safety. Hoboken: Wiley, 2014.
[10] M. Markel, Technical communication. Boston, Mass.: Bedford/St Martins, 2012. Pp. 587.
[11] Jacob L. Heller, 'Electrical injury: MedlinePlus Medical Encyclopedia', Nlm.nih.gov, 2015.
[Online]. Available: http://www.nlm.nih.gov/medlineplus/ency/article/000053.htm. [Accessed:
05- May- 2015].
[12] E. Oberg, F. Jones, H. Horton and H. Ryffel, Machinery's handbook. New York: Industrial
Press, 1988.
50
[13] 6535 Automated High Resistance 6535 Automated High Resistance Measurement System,
1st ed. Ontario, Canada: Guildline Instruments Limited, 2015, pp. 1-5.
[14] D. Jarrett, 'Automated guarded bridge for calibration of multimegohm standard resistors
from 10 MΩ to 1 TΩ', IEEE Trans. Instrum. Meas., vol. 46, no. 2, pp. 325-328, 1997.
[15] L. Henderson, 'A new technique for the automatic measurement of high value resistors',
Journal of Physics E: Scientific Instruments, vol. 20, no. 5, pp. 492-495, 1987.
[16] Sub-femtoamp Remote SourceMeter® SMU Instrument Data Sheet, 1st ed. USA:
KEITHLEY Corporation, 2015, pp. 1-4.
51
IV. Interview Summary
We interviewed Mr. Dean Jarrett, project leader of the metrology of the ohm project at
NIST, to obtain an expert’s opinion and feedback on our manual. When we asked why he
thought the Dual source bridge is the best measurement system compared to a regular
Wheatstone bridge, he mentioned that it depends on the user’s intentions of the measurement
system. However, for the lowest uncertainty, the Wheatstone works better. Since there was no
human operator, he mentioned that the automated modified Wheatstone bridge is more feasible.
He also said that the optimal equipment selection depends on user’s goals. He states that if users
do not have the money to buy the best Calibrators on the market, then they can buy used ones for
cheaper. He also says to use polyethylene cables because it does not have charge retention issues.
Furthermore, for high resistances he also recommends to use air baths because oil baths have
leakage currents at high resistance ranges. He believes that the manual will be useful to
companies such as NASA and defense contractors because they do not specialize in one area like
NIST. The smallest uncertainty he has obtained is .08 ppm from a measurement bridge. Mr.
Jarrett believes that anyone who would be around this system should have basic electronics
training along with knowledge of where the high voltage areas are. The limitations of the system
include the settling time of RC constant and issues with drift current if you wait too long to take
a measurement.
A complete transcript of the interview is included below.
Interview Questions and Responses
1. You have worked with high resistance bridges for over 20 years. Based on your experience,
why do you think the Dual source bridge is the best measurement system compared to a regular
Wheatstone bridge?
It depends on the situation, what your goals are. Overall, the dual source bridge, if you
want to have the lowest uncertainty, it is the best system. If you don’t need much accuracy there
are other systems. But I’d say, because of the demands of our customers and the work we do, we
have gone to the lowest uncertainty as possible that’s why we used the Wheatstone bridge.
When I first started, we had a guarded Wheatstone bridge and the Tera-ohm meter
system, the limitations were that the guarded Wheatstone bridge was manually operated. So, we
had someone looking up the detector and turning knobs. It was very inefficient. It was subjective
depends on the patient of the operator, we used to have two operators one operator in the
morning and operator at night. Automating will prevent leakage problems. The old bridges were
only able to use up to 10 Giga ohms. Because we replaced two resistors with the programmable
52
DC calibrators so we don’t have leakage problems and also its automated and it can go to higher
voltages. Plus it’s less labor. Before we used battery supply, but then we had to check the
voltages on the batteries.
2. Do you think an automated high resistance bridge system is more efficient and effective
compared to other systems in the industry? If so, why?
Certainly anything automated is more efficient than anything that is manually operated.
The main thing that we developed in our measurement bridges is that we have automated
guarded switching so we can attach a number of resistors and then automated software can
measure a lot of resistors at the same time. The measurement can run over night without an
operator. It eliminated the labor and the human factors to minimize systematic errors. The
timings are consistent and it’s not based on someone's judgements. There are also some other
systems in the industry, they don’t operate in the same wide range, and they don’t have the same
uncertainty range. There is another system that is widely used that is the binary voltage divider
bridge. Where a binary voltage bridge is used on two of the arms and that system is automated,
but it only works up to a 100 mega ohms. It has comparable uncertainties.
3. What are your recommendations for equipment to be used in an automated resistance
bridge?
There are some other Calibrators that are available and that don’t cost a lot. You could
build a bridge that wouldn’t be as accurate, but you can still make measurements. The other
option that some people have done is that they have bought used calibrators on the used
equipment market. The downside is that they are not supported by the manufacturer. We do have
two of those bridges in our oldest bridge. Those are probably 30 something years old, so at one
point there won't be repairable. That’s why we are building the new bridge to make the bridge
very robust and try to address some of the shortcomings, especially from 1 Tera-Ohm to 1 Kilo-
Ohm
4. Teflon and Polyvinyl cable are widely used in the measurement industry. In order to
minimize noise during a measurement, what cable would you recommend for the measurement
system? And why?
So, we use a lot of Teflon cable, we use that because Teflon is a high insulator. However,
we have also found that Teflon at the high resistance range Teflon retain charge so we looked at
some other alternatives. There is some cable that has. I forgot the name the insulator. Its poly
something. But not polyvinyl. That’s a low noise cable that’s the best one for the higher
resistance levels. It doesn’t have the charge retention issues and it’s a pretty good insulator.
53
5. Standard resistors are kept in either oil or air baths. For high resistance standards, what is
the ideal resistance storage method? And why?
At the resistance lab we use oil baths for basically anything from one mega ohm to
anything under less than an ohm. We use the oil because the oil is a pretty good insulator
because it’s clean and it’s not contaminated, it will have an insulation of 1014 ohms. Some of the
resistors have temperature coefficients that is they are susceptible to the changes of the
temperature. We can control oil better than air. The oil that we have in big baths in the labs, we
can control them with the covers on up to a one mili degree Celsius. We can't use oil after 10
mega ohm range because we get to the point where the resistance of the oil will create a leakage
path and create a leakage problem therefore we don’t use it for the high resistance range. We
could possibly use at 10 mega ohm then we would have to have a separate area.
6. What industries/work environments do you think, will benefit from this manual? And what
are some applications of high Resistance standards?
Often we have a lot of customers that will send us equipment and standard resistors for
calibration. Some of them are very capable of and outstanding experts in the field and some are
not. We have some customers that calibrate their resistors on the bench and wonder why the
value change by 50 ppm. I think a manual like this would be beneficial to people in the industry.
We are specialized in our labs because we are working with resistances. But some other labs like
in military labs and NASA and a lot of the defense contractors can't go into one discipline like
we can. So, they will be doing resistance, voltage, impedance, pressure, and temperature
measurements so a manual like this would be helpful to give them some things to consider. For
example the effect of temperature, the type of cable that they are using or the guarding. So I
think they will be useful for them.
Lot of the applications have to do with people who need to measure low currents. Some
of the applications that I have encountered are customers that calibrate standards for other labs.
In the department of energy and the nuclear power industry, they have equipment for measuring
background radiation dosimeters and the output is a really small current. In the pico-amp or in
the femto-amp range. To calibrate those dosimeters, they use a standard resistors to compare the
output of the dosimeters and calibrate them. Also, people making connectors use high resistance
standards because, that they want to check the purity of the material they are using. Another
54
application is that, we did a special test for Boeing where they wanted a resistor calibrated at 10
Tera-ohms and what they were using that for was to calibrate their fuel test sets. So that they can
grade the purity of the fuel going into their jet engines. A lot of the applications are dealing with
the purity. Also I have talked to someone who wanted to have some resistors Calibrators in the
10 Giga-ohm and 100 Giga-ohm and they were using the measurements to grade the purity of
raw textile material and to determine what it can be used for in the textile industry. One of the
other applications is for one of our most common customers, which is the Sandia national labs
and one of the things that they do at Sandia is that they support the nuclear weapons program.
They wanted to measure the resistivity of a ceramic coating on a neutron tube part of a nuclear
system.
7. You have implemented several automated high resistance measurement systems at NIST.
What is the highest level of accuracy that you were able to achieve from your systems?
One of the systems I have worked on, which is an automated double Kelvin bridge which
can measure up to 0.08 parts per million and those are using special resistors that are designed in
the 1970’s by a couple of manufacturers and they are stable and has low temperature
coefficients. Those are probably the lowest uncertainties that I have worked on. Also in high
resistance, we were able to achieve few PPM with the dual source bridge.
8. What safety measures would you recommend to improve the safe operation of the system?
One of the issues that we have with the high resistance is that sometimes we have to go
high voltages like 1000 volts. One of the things that we implemented in our dual sources bridges
is that, we limit the current. We build that into the code so that the Calibrators will limit the
current to 3 mA. Also we rope off the area. Another thing that I recommend is to have some kind
of a visual indicator when the system is running with a bright light. That will be useful to alert
people and also it will be good if the system gets stuck at one point. If the person is not
intimately familiar with the system a warning light will be very useful to know that the system
isn’t functioning properly. Another thing would be to have a display running off of a computer
saying the actual status of the measurement.
9. What type of training do you think people should receive, in order to operate this system?
Why?
The first thing that we do require is that basic electrical safety training and to go over our
standard operating procedures that we have in place and to be familiar with the instruments that
make up the bridge. For example, how the calibrators work, how the detector work, and how the
55
cables are connected. Also people should be aware of where the high voltage is at and where the
ground is when the system is operating.
10. What are the limitations of a bridge measurement system of this nature?
One of the limitation is the settling time of the RC time constant. Also if you wait too long what
happens is that it will take a long time to take a single measurement and then will have errors in
other components from drift.